ProfessionalPlus
The new generation
in field and lab water
quality instruments
Detachable field probes
and rugged cables mean
less downtime
Connect from field to
lab probes in seconds
Report parameters and
units to fit your needs
Log data and manage
data with desktop
software
USB connection for
software upgrades
ProfessionalPlus
DO, pH,
Conductivity, ORP,
Ammonium, Nitrate,
Chloride, Salinity,
Total Dissolved
Solids, Barometric
Pressure
2,000 data-set
memory with GLP
event logging
Interval or singleevent logging
capability
The rugged design and fieldworthiness of the Professional
Plus instrument allows you to
perform your job with confidence even in the harshest
field conditions and with the
The YSI Professional Plus accommodates
any combination of Pro Series probe, cable,
and accessory. This compact handheld includes
multiple languages, detailed GLP data, on-screen
Help functionality, auto-buffer recognition, milspec connectors, re-cal prompts, auto-stable and
security passwords to help make your job easier.
Finally, a professional instrument for a true
professional.
Rubber, over-molded
IP-67 water proof
case
Possible probe choices
3-year instrument
warranty; 2-year
cable warranty
User- upgradeable
software via USB
Backlit display AND
keypad
Graphic display with
detailed Help
Rugged cables, connectors, and probes
*
**
***
X=
Choice of ISE; pH, ORP, pH/ORP (not recommended on the 1010-X cable)
Choice of ISE; Ammonium, Nitrate, or Chloride for fresh water only
Choice of DO; Galvanic, Polarographic
Lab-grade pH and ORP cable/probe options available
Choice of cable length: 1, 4, 10, 20, 30-meter all (up to 100-meters on 20-X only)
Connect from field to lab probes in seconds
Data Manager
Software
Easy-to-use software to
manage all your data,
set up instruments, or
conduct real-time studies
View graphical and
tabular data
USB connection also
powers the instrument
lowest overall cost of ownership. Interchangeable probes
and cables reduce down-time
and keeps the Pro Plus ready
whenever and wherever you
are - field or lab.
if complexity is the problem...
ease of use is the solution;
productivity is the impact.
SPECIFICATIONS
(SYSTEM W/4-METER CABLE AND PROBES)
Dissolved Oxygen
Sensor Type
Polarographic or Galvanic
(%)
Range
0 to 500%
Accuracy
Resolution
Dissolved Oxygen
(mg/L)
Calibration
Sensor Type
Range
Accuracy
Resolution
Temperature
(Field rugged cables)
Temperature
(Lab-grade)*
Conductivity**
Calibration
Range
Accuracy
Resolution
Units
Range
Accuracy
Resolution
Units
Depth
Sensor Type
Range
Accuracy
Resolution
Salinity
Units
Calibration
Sensor Type
Range
Accuracy
pH
ORP
Resolution
Units
Calibration
Sensor Type
Range
Accuracy
Resolution
Units
Calibration
Sensor Type
Range
Accuracy
Resolution
Units
Calibration
0 to 200% 2% of reading or 2% air
saturation, whichever is greater
200% 500% 6% of reading
1% or 0.1% air saturation
(user selectable)
1 point and zero
Polarographic or Galvanic
0 to 50 mg/L
0 to 20 mg/L (2% of the reading or
0.2 mg/L, whichever is greater)
20 to 50 mg/L (6% of the reading)
0.1 or 0.01 mg/L (user selectable); 0.1%
air saturation
1 point and zero
--5 to 70C
0.2C
0.1C
C, F, K
--5 to 70C
0.35C
0.1C
C, F, K
4-meters submersion
Four electrode cell
0 to 200 mS/cm
0.5% of reading or 0.001 mS/cm,
whichever is greater (4-m cable)
1% of reading or 0.001 mS/cm,
whichever is greater (20-m cable)
0.001 mS/cm to 0.1 mS/cm (range
dependent)
S, mS
1 point
Calculated from conductivity and
temperature
0 to 70 ppt
1.0% of reading or 0.1 ppt,
whichever is greater
0.01 ppt
ppt, PSU
1 point
Glass Combination Electrode
0 to 14 units
0.2 units
0.01 units
mV, pH units
1, 2, 3, 4, 5, or 6 point (user selec
table); US, NIST or Custom Buffers
Platinum button
--1999 to +1999 mV
+/-20 mV in redox standards
0.1 mV
mV
1 point
Ammonium***
Nitrate***
Chloride***
Total Dissolved Solids
(TDS)
Sensor Type
Range
Accuracy
Resolution
Units
Calibration
Depth
Sensor Type
Range
Accuracy
Resolution
Units
Calibration
Depth
Sensor Type
Range
Accuracy
Resolution
Units
Calibration
Depth
Sensor Type
Range
Barometer
Resolution
Units
Sensor Type
Range
Accuracy
Resolution
Units
Calibration
Ion Selective Electrode
0 to 200 mg/L-N, 0 to 40C
10% of reading or 2 mg/L-N,
whichever is greater
0.1 mg/L
mg/L-N, mV
1, 2, or 3 point (user selectable)
17-meters submersion
Ion Selective Electrode
0 to 200 mg/L-N, 0 to 40C
10% of reading or 2 mg/L-N,
whichever is greater
0.1 mg/L
mg/L-N, mV
1, 2, or 3 point (user selectable)
17-meters submersion
Ion Selective Electrode
0 to 1000 mg/L, 0 to 40C
15% of reading or 5 mg/L,
whichever is greater
0.1 mg/L
mg/L-Cl-, mV
1, 2, or 3 point (user selectable)
17-meters submersion
Calculated from conductivity and
temperature
0 to 100 g/L
Cal constant range 0.30 to 1.00
(0.64 default)
0.001, 0.01, 0.1g/L
kg/L, g/L
Piezoresistive
375 to 825 mmHg
3 mmHg within 15C from
calibration point
0.1 mmHg
mmHg, inHg, mbar, psi, kPa, ATM
1 point
(INSTRUMENT ONLY AT AMBIENT TEMPERATURE)
pH
Range
-2.60 to 16.60
ORP
Conductivity
Dissolved Oxygen
Temperature
*
Lab-grade field cables include 605107, -108, -109, 605177, -178, -179
** Derived parameters can include resistivity, salinity, specific conductance, and total dissolved solids
*** ISE sensors for freshwater only
Accuracy
Resolution
Range
Accuracy
Resolution
Range
Accuracy
Resolution
Range
Accuracy
Resolution
Range
Accuracy
Resolution
Units
+/-0.1 mV (0.01 pH units)
0.1 mV (0.01 pH units)
-1999 to +1999 mV
+/-0.5 mV
0.01 mV
0.0 to 200 mS/cm
+/-0.3% FS +/-1 digit for each range
0.0001 mS/cm or 0.1 uS/cm to 0.1
mS/cm (range dependent)
0.00 to 90 mg/L; 0 to 550%
+/-0.2% FS (550% air saturation)
+/- 1 digit (with 1.25 PE membrane
at 10C)
0.01 mg/L; 0.1% air saturation
-10 to 100.00C
+/-0.2% FS (550% air saturation)
+/- 1 digit (with 1.25 PE membrane
at 10C)
0.1C
C, F, K
ProfessionalPlus
To order or for more
information, contact
YSI Environmental.
800 897 4151
www.YSI.com
YSI Environmental
1700/1725 Brannum Lane
Yellow Springs, Ohio 45387
+1 937 767 7241
+1 937 767 9353 fax
[email protected]SonTek/YSI
+1 858 546 8327
+1 858 546 8150 fax
[email protected]YSI Integrated Systems
& Services
+1 508 748 0366
+1 508 748 2543 fax
[email protected]YSI Gulf Coast
+1 225 753 2650
+1 225 753 8669 fax
[email protected]YSI Hydrodata
European Support Centre
+44 1462 673581
+44 1462 673582 fax
[email protected]YSI Hydrodata (Ireland)
+353 87 2430264
+353 46 9487637 fax
[email protected]YSI (Hong Kong) Limited
+852 2891 8154
+852 2834 0034 fax
[email protected]YSI Middle East (Bahrain)
+973 17536222
+973 17536333 fax
[email protected]STEP 1
ORDER INSTRUMENT
(INSTRUMENT INCLUDES DATA MANAGER SOFTWARE,
THE COMMUNICATIONS SADDLE, AND USB CABLE)
6050000
Professional Plus
STEP 4
ORDER SOLUTIONS
3161
1,000 us/cm conductivity solution (quart)
3163
10,000 us/cm conductivity solution (quart)
3169
50,000 us/cm conductivity solution (8 pints)
3682
Zobell, ORP solution (125 mL)
3824
pH 4, 7, 10 buffers (2 pints each)
3841
1 mg/L ammonium solution (500 mL)
3842
10 mg/L ammonium solution (500 mL)
3843
100 mg/L ammonium solution (500 mL)
3885
1 mg/L nitrate solution (500 mL)
3886
10 mg/L nitrate solution (500 mL)
3887
100 mg/L nitrate solution (500 mL)
STEP 2
ORDER CABLE(S)*
(ALL CABLES INCLUDE TEMPERATURE; CABLES OVER
1-METER INCLUDE CABLE MANAGEMENT KIT)
60510-1, 4, 10, 20 or 30-meter
ISE**/temp
60520-1, 4, 10, 20 or 30-meter*** DO/temp
60530-1, 4, 10, 20 or 30-meter
conductivity/temp
6051010-1, 4, 10, 20 or 30-meter ISE/ISE/temp
6051020-1, 4, 10, 20 or 30-meter ISE/DO/temp
6051030-1, 4, 10, 20 or 30-meter ISE/conductivity/temp
6052030-1, 4, 10, 20 or 30-meter DO/conductivity/temp
605107 1-meter single-junction probe and cable for pH/temp
605177 4-meter single-junction probe and cable for pH/temp
605108 1-meter single-junction probe and cable for ORP/temp
605178 4-meter single-junction probe and cable for ORP/temp
605109 1-meter single-junction probe and cable for pH/ORP/temp
605179 4-meter single-junction probe and cable for pH/ORP/temp
STEP 3
ORDER PROBE(S)
605202
Galvanic DO
605203
Polarographic DO
605780
Self-stirring BOD
605101
pH (ISE)
605102
ORP (ISE)
605103**** pH/ORP combination (ISE)
605104
Ammonium (ISE)
605105
Chloride (ISE)
605106
Nitrate (ISE)
STEP 5
ORDER ACCESSORIES
603059
Flow Cell, Standard, 203 mL
603077
Flow Cell Kit, 1 or 2 Port Probe (includes
603059 & 603078)
603078
Flow Cell Adapter, Single Port
603056
Flow Cell Mounting Spike
603075
Carrying Case, Soft-Sided
603074
Carrying Case, Hard-Sided
603069
Belt Clip
063517
Ultra Clamp
063507
Tripod Clamp
603062
Cable Management Kit
605978
Weight, Probe/Cable, 4.9 oz.
063019
Weight, Probe/Cable, 24 oz., 3
063020
Weight, Probe/Cable, 51 oz., 6
*
**
***
****
Cables with conductivity include probe (no need to order conductivity probe)
ISEs include pH, ORP, Ammonium, Nitrate, Chloride
Special order cables up to 100-meters are available with 60520 cables
Not recommended for 6051010 cables. Only pH would be reported.
Nanotech/YSI (Japan)
+81 44 222 0009
+81 44 221 1102 fax
[email protected]
YSI China
+86 10 5203 9675
+86 10 5203 9679 fax
[email protected]
ISO 9001
ISO 14001
(Yellow Springs facility)
Pure Data for a Healthy Planet and
Whos Minding the Planet are registered
trademarks of YSI Incorporated.
Whos Minding
the Planet?TM
2007 YSI Incorporated
Printed on recycled paper in USA
W14 0507
603059 Flow Cell
603069 Belt Clip
603075 Soft-sided Carrying Case
605978 Cable Weights