Liquiline CM442/CM444/CM448
Multiparameter transmitter with a maximum of eight measuring chan-
nels based on digital Memosens technology
TI00444C/07/EN/15.12
71189225
Liquiline CM442/CM444/CM448
Table of contents
2 Endress+Hauser
Liquiline CM442/CM444/CM448
Modular design The modular transmitter design means it can be easily adapted to suit your needs:
• Retrofit extension modules for new or extended range of functions, e.g. current outputs, relays and digital
communication
• Upgrade to maximum of eight-channel measurement
• Optional: M12 sensor connector for connecting any kind of Memosens sensor
a0016027 a0012440
a0012437 a0012438
Endress+Hauser 3
Liquiline CM442/CM444/CM448
Measuring system The following overview shows examples of the design and layout of a measuring system.
Other sensors and assemblies can be ordered for conditions specific to your application
(--> www.endress.com/products).
Conductivity
Inductive conductivity measurement
in wastewater treatment
• Immersion assembly
Dipfit CLA111
• Sensor Indumax CLS50D
with fixed cable
Conductive conductivity measure-
ment in power plant cooling water
• Immersion assembly
Dipfit CLA111
• Sensor Condumax CLS15D
a0012412
Oxygen
Oxygen in aeration basins
• Immersion assembly Dipfit
5 4 3 CYA112
• Holder CYH112
a0012411
a0015030 a0015077
If mounting outdoors, always use the weather protection cover (see "Accessories") to protect the transmitter
against weather conditions.
4 Endress+Hauser
Liquiline CM442/CM444/CM448
a0015819-de
Slot and port assignment for hardware and how this information appears on the display
Endress+Hauser 5
Liquiline CM442/CM444/CM448
You must be aware of the type of modules and the number of them supplied with the device you have ordered
to determine the delivery status of your Liquiline.
• Basic module
– Is always present and there is only ever one.
• Fieldbus module
– Optional and there can only be one.
• Input modules
– Must be clearly assigned to the number of optional inputs ordered.
– Examples:
2 current inputs = module 2AI
4 Memosens inputs = basic module with 2 inputs + module 2DS with 2 further inputs
• Current outputs and relays
– Various module combinations can exist.
The following table will help you find out which modules you get depending on the type and number of
outputs.
Relay
Current out- 0 2 4 6 8
puts
2 - 1 x 2R 1 x 4R 1 x 4R + 1 x 2R 2 x 4R
► Sum up the number of modules and sort them according to the specified sequence. This will give you the
slot assignment for your device.
a0015979
Creating a terminal diagram using the example of the NO contact (terminal 41) of a relay
6 Endress+Hauser
Liquiline CM442/CM444/CM448
a0015791
Endress+Hauser 7
Liquiline CM442/CM444/CM448
a0015827
8 Endress+Hauser
Liquiline CM442/CM444/CM448
a0015828
Endress+Hauser 9
Liquiline CM442/CM444/CM448
a0015032
10 Endress+Hauser
Liquiline CM442/CM444/CM448
a0016009
SD card
The exchangeable storage medium enables:
• Quick and easy software updates and upgrades
• Data storage of internal device memory (e.g. logs)
• Transfer of complete configurations to a device with an identical setup (backup function)
• Transfer of configurations without the TAG and bus address to devices with an identical setup (copy func-
tion)
Endress+Hauser offers industry-approved SD cards as accessories. These memory cards provide maximum data
security and integrity.
Other SD cards can also be used. However, Endress+Hauser does not accept any responsibility for the data
security of such cards.
Endress+Hauser 11
Liquiline CM442/CM444/CM448
Self-monitoring functions
Current inputs and outputs
Current inputs are deactivated in the event of overcurrent and reactivated once the overcurrent stops. The cur-
rent outputs are read to check if the value corresponds to the setpoint.
Board monitoring
Board voltages and board temperature are monitored.
Input
Measured variables --> Documentation of the connected sensor
Signal characteristic
Linear
Internal resistance
Non-linear
Test voltage
500 V
12 Endress+Hauser
Liquiline CM442/CM444/CM448
Cable length
Max. 100 m (330 ft)
Output
Output signal Depending on version:
• 1 to 8 x 0/4 to 20 mA, active, potentially isolated from one another and from the sensor circuits
• Thereof one current output with optional HART communication (exclusively via current output 1:1)
HART
PROFIBUS DP/RS485
Data transmission rate 9.6 kBd, 19.2 kBd, 45.45kBd, 93.75 kBd, 187.5 kBd, 500 kBd, 1.5
MBd, 6 MBd, 12 MBd
Connectors Spring-cage terminal (max. 1.5 mm) bridged internally in the con-
nector (T function), M12 optional
Modbus RS485
Connectors Spring-cage terminal (max. 1.5 mm) bridged internally in the con-
nector (T function), M12 optional
Endress+Hauser 13
Liquiline CM442/CM444/CM448
Linearization/transmission Linear
behavior
Test voltage
500 V
Cross-section
Max. 2.5 mm2 (14 AWG)
Relay outputs
Electrical specification Relay types
• 1 single-pin changeover contact (alarm relay)
• 2 or 4 single-pin changeover contacts (optional with extension modules)
0.5 A 450,000
0.5 A 650,000
0.5 A 350,000
14 Endress+Hauser
Liquiline CM442/CM444/CM448
Extension modules
0.5 A 450,000
2A 120,000
0.5 A 650,000
2A 170,000
0.5 A 350,000
2A 150,000
Protocol-specific data
HART Manufacturer ID 11h
Device variables 16 configured by the user and 16 pre-defined, dynamic variables PV,
SV, TV, QV
Supported features PDM DD, AMS DD, DTM, FieldXpert DD
Endress+Hauser 15
Liquiline CM442/CM444/CM448
Output data 16 measured values (value, unit, status), 8 digital values (value, sta-
tus)
Input data 4 setpoints (value, unit, status), 4 digital values (value, status), diag-
nostic information
TCP connections 3
Protocol RTU
Output data 16 measured values (value, unit, status), 8 digital values (value, sta-
tus)
Input data 4 setpoints (value, unit, status), 6 digital values (value, status), diag-
nostic information
Web server The web server enables full access to the device configuration, measured values, diagnostic messages, logbooks
and service data via standard WiFi/WLAN/LAN/GSM or 3G router with a user-defined IP address.
TCP port 80
16 Endress+Hauser
Liquiline CM442/CM444/CM448
Power supply
Supply voltage CM442
Depending on version:
100 to 230 V AC ± 15 %, 50/60 Hz
24 V AC/DC +20/-15 %, 50/60 Hz
CM444 and CM448
100 to 230 V AC ± 15 %, 50/60 Hz
NOTICE
The device does not have a mains switch.
► The customer must provide a protected circuit breaker in the vicinity of the device.
► This must be a switch or a power-circuit breaker and must be labeled as the circuit breaker for the device.
► At the supply point, the power supply for the 24 V versions must be isolated from dangerous live cables by
double or reinforced insulation.
Fuse CM442
5x20 mm, 250 V, 4.0 A, slow-blow (T4.0A)
CM444 and CM448
Fuse not exchangeable
Cable entry Identification of the cable entry on housing base Suitable gland
A, D, F, G M20x1.5 mm/NPT1/2"/G1/2
E Socket RJ45
) M12x1.5 mm
a0018025
Endress+Hauser 17
Liquiline CM442/CM444/CM448
a0012405 a0015871
Basic module BASE-H or -L (CM442) Basic module BASE-E (CM444 and CM448)
1 Power supply for digital fixed cable sensors with 1 Power supply for digital fixed cable sensors with
Memosens protocol Memosens protocol
2 SD card slot 2 SD card slot
3 Slot for display cable1) 3 Slot for display cable1)
4 Service interface 4 Service interface
5 Connections for 2 Memosens sensors 5 Connections for 2 Memosens sensors
6 Current outputs 6 Current outputs
7 Power connection 7 LEDs
8 Fuse 8 Socket for internal supply cable
9 Alarm relay connection 9 Alarm relay connection
18 Endress+Hauser
Liquiline CM442/CM444/CM448
a0015825
a0015872 a0015873
Connecting power supply with BASE-E Overall wiring diagram for BASE-E and extension power
A Internal supply cable supply unit
B Extension power supply unit
Endress+Hauser 19
Liquiline CM442/CM444/CM448
Connecting optional modules With extension modules you can purchase additional functions for your device.
NOTICE
Unacceptable hardware combinations (due to conflicts in power supply)
Incorrect measurements or total failure of the measuring point as a result of heat build-up or overloading
► If you are planning to extend your controller, make sure the resulting hardware combination is permitted
(Configurator on www.products.endress.com/cm442 or .../cm444 or .../CM448).
► Please note that if you are extending CM442 to CM444 or CM448, you must additionally fit an extension
power supply unit and an extension backplane. You must then also use basic module BASE-E.
► Please contact your Endress+Hauser sales center should you have any questions.
Module name
AOR 2R 4R 2DS
20 Endress+Hauser
Liquiline CM442/CM444/CM448
Module name
a0012454
Endress+Hauser 21
Liquiline CM442/CM444/CM448
Connection methods
1. Sensor cable connected directly to the terminal connector of the sensor module 2DS or of the basic
module L, H or E
2. Optional: plug connection of the sensor cable connected to the M12 sensor socket on the underside of
the device. With this type of connection, the device is already wired at the factory.
a0012459 a0012460
Sensors without additional supply voltage Sensors with additional supply voltage
22 Endress+Hauser
Liquiline CM442/CM444/CM448
a0016197
Endress+Hauser 23
Liquiline CM442/CM444/CM448
Performance characteristics
Response time Current outputs
t90 = max. 500 ms for an increase from 0 to 20 mA
Current inputs
t90 = max. 330 ms for an increase from 0 to 20 mA
Installation
Mounting plate
a0012426
Mounting plate
24 Endress+Hauser
Liquiline CM442/CM444/CM448
a0012428
Installation instructions You require the post mounting kit (optional) to mount the unit on a pipe, post or railing (square or circular,
span range 20 to 61 mm (0.79 to 2.40")).
Post mounting
a0012665
Endress+Hauser 25
Liquiline CM442/CM444/CM448
Rail mounting
a0012668
Wall mounting
Mount the controller in such a way that the wall support surface is at least the size of the rear housing panel.
a0012684 a0012685
1) The size of the drill holes depends on the wall plugs used. The wall plugs and screws must be provided by
the customer.
26 Endress+Hauser
Liquiline CM442/CM444/CM448
Environment
Ambient temperature range CM442
–20 to 60 ˚C (0 to 140 ˚F)
CM444
–20 to 55 ˚C (0 to 130 ˚F)
CM448
• –20 to 55 ˚C (0 to 130 ˚F) in general, with the exception of the packages listed under the 2nd bullet point
• –20 to 50 ˚C (0 to 120 ˚F) for the following packages:
– CM448-***6AA******+...
– CM448-***8A4******+...
– CM448-***8A5******+...
– CM448-**28A3******+...
– CM448-**38A3******+...
– CM448-**48A3******+...
– CM448-**58A3******+...
– CM448-**68A3******+...
Degree of protection IP 66/67, leak-tightness and corrosion resistance in accordance with NEMA TYPE 4X
Wall mounting
Frequency range 10 to 150 Hz (sinusoidal)
Amplitude 10 to 12.9 Hz: 0.75 mm
12.9 to 150 Hz: 0.5 g 1)
Test duration 10 frequency cycles/ spatial axis, in 3 spatial axes (1 oct./min)
Endress+Hauser 27
Liquiline CM442/CM444/CM448
Mechanical construction
Dimensions
a0012396
Cable mounting rail PBT GF30 FR, stainless steel 1.4301 (AISI304)
28 Endress+Hauser
Liquiline CM442/CM444/CM448
Operability
Operating concept The simple and structured operating concept sets new standards:
• Intuitive operation with the navigator and soft keys
• Fast configuration of application-specific measurement options
• Easy configuration
and diagnosis thanks to plain-text display
• All languages that can be ordered are available in every device
a0012445 a0012410
a0011764
Overview of operation
1 Display (with red display background in alarm condition)
2 Navigator (jog/shuttle function)
3 Soft keys (function depends on menu)
Endress+Hauser 29
Liquiline CM442/CM444/CM448
Remote operation Via HART (e.g. using HART modem and FieldCare)
a0015608
Via PROFIBUS DP
a0015874
PROFIBUS DP
T Terminating resistor
30 Endress+Hauser
Liquiline CM442/CM444/CM448
a0015875
Modbus RS485
T Terminating resistor
a0015876
Language groups The language selected in the product structure is the operating language preset at the factory.
All other languages can be selected using the menu.
• English (US)
• German
• French
• Spanish
• Italian
• Dutch
• Portuguese
• Polish
• Russian
• Turkish
• Chinese (Simplified, PR China)
• Japanese
• Czech
Endress+Hauser 31
Liquiline CM442/CM444/CM448
Ordering information
Product structure You can create a valid and complete order code using the Endress + Hauser Configurator tool on the Internet.
Enter the following addresses in your browser to access the relevant product page:
www.products.endress.com/cm442
www.products.endress.com/cm444
www.products.endress.com/cm448
1. The following options are available on the right-hand side of the page:
cCSAus The product complies with "CLASS 2252 05 - Process Control Equipment" and "CLASS 2252 85 - Process
Control Equipment - Certified to US Standards" requirements.
32 Endress+Hauser
Liquiline CM442/CM444/CM448
Accessories
The most important accessories that could be delivered at the time this document went to print are listed
below.
For information on accessories that are not listed here, please contact your local service or sales
representation.
Weather protection cover CYY101 weather protection cover for field devices, absolutely essential if operating the unit outdoors
• Material: stainless steel 1.4031 (AISI 304)
• Order No. CYY101-A
Pfaudler electrodes
Ceramax CPS341D
• pH electrode with pH-sensitive enamel
• Meets highest demands of measuring accuracy, pressure, temperature, sterility and durability
• Order as per product structure (--> online Configurator, www.products.endress.com/cps341d)
• Technical Information TI468C/07/EN
Endress+Hauser 33
Liquiline CM442/CM444/CM448
ORP sensors
Orbisint CPS12D
• ORP sensor with Memosens technology
• Dirt-repellent PTFE diaphragm;
• Order as per product structure (--> online Configurator, www.products.endress.com/cps12d)
• Technical Information TI367C/07/EN
Ceraliquid CPS42D
• ORP sensor with Memosens technology
• Ceramic diaphragm and KCl liquid electrolyte
• Order as per product structure (--> online Configurator, www.products.endress.com/cps42d)
• Technical Information TI373C/07/EN
Ceragel CPS72D
• ORP sensor with Memosens technology
• Double-chamber reference system and integrated bridge electrolyte;
• Order as per product structure (--> online Configurator, www.products.endress.com/cps72d)
• Technical Information TI374C/07/EN
Orbipac CPF82D
• Compact ORP sensor for installation or immersion operation in process water and wastewater
• Order as per product structure (--> online Configurator, www.products.endress.com/cpf82d)
• Technical Information TI191C/07/EN
Orbipore CPS92D
• ORP sensor with Memosens technology
• Open aperture diaphragm for media with a high dirt load
• Order as per product structure (--> online Configurator, www.products.endress.com/cps92d)
• Technical Information TI435C/07/EN
pH ISFET sensors
Tophit CPS471D
• Sterilizable and autoclavable ISFET sensor with Memosens technology
• For the food and pharmaceutical industries, process engineering, water treatment and biotechnology
• Order as per product structure (--> online Configurator, www.products.endress.com/cps471d)
• Technical Information TI283C/07/EN
Tophit CPS441D
• Sterilizable ISFET sensor with Memosens technology
• For media with
low conductivity, with liquid KCl electrolyte
• Order as per product structure (--> online Configurator, www.products.endress.com/cps441d)
• Technical Information TI352C/07/EN
Tophit CPS491D
• ISFET sensor with Memosens technology
• Open aperture diaphragm for media with high dirt load
• Order as per product structure (--> online Configurator, www.products.endress.com/cps491d)
• Technical Information TI377C/07/EN
34 Endress+Hauser
Liquiline CM442/CM444/CM448
Condumax CLS16D
• Hygienic conductivity sensor for measurement in pure and ultrapure water and in Ex applications
• With EHEDG and 3A certificates
• Order as per product structure (--> online Configurator, www.products.endress.com/cls16d)
• Technical Information TI227C/07/EN
Condumax CLS21D
• Two-electrode sensor in fixed cable and plug-in head version
• Order as per product structure (--> online Configurator, www.products.endress.com/cls21d)
• Technical Information TI085C/07/EN
Oxygen sensors
Oxymax COS51D
• Amperometric sensor for dissolved oxygen, with Memosens technology
• Order as per product structure (--> online Configurator, www.products.endress.com/cos51d)
• Technical Information TI413C/07/EN
Oxymax COS61D
• Optical oxygen sensor for drinking water and industrial water measurement
• Measuring principle: quenching
• Memosens protocol
• Material: stainless steel 1.4571 (AISI 316Ti)
• Order as per product structure (--> online Configurator, www.products.endress.com/cos61d)
• Technical Information TI387C/07/EN
Oxymax COS22D
• Sterilizable sensor for dissolved oxygen
• Order as per product structure (--> online Configurator, www.products.endress.com/cos22d)
• Technical Information TI446C/07/EN
Chlorine sensors
CCS142D
• Membrane-covered amperometric sensor for free chlorine
• Memosens technology
• Measuring range 0.01 to 20 mg/l
• Order as per product structure (--> online Configurator, www.products.endress.com/ccs142d)
• Technical Information TI419C/07/EN
Turbidity sensors
Turbimax CUS51D
• For nephelometric measurements of turbidity and solids in wastewater
• 4-beam scattered light method
• With Memosens protocol
• Order as per product structure (--> online Configurator, www.products.endress.com/cus51d)
• Technical Information TI461C/07/EN
Endress+Hauser 35
Liquiline CM442/CM444/CM448
Interface measurement
Turbimax CUS71D
• Immersion sensor for interface measurement
• Ultrasonic interface sensor
• Order as per product structure (--> online Configurator, www.products.endress.com/cus71d)
• Technical Information TI490C/07/EN
36 Endress+Hauser
Liquiline CM442/CM444/CM448
Cable glands
Kit CM44x: gland M
• Set, 6 pieces
• Order no. 71101768
Kit CM44x: gland NPT
• Set, 6 pieces
• Order no. 71101770
Kit CM44x: gland G
• Set, 6 pieces
• Order no. 71101771
Kit CM44x: dummy plug for cable gland
• Set, 6 pieces
• Order no. 71104942
Endress+Hauser 37
TI00444C/07/EN/15.12