SMP-088: Preventive Maintenance Program Supplement For All Standard Plants
SMP-088: Preventive Maintenance Program Supplement For All Standard Plants
SMP-088
Region:
Canada
United States
Issued: 1 Aug 2012 Revised: 15 Dec 2017 Effective: 5 Apr 2018 Page 1 of 14
UNCONTROLLED WHEN PRINTED. PRAXAIR BUSINESS CONFIDENTIAL. RESTRICTED FOR INTERNAL USE ONLY.
Praxair Operations Standard Standard No. SMP-088
This Standard, together with the following examples, provides the complete PM
program for CO2 facilities:
• SMP-074, regarding environmental preventive maintenance.
• SMP-075, regarding personnel safety preventive maintenance.
• SMP-080, regarding preventive maintenance program.
• SMP-080A, regarding electrical equipment preventative maintenance.
• SMP-088A, regarding N-Plant preventative maintenance program.
• SMP-088B, regarding VPSA preventative maintenance program.
• SMP-088C, regarding Membrane preventative maintenance program.
• SMP-088D, regarding PSA preventative maintenance program.
Issued: 1 Aug 2012 Revised: 15 Dec 2017 Effective: 5 Apr 2018 Page 2 of 14
UNCONTROLLED WHEN PRINTED. PRAXAIR BUSINESS CONFIDENTIAL. RESTRICTED FOR INTERNAL USE ONLY.
Praxair Operations Standard Standard No. SMP-088
STDSPO-0218- Automatic Valve Inspection and Service POLICY 1 Years Variable YES YES Minor
01
STDSPO-0221- Monthly Backup System Unit Inspection * 1 Months Variable NO NO Minor
01
STDSPO- KMAC - Visual Inspection (Probe and Main POLICY 1 Years Variable NO NO Minor
0222.1-01 Regulator)
STDSPO- LTPP Probe Replacement on Kaye McDonald POLICY 4 Years Fixed NO NO Minor
0223.1-01 Modules
STDSPO-0224- Electric Heater Trim/ Dryer Contactor Replacement * 5 Years Fixed NO NO Minor
01
STDSPO-0226- Annual Backup System Unit Inspection POLICY 1 Years Variable NO NO Minor
01
STDSPO-0258- Instrumentation Maintenance POLICY 1 Years Duplicate YES YES Minor
01
STDSPO- 480 Volt Motor Connections * 1 Years Fixed YES YES Major
0500.0-01
STDSPO- Sullair Control System Maintenance * 1 Years Fixed YES YES Minor
1101.1-01
STDSPO- Sullair Compressor Oil Sample, Additive and * 3 Months Variable NO NO Minor
1103.1-01 Bearing Fluid Filter
STDSPO- Sullair Compressor Six Month Maintenance * 6 Months Fixed YES YES Minor
1104.1-01
STDSPO- Sullair Compressor Annual Maintenance * 1 Years Fixed NO NO Minor
1106.1-01
STDSPO- Carbon Bead Change Out * 9 Months Fixed YES YES Minor
1107.2-01
STDSPO- Pre-Filter Maintenance (Only Plants with a Purafil * 3 Months Fixed NO NO Minor
1123.1-01 System)
STDSPO- Sullair Control System Maintenance (Two (2) * 2 Years Fixed YES YES Minor
1124.1-01 Year)
STDSPO-1129- Sullair Instrumentation Maintenance POLICY 4 Years Fixed NO NO Minor
01
STDSPO-1131- Teledyne 3290 Analyzer Maintenance * 2 Years Fixed NO NO Minor
01
STDSPO- Compressor Oil Purifier Cartridge Change out * 6 Months Variable NO NO Minor
1160.2-01 (Only Plants with COP)
STDSPO- Shunt Trip Testing POLICY 2 Years Fixed YES YES Minor
1165.1-01
STDSPO-1220- Sullair Oil Stop Valve Complete Rebuild * 3 Years Fixed NO NO Minor
01
STDSPO-1221- Trim Heater Contactor Replacement * 5 Years Fixed NO NO Minor
01
Issued: 1 Aug 2012 Revised: 15 Dec 2017 Effective: 5 Apr 2018 Page 3 of 14
UNCONTROLLED WHEN PRINTED. PRAXAIR BUSINESS CONFIDENTIAL. RESTRICTED FOR INTERNAL USE ONLY.
Praxair Operations Standard Standard No. SMP-088
PM
Description Task Instructions
Schedule
STDSPO- Reciprocating Task Instructions:
0003.1-01 Compressor – 1. Inspect the distance piece purge blower filter and mufflers
Periodic 2. Verify the pressure differential on the suction and discharge screens. Clean / replace the screens as required
Inspection 3. Verify schedule of next planned gas side overhaul as determined by compressor capacity and measured wear
rate of rider and piston rings
a. The pressure packing, rider and piston rings should be changed every 2 years when the compressor is new,
after cylinder reconditioning or a change in ring material. Gas side overhauls can be extended after wear rate is
established and documented
b. For horizontal and angled cylinders, wear rate of the rider rings can be established by piston to cylinder
clearance. Verify corresponding piston ring wear rate by removal of piston & inspection of the rider and piston
rings
c. For vertical cylinders, piston to cylinder clearance checks are of little value. Wear rate should be established by
removal of the piston and inspection of the rider and piston rings
d. For Sulzer machines visually check oil scrapers and piston rod surfaces. Check guide bearing - piston rod
clearance
e. For SIAD compressors with bronze piston guide rings, inspect and verify that screws are staked
4. Review and adjust the schedule of the following PM’s (as applicable) to align with the timing of the gas side
overhaul
a. SMP-080 STD-0196 Compressor – Gas Side Inspection (Wet Gas Service)
b. SMP-080 STD-0197 Main Drive Coupling Inspection
c. SMP-080 STD-0198 Main Drive Coupling Service (Grease or Oil Filled)
d. SMP-080 STD-0201 Piston Rod Packing Inspection
e. SMP-080 STD-0202 Compressor Suction & Discharge Valve Inspections
f. SMP-080 STD-0203 Compressor – Frame Inspection and Maintenance - Minor
g. SMP-080 STD-0204 Compressor – Frame Inspection and Maintenance – Major
h. SMP-0888B STDVPSA-0201 SIAD Oxygen Compressor (Recip) – 40k Hour Bearing Replacement
Interval: 1 Year
STDSPO- Analyzer - NOTE: This is a Policy PM. Extending the interval beyond what is specified requires a Variance per SOP-000.
0041.2-01 Calibration NOTE: This PM displaces the SMP-080 policy PMs STD-0810 (QA Analyzers in NonMedical service), 0820
Check (Analyzers in Safety service), and 0831 (Analyzers in Equipment Protection service).
Tasks:
[ ] Calibrate analyzers:
Every Month (see SMP-091) - Analyzers in QA service. Record the calibration as per the Quality Assurance
Program (has specific forms to use).
Every Month (see SMP-091) - Analyzers in Safety service and record same (typically recorded using QA forms or
use SMP-091 attch 4)
Every 5 Years (see SMP-091) - Analyzers in Equipment Protection service and record same (typically recorded
using QA forms or use SMP-091 attch 4)
[ ] Log monthly analyzer calibration information in QSS.
Interval: 1 Month
Issued: 1 Aug 2012 Revised: 15 Dec 2017 Effective: 5 Apr 2018 Page 4 of 14
UNCONTROLLED WHEN PRINTED. PRAXAIR BUSINESS CONFIDENTIAL. RESTRICTED FOR INTERNAL USE ONLY.
Praxair Operations Standard Standard No. SMP-088
STDSPO- Isolation Purity NOTE: This is a Policy PM. Extending the interval beyond what is specified requires a Variance per SOP-000.
0109.1-01 Solenoid
Valves Tasks:
[ ] Inspect (stroke) Solenoid valves:
Every 2 years (see SMP-091) - solenoid valves in QA service and record same using SMP-091 attch 5.
Every 2 years (see SMP-091) - Solenoid valves in Safety service and record same using SMP-091 attch 5.
Every 5 years (see SMP-091) - Solenoid valves in Equipment Protection service and record same using SMP-091
attch 5.
As/if desired (see SMP-091) - Solenoid valves in any other service and record same as/if desired
NOTE: Solenoid valves may be exempt from the above inspections when the solenoid vlv is coupled to automatic
valve and addressed as one using STDSPO-0218 (in such cases 0218 and SMP-091 call for SMP-091 attch 5 to
be filled out once for the auto vlv/solenoid vlv because both vlvs are interdependant thus one record will suffice).
[ ] Every 5 years replace solenoid valves in QA, Safety or Equipment Protection services; these include, but are
not limited to: 1)Solenoids on plant isolation valves For Membranes - (SV-107 and SV-842.), For Basic
Membranes (SV-123), For N-Plants – liquid make valves, liquid addition valves. 2) Solenoids on pipelines (and
purge systems), For Membranes - (SV-846, SV-864a and SV-864b), For N-Plants – gas make valves
Interval: 5 Years
STDSPO- Purity Isolation NOTE: This is a Policy PM. Extending the interval beyond what is specified requires a Variance per SOP-000.
0117.2-01 Circuit Test NOTE: It is recommended that this PM be coordinated with STDSPO-0119 (B/U test).
NOTE: This PM displaces the SMP-080 policy PMs covering maintenance of all instrumentation in QA service
which includes STD-0212 (make isolation), 0300 (P/L isolation), and STD-0810, 0811, 0812, and 0813.
Tasks:
[ ] Confirm the calibration (STDSPO-0041) records exist and have not expired for the analyzers used in this
control loop (thus in QA service).
[ ] Confirm the inspection (STDSPO-0109) records exist and have not expired for the solenoid valves used in this
control loop (thus in QA service).
[ ] Confirm the inspection (STDSPO-0218) records exist and have not expired for the automatic valves used in
this control loop (thus in QA service).
[ ] Test purity isolation loop(s) in accordance with local quality system instructions.
Interval: 2 Years
STDSPO- Pipeline NOTE: This is a Policy PM. Extending the interval beyond what is specified requires a Variance per SOP-000.
0119.2-01 Backup System
Test NOTE: It is recommended that this PM be coordinated with STDSPO-0117 (B/U test) and STDSPO-0109 (purity
solenoid).
Tasks:
[ ] Test plant backup system(s) in accordance with local quality system instructions.
Interval: 2 Years
STDSPO- Review Plant Tasks:
0152.1-01 Equipment List [ ] Please verify the following: 1) Proper equipment is listed in EAM for the site. 2) Proper model and serial
numbers are assigned to the equipment.
[ ] Fill in any blank spaces - if information is unavailable, please state so by placing a dash (---) in the space.
[ ] Make any necessary changes (additions, deletions, revisions) on the Equipment List and send it back.
[ ] If no changes (additions, deletions, revisions) are needed, please note in the Comments section of the work
order.
Interval: 2 Years
STDSPO- PLC Replace Tasks:
0200.1-01 Batteries – [ ] For PLC5, Control Logix, MicroLogix, and Compact Logix: While Processor is on, disconnect the old battery
Processor / and replace with a new one. Note that some of the newer generation processors use a memory card instead of a
BASIC battery in these cases this PM is not required.
Module(s) [ ] For SLC and SLC Basic modules: Turn off power to the processor and remove the card from rack. Remove
the old battery and replace with a new one. Reinstall module in rack and energize. Card must be energized within
30 min or memory will be lost.
Interval: 3 Years
STDSPO- Moisture Tasks:
0207.1-01 Separator [ ] Check the high-moisture separator switch per vendor instructions and record same using SMP-091
Switch Test attachment 10.
Inspection
Interval: 1 Years
Issued: 1 Aug 2012 Revised: 15 Dec 2017 Effective: 5 Apr 2018 Page 5 of 14
UNCONTROLLED WHEN PRINTED. PRAXAIR BUSINESS CONFIDENTIAL. RESTRICTED FOR INTERNAL USE ONLY.
Praxair Operations Standard Standard No. SMP-088
Interval: 5 Years
STDSPO- Automatic NOTE: This is a Policy PM. Extending the interval beyond what is specified requires a Variance per SOP-000.
0218-01 Valve NOTE: This PM displaces the SMP-080 policy PM STD-0886 for inspection of Auto/Solenoid Valves.
Inspection and NOTE: Solenoid valves coupled to automatic valves may be addressed as one in valve the PM (in such cases
Service
0218 and SMP-091 call for SMP-091 attch 5 to be filled out once for the auto vlv/solenoid vlv because both vlvs
are interdependent thus one record will suffice).
NOTE: Understanding there are a number of automatic valves in the facility, the intent of this PM is not to have
every automatic valve inspected at once each time this PM is due, instead the intent is to inspect a collection of
automatic valves when the PM is due and Year after Year do the same for a different collection of automatic
valves. Each facility will need to manage how this is accomplished so eventually all automatic valves are
inspected as prescribed (possibly by plant sections for example, cold box 1, BLOC 2, AC 1, FGC, etc.). Automatic
and solenoid valves may be grouped together in one inspection record when their operation interdependent.
Tasks:
[ ] Review the inspection records to identify which Automatic/Solenoid valves are due for inspection according to
the following:
1) Every 1 year - those having an average cycle time < 5 min and record same using SMP-091 attch 5.
2) Every 2 years (see SMP-091) - those in QA service and record same using SMP-091 attch 5.
3) Every 3 years - those having an average cycle time 5 - 60 min and record same using SMP-091 attch 5.
4) Every 4 years (see SMP-091) - those in Safety service and record same using SMP-091 attch 5.
5) Every 5 years (see SMP-091) - those in Equipment Protection Service and record same using SMP-091 attch
5.
6) Every 7 years - those having an average cycle time > 60 min and record same using SMP-091 attch 5.
7) As/if desired (see SMP-091) - all others not addressed above and record same as/if desired (may use SMP-
091 attch 5).
[ ] Whenever conducting above inspections also look and check the following: 1)Conduct leak check (if possible),
2)Check valve packing for leakage, 3)Inspect mounting brackets, couplings, and linkages for tightness and signs
of wear, 4)Check coupling bolt torque to manufacturer’s specifications, 5)Inspect actuator for signs internal
leakage, 6)Clean/replace solenoid valves, 7)Inspect tubing and hoses for damage and signs of leakage, 8)Inspect
actuator stops—locked in place and properly adjusted to minimize sticking and leakage.
[ ] Whenever conducting above inspections on valves with positioners also look and check the following: 1)
Inspect positioners for proper operation and signs of leakage, lubricate moving parts of positioners (roller bearing,
arms, etc.). 2) Inspect and clean the positioners pilot/shuttle valve (if applicable). 3) Inspect the instrumentation
pressure relief device.
Interval: 1 Year
STDSPO- Monthly Tasks:
0221-01 Backup System [ ] Visually inspect the LTPP and main regulator: Complete backup system unit inspection per SMP-088
Unit Inspection Attachment 2. Note: plants may use Attachment 2 or modify site specific reading sheets.
Interval: 1 Month
STDSPO- KMAC - Visual NOTE: This is a Policy PM. Extending the interval beyond what is specified in SMP-315 requires a Variance per
0222.1-01 Inspection SOP-000.
(Probe and NOTE: This PM displaces the SMP-080 policy PM STD-0881 (KMACs).
Main
Regulator) Tasks:
[ ] Visually inspect and service the LTPP and main regulator and record same using SMP-315 Attachment 3.
Interval: 1 Years
Issued: 1 Aug 2012 Revised: 15 Dec 2017 Effective: 5 Apr 2018 Page 6 of 14
UNCONTROLLED WHEN PRINTED. PRAXAIR BUSINESS CONFIDENTIAL. RESTRICTED FOR INTERNAL USE ONLY.
Praxair Operations Standard Standard No. SMP-088
STDSPO- LTPP Probe NOTE: This is a Policy PM. Extending the interval beyond what is specified in SMP-315 requires a documented
0223.1-01 Replacement replacement plan (see note below) or a variance per SOP-000.
on Kaye NOTE: This PM displaces SMP-080 policy PM STD-0882 (for KMACs probe replacements).
McDonald NOTE: Replacement of probes may need to be spread across a number of years. Per SMP-315, probes may be
Modules left in service past four years to permit phased replacement.
NOTE: Probes are very fragile and cannot be adjusted, but may be placed back into service for another four years
if the probe passes a bench test.
Tasks:
[ ] Determine the LTPP setpoint by using the following sequence. 1) Check the plant P&ID for the LTPP setpoint
listed on the print (this is the setpoint needed for the probe). 2) If the setpoint is not marked on the P&ID, use the
setpoint that is stamped on the probe. 3) If there is no setpoint stamped on the probe, use -40°F. 4) If the LTPP
probe has a setpoint other than -20°F or -40°F, consult with your TSE for guidance.
[ ] After determining the correct setpoint, order a LTPP from the Buffalo Service Center at 716-879-2870. -20°F
probe JDE part number #10061130, -40°F probe JDE part number # 10049700
[ ] Install the LTPP Probe -To avoid a quality incident, and to ensure that we supply our customers with a quality
product, follow these best practices:
Coordinate replacing the LTPP probe with the customer.
Have two Praxair technicians on-site to perform the work.
[ ] Upon installation, perform the checks as illustrated in SMP-315 Attachment 4 and record likewsie in same.
[ ] Test the backup system to verify that it functions as intended.
[ ] Send the old LTPP probe back to the Buffalo Service Center. You have approximately two weeks to send the
probe back, or your plant is charged for a new probe.
Praxair, Inc 175 E. Park Dr Tonawanda, NY 14150 Attn: Bldg - 62 / Paul Howell
Interval: 4 Years
STDSPO- Electric Heater Tasks:
0224-01 Trim/ Dryer [ ] Replace the contactor in the trim heater.
Contactor [ ] Contact and consult with TSE as needed.
Replacement
Interval: 5 Years
STDSPO- Annual Backup NOTE: This is a Policy PM. Extending the testing of this PM beyond what is specified requires a Variance per
0226-01 System Unit SOP-000.
Inspection
Tasks:
[ ] Perform annual backup inspection using the Customer Service Forms on Sharewaves:
http://airwaves.praxair.com/sites/naig/opsvc/de/cpe/info/Forms/Forms/AllItems.aspx
[ ] Nitrogen and Oxygen tanks use form # P-15-104 (Rev Q or latest revision). NOTE: As detailed in P-15-104:
check Hex 6-6 vaporizer inlet welds for leaks. Attention: Small leaks can take up to 10 minutes to form bubbles.
During inspection, check the piping leading to the Hex-6-6 vaporizer to confirm that there is at least a 12-inch
vertical run of pipe, or flex hose of any length, in the vaporizer inlet line.
[ ] 86C Customers use form # P-15-10
[ ] High pressure pumps include form # P-15--112
[ ] Gas receivers include form # P-15-108
[ ] Ground pumps include form # P-15-111
[ ] If you have additional equipment that is not mentioned above. Then look for that equipment form on the site.
[ ] Keep current hard copy on-site
Interval: 1 Years
STDSPO- Instrumentation NOTE: This is a policy PM. Extending the interval beyond what is specified in SMP-091 requires a Variance per
0258-01 Maintenance SOP-000.
NOTE: This PM displaces the SMP-080 policy PMs STD-0831, 0832, and 0833 for inspection, calibration and
function test of instrumentation in Equipment Protection service.
NOTE: The following instruments are exempt for this PM: All analyzers (covered in STDSPO-041). All
automatic/solenoid valves (covered in STDSPO-0218/0109/1129). All other instrumentation covered in STDSPO-
1129. All instrumentation in QA service (covered in STDSPO-0117). All instrumentation in Safety service (covered
in SMP-080 STD-0294, 0295, and 0296).
Tasks:
[ ] Identify the instrumentation in Equipment Protection service and complete the inspections, calibrations and
functional tests due per SMP-091.
Interval: 1 Year
Issued: 1 Aug 2012 Revised: 15 Dec 2017 Effective: 5 Apr 2018 Page 7 of 14
UNCONTROLLED WHEN PRINTED. PRAXAIR BUSINESS CONFIDENTIAL. RESTRICTED FOR INTERNAL USE ONLY.
Praxair Operations Standard Standard No. SMP-088
Interval: 1 year
STDSPO- Sullair Control Tasks:
1101.1-01 System [ ] For Membranes only: Check instrumentation for leaks - Transmitters and transducers, - E/P (605, 227, 108)
Maintenance and/or I/P (605, 227, 108)
[ ] For N-Plants and Membranes:
Functionally test the thermistor board. (Contact the TSE for the procedure if needed.)
Also verify and functionally test the high-discharge temperature shutdown switch using the attached procedure
(Non Supervisor plants).
Verify and functionally test the high condensate level shutdown switch.
Interval: 1 Years
STDSPO- Sullair Tasks:
1103.1-01 Compressor Oil [ ] Replace the bearing fluid filter if the dP is higher than 10 psid or higher than original installation psid, or
Sample, change once every 6 months.
Additive and [ ] Take oil sample following SMP-072
Bearing Fluid [ ] Add oil additive if machine has a COP System
Filter
For Sullair LS16 (1) Quart Gallon of PuroLube X-Tend (1gallons per year)
For Sullair TS20 (2) Quarts (2) Gallon of PuroLube X-Tend (4 gallons per year)
For Sullair TS32 (1) Gallon of PuroLube X-Tend (4 gallons per year)
Interval: 3 Month
STDSPO- Sullair Tasks:
1104.1-01 Compressor [ ] Replace the control line filter on the instrument line.
Six Month [ ] Grease the air compressor motor bearings per the manufacturer’s recommendations. Always read the
Maintenance nameplate and check the documentation to ensure that the proper grease and procedure is used. - Do not over
grease the bearings. - Remove the bottom plug, and make sure that the grease can pass out the plug port as new
grease is added.
[ ] Lubricate the Sullicon control valve linkage.
[ ] Inspect the Sullicon control valve diaphragm for leaks, cracks, deterioration, and replace as necessary.
[ ] Inspect the fin fan blades for cracks.
[ ] For M30 and M30HP (only), grease the Sullair enclosure fan motor (if applicable)
Interval: 6 Months
Issued: 1 Aug 2012 Revised: 15 Dec 2017 Effective: 5 Apr 2018 Page 8 of 14
UNCONTROLLED WHEN PRINTED. PRAXAIR BUSINESS CONFIDENTIAL. RESTRICTED FOR INTERNAL USE ONLY.
Praxair Operations Standard Standard No. SMP-088
Interval: 1 Years
STDSPO- Carbon Bead Tasks:
1107.2-01 Change Out [ ] Change out the carbon using Calgon carbon WS480 beads.
[ ] Open the blowdown valve on startup.
Interval: 9 Months
STDSPO- Pre-Filter Tasks:
1123.1-01 Maintenance [ ] Sample the air compressor pre-filter media and send to Clark Air Attn: Chris Clark 129 Elm Street East
(Only Plants Aurora, NY 14052 Phone number 716-652-7870
with a Purafil [ ] Send E-Mail to: [email protected] and notify him that a sample is in the mail.
System) [ ] Remove module near the center of the pass, and take out the rubber plug on side of module. Pour a
reasonable amount of media from the module into a clean container. Mix thoroughly, fill the provided plastic
Purafil sample bags halfway, and label each sample bag for the correct pass. Repeat for additional passes. Note:
A Media Life Analysis (MLA) report is provided after the quarterly sample is submitted; it indicates the remaining
life on the modules (Bed 1 and Bed 2).
[ ] Replace the modules in the equipment when the MLA is below 25 percent. Use the information provided in the
second tab to order new modules for the equipment.
[ ] Review the report to determine how to rearrange the new modules and discard the used modules. If the first
stage is below 25 percent but the second stage is above 25 percent, remove the first stage and discard. Move the
second stage modules into the first stage, and install the new modules into the second stage.
[ ] Inspect the unit for leaks.
Interval: 3 Months
STDSPO- Sullair Control Tasks:
1124.1-01 System [ ] Check the calibration - calibrate if necessary: - Load and unload switch - High pressure shutdown switch
Maintenance
(Two (2) Year) Interval: 2 Years
STDSPO- Sullair NOTE: This PM displaces the SMP-080 policy PM STD-0831, 0832, and 0886 for inspection and calibration of
1129-01 Instrumentation instrumentation in Equipment Protection service.
Maintenance NOTE: As per SMP-091, the instruments identified below are considered to be in Equipment Protection.
Tasks:
[ ] Every 5 years (see SMP-091) - Calibrate the following and record same in SMP-091 attch 4: - Transmitters
and transducers. - E/P (605, 227, 108) and/or I/P (605, 227, 108).
[ ] Every 5 years (see SMP-091) - Inspect (Check/Stroke) the following automatic valves and record same in
SMP-091 attch 5: - Spiral valve, - Sullicon control valve, - Membrane product valve
[ ] Every 4 years (see SMP-091) - Calibrate the high discharge temperature shutdown and record same in SMP-
091 attch 4.
Interval: 4 Years
Issued: 1 Aug 2012 Revised: 15 Dec 2017 Effective: 5 Apr 2018 Page 9 of 14
UNCONTROLLED WHEN PRINTED. PRAXAIR BUSINESS CONFIDENTIAL. RESTRICTED FOR INTERNAL USE ONLY.
Praxair Operations Standard Standard No. SMP-088
Interval: 2 Years
STDSPO- Compressor Oil Tasks:
1160.2-01 Purifier [ ] Change the COP cartridges:
Cartridge - Every 6 months, Sullair TS20 / COP 30, (1) PE-118-XL element
Change out - Every 6 months, Sullair TS32 / COP 60, (1) PE-127-XL element
(Only Plants
with COP) NOTE: EMAIL Fluid Metrics at [email protected] to order COP Elements
NOTE: This PM is frequency is changed to every 6 months except for LS16 machines on membranes
Interval: 6 Months
STDSPO- Shunt Trip NOTE: This is a Policy PM. Extending the interval beyond what is specified requires a Variance per SOP-000.
1165.1-01 Testing NOTE: You must be Level IV trained on electrical safety to complete this task.
Tasks:
[ ] Shut down the compressor using the OIU.
[ ] Disconnect the T/C wires from the SPA 2 unit.
[ ] Wire your Fluke 733 or equivalent instrument to the T/C input terminals on the SPA2 unit.
[ ] The temperature setpoint should be set between 250 and 275 degrees.
[ ] Raise the input temperature on your instrument until the breaker shunt trips.
[ ] If the breaker did not shunt within the range of 250 and 275 degrees, then troubleshoot why the breaker never
shunted. After repairing the loop, test it again.
[ ] Log the calibration data under the Comment section in the EAM database.
NOTE: If you need the program for the shunt trip, you can call your Membrane TSE.
Interval: 2 Years
STDSPO- Sullair Oil Stop Tasks:
1220-01 Valve Complete [ ] Rebuild the oil stop valve using the complete kit # 001680 which includes cylinder, piston, quad ring, nut, and
Rebuild o-ring.
[ ] Ensure proper integrity and alignment of the flexible process gas discharge piping assembly.
Interval: 3 Years
STDSPO- Trim Heater Tasks:
1221-01 Contactor [ ] Replace the contractor in the trim heater.
Replacement
Interval: 5 Years
Issued: 1 Aug 2012 Revised: 15 Dec 2017 Effective: 5 Apr 2018 Page 10 of 14
UNCONTROLLED WHEN PRINTED. PRAXAIR BUSINESS CONFIDENTIAL. RESTRICTED FOR INTERNAL USE ONLY.
Praxair Operations Standard Standard No. SMP-088
5.0 PM Revisions
This section will contain PMs revised that had numerous edits requiring the use of
MicroSoft® Word’s edit tracking functions to aid the editorial team in vetting their
updates over the lengthy (10-12 months) editorial process.
Issued: 1 Aug 2012 Revised: 15 Dec 2017 Effective: 5 Apr 2018 Page 11 of 14
UNCONTROLLED WHEN PRINTED. PRAXAIR BUSINESS CONFIDENTIAL. RESTRICTED FOR INTERNAL USE ONLY.
Praxair Operations Standard Standard No. SMP-088
Issued: 1 Aug 2012 Revised: 15 Dec 2017 Effective: 5 Apr 2018 Page 12 of 14
UNCONTROLLED WHEN PRINTED. PRAXAIR BUSINESS CONFIDENTIAL. RESTRICTED FOR INTERNAL USE ONLY.
Praxair Operations Standard Standard No. SMP-088
b. Apply a self-binding rubber tape to eliminate any sharp edges that may exist
with the bolted connection. 3M 130C tape is an example of such tape. If
varnished cambric tape is used in the first step, there may not be a need for
the rubber tape to smooth out edges.
c. Cover the full surface of the connection with half-lapped layer of vinyl tape
(3M 33+) to provide moisture barrier and insulating medium. The layers of
tape shall be installed without voids to preclude any contaminants from
entering the connection.
d. In lieu of the above, several commercially available motor termination kits
may be used, provided that they are sized correctly and are used in
accordance with manufacturer's instructions.
Interval: 1 year
6.0 References
SMP-074 regarding PM Program for Environmental compliance
SMP-075 regarding PM program for Personal Safety
SMP-080 regarding PM program for All Facilities
SMP-080A regarding PM program for Electrical Equipment
SMP-080B regarding Canadian PM program
SMP-088A regarding PM pogram for N-Plants
SMP-088B regarding PM program for VPSA Plants
SMP-088C regarding PM program for Membrane Plants
SMP-088D regarding PM program for N-PSA Plants
SMP-0324 regarding PM program for Pipelines
SOP-103 regarding master PM revisions
SOP-177 regarding documenting Work Orders In EAM
SOP-193 regarding EAM use for facility maintenance
Revisions
Date Author Description
1 Aug 2012 Jim Pfeil, Edward Initial Release.
M. Brown III,
Kimberly Henry
Bob Wojcik,
Kellina Anderson,
Tom Hart
26 Mar 2013 Tom Hart, Changed STDSPO-1107.
Jim Pfeil Made STDSPO-0041 Policy.
Mike Schunk
1 Oct 2013 Tom Hart Changed STDSPO-1107.
Jim Pfeil Made STDSPO-0041 Policy.
Chris Kulagowski Added STDSPO-0226-01 annual backup pm.
Changed STDSPO-0221. Added the word “Monthly” to the title
Added section 3.0.
STDSPO-0225 deleted (replaced by STD-0313 (see Alert US-2013-
03).
12 Jun 2014 Tom Hart STDSPO-1160, edited title and updated task instructions.
Jim Pfeil
Issued: 1 Aug 2012 Revised: 15 Dec 2017 Effective: 5 Apr 2018 Page 13 of 14
UNCONTROLLED WHEN PRINTED. PRAXAIR BUSINESS CONFIDENTIAL. RESTRICTED FOR INTERNAL USE ONLY.
Praxair Operations Standard Standard No. SMP-088
Issued: 1 Aug 2012 Revised: 15 Dec 2017 Effective: 5 Apr 2018 Page 14 of 14
UNCONTROLLED WHEN PRINTED. PRAXAIR BUSINESS CONFIDENTIAL. RESTRICTED FOR INTERNAL USE ONLY.