Users manual Contents of the Lagaay Oxygen Bag
(TYPE INTERNATIONAL / IVW COLUMN A/B)
Lagaay Oxygen Bag
204230 or 279750 Oxygen cylinder 2 liter, 200 bar,
bull nose or pin index connection 1 pce
Read this instruction carefully
210710 or 248960 Oxygen regulator, bull nose
Please make sure that all personnel handling the product are or pin index connection 1 pce
provided with the operating instructions & performance data 201390 Guedelairway, size 1 1 pce
and are fully trained to carry out the operation.
201400 Guedelairway, size 2 1 pce
In this manual you will find:
201410 Guedelairway, size 3 1 pce
1. An overview of the Lagaay Bag content 201420 Guedel airway, size 4 1 pce
2. Operating precautions
3. An Instruction for the oxygen flow meter/regulator 202750 Suction catheter, size CH12 1 pce
4. Picture manual for a step by step installation 246940 Oxygen mask rebreathing adult 2 pce
5. How to respond in an emergency
246950 Oxygen mask child 1 pce
210490 Res-Q-VacHand suction unit 1 pce
202910 Resuscitation bag white complete 1 pce
250320 Bag for oxygen set 1 pce
275430 Instructions for use oxygen set 1 pce
295090 Check and renew document 1 pce
999340 Proof of receipt and 1st check 1 pce
Lagaay Bag
999080 Certificate oxygen set 1 pce
Instruction Bullnose flow meter
4 Gauge
Inlet nut Operating precautions
2 • Never permit oil and grease to come in contact with Oxygen
cylinders, valves, regulators, gauges and fittings.
• Never use Oxygen from a cylinder without reducing the
pressure through a suitable regulator intended for that
1 Flow head purpose only. Stand to the valve side of the regulator when
opening cylinder valve.
Flow outlet 3
• Pressure regulators should never be changed by workers with
greasy hands. Never test the pressure in an Oxygen cylinder
Connecting the oxygen flow meter to the oxygen cylinder by holding a gloved hand over the nozzle. An oil soaked glove
may catch fire and cause the loss of one’s fingers or severe
a.Remove protection caps from inlet nut (2) and the yellow cap from the burns.
oxygen cylinder valve. Keep caps in a safe place for reuse during transport or
storage. • If valve is difficult to open, point the valve opening away from
b.Turn the knob on the flow in clockwise direction to make sure that the flow you and use greater force. Avoid, however, the use of a
head (1) says “ZERO“. wrench on valves equipped with hand wheels.
c.Manually screw the inlet nut (2) onto the cylinder valve. Turn the regulator
into the correct position for use and tighten the nut by hand – do not use any • Never permit Oxygen to enter the regulator suddenly. Open
tools! the cylinder valve slowly.
d.Connect the hose nipple to the flow outlet (3). Now you can connect any
accessory to the hose nipple (oxygen masks, resuscitator inhalator, • Before regulator is removed from a cylinder, close the cylinder
nebulizers). valve and release all gas from regulator.
e.Open the cylinder valve fully by turning the hand wheel anticlockwise .
f.Check the gauge (4) on the regulator to get a reading for the amount of • Avoid sparks or flame from welding or cutting torch coming in
oxygen in the oxygen cylinder. contact with cylinders.
g.Turn the knob of the flow head (1) to change the level of oxygen
being released. • Fully open the cylinder valve when cylinder is in use.
After use • Oxygen must never be used to blow dust out of clothing or to
a.Turn off the cylinder valve by turning the hand wheel in a clockwise direction freshen the air in closed places. This may cause explosive
to the stop position. Do not use excessive force. danger.
b.Reset the flow head (1) on the “ZERO” position – when the gas venting has
ceased. • Oxygen must be stored at least 20 feet from fuel gases such
c.Ensure the pressure gauge does not show any residual pressure. as Acetylene, Hydrogen, Methane, Propane, etc.
d.Refit pressure outlet and flow outlet protection caps. Before refitting the
caps, ensure they are clean.
USER MANUAL
LAGAAY OXYGEN BAG
1: Remove yellow cap from valve 2: Take out the flow meter 3: Remove protection caps
4: Connect the hose nipple to the flow outlet 5: Connect the flow meter to the cylinder
6: Manually screw the inlet nut to the oxygen valve 7: Ready for oxygen masks 8: Connect the oxygen mask onto the hose nipple
9: Set the oxygen flow using the knob of the flow head 10: Read the right quantity of oxygen on the gauge
How to respond if the victim is still slightly How to respond in case the victim does not
breathing. breathe independently.
Use: Oxygen supply with inhalation mask Use: oxygen supply together with resuscitator.
1. Slowly open the oxygen valve (turn to the left). 1. Slowly open the oxygen valve (turn to the left)
2. Connect oxygen tube of the inhalation mask to the 2. Connect oxygen tube of the inhalation mask to the
outlet/nipple of the regulator. outlet/nipple of the regulator.
3. Adjust the oxygen flow with one of the knobs at the front side 3. Adjust the oxygen flow with the knob at the front side of the
of the regulator. Possible settings: Adults 6-6 ltr./min. regulator. Possible settings: Adult 6-8 ltr./min.
4. Clean the mouth and throat of the victim using your fingers or 4. Clean the mouth and throat of the victim (narrow side on the
a cloth. Be aware of candy, chewing gum or vomit, etc. nose and wide side on the chin).
5. Bend the head of the victim backwards, in order to clear the 5. Bend the head of the victim backwards, in order to clear the
airway. airway.
6. Place the inhalation mask over mouth and nose of the victim 6. Place the mask of the resuscitator over the mouth and nose
(narrow side on the nose and wide side on the chin). of the victim (narrow side on the nose and wide side on the
chin).
7. Press slightly with your hand on the mask, avoid oxygen
leakage, (patient exhales by himself). 7. Press slightly on the mask, i.e. to avoid leakage of oxygen
during the resuscitation.
8. When finished, close the oxygen cylinder and let the oxygen
leave the system until the pressure gauge of the regulator 8. Use the other hand to squeeze the resuscitator slowly appr.
indicates 0 m/bar. Turn the knobs of the flow regulator to the 12-15 x per minute.
left till the indication o liters/minute.
9. When finished, close the oxygen cylinder and let the oxygen
9. Replace used masks. leave the system until the pressure gauge of the regulator
indicated 0 m/bar. Turn the knobs of the regulator to the left
to the indication o liters/minute.
10. Clean used mask and resuscitator with mild liquid soap or by
sterilizing