Below-shot-noise capacity in phase estimation using nonlinear interferometers

Cristofero Oglialoro Technische Universität Darmstadt, Fachbereich Physik, Institut für Angewandte Physik, Schloßgartenstr. 7, 64289 Darmstadt, Germany    Gerard J. Machado Imperial College London, Department of Physics, 1 Prince Consort Road, SW7 2AZ London, United Kingdom University of Oxford, Department of Physics, Parks Road, OX1 3PU Oxford, United Kingdom    Felix Farsch Technische Universität Darmstadt, Fachbereich Physik, Institut für Angewandte Physik, Schloßgartenstr. 7, 64289 Darmstadt, Germany    Daniel F. Urrego ICFO – Institut de Ciències Fotòniques, The Barcelona Institute of Science and Technology, 08860 Castelldefels, Spain    Alejandra A. Padilla ICFO – Institut de Ciències Fotòniques, The Barcelona Institute of Science and Technology, 08860 Castelldefels, Spain    Raj B. Patel Imperial College London, Department of Physics, 1 Prince Consort Road, SW7 2AZ London, United Kingdom    Ian A. Walmsley Imperial College London, Department of Physics, 1 Prince Consort Road, SW7 2AZ London, United Kingdom University of Oxford, Department of Physics, Parks Road, OX1 3PU Oxford, United Kingdom    Markus Gräfe Technische Universität Darmstadt, Fachbereich Physik, Institut für Angewandte Physik, Otto-Berndt-Straße 3, 64287 Darmstadt, Germany Fraunhofer Institute for Applied Optics and Precision Engineering IOF, Albert-Einstein-Str. 7, 07745 Jena, Germany    Juan P. Torres [email protected] ICFO – Institut de Ciències Fotòniques, The Barcelona Institute of Science and Technology, 08860 Castelldefels, Spain Department of Signal Theory and Communications, Universitat Politecnica de Catalunya, 08034 Barcelona, Spain    Enno Giese Technische Universität Darmstadt, Fachbereich Physik, Institut für Angewandte Physik, Schloßgartenstr. 7, 64289 Darmstadt, Germany [email protected]
Abstract

Over the past decade, several schemes for imaging and sensing based on nonlinear interferometers have been proposed and demonstrated experimentally. These interferometers exhibit two main advantages. First, they enable probing a sample at a chosen wavelength while detecting light at a different wavelength with high efficiency (bicolor quantum imaging and sensing with undetected light). Second, they can show quantum-enhanced sensitivities below the shot-noise limit, potentially reaching Heisenberg-limited precision in parameter estimation. Here, we compare three quantum-imaging configurations using only easily accessible intensity-based measurements for phase estimation: a Yurke‑type SU(1,1) interferometer, a Mandel‑type induced‑coherence interferometer, and a hybrid scheme that continuously interpolates between them. While an ideal Yurke interferometer can exhibit Heisenberg scaling, this advantage is known to be fragile under realistic detection constraints and in the presence of loss. We demonstrate that differential intensity detection in the Mandel interferometer provides the highest and most robust phase sensitivity among the considered schemes, reaching but not surpassing the shot-noise limit, even in the presence of loss. Intensity measurements in a Yurke-type configuration can achieve genuine sub-shot-noise sensitivity under balanced losses and moderate gain; however, their performance degrades in realistic high-gain regimes. Consequently, in this regime, the Mandel configuration with differential detection outperforms the Yurke-type setup and constitutes the most robust approach for phase estimation.

1 Introduction

We consider here a class of optical interferometers that can be termed as nonlinear interferometers [1] and which are based on original and fundamental ideas developed some decades ago [2, 3]. In this type of interferometers, at the input and output ports, instead of the beam splitters used in standard interferometers, one makes use of optical parametric amplifiers. These interferometers have gained increasing attention [4, 5] as a useful tool for imaging, parameter estimation, spectroscopy and optical coherence tomography, to name a few applications. They offer both practical advantages [6, 7, 8] and a way to achieve quantum-enhanced sensitivity [9, 10, 11], i. e., going below the shot-noise limit in parameter estimation.

We consider three different types of nonlinear interferometers, namely the Yurke, Mandel, and hybrid schemes, shown in Fig. 1. In all three cases, pumping the nonlinear medium AA with monochromatic light, alongside vacuum input, induces nondegenerate parametric down-conversion. The medium generates nn photons in each of two modes, called signal (blue) and idler (red) fields. The number of generated photons depends on the parametric gain, that is determined mainly by the intensity of the pump field and the nonlinear susceptibility of the medium [12].

Formally, the light generation process during parametric down-conversion can be described by a Bogoliubov transformation, whose coupling matrix is an element of the SU(1,1) group [13]. Due to the nondegenerate nature of the process, the signal and idler fields may be at different wavelengths, enabling bicolor sensing with undetected photons. For example, the signal mode might be in the visible range of the spectrum and the idler mode in the mid-infrared. The object is probed by the idler light and has a complex transmission coefficient ti=Tiexp(iφi)t_{\text{i}}=\sqrt{T_{\text{i}}}\exp(\text{i}\varphi_{\text{i}}), with transmittance 0Ti10\leq T_{\text{i}}\leq 1 and a phase φs\varphi_{\text{s}}. In the transmittance we may also include other loss channels of the idler arm. Loss in the signal arm is included through a transmission coefficient ts=Tsexp(iφs)t_{\text{s}}=\sqrt{T_{\text{s}}}\exp(\text{i}\varphi_{\text{s}}), with transmittance 0Ts10\leq T_{\text{s}}\leq 1, the phase shift experienced by the idler is φi\varphi_{\text{i}}. In general, the signal/idler transmittances may be different.

All three interferometers begin with a two‑mode‑squeezed vacuum interaction generated in medium AA, which produces correlated signal–idler beams, each with nn photons. The idler photons probe the sample, acquiring a phase shift to be determined, while the idler beam itself is never measured. The object’s information is retrieved solely from intensity (or photon‑number) measurements on the signal mode. Because of the identical initial squeezing process, the quantum Fisher information for phase estimation is the same for all of these configurations [10].

Although this feature is common to all three configurations, the transfer of phase information is different. In the Yurke [2] setup shown in Fig. 1 (a) both signal and idler modes are seeded into a second nonlinear medium BB, where the output signal mode intensity (NsN_{\text{s}}) is detected. The two-mode squeezed state generated in medium AA is either further squeezed/amplified by medium BB or antisqueezed/deamplified [14]. Because squeezing is a phase-sensitive process that depends on the phases of the pump, signal, and idler beams, these phases determine the detected intensity.

The sensitivity of phase estimation achievable in a Yurke configuration depends on the losses in the signal and idler arms of the interferometer being equal (balanced) or not (unbalanced). This effect is a manifestation of the strong dependence of the degree of squeezing on the presence of loss. Indeed, for a particular value of the gain in the low-gain regime, unbalanced loss can even provide a better sensitivity compared to a balanced situation [15]. Complementing these results, we demonstrate below that balanced and unbalanced situations in a Yurke configuration lead to radically different behaviors of the sensitivity in the high-gain regime.

Refer to caption
Figure 1: Sketches of different nonlinear interferometers: In the Yurke setup (a), a nonlinear medium AA with vacuum input generates signal and idler fields j=s,ij=\text{s},\text{i} (blue and red). They may be subject to loss, encoded in a non-unit transmittance TjT_{j}, before they seed a second nonlinear medium BB, whose signal output NsN_{\text{s}} is detected. In the Mandel setup (b), only the idler seeds medium BB, while the signal modes of both media are interfered on a 50:50 beam splitter, whose outputs NsN_{\text{s}} and NsN_{\text{s}}^{\prime} are detected. In a hybrid configuration (c), only a fraction 1ϱ1-\varrho of the signal mode is seeded into medium BB, while the remaining part is interfered with the signal output of medium BB on a beam splitter, whose transmittance is 1ϱ/21-\varrho/2. The variable parameter ϱ\varrho tunes the setup from a Yurke (ϱ=0\varrho=0) to a Mandel configuration (ϱ=1\varrho=1).

It is well-known that squeezed states can enable an interferometer to achieve a sensitivity below shot noise, which in the ideal case is such that the phase uncertainty has a Heisenberg 1/n1/n-scaling. Moreover, seeding both idler and signal into medium BB gives rise to a compact and symmetric form, which is robust for many applications like spectroscopy [16, 17], microscopy [18, 19], and holography [20, 21, 22]. However, under loss, the squeezed quantum state deteriorates quickly [23, 24], losing its sub-shot-noise capability. Nevertheless, this degradation can be partially compensated by using different gains [9, 25, 26] in media AA and BB.

In contrast, in the Mandel [27] setup shown in Fig. 1 (b), only the idler mode seeds medium BB. The signal modes emitted by media AA and BB interfere on a 50:50 beam splitter. The intensities measured at the two exits of the beam splitter, NsN_{\text{s}} and NsN_{\text{s}}^{\prime}, depend on the phases of the pump, signal, and idler beams inside the interferometer. However, because the nonlinear process in medium BB is partially seeded, the intensities of the two signal modes can differ significantly [28, 29, 30]. In fact, such an interferometer does not display a pure SU(1,1) symmetry, because there are three input and output modes.

Although the quantum Fisher information, a measure of the best sensitivity that can be reached for any measurement, coincides for both the Yurke and Mandel configurations and implies a Heisenberg scaling [10], in the Mandel setup such a sensitivity cannot be obtained from simple intensity or photon-number measurements. In fact, to extract a phase with such a high precision, one needs to rely on much more involved experimental methods like homodyning, similar to truncated SU(1,1) interferometers [31, 32, 33, 34].

We therefore refrain from using the quantum Fisher information as a metric for phase sensitivity and instead directly obtain the phase uncertainty via Gaussian uncertainty analysis of the intensity measurement. For high gain, which is the most interesting regime for quantum metrology, signal modes of different intensity interfere, such that the contrast of NsN_{\text{s}} decreases significantly [28, 29, 30] with adverse consequences for the sensitivity. Here, we demonstrate below that the difference of both exits [35], namely N=NsNsN_{-}=N_{\text{s}}-N_{\text{s}}^{\prime}, is robust against this deleterious effect and in fact is limited by shot noise.

A comparison of the Yurke and Mandel configurations [36], based on the quantum Fisher information [10] and intensity as a phase estimator [35], suggests different implications for the attainable phase sensitivity. These previous studies have explored both configurations with additional coherent seeds in the signal and/or idler modes. In our comparative study, however, we focus on the vacuum-input scenario to isolate the fundamental performance limits and facilitate a clear discussion of the underlying physics.

We also discuss a hybrid setup that has been already implemented in the low gain regime [37]. This scheme is sketched in Fig. 1 (c). A beam splitter with transmittance 1ϱ1-\varrho couples a fraction of the signal beam out of the Yurke configuration. This fraction interferes with the signal mode emitted by medium BB on a flexible beam splitter with transmittance 1ϱ/21-\varrho/2. Such a configuration allows for tuning from the Yurke case (ϱ=0\varrho=0) to the Mandel case (ϱ=1\varrho=1) and studying the transition between both configurations.

Here, we analyze the phase estimation sensitivity of the different configurations by approximating the signal and idler modes as single-mode beams. Despite its simplicity, this approach provides useful and qualitatively accurate results. Heuristically, one can think of NsN_{\text{s}} as the number of signal photons generated per unit bandwidth (photons/s/Hz\textrm{photons}/\textrm{s}/\textrm{Hz}), so that NsN_{\text{s}} can be related to the flux rate Rs=Ns×R_{\text{s}}=N_{\text{s}}\times\mathcal{B} of signal photons observed in an experiment, where \mathcal{B} is the spectral bandwidth of detection. More rigorously, it can be demonstrated that the validity of this result assumes a single transverse mode and a detection bandwidth \mathcal{B} much smaller than the bandwidth of parametric down-conversion [38] in media AA and BB. For the variance, one can derive a similar expression, even when one includes the spatial/frequency multimode character of signal/idler modes [39].

2 Phase uncertainty: the lossless case

We begin our discussion by studying these configurations without any loss. The mathematical expressions of the interference patterns and variances are derived in Appendix A. We find that detecting only one exit of the (Mandel or Yurke) interferometer, the signal photon number NsN_{\text{s}} depends on the phase ϕ\phi to be determined and follows thermal statistics as expected for squeezed light. Therefore, it has a variance Var(Ns)=Ns[1+Ns]\text{Var}\left(N_{\text{s}}\right)=N_{\text{s}}\left[1+N_{\text{s}}\right]. These statistics imply through Gaussian uncertainty propagation that the phase uncertainty for the Yurke (Y) configuration and the uncertainty observed for detecting a single exit of the Mandel configuration (SM) takes the form

σY/SM2=Ns[1+Ns](Ns/ϕ)2.\displaystyle\sigma^{2}_{\text{Y/SM}}=\frac{N_{\text{s}}\left[1+N_{\text{s}}\right]}{\left(\partial N_{\text{s}}/\partial\phi\right)^{2}}. (1)

In the Yurke configuration, the interference pattern takes the form Ns=2n(1+n)(1+cosϕ)N_{\text{s}}=2n(1+n)(1+\cos\phi), where nn corresponds to photons generated in nonlinear medium AA and depends on the parametric gain. Furthermore we find that in the case of balanced gain and no losses, when all light generated in AA is converted back to the pump in BB, the interference pattern shows perfect contrast.

Because the statistics are thermal, the variance vanishes at destructive interference, namely for ϕ=π\phi=\pi. Therefore, it is possible to work at the dark fringe with minimal intensity fluctuations. Taking this limit, we find [2]

σY2|π=14n(n+1)hg14n2,\displaystyle\left.\sigma^{2}_{\text{Y}}\right|_{\pi}=\frac{1}{4n(n+1)}\overset{\text{hg}}{\longrightarrow}\frac{1}{4n^{2}}, (2)

where we have expanded the result in the last step in orders of 1/n1/n and observe Heisenberg scaling in the high-gain (hg) regime 1n1\ll n. One can show that this expression saturates the quantum Cramér-Rao bound and therefore constitutes an optimal measurement, which implies that the mean photon number obtained from an intensity measurement is the optimal estimator for the interferometer phase [10].

In contrast, the interference pattern Ns=n(n+2+n+1cosϕ)/2N_{\text{s}}=n(n+2+\sqrt{n+1}\cos\phi)/2 observed in one of the two exits of the Mandel configuration does not display perfect destructive interference for arbitrary nn, so that the thermal variance does never vanish. Hence, there is no benefit from working at the dark fringe and the working point of highest sensitivity is obtained for a different phase. Details on the optimization can be found in Sec. 3. After this procedure, we find that the minimal phase uncertainty

σSM2|ϕmin=14nn+2n+1+n2+4(n+1)2+n48(n+1),\displaystyle\left.\sigma_{\text{SM}}^{2}\right|_{\phi_{\text{min}}}=\frac{1}{4n}\frac{n+2}{n+1}+\frac{n^{2}+\sqrt{4(n+1)^{2}+n^{4}}}{8(n+1)}, (3)

not only scales worse than shot noise, but even deteriorates in the high-gain regime, as σSM2|ϕminhgn/4\sigma_{\text{SM}}^{2}|_{\phi_{\text{min}}}\!\!\overset{\text{hg}}{\longrightarrow}n/4. This behavior is a consequence of the fact that photon-number measurements do not necessarily constitute optimal measurements.

Refer to caption
Figure 2: Phase uncertainty of the hybrid setup nσH2n\,\sigma^{2}_{\text{H}} for n=10n=10, as a function of the mixing parameter ϱ\varrho and the phase ϕ\phi. The phases minimizing the sensitivity for discrete values of ϱ\varrho are represented by white dots. The optimal working point for the Yurke setup (ϱ=0\varrho=0) is at ϕmin=π\phi_{\text{min}}=\pi, while the sensitivity is minimized at two phases for ϱ>0\varrho>0. For the Mandel setup (ϱ=1\varrho=1), these phases are ϕmin=π/2, 3π/2\phi_{\text{min}}=\pi/2,\,3\pi/2. The margin shows the minimal phase uncertainties, ranging from the optimal uncertainty of the Yurke setup to the shot-noise limit exhibited by the Mandel configuration.

One way to enhance the sensitivity is to maximize the contrast that is below unity because signal beams of different intensities interfere at the beam splitter [29]. However, in the Mandel configuration there is also the option to detect the intensity difference of both exits, namely N=NsNsN_{-}=N_{\text{s}}-N_{\text{s}}^{\prime}, which suffers no problem [35] from interfering different intensities since N=2n1+ncosϕN_{-}=2n\sqrt{1+n}\cos\phi. Here, the contrast can be viewed as perfect. Without loss and following Appendix A, the variance can be written as Var(N)=N++N2\operatorname{Var}(N_{-})=N_{+}+N_{-}^{2}, where also the sum N+=Ns+NsN_{+}=N_{\text{s}}+N_{\text{s}}^{\prime} of both exits contributes. Consequently, we find the phase uncertainty obtained from such a measurement

σ2=N++N2(N/ϕ)2.\displaystyle\sigma^{2}_{-}=\frac{N_{+}+N_{-}^{2}}{(\partial N_{-}/\partial\phi)^{2}}. (4)

Because the sum of the intensities N+=n(n+2)N_{+}=n(n+2) is independent of the interferometer phase and scales with n2n^{2}, the variance will never vanish. Therefore, the only option to minimize the phase uncertainty is to work at mid fringe ϕ=π/2\phi=\pi/2, where N=0N_{-}=0, while N/ϕ\partial N_{-}/\partial\phi is maximal. In fact, an optimization procedure finds exactly this phase and we obtain the phase uncertainty

σ2|π/2=14nn+2n+1hg14n,\displaystyle\left.\sigma^{2}_{-}\right|_{\pi/2}=\frac{1}{4n}\frac{n+2}{n+1}\overset{\text{hg}}{\longrightarrow}\frac{1}{4n}, (5)

which does not suffer from the severe sensitivity deterioration like the single exit detection scheme, but is still shot-noise limited.

For the lossless Yurke configuration it is optimal to work at destructive interference, whereas for the Mandel setup the working point lies at mid fringe. In the former case, a two-mode squeezed state is generated and subsequently antisqueezed, while the latter displays conventional interference, which is characterized by the shot-noise limit. This result is obtained by estimating the phase from the intensity difference [35], which is beneficial due to the unbalanced intensity of the two signal intensities, while still assuming equal gain.

Even though both configurations rely on squeezing and allow for bicolor imaging, the resulting sensitivities are quite different. To gain further insights, we study a hybrid setup that allows for a continuous transition from the Yurke to the Mandel case. For that, a part of the signal arm is coupled out of the Yurke interferometer and interferes on a beam splitter with the signal output of the second nonlinear medium. This hybrid configuration has been effectively implemented in the low-gain regime [37]. Because it is necessary to achieve optimal sensitivity in the Mandel configuration, we resort to a detection of the differential intensity NN_{-} between both exits.

Following Fig. 1 (c), the reflectivities of the two beam splitters can be tuned by a parameter ϱ\varrho in such a way that ϱ=0\varrho=0 reproduces the Yurke configuration, while ϱ=1\varrho=1 corresponds to the Mandel setup. We choose the two phases of both setups to be the same and denote them by ϕ\phi. The interference pattern as well as the variances and relevant transformations are given in Appendix A.

The resulting phase uncertainty is plotted as a function of the phase and of the tuning parameter ϱ\varrho in Fig. 2 for a fixed photon number nn. Indeed, we observe a smooth transition from a single phase with minimal uncertainty in the Yurke case at the dark fringe ϕ=π\phi=\pi, towards optimal phases at two mid-fringe positions ϕ=π/2,3π/2\phi=\pi/2,3\pi/2 in the Mandel case. The optimal phases are denoted by white dots in the density plot. In fact, for ϱ0\varrho\neq 0 the phase uncertainty diverges at dark fringe ϕ=π\phi=\pi, which is a known feature of losses [23, 40], where no perfect destructive interference is possible.

This feature demonstrates that the Mandel setup can be interpreted as the Yurke configuration including loss, which is recycled by a beam splitter before detection. The margin of the plot shows for each parameter setting of ϱ\varrho the optimal phase and the associated phase uncertainty. Here, we see that the phase uncertainty moves from the Heisenberg scaling to shot noise, but also demonstrates that gradual transitions between both regimes are possible.

3 Phase sensitivity in a Yurke setup in the presence of loss

In the hybrid setup, we observed that recycling losses from the signal beam in the Yurke configuration, embedded within a Mandel scheme, causes a gradual transition from Heisenberg-limited to shot-noise scaling. This observation underlines the role that optical losses play in determining the sensitivities of nonlinear interferometers. Under realistic conditions, photon losses can occur in both arms of the interferometer, and the magnitude of loss may differ between the two paths. Such a scenario is not expected to improve the sensitivity of the Mandel configuration. However, it can reduce the capacity of the Yurke configuration to maintain Heisenberg-limited sensitivity, as has been demonstrated [23] for the case of balanced losses in both arms. In this Section we analyze the phase uncertainty of the Yurke configuration in the presence of loss.

Following Appendix A, the detected number of signal photons leaving the interferometer can be written as

Ns=a+bcosϕ,N_{\text{s}}=a+b\cos\phi, (6)

where

a=n(1+Ts)+n2(Ts+Ti)b=2n(n+1)TsTi.\displaystyle\begin{split}a=&n\left(1+T_{\text{s}}\right)+n^{2}\left(T_{\text{s}}+T_{\text{i}}\right)\\ b=&2n\left(n+1\right)\sqrt{T_{s}T_{i}}.\end{split} (7)

Here, TsT_{\text{s}} and TiT_{\text{i}} are the overall losses in the signal/idler paths of the interferometer, ϕ\phi is the phase to be estimated, and nn is the number of photons probing the sample located in the idler arm of the interferometer. The contrast of the interference pattern measured by varying the phase ϕ\phi is C=b/aC=b/a.

The uncertainty of phase estimation, using the propagation of errors equation, reads

σY2=a(1+a)+b(1+2a)cosϕ+b2cos2ϕb2sin2ϕ.\sigma_{\text{Y}}^{2}=\frac{a(1+a)+b(1+2a)\cos\phi+b^{2}\cos^{2}\phi}{b^{2}\sin^{2}\phi}. (8)

The optimum phase ϕmin\phi_{\text{min}} that provides the minimum of the variance [40] given by Eq. (8) can be calculated to be

ϕmin=cos1{a(1+a)+b2b(1+2a)+[a(1+a)+b2b(1+2a)]21}.\displaystyle\begin{split}\phi_{\text{min}}=\operatorname{cos}^{-1}&\left\{-\frac{a(1+a)+b^{2}}{b(1+2a)}\right.\\ &\left.+\sqrt{\left[\frac{a(1+a)+b^{2}}{b(1+2a)}\right]^{2}-1}\right\}.\end{split} (9)

Substituting the angle given in Eq. (9) into the expression of the variance given by Eq. (8), we obtain that the minimum variance, for a given value of the parametric gain, is

σY2|min=a(1+a)+b22b2+(a2b2)[(1+a)2b2]2b2.\displaystyle\begin{split}\left.\sigma_{\text{Y}}^{2}\right|_{\text{min}}=&\frac{a(1+a)+b^{2}}{2b^{2}}\\ &+\frac{\sqrt{\left(a^{2}-b^{2}\right)\left[\left(1+a\right)^{2}-b^{2}\right]}}{2b^{2}}.\end{split} (10)

In order to determine if we can observe a quantum-enhanced sensitivity in phase estimation even under the presence of loss, it is helpful to make use of the concept of the normalized classical Fisher information (Fc/nF_{\text{c}}/n). The quantum nature of output signal photons in the Yurke setup corresponds to thermal light, so the classical Fisher information associated to the phase estimation protocol can be written as (see Appendix B)

Fc=1/σY2,F_{\text{c}}=1/\sigma_{\text{Y}}^{2}, (11)

where σY2\sigma_{\text{Y}}^{2} is given by Eq. (8). This connection implies that the mean photon number, i. e., the detected intensity, is an optimal phase estimator in the Yurke-type setup. Such a behaviour is expected since the probability distribution that describes signal photons (thermal light) belongs to the exponential family of probability distributions.

We consider as benchmark case for comparison the case of phase estimation using coherent light. For a coherent state, the Fisher information FccohF_{\text{c}}^{\text{coh}} scales linearly with nn. A feature of phase estimation with a coherent light is thus that the normalized Fisher information Fccoh/nF_{\text{c}}^{\text{coh}}/n remains constant when increasing the number of photons nn. A quantum sensing protocol scheme for which Fc/nF_{\text{c}}/n increases with nn is said to have quantum-enhanced sensitivity, because it exceeds the sensitivity improvement achievable by a coherent state when increasing nn. On the contrary, if the protocol shows that Fc/nF_{\text{c}}/n decreases with nn, it would do worse than shot noise.

Refer to caption
Figure 3: Normalized classical Fisher information Fc/n=1/(nσY2|min)F_{\text{c}}/n=1/(n\sigma_{\text{Y}}^{2}|_{\text{min}}) at the optimal phase setting ϕmin\phi_{\text{min}} as a function of nn and TiT_{\text{i}}, assuming Ts=0.8T_{\text{s}}=0.8. An increase of the normalized Fisher information with the number of probing photons implies better than shot-noise scaling, while a constant behavior corresponds to a shot-noise limited measurement. A decrease of the normalized Fisher information results in a deterioration of the phase sensitivity with the photon number, therefore worse than shot-noise scaling. We observe a resonance for Ti=TsT_{\text{i}}=T_{\text{s}}. The cuts along the dashed lines at Ti=0.7T_{\text{i}}=0.7 (black) and Ti=0.8T_{\text{i}}=0.8 (blue) are displayed in Fig. 4.

Using Eq. (10), we plot in Fig. 3 the normalized classical Fisher information Fc(ϕmin)/n=1/(nσY2|min)F_{\text{c}}(\phi_{\text{min}})/n=1/(n\sigma^{2}_{\text{Y}}|_{\text{min}}) at the optimal measurement point as a function of nn and TiT_{\text{i}}, assuming Ts=0.8T_{\text{s}}=0.8. In this plot, we observe a peak close to Ti=TsT_{\text{i}}=T_{\text{s}}, but even at this point the normalized classical Fisher information does not grow indefinitely, but saturates, indicating a shot-noise behavior.

To study this behavior in more detail, we plot in Fig. 4 the dependence of the normalized classical Fisher information for two cases that are representative of the general trend we have found exploring the whole parameter space. Only for small photon numbers, we observe a sub-shot-noise scaling for both cases, identified from an increasing normalized classical Fisher information. When losses inside the interferometer are equal (Ts=TiTT_{\text{s}}=T_{\text{i}}\equiv T), the normalized classical Fisher information tends to a constant value for large values of nn, as shown by the thick solid blue line, approaching shot-noise scaling. When losses inside the interferometer are unbalanced (TsTiT_{\text{s}}\neq T_{\text{i}}), the normalized classical Fisher information decreases for large values of nn as shown by the thick solid black line. This behavior indicates a scaling behavior worse than shot noise.

For a particular value of the phase (dotted, dashed, and solid thin lines) that does not coincide with ϕmin\phi_{\text{min}}, we see in Fig. 4 that there are regions of the parametric gain when the normalized classical Fisher information grows with increasing values of nn, which corresponds to quantum-enhanced phase estimation sensitivity. For large values of nn, the variance decreases. Therefore, for a chosen phase ϕ\phi, eventually phase estimation for large enough values of nn is worse than the shot-noise limit, since the normalized classical Fisher information decreases with increasing values of nn. Hence, only for the optimal phase ϕmin\phi_{\text{min}} one can approach shot noise at best.

In order to get further physical insight into the regime of high parametric gain, and thus large values of nn, we expand Eq. (10) as a function of 1/n1/n and we obtain

σY2|minhg(TsTi)24TsTi+(Ts+Ti)(1Ts)2TsTi1n+1(3Ti+Ts)(1Ti)4TsTi1n2.\displaystyle\begin{split}\left.\sigma^{2}_{\text{Y}}\right|_{\text{min}}\overset{\text{hg}}{\longrightarrow}&\frac{(T_{\text{s}}-T_{\text{i}})^{2}}{4T_{\text{s}}T_{\text{i}}}+\frac{(T_{\text{s}}+T_{\text{i}})(1-T_{\text{s}})}{2T_{\text{s}}T_{\text{i}}}\frac{1}{n}\\ &+\frac{1-(3T_{\text{i}}+T_{\text{s}})(1-T_{\text{i}})}{4T_{\text{s}}T_{\text{i}}}\frac{1}{n^{2}}.\end{split} (12)

Equations (8), (10), and (12) are the main results of this section. Equation (8) is the general expression that determines the uncertainty of phase estimation for any phase ϕ\phi and any parametric gain. Equation (10) is the uncertainty of phase estimation when ϕ=ϕmin\phi=\phi_{\text{min}}, and thus gives the minimum variance for a given value of the parametric gain. Equation (12) is the minimum variance for the case of high parametric gain (n1n\gg 1), and it highlights the dependence of the variance of phase estimation on the value of nn.

The first, constant contribution to Eq. (12) only arises if there is loss imbalance TsTi0T_{\text{s}}-T_{\text{i}}\neq 0 and leads to a term that is not suppressed by increasing photon number. Therefore, the value (TsTi)2/(4TsTi)(T_{\text{s}}-T_{\text{i}})^{2}/(4T_{\text{s}}T_{\text{i}}) poses the ultimate limit to the phase uncertainty, even if the phase is optimal. However, this term vanishes for balanced loss Ts=TiT_{\text{s}}=T_{\text{i}}, so that in this case the second contribution dominates.

This contribution can be associated with shot noise due to its scaling with 1/n1/n and is the best one can achieve in a Yurke setup including loss. However, in the regime of small photon numbers, the third contribution might dominate, so that the phase uncertainty will approach a Heisenberg scaling. The optimal phase uncertainty from Eq. (10) and these three contributions are plotted in Fig. 5 for balanced loss and a loss imbalance. Clearly, we observe that for unbalanced loss the phase uncertainty levels off, while for balanced loss we find shot-noise scaling. For low photon numbers, both cases approach the Heisenberg scaling given by the red dashed line.

Refer to caption
Figure 4: Normalized classical Fisher information as a function of nn. The blue lines correspond to a case with equal losses (Ts=Ti=0.8T_{\text{s}}=T_{\text{i}}=0.8). The thick blue solid line denotes the maximal classical Fisher information for each value of nn, i. e. Fc/nF_{\text{c}}/n evaluated at ϕmin\phi_{\text{min}}. It approaches a constant value for high values of nn, which corresponds to shot-noise scaling. The thin light blue lines are the normalized classical Fisher information for specific values of the phase, ϕ=0.97π,0.95π,0.9π\phi=0.97\pi,0.95\pi,0.9\pi (dotted, dashed, solid). The black lines corresponds to a case with non equal losses (Ts=0.8T_{\text{s}}=0.8 and Ti=0.7T_{\text{i}}=0.7). The thick black line denotes the maximal classical Fisher information, evaluated at ϕmin\phi_{\text{min}}, whereas the thin gray lines corresponds to phases ϕ=0.97π,0.95π,0.9π\phi=0.97\pi,0.95\pi,0.9\pi (dotted, dashed, solid). The decrease of the normalized classical Fisher information signifies a scaling worse than shot noise, whereas an increase corresponds to sub-shot-noise behavior.

We can get a clearer physical explanation of the different behavior of the minimum variance in the two regimes described above by rewriting Eq. (6) as

Ns=n[1+Ts+2TsTicosϕ]+n2[Ts+Ti+2TsTicosϕ].\displaystyle\begin{split}N_{\text{s}}=&n\left[1+T_{s}+2\sqrt{T_{s}T_{i}}\cos\phi\right]\\ &+n^{2}\left[T_{s}+T_{i}+2\sqrt{T_{s}T_{i}}\cos\phi\right].\end{split} (13)

In the lossless case, the minimum variance is associated to the phase ϕmin=π\phi_{\text{min}}=\pi that yields Ns=0N_{\text{s}}=0. Including loss, inspection of Eq. (9) shows that in the high-gain regime, ϕminπ\phi_{\text{min}}\rightarrow\pi. For this setting of ϕ\phi, the second term in Eq. (13), scaling with n2n^{2}, vanishes if Ts=TiT_{\text{s}}=T_{\text{i}}. However, this is not the case if TsTiT_{\text{s}}\neq T_{\text{i}}. Therefore the variance of NsN_{\text{s}}, that is Ns(Ns+1)N_{\text{s}}(N_{\text{s}}+1), is lower in the case with Ts=TiT_{\text{s}}=T_{\text{i}} compared to the case with TsTiT_{\text{s}}\neq T_{\text{i}}. This reduced variance translates into better sensitivity in phase estimation for the case with Ts=TiT_{\text{s}}=T_{\text{i}}.

In a realistic scenario, balanced loss is unfeasible. We thus expect a sensitivity worse than shot noise for such a Yurke setup. The question arises whether the Yurke configuration is in fact, for realistic experimental implementations, better suited than the Mandel configuration, which also allows for bicolor quantum imaging with shot-noise limited phase uncertainties.

Refer to caption
Figure 5: Minimal phase uncertainty σY2|min\sigma_{\text{Y}}^{2}|_{\text{min}} as a function of nn. For Ts=0.9T_{\text{s}}=0.9 and Ti=0.85T_{\text{i}}=0.85 (black), we observe a constant limit (gray line) in the high-gain regime, while for Ts=Ti=0.9T_{\text{s}}=T_{\text{i}}=0.9 (blue) we observe shot-noise scaling. For low photon numbers, both cases approach a Heisenberg scaling (dashed, red line). The respective limits and scalings plotted in the figure correspond to the respective terms of Eq. (12).

4 Comparison of Yurke and Mandel setups in the high-gain regime

In a realistic experimental implementation of phase estimation with a Yurke configuration [20, 41, 26], the likely scenario should correspond to the case TsTiT_{s}\neq T_{i}. We have seen in Sec. 3 that in the case of imbalanced losses inside the interferometer, going to the high-gain regime yields a sensitivity in phase estimation that is even worse than the shot-noise limit.

In Sec. 2, we saw that in the lossless case, the Mandel configuration can provide a phase sensitivity equivalent to the shot-noise limit when measuring the differential signal. In this section, we will explore under which conditions, in the presence of loss, the Mandel configuration can still show a sensitivity equivalent to the shot-noise limit, thus surpassing the performance of the Yurke configuration in the high-gain regime.

We start the discussion by focusing on the detection of a single interferometer exit in the Mandel configuration, which even in the lossless case scales worse than shot noise. The photon statistics is also thermal and the interference pattern has exactly the same form as in the Yurke case, but with a modified baseline and contrast, so that we can also resort to formally inserting the respective expressions for aa and bb into Eq. (9) to obtain the optimal phase uncertainty. Indeed, we find in Appendix A the baseline a=n(1+Ts+nTi)/2a=n(1+T_{\text{s}}+nT_{\text{i}})/2 and the parameter b=nTsTi(1+n)b=n\sqrt{T_{\text{s}}T_{\text{i}}(1+n)} that is connected to the contrast through C=b/aC=b/a. Because the contrast is smaller than unity, this experiment inhibits perfect destructive interference. Indeed, we find in the high-gain limit

σSM2|ϕminhgTin4Ts,\left.\sigma^{2}_{\text{SM}}\right|_{\phi_{\text{min}}}\overset{\text{hg}}{\longrightarrow}\frac{T_{\text{i}}n}{4T_{\text{s}}}, (14)

which is a generalization of the case without loss from Eq. (3). Not only is this scaling worse than shot noise, it is even increasing with increasing photon numbers. In contrast to the Yurke setup, the loss imbalance has no effect and this limit cannot be suppressed, as it will always be dominant for sufficiently high nn.

To overcome this issue, from now on we consider measuring the photon-number difference NN_{-} between both exits of the interferometer [35]. Appendix A gives the variance of NN_{-} under the influence of photon loss, which in turn leads to the phase uncertainty

σ2=N++N2+2(1Ti)Tsn2(N/ϕ)2,\displaystyle\sigma^{2}_{-}=\frac{N_{+}+N_{-}^{2}+2(1-T_{\text{i}})T_{\text{s}}n^{2}}{(\partial N_{-}/\partial\phi)^{2}}, (15)

where the photon-number difference and sum of both exits take the explicit form

N=2n(n+1)TiTscosϕN+=n[1+Ts+nTi].\displaystyle\begin{split}N_{-}=&2n\sqrt{(n+1)T_{\text{i}}T_{\text{s}}}\cos\phi\\ N_{+}=&n[1+T_{\text{s}}+nT_{\text{i}}].\end{split} (16)

This phase uncertainty is always minimized at mid-fringe for ϕ=π/2\phi=\pi/2, where N=0N_{-}=0, so that no optimization depending on the transmittance is necessary. In fact, for this phase setting N=0N_{-}=0 and therefore the variance becomes minimal, since N+N_{+} is independent of the phase. At the same time, the slope of the photon-number difference in the denominator is maximal and we find

σ2|π/2=14TiTsn1+Ts+n[Ti+2Ts2TiTs]n+1,\displaystyle\left.\sigma^{2}_{-}\right|_{\pi/2}=\frac{1}{4T_{\text{i}}T_{\text{s}}n}\frac{1+T_{\text{s}}+n[T_{\text{i}}+2T_{\text{s}}-2T_{\text{i}}T_{\text{s}}]}{n+1}, (17)

which retains its shot-noise scaling in the high-gain regime, as the factor Ti+2Ts2TiTsT_{\text{i}}+2T_{\text{s}}-2T_{\text{i}}T_{\text{s}} never vanishes. Indeed, taking the high-gain limit leads to

σ2|π/2hg14n(2Ti+1Ts2),\displaystyle\left.\sigma^{2}_{-}\right|_{\pi/2}\overset{\text{hg}}{\longrightarrow}\frac{1}{4n}\left(\frac{2}{T_{\text{i}}}+\frac{1}{T_{\text{s}}}-2\right), (18)

which shows no exceptional point for equal loss Ti=TsT_{\text{i}}=T_{\text{s}}.

We compare this optimal phase uncertainty to that of the Yurke configuration in Fig. 6. For unbalanced loss, the phase uncertainty of the Yurke setup saturates, whereas the Mandel setup always displays shot-noise scaling and surpasses the Yurke sensitivity at sufficiently high photon numbers. Even for balanced loss, the Yurke interferometer does not exhibit Heisenberg scaling for this range of parameters, providing at best only a fourfold improvement compared to the Mandel configuration. Since perfectly balanced loss is unlikely, the Mandel setup appears advantageous in high-gain quantum imaging.

Refer to caption
Figure 6: Minimal phase uncertainty σj2|min\sigma_{j}^{2}|_{\text{min}} as a function of nn. For the Yurke setup (Y, black) we observe for σY2|min\sigma_{\text{Y}}^{2}|_{\text{min}} a saturation (gray line) in the high-gain regime for Ts=0.8T_{\text{s}}=0.8 and Ti=0.7T_{\text{i}}=0.7 . In contrast, in the differential Mandel setup (M, red), there is no saturation and the sensitivity σ2|π/2\sigma_{-}^{2}|_{\pi/2} surpasses the one of the Yurke case, displaying clear shot-noise scaling. For reference, the dashed blue line denotes the phase uncertainty of the Yurke setup for equal loss Ts=Ti=0.8T_{\text{s}}=T_{\text{i}}=0.8.

5 Conclusions

From a practical perspective, our results provide design guidelines for quantum sensors based on nonlinear interferometers. To achieve genuine Heisenberg scaling, experiments should aim for moderate parametric gain, carefully balanced internal losses, and operation near destructive interference in a Yurke-type geometry. In contrast, when large photon fluxes are needed or when asymmetric and sample-induced losses are unavoidable, a Mandel-type interferometer with differential detection offers a more robust, though shot-noise-limited, solution that can outperform the Yurke setup in the relevant parameter regime.

This trade-off between ultimate scaling and robustness directly affects potential applications of nonlinear interferometers in quantum-enhanced metrology and imaging, especially when the sample itself is a major source of loss. The parameter-space studies and scaling laws developed here serve as a guide for choosing architectures and operating points in future experiments, and they encourage further research on engineered-loss platforms, alternative detection strategies, and multimode extensions aimed at recovering quantum-enhanced sensitivity under more realistic experimental conditions.

Looking ahead, our analysis can be extended in several directions that are relevant for practical quantum sensors. Beyond the internal losses considered here, external detection loss and detector noise can be incorporated using quantum-imaging distillation and detection-loss-tolerant schemes [42, 7], yielding a full error budget from the nonlinear crystals to the camera.

It will also be important to revisit these architectures from the perspective of quantum Fisher information [10]. Yurke and Mandel configurations have the same ultimate quantum Fisher information, but realistic measurement schemes access very different fractions of it once loss and noise are present. Quantifying how alternative readout strategies, such as balanced detection, homodyne or truncated SU(1,1) protocols, approach these bounds in realistic multimode imaging scenarios will guide the choice of architectures for bicolor quantum imaging, nonlinear interferometric spectroscopy, and tomography.

In summary, our results provide guidance on what can be achieved with simple and experimentally easily accessible intensity measurements in the Yurke-type and Mandel-type setups, the two configurations most widely used in nonlinear interferometry. We identify the configuration that provides the best sensitivity for phase estimation under different gain and loss regimes, and the conditions under which genuine Heisenberg-limited precision can be achieved with intensity measurements, surpassing the shot-noise limit.

Acknowledgments

The authors thank P. Schach for stimulating discussions. D.F.U., A.A.P. and J.P.T. acknowledge support from the project NOVISLIGHT (PID2023-149780NB-I00) funded by Ministerio de Ciencia, Innovación y Universidades (Proyectos de generación de conocimiento 2023). This work was partially funded by CEX2024-001490-S [MICIU/AEI/10.13039/501100011033]. G.J.M., R.B.P. and I.A.W. acknowledge support from the National Research Council Canada (project QSP 062-2) and the UK Research and Innovation Future Leaders Fellowship (project: MR/W011794/1), Department of Science, Innovation & Technology Tactical Fund. Support is acknowledged from the German Federal Ministry of Education and Research (BMBF) within the funding program “quantum technologies – from basic research to market” with contract number 13N16496 (QUANCER). Moreover, this work is also funded by the Deutsche Forschungsgemeinschaft (DFG, German Research Foundation) – Projektnummer 552245798.

Appendix A Interference patterns

A.1 Interferometer output operator

In this appendix, we derive the bosonic operators that account for the evolution of a photon through the setup. To describe parametric down-conversion, we use Bogoliubov transformations that have coupling matrices that are elements of the SU(1,1) group. For our two nonlinear media j=A,Bj=A,B, they take the form

b^s/i=uja^s/i+vja^i/s,\hat{b}_{\text{s/i}}=u_{j}\hat{a}_{\text{s/i}}+v_{j}\hat{a}^{\dagger}_{\text{i/s}}, (19)

where b^s/i\hat{b}_{\text{s/i}} is the output annihilation operator of that particular medium in signal/idler mode and a^s/i\hat{a}_{\text{s/i}} is the respective input annihilation operator, which of course fulfill the usual bosonic commutation relations. The complex coefficients uju_{j} and vjv_{j} increase exponentially with parametric gain, but also include the phase of the pump laser at interaction, as parametric amplification is a phase-sensitive process. They fulfill the property |uj|2|vj|2=UjVj=1|u_{j}|^{2}-|v_{j}|^{2}=U_{j}-V_{j}=1, which corresponds to the hyperbolic identity, so that the coefficients can be parametrized by hyperbolic functions that scale exponentially with gain.

The loss on each arm j=s,ij=\text{s},\text{i} is included by conventional SU(2) beam-splitter matrices, namely

b^j=tja^j+rjl^j,\hat{b}_{j}=t_{j}\hat{a}_{j}+r_{j}\hat{l}_{j}, (20)

where the bosonic operators l^j\hat{l}_{j} describe vacuum input modes of the respective loss channels. The complex transmission coefficients fulfill the relation |tj|2+|rj|2=Tj+Rj=1|t_{j}|^{2}+|r_{j}|^{2}=T_{j}+R_{j}=1 and imply a transmittance TjT_{j} on arm jj. This transmittance may include the transmittance of the object used in an imaging setup, but also any other loss on that arm. Moreover, the phase acquired on each arm can be included as the phase of tjt_{j}, so that the phase shift induced by the imaged object is accounted for.

To find the transformations for the different setups, one can step by step follow the sketches in Fig. 1 and insert the corresponding SU(1,1) and SU(2) transformations. The Yurke configuration is characterized by the output operator

b^Y=αYa^sA+κYl^s+βYa^i+λYl^i,\displaystyle\hat{b}_{\text{Y}}=\alpha_{\text{Y}}\hat{a}_{\text{s}A}+\kappa_{\text{Y}}\hat{l}_{\text{s}}+\beta_{\text{Y}}\hat{a}_{\text{i}}^{\dagger}+\lambda_{\text{Y}}\hat{l}_{\text{i}}^{\dagger}, (21)

in the signal mode, which effectively describes the transformation of the entire interferometer. Here, the annihilation operators a^sA\hat{a}_{\text{s}A} and a^i\hat{a}_{\text{i}} denote vacuum input at the first nonlinear medium AA, while the annihilation operators operators l^s\hat{l}_{\text{s}} and l^i\hat{l}_{\text{i}} are the vacuum inputs associated with the loss channels in the signal and idler arm, respectively. The complex path coefficients contain all the information of the interferometer, that is, the gain, loss, sample properties, and phases, and are given by

αY\displaystyle\alpha_{\text{Y}} =uBtsuA+vBtivA and κY=uBrs\displaystyle=u_{B}t_{\text{s}}u_{A}+v_{B}t_{\text{i}}^{\ast}v_{A}^{\ast}\text{ and }\kappa_{\text{Y}}=u_{B}r_{\text{s}} (22a)
βY\displaystyle\beta_{\text{Y}} =uBtsvA+vBtiuA and λY=vBri.\displaystyle=u_{B}t_{\text{s}}v_{A}+v_{B}t_{\text{i}}^{\ast}u_{A}^{\ast}\text{ and }\lambda_{\text{Y}}=v_{B}r_{\text{i}}^{\ast}. (22b)

For the Mandel configuration, the signal mode of medium AA is not used as an input for medium BB. Hence, an additional (vacuum) input channel associated with operator a^sB\hat{a}_{\text{s}B} has to be included into the description. Moreover, the two signal modes interfere on a 50:50 beam splitter through the SU(2) transformation b^M(±)=(b^sA±b^sB)/2\hat{b}_{\text{M}}^{(\pm)}=(\hat{b}_{\text{s}A}\pm\hat{b}_{\text{s}B})/\sqrt{2}. As a consequence, we find two output operators after the beam splitter, namely

b^M(±)=αM(±)a^sA+γM(±)a^sB+κMl^s+βM(±)a^i+λM(±)l^i,\displaystyle\begin{split}\hat{b}^{(\pm)}_{\text{M}}=&\,\alpha^{(\pm)}_{\text{M}}\hat{a}_{\text{s}A}+\gamma^{(\pm)}_{\text{M}}\hat{a}_{\text{s}B}+\kappa_{\text{M}}\hat{l}_{\text{s}}\\ &+\beta^{(\pm)}_{\text{M}}\hat{a}_{\text{i}}^{\dagger}+\lambda^{(\pm)}_{\text{M}}\hat{l}_{\text{i}}^{\dagger},\end{split} (23)

with path coefficients

αM(±)\displaystyle\alpha_{\text{M}}^{(\pm)} =tsuA±vBtivA2 , γM(±)=±uB2 , κM=rs2\displaystyle=\frac{t_{\text{s}}u_{A}\pm v_{B}t_{\text{i}}^{\ast}v_{A}^{\ast}}{\sqrt{2}}\text{ , }\gamma_{\text{M}}^{(\pm)}=\frac{\pm u_{B}}{\sqrt{2}}\text{ , }\kappa_{\text{M}}=\frac{r_{\text{s}}}{\sqrt{2}} (24a)
βM(±)\displaystyle\beta_{\text{M}}^{(\pm)} =tsvA±vBtiuA2 and λM(±)=±vBri2.\displaystyle=\frac{t_{\text{s}}v_{A}\pm v_{B}t_{\text{i}}^{\ast}u_{A}^{\ast}}{\sqrt{2}}\text{ and }\lambda_{\text{M}}^{(\pm)}=\pm\frac{v_{B}r_{\text{i}}^{\ast}}{\sqrt{2}}. (24b)

Modifying the Mandel setup such that the lost fraction of intensity of the signal mode couples back into the second nonlinear medium leads to a hybrid between the Mandel and Yurke configuration, where the path coefficients are altered by the modified seed of the second medium. We therefore arrive at

d^H(±)=αH(±)a^sA+γH(±)a^sB+βH(±)a^i,\displaystyle\hat{d}^{(\pm)}_{\text{H}}=\alpha^{(\pm)}_{\text{H}}\hat{a}_{\text{s}A}+\gamma^{(\pm)}_{\text{H}}\hat{a}_{\text{s}B}+\beta^{(\pm)}_{\text{H}}\hat{a}_{\text{i}}^{\dagger}, (25)

for a configuration without loss, where we have defined the path coefficients

αH(+)=rrsuA+tαY\displaystyle\alpha_{\text{H}}^{(+)}=-rr_{\text{s}}^{\ast}u_{A}+t\alpha_{\text{Y}} αH()=trsuArαY\displaystyle\text{ , }\alpha_{\text{H}}^{(-)}=-t^{\ast}r_{\text{s}}^{\ast}u_{A}-r^{\ast}\alpha_{\text{Y}} (26a)
βH(+)=rrsvA+tβY\displaystyle\beta_{\text{H}}^{(+)}=-rr_{\text{s}}^{\ast}v_{A}+t\beta_{\text{Y}} βH()=trsvArβY\displaystyle\text{ , }\beta_{\text{H}}^{(-)}=-t^{\ast}r_{\text{s}}^{\ast}v_{A}-r^{\ast}\beta_{\text{Y}} (26b)
γH(+)=tuBrs+rts\displaystyle\gamma_{\text{H}}^{(+)}=tu_{B}r_{\text{s}}+rt_{\text{s}}^{\ast} γH()=ruBrs+tts.\displaystyle\text{ , }\gamma_{\text{H}}^{(-)}=-r^{\ast}u_{B}r_{\text{s}}+t^{\ast}t_{\text{s}}^{\ast}. (26c)

We now introduce the tuning parameter ϱ=1Ts\varrho=1-T_{\text{s}} to tune the fraction that couples out of the Yurke setup. At the same time, we choose the transmittance of the final beam splitter to be |t|2=1ϱ/2|t|^{2}=1-\varrho/2, in agreement with Fig. 1. These choices lead for ϱ=0\varrho=0 to the Yurke and for ϱ=1\varrho=1 to the Mandel configuration.

A.2 Interference signal

In contrast to configurations that use a coherent input to boost the performance of the interferometers [43], we assume no seed, that is, vacuum input. In this case, the detected photon numbers are always of the form |βj|2+|λj|2|\beta_{j}|^{2}+|\lambda_{j}|^{2} for all three considered setups, where |λj|2|\lambda_{j}|^{2} can be interpreted as vacuum noise introduced by loss.

As reported before [9], we observe for the Yurke configuration that the detected signal photons can be described by the interference pattern Ns=a+bcosϕN_{\text{s}}=a+b\cos\phi, with

a\displaystyle a =TsVA+VB+(Ts+Ti)VAVB\displaystyle=T_{\text{s}}V_{A}+V_{B}+(T_{\text{s}}+T_{\text{i}})V_{A}V_{B} (27a)
b\displaystyle b =2TsTiVAVB(1+VA)(1+VB).\displaystyle=2\sqrt{T_{\text{s}}T_{\text{i}}V_{A}V_{B}(1+V_{A})(1+V_{B})}. (27b)

Here, aa corresponds to the baseline of the pattern, the contrast is determined by C=b/aC=b/a, and the interferometer phase has been defined as ϕ=arg(tstivAvBuAuB)\phi=\operatorname{arg}(t_{\text{s}}t_{\text{i}}v_{A}v_{B}^{\ast}u_{A}u_{B}). For equal gain, VA=VB=nV_{A}=V_{B}=n corresponds to the number of photons generated in nonlinear medium AA and is used to determine whether we observe sub-shot-noise sensitivities.

For the Mandel configuration, the two exits show the photon numbers

Ns\displaystyle N_{\text{s}} =b^M(+)b^M(+)=a+bcosϕ\displaystyle=\big\langle\hat{b}^{(+)\dagger}_{\text{M}}\hat{b}^{(+)}_{\text{M}}\big\rangle=a+b\cos\phi (28a)
Ns\displaystyle N_{\text{s}}^{\prime} =b^M()b^M()=abcosϕ,\displaystyle=\big\langle\hat{b}^{(-)\dagger}_{\text{M}}\hat{b}^{(-)}_{\text{M}}\big\rangle=a-b\cos\phi, (28b)

with the interferometer phase ϕ=arg(tstivAvBuA)\phi=\operatorname{arg}(t_{\text{s}}t_{\text{i}}v_{A}v_{B}^{\ast}u_{A}), which effectively corresponds to the phase observed for the Yurke setup as uBu_{B} can be assumed as real. Here, we find the parameters

a\displaystyle a =(TsVA+VB+TiVAVB)/2\displaystyle=(T_{\text{s}}V_{A}+V_{B}+T_{\text{i}}V_{A}V_{B})/2 (29a)
b\displaystyle b =TsTiVAVB(1+VA),\displaystyle=\sqrt{T_{\text{s}}T_{\text{i}}V_{A}V_{B}(1+V_{A})}, (29b)

as a slight generalization of established results [29].

Next, we turn to the hybrid configuration without loss and rewrite the tuning parameter as ϱ=1τ\varrho=1-\tau. Because the setup effectively corresponds to three-path interference [37], in general a beating occurs. However, we can identify the phases associated with the Mandel and Yurke setups, as above, and choose in the following ϕM=ϕY=ϕ\phi_{\text{M}}=\phi_{\text{Y}}=\phi. Here, the phase of the Mandel setup includes an additional shift of π\pi imprinted upon reflection by the beam splitter that couples light out of the Yurke setup. This phase shift is necessary so that the Mandel result can be recovered without the two exits changing their role. With the identification Ts=τT_{\text{s}}=\tau and |t|2=(1+τ)/2|t|^{2}=(1+\tau)/2, we find for the sum of the photon numbers of both exits

N+=n[n+2+nτ+2(1+n)τcosϕ],N_{+}=n\left[n+2+n\tau+2(1+n)\sqrt{\tau}\cos\phi\right], (30)

and a photon-number difference in the usual form N=a+bcosϕN_{-}=a+b\cos\phi, however, with

a\displaystyle a =2nϱ(n+1)(τ+τ2)+n2τ+n(2+n)τ2\displaystyle=2n\varrho\sqrt{(n+1)(\tau+\tau^{2})}+n^{2}\tau+n(2+n)\tau^{2} (31a)
b\displaystyle b =2n[(1+n)τ3/2+ϱ(1+n)(1+τ)].\displaystyle=2n\bigg[(1+n)\tau^{3/2}+\varrho\sqrt{(1+n)(1+\tau)}\bigg]. (31b)

A.3 Variances of detected photon numbers

With the help of basic bosonic commutation relations, we bring the variance to normal order, that is, Var(Ns)=b^jb^jb^jb^jNs2=b^j2b^j2+NsNs2\operatorname{Var}(N_{\text{s}})=\big\langle\hat{b}_{j}^{\dagger}\hat{b}_{j}\hat{b}_{j}^{\dagger}\hat{b}_{j}\big\rangle-N_{\text{s}}^{2}=\big\langle\hat{b}_{j}^{\dagger 2}\hat{b}_{j}^{2}\big\rangle+N_{\text{s}}-N_{\text{s}}^{2}. The second-order correlation function, when acting on the vacuum input, can then be expressed by

b^j2b^j2=2(|βj|4+|λj|4)+4|βj|2|λj|2=2Ns2.\big\langle\hat{b}_{j}^{\dagger 2}\hat{b}_{j}^{2}\big\rangle=2(|\beta_{j}|^{4}+|\lambda_{j}|^{4})+4|\beta_{j}|^{2}|\lambda_{j}|^{2}=2N_{\text{s}}^{2}. (32)

As a consequence, we find the variance

Var(Ns)=Ns[1+Ns],\operatorname{Var}(N_{\text{s}})=N_{\text{s}}[1+N_{\text{s}}], (33)

in agreement with thermal statistics. The same calculation holds for the other exit and the variance of NsN_{\text{s}}^{\prime}.

To calculate the variance of the photon-number difference N=NsNsN_{-}=N_{\text{s}}-N_{\text{s}}^{\prime} in the Mandel setup, we first identify Var(N)=Var(Ns)+Var(Ns)2Cov(Ns,Ns)\operatorname{Var}(N_{-})=\operatorname{Var}(N_{\text{s}})+\operatorname{Var}(N_{\text{s}}^{\prime})-2\operatorname{Cov}(N_{\text{s}},N_{\text{s}}^{\prime}), where we introduce the covariance

Cov(Ns,Ns)=b^M(+)b^M()b^M(+)b^M()+c.c.2NsNs,\operatorname{Cov}(N_{\text{s}},N_{\text{s}}^{\prime})=\frac{\big\langle\hat{b}_{\text{M}}^{(+)\dagger}\hat{b}_{\text{M}}^{(-)\dagger}\hat{b}_{\text{M}}^{(+)}\hat{b}_{\text{M}}^{(-)}\big\rangle+\text{c.c.}}{2}-N_{\text{s}}N_{\text{s}}^{\prime}, (34)

already normally ordered. When the respective operators act on the input vacuum state, we find for the cross-correlation function the expression

b^M(+)b^M()b^M(+)b^M()=2NsNsVBVATs(1Ti).\big\langle\hat{b}_{\text{M}}^{(+)\dagger}\hat{b}_{\text{M}}^{(-)\dagger}\hat{b}_{\text{M}}^{(+)}\hat{b}_{\text{M}}^{(-)}\big\rangle=2N_{\text{s}}N_{\text{s}}^{\prime}-V_{B}V_{A}T_{\text{s}}(1-T_{\text{i}}). (35)

With that, the variance of the photon-number difference reduces to

Var(N)=N++N2+2VBVATs(1Ti),\operatorname{Var}(N_{-})=N_{+}+N_{-}^{2}+2V_{B}V_{A}T_{\text{s}}(1-T_{\text{i}}), (36)

with N+=NsNsN_{+}=N_{\text{s}}-N_{\text{s}}^{\prime}. For the lossless hybrid setup, we also find the respective limit Var(N)=N++N2\operatorname{Var}(N_{-})=N_{+}+N_{-}^{2}.

Appendix B Classical Fisher information associates to thermal light

The photon-number probability distribution associated to thermal light is

pm=11+Ns(ϕ)[Ns(ϕ)1+Ns(ϕ)]m,p_{m}=\frac{1}{1+N_{\text{s}}(\phi)}\,\left[\frac{N_{\text{s}}(\phi)}{1+N_{\text{s}}(\phi)}\right]^{m}, (37)

where mm is the number of photons and ϕ\phi is the phase to be estimated. Ns(ϕ)N_{\text{s}}(\phi) is the mean value of the probability distribution, and the variance is Ns(ϕ)[1+Ns(ϕ)]N_{\text{s}}(\phi)\left[1+N_{\text{s}}(\phi)\right].

The classical Fisher information associated to a probability distribution pmp_{m} can be written as

Fc(ϕ)=mpm(2lnpmϕ2).F_{\text{c}}(\phi)=\sum_{m}p_{m}\,\left(\frac{\partial^{2}\ln p_{m}}{\partial\phi^{2}}\right). (38)

From Eq. (37), we obtain that

2lnpmϕ2={1+m[1+Ns]2mNs2}[Nsϕ]2+{1+m1+Ns+mNs}2Nsϕ2.\displaystyle\begin{split}\frac{\partial^{2}\ln p_{m}}{\partial\phi^{2}}=\left\{\frac{1+m}{\left[1+N_{\text{s}}\right]^{2}}-\frac{m}{N_{\text{s}}^{2}}\right\}\left[\frac{\partial N_{\text{s}}}{\partial\phi}\right]^{2}\\ +\left\{-\frac{1+m}{1+N_{\text{s}}}+\frac{m}{N_{\text{s}}}\right\}\frac{\partial^{2}N_{\text{s}}}{\partial\phi^{2}}.\end{split} (39)

Making use of mpm=1\sum_{m}p_{m}=1 and mmpm=Ns(ϕ)\sum_{m}m\,p_{m}=N_{\text{s}}(\phi), substitution of Eq. (39) into Eq. (38) yields

F(ϕ)=1Ns(ϕ)[1+Ns(ϕ)][Ns(ϕ)ϕ]2.F(\phi)=\frac{1}{N_{\text{s}}(\phi)\left[1+N_{\text{s}}(\phi)\right]}\left[\frac{\partial N_{\text{s}}(\phi)}{\partial\phi}\right]^{2}. (40)

We have thus demonstrated that the classical Fisher information is F(ϕ)=1/σY2(ϕ)F(\phi)=1/\sigma_{Y}^{2}(\phi), where σY2(ϕ)\sigma_{Y}^{2}(\phi) is obtained using the propagation of errors equation.

References

  • [1] M. V. Chekhova and Z. Y. Ou. “Nonlinear interferometers in quantum optics”. Adv. Opt. Photon. 8, 104 (2016).
  • [2] B. Yurke, S. L. McCall, and J. R. Klauder. “SU(2) and SU(1,1) interferometers”. Phys. Rev. A 33, 4033–4054 (1986).
  • [3] X. Y. Zou, L. J. Wang, and L. Mandel. “Induced coherence and indistinguishability in optical interference”. Phys. Rev. Lett. 67, 318–321 (1991).
  • [4] A. Hochrainer, M. Lahiri, M. Erhard, M. Krenn, and A. Zeilinger. “Quantum indistinguishability by path identity and with undetected photons”. Rev. Mod. Phys. 94, 025007 (2022).
  • [5] J. Fuenzalida, E. Giese, and M. Gräfe. “Nonlinear interferometry: A new approach for imaging and sensing”. Adv. Quantum Technol. 7, 2300353 (2024).
  • [6] G. Barreto Lemos, V. Borish, G. D. Cole, S. Ramelow, R. Lapkiewicz, and A. Zeilinger. “Quantum imaging with undetected photons”. Nature 512, 409–412 (2014).
  • [7] J. Fuenzalida, M. Gilaberte Basset, S. Töpfer, J. P. Torres, and M. Gräfe. “Experimental quantum imaging distillation with undetected light”. Sci. Adv. 9, eadg9573 (2023).
  • [8] A. Vallés, G. Jiménez, L. J. Salazar-Serrano, and J. P. Torres. “Optical sectioning in induced coherence tomography with frequency-entangled photons”. Phys. Rev. A 97, 023824 (2018).
  • [9] E. Giese, S. Lemieux, M. Manceau, R. Fickler, and R. W. Boyd. “Phase sensitivity of gain-unbalanced nonlinear interferometers”. Phys. Rev. A 96, 053863 (2017).
  • [10] J. Kranias, G. Thekkadath, K. Heshami, and A. Z. Goldberg. “Metrological Advantages in Seeded and Lossy Nonlinear Interferometers”. Quantum 9, 1619 (2025).
  • [11] J. Flórez, E. Giese, D. Curic, L. Giner, R. W. Boyd, and J. S. Lundeen. “The phase sensitivity of a fully quantum three-mode nonlinear interferometer”. New J. Phys. 20, 123022 (2018).
  • [12] J. P. Torres, K. Banaszek, and I. A. Walmsley. “Engineering nonlinear optic sources of photonic entanglement”. Progress in Optics 56, 227–331 (2011).
  • [13] L. C. Biedenharn, J. Nuyts, and N. Straumann. “On the unitary representations of SU(1,1)SU(1,1) and SU(2,1)SU(2,1)”. Ann. Inst. Henri Poincaré A 3, 13–39 (1965). url: https://www.numdam.org/item/AIHPA_1965__3_1_13_0/.
  • [14] T. J. Herzog, J. G. Rarity, H. Weinfurter, and A. Zeilinger. “Frustrated two-photon creation via interference”. Phys. Rev. Lett. 72, 629–632 (1994).
  • [15] Y. Michael, I. Jonas, L. Bello, M.-E. Meller, E. Cohen, M. Rosenbluh, and A. Pe’er. “Augmenting the sensing performance of entangled photon pairs through asymmetry”. Phys. Rev. Lett. 127, 173603 (2021).
  • [16] C. Lindner, J. Kunz, S. J. Herr, S. Wolf, J. Kießling, and F. Kühnemann. “Nonlinear interferometer for Fourier-transform mid-infrared gas spectroscopy using near-infrared detection”. Opt. Express 29, 4035–4047 (2021).
  • [17] T. Tashima, Y. Mukai, M. Arahata, N. Oda, M. Hisamitsu, K. Tokuda, R. Okamoto, and S. Takeuchi. “Ultra-broadband quantum infrared spectroscopy”. Optica 11, 81–87 (2024).
  • [18] I. Kviatkovsky, H. M. Chrzanowski, E. G. Avery, H. Bartolomaeus, and S. Ramelow. “Microscopy with undetected photons in the mid-infrared”. Sci. Adv. 6, eabd0264 (2020).
  • [19] M. Gilaberte Basset, A. Hochrainer, S. Töpfer, F. Riexinger, P. Bickert, J. R. León-Torres, F. Steinlechner, and M. Gräfe. “Video-rate imaging with undetected photons”. Laser Photonics Rev. 15, 2000327 (2021).
  • [20] S. Töpfer, M. Gilaberte Basset, J. Fuenzalida, F. Steinlechner, J. P. Torres, and M. Gräfe. “Quantum holography with undetected light”. Sci. Adv. 8, eabl4301 (2022).
  • [21] J. R. León-Torres, F. Krajinić, M. Kumar, M. Gilaberte Basset, F. Setzpfandt, V. Flavio Gili, B. Jelenković, and M. Gräfe. “Off-axis holographic imaging with undetected light”. Opt. Express 32, 35449–35461 (2024).
  • [22] S. Töpfer, S. Tovar-Perez, J. R. León Torres, D. Derr, E. Giese, J. Fuenzalida, and M. Gräfe. “Synthetic quantum holography with undetected light”. Optica Quantum 3, 129–136 (2025).
  • [23] A. M. Marino, N. V. Corzo Trejo, and P. D. Lett. “Effect of losses on the performance of an SU(1,1) interferometer”. Phys. Rev. A 86, 023844 (2012).
  • [24] Q.-K. Gong, X.-L. Hu, D. Li, C.-H. Yuan, Z. Y. Ou, and W. Zhang. “Intramode-correlation-enhanced phase sensitivities in an SU(1,1) interferometer”. Phys. Rev. A 96, 033809 (2017).
  • [25] M. Manceau, F. Khalili, and M. Chekhova. “Improving the phase super-sensitivity of squeezing-assisted interferometers by squeeze factor unbalancing”. New J. Phys. 19, 013014 (2017).
  • [26] M. Manceau, G. Leuchs, F. Khalili, and M. Chekhova. “Detection loss tolerant supersensitive phase measurement with an SU(1,1) interferometer”. Phys. Rev. Lett. 119, 223604 (2017).
  • [27] L. J. Wang, X. Y. Zou, and L. Mandel. “Induced coherence without induced emission”. Phys. Rev. A 44, 4614–4622 (1991).
  • [28] H. M. Wiseman and K. Mølmer. “Induced coherence with and without induced emission”. Phys. Lett. A 270, 245–248 (2000).
  • [29] M. I. Kolobov, E. Giese, S. Lemieux, R. Fickler, and R. W. Boyd. “Controlling induced coherence for quantum imaging”. J. Opt. 19, 054003 (2017).
  • [30] G. J. Machado, L. Sendra, A. Vallés, and J. P. Torres. “Complementarity relationship between coherence and path distinguishability in an interferometer based on induced coherence”. Phys. Rev. A 110, 012421 (2024).
  • [31] B. E. Anderson, B. L. Schmittberger, P. Gupta, K. M. Jones, and P. D. Lett. “Optimal phase measurements with bright- and vacuum-seeded SU(1,1) interferometers”. Phys. Rev. A 95, 063843 (2017).
  • [32] B. E. Anderson, P. Gupta, B. L. Schmittberger, T. Horrom, C. Hermann-Avigliano, K. M. Jones, and P. D. Lett. “Phase sensing beyond the standard quantum limit with a variation on the SU(1,1) interferometer”. Optica 4, 752–756 (2017).
  • [33] P. Gupta, B. L. Schmittberger, B. E. Anderson, K. M. Jones, and P. D. Lett. “Optimized phase sensing in a truncated SU(1,1) interferometer”. Opt. Express 26, 391–401 (2018).
  • [34] C. M. Caves. “Reframing SU(1,1) interferometry”. Adv. Quantum Technol. 3, 1900138 (2020).
  • [35] N. R. Miller, S. Ramelow, and W. N. Plick. “Versatile Super-Sensitive Metrology Using Induced Coherence”. Quantum 5, 458 (2021).
  • [36] F. Roeder, R. Pollmann, V. Quiring, C. Eigner, B. Brecht, and C. Silberhorn. “Towards integrated sensors for optimized OCT with undetected photons” (2025). arXiv:2508.05320.
  • [37] N. R. Gemmell, Y. Ma, E. Pearce, J. Flórez, O. Czerwinski, M. S. Kim, R. F. Oulton, A. S. Clark, and C. C. Phillips. “Coupling undetected sensing modes by quantum erasure”. APL Quantum 1, 036121 (2024).
  • [38] V. Balic, D. A. Braje, P. Kolchin, G. Y. Yin, and S. E. Harris. “Generation of paired photons with controllable waveforms”. Phys. Rev. Lett. 94, 183601 (2005).
  • [39] J. M. Gené, M. H. M. Passos, D. F. Urrego, and J. P. Torres. “Sensitivity of phase estimation with an SU(1,1) interferometer beyond the single-mode approximation”. Article in preparation (2026).
  • [40] J. Xin, X.-M. Lu, X. Li, and G. Li. “Optimal phase point for SU(1,1) interferometer”. J. Opt. Soc. Am. B 36, 2824–2833 (2019).
  • [41] D. A. Kalashnikov, A. V. Paterova, S. P. Kulik, and L. A. Krivitsky. “Infrared spectroscopy with visible light”. Nat. Photonics 10, 98–101 (2016).
  • [42] S. Schaffrath, D. Derr, M. Gräfe, and E. Giese. “Quantum imaging beyond the standard-quantum limit and phase distillation”. New J. Phys. 26, 023018 (2024).
  • [43] W. N. Plick, J. P. Dowling, and G. S Agarwal. “Coherent-light-boosted, sub-shot noise, quantum interferometry”. New J. Phys. 12, 083014 (2010).