ECT Inspection Technique: Setup and Calibration
ECT Inspection Technique: Setup and Calibration
To optimize depth-curve measurements when calibration flaws deviate from standard depths (20%, 40%, 60%, 80%), make targeted corrections. Access the Analysis setup, navigate to Calibration Points, and double-click on depth values requiring correction to edit them. This approach allows for precise adjustments to match actual flaw metrics, ensuring accurate representation and analysis of defect sizes during inspection .
Depth values of calibration points might need modification during ECT setup when depths of calibration flaws do not align with expected values (20%, 40%, 60%, 80%). In such cases, optimize depth-curve measurements by accessing the Analysis setup option and altering depth values in the Calibration Points tab. This corrective measure ensures accurate depth assessments by editing incorrect values for all defects requiring adjustment .
During manual calibration of an Eddy Current Testing (ECT) setup, adjust rotation and gain by selecting the Hole defect with red cursors and using the impedance plane. To set rotation, press and hold the CTRL key, and left-click and drag the mouse. To adjust gain, perform the same action but use the right-click. These actions display the phase and amplitude in the upper-left corner of the impedance plane, providing a visual reference for making precise adjustments .
The calibration process for channels and depth curves in ECT setup entails inserting the probe into a calibration tube and ensuring placement away from defects. Balance the probe and start the acquisition by moving it through the tube. For defect signal calibration, use the System Calibration icon, select the curve channel and defect signal with red cursors, match it to the reference name, and click Match Maximum. Repeat for all reference defects, including measuring the support plate for the MIX channel. Partial calibrations can be marked if not all defects are measured. Finalize by clicking Calibrate .
The configuration of channels and signals in an ECT setup involves several key steps. After selecting appropriate frequencies and channels, you typically use 4 frequencies for differential and absolute coils. Select the channel for the MIX1 signal, using Channel 1 as D90 and Channel 2 as D90/2 by default. If needed, rename each channel or manually edit the frequency value. Configure depth curves by associating default codes with defect characterizations and adjust the channel settings accordingly for optimized signal measurements during inspections .
If calibration fails during ECT setup, investigate the issue by examining calibration details. Click Details, navigate to the Processed Channels tab, and identify problematic channels, often indicated by a gain value of 15. If necessary, increase the gain in MS5800 or TC5700 settings to resolve gain-related issues. Further, revise calibration measurements to ensure all defects in the tube, including the support plate for the MIX channel, have been accounted for. For partial calibrations, select the Partial calibration check box and recalibrate as needed .
To load an existing ECT setup, click the Setup menu, select Load, then use the scroll bar to find the desired setup. Click the setup name you wish to load and then click Load. This process enhances efficiency by saving time on configuration, allowing technicians to quickly re-use previously optimized parameters and settings for consistent and reliable inspection outcomes across similar tests .
To set up an Eddy Current Testing inspection technique, follow these steps: First, click the Setup menu and select Wizard. Enter a name and description for the setup, then choose the equipment to be used for the inspection. Next, select Eddy Current technology and click Create. For frequency selection, choose the material to inspect from the Material list, input the wall thickness of the tube, and the F90 frequency will be automatically calculated. Adjust if needed. Select Bobbin or AC in the Type list and Internal or External in the Reference list, as well as the appropriate Connector type (4pins or 41pins). For channel selection, choose the frequencies by checking the appropriate boxes, usually using 4 frequencies for differential and absolute coils. Assign channels for the MIX1 signal, with Channel 1 typically as D90 and Channel 2 as D90/2, and rename channels if desired. Select calibration points detected in the calibration tube (hole, ID groove, OD defect) and define depth curves using default codes for defects (e.g., PIT, WLL). Follow ASME standard for calibration, setting differential and absolute signals on significant defects like the Hole at 40 deg and 1 V, and consider calibrating large defects separately. Finally, finish the setup by clicking Finish, or load a previously stored setup if preferred .
Using a LP (Low Pass) filter in the ABS channels during ECT setup reduces noise by attenuating high-frequency components that do not contribute to the necessary measurement, thus clarifying the signal. By entering a cutoff frequency and applying the filter, it allows only frequencies below this threshold to pass through, minimizing interference and enhancing the signal-to-noise ratio, resulting in clearer defect detection .
Calibrating large defects separately during ECT setup is important to prevent saturation of the absolute channel, especially when dealing with large defects, as it might distort the signal. This precaution ensures the absolute channel is correctly set for accurate readings. Calibration can be done on a larger diameter hole or on an IDGROOVE at specific settings (e.g., 10 deg and 5 V), thus maintaining the integrity of the setup and allowing reliable test outcomes .