Data sheet
MX1615
Bridge/strain gauge
amplifier
Special features
16 individually configurable
inputs
Connection of strain gauges in
quarter- , half- and full-bridge
Bridge excitation : DC or carrier
frequency
Internal shunt resistor
Connection of standard voltage,
PT100, resistor
Data rate up to 19200 Hz per
channel
3000 Hz bandwidth
24bit A/D converter per channel
for synchronous, parallel
measurements
Active low pass filter
TEDS support
Digital platform
Signal
A/D
Bridge
excitation
TEDS
Electrically isolated
Ethernet
Connectors 1.16
Connector sockets (Plug terminal, 8 Pin)
Block diagram
IEEE1394b
FireWire
IEEE1394b
FireWire
Supply
voltage
B3425-3.0 en
Specifications MX1615
General specifications
Inputs
Number
16, electrically isolated from the supply
Strain gauges in full- , half- or quarter-bridge configuration.
Selectable bridge excitation voltage : DC voltage or carrier
frequency (AC/CF, 1200 Hz)
Transducer technologies, can be adjusted individually
SG-quarter bridges
SG-half bridges
SG-full bridges
Three wire and four wire
five wire
six wire
Resistor, Resistance thermometer (PT100), connection in
four-wire configuration
Voltage ("10 V differential, 0 30 V unipolar) without
transducer supply
24 Bit Delta Sigma converter
A/D converter per channel
Data rate per channel, max.
Hz
0.1 19,200, can be individually adjusted per channel
Bandwidth
kHz
Active low-pass filter (Bessel/Butterworth, can be switched
off)
Hz
3.0
0.4 using carrier frequency
0.01 3,000
Transducer identification (TEDS, IEEE 1451.4)
max. distance of the TEDS module
100
Phnix Contact FMC1,5/8ST3,5RF plug terminal
Plug included
10 ... 30 (24 V nominal (rated) voltage)
Transducer connection
Supply voltage range (DC)
max. 5 ms at 24 V
Supply voltage interruption
Power consumption
Ethernet (data link)
Protocol/addressing
Connection
Max. cable length to module
Synchronization options
EtherCAT1)
NTP
IRIGB (B000 to B007; B120 to B127)
IEEE1394b FireWire (module synchronization, data link,
optional supply voltage)
Baud rate
Max. current from module to module
Max. cable length between the nodes
Max. number of modules connected in series (daisy chain)
Max. number of modules in a IEEE1394b FireWire system
(including hubs2), backplane)
Max. number of hops3)
Nominal (rated) temperature range
< 12
10Base-T / 100Base-TX
TCP/IP (direct IP address or DHCP)
8P8C plug (RJ-45) with twisted pair cable (CAT-5)
100
IEEE1394b FireWire (only QuantumX, automatically,
recommended)
via CX27
via Ethernet
via MX440A - or MX840A input channel
IEEE 1394b (HBM modules only)
MBaud
A
m
400 (approx. 50 MByte/s)
1.5
5
12 (=11 hops)
C [F]
24
14
20 .... +60 [4 ... +140]
Operating temperature range
C [F]
20 ... +65 [4 ... +149]
Storage temperature range
C [F]
40 ... +75 [40 ... +167]
5 ... 95 (non condensing)
Rel. humidity
III
Protection class
IP20 per EN 60529
Degree of protection
Mechanical tests4)
Vibration (30 min)
Shock (6 ms)
EMC requirements
Max. input voltage at transducer socket to ground,
transient free
Pin 6 and 7 to Pin 1, 2, 3, 4 or 5
Dimensions, horizontal (W x H x D)
m/s2
m/s2
50
350
per EN 613261
"18
mm
52.5 x 200 x 122 (with case protection)
44 x 174 x 119 (without case protection)
980
Weight, approx.
1) EtherCAT is a registered trademark and patented technology, licensed by Beckhoff Automation GmbH, Germany
2) Hub: IEEE1394b FireWire node or distributor
3) Hop: Transition from module to module or signal conditioning / distribution via IEEE1394b FireWire (hub, backplane)
4) Mechanical stress is tested according to European Standard EN6006826 for vibrations and EN60068227 for shock. The equipment is subjected to an
acceleration of 50 m/s2 in a frequency range of 5...65 Hz in all 3 axes. Duration of this vibration test: 30min per axis. The shock test is performed with a nominal
acceleration of 350 m/s2 for 6 ms, half sine pulse shape, with 3 shocks in each of the 6 possible directions.
HBM
B3425-3.0 en
Specifications MX1615 (Continued)
4 mV/V CF strain gauge full and half bridge with 1 V; 2.5 V or 5 V excitation (AC, square)
0.05 1)
Accuracy class
Carrier frequency (square)
Hz
1,200 "2
Bridge excitation voltage (effective)
1; 2.5; 5 ("5 %)
strain gauge full and half bridges
Transducers that can be connected
Permissible cable length between MX1615
and transducer
< 100
Measuring ranges
at 5 V excitation
at 2.5 V excitation
at 1 V excitation
mV/V
mV/V
mV/V
"4
"8
"20
Control signal (Shunt)
Full bridge
Half bridge
mV/V
mV/V
1.066 "0.3% (at 350 Ohm)
+1.008 "0.3% (at 350 Ohm) 2)
Measurement frequency range (3 dB)
Hz
0 ... 400
Transducer impedance
at 5 V excitation
at 2.5 V excitation
at 1 V excitation
W
W
W
300 ... 1,000
110 ... 1,000
80 1,000
mV/V
mV/V
mV/V
< 0.2
< 0.5
< 1.5
Noise at 25 5C and 2.5 V excitation (peak to peak)
with filter 1 Hz Bessel
with filter 10 Hz Bessel
with filter 100 Hz Bessel
< 0.02 of full scale
Zero drift (Full bridge with 5 V excitation)
% / 10 K
< 0.02 1) of full scale
Full-scale drift (5 V excitation)
% / 10 K
< 0.05 of measurement value
Linearity error
1)
2)
0.1 with half bridge
When using a half bridge, the control signal may only be used when signals 1 (Pin 6) and 4 (Pin 7) are bridged.
4 mV/V CF strain gauge full and half bridge with1 V; 2.5 V or 5 V excitation (DC)
0.05 1)
Accuracy class
Bridge excitation voltage (DC)
<
strain gauge half and full bridges
Transducers that can be connected
Permissible cable length between MX1615
and transducer
1; 2.5; 5; ("5 %)
< 100
Measuring ranges
at 5 V excitation
at 2.5 V excitation
at 1 V excitation
mV/V
mV/V
mV/V
"4
"8
"20
Control signal (Shunt)
Full bridge
Half bridge
mV/V
mV/V
1.066 "0.3% (at 350 Ohm)
+1.008 "0.3% (at 350 Ohm) 2)
Measurement frequency range (3 dB)
Hz
0 ... 3,000
Transducer impedance
at 5 V excitation
at 2.5 V excitation
at 1 V excitation
at 0.5 V excitation
W
W
W
W
300 ... 1,000
110 ... 1,000
80 ... 1,000
80 ... 1,000
mV/V
mV/V
mV/V
mV/V
< 0.2
< 0.4
<1
<3
Noise at 25 5C and 2.5 V excitation (peak to peak)
with filter 1 Hz Bessel
with filter 10 Hz Bessel
with filter 100 Hz Bessel
with filter 1 kHz Bessel
< 0.02 of full scale
Zero drift (Full bridge with 5 V excitation)
% / 10 K
< 0.05 1) of full scale
Full-scale drift (5 V excitation)
% / 10 K
< 0.05 of measurement value
Linearity error
1)
with half bridge : 0.1; with 10 V/m electromagnetic field strength (EN6100043) : 0.2
2) When using a half bridge, the control signal may only be used when signals 1 (Pin 6) and 4 (Pin 7) are bridged.
B3425-3.0 en
HBM
Specifications MX1615 (Continued)
5 mV/V CF strain gauge single bridge with 0.5 V; 1 V; 2.5 V or 5 V excitation (AC, square)
0.1
Accuracy class
Carrier frequency (square)
Hz
1200"2
Bridge excitation voltage (effective)
0.5; 1; 2.5; 5 ("5 %)
SG quarter bridge in four wire circuit
Transducers that can be connected
Permissible cable length between MX1615 and
transducer
< 100
Measuring ranges
at 5 V excitation (only at 350 Ohm strain gauge)
at 2.5 V excitation
at 1 V excitation
at 0.5 V excitation
mV/V
mV/V
mV/V
mV/V
"4
"8
"20
"40
Control signal (Shunt)
mV/V
1.008 "0,1 % (at 350 Ohm)
Measurement frequency range (3 dB)
Hz
0 ... 400
Internal completion resistors
120 and 350
mV/V
mV/V
mV/V
< 0.3
< 0.6
< 1.5
< 0.05 of full scale
Noise at 25 5C and 5 V excitation (peak to peak)
with filter 1 Hz Bessel
with filter 10 Hz Bessel
with filter 100 Hz Bessel
Linearity error1)
1)
Zero drift1) (5 V excitation)
% / 10 K
< 0.1 of full scale
Full-scale1) drift (5 V excitation)
% / 10 K
< 0.1 of measurement value
With 350 ohm resistor
5 mV/V CF strain gauge single bridge with 0.5 V; 1 V; 2.5 V or 5 V excitation (DC)
0.11)
Accuracy class
Bridge excitation voltage (DC)
SG quarter bridges in four wire or three wire circuit
Transducers that can be connected
Permissible cable length between MX1615
and transducer
0.5; 1; 2.5; 5 ("5 %)
< 100
Measuring ranges
at 5 V excitation (only at 350 Ohm strain gauge)
at 2.5 V excitation
at 1 V excitation
at 0.5 V excitation
mV/V
mV/V
mV/V
mV/V
"4
"8
"20
"40
Control signal (Shunt)
mV/V
1.008 "0,1 % (at 350 Ohm)
Measurement frequency range (3 dB)
Hz
0 ... 3,000
Internal completion resistors
120 and 350
mV/V
mV/V
mV/V
mV/V
< 0.4
< 0.6
< 1.5
<3
Noise at 25 5C and 5 V excitation (peak to peak)
with filter 1 Hz Bessel
with filter 10 Hz Bessel
with filter 100 Hz Bessel
with filter 1 kHz Bessel
Linearity error2)
< 0.05 of full scale
Zero drift2) (5 V excitation)
% / 10 K
<0.1 of full scale
Full-scale2)
% / 10 K
< 0.1 of measurement value
drift (5 V excitation)
1)
with 10 V/m electromagnetic field strength (EN6100043) : 0.2
The accuracy class does not take into account measurement errors resulting from asymmetrical cable resistances when using a three-wire
circuit.
2) With 350 ohm resistor and connection using a four-wire circuit.
HBM
B3425-3.0 en
Specifications MX1615 (Continued)
Voltage 10 V (DC)
0.05
Accuracy class
Voltage sources
Transducers that can be connected
Permissible cable length between MX1615 and
transducer
100
Measuring range
V
V
"10 differential,
0 30 unipolar
Measurement frequency range (3 dB)
Hz
0 ... 3,000
< 500
Input impedance (symmetrical)
MW
> 1.5
Noise at 25 5C (peak to peak)
at 1 Hz Bessel filter
at 10 Hz Bessel filter
at 100 Hz Bessel filter
at 1 kHz Bessel filter
mV
mV
mV
mV
150
300
600
2000
Linearity error
< 0.02 of full scale
Common-mode rejection
at DC common-mode
at 50 Hz common-mode, typically
dB
dB
> 100
75
Max. common-mode voltage
Channel from housing and supply ground
Channel from channel
V
V
"60
"5
Internal resistance of the connected voltage source
Zero drift
% / 10 K
< 0.03 of full scale
Full-scale drift
% / 10 K
< 0.05 of measurement value
B3425-3.0 en
HBM
Specifications MX1615 (Continued)
Resistance
0.1
Accuracy class
PTC, NTC, KTY, TT-3, resistances generally
(connection with four wire configuration)
Transducers that can be connected
Permissible cable length between MX1615
and transducer
100
Measuring range
0 ... 1,0001)
Excitation current
mA
0.37 ... 1.43
Measurement frequency range (3 dB)
Hz
0 ... 3,000
Noise at 25 5C (peak to peak)
with filter 1 Hz Bessel
with filter 10 Hz Bessel
with filter 100 Hz Bessel
with filter 1 kHz Bessel
K
K
K
K
< 0.1
< 0.2
< 0.5
< 1.5
Linearity error
< 0.05 of full scale
Zero drift
Full-scale drift
1)
%/10K
< 0.02 of full scale
% / 10 K
< 0.1 of measurement value
Measuring range can be modulated up to 5 kW, in this case: accuracy class 2
Resistance thermometer (PT100)
0.1
Accuracy class
PT100 (connection with four wire configuration)
Transducers that can be connected
Permissible cable length between MX1615
and transducer
100
oC [oF]
200 ... +848 [328 ... +1,558.4]
Excitation current
mA
0.67 ... 1.36
Measurement frequency range (3 dB)
Hz
0 ... 3,000
Noise at 25 5C (peak to peak)
with filter 1 Hz Bessel
with filter 10 Hz Bessel
with filter 100 Hz Bessel
with filter 1 kHz Bessel
K
K
K
K
< 0.02
< 0.04
< 0.1
< 0.3
Linearization range
<"0.3
Zero drift
K / 10 K
< 0.2
Full-scale drift
K / 10 K
< 0.5
Linearity error
HBM
B3425-3.0 en
Active low pass filter data MX1615
(4th order Bessel/Butterworth, with CF supply only valid for fg v100 Hz)
Type
1dB (Hz)
3dB (Hz)
20dB (Hz)
1000
1575
3611
Bessel
Butterworth
Rise time (ms)
Overshoot (%)
Data rate (Hz)
0.11
0.2
1.4
19200
1.4
9600
1000
1575
3612
0.11
0.2
500
812
2079
0.3
0.38
1.3
9600
860
0.9
1.05
0.8
9600
2.11
4.18
0.8
9600
0.8
9600
10.4
0.8
9600
21.0
0.8
9600
41.4
0.8
2400
200
335
100
168
427
1.8
50
213
85
3.8
20
84
33.7
10
16.6
43
9.6
19.5
8.4
21
39
3.4
8,6
97
102
0.8
2400
215
0.8
2400
1.6
4,2
197
0.5
0.84
2,1
390
418
0.8
300
1033
0.8
300
0.2
0.34
0,85
980
0.1
0.17
0,43
1950
2090
0.8
300
0.05
0.085
0,21
3860
4170
0.8
20
0.02
0.036
0,088
9800
10560
0.8
20
0.01
2000
0.017
0,044
19500
21200
0.8
20
3053
5083
0.144
8.5
19200
1000
1170
2077
0.27
0.344
11
19200
1000
1171
2078
0.27
0.378
11
9600
500
587
1048
0.64
0.652
11
9600
200
237
420
1.76
1.64
11
9600
100
118
210
3.65
3.28
11
9600
50
59
105
7.49
6.29
11
9600
20
24
42
18.8
16.15
11
9600
10
12
21
37.7
32.29
11
9600
5.95
10.5
74.9
65.92
11
2400
2.37
4.24
188
163.6
11
2400
1.26
0.59
2.12
370
315
11
2400
1.05
756
656
11
300
1640
11
300
0.5
*)
Phase delay (ms)
0.2
0.241
0.419
1900
0.1
0.122
3770
3280
11
300
0.05
0.060
0.210
0.106
7490
6596
11
20
0.02
0.0245
0.042
18900
16200
11
20
0.01
0.012
0.021
37700
32383
11
20
The analog-to-digital converters delay time is 128 s for all data rates and has not been accounted for in the Phase delay column!
B3425-3.0 en
HBM
Amplitude response of MX1615 Butterworth filter
10 Hz
200 Hz
2 kHz
20 Hz
50 Hz
500 Hz
1 kHz
100 Hz
0.01 Hz
0.1 Hz
0.02 Hz
0.2 Hz
0.5 Hz
HBM
2 Hz
0.5 Hz
1 Hz
5 Hz
B3425-3.0 en
Amplitude response of MX1615 Bessel filter
10 Hz
1 kHz
200 Hz
20 Hz
500 Hz
50 Hz
100 Hz
0.01 Hz
0.02 Hz
0.1 Hz
0.5 Hz
B3425-3.0 en
0.5 Hz
0.2 Hz
2 Hz
1 Hz
5 Hz
HBM
Specifications Power pack NTX001
NTX001
Nominal input voltage (AC)
100 ... 240 ("10%)
Stand-by power consumption at 230 V
0.5
Nominal load
UA
IA
V
A
24
1.25
V
A
mV
24" 4%
0 1.25
v120
Static output characteristics
UA
IA
UBr (Output voltage ripple; peak to peak)
Current limiting, typically from
1.6
galvanically, by optocoupler and converter
Primary secondary separation
Creep distance and clearance
mm
w8
High-voltage test
kV
w4
Ambient temperature range
oC
[F]
0... +40 [+32 ... +104]
Storage temperature
oC
[F]
40 ... +70 [40 ... +158]
Accessories, to be ordered separately
MX1615 accessories
Article
Description
Order No.
AC-DC power supply / 24 V
Input : 100 ... 240 V AC ("10%), 1.5 m cable
Output: 24 V DC, max. 1.25 A, 2 m cable with ODU connector
1NTX001
3m cable QuantumX supply
3 m cable for voltage supply of QuantumX modules;
Suitable plug (ODU Medi-Snap S11M08P04MJGO5280) on
one side and open strands on the other end.
1KAB2713
IEEE1394b FireWire cable (module-to-module)
FireWire connection cable for QuantumX modules; with
matching plugs on both sides.
Lengths 0.2 m/2 m/5 m
Note: The cable enables QuantumX modules to be supplied
with voltage (max. 1.5 A, from the source to the last drain).
1KAB2690.2
1KAB2692
1KAB2695
IEEE1394b FireWire IEEE PCCard
FireWire IEEE 1394b PCCard (PCMCIA adapter) to connect
QuantumX modules to a notebook or PC
1IF001
IEEE1394b FireWire IEEE ExpressCard
FireWire IEEE 1394b ExpressCard (ExpressCard/34) to
connect QuantumX modules to a notebook or PC
1IF002
IEEE1394b FireWire cable PC-to-module
Firewire connection cable from the PC to the first module
for data transfer from QuantumX modules to the PC;
With matching plugs on both sides; Length: 3 m.
1KAB2703
IEEE1394b FireWire cable, hub-to-module, 3 m
FireWire connection cable between HUB and module.
1KAB2753
FireWire Extender SCMFW
Package including 2 in-line elements for extension of the
FireWire connection up to 40 m; required parts: 2 x
1-KAB269-x and Industrial Ethernet cable (M12, CAT5e/6,
max. 30 m). KAB270-3 connection is not possible!
1SCM-FW
Ethernet cross over cable
Ethernet cross-over cable for direct operation of devices on a
PC or notebook, length 2 m, type CAT5+
1KAB2392
Connecting elements for QuantumX modules
Connecting elements (clips) for QuantumX modules; Set
comprising 2 case clips including mounting material for fast
connection of 2 modules.
1CASECLIP
Fitting panel for QuantumX modules
Fitting panel for mounting of QuantumX modules using case
clips (1CASECLIP), lashing strap or cable tie. Basic
fastening by 4 screws.
1CASEFIT
QuantumX backplane (Standard)
QuantumX backplane Standard for a maximum of 9
modules, IP 20 version;
Mounting on wall or control cabinet (19)
- Connection of external modules by FireWire possible;
- Power supply: 24 V DC / max. 5 A (150 W);
1BPX001
Voltage supply
Communication
Mechanical data
HBM
10
B3425-3.0 en
Accessories, to be ordered separately (continued)
MX1615 accessories
Article
Description
Order No.
Push-In connector (8 pins)
10 Push-In-connectors, Phnix Contact, 8 pins
1CONS1005
1WireEEPROMS
Package of 10 pieces of 1-wire EEPROMS DS24B33
(IEEE1451.4 TEDS)
1TEDSPAK
Package including:
- MX1615 amplifier (1-MX1615)
- Power supply (1-NTX001)
- 4 transducer plugs with TEDS (1-SUBHD15-MALE)
- Ethernet Cross-over cable (1-KAB239-2)
- catmanEasy software from HBM (1-CATMAN-EASY)
- Including software maintenance for the first 12 months
1MX1615PAKEASY
Package including:
- MX1615 amplifier (1-MX1615)
- Power supply (1-NTX001)
- 4 transducer plugs with TEDS (1-SUBHD15-MALE)
- Ethernet Cross-over cable (1-KAB239-2)
- catmanAP software from HBM (1-CATMAN-AP)
- Including software maintenance for the first 12 months
QuantumX driver for DIAdem software from National
1MX1615PAKAP
Transducer side
Software and product packages
MX1615 + catmanEASY
MX1615 + catmanAP
DIAdem driver
CANape driver
QuantumX system CD
B3425-3.0 en
Instruments. German user interface only.
QuantumX driver for CANape software from Vector
1DIADEMDRIVER
1CANAPEDRIVER
Informatik. CANape versions from 10.0 are supported.
QuantumX Assistant, LabVIEW driver, programming interface
(.NETAPI), TEDS editor
11
free
HBM
Hottinger Baldwin Messtechnik GmbH
Im Tiefen See 45 64293 Darmstadt Germany
Tel. +49 6151 803-0 Fax: +49 6151 803-9100
Email: info@[Link] [Link]
B34253.0 en
Hottinger Baldwin Messtechnik GmbH.
Subject to modifications. All product descriptions are for
general information only. They are not to be understood as a
guarantee of quality or durability.