Peng2019 PDF
Peng2019 PDF
Original article
A R T I C L E I N F O A B S T R A C T
Article history: Purpose: Polyetheretherketone (PEEK), a high-strength, aesthetic, and non-allergic thermoplastic
Received 9 April 2019 polymer, recently became a candidate for replacing metallic components in dental prosthesis. However,
Received in revised form 17 July 2019 as PEEK is flexible, the need for retention presents a key challenge in terms of its clinical application. In
Accepted 26 July 2019
this study, clasps prepared using PEEK were optimized and evaluated to provide the mechanical
Available online xxx
properties required by dentures.
Methods: Seventy-two three-dimensional rod-shape models, based on four thickness/width ratios, three
Keywords:
base widths, and six taper ratios were created. These models were analyzed using finite-element
Polyetheretherketone
Dental clasp
methods to determine which modified clasp arm shape provided the most appropriate mechanical
Shape optimization properties. Three shape-optimized PEEK specimens and one standard-shape Co–Cr alloy specimen were
Fatigue then fabricated. Constant-displacement fatigue testing was performed to calculate load values and
CAD/CAM deformations after ten years of clinical use.
Results: Shape optimization indicated a maximum stress concentration that was consistently located at
the base of the specimen, a correlation between mean load values and thickness that was greater than
that with the width, and a correlation between taper ratio and mean load values. Fatigue testing showed
that although PEEK exhibited significantly lower average load values than the Co–Cr alloy, these were
sufficient for clinical use. All specimens exhibited significant deformation during the first period of
cycling; however, there was no significant difference in the deformation between the two materials after
fatigue testing.
Conclusions: PEEK exerts fewer stresses on abutments compared to standard-alloy clasps, provides
adequate retention, and satisfy aesthetic demands, indicating that PEEK presents a promising alternative
to conventional metal clasps.
© 2019 Japan Prosthodontic Society. Published by Elsevier Ltd. All rights reserved.
[Link]
1883-1958/ © 2019 Japan Prosthodontic Society. Published by Elsevier Ltd. All rights reserved.
Please cite this article in press as: T.-Y. Peng, et al., Finite-element analysis and optimization of the mechanical properties of
polyetheretherketone (PEEK) clasps for removable partial dentures, J Prosthodont Res (2019), [Link]
G Model
JPOR 595 No. of Pages 7
Fig. 1. Schematic illustration of (A) the specimens and (B) the constant displacement fatigue test.
Fixed area = portion for fixation to the machine, loading point = the loading location used by the testing machine, Wb = width of the base, Wt = tip width, Tb = thickness of the
base, and Tt = tip thickness (all in mm).
such as yttria-stabilized tetragonal zirconia polycrystal (Y-TZP) and cross-sectional area was defined using six taper ratios, as shown in
ceria-stabilized zirconia/alumina nanocomposite (Ce-TZP/A), have Table 1. Overall, a total of 72 differently shaped 3D models were
been recently developed and applied within mainstream restor- created.
ative dentistry. Although the application of zirconia ceramics to
RPD clasps has recently been attempted, their clinical lifespan 2.2. Shape optimization
must still be evaluated [7,8].
Developments in materials science have led to the introduction Shape optimization was performed using the finite element
of polyetheretherketone (PEEK), a high-performance thermoplas- method (FEM). Load values for various clasp arm shapes at
tic polymer, to restorative dentistry, as a candidate for replacing displacements of 0.25 mm and 0.50 mm were measured, and stress
metallic components in dental prosthesis. The mechanical distributions were calculated under each loading condition. The
properties of PEEK do not change during the sterilization process material used for optimization in the present study was assumed
and its elastic modulus is similar to those of human bone, enamel, to be linearly elastic, homogeneous, and isotropic, so as to
and dentin, suggesting it to be a suitable restorative material. PEEK approximate the real properties of PEEK (VESTAKEEP DC4450 R
features stable chemical properties, and is biocompatible, wear- (Lot No.: 57781699), Evonik Japan Co., Tokyo, Japan). The following
resistant, stable at high temperatures, insoluble in water. This data were used, as supplied by manufacturers: tensile strength
material also presents low reactivity with other materials, is non- stress at yield = 110 MPa; modulus of elasticity = 4.8 GPa; specific
allergic, and has lower plaque affinity than other materials such as gravity = 1.52 g/cm3; linear thermal expansion = 0.45 10 4 K 1,
metals and resins. Furthermore, PEEK can be processed using and each model was assembled using linear FEM (Autodesk
computer-aided design and computer-aided manufacturing (CAD/ Nastran In-CAD, AutoDesk Inc., San Rafael, CA, USA). A finer mesh
CAM), rendering it easily reproducible in the event of failure, and was generated at the material interface to ensure the accuracy of
easily relined in the case of resorption [9–16]. force transfer, and all nodes in fixed area were restrained. A
The combination of these unique mechanical and physical concentrated load was applied to the loading points of the models.
properties renders PEEK a promising material for replacing metal Clasp arm shapes with tapered uniformly from the tip to the base
frameworks. To date, there have been few clinical studies that discuss and thinner in width or thickness exhibited better functionality
the application of PEEK as a framework material for RPDs [17,18]; and lower abnormal sensation [22]. Due to this, the appropriately
nevertheless, according to its superior flexible properties, obtaining three shape-optimal 3D models with the required mechanical and
the necessary retentive force and fatigue resistance will be key aesthetic properties could be extracted and fabricated into test-
challenges in the development of PEEK RPD clasps [19,20]. Therefore, specimens.
the objective of this study is to optimize PEEK clasp design in order to
provide the mechanical properties required by RPDs. 2.3. Test-specimen fabrication
2. Materials and methods Three PEEK (VESTAKEEP DC4450 R (Lot No.: 57781699), Evonik
Japan Co) test specimens were fabricated for the shape-optimized
2.1. Generating three-dimensional models 3D models using CORiTEC 250I/DRY (imes-icore, Eiterfeld,
Germany). One standard-shape (base thickness/width of 1.00/
Three-dimensional (3D) models with clasp arms in the form of a 2.00 mm (Group A1), taper ratio of 0.8) cobalt–chromium (Co–Cr)
rod-shape were designed by SolidWorks 2013 (Dassault Systèmes alloy specimen (Co 63%; Cr 30%; Mo 5%; Wironit Extra-hard, Bego,
SolidWorks, Waltham, MA, USA). The 3D models were 15 mm in Bremen, Germany) was also prepared for comparison, and the
length and the loading point was set at 3 mm from the tip (Fig. 1). plastic pattern (SHERAprint-vast, SHERA Werkstoff-Technologie,
These models were classified into four groups based on thickness/ Lemförde, Germany) for the cast was created using Rapid Shape
width ratios (Tb/Wb), and each group was then divided into three D30 II (Rapid Shape, Heimsheim, Germany), then invested and cast
subgroups according to the base width. In each subgroup, the according to the manufacturer’s instructions.
Please cite this article in press as: T.-Y. Peng, et al., Finite-element analysis and optimization of the mechanical properties of
polyetheretherketone (PEEK) clasps for removable partial dentures, J Prosthodont Res (2019), [Link]
G Model
JPOR 595 No. of Pages 7
Table 1. 3D models with 72 different shapes (4 groups, 3 subgroups, and 6 taper ratios for each subgroup).
Wt Tt Wt Tt Wt Tt Wt Tt Wt Tt Wt Tt
A Group A1 2.00 1.00 1.00 0.50 1.20 0.60 1.40 0.70 1.60 0.80 1.80 0.90 2.00 1.00
Group A2 2.50 1.25 1.25 0.63 1.50 0.75 1.75 0.88 2.00 1.00 2.25 1.13 2.50 1.25
Group A3 3.00 1.50 1.50 0.75 1.80 0.90 2.10 1.05 2.40 1.20 2.70 1.35 3.00 1.50
B Group B1 2.00 1.25 1.00 0.63 1.20 0.75 1.40 0.88 1.60 1.00 1.80 1.13 2.00 1.25
Group B2 2.50 1.56 1.25 0.78 1.50 0.94 1.75 1.09 2.00 1.25 2.25 1.41 2.50 1.56
Group B3 3.00 1.88 1.50 0.94 1.80 1.13 2.10 1.31 2.40 1.50 2.70 1.69 3.00 1.88
C Group C1 2.00 1.50 1.00 0.75 1.20 0.90 1.40 1.05 1.60 1.20 1.80 1.35 2.00 1.50
Group C2 2.50 1.88 1.25 0.94 1.50 1.13 1.75 1.31 2.00 1.50 2.25 1.69 2.50 1.88
Group C3 3.00 2.25 1.50 1.13 1.80 1.35 2.10 1.58 2.40 1.80 2.70 2.03 3.00 2.25
D Group D1 2.00 1.75 1.00 0.88 1.20 1.05 1.40 1.23 1.60 1.40 1.80 1.58 2.00 1.75
Group D2 2.50 2.19 1.25 1.09 1.50 1.31 1.75 1.53 2.00 1.75 2.25 1.97 2.50 2.19
Group D3 3.00 2.63 1.50 1.31 1.80 1.58 2.10 1.84 2.40 2.10 2.70 2.36 3.00 2.63
Wb = width of the base, Wt = tip width, Tb = thickness of the base, Tt = tip thickness (as Fig. 1).
a
Groups are divided based on the thickness/width (Tb/Wb) ratio (Group A: 0.500; Group B: 0.625; Group C: 0.750; Group D: 0.875). b Subgroups are divided based on the
width of the base (Wb) (Subgroup 1: 2.00 mm; Subgroup 2: 2.50 mm; Subgroup 3: 3.00 mm). c Taper ratios are the ratio of the cross-sectional dimensions at the tip to those of
the base [Wt/Wb = Tt/Tb].
2.4. Constant-displacement fatigue testing (0.5- to 0.7-taper), Group D2 (0.7- to 0.8-taper), and Group D3
(0.5- to 0.6-taper) were greater than the lowest acceptable
Test conditions were maintained at room temperature using a retentive force (1.6 N) of RPDs (Fig. 2), and the maximum stresses
Servopulser testing machine (EHF-FD5KN-4LA, Shimadzu Corp., were less than the yield stress (110 MPa) of the employed PEEK
Kyoto, Japan) and carried out under two conditions: one with a material (Fig. 3). Among those, Group B3 (0.9-taper), Group C3
constant displacement of 0.25 mm for both the PEEK and Co–Cr (0.5-taper), and Group D2 (0.7-taper) were slender and with
alloy specimens, the other at a constant displacement of 0.50 mm uniformly tapered; thus, identified as the appropriately three
for only the PEEK specimens. These constant displacements were modified shape-optimal 3D models.
maintained at the loading point of the specimens with a sinusoidal The von Mises stress analyses indicated that maximum stress
wave frequency of 5 Hz. In this way, the present study may assume concentrations were consistently located at the base of each model
that the clasps were inserted and removed from the abutment with (Fig. 4). For each group, there was a significant difference between
an undercut of 0.25 mm or 0.50 mm. The load values were recorded the maximum stress values with varying both cross-sectional
in order to analyze deformation during testing; the deformation of dimensions and taper ratios (P < 0.05). There was also a strong
the specimens in the direction of the load was observed using a positive correlation between the thickness (r > 0.8) and stress
digital microscope (KH-1300, Hirox Co., Tokyo, Japan) every 3000 concentration (Table 2). Additionally, there was a significant
cycles. A total of 15,000 cycles were performed, representing the difference in the mean load values that were obtained for different
simulated insertion and removal of the RPDs over ten years, with cross-sectional dimensions (P < 0.05), and the correlation between
the assumption that the patient would perform four complete mean load values and thickness (r > 0.9) was greater than that
cycles per day. between the mean load values and the width (r > 0.6) (Table 2). For
identical dimensions, a larger taper ratio was associated with
2.5. Statistical analysis increased mean load values (Fig. 2).
The results of shape optimization were recorded, and the 3.2. Constant-displacement fatigue tests
correlation between the cross-sectional dimensions, taper ratio,
load values, and maximum stress were obtained via Pearson Three 3D models (Group B3 (0.9-taper), Group C3 (0.5-taper),
correlation and linear regression. The mean interval load values and Group D2 (0.7-taper)) were selected via shape optimization
and deformations were calculated for the constant displacement and used in the fabrication of PEEK specimens. The Shapiro–Wilk
fatigue tests, and the normality of the distribution and the test and the Levene’s test run on all variables showed normally
homogeneity of variance were primarily analyzed using the distributed and homoscedasticity. Table 3 shows the average load
Shapiro–Wilk test and Levene’s test. Data comparisons were values and deformation in the direction of the loads for PEEK and
conducted using one-way analysis of variance (ANOVA) with the Co–Cr alloy. The Co–Cr alloy (mean: 8.26 N) exhibited
post-hoc Scheffé tests. All statistical analyses were performed consistently and significantly higher average load values than
using IBM SPSS statistical software (SPSS version 24; IBM Corp., PEEK (mean: 2.06–3.67 N) during 15,000 cycles (P < 0.05).
Armonk, NY, USA), and the level of statistical significance was set The deformations in the direction of the loads in all the interval
at 5%. cycles are presented in Fig. 5. All specimens exhibited significantly
greater deformation during the first period of cycling (P < 0.05),
3. Result which then remained unchanged until the end of cyclic testing. The
final extent of deformation after 15,000 cycles was small, showing
3.1. Shape optimization no significant differences between each cycle interval. There were
no significant differences between the deformation of the PEEK
When the displacement was set at either 0.25- or 0.50-mm, (mean: 0.011–0.017 mm) and Co–Cr alloy (mean: 0.017 mm)
the mean load values for Group B3 (0.9-taper), Group C3 samples after 15,000 cycles (P = 0.11) (Table 3).
Please cite this article in press as: T.-Y. Peng, et al., Finite-element analysis and optimization of the mechanical properties of
polyetheretherketone (PEEK) clasps for removable partial dentures, J Prosthodont Res (2019), [Link]
G Model
JPOR 595 No. of Pages 7
Fig. 2. Load values when constant displacements of (A) 0.25 mm and (B) 0.50 mm were applied to the 3D models.
The horizontal dashed line represented the lowest acceptable retentive force (1.6 N) of RPDs.
Please cite this article in press as: T.-Y. Peng, et al., Finite-element analysis and optimization of the mechanical properties of
polyetheretherketone (PEEK) clasps for removable partial dentures, J Prosthodont Res (2019), [Link]
G Model
JPOR 595 No. of Pages 7
Fig. 4. Representative von Mises stresses (MPa) and distributions, obtained when constant displacements of (A) 0.25 mm and (B) 0.50 mm were applied to the 3D models.
Table 2. Pearson’s correlations between variables: width, thickness, and taper ratio.
Table 3. Pairwise comparison of the load value and deformation of specimens (n = 7).
SD: Standard deviation. Within the same column, different letters indicate groups that are statistically different (P < 0.05).
Frank et al. [32], suggested that a retention of 2.94–7.35 N was retentive force for a clasp will actually decide on the type and
necessary in the case of a distal extension RPD to protect against number of clasps in the RPD; however, the lowest acceptable
removal during the chewing of food, and Ahmad et al. [33] asserted retentive force for one clasp was determined to be approximately
that the guiding planes would also provide some retentive force 1.6 N [34,35]. The retentive force required for each component, will
(mean retention = 2.41 N). Various studies have discussed the differ considerably with varying RPD design. The obtained
necessary retentive force for RPD clasps, showing that the suitable constant-displacement fatigue test data (Table 3) indicated that
Please cite this article in press as: T.-Y. Peng, et al., Finite-element analysis and optimization of the mechanical properties of
polyetheretherketone (PEEK) clasps for removable partial dentures, J Prosthodont Res (2019), [Link]
G Model
JPOR 595 No. of Pages 7
Conflict of interest
Acknowledgements
Fig. 5. Deformation of the specimens for every 3000 cycles, in the direction of
loading. The authors thank Mr. Katsuhiko Sugawara (KS dental,
Yokohama, Japan) for technical assistance with manufacturing
the specimens, and Evonik Japan Co., Ltd. for supplying the free
the average load values of the PEEK clasp (mean: 2.06–3.67 N) were PEEK materials for specimen manufacture in this study.
smaller than those of the Co–Cr alloy (mean: 8.26 N); although,
considering all the factors mentioned above, PEEK clasps may References
provide a sufficient retentive force for clinical use.
Metal materials with larger rigidity are unsuitable for large [1] Baba K. Paradigm shifts in prosthodontics. J Prosthodont Res 2014;58:1–2.
undercuts, due to the fact that the clasps would place a large stress [2] Joda T, Zarone F, Ferrari M. The complete digital workflow in fixed
prosthodontics: a systematic review. BMC Oral Health 2017;17:124.
on the abutments [36,37]. However the lower elastic modulus of [3] Hamann CP, DePaola LG, Rodgers PA. Occupation-related allergies in dentistry.
PEEK renders this material suitable for use with a larger undercut, J Am Dent Assoc 2005;136:500–10.
which may be advantageous in clinical situations in which more [4] Poggio CE, Ercoli C, Rispoli L, Maiorana C, Esposito M. Metal-free materials for
fixed prosthodontic restorations. Cochrane Database Syst Rev 2017;12:
aesthetically pleasing results or improved periodontal health are
CD009606.
required [19,20,38]. Moreover, when reinforced by plaque control [5] Kishita C, Hamano T, Tsuru K, Nishi Y, Nagaoka E. Application of a glass-
and regular denture maintenance, there is no evidence suggesting reinforced composite material to clasps—the effects of immersion and
repeated loading. Dent Mater J 2004;23:528–32.
that a deeper undercut will result in microbiological risks related
[6] Fueki K, Ohkubo C, Yatabe M, Arakawa I, Arita M, Ino S, et al. Clinical
to periodontitis in the abutment teeth of RPD wearers [39,40]. application of removable partial dentures using thermoplastic resin-part I:
Within the limitations of this in vitro study, it was possible to definition and indication of non-metal clasp dentures. J Prosthodont Res
assert that PEEK specimens with a width of 3.00 mm, a thickness of 2014;58:3–10.
[7] Tanimoto Y. Dental materials used for metal-free restorations: recent advances
2.25 mm at the base (Group C3), a taper of 0.5, and with an and future challenges. J Prosthodont Res 2015;59:213–5.
undercut of 0.50 mm exhibit the best mechanical properties. [8] Urano S, Hotta Y, Miyazaki T, Baba K. Bending properties of Ce-TZP/A
However, the curved clasp-arm shape was not considered in the nanocomposite clasps for removable partial dentures. Int J Prosthodont
2015;28:191–7.
present study; and the various paths of insertion required by actual [9] Schwitalla A, Müller WD. PEEK dental implants: a review of the literature. J
clinical usage may produce greater loads on the abutment teeth, Oral Implant 2013;39:743–9.
resulting in permanent deformation of the clasp over a short [10] Stawarczyk B, Beuer F, Wimmer T, Jahn D, Sener B, Roos M, et al.
Polyetheretherketone — a suitable material for fixed dental prostheses? J
period of time. The effects of these factors should be confirmed by Biomed Mater Res B Appl Biomater 2013;101:1209–16.
further research in order to determine the optimal design and [11] Ma R, Tang T. Current strategies to improve the bioactivity of PEEK. Int J Mol Sci
clinical efficacy of PEEK clasps. However, the superior flexibility 2014;15:5426–45.
[12] Garcia-Gonzalez D, Rusinek A, Jankowiak T, Arias A. Mechanical impact
and lower elastic modulus of PEEK, which engages a deeper
behavior of polyether–ether–ketone (PEEK). Compos Struct 2015;124:88–99.
undercut than the Co–Cr alloy but also exerts lower stresses on the [13] Najeeb S, Zafar MS, Khurshid Z, Siddiqui F. Applications of polyetheretherke-
abutments, suggest that this material is appropriate for use in tone (PEEK) in oral implantology and prosthodontics. J Prosthodont Res
RPDs. Removable partial dentures with PEEK clasps are recom- 2016;60:12–9.
[14] Skirbutis G, Dzingute_ A, Masiliunaite_ V, Šulcaite_ G, Žilinskas J. A review of
mended in clinical cases with sufficient residual tooth, aesthetic PEEK polymer’s properties and its use in prosthodontics. Stomatologija
concerns, or even predominant concerns about periodontal health. 2017;19:19–23.
Its benefits range from the maintenance of periodontal health, [15] Wagner C, Stock V, Merk S, Schmidlin PR, Roos M, Eichberger M, et al.
Retention load of telescopic crowns with different taper angles between
satisfying aesthetic demands, and the improvement of patient cobalt-chromium and polyetheretherketone made with three different
quality of life. manufacturing processes examined by pull-off test. J Prosthodont
2018;27:162–8.
[16] Arnold C, Hey J, Schweyen R, Setz JM. Accuracy of CAD-CAM-fabricated
5. Conclusion removable partial dentures. J Prosthet Dent 2018;119:586–92.
[17] Zoidis P, Papathanasiou I, Polyzois G. The use of a modified Poly-Ether-Ether-
A comprehensive consideration of the results of the present in Ketone (PEEK) as an alternative framework material for removable dental
prostheses. A clinical report. J Prosthodont 2016;25:580–4.
vitro study prompts the following conclusions: [18] Dawson JH, Hyde B, Hurst M, Harris BT, Lin WS. Polyetherketoneketone (PEKK),
a framework material for complete fixed and removable dental prostheses: a
1. The differences between the mean load values provided by clinical report. J Prosthet Dent 2018;119:867–72.
[19] Arda T, Arikan A. An in vitro comparison of retentive force and deformation of
clasps with various cross-sectional dimensions were significant,
acetal resin and cobalt-chromium clasps. J Prosthet Dent 2005;94:267–74.
and the effect of thickness was greater than that of width. For [20] Meenakshi A, Gupta R, Bharti V, Sriramaprabu G, Prabhakar R. An evaluation of
clasps of the same dimension, a higher taper ratio was retentive ability and deformation of acetal resin and cobalt-chromium clasps. J
Clin Diagn Res 2016;10:ZC37–41.
associated with higher mean load values.
[21] Sato Y, Yuasa Y, Akagawa Y, Ohkawa S. An investigation of preferable taper and
2. The load values required for deflecting the Co–Cr alloy (mean thickness ratios for cast circumferential clasp arms using finite element
SD: 8.26 0.55) were significantly higher (P < 0.05) than those analysis. Int J Prosthodont 1995;8:392–7.
Please cite this article in press as: T.-Y. Peng, et al., Finite-element analysis and optimization of the mechanical properties of
polyetheretherketone (PEEK) clasps for removable partial dentures, J Prosthodont Res (2019), [Link]
G Model
JPOR 595 No. of Pages 7
[22] Carr AB, Brown DT. McCracken’s removable partial prosthodontics. 12th ed. St. [33] Ahmad I, Sherriff M, Waters NE. The effect of reducing the number of clasps on
Louis: Elsevier/Mosby; 2010. removable partial denture retention. J Prosthet Dent 1992;68:928–33.
[23] Merk S, Wagner C, Stock V, Schmidlin PR, Roos M, Eichberger M, et al. Retention [34] Soo S, Leung T. Hidden clasps versus C clasps and I bars: a comparison of
load values of telescopic crowns made of Y-TZP and CoCr with Y-TZP secondary retention. J Prosthet Dent 1996;75:622–5.
crowns: impact of different taper angles. Materials 2016;9:354. [35] Helal MA, Baraka OA, Sanad ME, Ludwig K, Kern M. Effects of long-term
[24] Davenport JC, Basker RM, Heath JR, Ralph JP, Glantz PO. Prosthetics: retention. simulated RPD clasp attachment/detachment on retention loss and wear for
Br Dent J 2000;189:646–57. two clasp types and three abutment material surfaces. J Prosthodont
[25] Tannous F, Steiner M, Shahin R, Kern M. Retentive forces and fatigue resistance 2012;21:370–7.
of thermoplastic resin clasps. Dent Mater J 2012;28:273–8. [36] Bridgeman JT, Marker VA, Hummel SK, Benson BW, Pace LL. Comparison of
[26] Wang L-F, Shi L-Y. Simulation optimization: a review on theory and titanium and cobalt-chromium removable partial denture clasps. J Prosthet
applications. Acta Autom Sin 2013;39:1957–68. Dent 1997;78:187–93.
[27] Su J, Chen Z, Wang Z, Xi G. An adaptive finite element method for shape [37] Savitha PN, Lekha KP, Nadiger RK. Fatigue resistance and flexural behavior of
optimization in stationary incompressible flow with damping. Int J Numer acetal resin and chrome cobalt removable partial denture clasp: an in vitro
Anal Model 2014;5:79–96. study. Eur J Prosthodont Restor Dent 2015;3:71–6.
[28] Amaran S, Sahinidis NV, Sharda B, Bury SJ. Simulation optimization: a review [38] Rodrigues RC, Ribeiro RF, de Mattos Mda G, Bezzon OL. Comparative study of
of algorithms and applications. 4OR-Q J Oper Res 2014;12:301–33. circumferential clasp retention force for titanium and cobalt-chromium
[29] Fitton JS, Davies EH, Howlett JA, Pearson GJ. The physical properties of a removable partial dentures. J Prosthet Dent 2002;88:290–6.
polyacetal denture resin. Clin Mater 1994;17:125–9. [39] Mine K, Fueki K, Igarashi Y. Microbiological risk for periodontitis of abutment
[30] Turner JW, Radford DR, Sherriff M. Flexural properties and surface finishing of teeth in patients with removable partial dentures. J Oral Rehabil
acetal resin denture clasps. J Prosthodont 1999;8:188–95. 2009;36:696–702.
[31] Rabinowitz S, Beardmore P. Cyclic deformation and fracture of polymers. J [40] Shimura Y, Wadachi J, Nakamura T, Mizutani H, Igarashi Y. Influence of
Mater Sci 1974;9:81–99. removable partial dentures on the formation of dental plaque on abutment
[32] Frank RP, Nicholls JI. A study of the flexibility of wrought wire clasps. J Prosthet teeth. J Prosthodont Res 2010;54:29–35.
Dent 1981;45:259–67.
Please cite this article in press as: T.-Y. Peng, et al., Finite-element analysis and optimization of the mechanical properties of
polyetheretherketone (PEEK) clasps for removable partial dentures, J Prosthodont Res (2019), [Link]