This document describes a method for determining filterable residue in water and wastewater samples. The method involves filtering a sample, drying the filtrate in an evaporating dish at 180°C, and weighing the residue. Filterable residue is defined as solids capable of passing through a glass fiber filter. Drying is continued until a constant weight is obtained. The amount of filterable residue is calculated based on the weight of dried residue and sample volume filtered.
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
573 views3 pages

EPA Method 160 1

This document describes a method for determining filterable residue in water and wastewater samples. The method involves filtering a sample, drying the filtrate in an evaporating dish at 180°C, and weighing the residue. Filterable residue is defined as solids capable of passing through a glass fiber filter. Drying is continued until a constant weight is obtained. The amount of filterable residue is calculated based on the weight of dried residue and sample volume filtered.
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

METHOD #: 160.1 TITLE: ANALYTE: INSTRUMENTATION: STORET No.

Approved for NPDES (Issued 1971) Residue, Filterable (Gravimetric, Dried at 180C) Residue, Filterable Drying Oven 70300

1.0

Scope and Application 1.1 1.2 This method is applicable to drinking, surface, and saline waters, domestic and industrial wastes. The practical range of the determination is 10 mg/L to 20,000 mg/L

2.0

Summary of Method 2.1 2.2 A well-mixed sample is filtered through a standard glass fiber filter. The filtrate is evaporated and dried to constant weight at 180C. If Residue, Non-Filterable is being determined, the filtrate from that method may be used for Residue, Filterable.

3.0

Definitions 3.1 Filterable residue is defined as those solids capable of passing through a glass fiber filter and dried to constant weight at 180C.

4.0

Sample Handling and Preservation 4.1 Preservation of the sample is not practical; analysis should begin as possible. Refrigeration or icing to 4C, to minimize micro-biological decomposition of solids, is recommended.

5.0

Interferences 5.1 Highly mineralized waters containing significant concentrations of calcium, magnesium, chloride and/or sulfate may be hygroscopic and will require prolonged drying, desiccation and rapid weighing. Samples containing high concentrations of bicarbonate will require careful and possibly prolonged drying at 180C to insure that all the bicarbonate is converted to carbonate. Too much residue in the evaporating dish will crust over and entrap water that will not be driven off during drying. Total residue should be limited to about 200 mg.

5.2

5.3

6.0

Apparatus

6.1 6.2 6.3 6.4 6.5 6.6 6.7 6.8 6.9 7.0

Glass fiber filter discs, 4.7 cm or 2.1 cm, without organic binder, Reeve Angel type 934-AH, Gelman type A/E, or equivalent Filter holder, membrane filter funnel or Gooch crucible adapter Suction flask, 500 mL Gooch crucibles, 25 mL (if 2.1 cm filter is used) Evaporating dishes, porcelain, 100 mL volume. (Vycor or platinum dishes may be substituted) Steam bath Drying oven, 180C 2C Desiccator Analytical balance, capable of weighing to 0.1 mg

Procedure 7.1 Preparation of glass fiber filter disc: Place the disc on the membrane filter apparatus or insert into bottom of a suitable Gooch crucible. While vacuum is applied, wash the disc with three successive 20 mL volumes of distilled water. Remove all traces of water by continuing to apply vacuum after water has passed through. Discard washings. Preparation of evaporating dishes: If Volatile Residue is also to be measured heat the clean dish to 550 50C for one hour in a muffle furnace. If only Filterable Residue is to be measured heat the clean dish to 180 2C for one hour. Cool in desiccator and store until needed. Weigh immediately before use. Assemble the filtering apparatus and begin suction. Shake the sample vigorously and rapidly transfer 100 mL to the funnel by means of a 100 mL graduated cylinder. If total filterable residue is low, a larger volume may be filtered. Filter the sample through the glass fiber filter, rinse with three 10 mL portions of distilled water and continue to apply vacuum for about 3 minutes after filtration is complete to remove as much water as possible. Transfer 100 mL (or a larger volume) of the filtrate to a weighed evaporating dish and evaporate to dryness on a steam bath. Dry the evaporated sample for at least one hour at 180 plus or minus 2C. Cool in a desiccator and weigh. Repeat the drying cycle until a constant weight is obtained or until weight loss is less than 0.5 mg.

7.2

7.3

7.4

7.5 7.6

8.0

Calculation 8.1 Calculate filterable residue as follows:

where: A = weight of dried residue + dish in mg B = weight of dish in mg C = volume of sample used in mL

9.0

Precision and Accuracy 9.1 Precision and accuracy are not available at this time.

Bibliography 1. Standard Methods for the Examination of Water and Wastewater, 14th Edition, p 92, Method 208B, (1975).