Instrument Control Parameters Overview

This document contains the instrument control parameters for GC-MS analysis including: 1) GC oven temperature ramps starting at 35°C and ramping to 250°C over 23.5 minutes; 2) Split/splitless front inlet parameters including a split ratio of 50:1; 3) Splitless back inlet parameters including a 1 minute purge time; 4) Column specifications including a 30m Rtx-5 column connected to the front inlet and a 40m column connected to the back inlet; 5) MS acquisition parameters including a scan range of 40-200m/z.

Uploaded by

holykarmor
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as TXT, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
118 views4 pages

Instrument Control Parameters Overview

This document contains the instrument control parameters for GC-MS analysis including: 1) GC oven temperature ramps starting at 35°C and ramping to 250°C over 23.5 minutes; 2) Split/splitless front inlet parameters including a split ratio of 50:1; 3) Splitless back inlet parameters including a 1 minute purge time; 4) Column specifications including a 30m Rtx-5 column connected to the front inlet and a 40m column connected to the back inlet; 5) MS acquisition parameters including a scan range of 40-200m/z.

Uploaded by

holykarmor
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as TXT, PDF, TXT or read online on Scribd

INSTRUMENT CONTROL PARAMETERS:

Instr

ument #1
---------------------------------------------C:\MSDCHEM\1\METHODS\UNDERGRADYEASTPROJECT\SPME_and_StaticHeadspace.M
Control Information
------- ----------Sample Inlet
:
Injection Source :
Injection Location:
Mass Spectrometer :

GC
Manual
Rear
Enabled

=============================================================================
6890 GC METHOD
=============================================================================
OVEN
Initial temp: 35 'C (On)
Initial time: 2.00 min
Ramps:
# Rate Final temp Final time
1 10.00
250
0.00
2 0.0(Off)
Post temp: 50 'C
Post time: 0.00 min
Run time: 23.50 min

CRYO (N2)
Cryo: Off
Cryo fault: Off
Cryo timeout: 120.00 min (Off)
Quick cryo cool: Off
Ambient temp: 30 'C

FRONT INLET (SPLIT/SPLITLESS)


Mode: Split
Initial temp: 230 'C (On)
Pressure: 8.06 psi (On)
Split ratio: 50:1
Split flow: 71.6 mL/min
Total flow: 81.6 mL/min
Gas saver: On
Saver flow: 20.0 mL/min
Saver time: 2.00 min
Gas type: Hydrogen

BACK INLET (SPLIT/SPLITLESS)


Mode: Splitless
Initial temp: 250 'C (On)
Pressure: 17.53 psi (On)
Purge flow: 6.5 mL/min
Purge time: 1.00 min
Total flow: 14.5 mL/min
Gas saver: On
Saver flow: 15.0 mL/min
Saver time: 1.00 min
Gas type: Hydrogen

COLUMN 1
Capillary Column
Model Number: Restek Rtx-5

COLUMN 2
Capillary Column
Nominal length: 30.0 m
Nominal diameter: 250.00 um
Nominal film thickness: 0.50 um
Mode: constant pressure
Pressure: 8.06 psi
Nominal initial flow: 1.4 mL/min
Average velocity: 39 cm/sec
Inlet: Front Inlet
Outlet: Front Detector
Outlet pressure: ambient

Max temperature: 330 'C


Nominal length: 40.0 m
Nominal diameter: 180.00 um
Nominal film thickness: 0.40 um
Mode: constant flow
Initial flow: 1.0 mL/min
Nominal init pressure: 17.55 psi
Average velocity: 46 cm/sec
Inlet: Back Inlet

Maximum temp: 330 'C


Equilibration time: 0.50 min

Outlet: MSD
Outlet pressure: vacuum
FRONT DETECTOR (FID)
Temperature: 250 'C (Off)
Hydrogen flow: 40.0 mL/min (Off)
Air flow: 450.0 mL/min (Off)
Mode: Constant makeup flow
Makeup flow: 45.0 mL/min (Off)
Makeup Gas Type: Nitrogen
Flame: Off
Electrometer: Off
Lit offset: 2.0

BACK DETECTOR (TCD)


Temperature: 250 'C (Off)
Reference flow: 20.0 mL/min (Off)
Mode: Constant makeup flow
Makeup flow: 7.0 mL/min (Off)
Makeup Gas Type: Nitrogen
Filament: Off
Negative polarity: Off

SIGNAL 1
Data rate: 50 Hz
Type: test plot
Save Data: On
Zero: 0.0 (Off)
Range: 0
Fast Peaks: Off
Attenuation: 0

SIGNAL 2
Data rate: 20 Hz
Type: back detector
Save Data: Off
Zero: 0.0 (Off)
Range: 0
Fast Peaks: Off
Attenuation: 0

COLUMN COMP 1
Derive from front detector

COLUMN COMP 2
Derive from back detector

THERMAL AUX 1
Use: MSD Transfer Line Heater
Description:
Initial temp: 270 'C (On)
Initial time: 0.00 min
# Rate Final temp Final time
1 0.0(Off)
VALVES
Valve 7 Remote Start
TIME TABLE
Time

POST RUN
Post Time: 0.00 min

Specifier

Parameter & Setpoint

GC Injector
Front Injector:
No parameters specified
Back Injector:
Injector not configured, use these parameters if it becomes configured
Sample Washes
0
Sample Pumps
5
Injection Volume
0.50 microliters

Syringe Size
PostInj Solvent A Washes
PostInj Solvent B Washes
Viscosity Delay
Plunger Speed

5.0 microliters
3
0
1 seconds
Fast

Column 1 Inventory Number : AB008


Column 2 Inventory Number : AB001
MS ACQUISITION PARAMETERS
General Information
------- ----------Tune File
Acquistion Mode

: atune.u
: Scan

MS Information
-- ----------Solvent Delay

: 0.00 min

EM Absolute
Resulting EM Voltage

: True
: 1952.9

[Scan Parameters]
Low Mass
High Mass
Threshold
Sample #

:
:
:
:

40.0
200.0
0
1

A/D Samples

[MSZones]
MS Quad
MS Source
Timed Events
----- ------

: 150 C
: 230 C

maximum 200 C
maximum 250 C

[Timed MS Detector Entries]


Time (min)
0.01

State (MS on/off)


On
END OF MS ACQUISITION PARAMETERS
TUNE PARAMETERS for SN: US91922553
-----------------------------

EMISSION
ENERGY
REPELLER
IONFOCUS
ENTRANCE_LE
EMVOLTS
AMUGAIN

:
:
:
:
:
:
:

34.610
69.922
29.955
90.157
0.000
2941.176
2483.000

AMUOFFSET
FILAMENT
DCPOLARITY
ENTLENSOFFS
MASSGAIN
MASSOFFSET

:
:
:
:
:
:

120.000
2.000
1.000
25.098
273.000
-10.000
END OF TUNE PARAMETERS
------------------------------------

END OF INSTRUMENT CONTROL PARAMET


ERS
-----------------------------------