User Manual
SPD3000 Series Programmable DC Power Supply
2012
SIGLENT TECHNOLOGIES CO., LTD
SIGLENT
Copyright information
1. SIGLENT TECHNOLOGIES CO., LTD. All Rights Reserved. 2. Information in this publication replaces all previous corresponding material. 3. SIGLENT reserves the right to change the specifications and price of the product. 4. Any way of copying, extracting and translating the content of the manual is not allowed without permission of SIGLENT.
Note:
SIGLENT is the registered trademark of SIGLENT TECHNOLOGIES CO., LTD.
SPD3000 User Manual
SIGLENT
General Safety Summary
Please review the following safety precautions carefully to avoid personal injury or damage to this product or any product connected to it. To prevent potential danger, please use the instrument as specified. Use proper power cord Only the power cord designed for the instrument and authorized by local country could be used. Power supply AC Input Voltages: 100V/110V/220V/230V 10%, 50/60Hz. Use proper fuse The fuse types: 100V/110V: T6.3A/250V; 220V/230V: T3.15A/250V; Make sure to use the correct type of fuse before turning on the instrument. Do not connect the power cord before replacing the fuse. Find out the reason why the fuse burned out before replacing the fuse. Ground the instrument The instrument is grounded through the protective terra conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to the earth. Make sure that the instrument is properly grounded before any inputs or outputs. Observe all terminal ratings To avoid fire or electric shock, please observe all ratings and symbols on the instrument. Read this guide carefully to know more details about the ratings before connection. Keep proper ventilation Inadequate ventilation may cause an increase of temperature,, which will lead to further damage. Please keep proper ventilation and check the fan and air-vents regularly when using the instrument.
Operate condition Location: indoor, no strong light, almost no Interfering pollution; Comparative humidity: <80% Altitude: <2000m Temperature: 0 to 40
II SPD3000 User Manual
SIGLENT
Do not operate in an explosive atmosphere To avoid personal injury or damage to instrument, please do not operate in an explosive atmosphere. Keep surface of the product clean and dry To avoid dust or moisture in the air influence the performance of the instrument , please keep surface of the product clean and dry.
SPD3000 User Manual
III
SIGLENT
Safety Terms and Symbols
Terms may appear on the product: DANGER: Indicates direct injury or hazard that may happen. WARNING: Indicates potential injury or hazard that may happen. CAUTION: Indicates potential damage to the instrument or other property that may happen.
Symbols may appear on the product:
Hazardous Vlotage
Protective Earth Ground
Warning
Earth Ground
Power Switch
IV SPD3000 User Manual
SIGLENT
SPD3000 Series Brief Introduction
SPD3000 series Programmable DC Power Supply is convenient, flexible and multi-function. It has three independent outputs, two sets of adjustable voltage value and a fixed set of selectable voltage value of 2.5V, 3.3V, and 5V ,and it also provides output short circuit and overload protection at the same time.
Main features of SPD3000 Series
4.3 TFT color LCD display with content of 16M; Three independent outputs, two of which are adjustable, and the total power up to 195W; Four kinds of input voltage values includes 100V, 110V, 220V and 230V to satisfy different requirements; Function of storage and call setting parameters; Function of timing output ; Function of waveform display, Real-time display of voltage or current waveforms with digital display of voltage, current and power values Start up screen protective procedure every fixed interval(30 minutes) Perfect PC software to realize the real-time control through USBTMC..
SPD3000 User Manual
SIGLENT
Content
General Safety Summary ....................................................................................................II SPD3000 Series Brief Introduction..................................................................................... V Chapter 1 Start Guide ..........................................................................................................1 The front panel ..........................................................................................................................2 The rear panel ...........................................................................................................................6 Notice in first use of SPD3000 Series .................................................................................7 Input power requirement ...........................................................................................................7 Electrical check .........................................................................................................................7 Output check .............................................................................................................................8 Chapter 2 Control panel operation ......................................................................................9 2.1 Output summary ............................................................................................................10 2.2 CH1/CH2 Independent Output.....................................................................................11 2.3 CH3 Independent mode................................................................................................12 2.4 CH1/CH2 Series mode .................................................................................................13 2.5 CH1/CH2 Parallel mode ...............................................................................................15 2.7 Timer ................................................................................................................................17 2.9 Version information ........................................................................................................20 2.10 Upgrade firmware ........................................................................................................21 Chapter 3 Remote control .................................................................................................23 3.1 Command list..................................................................................................................23 3.2 Command description ...................................................................................................24 Chapter 4 Specification......................................................................................................30 Chapter 5 Troubleshooting ................................................................................................32 Chapter 6 Service and Support .........................................................................................33 Maintain summary ................................................................................................................33 Contact SIGLENT .................................................................................................................33 Appendix A: Accessories ........................................................................................................34
VI SPD3000 User Manual
SIGLENT
Chapter 1 Start Guide
In this chapter, we mainly introduce the panel and Display Interface of SPD3000, and also the notes about how to check and operate it at the first use. You can quickly understand how to operate it after reading the following steps. Please check the instrument according to the following steps. 1. The machine and packing inspection If the packing or cushioning material is seriously damaged, please retained them for follow-up inspection. If the instrument is damaged during shipment, the compensation will be provided by consigner or carrier, and SIGLENT will not or replacement. If there is a mechanical damage or loss please contact with your sales representative. [Link] examination Information about the accessories is introduced in detail at the end of the manual, you can refer to this description to check whether the accessories are completed t. If there are accessories damaged or lacked, please contact with your SIGLENT sales representative. undertake free repair
SPD3000 User Manual
SIGLENT
The front panel
6 7 8 9
10
11
12
13
14
15
NO. 1 2 3 4 5 6 7 8
Description Logo Display Area Model System parameter configuration button Multi-function knob Fine Adjust button Right/Left Direction button Channel Control button
NO. 9 10 11 12 13 14 15
Description CH3 DIP Switch Power Switch CH1 Output Terminal Ground Terminal CH2 Output Terminal CV/CC indicator light CH3 Output Terminal
2 SPD3000 User Manual
SIGLENT
Instruction for buttons
Buttons for setting parameters
WAVEDISP
: Press the button to turn on/off the waveform display interface;
SER
: Press the button to set series mode of CH1/CH2, and the logo displays at the same time; :Press the butten to set parallel mode of CH1/CH2, and the logo displays at the same time; :Press the butten to enter the storage function interface
PARA
RECALL/SAVE
TIMER
:Press the butten to enter timer setting interface : Press the button longer to lock it and shorter to turn to system information interface.
LOCK/VER
Buttons for controlling the channel
ALL ON/OFF
: Press the button to turn on/off all the channels;
CH1 CH2 ON/OFF
: Press the button to select CH1 as the current channel; : Press the button to select CH2 as the current channel; : Press the button to turn on/off the CH1 output; : Press the button to turn on/off the CH3 output.
CH3 ON/OFF
Other buttons
FINE
: Press the button to open the fine tuning function and change the parameters with the minimum step; : To move the cursor around.
SPD3000 User Manual
SIGLENT
The output terminal on front panel
The output terminals of CH1CH2 and CH3 include positive terminal and negative terminal, and a common ground for CH1 and CH2 additional. Each channel has its own logo. The operation details are introduced in the follow-up control panel operation.
4 SPD3000 User Manual
SIGLENT
User interface
1 2 3 4 5
10 6 7
9 1
Parallel/Series logo: The logo will be displayed when the corresponding mode is on
2 3
Channel logo; Operating mode logo: The corresponding logo will display when working in CV or CC mode;
4 5 6 7 8 9 10
LOCK logo: It will display when the LOCK is turned on USB logo: It will display when there is a USB connection; Voltage value; Current value; Power value; Timer lD: Timer state identification Channel on/off logo;
SPD3000 User Manual
SIGLENT
The rear panel
Description:
1 2 3 4 5 6 7
warning message The DIP switch of the AC power and its identification . The description of the AC input voltage AC power socket The fan air vents CE certification mark USB interface and identification
6 SPD3000 User Manual
SIGLENT
Notice in first use of SPD3000 Series
To ensure that the instrument can work normally,, you need to undertake the necessary inspection before using the SPD3000.
Input power requirement
The SPD3000 series allows two kinds of frequency that are 50hz and 60hz, four kinds of AC power that are 100v110v220v and 230v. You can choose the different input power with the DIP switch at the rear panel according to your actual requirement. Warning Pease disconnect the power cords at first and then dial the code to the corresponding gear if you want to change a new power supply.
Electrical check
Please use power cord provided as accessories and connect the instrument to AC power first, and then starting power check according to the following steps. 1. Connect the power supply Warning To avoid electrical shock, please make sure that the instrument has been properly earthing. 2. Turn on the power switch Press the button POWER to enter boot interface, and the default setup will show automatically after a while.
SPD3000 User Manual
SIGLENT
Output check
The output check includes voltage check in condition of all channels with no load and current check in condition of short circuit so as to make sure that the instrument correctly responds to operation of the front panel.
1. Voltage output check (1) Within no load, turn on the power, and make sure the setting current values of all channels are not zero; (2) Check the voltage output of CH1/CH2 Turn on CH1/CH2 and the instrument work in CV mode. Check whether the voltage value could be changed from 0V to 32V.
2. Current output check (1) Turn on the power. (2) Check the current output of CH1/CH2 Use an insulated wire to connect the positive and negative terminal of CH1/CH2; Turn off CH1 and CH2; Revolve the knob to set the voltage value to 32V; Revolve the knob to set the current value to 0A; Adjust the current parameters to check whether the current value can be changed from 0A to 3.2A.
8 SPD3000 User Manual
SIGLENT
Chapter 2 Control panel operation
In this chapterthe function and operation of SPD3000 series control panel will be introduced in detail to giving you an all-around understanding of it, which will eventually lead to easier work.
Brief introduction:
Output summary CH1/CH2 independent output CH3 independent output Parallel output Series output Waveform display Timer Save and recall
SPD3000 User Manual
SIGLENT
2.1 Output summary
Summary
SPD3000 series have three independent outputs, two of which are adjustable in voltage value and the other one includes selectable 2.5V, 3.3V or 5.0V
Independent/Parallel/Series
SPD3000 series have three output modes: independentparallel and series, which can be selected through the track switch on the front panel. In the independent mode, the output current and voltage are controlled respectively. In the parallel mode, the current value is twice that of the single channel. In
the series mode, the voltage value is twice that of the single channel.
Constant voltage/current
In the constant current mode (independent or tracking mode), the current value is rated and controlled through the front panel. The indicator light displays red and the voltage value is under rating. It will return to constant voltage mode when the current value is under dating. In the constant voltage mode, the current value is less than the setting value, and the voltage value is controlled through the front panel. The indicator light displays green and the current value is maintained at the set value. It will return to constant current mode when the voltage value is under dating.
10 SPD3000 User Manual
SIGLENT
2.2 CH1/CH2 Independent Output
Instruction CH1 and CH2 are working in the independent mode, at the same timethey are insulated from the ground.
Output rating
0~30V/0~3A (max:32V, 3.2A)
Operation steps 1. Make sure that parallel/series mode is off. 2. Connect load to the positive and negative terminals of CH1/CH2. 3. Set voltage and current value of CH1/CH2: Firstly, move the cursor to select the wanted parameter(voltage,current), and then revolve the multi-function knob to change the corresponding parameter(you can press FINE to make accurate adjusting). Coarse adjusting: 0.1V or 0.1A each step. Fine adjusting: the least precision each step. 4. Open output Press ON/OFF button to turn on the output, the corresponding indicator light gets lighted immediately and CC or CV is showed on the current interface.
SPD3000 User Manual
11
SIGLENT
2.3 CH3 Independent mode
Instruction CH3 is independent from CH1 and CH2, and it works neither in parallel mode nor in series mode. Its voltage and current ratings are respectively 2.5V,3.3V, 5V and 3A.
Output ratings Operation steps:
2.5V/3.3V/5V, 3A
1. Connect the load to the positive and negative terminals of CH3 on the front panel. 2. Select the wanted voltage value by moving CH3 DIP switch. 3. Open output Press ON/OFF button to turn on the output, the corresponding indicator light gets lighted immediately.
CV
CC
When the current value is higher than 3A, the overload indicator light turns red and the working mode turns to CC from CV Note: overload does not mean abnormal operation.
12 SPD3000 User Manual
SIGLENT
2.4 CH1/CH2 Series mode
Instruction In the series mode, CH1 and CH2 are linked internally into one channel which is controlled by CH1, and the output voltage value is twice compared with that of single channel.
Output rating
0~60V/0~3A (max: 64V,3.2A)
Operation steps 1. Press the SER button to start the Series mode, and the indicator light turns bright immediately; 2. Connect the load to the positive terminal of CH2 and the negative terminal of CH1; 3. Turn on CH2, and then adjust its current value by revolving the multi-function button to 3A. In the default setup, the instrument works in Coarse, you can turn to Fine by pressing FINE button; Coarse: 0.1V or 0.1A each step; Fine: the least precision each step; 4. Turn on CH1, and then set the output voltage/current value by revolving the multi-function knob; 5. Press ON/OFF button to open the output. Note: You can identify the current working state CC or CV by referring to indicator light of CH1/CH2.(red means CV, green means CC).
SPD3000 User Manual
13
SIGLENT
14 SPD3000 User Manual
SIGLENT
2.5 CH1/CH2 Parallel mode
Instruction In the parallel mode, CH1 and CH2 are linked internally into one channel which is controlled by CH1. Its output current value is twice as much as the single channel.
Output rating
0~30V/0~6A
Operation steps: 1. Press the PAR button to start Parallel mode, and the indicator light turns bright immediately 2. Connect the load to the positive and negative terminal of CH1 3. Turn on CH1, and then set the output voltage/current value by revolving the multi-function knob 4. Press ON/OFF button to open the output. Note: You can identity the current working state CC or CV by referring to indicator light of CH1/CH2.(red means CV, green means CC) In parallel mode, CH2 only works in CC mode.
SPD3000 User Manual
15
SIGLENT
2.6 Save and Recall
Save setup
Five group setups can be saved in memory. The contents of the setup file including: Independent/series/parallel mode Output voltage/current value Timer setup Operation steps 1. Set the state that you wanted; 2. Press STORE to enter Save/Recall interface; 3. Select FILE CHOICE by revolving the multi-function knob; 4. Select OPEN CHOICE by moving the cursor; 5. Move the cursor to STORE, and then press it to save the current setup.
Recall setup
Operation steps: 1. Press STORE to enter Save/Recall interface; 2. Select FILE CHOICE by revolving the multi-function knob; 3. Select OPEN CHOICE by moving the cursor; 4. Move the cursor to RECALL, and then press it to read the saved setup.
Note:
if you want to delete the file that has been saved, please go on with step 3 above, select DELETE and then press it.
16 SPD3000 User Manual
SIGLENT
2.7 Timer
The timer works in the Independent mode, and can save five timing setups, each of which is independent from each other. You can set any voltage/current value within the range as you want. The timer supports consecutive output, and the longest time of each group is 10000s. Setup steps: 1. Press CH1/CH2 to select the wanted channel; 2. Press TIMER to enter the Timer Setup interface, and the indicator light turns bright immediately; 3. Move the cursor to select the wanted parameter(voltage/current/time) by pressing the direction button; 4. Revolve the multi-function knob to set the corresponding value and then press it. 5. Press Timer again to exit current interface.
SPD3000 User Manual
17
SIGLENT
Start the Timer 1. Move cursor to TIMER menu by pressing the direction button; 2. Revolve the multi-function knob to make theTimer state to ON; 3. Press the knob to start theTimer.
Note:
If press the TIMER button when the timer is running, then you can observe the decreasing of the time ,and the changes of the curve .The Timer will automatically turn off when the time reduces to 0.
18 SPD3000 User Manual
SIGLENT
2.8 Waveform display
Display the changing of current voltage and current in the form of curve. Operation steps: 1. Select CH1/CH2, and then set voltage/current parameter; 2. Press WAVE button to enter Waveform Display interface, and the indicator light turns bright immediately. 3. Press CH1/CH2 ON/OFF button to turn on CH1/CH2 channel, the corresponding indicator light turns bright immediately and you can observe the real-time changing of current voltage/current. Waveform interface:
Note:
Yellow line means voltage, green line means current, and axis of ordinate means voltage or current value (0~30V/0~3A).
SPD3000 User Manual
19
SIGLENT
2.9 Version information
Press LOCK button quickly to enter Version Information interface, which is shown below:
20 SPD3000 User Manual
SIGLENT
2.10 Upgrade firmware
The software of the instrument is upgraded with a fixed name file via PC management software with USBTMC. The upgrade method is below: Upgrade in normal Interface 1. Open the EsayPower software after USB line having been connected perfectly, and make sure the instrument is connected with the software correctly. 2. Click Version and then choose Upgrade in the drop-down menu to enter the USB firmware upgrade dialogue. See figure 1:
Figure 1
3. Figure 2 shows the firmware upgrade dialogue. Click file choosing icon and then select the file to be upgraded which must have a postfix ugf.
Figure 2 SPD3000 User Manual 21
SIGLENT
Figure 3
4. As it shows in figure 4, click Upgrade button to start upgrading. The upgrade is finished when the progress bar complete and the instrument will run the version after upgrade.
Figure 4
Upgrade Via Guide Procedure Upgrade via guide procedure also can be used if the method above is failed. Specific steps are below: 1. Press the knob before powering on the instrument, and the turn on the instrument and it will enter the guide procedure mode. 2. After enter the guide procedure mode, the upgrade method is the same as the first one.
22 SPD3000 User Manual
SIGLENT
Chapter 3 Remote control
The SCPI Commands are used to perform remote control through USBTMC. Connect the USB device on the rear panel to computer which has been installed with EasyPower software or NIMeasurement & Automation.
3.1 Command list
1. *IDN? 2. *SAV 3. *RCL 4. INSTrument {CH1|CH2} 5. INSTrument ? 6. MEASure:CURRent? 7. MEAsure:VOLTage? 8. MEASure:POWEr? 9. [SOURce:]CURRent <current> 10. [SOURce:]CURRent ? 11. [SOURce:]VOLTage <volt> 12. [SOURce:] VOLTage? 13. OUTPut 14. OUTPut:TRACk 15. OUTPut:WAVE 16. TIMEr:SET 17. TIMEr:SET? 18. TIMEr 19. SYSTem:ERRor? 20. SYSTem:VERSion? 21. SYSTem: STATus?
SPD3000 User Manual
23
SIGLENT
3.2 Command description
1.*IDN? Command format Description *IDN? Query the manufacturer, product type, series NO. and software version. Return Info Example Manufacturer, product type, series NO., software version. Siglent, SPD3303S, SPD00001130025, [Link].
2.*SAV Command format Description *SAV <name> Save current state in nonvolatile memory with the specified name. Example *SAV 1
3.*RCL Command format Description *RCL <name> Recall state that had been saved from nonvolatile memory. Example *RCL 1
[Link] Command format Description Example INSTrument <CH1|CH2> Select the channel that will be operated. INSTrument CH1
24 SPD3000 User Manual
SIGLENT
Command format Description Example Return Info
INSTrument? Query the current operating channel INSTrument? CH1
[Link] Command format Description MEASure: CURRent? < CH1|CH2> Query current value for specified channel, if there is no specified channel, query the current channel. Example Return Info MEASure: CURRent? CH1 3.000
Command format Description
MEASure: VOLTage? < CH1|CH2> Query voltage value for specified channel, if there is no specified channel, query the current channel.
Example Return Info
MEASure: VOLTage? CH1 30.000
Command format Description
MEASure: POWEr? < CH1|CH2> Query power value for specified channel, if there is no specified channel, query the current channel.
Example Return Info
MEASure: POWEr? CH1 90.000
SPD3000 User Manual
25
SIGLENT
[Link]
Command format
<SOURce:>CURRent <value> <SOURce>:={CH1 | CH2}
Description Example
Set current value of the selected channel CH1:CURRent 0.5
Command format
<SOURce>: CURRent? <SOURce>:={CH1 | CH2}
Description Example Return Info
Query the current value of the selected channel. CH1: CURRent? 0.500
Command format
<SOURce>: VOLTage <value> <SOURce>:={CH1 | CH2}
Description Example
Set voltage value of the selected channel CH1: VOLTage 25
Command format
<SOURce>:CURRent? <SOURce>:={CH1 | CH2}
Description Example Return Info
Query the voltage value of the selected channel. CH1: VOLTage? 25.000
26 SPD3000 User Manual
SIGLENT
[Link]
Command format
OUTPut <SOURce>, <state> <SOURce>:={CH1 | CH2}; <state>:={ON | OFF}
Description Example
Turn on/off the specified channel. OUTPut CH1, ON
Command format
OUTPut: TRACK <NR1> <NR1>:={0[Independent] | 1[Series] | 2[Parallel]}
Description Example
Select operation mode OUTPut: TRACK 0
Command format
OUTPut: WAVE<SOURce>, <state> <SOURce>:={CH1 | CH2}; <state>:={ON | OFF}
Description
Turn on/off the Waveform Display function of specified channel.
Example
OUTPut: WAVE CH1, ON
[Link]
Command format
TIMEr: SET <SOURce>, <secnum>, <volt>, <curr>, <time> <SOURce>:={CH1 | CH2}; < secnum >;=1 to 5;
Description Example
Set timing parameters of specified channel TIMEr: SET CH1, 2, 3, 0.5, 2
SPD3000 User Manual 27
SIGLENT
Command format
TIMEr: SET? <SOURce>, <secnum> <SOURce>:={CH1 | CH2}; < secnum >;=1 to 5;
Description
Query the voltage/current/time parameters of specified group of specified channel.
Example Return Info
TIMEr: SET? CH1, 2 3, 0.5, 2
Command format
TIMEr <SOURce>, <state> <SOURce>:={CH1 | CH2}; < state >;={ON | OFF};
Description Instruction
Turn on/off Timer function of specified channel The command works effectivly only when <secnum> starts from 1.
Example
TIMEr CH1, ON
[Link]
Command format Description
SYSTem: ERRor? Query the error code and the information of the equipment.
Command format Description Example Return Info
SYSTem: VERSion? Query the software version of the equipment. SYSTem: VERSion? [Link]
28 SPD3000 User Manual
SIGLENT
Command format Description Instruction
SYSTem: STATus? Query the current working state of the equipment. The return info is Hexadecimal format, but the actual state is binary , so you must change the return info into a binary. The state correspondence relationship is as follow.
Example Return info
SYSTem: STATus? 0x0224
Bit NO. 0 1 2,3
Corresponding state 0: CH1 CV mode; 0: CH2 CV mode; 1: CH1 CC mode 1: CH2 CC mode 10: Parallel mode
01: Independent mode; 11: Series mode
4 5 6 7 8 9
0: CH1 OFF 0: CH2 OFF
1: CH1 ON 1: CH2 ON 1: TIMER1 ON 1: TIMER2 ON 1: CH1 waveform diplay 1: CH2 waveform diplay
0: TIMER1 OFF 0: TIMER2 OFF
0: CH1 digital display; 0: CH2 digital display;
SPD3000 User Manual
29
SIGLENT
Chapter 4 Specification
Test condition:Heat for 30minitus, temperature between +20~+30.
Output rating CH1/CH2 independent CH1/CH2 series CH1/CH2 parallel CH3 030V , 03A 060V , 03A 030V , 06A 2.5V/3.3V/5.0V , 03A 0.01%+3mV 0.01%+3mV(rating current 3A) 0.02%+5mV(rating current 3A) 2mVrms5Hz 1MHz 100s(50% load change ,minimum load 0.5A) 300ppm/ 0.2%+3mA 0.2%+3mA 3mArms 5mV 15mV 2mVrms5Hz 1MHz 0.5%+10mV of Master(No Load) Line 0.01%+3mV Load 0.01%+3mV(rating current3A) 0.02%+5mV(rating current3A) Line 0.01%+5mV Load 300mV Voltage: 1mV Current: 1mA 3.2A full scale, display 4 digits 0.4 LED
Constant Voltage coefficient Voltage Mode Load coefficient Ripple and noise Recover time Temperature coefficient Constant Power coefficient Current Mode Load coefficient Ripple and noise CH3 Power coefficient Load coefficient Ripple and noise Tracking operation Track error Parallel coefficient
Series coefficient Resolution Precision Voltage Current Ammeter Voltmeter Set precision Read precision Insulation Between base and terminal Between base and AC power cord
32V full scale, 5 digits 0.4 LED display Voltage 0.03% of reading + 10mV Current (0.3% of reading + 10mA) Voltage 0.03% of reading + 10mV Current (0.3% of reading + 10mA) 20Mor above (DC 500V) 30Mor above (DC 500V)
30 SPD3000 User Manual
SIGLENT Operating environment Indoor Altitude Environment temperature Comparative humidity Installation level Pollution level Environment temperature Comparative humidity
2000 m 0 ~ 40 80% 2 -10 ~ 70 70%
Storage environment
Power supply AC 100V/120V/220V/230V10%, 50/60HZ Volume 275mm x 225mm x 136mm Weight 8kg
SPD3000 User Manual
31
SIGLENT
Chapter 5 Troubleshooting
Question 1: what to do if there occurs a short circuit on output terminal? Answer1: There are over current protection and short circuit protection inside the power, so current will be controlled in safety range. Question 2: Is it abnormal that the CH3 overload indicator is lit ? Answer 2: No, that only means the current value reaches 3A, which is the maximum value within its range, and now power supply can be used continuously with the suggestion that decrease the output load. Question 3: Is it normal that in the series mode, voltage and current value of a channel is respective 0V and not 0A, while that of another channel is both not 0? Answer 3: Yes, because when current output load is higher than the limited current value, the working mode turns to CC from CV. Question 4: How to deal with the upgrade failed? Answer 4: please do it again. Question 5: why the practical value is not the same with the performance standard) and even changs slowly on startup? Answer 5: Its normal. On startup, components inside the instrument are in the process of getting stable, readings will be stable about 30 minutes later. set value (over
32 SPD3000 User Manual
SIGLENT
Chapter 6 Service and Support
Maintain summary
SIGLENT warrants that the products that it manufactures and sells will be free from defects in materials and workmanship for a period of three years from the date of shipment from an authorized SIGLENT distributor. If a product or CRT proves defective within the respective period, SIGLENT will provide repair or replacement as described in the complete warranty statement. To arrange for service or obtain a copy of the complete warranty statement, please contact your nearest SIGLENT sales and service office. Except as provided in this summary or the applicable warranty Statement, SIGLENT makes no warranty of any kind, express or implied, including without limitation the implied warranties of merchantability and fitness for a particular purpose. In no Event shall SIGLENT be liable for indirect, special or Consequential damages
Contact SIGLENT
MTR Add: 3/F, Building 4, Antongda Industrial Zone, 3rd Liuxian Road, 68 District, Baoan District, Shenzhen, P.R. CHINA Tel: +86-755-36615186 E-mail:sales@[Link] [Link]
SPD3000 User Manual
33
SIGLENT
Appendix A: Accessories
SPD3000 Series Programmable DC Power Supply Accessories: Standard Accessories: A User Manual A Certification A Guaranty Card A CD(including EasyPower computer software system) A Power Cord that fits the standard of destination country A USB Cable
34 SPD3000 User Manual