Type 5974 8-Channel Charge Amplifier

The Type 5974 is an 8-channel charge amplifier that conditions accelerometer signals for recording. It provides filtering, gain adjustment, and buffering between accelerometers and recording equipment like the SONY PC 208 A DAT recorder. Key features include eight independent channels, adjustable input sensitivity and filtering per channel, and a tachometer input. It is portable and battery powered, making it suitable for multi-channel vibration measurements in the field.

Uploaded by

jhon vargas
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
344 views4 pages

Type 5974 8-Channel Charge Amplifier

The Type 5974 is an 8-channel charge amplifier that conditions accelerometer signals for recording. It provides filtering, gain adjustment, and buffering between accelerometers and recording equipment like the SONY PC 208 A DAT recorder. Key features include eight independent channels, adjustable input sensitivity and filtering per channel, and a tachometer input. It is portable and battery powered, making it suitable for multi-channel vibration measurements in the field.

Uploaded by

jhon vargas
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

Product Data

8-channel Charge Amplifier — Type 5974


USES: FEATURES:
❍ Multichannel accelerometer preamplifier ❍ Eight identical channels
❍ Input module for the SONY PC 204/208 A DAT ❍ Individual settings for input sensitivity, high-pass
instrumentation cassette recorder filtering and low-pass filtering
❍ In high-quality field and laboratory measurement ❍ Internal selection of floating or grounded input and
systems output, acceleration, velocity, or displacement
❍ Indication of overload for each channel
❍ Internal rechargeable battery for portability
❍ Tacho input channel supporting Brüel & Kjær
Photoelectric Probe MM 0024

Type 5974 8-channel Charge Ampli-


fier is a conditioning amplifier provid-
ing filtering and gain to accelerometer
signals, and buffering from other
equipment in the configuration.
The unit is primarily for use with a
SONY PC 208 A DAT Instrumentation
Cassette Recorder and is housed in a
similar case, powered by the same type
of battery or power source. Portability Fig.1 Type 5974 8-channel Charge Amplifier front panel
is a key feature. Up to 8 accelerometer
signals can be recorded on the digital
tape recorder. These are then replayed access four additional switches per where a number of signals have to be
into a vibration analyzer for processing channel by simply removing the top measured simultaneously. The 5974
of field measurements in a laboratory
environment.
plate. These switches allow you to is normally used with a DAT record-
choose between each of the following er, although it can be used on its own
options: as a multichannel accelerometer con-
❍ grounded or floating input ditioning amplifier.
Overview ❍ semi-floating or grounded output Attaching calibrated accelerome-
❍ measuring acceleration directly or ters allows high-quality vibration
integrating measurements to be made on moving
All eight vibration conditioning chan- ❍ (if integration is chosen) measur- machines while under non-laboratory
nels of Type 5974 are electrically ing velocity or measuring displace- conditions. For example, the charac-
identical, but operate entirely inde- ment. teristics of a motor can be recorded
pendently, giving minimal cross- The accelerometer signals are input in a real-life situation for future ana-
channel differential gain. The active via TNC connectors on the front pan- lysis.
circuitry conforms to the same high el, and the output signals are sent A special feature of the Type 5974
quality expected of digital recording via BNC connectors on the side pan- is its built-in tachometer circuitry.
systems, and will not unduly degrade el. The outputs are all low impedance Tachometer pulses, for example from
analogue signals before conversion to and electronically semi-balanced. a Brüel & Kjær Photoelectric Probe
a digital format. MM 0024, are converted to 5 V logic
With its four-position input sensi- levels in the unit. These tachometer
tivity switches (0 dB to +30 dB gain) Applications pulses can be recorded by a DAT re-
the amplifier can be used with a wide corder on an independent digital
variety of Brüel & Kjær’s accelerome- channel, so that important measure-
ters. Four low-pass and four high- The Type 5974 is built into a small ment parameters (such as engine
pass filters per channel are also se- cabinet. This makes it suitable for a speed) can be directly related to the
lectable from the front panel. You can wide variety of portable applications measured data.

Brüel & Kjær B K


Recording on DAT

The Type 5974 is primarily for use


with a SONY PC 208 A (or four-chan-
nel version PC 204 A) DAT recorder.
For this reason the case size and con-
nector positioning are set to match
those of the DAT recorder. By adding
side frames, the two units are incor-
porated into one portable system (the
Type 9686 as shown in Fig. 2). When
the 5974 is connected to the DAT re-
corder, it acts as a buffer amplifier
for the DAT recording inputs.
In this combined mode, the frequen- Fig.2 Type 5974 and DAT recorder as one package, the Type 9686
cy response limit of the system is de-
fined by the upper limit of the DAT
recorder. When running at twice the The Type 5974 and the DAT record- each unit. The internal battery can
standard tape speed, the eight DAT er operate using the same type of re- be trickle-charged via the external
channels have a maximum recordable chargeable battery or external power supply. A full recharge takes a max-
frequency of 10 kHz. This is sufficient source (a 12 V vehicle battery for ex- imum of 14 hours.
for most measurement situations in ample). This avoids the problem of
which an accelerometer is used requiring different power systems for

2
Specifications 5974
Amplifier Stages Data Output: 9-pin, sub-D type socket RADIATED RF: (3 to 10 V/m, 80% AM, 1 kHz)
INPUT TYPE: Grounded or Floating EXTERNAL SUPPLY INPUT: CONDUCTED RF: (3 to 10 V, 80% AM, 1 kHz)
MAX. INPUT LEVEL: 7 V RMS 2-pin (cable WL 1192)
CMRR (@ 100 Hz, floating input, relative to
ground): >40 dB Dimensions and Weight
INPUT SENSITIVITY (±0.2 dB): Height: 70 mm (2.8″)
(output = 1 mv/1 pC at 0 dB gain) Width: 297 mm (11.7″)
0.316 pC/m/s2 (30 dB gain) Depth: 220 mm (8.7″)
1.00 pC/m/s2 (20 dB gain) Weight: 2.4 kg (1.1 lb.) without battery (3.2 kg
3.16 pC/m/s2 (10 dB gain) (1.5 lb.) with battery)
10.00 pC/m/s2 (0 dB gain) 1
Measured with max. gain and 1 nF termination
HIGH-PASS FILTERS (–1 dB):
0.3, 1, 3, or 10 Hz EMC
LOW-PASS FILTERS (–1 dB): SUSCEPTIBILITY TO DISTURBANCES
1, 5, or 10 kHz, or Off (40 kHz) SPECIFIED IN EN 50082–2:
INTEGRATOR GAIN (Input Sensitivity = 10 pC/ Measured using Accelerometer Cable AO 1382
m/s2): with Ferrite Cable Clamp LK 0014 mounted on
Velocity: 0 dB @ 16 Hz accelerometer cable and Ferrite Cable Clamp
Displacement: 0 dB @ 4 Hz LK 0013 mounted on external DC cable. Note: All values are typical at 25°C (77°F), un-
INTEGRATOR LOWER LIMITING At magnetic field 30 A/m input noise can in- less measurement uncertainty or tolerance field
FREQUENCY: 1 Hz crease to 300 µV (30 dB gain) is specified. All uncertainty values are specified
FIXED OUTPUT CALIBRATION: 10 mV/m/s2 At magnetic field 80 A/m input noise can in- at 2σ (i.e. expanded uncertainty using a cover-
OUTPUT IMPEDANCE: crease to 660 µV (30 dB gain) age factor of 2)
50 Ω (100 Ω with semi-floating output)
MAX. OUTPUT LEVEL: 7 V RMS COMPLIANCE WITH STANDARDS:
OVERLOAD LEVEL: 10.5 V peak
CHANNEL SEPARATION (10 kHz): >80 dB
OUTPUT NOISE (0.3 Hz to 22 kHz, with CE-mark indicates compliance with: EMC Directive.
shorted input):
<12 µV (0 dB gain)
<35 µV (10 dB gain) Safety EN 61010–1 and IEC 1010–1: Safety requirements for electrical equipment for
<60 µV (20 dB gain) measurement, control and laboratory use.
<200 µV (30 dB gain)
EMC Emission EN 50081–1: Generic emission standard. Part 1: Residential, commercial and
Tachometer light industry.
MAXIMUM PULSE FREQUENCY: 96 kHz EN 50081–2: Generic emission standard. Part 2: Industrial environment.
MAXIMUM PULSE AMPLITUDE: 10 V CISPR 22: Radio disturbance characteristics of information technology
THRESHOLD LEVEL: equipment. Class B Limits.
Positive-going: 1.9 V FCC Rules, Part 15: Complies with the limits for a Class B digital device.
Negative-going: 0.9 V
EMC Immunity EN 50082–1: Generic immunity standard. Part 1: Residential, commercial and
light industry.
Power Supply
EN 50082–2: Generic immunity standard. Part 2: Industrial environment.
BATTERY:
Type: NP–1B rechargeable (WQ 1129) Note 1: The above is guaranteed using accessories listed in this Product Data
Capacity: 2.3 Ah sheet only.
Life: Approx. 5 hours Note 2: See “EMC”
Charge: Max. 14 hours via external supply
EXTERNAL SUPPLY: 12 V to 20 V DC Temperature IEC 68–2–1 & IEC 68–2–2: Environmental Testing. Cold and Dry Heat.
SUPPLY CURRENT: Operating Temperature: 0 to +50°C
450 mA at 12 V (550 mA when tacho probe is Storage Temperature: –25 to +70°C
connected)
550 mA at 18 V and charging battery Humidity IEC 68 –2 –3: 90% RH (non-condensing at 40°C)

Enclosure IEC 529: IP 20


Connectors
ACCELEROMETER INPUT: TNC Mechanical Non-operating:
SIGNAL AND REGENERATED TACHO IEC 68–2–6: Vibration: 0.3 mm, 20 m/s2, 10–500 Hz
OUTPUT: BNC
IEC 68–2–27: Shock: 1000 m/s2
TACHO:
Input: BNT IEC 68–2–29: Bump: 1000 bumps at 250 m/s2

Ordering Information
Type 5974 8-channel Charge Amplifier LK 0013 Ferrite cable clamp for ext. DC or Optional Accessories
Includes the following accessories: BNC cable
WB 1375: Mains adaptor, 100 to 127 VAC LK 0014: Ferrite cable clamp for MM 0024: Photoelectric Probe
or accelerometer cable WQ 1128: SONY 4-cell battery charger
WB 1376: Mains adaptor, 200 to 240 VAC WQ 1129: SONY rechargeable battery pack
WU 0484: Guard frames and screw set To make a combined accelerometer amplifier/ (NP-1B)
DH 0541: Carrying strap DAT recorder package, you should additionally
WL 1141: Tachometer data interface cable order a SONY PC 204/208 A DAT recorder. This
8 × WL 1157: 0.09 m BNC to BNC cables must be done via a SONY agent.
WL 1192: External supply cable)

Brüel & Kjær reserves the right to change specifications and accessories without notice

3
Brüel & Kjær B K

WORLD HEADQUARTERS:
DK-2850 Naerum · Denmark · Telephone: +45 45 80 05 00 · Fax: +45 45 80 14 05 · Internet: [Link] · e-mail: info@[Link]
Australia (02 ) 9450-2066 · Austria 00 43-1-865 74 00 · Belgium 016/44 92 25 · Brazil (011) 246-8166 · Canada: (514) 695-8225 · China 10 6841 9625 / 10 6843 7426
Czech Republic 02-67 021100 · Finland 90-229 3021 · France (01) 69 90 69 00 · Germany 0610 3/908-5 · Holland (0)30 6039994 · Hong Kong 254 8 7486
Hungary (1) 215 83 05 · Italy (02) 57 60 4141 · Japan 03-3779-8671 · Republic of Korea (02) 3473-0605 · Norway 66 90 4410 · Poland (0-22) 40 93 92 · Portugal (1) 47114 53
Singapore (65) 275-8816 · Slovak Republic 07-37 6181 Proprietary — 00
· Spain (91) 36810 Authorized
· Sweden (08) Agency Staff Only
71127 30 · Switzerland 01/94 0 09 09 · Taiwan (02) 713 9303
United Kingdom and Ireland (0181) 954-236 6 · USA 1 - 800 - 332 - 2040
Local representatives and service organisations worldwide
BP 1419 – 14 96/08

You might also like