U 933 Humidity Controller

This calibration certificate summarizes the calibration of a Shivali humidity indicator with sensor. The instrument was calibrated at ARC Testing Laboratory on June 2, 2022 against a digital thermo-hygrometer standard. Measurements were taken at five humidity points between 10-95% RH at 25°C. All measurements for the device under calibration were within ±0.89% of the standard values. The next calibration is recommended for June 1, 2023.

Uploaded by

Amit Kumar
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
667 views17 pages

U 933 Humidity Controller

This calibration certificate summarizes the calibration of a Shivali humidity indicator with sensor. The instrument was calibrated at ARC Testing Laboratory on June 2, 2022 against a digital thermo-hygrometer standard. Measurements were taken at five humidity points between 10-95% RH at 25°C. All measurements for the device under calibration were within ±0.89% of the standard values. The next calibration is recommended for June 1, 2023.

Uploaded by

Amit Kumar
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd

Calibration Certificate

ULR No.: CC309622000000933F

Certificate No.: CCTPL/RPM/0116/41 Field: Thermal


SRF No.: 100/22 Date of Receipt: 02/06/2022

SRF Date: 29/05/2022 Date of Calibration: 02/06/2022


Company Name
& Address M/s. ARC Testing Laboratory Recommended Due Date.:
01/06/2023
Kh. No.42/9/4, Village-Badli, Near Astha (As per agreed by customer)
Hospital, New Delhi-110042 Certificate Issue Date: 05/06/2022

DUC* Details

Instrument Name: Humidity Indicator With Sensor Instrument Sr. No.: ---

Make: Shivali Instrument Id No.: ATL/HC-02

Model No.: --- Location: Humidity Chamber

Range: 10-100 % RH DUC Condition: OK

Readability: 1% RH Calibration Performed at: At Lab

Standard Equipments Used(Traceable to National Standard)

Sr. No Instrument Name Cert. No Cal. By Calibration Due Date

Dig. Thermo-Hygrometer With


1 FL/C/TH/25082021-C002 Fare Lab 27/08/2022
Sensor

Temperature Humidity Reference Standard Calibration Procedure No.


Environment Condition
25±5ºC 55±10%RH DKD-R5-7 CCTPL/TH/CP/04S

Calibration Result
Humidity (%RH) @ 25 °C
Sr. No. Expanded Uncertainty ±
Standard (%RH) DUC* (%RH) Deviation (%RH)
(%RH)

1 10.13 10 -0.13 1.6

2 50.28 50 -0.28 1.6

3 70.47 70 -0.47 1.6

4 90.68 90 -0.67 1.6

5 95.89 95 -0.89 1.6


Note : DUC* : Device Under Calibration
Uncertainty of measurment at approx 95% Confidence Level with coverage factor k = 1.96.

Calibrated By : Amit Pal Approved By : Sandeep Sharma


Calibration Engineer Director

Note. [Link] Certificate only to the particular instrument submitted for calibration.
[Link] Calibration Result reported in this certificate valid at the time of and under the stated condition of measurement.
[Link] particular certificate can not be reproduce except in full written, without prior permission of CCTPL.

`--- End of Certificate --- Page 1 of 1


Date Entry Sheet f
Env. Temp. & Humidity Before Env. Temp. & Humidity After
25.4 °C 55 %RH 26.5 °C 49 %RH

Instument Name Make Model


Dig. Thermo-Hygrometer with
Master Instrument Sensor
Rotronic HP32

DUC Name Humidity Indicator ---- ---

Humidity (%RH) @ 25°C

10 %RH 50 %RH 70 %RH 90


Standard
Calibraiton Standard DUC Standard DUC Standard DUC Standard
Point Readings Reading Readings Reading Readings Reading Readings

%RH %RH %RH %RH %RH %RH %RH %RH

10.06 10 50.22 50 70.42 70 90.62


10
10.08 10 50.25 50 70.44 70 90.64
10.09 10 50.26 50 70.45 70 90.65
50
10.11 10 50.26 50 70.46 70 90.65

10.12 10 50.28 50 70.47 70 90.66


70
10.14 10 50.29 50 70.48 70 90.68
10.15 10 50.29 50 70.48 70 90.69
90
10.16 10 50.31 50 70.49 70 90.71

10.16 10 50.32 50 70.51 70 90.72


95
10.18 10 50.33 50 70.52 70 90.73
Average 10.13 10 50.28 50 70.47 70 90.68
Date Entry Sheet for DUC & Observation
[Link]. SRF No Cal. Date
At Site 44/20 06/06/2020
Model Range L. Count Sr. NO.

HP32 0 to 50 °C & 10 to 95%RH 0.01°C & .01%RH

--- Upto 99 %RH 1% RH ---

%RH 95 %RH
DUC Standard DUC Reading
Reading Readings Error Resolution Uniformity

%RH %RH %RH %RH %RH %RH

90 95.84 95
-0.125 0.1 0
90 95.85 95
90 95.86 95
-0.281 0.1 0
90 95.87 95

90 95.88 95 -0.472 0.1 0


90 95.89 95
90 95.91 95
-0.675 0.1 0
90 95.92 95

90 95.93 95
-0.890 0.1 0
90 95.95 95
90 95.89 95
Instrument ID Inst. Loc.

CCTPL/TH/R-1914

JEH/OT-4/HM-01 OT-4 (Ist Floor)

Stability Master Expanded


Uncertainty Uncertainty

%RH ±%RH ±%RH

0 0.6 0.59

0 0.6 0.59

0 0.6 0.59

0 0.6 0.59

0 0.6 0.59
Calibration Observation Sheet for Thermal
Calyss Calibration & Testing Pvt. Ltd.
CCTPL/QF/6.4/05/T-01
Issue Date:08/07/2019 Amd Date:00
F-40, Sector-9, Noida-201301,UP
Issue No: 01 Amd No: 00

1 S.R.F. No.: IQC 2 Job Order Number:


3 Calibration Date: 02//06//2022 4 Rcoommended Date of Next Calibration : 01//06//2023
5 Instrument Details(DUC) :
Shivali
Name: Humidity Indicator With Sensor Readablity : 1% RH Make/Model :
---

Range : 10-100 % RH ID/[Link]: ATL/HC-02 / --- Location: At Lab

6 Instrument Details (Standard):


Dig. Thermo-Hygrometer With
Name: ID/[Link]: CCTPL/TH/R-1913 Calibrated By: Fare Lab
Sensor
Make/Model : Rotronic/HP-32 Cert. No: FL/C/TH/25082021-C002 [Link] upto: 27//08//2022
7 Instrument Details (Source):
Name : Err:509 ID/Sr No.: CCTPL/TH/R-1915 Make/Model : Enrico Sceintific
8 Enviromental Condition :
Before: 25.4 Before: 55
Temperature (ºC): Humidity(%RH):
After : 26.5 After: 49

9 Observations:
Set Value: 20 ˚C Set Value: 25 ˚C Set Value: 30 ˚C
Display Value: Display Value: Display Value:
Sr. No.
Reading of Std Reading of DUC Reading of Std Reading of DUC Reading of Std Reading of DUC
(˚C) (˚C) (˚C) (˚C) (˚C) (˚C)

1 30.17 31.0 50.42 51.0 89.81 91.0


2 30.21 31.0 50.42 51.0 89.83 91.0
3 30.25 31.0 50.48 51.0 89.86 91.0
4 30.25 31.0 50.48 51.0 89.86 91.0
5 30.25 31.0 50.53 51.0 89.88 91.0
6 30.26 31.0 50.54 52.0 89.93 91.0
7 30.28 31.0 50.58 52.0 89.93 92.0
8 30.30 31.0 50.58 52.0 89.96 92.0
9 30.30 31.0 50.61 52.0 89.99 92.0
10 30.33 31.0 50.60 52.0 90.16 92.0

Remarks: ..........................................................................................................................................................

Calibrated by: Checked by:

Prepared By: Approved By: Issued By:


Quality Manager Director Quality Manager
Calibration Observation Sheet for Thermal
Calyss Calibration & Testing Pvt. Ltd.
CCTPL/QF/6.4/05/T-01
Issue Date:08/07/2019 Amd Date:00
F-40, Sector-9, Noida-201301,UP
Issue No: 01 Amd No: 00

10 Observations:
Set Value: 50 ˚C Set Value: 0 ˚C Set Value: 10 %RH
Display Value: Display Value: Display Value:
Sr. No.
Reading of Std Reading of DUC Reading of Std Reading of DUC Reading of Std Reading of DUC
(˚C) (˚C) (˚C) (˚C) (%RH) (%RH)

1 0.00 0 0.00 0 10.06 10


2 0.00 0 0.00 0 10.08 10
3 0.00 0 0.00 0 10.09 10
4 0.00 0 0.00 0 10.11 10
5 0.00 0 0.00 0 10.12 10
6 0.00 0 0.00 0 10.14 10
7 0.00 0 0.00 0 10.15 10
8 0.00 0 0.00 0 10.16 10
9 0.00 0 0.00 0 10.16 10
10 0.00 0 0.00 0 10.18 10

11 Observations:
Set Value: 50 %RH Set Value: 70 %RH Set Value: 90 %RH Set Value: 95 %RH
Disp. Value: Disp. Value: Disp. Value: Disp. Value:
[Link].
Reading of Reading of Reading of Reading of Reading of Reading of Reading of Reading of
Std (%RH) DUC(%RH) Std (%RH) DUC(%RH) Std (%RH) DUC(%RH) Std (%RH) DUC(%RH)
1 50.22 50 70.42 70 90.62 90 95.84 95
2 50.25 50 70.44 70 90.64 90 95.85 95
3 50.26 50 70.45 70 90.65 90 95.86 95
4 50.26 50 70.46 70 90.65 90 95.87 95
5 50.28 50 70.47 70 90.66 90 95.88 95
6 50.29 50 70.48 70 90.68 90 95.89 95
7 50.29 50 70.48 70 90.69 90 95.91 95
8 50.31 50 70.49 70 90.71 90 95.92 95
9 50.32 50 70.51 70 90.72 90 95.93 95
10 50.33 50 70.52 70 90.73 90 95.95 95

Remarks: ….......................................................................................................................................................

Calibrated by: Checked by:

Prepared By: Approved By: Issued By:


Quality Manager Director Quality Manager
Uncertainity Calculation Sheet
1. Laboratory : CCTPL

2. Standard / Equipment Used for Calibration

Details of Standard Used Standard1 Source


Instruments Name : Dig. Thermo Hygrometer with Sensor Temp. Humidity Calibrator
Manufacturer/Model : Rotronic/ HP-32 Scientific Enirco
[Link]./[Link]: CCTPL/TH/R-1913 CCTPL/TH/R-1915
Range : 10 to 50 °C & 10 to 95%RH (-5 to 95)˚C & 10 to 95 %RH
Certificate No. FL/CAL/TH/22082019-C002
CalibrationDate&validity 28.08.2019 & 28.08.2020
Traceability: Farelabs Gurugram

[Link] Condition During Calibration


Temperature 25±2 ˚C Humidity 55±10 % RH

4. Evaluation
20 °C 25 °C 30 °C 50
Standard Standard DUC Standard DUC Standard DUC Standard
Calibraiton Readings Reading Readings Reading Readings Reading Readings
Point

°C °C °C °C °C °C °C °C
30.170 31.000 50.420 51.000 89.810 91.0 0
20 30.210 31.000 50.420 51.000 89.830 91.0 0
30.250 31.000 50.480 51.000 89.860 91.0 0
25 30.250 31.000 50.480 51.000 89.860 91.0 0
30.250 31.000 50.530 51.000 89.880 91.0 0
30 30.260 31.000 50.540 52.000 89.930 91.0 0
30.280 31.000 50.580 52.000 89.930 92.0 0
50 30.300 31.000 50.580 52.000 89.960 92.0 0
30.300 31.000 50.610 52.000 89.990 92.0 0
0 30.330 31.000 50.600 52.000 90.160 92.0 0
Average 30.260 31.000 50.524 51.500 89.921 91.400 0.000
Standard 0.05 °C 0.47 °C 0.44 °C 0.00
Deviation

TYPE B UNCERTAINTY EVALUATION

Ub1: Source: Uncertainty of Reference Instruments Dig Hygrometer


(Taken from the calibration certificate)

b1 = ˚C
num Diviser-k: 2
Assuming Normal Distribution,Ub1= b1/k=
Degree of freedom V1 = infinite b1

Ub2: Source: Uncertainty due to Uniformity of Chamber (Source)


(Taken from Manufacturer's Specification of Calibrator)

b2 = num ˚C Diviser-k: 1.7320508


Assuming Rectangular Distribution,Ub2= b2/k=
Degree of freedom V1 = infinite

Ub3: Source: Uncertainty due to Stability of Chamber (Source)


(Taken from Manufacturer's Specification of Calibrator)

b3 = num #REF! Diviser-k: 1.7320508


Assuming Rectangular Distribution,Ub3= b3/k=
Degree of freedom V1 = infinite

Ub4: Source: Uncertainty due to Reolution of DUC)


Since the limits given are half the count, and assuming Rectangular Distribution

Resolution Multiplier b4 Diviser Ub5


o
C Sqrt(3)
0.01 0.5 0.005 1.732051 0.0028868

Uncertainty,Ub5 = b4 / √3

Combined Uncertainty Range


Ua
𝑼𝒄=√(∑129_(𝒊 Ub1
=𝟏)^𝑵▒𝑼_𝒊^𝟐
) Ub2
Ub3
Ub4
Uc = √ [(Ua)2+(U1)2+(U2)2+…+(Un)2] Uc
Coverage
Factor
Expanded
Expanded Uncertainty at 95% Confidence level :- Uncertainity
Expanded Uncertainty U = k x Uc

Effective degree of freedom :- Range


Veff
Veff = Uc4(Y)
[Σ{(ui4(Y / Vi)}] Range
Final Result= 20
25
30
50
0
Uc=√(∑1
𝑼𝒄=√(∑129_(𝒊 With Coverage Factor K=1.96 at 95 %Confidence Level
29_(i=1)^
=𝟏)^𝑵▒𝑼_𝒊^𝟐
N▒U_i^2
) )
˚C

0 to 95 %RH

°C 0 °C
DUC Standard DUC Standard Type A Uncertainty
Reading Readings Reading Deviation Uncertainty of Standard
Error Instrument Uniformity
(±)

°C °C °C °C °C °C °C °C
0 0 0
0.740 0.05 0.0146758 0.2 0.5
0 0 0
0 0 0
0.976 0.47 0.1477701 0.2 0.5
0 0 0
0 0 0
2.000 0.44 0.1406844 0.2 0.5
0 0 0
0 0 0
0.000 0.00 0.0000000 0.2 0.5
0 0 0
0 0 0
0.000 0.00 0.0000000 0.2 0.5
0 0 0
0.000 0.000 0.000
°C 0.00 °C

Hygrometer

Range
20 25 30 50 0 °C
0.2 0.2 0.2 0.2 0.2 °C
0.10000 0.10000 0.10000 0.10000 0.10000 °C

Range
20 25 30 50 0 °C
0.5 0.5 0.5 0.5 0.5 °C
0.28868 0.28868 0.28868 0.28868 0.28868 °C

Range
20 25 30 50 0 °C
0.5 0.5 0.5 0.5 0.5 °C
0.288675 0.288675 0.288675 0.288675 0.288675 °C

ar Distribution

Range
20 25 30 50 0 °C
0.1 0.1 0.1 0.1 0.1 °C
0.02887 0.02887 0.02887 0.02887 0.02887 °C

20 25 30 50 0 °C
0.014676 0.147770 0.140684 0.000000 0.000000 °C
0.100000 0.100000 0.100000 0.100000 0.100000 °C
0.288675 0.288675 0.288675 0.288675 0.288675 °C
0.288675 0.288675 0.288675 0.288675 0.288675 °C
0.028868 0.028868 0.028868 0.028868 0.028868 °C
0.421563 0.446471 0.444176 0.421307 0.421307 °C

1.96 1.96 1.96 1.96 1.96

0.8262635 0.8750824 0.8705845 0.8257627 0.82576268


°C

20 25 30 50 0
6.13E+06 7.50E+02 8.94E+02 #DIV/0! #DIV/0!

31.000 ± 0.83 °C
51.500 ± 0.88 °C
91.400 ± 0.87 °C
0.000 ± 0.83 °C
0.000 ± 0.83 °C
nfidence Level
Stability Resolution

°C °C
0.5 0.1

0.5 0.1

0.5 0.1

0.5 0.1

0.5 0.1
Uncertainity Calculation Sheet
1. Laboratory : CCTPL

3. Standard / Equipment Used for Calibration

Details of Standard Used Standard1 Source


Instruments Name : Dig. Thermo Hygrometer with Sensor Temp. Humidity Calibrator
Manufacturer/Model : Rotronic/ HP-32 Scientific Enirco
[Link]./[Link]: CCTPL/TH/R-1913 CCTPL/TH/R-1915
Range : 10 to 50 °C & 10 to 95%RH (-5 to 95)˚C & 10 to 95 %RH
Certificate No. FL/CAL/TH/22082019-C002
CalibrationDate&validity 28.08.2019 & 28.08.2020
Traceability: Farelabs Gurugram

[Link] Condition During Calibration


Temperature 25±2 °C Humidity 55±10 % RH

5. Evaluation
10 %RH 50 %RH 70 %RH 90
Standard Standard DUC Standard DUC Standard DUC Standard
Calibraiton Readings Reading Readings Reading Readings Reading Readings
Point

%RH %RH %RH %RH %RH %RH %RH %RH


10.060 10.000 50.220 50.000 70.420 70.000 90.620
10 10.080 10.000 50.250 50.000 70.440 70.000 90.640
10.090 10.000 50.260 50.000 70.450 70.000 90.650
50 10.110 10.000 50.260 50.000 70.460 70.000 90.650
10.120 10.000 50.280 50.000 70.470 70.000 90.660
70 10.140 10.000 50.290 50.000 70.480 70.000 90.680
10.150 10.000 50.290 50.000 70.480 70.000 90.690
90 10.160 10.000 50.310 50.000 70.490 70.000 90.710
10.160 10.000 50.320 50.000 70.510 70.000 90.720
95 10.180 10.000 50.330 50.000 70.520 70.000 90.730
Average 10.125 10.000 50.281 50.000 70.472 70.000 90.675
Standard 0.04 %RH 0.03 %RH 0.03 %RH 0.04
Deviation

TYPE B UNCERTAINTY EVALUATION

Ub1: Source: Uncertainty of Reference Instruments Dig. Hygromter


(Taken from the calibration certificate)

b1 = num 0 Diviser-k: 2
Assuming Normal Distribution,Ub1= b1/k=
Degree of freedom V1 = infinite

Ub2: Source: Uncertainty due to Uniformity of Calibrator (Source)


(Taken from Manufacturer's Specification of Bath)
b2 = num 0 Diviser-k: 1.7320508
Assuming Rectangular Distribution,Ub2= b2/k=
Degree of freedom V1 = infinite

Ub3: Source: Uncertainty due to Stability of CAlibrator (Source)


(Taken from Manufacturer's Specification of Bath)

b3 = num 0 Diviser-k: 1.7320508


Assuming Rectangular Distribution,Ub3= b3/k=
Degree of freedom V1 = infinite

Ub4: Source: Uncertainty due to Reolution of DUC)


Since the limits given are half the count, and assuming Rectangular Distribution

Resolution Multiplier b4 Diviser Ub5


o
C Sqrt(3)
0.01 0.5 0.005 1.732051 0.0028868

Uncertainty,Ub5 = b4 / √3

Combined Uncertainty Range


Ua
𝑼𝒄=√(∑129_(𝒊 Ub1
=𝟏)^𝑵▒𝑼_𝒊^𝟐
) Ub2
Ub3
Ub4
Uc = √ [(Ua)2+(U1)2+(U2)2+…+(Un)2] Uc
Coverage
Factor
Expanded
Expanded Uncertainty at 95% Confidence level :- Uncertainity
Expanded Uncertainty U = k x Uc

Effective degree of freedom :- Range


Veff
Veff = Uc4(Y)
[Σ{(ui4(Y / Vi)}] Range
Final Result= 10
50
70
90
95
Uc=√(∑1
𝑼𝒄=√(∑129_(𝒊 With Coverage Factor K=2 at 95 %Confidence Level
29_(i=1)
=𝟏)^𝑵▒𝑼_𝒊^𝟐
^N▒U_i^
)
2)
˚C

0 to 95 %RH

%RH 95 %RH
DUC Standard DUC Standard Type A Uncertainty
Reading Readings Reading Deviation Uncertainty of Standard
Error Instrument UniformityStability
(±)

%RH %RH %RH %RH %RH %RH %RH %RH %RH


90.000 95.840 95.000
-0.125 0.04 0.0124944 0.6 0 0
90.000 95.850 95.000
90.000 95.860 95.000
-0.281 0.03 0.0107961 0.6 0 0
90.000 95.870 95.000
90.000 95.880 95.000
2.000 0.03 0.0097525 0.6 0 0
90.000 95.890 95.000
90.000 95.910 95.000
-0.675 0.04 0.0116667 0.6 0 0
90.000 95.920 95.000
90.000 95.930 95.000
-0.890 0.04 0.0115470 0.6 0 0
90.000 95.950 95.000
90.000 95.890 95.000
%RH 0.04 %RH

Hygromter

Range
10 50 70 90 95 %RH
0.6 0.6 0.6 0.6 0.6 %RH
0.30000 0.30000 0.30000 0.30000 0.30000 %RH
Range
10 50 70 90 95 %RH
0 0 0 0 0 %RH
0.00000 0.00000 0.00000 0.00000 0.00000 %RH

Range
10 50 70 90 95 %RH
0 0 0 0 0 %RH
0.000000 0.000000 0.000000 0.000000 0.000000 %RH

ar Distribution

Range
10 50 70 90 95 %RH
0.1 0.1 0.1 0.1 0.1 %RH
0.02887 0.02887 0.02887 0.02887 0.02887 %RH

10 50 70 90 95
0.012494 0.010796 0.009752 0.011667 0.011547 %RH
0.300000 0.300000 0.300000 0.300000 0.300000 %RH
0.000000 0.000000 0.000000 0.000000 0.000000 %RH
0.000000 0.000000 0.000000 0.000000 0.000000 %RH
0.028868 0.028868 0.028868 0.028868 0.028868 %RH
0.301645 0.301579 0.301543 0.301611 0.301607 %RH

1.96 1.96 1.96 1.96 1.96

0.5912234 0.5910948 0.5910251 0.5911584 0.59114934


%RH

10 50 70 90 95
3.06E+06 5.48E+06 8.23E+06 4.02E+06 4.19E+06

10.000 ± 0.59 %RH


50.000 ± 0.59 %RH
70.000 ± 0.59 %RH
90.000 ± 0.59 %RH
95.000 ± 0.59 %RH
Resolution

%RH
0.1

0.1

0.1

0.1

0.1