Outside Micrometer Calibration Procedure
Memorial University Technical Services
Document No.: TS-0049
Revision: 3
1.0 Standards and Equipment
The following equipment is required:
Master Gauge Block Set
NOTE: Standards and equipment used must have a valid calibration certificate
2.0 Calibration Procedure
“TS-0071 Outside Micrometer Calibration Record Sheet” must be used.
Clean the Outside Micrometer’s measuring surfaces and the gauge blocks to be used.
NOTE: 1. Digital Outside Micrometers only need one scale to be verified.
2. Zero the Outside Micrometer at the start and adjust as required by the
manufacturers’ specifications. If you cannot zero it then use a gauge
block and adjust the micrometer to that value.
3. When testing the Outside Micrometer, one of the points must be near
the lower limit that the instrument can measure, another somewhere in
the middle, and the third near the upper limit.
4. Use a conversion factor of 25.40 mm/in to convert gauge block lengths
to Metric from Imperial.
Step 1:
Measure the ambient temperature and record it. If the temperature is <18oC or >24oC,
see the Division Manager for further instructions.
Page 2 of 2
Step 2:
Test Characteristic: Anvils and Thimble
Test Method: Visual, Touch
Acceptable Limit: No damage, nicks, or burrs. Should have straight and parallel
faces with no free play over the whole length
Step 3:
Test Characteristic: Ratchet
Test Method: Turn the ratchet with the instrument locked, to insure if it is
functional
Acceptable Limit: Good working order
Step 4:
Test Characteristic: Measuring Scale
Test Method: Using gauge blocks and the smallest anvil, measure and record
5 different lengths. For micrometers with more than one anvil,
make a separate measurement with each anvil
Acceptable Limit: +/- 0.0002” or 0.005 mm
Step 5:
Test Characteristic: Rod Inserts
Test Method: For micrometers with a large measuring range, and
different rod inserts, make one measurement for each rod
individually.
Acceptable Limit: +/- 0.0002” or 0.005mm