0% found this document useful (0 votes)
194 views4 pages

Spec Mantis

The document describes the MANTIS compact flaw detector which uses phased array ultrasonic testing (PAUT), conventional UT, and total focusing method (TFM) for inspection. It has 16 to 64 channels, supports linear, sectorial and compound scanning, and provides real-time TFM imaging. The device is portable and lightweight, has an intuitive interface, and includes onboard analysis tools and reporting capabilities.

Uploaded by

assurendran
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
194 views4 pages

Spec Mantis

The document describes the MANTIS compact flaw detector which uses phased array ultrasonic testing (PAUT), conventional UT, and total focusing method (TFM) for inspection. It has 16 to 64 channels, supports linear, sectorial and compound scanning, and provides real-time TFM imaging. The device is portable and lightweight, has an intuitive interface, and includes onboard analysis tools and reporting capabilities.

Uploaded by

assurendran
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

MANTIS

COMPACT FLAW DETECTOR WITH TFM

LIGHTWEIGHT 16:64PR PORTABLE


PHASED-ARRAY FLAW DETECTOR

PAUT | sectorial, linear & compound scanning


Conventional UT | pulse-echo & dual techniques
TOFD | time of flight diffraction with lateral wave straightening
TFM | total focusing method in real-time for expertise
MANTIS
FULLY LOADED

Real-time TFM
Onboard PAUT calculator for all geometries
Compound scanning
Multi-group configurations
High PRF
Up to 3 encoded axes
ISO & ASTM code compliant

QUICK SETUP TIME

Intuitive interface, step by step app design


3-click TCG, TCG save and import
3-click material velocity
3-click probe balancing
3-click wedge calibration
Onboard library of probes, wedges & scanners
Onboard library of geometries and weld preps
Application oriented templates

ONBOARD ANALYSIS

800% amplitude dynamic


Cumulated volume views
TOFD lateral wave linearization
Customizable inspection report
Dedicated analysis tools
Fast Ethernet file transfer
PC data analysis with CAPTURE

@M2M_ndt m2m_ndt [Link]


MANTIS
MANTIS comes in 3 cost-effective packages: Adept, Expert and Master.

MASTER
Key features: Typical applications:
Hi-resolution TFM Thick pipes, austenitic welds,
Matrix limited probe access, 3D
and dual arrays (DLA, DMA) focusing
FMC recording R&D
+ EXPERT features + EXPERT Typical applications

EXPERT
Key features: Typical applications:
3-axis encoding Nozzle inspection
live 3D-overlays T-, K-, Y-Joints
20kHz PRF + ADEPT Typical applications
+ ADEPT features

ADEPT Key features:


PAUT + TOFD + PE + TFM
Weld prep overlay
CAD import
Multi-group
Quick TCG/DAC/AVG
Quick calibration wizards
Comes with CAPTURE for PC
CIVA and ENLIGHT compatible
Free data viewer
Fast Ethernet file transfer + USB 3.0
Unique flaw-detector with real-time TFM

Typical applications:
Composite, Thin pipes, Corrosion
Rope access inspections

M2M
1 rue de Terre-Neuve . Bt. H. Miniparc du Verger . 91940 Les Ulis . France | t. +33 (0)1 60 92 39 65 contact@[Link]
MANTIS
general
L x W x H: 320mm x 220mm x 100mm 8.4 high contrast resistive screen - resolution 1024x768 px
Operating temperature range: from -10C to 45C | 14F to 113F Weight: 4,4kg with battery
Storage temperature range: -10C to 60C | 14F to 140F with battery IP65 according to CEI60529
Operating time: 4h (hot swappable battery) Shock resistance according to MIL-STD-810G

standard phased-array
Linear scanning, sectorial scanning, compound Linear, matrix*, DLA and DMA* probes
Maximum active aperture: 16 channels Up to 6 probes | Up to 8 groups | Up to 2048 delay-laws
Phased array computation delay laws on plate, cylinder, T* & Y*, nozzle* CIVA fueled phased-array calculator
Focusing mode: true depth, sound path, projection

real-time TFM
Reconstruction channels: 16 up to 64* Max number of points of reconstructed image: up to 65k
Max refresh rate: up to 80fps Sound paths: direct (L or S), indirect* and converted* modes

pulsers
64 phased-array channels: UT-TOFD:
Negative square pulse, width: 35ns to 1250ns Negative square pulse, width: 30ns to 1250ns
Voltage: 12V 90V with 1V step Voltage: 12V to 200V with 1V step
Max. PRF: 12kHz up to 20kHz* Max. PRF: 12kHz up to 20kHz*

receivers
16 phased-array channels: UT-TOFD:
Input impedance: 50 Input impedance: 50
Frequency range: 0.4 to 20MHz Frequency range: 0.6 to 25MHz
Max. input signal: 2Vpp | TCG ACG DGS calibration wizard Max. input signal: 2Vpp
Gain: up to 120dB (0.1dB step) TCG DAC calibration wizard

Indicated values may change without notice. 2017 M2M. All rights reserved. Ref: PRO_MANTIS_SPEC_V1_EN
Cross-talk between two channels < 50 dB Gain: up to 120dB (0.1dB step)

digitizer
Digitizing and real-time summation on 16 channels Resolution: 16bits
FIR filters Max. sampling frequency: 100 MHz
Real-time averaging up to x32 Digitizing depth up to 16k points
Rectified, RF, envelope A-scan range or delay max 65k points

acquisition
Hardware acquisition gates Max. data flow 150 MB/s on a 128Gb SSD
A-Scan/Peak data recording Inspection data file size: up to 10Gb
FMC recording Data transfer through Ethernet
Acquisition trigger on time, event, encoder 800% amplitude range

wizards
CAD overlay and 3D view Scanner calibration
Real-time phased array calculator Amplitude calibration (TCG, DAC, DGS)
Base-time calibration for conventional UT Probe design | Weld geometry design
Wedge calibration (angle, height) Amplitude balancing
Velocity calibration Part geometry with parametric shapes: plate, cylinder, T* & Y*, nozzle*

analysis
Capture software with analysis and reporting tools Free viewer Amplitude range: 800%
A-Scan, B-Scan, C-Scan, D-Scan, Echodynamic, Top view, Side view, 3D view Overlay part geometry: plate, cylinder, T* or Y* section, nozzle*
Analysis gates Overlay weld geometry
Compatibility with CIVA analysis and ENLIGHT Customizable inspection report

I-O
Encoder inputs: 2 axes up to 3 axes* 2 LEMO 00 connectors for UT-TOFD (1 PR 1R)
1 IPEX connector for phased-array probe - can be upgraded to 2 with splitter* 1 external trigger
1 USB 2.0 + 1 USB 3.0 + 1 mini display port + 1 RJ45 Ethernet 7 TTL inputs/outputs

* Option
Standard: EN ISO 18563-1 for phased array channels
Standard: EN ISO 12668-1 for conventional channel