0 ratings0% found this document useful (0 votes) 693 views208 pagesEddy Current
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content,
claim it here.
Available Formats
Download as PDF or read online on Scribd
- Pece F523
AECL-7523 6 Feu Fz
ATOMIC ENERGY LENERGIE ATOMIQUE
OF CANADA LIMITED DU CANADA LIMITEE
EDDY CURRENT TESTING
MANUAL ON EDDY CURRENT METHOD
Essais par courant de Foucault
Manuel des méthodes d’essai par courant de Foucault
Volume 1
V.S. CECCO, G. VAN DRUNEN and F.L. SHARP
CERN LIBRARIES, GENEVA
(OA
Cm-P00067661
Chalk River Nuclear Laboratories Laboratoires nucléaires de Chalk River
Chalk River, Ontario
November 1981 novembreATOMIC ENERGY OF CANADA LIMITED
EDDY CURRENT TESTING
VOLUME 1
MANUAL ON EDDY CURRENT METHOD
Cecco, G. Van Drunen and F.L. Sharp
Chalk River Nuclear Laboratories
Chalk River, Ontario KOI 150
1981 NOVEMBER
AECL-7523LYENERGIE ATOMIQUE DU CANADA, LIMITEE
Essais par courant de Foucault
Volume 1
Manuel des méthodes d'essai par courant de Foucault
V.S. Cecco, G. Van Drunen et F.L. Sharp
Résumé
Ce manuel de référence et d'instruction a pour but de fournir
& ceux qui font des essais par courant de Foucault les principes
fondamentaux de la technique et les connaissances voulues pour
interpréter conme il faut les résultats souvent compliqués de ces
essais. Une approche non rigoureuse est employée pour simplifier
les complexes phénoménes physiques. L'accent est mis sur un choix
approprié de fréquences d'essai et sur 1'interprétation des signaux.
La détection et le diagnostic des défauts font 1'objet d'une attention
particuliére. La conception et la réalisation des sondes sont traitées
de facon approfondie car les sondes jouent un réle clé dans les essais
Par courant de Foucault. Les avantages et les limitations des divers
types de sondes sont indiqués.
La théorie électromagnétique, 1'instrumentation, les méthodes
d’essai et les analyses de signaux sont décrites. Les réponses des
sondes permettent d'avoir une compréhension fondamentale du comportement
des courants de Foucault, & condition d'avoir recours aux déductions
simplifiges indiquées dans le manuel pour tester les paramétres. Les
signaux des courants de Foucault sont présentés sur des diagrammes de
plans d'impédance tout au long du manuel, car il s'agit 18 de 1'infor-
mation la plus commune affichée sur les instruments universels modernes.
L'emploi du "retard de phase" dans 1'analyse des signaux est décrit en
détail. Pour compléter 1a théorie, des exemples pratiques sont donnés.
Ces exemples ont pour but de rendre les inspections plus performantes
et ils montrent comment les principes de base s'appliquent au diagnostic
des signaux réels.
Laboratoires nucléaires de Chalk River
Chalk River, Ontario KOJ 1J0
Novembre 1981
AECL-7523ATOMIC ENERGY OF CANADA LIMITED
EDDY CURRENT TESTING
VOLUME 1
NANUAL ON EDDY CURRENT METHOD
V.S. Ceceo, G- Van Drunen and F.L. Sharp
ABSTRACT
This training and reference manual was assembled to provide
those involved in eddy current testing with both the
fundamental principles of the technique as well as the
knowledge to deal with often complicated test results. A
non-rigorous approach is used to simplify complex physical
phenomena. Emphasis is placed on proper choice of test
frequency and signal interpretation. Defect detection and
diagnosis receive particular attention. Design and
construction of probes are covered extensively since probes
play a key role in eddy current testing. The advantages and
limitations of various probe types are discussed.
Electromagnetic theory, instrumentation, test methods and
signal analysis are covered. Simplified derivations of probe
response to test parameters are presented to develop a basic
understanding of eddy current behaviour. Eddy current
signals are presented on impedance plane diagrams throughout
the manual since this is the most commen display on modern,
general purpose instruments. The use of "phase lag” in
signal analysis is covered in detail. To supplement theory,
practical examples are presented to develop proficiency in
performing inspections, and to illustrate how basic
principles are applied to diagnose real signals.
Chalk River Nuclear Laboratories
Chalk River, Ontario KOs 10
1981 NOVEMBER
AECL-7523witi-
ACKNOWLEDGEMENTS
This manual is an accumulation of knowledge and experience
obtained by the NDT Development Branch (formerly Quality
Control Branch) of CRNL through its 10 years of existence.
The authors are indebted to the other members of the
Nondestructive Testing Development Branch especially
C.R. Bax, H.W. Ghent, J.R. Carter, G.A. Leakey and
W. Pantermoller who assisted in collecting some of the data
in the manual and made many constructive criticisms.
All rights reserved. No part of this report may be
reproduced by any means, nor transmitted, nor translated into
a machine language without the written permission of Atomic
Energy of Canada Limited Research Company.TABLE OF CONTENTS
CHAPTER 1 - INTRODUCTION PAGE
del EDDY CURRENT TESTING 1
1.2 PURPOSE OF THIS MANUAL 1
1.3 HISTORICAL PERSPECTIVE 2
CHAPTER 2 - EDDY CURRENT FUNDAMENTALS
2el BASIC EQUIPMENT 5
2.2 GENERATION OF EDDY CURRENTS 6
2.2.1 Introduction 6
2.2.2 Magnetic Field Around a Coil 6
2.2.3 Equations Governing Generation cf Eddy
Currents 8
2.3 FUNDAMENTAL PROPERTIES OF EDDY CURRENT FLOW 10
2.4 SKIN EFFECT 11
2.4.1 Standard Depth of Penetration 12
2.4.2 Depth of Penetration in Finite Thickness
Samples 13
2.4.3 Standard Phase Lag 14
2.4.4 Phase Lag in Finite Thickness Samples 16
2.5 SUMMARY a7
2.6 WORKED EXAMPLES 18
2.6.1 Standard Depth of Penetration and Phase Lag 18
CHAPTER 3 - ELECTRICAL CIRCUITS AND PROBE IMPEDANCE
3el INTRODUCTION 19
3.2 IMPEDANCE EQUATIONS AND DEFINITIONS 19
3.3 SINUSOIDS, PHASORS AND ELECTRICAL CIRCUITS 21
3.4 MODEL OF PROBE IN PRESENCE OF TEST MATERIAL 23
3.5 SIMPLIFIED IMPEDANCE DIAGRAMS 25
3.5.1 Derivation of Probe Impedance for Probe/
Sample Combination 25
3.5.2 Correlation Between Coil Impedance and
Sample Properties 28
3.6 SUMMARY 30
3.7 WORKED EXAMPLES 31
3.7.1 Probe Impedance in Air 31
3.7.2 Probe Impedance Adjacent to Sample 32
3.7.3 Voltage-Current Relationship 32CHAPTER 4 - INSTRUMENTATION
INTRODUCTION
BRIDGE CIRCUITS
4.2.1 Simple Bridge Circuit
4.2.2 Typical Bridge Circuit in Eddy Current
Instruments
4.2.3 Bridge Circuit in Crack Detectors
RESONANCE CIRCUIT AND EQUATIONS
EDDY CURRENT INSTRUMENTS
4.4.1 General Purpose Instrument (Impedance Method)
4.4.2 Crack Detectors
4.4.3 Material Sorting and Conductivity
Instruments
SEND-RECEIVE EDDY CURRENT SYSTEMS
4.5.1 Hall-Effect Detector
4.5.2 Send-Receive Coils and Lift-Off Compensation
MULTIFREQUENCY EQUIPMENT
PULSED EDDY CURRENT EQUIPMENT
SPECIAL TECHNIQUES
RECORDING EQUIPMENT
4.9.1 Frequency Response
SUMMARY
WORKED EXAMPLES
4.11.1 Impedance at Resonance
PAGE
33
34
34
36
37
38
40
40
42
4h
45
46
47
48
49
51
53
53
54
34eta)
CHAPTER 5 - TESTING WITH SURFACE PROBES
INTRODUCTION
SURFACE PROBES
5.2.1 Probe Types
2. Directional Properties
[Link] Sensitivity at Centre of a Coil
5.2.3. Probe Inductance
PARAMETERS AFFECTING SENSITIVITY TO DEFECTS
1 Sensitivity with Lift-off and Defect Depth
2 Effect of Defect Length
COMPARISON BETWEEN SURFACE AND THROUGH-WALL INSPECTION
IMPEDANCE GRAPH DISPLAY
Effect of Resistivity
Effect of Permeability
Effect of Thickness
Effect of Frequency
Effect of Probe Diameter
Comparison of Experimental and Computer
Impedance Diagrams
BUGuae
wun
CHARACTERISTIC PARAMETER
DEFINITION OF “PHASE” TERMINOLOGY
SELECTION OF TEST FREQUENCY
5.8.1 Inspecting for Defects
5.8.2 Measuring Resistivity
5.8.3 Measuring Thickness
5.8.4 Measuring Thickness of a Non-conducting Layer
on a Conductor
5.8.5 Measuring Thickness of a Conducting Layer on
a Conductor
PROBE-CABLE RESONANCE
SUMMARY
WORKED EXAMPLES
5.11.1 Effective Probe Diameter
5.11.2 Characteristic Parameter
PAGE
55
356
59
60
61
65
65
66
67
69
72
72
72
73
73
74
77
78
78
80
83
84
84
85
86
88
88
88vii-
CHAPTER 6 - SURFACE PROBE SIGNAL ANALYSIS
PAGE
6.1 INTRODUCTION 89
6.2 EDDY CURRENT SIGNAL CHARACTERISTICS 89
6.2.1 Defect Signal Amplitude 89
6.2.2 Defect Signal Phase 91
6.3 EFFECT OF MATERIAL VARIATIONS AND DEFECTS IN A FINITE
THICKNESS 93
6.4 COIL IMPEDANCE CHANGES WITH DEFECTS 97
6.4.1 Surface Defect Measurement 97
6.4.2 Subsurface Defect Measurement 97
6.5 COIL IMPEDANCE CHANGES WITH OTHER VARIABLES 98
6.5.1 Ferromagnetic Indications 98
6.5.2 Electrical Resistivity 100
6.5.3 Signals from Changes in Surface Geometry 100
6.6 CALIBRATION DEFECTS 101
6.7 SUMMARY 104
CHAPTER 7 - TESTING OF TUBES AND CYLINDRICAL COMPONENTS
Tad INTRODUCTION 105
72 PROBES FOR TUBES AND CYLINDRICAL COMPONENTS 105
7.2.1 Probe Types 105
7.2.2 Comparing Differential and Absolute Probes 107
7.2.3 Directional Properties 109
7.2.4 Probe Inductance 110
7.2.5 Probe-Cable Resonance 112
7.3 IMPEDANCE PLANE DIAGRAMS 13
7.3.1 Solid Cylinders as
[Link] Sensitivity in Centre of a Cylinder 116
7.3.2 Tubes 118
7.3.3 Characteristic Frequency for Tubes 120
7.3.4 Computer Generated Impedance Diagrans 322
7.4 CHOICE OF TEST FREQUENCY 123
7.4.1 Test Frequency for Solid Cylinders 123
7.4.2 Test Frequency for Tubes 124
765 PROBES FOR DETECTING CIRCUMFERENTIAL CRACKS 125
7.6 SUMMARY 128
17 WORKED EXAMPLES 129
7.7.1 Calculate f/f, to operate at knee location,
for a cylinder 129
7.7.2 (a) Calculate optimum test frequency for tube
inspection 129
(b) Determine operating point for above frequency 130
(e) Caleulate frequency to discriminate ferro-
magnetic indications 130CHAPTER
8.5
CHAPTER
CHAPTER
10-1
10.2
10.3
-viti-
8 - TUBE TESTING - SIGNAL ANALYSIS
INTRODUCTION
EDDY CURRENT SIGNALS
Defect Signal Characteristics
Effect of Test Frequency
Calibration Tubes and Simple Defects
Vectorial Addition and Defects at Baffle Plates
Tube Inspection at Tubesheets
Testing Tubes with Internal Surface Probes
aUboRE
ANOMALOUS EDDY CURRENT SIGNALS
8.3.1 Ferromagnetic Inclusions and Deposits
3.2 Conducting Deposits
MULTIFREQUENCY EDDY CURRENT TESTING
8.4. Background
8.4.2 Multifrequency Testing of Dented Tubes
SUMMARY
9 - METALLURGICAL PROPERTIES AND TESTING FERRO-
MAGNETIC MATERIALS
INTRODUCTION
ELECTRICAL CONDUCTIVITY
9.2.1 Factors Affecting Resistivity
9.2.2 Material Sorting by Resistivity
MAGNETIC PROPERTIES
Magnetic Hysteresis
Magnetic Permeability
Factors Affecting Magnetic Permeability
aoe
oD
oto
TESTING MAGNETIC MATERIALS
9.4.1 Simplified Impedance Diagrans
9.4.2 Impedance Diagrams
9.4.3 Material Sorting by Magnetic Permeability
9.4.4 Testing for Defects in Magnetic Materials
SUMMARY
WORKED EXAMPLES
1 Calculate Conductivity
2 Calculate Magnetic Permeability
3 Calculate Standard Depth of Penetration
10 - DEFINITIONS, REFERENCES AND INDEX
DEFINITIONS
REFERENCES
INDEX
PAGE
131
131
131
135
138
142
146
147
149
149
153
155
155
158
162
163
163
163
166
168
169
170
172
176
174
176
178
178
184
185
185
185
186
187
194
195SYMBOL
ouverune
Bp ggngn ms
é
ONKMM Same maH
or or
eewseane
nix
NOMENCLATURE
QUANTITY
Cross-Sectional area
Radius
Length
Thickness
width
Diameter
Magnetic flux density
Capacitance
Test frequency
Optimum tube testing frequency
Characteristic or Limit
frequency :
Resonant frequency
Magnetic field intensity
(Ma gnetizing force)
Current
Current density
Self Inductance
Number of turns (Windings)
Characteristic Parameter
Resistance
Resistive load
Electric potential
Depth below the surface
Inductive Reactance
Capacitive Reactance
Impedance
Standard Depth
of Penetration
Permeability
Resistivity
Conductivity
Magnetic flux
Fill Factor
Phase Lag
Angular frequency
Angle between Z & R
SI UNIT
metre”
metre
metre
metre
metre
metre A
weber/meter
farads
hertz
hertz
or tesla
hertz
hertz
amperes/meter or
lenze
amperes p
amperes/meter
henry
dimensionless
dimensionless
ohm
oho
volt
metre
ohm
oho
ohm
metre
henry/meter
nicrohm-centimetre
stemens/meter
weber
dimensionless
radians
radians/second
degreesCHAPTER 1 - INTRODUCTION
EDDY CURRENT TESTING
Eddy current testing (ET) is a nondestructive test technique
based on inducing electrical currents in the material being
inspected and observing the interaction between those
currents and the material. Eddy currents are generated by
electromagnetic coils in the test probe, and monitored
simultaneously by measuring probe electrical impedance.
Since it's an electromagnetic induction process, direct
electrical contact with the sample is not required; however,
the sample material has to be conductive.
Eddy current testing is a versatile technique. It's mainly
used for thin materials; in thick materials, penetration
constraints limit the inspected volume to thin surface
layers. In addition to flaw inspection, ET can be used to
indirectly measure mechanical and metallurgical
characteristics which correlate with electrical and magnetic
properties. Also, geometric effects such as thickness,
curvature and probe-to-material spacing influence eddy
current flow and can be measured.
The large number of potentially significant variables in ET
is both a strength and a weakness of the technique since
effects of otherwise trivial parameters can mask important
information or be misinterpreted. Virtually everything that
affects eddy current flow or otherwise influences probe
impedance has to be taken into account to obtain reliable
results. Thus, credible eddy current testing requires a high
level of operator training and awareness.
PURPOSE OF THIS MANUAL
The purpose of this manual is to promote the development and
use of eddy current testing by providing a scientifically
sound training and reference manual. The selection of
material presented is based on the premise that a sound
understanding of basic principles is essential to obtaining
valid data and interpreting it correctly. A non-rigorous
approach has been used to present complex physical phenomena
in a document oriented towards application of eddy current
techniques, especially for defect detection and diagnosis.
The presentation moves from theory (including a review of
basic electrical concepts) to test methods and signal
analysis. Simplified derivations of probe response to test
parameters are presented to develop a basic understanding of
eddy current test principles. Thus, eddy current signals areconsistently illustrated on impedance plane diagrams (the
display used in modern eddy current test instruments) and to
aid explanation, the parameter “eddy current phase lag” is
introduced.
Since probes play a key role in eddy current testing,
technical aspects of probe design are introduced with
pertinent electrical impedance calculations. While knowledge
of basic electrical circuits is required for a complete
understanding of eddy current test principles, a good
technical base for inspection can still be obtained if
sections of this manual requiring such a background are
skipped.
From an applications point of view, the material in this
manual provides an inspector with the necessary background to
decide:
1) what probe(s) to use,
2) what test frequencies are suitable,
3) what calibration defects or standards are required for
signal calibration and/or simulation,
4) what tests are required to differentiate between
significant signals and false indications.
5) how to estimate depth of real defects
To supplement theory, practical examples are presented to
develop proficiency in performing inspections, and to
illustrate how basic principles are applied to diagnose real
signals.
A number of laboratory demonstrations, practical test
multiple choice questions are included in Volume 2,
Current Course Supplement”. They are divided into groups
corresponding to the chapters in this manual. The
demonstrations are intended for use in eddy current courses
to help clarify some of the more difficult concepts. The
practical tests are to give students practice in using
equipment and performing typical tests. The multiple choice
questions are intended to check students’ understanding of
the course material and prepare for certification exams.
HISTORICAL PERSPECTIVE
Electromagnetic testing -- the interaction of magnetic fields
with circulating electrical currents -- had its origin in
1831 when M. Faraday discovered electromagnetic induction.
He induced current flow in a secondary coil by switching a
battery on and off. D.E. Hughes performed the first recorded
eddy current test in 1879. He was able to distinguish
between different metals by noting a change in excitation
frequency resulting from effects of test material resistivity
and magnetic permeability.THERE MUST BE DEFECTS IN
THESE TUBES SOMEWHERE —
I SAW SQUIGGLES ON THE EDoy
CURRENT SCREENS
Fig. 1.1:Misinterpreted Signals
Initially, the extreme sensitivity to many material
properties and conditions made ET difficult and unreliable.
Figure 1.1 illustrates this point. It took until 1926 before
the first eddy current instrument was developed to measure
sample thickness. By the end of World War II further
research and improved electronics made industrial inspection
possible, and many practical instruments were developed. A
major breakthrough came in the 1950's when Forster developed
instruments with impedance plane signal displays. These made
it possible to discriminate between different parameters,
though the procedure was still empirical. During the 1960
progress in theoretical and practical uses of eddy current
testing advanced the technology from an empirical art to an
accepted engineering discipline.During that time, other nondestructive test techniques such
as ultrasonics and radiography became well established and
eddy current testing played a secondary role, mainly in the
aircraft industry. Recent requirements -- particularly for
heat exchanger tube inspection in the nuclear industry -~
have contributed significantly to further development of ET
as a fast, accurate and reproducible nondestructive test
technique.
Until recently, eddy current testing vas a technology where
he basic principles were known only to researchers, and a
black box” approach to inspection was often followed. The
authors’ objective in compiling this manual’is to draw upon
research, laboratory and industrial inspection experience to
bridge that gap and thereby permit the full potential of eddy
current testing to be realized.CHAPTER 2 ~ EDDY CURRENT FUNDAMENTALS
BASIC EQUIPMENT
Basic eddy current test equipment consists of an alternating
current source (oscillator), a probe containing a coil
connected to the current source, and a voltmeter which
measures the voltage change across the coil, as shown in
Figure 2.1.
OscILLAToR voumerer
consent PRORE
PROBE
‘MOVEMENT
TEST PLATE
Fig. 2
Eddy Current Test Equipment
The oscillator must be capable of generating a time varying
(usually sinusoidal) current at frequencies ranging from
about 1 kHz (1000 cycles per second) to about 2 MHz
(2,000,000 cycles per second). Oscillators which operate at
higher or lower frequencies, or with pulsed currents are used
for specialized applications
The coil within the probe is an insulated copper wire wound
onto a suitable form. The wire diameter, the number of turns
and coil dimensions are all variables which must be
determined in order to obtain the desired inspection results.
Coil variables are discussed in later chapters.2.21
Depending upon the type of inspection, an eddy current probe
can consist of a single test coil, an excitation coil with a
separate receive (sensing) coil, or an excitation coil with a
Hall-effect sensing detector, as shown in Figure 2.2.
‘voLTHeTER vouTereR VOLTAETER
‘osciLLaroR osciciator: oscrutaTor
sensine wate,
cot DeTEcroR
TEST ARTICLE
TEST ARTICLE TEST ARTICLE
7
con excitation EXCITATION
cont tole
(A) SELF-INDUCTANCE (B) SEND-RECEIVE (©) MAGNETIC REACTION
Fig. 2.2: Eddy Current Inspection System:
The voltmeter measures changes in voltage across the coil
which result from changes in the electrical conditions and
properties of the conducting material tested and/or changes
in relative position between the coil and the material
tested. This voltage change consists of an amplitude
variation and a phase variation relative to the current
passing through the coil. The reason for amplitude and phase
changes in this voltage is discussed in Chapter 3.
GENERATION OF EDDY CURRENTS
Introduction
In this section the topic of the magnetic field surrounding a
coil carrying current is introduced together with the
mechanism by which eddy currents are induced and how they are
measured.
Magnetic Field Around A Coil
Oersted discovered that whenever there is an electric
current, a magnetic field exists. Consider electric current
directed along a wire, a magnetic field is created in such a
direction that if your right-hand thumb points in the
direction of current, your curled fingers point in thedirection of the magnetic field. This is the “right-hand
rule".
Associated with a magnetic field is magnetic flux density.
It has the same direction as the magnetic field and its
magnitude depends upon position and current. It is therefore
a field vector quantity and is given the symbol B. Its unite
in the SI system is the tesla (T) or webers per square metre
(wb/m2
The B-field distribution around a long straight wire is shown
in Figure 2.3(a). In Figure 2.3(b) the B-field in the axial
direction of a single turn is shown as a function of radius.
As more windings are added, each carrying the same current,
the flux density rapidly increases and its associated
distribution is altered.
B
(a) Straight Wire (b) Single Turn Coil
Current Flowing into page
Fig Magnetic Flux Distribution
Flux density varies linearly with electric current in the
coil, i-e-, if coil current doubles, flux density doubles
everywhere. The total magnetic flux, ¢p, contained within
the loop is the product of B and area of the coil. The unit
in the SI system for magnetic flux is the weber (Wb).2.2.3 Equations Governing Generation of Eddy Currents
In any electrical circuit, current flow is governed by Ohm's
Law and is equal to the driving (primary circuit) voltage
divided by primary circuit impedance.
I= vi/z (2.1)
Pp PP
The eddy current coil is part of the primary circuit. The
current passing through the coil normally varies sinusoidally
with time and is given by:
= 2.2
1, T, sin(wt) (2.2)
where I, is the peak current value in the circuit and w
(omega) is the frequency in radians/s (w equals 2nf when £
is frequency in hertz).
From Oersted's discovery, a magnetic flux (4) exists around
a coil carrying current (see Figure 2.4) proportional to the
number of turns in the coil (Np) and the current (Ip).
¢, = NOT (2.3)
PROBE
(primary
efreuie)
SAMPLE
(secondary
elreust)
Coil Carrying Alternating Current AdjacentFaraday's Law states a voltage (Vg) 1s created or induced in
a region of space when there is a changing magnetic field.
When we apply this to our coil,
ox te (2.4)
‘8 p de
ap
where —g@ is the rate of change in @, with time.
Since coil current varies sinusoidally with time, total
magnetic flux in the coil also varies sinusoidally,
= 6, sin(ut
t) 7% stm(we)
where $, is the magnetic flux corresponding to Ig.
The induced voltage as described by equation 2.4 results in
= = N,w6, cos (ue) (2.5)
which also varies periodically with time, If we bring the
coil close to a test sample, Ohm's Lav states that if there
is a driving voltage (V,) and the sample's impedance is
finite, current will flow,
I, = V,/2, (2.6)
where I, is current flowing through the sample, V, is
induced voltage and Z, is the sample's impedance
or opposition to the flow of current.
These induced currents are known as eddy currents because of
their circulatory paths. They, in turn, generate their own
magnetic field according to Lenz's Law, which opposes the
primary field,
o, 8-1 (2-7)
(2.8)
and 27 4
where ¢; is the equilibrium magnetic flux surrounding the
coil in the presence of a test sample.
The flow of eddy currents results in resistive (Ohmic) losses
and a decrease in magnetic flux. This is reflected as a
decrease in probe impedance. In equation form,
2 oy (2.9)
an v= 21, :
a 7 (2.10)-10-
Equation 2.9 indicates a coil's impedance is a function of
the magnetic field surrounding it and in turn the magnetic
field is governed by induced current in the epecimen
(equations 2.8 and 2.7). The relations between probe
impedance and sample properties will be derived in Chapter 3.
To summarize, flux is set up by pasuing alternating current
through the test coil. When this coil is brought close to a
conductive sample, eddy currents are induced. In addition,
the magnetic flux associated with the eddy currents oppo
coil's magnetic flux, thereby decreasing net flux. This
results in a change in coil impedance and voltage drop. It is
the opposition between the primary (coil) and secondary (eddy
current) fields that provides the basis for extracting
information during eddy current testing.
the
It should be noted that if a sample is ferromagnetic,
equation 2.9 still applies but the magnetic flux is
strengthened despite opposing eddy current effects. The high
magnetic permeability of ferromagnetic materials
distinguishes them from non-ferromagnetic materials and
strongly influences eddy current test parameters.
Ferromagnetic specimen inspection is discussed in Chapter 9
and unless specified the rest of the manual is restricted to
non-ferromagnetic materials.
FUNDAMENTAL PROPERTIES OF EDDY CURRENT FLOW
Eddy currents are closed loops of induced current circulating
in planes perpendicular to the magnetic flux. They normally
travel parallel to the coil's winding and parallel to the
surface. Eddy current flow is limited to the area of the
inducing magnetic field.
Test frequency determines depth of penetration into the
specimen; as frequency is increased, penetration decreases and
the eddy current distribution becomes denser near the
specimen's surface. Test frequency also affects the
sensitivity to changes in material properties and defects.
Figure 2.5(a) shows the algebraic relationships and Figure
2.5(b) the oscilloscope display of eddy current and magnetic
field distribution with depth into the specimen. Both the
eddy currents and magnetic flux get weaker with depth because
of “skin effect". In addition to this attenuation, the eddy
currents lag in phase with depth. Eddy currents’ phase lag
ie the key parameter that makes eddy current testing a useful
NDT method. The parameters skin depth and phase lag are
discussed in the next section.1g
on,
con $9, sinlat)
Tetsinwt) /\
t
Te-1y(nsin(atp )
@ @
Fig. 2.5: Eddy Current and Magnetic Flux Distribution
With Depth Into a Conductor
SKIN EFFECT
Eddy currents induced by a changing magnetic field concentrate
near the surface adjacent to the excitation coil. The depth
of penetration decreases with test frequency and is a function
of electrical conductivity and magnetic permeability of the
specimen. This phenomenon is known as the skin effect and is
analogous to the situation in terrestrial heat conduction where
daily surface temperature fluctuations are not appreciable below
the earth's surface. Skin effect arises as follows: the eddy
currents flowing in the test object et any depth produce
magnetic fields which oppose the primary field, thus reducing
net magnetic flux and causing a decrease in current flow as
depth increases. Alternatively, eddy currents near the surface
can be viewed as shielding the coil's magnetic field thereby
weakening the magnetic field at greater depths and reducing
induced currents.
The equation for flow of induced currents is
25 2 on BS (2.11)
v's OU oe
where J 1s current density, o is conductivity, wu is magnetic
permeability and y2 is a differential operator
of second order.-12-
For a semi-infinite (thick) conductor the solution to the
above equation is
J,/Jg = &® sin(we-B) (2.128)
where Jx/Jg is the ratio of eddy current density Jy at
depth x to the surface density Jo,and e = 2.718 is the base
of natural logarithms. § is given by x/$ where 6 = (mfuc)-1/2,
the standard depth of penetration (see next section).
Equation 2-12(a) can be separated into two components:
« ex/é
I,/Iy =e (2,12b)
which describes the exponential decrease in eddy current
density with depth,and
Jg/Jo © sin (wt-x/6) (2.120)
denoting the increasing time or phas
signal with depth.
lag of the sinusoidal
2.4.1 Standard Depth of Penetration
Figure 2.6 illustrates the change in eddy current density in
a semi-infinite conductor. Eddy current density decreases
exponentially with depth.
HTP EEL
0.2 0.4 0.8 0.8 1.1
$08 to yy
3 85007, m
the
Fig. 2.6: Eddy Current and Magnetic Flux Distribution
qiehapeh-in-a Thick Plate. -13-
The depth at which eddy current density has decreased to 1/e or
36.8% of the surface density is called the standard depth of
penetration. The word ‘standard’ denotes plane wave
electromagnetic field excitation within the test sample
(conditions which are rarely achieved in practice). The
standard depth of penetration is given by
és 50vp/fu > mm (2.13a)
or 8 = 2e7ER, , inches (2.138)
where p 46 electrical resistivity in microhn-centimetres
£ is test frequency in hertz, and u, is relative magnetic
permeability (dimensionless)*.
The skin depth equation is strictly true only for infinitely
thick material and planar magnetic fields. Using the
standard depth, 6, calculated from the above equation
makes it a material/test parameter rather than a true
measure of penetration.
Depth of Penetration in Finite Thickness Samples
Sensitivity to defects depends on eddy current density at
defect location. Although eddy currents penetrate deeper
than one standard depth of penetration they decrease rapidly
vith depth. At two standard depths of penetration (25), eddy
current density has decreased to (1/e)? or 13.5% of the
surface density. At three depths (36) the eddy current density
is down to only 5% of the surface density. However, one should
keep in mind these values only apply to thick samples
(thickness, t >56) and planar magnetic excitation fields.
Planar field conditions require large diameter probes (diameter
>10t) in plate testing or long coils (length >5t) in tube
testing. Real test coils will rarely meet these requirements
since they would possess low defect sensitivity. For thin plate
or tube samples, current density drops off less than calculated
from equation 2-12(b) as shown in Figure 2.7. For solid
cylinders the overriding factor is a decrease to zero at the
centre resulting from geometry effects as shown in Fig. 2.7(c)
and discussed in Section 7.3.1.
One should also note, that the magnetic flux is attenuated
across the sample, but not completely. Although the currents
are restricted to flow within specimen boundaries, the
magnetic field extends into the air space beyond. This
allows the inspection of multi-layer components separated by
an air space.
‘See Chapter 9 for a description of electrical and magnetic
properties. uU, ™ uy , incremental permeability, at zero
biasing magnetization flux.(a) PLATE (LARGE coIL, 9108)
Tae EAUATION 2.12
pert
(0) TUBE (LONG ENCERCLING COIL, 288)
Jy © EDDY CURRENT DENSITY AT SURFACE
OR J_ * EDDY CURRENT DENSITY AT LOCATION
1 OR + BELOF THE SURFACE
Rane aT PT
et
LH
az TUBE AND ROD GEONETRY
2 (ey #0 FoR oD)
(©) ROD CENCIRCLING COIL. £>5 rq)
Fig. 2.7: Eddy Current Distribution With Depth in
Various Samples
The sensitivity to a subsurface defect depends on the eddy
current density at that depth, it is therefore important to
know the effective depth of penetration. The effective depth
of penetration is arbitrarily defined as the depth at which
eddy current density decreases to 5% of the surface density.
For large probes and thick samples, this depth is about three
standard depths of penetration. Unfortunately, for most
components and practical probe sizes, this depth will be les
than 35, the eddy currents being attenuated more than
predicted by the skin depth equation. The effect of probe
diameter on the decrease in eddy current density or defect
sensitivity with depth is discussed in Section 5.3.1.
Standard Phase Lag
The signal produced by a flaw depends on both amplitude and
phase of the currents being obstructed. A small surface
defect and large internal defect can have a similar effect
on the magnitude of test coil impedance. However, because of
the increasing phase lag with depth, there will be a
characteristic difference in the test coil impedance vector.
This effect allows location and extent of a defect to be
determined.Phase lag is derived from equation 2.12(c) for infinitely
thick material. It represents a phase angle lag of x/é
radians between the sinusoidal eddy currents at the surface
and those below the surface. It is éenoted by the symbol g
(beta) and is given by:
B= x/6 radians (2.14a)
oo B= x/6 x 57 degrees (2.14)
where x is distance below the surface in the same units as 6.
TT
B (OEGREES)
x
3
B= 3 x 57, DEGREES
Fig. 2.8: Eddy Current Phase Lag Variation With Depth
in Thick Samples
When x is equal to one standard depth of penetration,
phase lag is 57° or one radian. This means that the eddy
currents flowing below the surface, at one standard depth of
penetration, lag the surface currents by 57°. At two
standard depths of penetration they leg the surface currents
by 114%. This is illustrated in Figure 2.8.Phase Lag in Finite Thickness Samples
For thin samples, eddy current phase decreases slightly less
rapidly with depth than stated above. See Figure 2.9(a), (b)
and (c) for the plots of phase lag with depth for a plate,
tube, and cylinder, respectively. The phase lag illustrated
in these plots does not change significantly with cofl
diameter or length. For thick samples and practical probe
sizes, equation 2.14 is sufficiently accurate.
0 .
~
S
ah SS. 0
ON
wh NN ONS
Pr \ ALE
a0 \ tte.
a
7 \ t
0 7 oo
NL pete
100 100"
2 4 8 80
co) pute . co) Tue
goes
PLATE GtOVETRY
WOE AND ROD (
0) eowermY
By A, = PHASE MITA DEPTH x, OR + RELATIVE
ee” To sunrAce CURRENT
—— ror cuves
= CALGULATED, EQUATION 2.14 (4)
Fig. 2.9: Eddy Current Phase Lag in Various Samples
ce) roo
Phase lag can be visualized as a shift in time of the
sinusoidally varying current flowing below the surface. This
was illustrated in Figure 2.5. Phase lag plays a key role in
the analysis of eddy current test signals. It will be used
throughout the manual to link theory and observations. It
should not be misinterpreted or confused with the phase angle
between voltage and current in AC theory. Both the voltage
and current (and magnetic field) have this phase shift or lag
with depth.SUMMARY
Eddy current testing is based on inducing electrical currents
in the material being inspected and observing the interaction
between these currents and the material.
This process occurs as follows: When a periodically varying
magnetic field intersects an electrical conductor, eddy cur~
rents are induced according to Faraday's and Ohm's Laws. The
induced currents (known as eddy currents because of their
circulatory paths) generate their own magnetic field which
opposes the excitation field. The equilibrium field is re-
duced resulting in a change of coil impedance. By monitoring
coil impedance, the electrical, magnetic and geometric pro-
perties of the sample can be measured. Eddy currents are
closed loops of induced current circulating in planes perpen-
dicular to the magnetic flux. They normally travel parallel
to the coil's winding and parallel to the surface. Eddy cur-
rent flow is limited to the area of the inducing magnetic
field.
Depth of penetration into a material depends on its electri-
cal resistivity, magnetic permeability and on test frequency.
The basic equation of ET is the standard depth of penetration
given by
6 = sovp/fu, , mm (2.134)
where p is electrical resistivity, microhm-centimetres;
f is test frequency, hertz;
and ul, is relative magnetic permeability, dimensionless.
It states that in thick materials eddy current density
decreases to 37% of the surface density at a depth of one
standard depth of penetration. In most eddy current tests,
especially with surface probes, the actual eddy current
density (at a depth equal to that calculated by equation
2.13a) is much less than 37%.
Eddy currents also lag in phase with depth into the material.
Phase lag depends on the same material properties that
govern depth of penetration and is given by
B= x/6 = __X___, radians (2.14a)
50¥p7Eu,
where x is distance below surface, um.
Phase lag is the parameter that makes it possible to deter-
mine the depth of a defect. It also allows discrimination
between defect signals and false indications. It is the key
parameter in eddy current testing.2.6 WORKED EXAMPLES
2.6.1 Standard Depth of Penetration and Phase Lag
PROBLEM: (a) Calculate the standard depth of penetration
in a thick 304 SS sample, at a test frequency of
100 kHz.
(b) Determine the eddy current phase lag at a
depth of 1.5 mm in 304 SS at 100 kHz.
SOLUTION: 304 SS properties: 9 = 72 microhm ~ centimetres
and upd
(a) from equation 2.13(a),
6 = 50
= 50(7.2 x10
Therefore the standard depth of penetration is 1.3 mm.
(b) from equation 2.14(b),
8 = x/S x 57
= 2:3 x 57. 68 degrees
Therefore the phase lag is 64 degrees.-19-
CHAPTER 3 - ELECTRICAL CIRCUITS AND PROBE IMPEDANCE
INTRODUCTION
Eddy current testing consists of monitoring the flow and
distribution of eddy currents in test material. This is
achieved indirectly by monitoring probe impedance during a
test. An understanding of impedance and associated
electrical quantities is therefore imperative for a
fundamental appreciation of eddy current behaviour.
The first two sections review the electrical quantities
important in eddy current testing. This is followed by
presentation of a model of a test coil coupled to test
material and its equivalent electrical circuit. The
equivalent circuit approach permits derivation of simplified
impedance diagrams to show the effect of test and material
parameters on probe impedance in graphical form. Once the
simple impedance diagram concepts of this chapter are
understood, the more complex diagrans of subsequent chapters
should present little difficulty.
IMPEDANCE EQUATIONS AND DEFINITIONS
All information about a sample is obtained through changes in
electrical characteristics of the coil/sample combination.
Therefore a basic understanding of electrical quantities is
important for eddy current inspection.
RESISTANCE: (symbol: R, units: ohm, 2)
Opposition to the flow of electrical current is
called resistance. It is constant for both
direct and alternating current. The electrical
component is called a resistor.
V=IR Ohm's Law Gel)
where, V is voltage drop across resistor (volt), and
I is current through resistor (ampere)
INDUCTANC
: (symbol: L, unite: henry, H)
The property of an electric circuit by virtue of
which a varying current induces an electromotive
force in that circuit (self) or in a
neighbouring circuit (mutual) is called
inductance. The electrical component is called
an inductor.-20-
= total flux linkages
current through coil (3.24)
No
= a2
a (3.20)
= ky (NP A/2) (3.3)
where, N is number of coil turns
bp 18 magnetic flux (weber)
I” is current (ampere)
k, is a geometric factor
AY is coil's planar surface area (mm’)
% 46 coil's axial length (mm)
The self-inductance of a coil is proportional to coil
windings squared (N2)and planar surface area (A), and
inversely proportional to coil length (2),
INDUCTIVE REACTANCE: (symbol: X,, unit
ohm, 2)
Opposition to changes in alternating current
flow through a coil is called inductive
reactance.
Lt ob (34a)
or x, = 2neL (3-4)
where, £ is frequency of alternating current
(hertz), and w is angular frequency
(radians/second)
CAPACITIVE REACTANCE (symbol: Xe, unit.
Opposition to changes in alternating voltage
across a capacitor is called capacitive
reactance.
Eddy current coil capacitive reactance is
normally negligible. Hovever, capacitance can
be important when considering impedance of probe
cables (Sections 5.9 and 7.2.5).
Xe" 3G (3.5)
where, © 4s capacitance (farad)
IMPEDANCE: (symbol: Z, units: ohm, 2)
The total opposition to alternating current flow
is called IMPEDANCE. For a coil,
Iz] “Wee (3-6)-21-
x,
and @ = Arctan + (3.7)
where |Z|is magnitude of Z, and 6 is phase of Z
(described in next section).
SINUSOIDS, PHASORS AND ELECTRICAL CIRCUITS
In a direct current (DC) circuit, such as a battery and light
bulb, current and voltage are described completely by their
respective magnitudes, Figure 3.1(a). Analysis of
alternating current (AC) circuits is more complex. Since
voltage and current amplitude vary with time, the phase (or
time delay) relationship between them must also be taken into
account. A typical AC circuit, an inductor in series with a
resistor, is presented in Figure 3.1(b). This is a
simplified model of a probe assembly: the inductor is the
reactive part of the assembly (coil) while the resistor
models both coil wire and cable resistance. Figure 3.1(c)
shows voltage across the inductor (Vj) leads the current
(1) by 90°, while voltage across the resistor (Vg) is in
phase with current. Since the current is common to both
inductor and resistor, it is possible to use current as a
point of reference. Hence, we deduce the voltage across the
inductor leads the voltage across the resistor by 90°.
If one measures the voltage drop, Vz, across both the
inductor and resistor, we find Vp leads current (or Va)
by an angle less than 90°, as shown in Figure 3.1(d).
To evaluate the total voltage Vp, we add vectorially the
separate voltages Vp and Vz,
Ve = VR + Vy (3.8)
= I(R + jwL) (3-9a)
where j is a mathematical operator (rotates a vector
cow by 90°)
or Vp = IR sin ( wtt0) + j IWL sin (wetm/2) (3.9b)
Representing voltage waveforms as in Figure 3.1(d) or
equation 3.9(b) can be both time consuming and confusing. A
simpler voltage representation is available by means of
phasor diagrams. In phasor diagrams the voltage is
represented by its peak value (amplitude) and phase shift (8)
relative to the current. The two terms in equation 3.9(b)
both contain an amplitude and phase shift so they can be~22-
DIRECT cumReNT
velk
(2 sure eager e1neutr
ALTERING cue
‘yc anne Car 90°
(©) avremuarine consent crneurr
vats) resistance, R
(4) voLtAse Gaara SPLAT OF masons (Octane anon nisms
Representation of Direct Current and
Alternating Current Circuit Parameters- =23-
represented by phasors. The first term's amplitude is IR and
its phase shift is 0. The amplitude of the second term is IwL
and its phase shift is ™/2 or 90°. Each phasor can be
represented by an arrow starting at the origin. The phasor's
amplitude is indicated by the length of the arrow OP and the
phase shift by the direction of the arrow, see Figure 3.1(e).
Phasors are displayed graphically with the resistive
component (Va), having a phase shift 6 = 0, along the
horizontal axis. As @ increases the phasor rotates
counter-clockwise. The reactive component (V;), having a
phase shift 9 = 90°, will be represented along the vertical
axis.
Current is common to both voltage components and since V=1Z,
the voltage graph of Figure 3.1(e) can be converted to an
impedance graph display, as in Figure 3.1(f). If this
approach is applied to eddy current testing, it is found that
any changes in resistance or inductive reactance will cause a
change in the position of the end of the vector (point P)
which represents the total impedance vector.
To obtain the reactive and resistive components from this
graph requires knowledge of trigonometry.
Reactive component: X, 7 wh = [2] sin 0 (3-10)
Resistive componen R = |2| cos 8 (3.11)
Amplitude of impedance: [Z| = yk? + x;?
Phase angle: 6 = Arctan X,/R (3-7)
Note the x axis component represents pure resistance (phase
shift = 0°) while the y axis component represents pure
inductive reactance (phase shift = + 90°). In these
calculations it is assumed coil capacitance is negligible.
MODEL OF PROBE IN PRESENCE OF TEST MATERIAL
The test probe contains a cotl which when placed on or close
to a test sample can be considered as the primary winding of
a transformer. The field created by alternating current in
the coil induces eddy currents in the test sample which acts
as a single turn secondary winding,Ns = 1, Figure 3.2(b)-
Eddy currents align to produce a magnetic field which tends
to weaken the surrounding net magnetic flux $p, according
to Lenz's Law.(b)
©)
Q-\) ©
3 seconoary
& RECEIVE COIL,
2
(ce)
Fig. 3.2: Model of a Coil with Test Object
There are two methods of sensing changes in the secondary
current, Ig- The “impedance method” of eddy current
testing consists of monitoring the voltage drop across the
primary coil (V, = TpZp)- The impedance Z, is
altered by the Load b£"the secondary of the transformer.
Therefore, changes in secondary resistance, Rg, or
inductance Lg can be measured as changes in V5.
The “send-receive” method of eddy current testing uses two
coils. Eddy current flow in the sample is altered by defects
and these variations are detected by monitoring the voltage
across a secondary receive coil, see Figure 3.2(c).~25-
SIMPLIFIED IMPEDANCE DIAGRAMS
Derivation of Probe Impedance for Probe/Sample Combination
We now consider how changes in the test sample affect coil
impedance on the impedance graph display.
From the previous section the probe and test sample can be
modelled as a transformer with a multi-turn primary (coil)
and single turn secondary (sample), Figure 3.3(a). This
circuit can be simplified to an equivalent circuit where the
secondary circuit load is reflected as a resistive load in
parallel with the coil's inductive reactance, Figure 3.3(b).
This circuit is an approximate model of a real coil adjacent
to a conductor. It is assumed that all of the magnetic flux
from the primary coil links the test sample; the coupling is
perfect (100%). It is also assumed that there is no skin
depth attenuation or phase lag across the sample thicknes
OE Ome
(©) cavivacenr pana exneutt
(> coutvanr senies etacat?
Fig. 3.3 Equivalent Circuit
The equivalent circuit concept can be used to obtain
simplified impedance diagrams applicable to eddy current
testing. These diagrams serve as an introduction to the more
detailed diagrams which include variations caused by the skin
effect. The coil/sample circuit model can be transformed
into the simpler series circuit by the following mathematical
manipulations. The load resistance R, can be transferred
from the secondary back to the primary winding by multiplying
it by the turns ratio squared, (ap /¥g)?, Figure 3.3(b).The total impedance of this parallel circuit can be evaluated
and transformed into an equivalent series circuit as follows:
2 =
P
2
where 2, = NoRy
and 227 io»
Where Xo™ WL, coll inductive reactance in air,
ine x
Therefore 2p “zh
wR, +4X,
pes * 3X,
which transforms to
wep x? cv?r_)?x (3.128)
"ar Tt
P
ompR) +)? aR)? + Ox)
This can be viewed as a series combination, in the primary
circuit, of resistance Ry and inductive reactance X, or
z= R, + 45x 3.12b
pT Rt IX, (3.129)
The series circuit in Figure 3.3(c) is therefore fully
equivalent to the parallel one of Figure 3.3(b). Rp can be
considered as coil wire and cable resistance while
Zp=Ry+jXp is the total impedance of the probe/sample
combination.
When the probe is far from the sample (probe in air), Ry is
very large and by substituting R, = © into equation 3.124
results in
Ry=0, Xp"Xq and 2p"Xo: -27-
The above results can be obtained by removing component
NR, from Figure 3.3(b), since Rg= * implies an open
cbreuit.
One last transformation in the equation is required before
impedance graphs can be obtained. Equation 3.12(a) can be
simplified by setting
Cy = XG
where G = 1/NoR, 48 equivalent circuit conductance.
Substitution in 3.12(a) yields
KC x
oro 5 y 2
P 2 2
1+co 1+¢5
Normalizing with respect to Xo, the coil's inductive
reactance when far removed from the sample (coil in air)
results in
z cy 1
a z+3 z (3413)
0 «1tce +02
By varying Cy, in equation 3.13, from 0 to infinity the
impedance curve of Figure 3.4 is obtained. The impedance
locus is that of a semi-circle with center at Xp/Xo™k
and Ri/Xo = 0; its radius is b. With the help of
equation 3.13 and Figure 3.4, impedance changes can be
related to changes in the sample characteristics.
'
Ea onsearine rou
vom 260
MDCT
mcrae
RG
NORMALIZED RESISTANCE
Impedance Graph Displa-28-
3.5.2 Correlation Between Coil Impedance and Sample Properties
The effect of test parameter variations on probe impedance
can be derived from equation 3.13. Each parameter is
substituted in turn into Co=Xo/N3R,; if an increase in
the parameter results in an increase in C,, the operating
point (position on impedance diagram) moves DOWN the impedance
curve, if Cg decreases, the operating point moves UP the
impedance curve. These correlations are useful in obtaining a
qualitative appreciation of the effect of the various test
parameter: It is also useful to know that probe/sample
effects can be derived from the sinple equivalent parallel
circuit where the sample is treated as a resistor in parallel
with an inductor (coil). The complete effect can then be
obtained by adding the effect of 'phase lag’, which will be
treated in later chapters.
Study of equation 3.13 reveals the following:
An increase in R, results in a decrease in Cy.
Therefore an increase in resistance to eddy current flow
moves the operating point, P, UP the impedance curve
(along the semi-circle), see Figure 3.5(a).
2. Re = o8/A
where, p is electrical resistivity, 2 1s eddy current
flow distance and A is cross-sectional area to current
flow.
Therefore, = constant x Rg
An increase in electrical resistivity will move the
operating point UP the impedance curve. The opposite is
true for an increase in electrical conductivity. See
Figure 3.5(a).
3. For thin wall tubes or plates of thickness ¢,
Rg = p2/A = pnD/tw
and for constant probe or tube diameter, D, and coil
width, w,
Rg = constant/t
An increase in tube wall (or plate thickness) will move
the operating point DOWN the impedance curve, see Figure
3.5(b).-29-
Co = Who/NpRy = constant x &
for constant sample properties.
An increase in test frequency will move the operating
point DOWN the impedance curve, see Figure 3.5(c)-
Lo = constant x D?; probe inductance increases
proportional to probe or tube diameter squared.
Also Rg = p™D/tw = constant x D, for constant
thickness, t, and cpil width, v. Substituting L, and
R, into Cj = WLo/N5R, results in Co=constant x D.
An increase in probe diameter or tube diameter will
move the operating point DOWN the impedance curve, see
Figure 3.5(4).
In the equivalent circuit derivation perfect coupling wa:
assumed for sake of simplification. However, it can be
shown that when mutual coupling between coil and sample
is decreased, the impedance point traces smaller
semi-circles as Cg increases from 0 to infinity, see
Figure 3.5(e).
Prune
Wise
= DEGREASING FILL FACTOR
Om INCREASING LIFT-OFF
Tn
Fig 5: Simplified Impedance Diagrams- -30-
SUMMARY
The impedance method of eddy current testing consists of
monitoring the voltage drop across a test coil. The
impedance has resistive and inductive components; the
impedance magnitude is calculated from the equation
Iz] mye t (wL)“ , ohms (3.6)
and the impedance phase is calculated from
@ = Arctan 4 » degrees (3.7)
The voltage across the test coil is V = IZ where I is the
current through the coil and Z is the impedance.
A sample's resistance to the flow of eddy currents is
reflected as a resistive load and is equivalent to a
resistance in parallel to the coil inductive reactance. This
load results in a resistive and inductive impedance change in
the test coil. Coil impedance can be displayed on normalized
impedance diagrams. These are two-dimensional plots with the
inductive reactance displayed on the vertical axis and
resistance on the horizontal axis as in Figure 3.6.
wl
oly
\ for
NORMALIZED OPERATING POINT
INDUCT ANCE
REACTANCE
NORMALIZED RESISTANCE, a
whe
Fig. 3 Impedance Graph Display-31-
With this display we can analyze the effect of sample and
test parameters on coil impedance. The equivalent circuit
derivation of coil impedance is useful for a qualitative
understanding of the effect of various test parameters. It
is valid only for non-ferromagnetic material and for the
condition of no skin depth attenuation or phase lag across
the sample. (Ferromagnetic materials will be covered in
Section 9.4).
Note that all test parameters result in a semicircle display
as they increase or decrease. A resistance increase to the
eddy current flow or increase of sample's electrical
resistivity moves the operating point UP the impedance
diagram, i-e., increase in coil inductance and a change in
coil resistance.
An increase in a sample's electrical conductivity, thickness
or tube diameter, moves the operating point DOWN the
impedance curve. An increase in test frequency or probe
diameter also moves the operating point DOWN the impedance
curve. Although not shown in the above figure, a decrease in
£1ll-factor or increase in lift-off results in a decrease in
semicircle radius and a smaller change in coil impedance.
In some test requirements it is advantageous to operate at
specific locations on the impedance diagram. By choosing the
appropriate test parameters this is usually possible.
WORKED EXAMPLES
Probe Impedance in Air
PROBLEM: An eddy current test is carried out at a test
frequency of 50 kHz. Coil resistance is 15 ohms
while its inductance is 60 microhenries.
a) What is the inductive reactance of the test
coil?
b) What is the impedance of the test coil?
c) What is the angle, 0, between the total
impedance vector and the resistance vector?
SOLUTION:
a) Xp = 2 TEL = (27) (50 x 103) (60 x 107
Xz = 18.8 ohms
r?+ (2 TEL)? = Vis? + (18.8)?
24.1 ohms
b)
c) 9 = Aretan 2th .
Arctan 1.253
51.4 degrees=32-
3.7.2 Probe Impedance Adjacent to Sample
PROBLE
SOLUTION:
An eddy current test is carried out on brass using
a surface probe at 50 kHz. Coil resistance in air
is 15 ohms and its inductance in air is 60
microhenries. Probe impedance with the probe on
the brass sample is measured as Z, = 24.5 ohms
and 6 = 35 degrees.
Calculate: a) Xp, inductive reactance
and b) RE, resistive load
a) X= Z, sind
ep “Pp
24.5 sin 35° = 14.1 ohms
b) RL = 2, cos - Rye
= 24.5 cos 35° - 15.0 = 5.1 ohms
3.7.3 Voltage - Current Relationship
PROBLEM:
SOLUTION:
therefore,
For the above probe impedance problem calculate
voltage drop across the probe if test current is
100 milliamperes.
Probe impedance [Z| = 24.5 ohms
Ohm's Law states that = v= 1 |2|
V = (0.10) (24.5) = 2.45 volts.
Voltage across the probe is 2.45 volts.=33-
CHAPTER 4 - INSTRUMENTATION
INTRODUCTION
All the information about a test pert is transmitted to the
test coil through the magnetic field surrounding it. The
impedance eddy current method monitors voltage drop across
the primary coil, V, = IpZp} as coil impedance changes
80 will the voltage across the coil if current remains rea-
sonably constant. The send-receive eddy current method moni-
tors voltage developed across a sensing coil (or Hall effect
detector) placed close to the excitation coil, see Figure 2.2.
In most inspections, probe impedance (or voltage) changes
only slightly as the probe passes « defect, typically less
than 1%. This small change is difficult to detect by measur-
ing absolute impedance or voltage. Special instruments have
been developed incorporating various methods of detecting and
amplifying small impedance changes.
The main functions of an eddy current instrument are illus~
trated in the block diagram of Figure 4.1. A sine wave
foto
FASE [Link] METER
sensitive |
>
ac TO pe
AMPLIFIER CONVERTOR PHASE
(rus x ROTATION
FiLtering)—>—]
12, 27)
y
a
BALANCE XY
MONITOR
LU
OSCILLATOR
‘TRANSFORMER
Block Diagram of Eddy Current Instrument=34-
oscillator generates sinusoidal current, at a specified fre-
quency, that passes through the test coils. Since the impe-
dance of two coils is never exactly equal, balancing is
required to eliminate the voltage difference between them.
Most eddy current instruments achieve this through an AC
bridge or by subtracting a voltage equal to the unbalance
voltage. In general they can tolerate an impedance mismatch
of 5%. Once balanced, the presence of a defect in the vici-
nity of one coil creates a small unbalance signal which is
then amplified.
Since the sinusoidal unbalance voltage signal is too diffi-
cult and inefficient to analyze, it is converted to a direct
current (DC) signal retaining the amplitude and phase charac-
teristics of the AC signal. This is normally achieved by
resolving the AC signal into quadrature components and then
rectifying them while retaining the approximate polarity. In
general purpose instruments, these signals are normally dis~
played on X-Y monitors. Simpler instruments, such as crack
detectors, however, have a meter to display only the change
in voltage amplitude. To decrease electrical instrument
noise, filtering is used at the signal output; however, this
decreases the frequency response and thereby restricts the
inspection speed.
The most troublesome parameter in eddy current testing is
lift-off (probe-to-specimen spacing). A small change in
lift-off creates a large output signal. The various methods
used to decrease this effect are discussed in the individual
sections on specific eddy current instruments.
BRIDGE CIRCUITS
Most eddy current instruments use an AC bridge to sense the
slight impedance changes between the coils or between a
single coil and reference impedance. In this section the
important characteristics of bridge balance are discussed.
Simple Bridge Circuit
A common bridge circuit is shown in general form in Figure
4.2, the arms being indicated as impedance of unspecified
sorts. The detector is represented by a voltmeter. Balance
is secured by adjustments of one or more of the bridge arms.
Balance is indicated by zero response of the detector, which
means that points A and C are at the sane potential (have the
same instantaneous voltage). Current will flow through the
detector (voltmeter) if points A and C on the bridge arms are
at different voltage levels (there is a difference in voltage
drop from B to A and B to C). Current nay flow in either
direction, depending on whether A or C is at higher
potential.=35-
Fi
Common Bridge Circuit
If the bridge is made up of four impedance arms, having
inductive reactance and resistive components, the voltage
from B to A must equal the voltage from B to C in both
amplitude and phase for the bridge to be balanced.
At balance,
42 = 1222
and 1423 = 1224
from which the following relationship is obtained:
2, «2
(4.1)
sy
2 74
Equation 4.1 states that the ratio of impedances of one pair
of adjacent arms must equal the ratio of impedances of the
other pair of adjacent arms for bridge balance.
If the bridge was made up of four resistance arms, bridge
balance would occur if the magnitude of the resistors satisfies
equation 4.1 (with Z1 replaced with Rj, etc). However, if the
impedance components are eddy current probes consisting of both
inductive reactance and resistance, the magnitude and phase of
the impedance vectors must satisfy equation 4.1.- -36-
In practice, this implies the ratio of inductive reactance of
one pair of adjacent arms must equal the ratio of inductive
reactance of the other pair of adjacent arms; the same being
true for the resistive component of impedance.
Figure 4.2 and equation 4.1 can be used to illustrate the
characteristic 'figure 8’ signal of a differential probe. If
Z > zB, point c ts at a higher potential than point a.
This implies that when Z, increases (i.e., coil moving
across a defect) with Zz, 23 & Z, constant, the bridge
voltage unbalance increases,and the opposite happens when
23 increases. It is this bridge unbalance characteristic
that results in @ plus-minus or ‘figure 8' signal as the
differential probe moves across a localized defect. This
signal occurs independent of whether the two coils are wound
in opposition or in addition.
Typical Bridge Circuit in Eddy Current Instruments
Figure 4.3 illustrates a typical AC bridge used in eddy
current instruments. It is similar to the bridge in Figure
4.2 except for two additional arms. In this bridge the probe
coils are placed in parallel with variable resistors. The
balancing, or matching of voltage vector phase and amplitude,
is achieved by varying these resistors until a null is
achieved. Potentiometer R2 balances the reactive component
of the coils to make the phase angle of each coil circuit
equal. Potentiometer Ry balances the resultant voltage
with an equal voltage amplitude to null the instantaneous
voltage between Rj and R2-
TEST COIL
REFERENCE
Corl.
Fig. 4.3: Typical Bridge Circuit Used in Eddy
Current Instruments-37-
The inductive voltage drop across each coil is equalized by
controlling the current passing through the coils. This is
done by varying potentiometer Ry. However, when the test
coil inductance differs significantly from reference coil
inductance, potentiometer Ry will have to be rotated to one
extremity. This means less current passes through one coil
making it less sensitive than the other coil. When this
occurs, a distorted (unsymmetrical) signal results if a
differential probe is used. In addition, the common cable
lead carries the unbalanced current, resulting in cable
noise, especially if the cable is not properly shielded and
grounded.
In the Figure 4.3 circuit, the output voltage for large
(210%) off-null (off-balance) conditions is a nonlinear func-
tion of the change in coil impedance. However, for defect
detection, close to balance, this discrepancy is small.
Bridge Circuit in Crack Detectors
Portable eddy current instruments are often used to inspect
for surface defects. A typical crack detector circuit is
shown in Figure 4.4. An oscillator supplies an alternating
current to an AC Bridge, containing a single eddy current
probe coil as one arm of the bridge. A capacitor is connec-
ted in parallel with the coil so the L-C (inductance-
capacitance) circuit is near resonance. When the coil is
placed on a test sample, the bridge is unbalanced and the
pointer on the meter swings off-scale. The bridge can be
balanced by adjusting potentiometer Ry-
VARIABLE
RESISTOR
Vo(7Ny
‘OscILLaToR
FL Simplified Circuit of Crack Detector