Ti00444cen 1512

The Liquiline CM442/CM444/CM448 is a multiparameter transmitter that can connect up to 8 digital sensors using Memosens technology. It has various benefits including maximum process safety, fast commissioning, and minimum inventory requirements. It is suited for non-hazardous area applications in water and wastewater, power stations, chemical industry, and other industrial uses.

Uploaded by

Tung
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
309 views38 pages

Ti00444cen 1512

The Liquiline CM442/CM444/CM448 is a multiparameter transmitter that can connect up to 8 digital sensors using Memosens technology. It has various benefits including maximum process safety, fast commissioning, and minimum inventory requirements. It is suited for non-hazardous area applications in water and wastewater, power stations, chemical industry, and other industrial uses.

Uploaded by

Tung
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 38

Technical Information

Liquiline CM442/CM444/CM448
Multiparameter transmitter with a maximum of eight measuring chan-
nels based on digital Memosens technology

Application Your benefits


Liquiline CM442/CM444/CM448 is an extensible • Maximum process safety thanks to:
multiparameter controller for monitoring and controlling pro- – Simple and transparent menu guidance via a graphic dis-
cesses in industry and the environmental sector. play
Depending on the version ordered, up to 8 digital sensors with – Standardized intuitive operating concept for all the
Memosens technology can be connected. devices of the new Liquiline, sampler and analyzer plat-
The measuring signals can be linked by means of the mathema- form
tical function and new measured values calculated (e.g. pH cal- • Fast commissioning thanks to:
culation based on differential conductivity measurement). – Memosens: use of lab-calibrated sensors thanks to plug-
Furthermore, up to eight 0/4 to 20 mA analog outputs are and-play capabilities
available. Digital Fieldbuses (HART, PROFIBUS, Modbus) and – Preconfigured Liquiline transmitter
the integrated web server offer additional options for connec- – Easy to expand and adapt system to meet new require-
ting to process control systems. A cleaning function, controller ments
and alarm relay can be selected. • Minimum inventory:
Optional current inputs are available for processing measured – Cross-platform, modular concept (e.g. identical modules
values from other devices, e.g. flow or level. irrespective of parameters)
– Integration into Fieldcare and W@M facilitates effective
The rugged plastic version is tailored to the following non-
asset management
hazardous area applications:
• Water and wastewater
• Power stations
• Chemical industry
• Other industrial applications

TI00444C/07/EN/15.12
71189225
Liquiline CM442/CM444/CM448

Table of contents

Function and system design. . . . . . . . . . . . . . . . . . . . . 3 Performance characteristics. . . . . . . . . . . . . . . . . . . . 24


Memosens technology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Response time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Modular design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Reference temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Maximum measured error of sensor inputs . . . . . . . . . . . . . . . . . 24
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Maximum measured error of current inputs and outputs . . . . . . 24
Device architecture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Resolution of current inputs and outputs . . . . . . . . . . . . . . . . . . 24
Communication and data processing . . . . . . . . . . . . . . . . . . . . . . 10 Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Serviceability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Reliability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Mounting plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Weather protection cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Measured variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Measuring ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Input types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Input signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Current input, passive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Relative humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Pollution degree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Pressure compensation to environment . . . . . . . . . . . . . . . . . . . 27
Linearization/transmission behavior . . . . . . . . . . . . . . . . . . . . . . 14
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 28
Current outputs, active . . . . . . . . . . . . . . . . . . . . . . . 14 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Signal characteristic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Electrical specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Operability. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Relay outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Electrical specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Language groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Protocol-specific data. . . . . . . . . . . . . . . . . . . . . . . . . 15 Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . 32


HART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Product structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
PROFIBUS DP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Modbus RS485 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Modbus TCP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Certificates and approvals . . . . . . . . . . . . . . . . . . . . . 32
Web server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4 mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
cCSAus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Energy supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 FM/CSA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 MCERTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Fieldbus connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Weather protection cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Cable entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Post mounting kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Measuring cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Connecting optional modules . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Additional functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Protective ground connection . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Other accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Sensor connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

2 Endress+Hauser
Liquiline CM442/CM444/CM448

Function and system design


Memosens technology Memosens makes your measuring point safer and more reliable:
• Non-contact, digital signal transmission enables optimum galvanic isolation
MEMO SENS
• No contact corrosion
• Completely watertight
• Laboratory sensor calibration possible, thus increasing measured value availability
• Predictive maintenance thanks to recording of sensor data, e.g.:
– Total hours of operation
– Hours of operation
with very high or very low measured values
– Hours of operation with high temperatures
– Number of steam sterilizations
– Sensor condition

Modular design The modular transmitter design means it can be easily adapted to suit your needs:
• Retrofit extension modules for new or extended range of functions, e.g. current outputs, relays and digital
communication
• Upgrade to maximum of eight-channel measurement
• Optional: M12 sensor connector for connecting any kind of Memosens sensor

a0016027 a0012440

CM448: Sample Fitting the extension module

Display Graphic display:


• Back light with switch-off function
• Red display background for alarms alerts users to errors
• Transflective display technology for maximum contrast even in bright environments
• User-definable measuring menus mean you can always keep track of the values that are important for your
application.
• Load curve display

a0012437 a0012438

Backlit display Red background indicates an error

Endress+Hauser 3
Liquiline CM442/CM444/CM448

Measuring system The following overview shows examples of the design and layout of a measuring system.
Other sensors and assemblies can be ordered for conditions specific to your application
(--> www.endress.com/products).

Measuring point pH value or ORP


A complete measuring system consists of: pH measurement in drinking water
• Liquiline transmitter • Retractable assembly Cleanfit
• Sensors with Memosens technology CPA471
• Assemblies to suit the sensors used • Sensor Orbisint CPS11D
• Post or rail mounting (optional) • Measuring cable CYK10
• Weather protection cover (optional) --> graphic
ORP in drinking water
• Immersion assembly Dipfit
CYA112
1 2 • Sensor Orbisint CPS12D
• Measuring cable CYK10
a0012413

Conductivity
Inductive conductivity measurement
in wastewater treatment
• Immersion assembly
Dipfit CLA111
• Sensor Indumax CLS50D
with fixed cable
Conductive conductivity measure-
ment in power plant cooling water
• Immersion assembly
Dipfit CLA111
• Sensor Condumax CLS15D
a0012412

Oxygen
Oxygen in aeration basins
• Immersion assembly Dipfit
5 4 3 CYA112
• Holder CYH112
a0012411

Measuring system (e.g. two-channel device) • Sensor


1 Liquiline M
– COS61D (optical)
2 Weather protection cover CYY101 (optional) with fixed cable,
3, 5 Sensor cable CYK10 or fixed cable (digital fixed cable sensors – COS51D (amperometric)
with Memosens protocol) cable CYK10
4 Power supply cable (to be provided by the customer, not part
of the scope of supply) Figure: CYA112 with COS61D
a0012415

Nitrate and SAC Turbidity and interface


Nitrate in wastewater Turbidity in industrial water
• Sensor CAS51D-**A2 with fixed • Sensor Turbimax CUS51D
cable with fixed cable
• Assembly CYA112 • Assembly Flowfit CUA250
• Holder CYH112 • Spray head CUR3 (optional)
SAC in the wastewater treatment out- Interface in the primary clarifier
let • Sensor Turbimax CUS71D
• Sensor CAS51D-**2C2 with fixed • Assembly CYA112
cable • Holder CYH112
• Assembly CYA112
Figure: CUA250 with CUS51D
• Holder CYH112
a0012787 a0012788

Chlorine Ion selective electrodes


Chlorine (and pH) in drinking water Ammonium and nitrate measurement
• Sensor CCS142D in the aeration basin
• Sensor CPS11D • Sensor CAS40D with fixed cable
• Measuring cable CYK10 • Holder CYH112
• Flow assembly CCA250

a0015030 a0015077

If mounting outdoors, always use the weather protection cover (see "Accessories") to protect the transmitter
against weather conditions.

4 Endress+Hauser
Liquiline CM442/CM444/CM448

Device architecture Slot and port assignment

a0015819-de

Slot and port assignment for hardware and how this information appears on the display

The electronic components are modular in design:


• There are several points available to connect electronic modules. They are known as "slots".
• The slots are numbered consecutively in the housing. Slots 0 and 1 are always reserved for the basic module.
• Each electronic module has one or more input(s) and output(s) or relays, collectively termed "ports" here.
• The ports are numbered consecutively for each electronic module and are automatically detected by the
software.
• Outputs and relays are named according to their function, e.g. "current output", and are displayed with the
slot and port numbers in ascending order
Example,  å :
– "Current output 2:1" means:
Slot 2 (e.g. module AOR) : Port 1 (current output 1 of module AOR)
• Inputs are assigned to measuring channels in ascending order "Slot:Port number"
Example:
– "CH1: 1:1 pH glass" means:
Channel 1 (CH1) is slot 1 (basic module) : Port 1 (input 1) and a pH glass sensor is connected to it
The unique terminal name can be explained as follows:
Slot no. : Port no. : Terminal

Electronic components ex works


Depending on the version ordered, the device is supplied with a number of electronic modules, which are assi-
gned in a specific sequence in ascending order to slots 0 to 7.
If you do not have a particular module, the next moves up automatically:
• The basic module (which is always present) always occupies slots 0 and 1
• Fieldbus module 485
• Memosens input module 2DS (DS = digital sensor)
• Current input module 2AI (AI = analog input)
• Current output module 4AO or 2AO (AO = analog output)
• Relay modules AOR, 4R or 2R (AOR = analog output + relay, R = relay)
Modules with 4 ports are connected before modules of the same type with 2 ports.

Endress+Hauser 5
Liquiline CM442/CM444/CM448

You must be aware of the type of modules and the number of them supplied with the device you have ordered
to determine the delivery status of your Liquiline.
• Basic module
– Is always present and there is only ever one.
• Fieldbus module
– Optional and there can only be one.
• Input modules
– Must be clearly assigned to the number of optional inputs ordered.
– Examples:
2 current inputs = module 2AI
4 Memosens inputs = basic module with 2 inputs + module 2DS with 2 further inputs
• Current outputs and relays
– Various module combinations can exist.
The following table will help you find out which modules you get depending on the type and number of
outputs.

Relay

Current out- 0 2 4 6 8
puts

2 - 1 x 2R 1 x 4R 1 x 4R + 1 x 2R 2 x 4R

4 1 x 2AO 1 x AOR 1 x 2AO + 1 x 4R 1 x AOR + 1 x 4R 1 x 2AO + 2 x 4R

6 1 x 4AO 1 x 4AO + 1 x 2R 1 x 4AO + 1 x 4R 1 x 4AO + 1 x 4R 1 x 4AO + 2 x 4R


+ 1 x 2R
8 1 x 4AO + 1 x 4AO + 1 x 4AO + 1 x 4AO + 1 x 4AO +
1 x 2AO 1 x 2AO + 1x 2AO + 1 x 4R 1 x 2AO + 1 x 2AO +
1 x 2R 1 x 4R + 1 x 2R 2 x 4R

► Sum up the number of modules and sort them according to the specified sequence. This will give you the
slot assignment for your device.

Example for creating a terminal diagram:


Device with 4 inputs for digital sensors, 4 current outputs and 4 relays
• Basic module BASE-E (contains 2 sensor inputs, 2 current outputs)
• Module 2DS (2 sensor inputs)
• Module 2AO (2 current outputs)
• Module 4R (4 relays)

a0015979

Creating a terminal diagram using the example of the NO contact (terminal 41) of a relay

6 Endress+Hauser
Liquiline CM442/CM444/CM448

Block diagram CM442

a0015791

Block diagram CM442

1 Current output 1:1, + HART (optional) 6 Power supply


2 Current outputs (2 x optional) 7 Service interface
3 2 x Memosens input (1 x optional) 8 Power supply, fixed cable sensors
4 PROFIBUS DP/Modbus/Ethernet (optional) 9 Alarm relay
5 2 x current input (optional) 10 2 or 4 x relays (optional)

Endress+Hauser 7
Liquiline CM442/CM444/CM448

Block diagram CM444

a0015827

Block diagram CM444

1 Current output 1:1 + HART (both optional) 6 Power supply


2 Max. 7 x current output (optional) 7 Service interface
3 Memosens input (2 x standard + 2 x optional) 8 Power supply, fixed cable sensors
4 PROFIBUS DP/Modbus/Ethernet (optional) 9 Alarm relay
5 2 x current input (optional) 10 2 or 4 x relays (optional)

8 Endress+Hauser
Liquiline CM442/CM444/CM448

Block diagram CM448

a0015828

Block diagram CM448

1 Current output 1:1 + HART (both optional) 7 Service interface


2 Max. 7 x current output (optional) 8 Power supply, fixed cable sensors
3 Max. 8 x Memosens input (of which 2 x optional) 9 Alarm relay
4 PROFIBUS DP/Modbus/Ethernet (optional) 10 2 or 4 x relays (optional)
6 Power supply

Endress+Hauser 9
Liquiline CM442/CM444/CM448

Communication and data pro- Types of communication:


cessing • Fieldbuses
– HART
– PROFIBUS DP (Profile 3.02)
– Modbus TCP or RS485
• Configuration via Ethernet
Only one type of Fieldbus communication can ever be active. The last activation code entered decides
which bus is used.
Extension module 485 and current outputs
For communication types PROFIBUS DP, Modbus and Ethernet:
• CM442:
Current outputs cannot be used in parallel. Any existing current outputs are deactivated with the installation
of 485.
• CM444/CM448:
Max. of 2 current outputs can be used in parallel.
Bus termination on the device
• Via slide switch at bus module 485
• Displayed via LED "T" on bus module 485

Serviceability Data logger function


• Adjustable scan time: 1 to 3600 s (6 h)
• Data logbooks:
– Max. 8 data logbooks
– 150,000 entries per logbook
– Graphic display (load curves) or numerical list
• Calibration logbook: max. 75 entries
• Hardware version logbook:
– Hardware configuration and modifications
– Max. 125 entries
• Version logbook:
– E.g. software updates
– Max. 50 entries
• Operation logbook: max. 250 entries
• Diagnostics logbook: max. 250 entries

a0015032

Data logbook: Graphic display

FieldCare and Field Data Manager


FieldCare
Configuration and asset management software based on FDT/DTM technology
• Complete device configuration when connected via FXA291 and service interface
• Access to a number of configuration parameters and identification, measuring and diagnostic data when con-
nected via HART modem
• Logbooks can be downloaded in CSV format or binary format for "Field Data Manager" software

10 Endress+Hauser
Liquiline CM442/CM444/CM448

Field Data Manager


Visualization software and database for measuring, calibration and configuration data
• SQL database which is protected against manipulation
• Functions to import, save and print out logbooks
• Load curves to display measured values

a0016009

Field Data Manager: Load curves

SD card
The exchangeable storage medium enables:
• Quick and easy software updates and upgrades
• Data storage of internal device memory (e.g. logs)
• Transfer of complete configurations to a device with an identical setup (backup function)
• Transfer of configurations without the TAG and bus address to devices with an identical setup (copy func-
tion)
Endress+Hauser offers industry-approved SD cards as accessories. These memory cards provide maximum data
security and integrity.
Other SD cards can also be used. However, Endress+Hauser does not accept any responsibility for the data
security of such cards.

Virtual process values (mathematical functions)


In addition to "real" process values, which are provided by connected physical sensors or analog inputs,
mathematical functions can be used to calculate a maximum of 6 "virtual" process values.
The "virtual" process values can be:
• Output via a current output or a Fieldbus
• Used as a regulating control variable
• Assigned as a measured variable to a limit contactor
• Used as a measured variable to trigger cleaning
• Displayed in user-defined measuring menus.
The following mathematical functions are possible:
• pH calculation based on two conductivity values acc. to VGB 405 RL, e.g. in boiler feed water
• Difference between two measured values from different sources, e.g. for membrane monitoring
• Differential conductivity , e.g. monitoring the efficiency of ion exchangers
• Degassed conductivity, e.g. for process controls in power plants
• Redundancy for monitoring two or three redundantly measuring sensors
• rH calculation based on the measured values of a pH and an ORP sensor

Endress+Hauser 11
Liquiline CM442/CM444/CM448

Reliability Data security


All settings, logbooks etc. are stored in a non-volatile memory to ensure that the data are retained even in the
event of a disruption to the power supply.

Sensor Check System (SCS)


The Sensor Check System (SCS) monitors the high impedance of the pH glass.
An alarm is issued if a minimum impedance value is undershot or a maximum impedance is exceeded.
• Glass breakage is the main reason for a drop in high impedance values.
• The reasons for increasing impedance values include:
– Dry sensor
– Worn pH glass membrane.

Process Check System (PCS)


The process check system (PCS) checks the measuring signal for stagnation. An alarm is triggered if the
measuring signal does not change over a specific period (several measured values).
The main causes of stagnating measured values are:
• Contaminated sensor, or sensor outside of medium
• Sensor failure
• Process error (e.g. through control system)

Self-monitoring functions
Current inputs and outputs
Current inputs are deactivated in the event of overcurrent and reactivated once the overcurrent stops. The cur-
rent outputs are read to check if the value corresponds to the setpoint.
Board monitoring
Board voltages and board temperature are monitored.

Input
Measured variables --> Documentation of the connected sensor

Measuring ranges --> Documentation of the connected sensor

Input types • Digital sensor inputs


• Analog current inputs (optional)

Input signal Depending on version


• Max. 8 x binary sensor signal
• 2 x 0/4 to 20 mA (optional), passive, potentially isolated from one another and from the sensor inputs

Current input, passive Span


> 0 to 20 mA

Signal characteristic
Linear

Internal resistance
Non-linear

Test voltage
500 V

12 Endress+Hauser
Liquiline CM442/CM444/CM448

Cable specification Cable type


Memosens data cable CYK10 or sensor fixed cable, each with cable end sleeves or M12 round-pin connector

Cable length
Max. 100 m (330 ft)

Output
Output signal Depending on version:
• 1 to 8 x 0/4 to 20 mA, active, potentially isolated from one another and from the sensor circuits
• Thereof one current output with optional HART communication (exclusively via current output 1:1)

HART

Signal coding FSK ± 0.5 mA above current signal

Data transmission rate 1200 baud

Galvanic isolation Yes


Load (communication resistance) 250 

PROFIBUS DP/RS485

Signal coding EIA/TIA-485, PROFIBUS DP-compliant acc. to IEC 61158

Data transmission rate 9.6 kBd, 19.2 kBd, 45.45kBd, 93.75 kBd, 187.5 kBd, 500 kBd, 1.5
MBd, 6 MBd, 12 MBd

Galvanic isolation Yes

Connectors Spring-cage terminal (max. 1.5 mm) bridged internally in the con-
nector (T function), M12 optional

Bus termination Internal slide switch with LED display

Modbus RS485

Signal coding EIA/TIA-485


Data transmission rate 2,400, 4,800, 9,600, 19,200, 38,400, 57,600 and 115,200 baud

Galvanic isolation Yes

Connectors Spring-cage terminal (max. 1.5 mm) bridged internally in the con-
nector (T function), M12 optional

Bus termination Internal slide switch with LED display

Ethernet and Modbus TCP

Signal coding IEEE 802.3 (Ethernet)


Data transmission rate 10/100 MBd

Galvanic isolation Yes

Connection RJ45, M12 optional


IP address DHCP or configuration using menu

Endress+Hauser 13
Liquiline CM442/CM444/CM448

Signal on alarm Adjustable, as per NAMUR Recommendation NE 43


• In measuring range 0 to 20 mA (HART is not available with this measuring range):
Error current from 0 to 23 mA
• In measuring range 4 to 20 mA:
Error current from 2.4 to 23 mA
• Factory setting for error current for both measuring ranges:
21.5 mA

Load Max. 500 

Linearization/transmission Linear
behavior

Current outputs, active


Span 0 to 23 mA
2.4 to 23 mA for HART communication

Signal characteristic Linear

Electrical specification Output voltage


Max. 24 V

Test voltage
500 V

Cable specification Cable type


Recommended: shielded cable

Cross-section
Max. 2.5 mm2 (14 AWG)

Relay outputs
Electrical specification Relay types
• 1 single-pin changeover contact (alarm relay)
• 2 or 4 single-pin changeover contacts (optional with extension modules)

Relay switching capacity

Basic module (alarm relay)

Switching voltage Load (max.) Switching cycles


(min.)
230 V AC, cos = 0.8 to 1 0.1 A 700,000

0.5 A 450,000

115 V AC, cos = 0.8 to 1 0.1 A 1,000,000

0.5 A 650,000

24 V DC, L/R = 0 to 1 ms 0.1 A 500,000

0.5 A 350,000

14 Endress+Hauser
Liquiline CM442/CM444/CM448

Extension modules

Switching voltage Load (max.) Switching cycles


(min.)

230 V AC, cos = 0.8 to 1 0.1 A 700,000

0.5 A 450,000
2A 120,000

115 V AC, cos = 0.8 to 1 0.1 A 1,000,000

0.5 A 650,000

2A 170,000

24 V DC, L/R = 0 to 1 ms 0.1 A 500,000

0.5 A 350,000
2A 150,000

Minimum load (typical)


• Min. 100 mA with 5 V DC
• Min. 1 mA with 24 V DC
• Min. 5 mA with 24 V AC
• Min. 1 mA with 230 V AC

Cable specification Cross-section


Max. 2.5 mm2 (14 AWG)

Protocol-specific data
HART Manufacturer ID 11h

Device type 119Ch (CM44x), 119Dh (CSFxx)


Device revision 001h

HART version 7.2

Device description files (DD/DTM) www.endress.com


Device Integration Manager (DIM)

Device variables 16 configured by the user and 16 pre-defined, dynamic variables PV,
SV, TV, QV
Supported features PDM DD, AMS DD, DTM, FieldXpert DD

Endress+Hauser 15
Liquiline CM442/CM444/CM448

PROFIBUS DP Manufacturer ID 11h

Device type 155Dh (CM44x), 155Ch (CSFxx)

Profile version 3.02

GSD files www.products.endress.com/profibus


Device Integration Manager DIM

Output variables 16 AI blocks, 8 DI blocks


Input variables 8 AO blocks, 4 DO blocks

Supported features • 1 MSCY0 connection (cyclical communication, master class 1 to


slave)
• 1 MSAC1 connection (acyclical communication, master class 1 to
slave)
• 2 MSAC2 connections (acyclical communication, master class 2
to slave)
• Device lock: The device can be locked using the hardware or soft-
ware.
• Addressing using DIL switches or software
• GSD, PDM DD, DTM

Modbus RS485 Protocol RTU/ASCII

Function codes 03, 04, 06, 08, 16, 23


Broadcast support for function codes 06, 16, 23

Output data 16 measured values (value, unit, status), 8 digital values (value, sta-
tus)
Input data 4 setpoints (value, unit, status), 4 digital values (value, status), diag-
nostic information

Supported features Address can be configured using switch or software

Modbus TCP TCP port 502

TCP connections 3

Protocol RTU

Function codes 03, 04, 06, 08, 16, 23

Broadcast support for function codes 06, 16, 23

Output data 16 measured values (value, unit, status), 8 digital values (value, sta-
tus)

Input data 4 setpoints (value, unit, status), 6 digital values (value, status), diag-
nostic information

Supported features Address can be configured using DHCP or software

Web server The web server enables full access to the device configuration, measured values, diagnostic messages, logbooks
and service data via standard WiFi/WLAN/LAN/GSM or 3G router with a user-defined IP address.

TCP port 80

Supported features • Remote-controlled device configuration


• Save/restore device configuration
• Logbook export (file formats: CSV, FDM)
• Access to web server via DTM or Internet Explorer

16 Endress+Hauser
Liquiline CM442/CM444/CM448

Power supply
Supply voltage CM442
Depending on version:
100 to 230 V AC ± 15 %, 50/60 Hz
24 V AC/DC +20/-15 %, 50/60 Hz
CM444 and CM448
100 to 230 V AC ± 15 %, 50/60 Hz

NOTICE
The device does not have a mains switch.
► The customer must provide a protected circuit breaker in the vicinity of the device.
► This must be a switch or a power-circuit breaker and must be labeled as the circuit breaker for the device.
► At the supply point, the power supply for the 24 V versions must be isolated from dangerous live cables by
double or reinforced insulation.

Fieldbus connection Supply voltage Not applicable

Power consumption CM442


Depending on supply voltage
• 100 to 230 V AC and 24 V AC:
Max. 55 VA
• 24 V DC:
Max. 22 W
CM444 and CM448
Max. 73 VA

Fuse CM442
5x20 mm, 250 V, 4.0 A, slow-blow (T4.0A)
CM444 and CM448
Fuse not exchangeable

Cable entry Identification of the cable entry on housing base Suitable gland

B, C, H, I, 1-8 M16x1.5 mm/NPT3/8"/G3/8

A, D, F, G M20x1.5 mm/NPT1/2"/G1/2

E Socket RJ45

) M12x1.5 mm

Recommended assignment of cable entries

1-8 Sensors 1-8


A Energy supply
B RS485 Out
C Ethernet 2
D,F,G Current outputs and inputs, relays
H Ethernet 1
I RS485 In
E Do not use

a0018025

Endress+Hauser 17
Liquiline CM442/CM444/CM448

Cable specification Cable gland Permitted cable diameter

M16x1.5 mm 4 to 8 mm (0.16 to 0.32")

M12x1.5 mm 2 to 5 mm (0.08 to 0.20")


M20x1.5 mm 6 to 12 mm (0.24 to 0.48")

NPT3/8" 4 to 8 mm (0.16 to 0.32")

G3/8 4 to 8 mm (0.16 to 0.32")

NPT1/2" 6 to 12 mm (0.24 to 0.48")

G1/2 7 to 12 mm (0.28 to 0.48")

Electrical connection Basic module

a0012405 a0015871

Basic module BASE-H or -L (CM442) Basic module BASE-E (CM444 and CM448)
1 Power supply for digital fixed cable sensors with 1 Power supply for digital fixed cable sensors with
Memosens protocol Memosens protocol
2 SD card slot 2 SD card slot
3 Slot for display cable1) 3 Slot for display cable1)
4 Service interface 4 Service interface
5 Connections for 2 Memosens sensors 5 Connections for 2 Memosens sensors
6 Current outputs 6 Current outputs
7 Power connection 7 LEDs
8 Fuse 8 Socket for internal supply cable
9 Alarm relay connection 9 Alarm relay connection

1) Internal device connection. Do not disconnect the plug!

18 Endress+Hauser
Liquiline CM442/CM444/CM448

Connecting supply voltage for CM442

a0015825

Connecting power supply on the BASE-H or -L


H Power unit 100 to 230 VAC
a0012404
L Power unit 24 VAC or 24 VDC
Overall wiring diagram for BASE-H or -L

Connecting supply voltage for CM444 and CM448

a0015872 a0015873

Connecting power supply with BASE-E Overall wiring diagram for BASE-E and extension power
A Internal supply cable supply unit
B Extension power supply unit

Endress+Hauser 19
Liquiline CM442/CM444/CM448

Connecting optional modules With extension modules you can purchase additional functions for your device.

NOTICE
Unacceptable hardware combinations (due to conflicts in power supply)
Incorrect measurements or total failure of the measuring point as a result of heat build-up or overloading
► If you are planning to extend your controller, make sure the resulting hardware combination is permitted
(Configurator on www.products.endress.com/cm442 or .../cm444 or .../CM448).
► Please note that if you are extending CM442 to CM444 or CM448, you must additionally fit an extension
power supply unit and an extension backplane. You must then also use basic module BASE-E.
► Please contact your Endress+Hauser sales center should you have any questions.

Module name

AOR 2R 4R 2DS

a0015747 a0015748 a0015749 a0015754

• 2 x 0/4 to 20 mA ana- • 2 relays • 4 relays • 2 digital sensor inputs


log outputs • Order no. 71125375 • Order no. 71125376 • 2 power supply systems
• 2 relays for digital sensors
• Order no. 71111053 • Order no. 71135631

a0015755 a0015758 a0015757 a0015756

20 Endress+Hauser
Liquiline CM442/CM444/CM448

Module name

2AO 4AO 2AI 485

a0015750 a0015751 a0015752 a0015753

• 2 x 0/4 to 20 mA ana- • 4 x 0/4 to 20 mA ana- • 2 x 0/4 to 20 mA ana- • Ethernet (web server or


log outputs log outputs log inputs Modbus TCP)
• Order no. 71135632 • Order no. 71135633 • Order no. 71135639 • 5V power supply for
PROFIBUS DP termina-
tion
• RS485 (PROFIBUS DP
or Modbus RS485)
• Order no. 71135634

a0015759 a0015760 a0015761 a0015762

PROFIBUS DP (module 485)


Contacts A - A’, B - B’ and C - C’ are bridged in the connector. This ensures that PROFIBUS communi-
cation is not interrupted if the connector is disconnected.

Protective ground connection

a0012454

Cable mounting rail and associated function


1 Cable mounting rail 3 Additional threaded bolts for ground connections
2 Threaded bolt (protective ground connection, 4 Cable clamps (fixing and grounding the sensor
central grounding point) cables)

Endress+Hauser 21
Liquiline CM442/CM444/CM448

Sensor connection Sensors with Memosens protocol

Sensor types Sensor cables Sensors

Digital sensors without CYK10 with plug-in • pH sensors


additional power supply connection and inductive • ORP sensors
signal transmission • Combi sensors
• Amperometric oxygen sensors
• Conductive conductivity sensors
• Chlorine sensors

Fixed cable Inductive conductivity sensors

Digital sensors with additional Fixed cable • Turbidity sensors


power supply • Sludge level sensors
• Sensors for measuring the spectral absorption
coefficient (SAC)
• Nitrate sensors
• Optical oxygen sensors
• Ion-sensitive sensors

Connection methods
1. Sensor cable connected directly to the terminal connector of the sensor module 2DS or of the basic
module L, H or E
2. Optional: plug connection of the sensor cable connected to the M12 sensor socket on the underside of
the device. With this type of connection, the device is already wired at the factory.

1. Sensor cable connected directly

a0012459 a0012460

Sensors without additional supply voltage Sensors with additional supply voltage

22 Endress+Hauser
Liquiline CM442/CM444/CM448

a0016197

Sensors with and without additional supply voltage at sensor


module 2DS

2. Connection via M12 plug-in connection


Device versions with a pre-installed M12
socket are ready-wired upon delivery.
Install an M12 socket, which is available as an
accessory, in a suitable cable gland opening in the
base of the housing, and connect the cables to the
Memosens terminals of the sensor or basic module
as per the wiring diagram ( å ) .
Connecting the sensor
► Connect the sensor cable plug (pos. 1) directly
to the M12 socket.
Please note the following for these device versions:
• The internal device wiring is always the same
regardless of what kind of sensor you connect to
the M12 socket (plug&play).
• The signal and power supply cables are assigned
a0018021
in the sensor plug-in head in such a way that the
M12 assignment PK and GY power supply cables are either used
1 PK (24 V) (e.g. optical sensors) or not (e.g. pH or ORP
2 GY (Shield 24 V) sensors).
a0018019
3 BN (3 V)
M12 plug-in connection (e.g. at 4 WH (Shield 3 V)
sensor module) 5 GN (Memosens)
1 Sensor with M12 6 YE (Memosens)
connector 7, NC Not connected

Endress+Hauser 23
Liquiline CM442/CM444/CM448

Performance characteristics
Response time Current outputs
t90 = max. 500 ms for an increase from 0 to 20 mA
Current inputs
t90 = max. 330 ms for an increase from 0 to 20 mA

Reference temperature 25 ˚C (77 ˚F)

Maximum measured error of --> Documentation of the connected sensor


sensor inputs

Maximum measured error of Typical measured errors:


current inputs and outputs < 20 μA (with current values < 4 mA)
< 50 μA (with current values 4 to 20 mA)
at 25 ˚C (77˚ F) each
Additional measured error depending on the temperature:
< 1.5 μA/K

Resolution of current inputs < 5 μA


and outputs

Repeatability --> Documentation of the connected sensor

Installation
Mounting plate

a0012426

Mounting plate

24 Endress+Hauser
Liquiline CM442/CM444/CM448

Weather protection cover

a0012428

Weather protection cover for CM44x

Installation instructions You require the post mounting kit (optional) to mount the unit on a pipe, post or railing (square or circular,
span range 20 to 61 mm (0.79 to 2.40")).

Post mounting

a0012665

Post mounting (exploded view)


1 Weather protection cover (optional) 5 Spring washers, nuts (post mounting kit)
2 Post mounting plate (post mounting kit) 6 Pipe or railing (circular/square)
3 Spring washers, nuts (post mounting kit) 7 Mounting plate
4 Pipe clamps (post mounting kit) 8 Threaded rods (post mounting kit)

Endress+Hauser 25
Liquiline CM442/CM444/CM448

Rail mounting

a0012668

Rail mounting (exploded view)


1 Weather protection cover (optional) 6 Pipe or railing (circular/square)
2 Post mounting plate (post mounting kit) 7 Mounting plate
3 Spring washers, nuts (post mounting kit) 8 Threaded rods (post mounting kit)
4 Pipe clamps (post mounting kit) 9 Screws (post mounting kit)
5 Spring washers, nuts (post mounting kit)

Wall mounting
Mount the controller in such a way that the wall support surface is at least the size of the rear housing panel.

a0012684 a0012685

Wall mounting Attach the Liquiline and click it into place


1 Wall
2 4 drill holes1)
3 Mounting plate
a0012686 4 Screws Ø 6 mm (not part of
Minimum spacing required for mounting scope of supply)

1) The size of the drill holes depends on the wall plugs used. The wall plugs and screws must be provided by
the customer.

26 Endress+Hauser
Liquiline CM442/CM444/CM448

Environment
Ambient temperature range CM442
–20 to 60 ˚C (0 to 140 ˚F)
CM444
–20 to 55 ˚C (0 to 130 ˚F)
CM448
• –20 to 55 ˚C (0 to 130 ˚F) in general, with the exception of the packages listed under the 2nd bullet point
• –20 to 50 ˚C (0 to 120 ˚F) for the following packages:
– CM448-***6AA******+...
– CM448-***8A4******+...
– CM448-***8A5******+...
– CM448-**28A3******+...
– CM448-**38A3******+...
– CM448-**48A3******+...
– CM448-**58A3******+...
– CM448-**68A3******+...

Storage temperature –40 to 80 ˚C (–40 to 175 ˚F)

Relative humidity 10 to 95%,


not condensing

Degree of protection IP 66/67, leak-tightness and corrosion resistance in accordance with NEMA TYPE 4X

Vibration resistance Environmental tests


Vibration test based on DIN EN 60068-2, October 2008
Vibration test based on DIN EN 60654-3, August 1998

Post or pipe mounting


Frequency range 10 to 500 Hz (sinusoidal)
Amplitude 10 to 57.5 Hz: 0.15 mm
57.5 to 500 Hz: 2 g 1)
Test duration 10 frequency cycles/ spatial axis, in 3 spatial axes (1 oct./min)

Wall mounting
Frequency range 10 to 150 Hz (sinusoidal)
Amplitude 10 to 12.9 Hz: 0.75 mm
12.9 to 150 Hz: 0.5 g 1)
Test duration 10 frequency cycles/ spatial axis, in 3 spatial axes (1 oct./min)

1) g ... gravitational acceleration (1 g is approximately 9.81 m/s2)

Electromagnetic compatibility Interference emission


and interference immunity as per EN 61326-1: 2006, class A for industry

Pollution degree The product is suitable for pollution degree 4.

Pressure compensation to Filter made of GORE-TEX used as pressure compensation element


environment Ensures pressure compensation to environment and guarantees IP protection.

Endress+Hauser 27
Liquiline CM442/CM444/CM448

Mechanical construction
Dimensions

a0012396

Dimensions of field housing

Weight Approx. 2.1 kg (4.63 lbs), depending on the version

Material Lower housing section PC-FR

Display cover PC-FR

Display film and soft keys PE

Housing seal EPDM

Module side panels PC-FR

Module covers PBT GF30 FR

Cable mounting rail PBT GF30 FR, stainless steel 1.4301 (AISI304)

Clamps Stainless steel 1.4301 (AISI304)

Screws Stainless steel 1.4301 (AISI304)

28 Endress+Hauser
Liquiline CM442/CM444/CM448

Operability
Operating concept The simple and structured operating concept sets new standards:
• Intuitive operation with the navigator and soft keys
• Fast configuration of application-specific measurement options
• Easy configuration
and diagnosis thanks to plain-text display
• All languages that can be ordered are available in every device

a0012445 a0012410

Easy operation Plain-text menu

Local operation Human interface

a0011764

Overview of operation
1 Display (with red display background in alarm condition)
2 Navigator (jog/shuttle function)
3 Soft keys (function depends on menu)

Endress+Hauser 29
Liquiline CM442/CM444/CM448

Remote operation Via HART (e.g. using HART modem and FieldCare)

a0015608

HART using modem


1 Device module Base-L, -H or -E: current output 1 with HART
2 HART modem to connect to PC, e.g. Commubox FXA191 (RS232) or FXA1951) (USB)
3 HART handheld terminal

Via PROFIBUS DP

a0015874

PROFIBUS DP
T Terminating resistor

1) Switch position "on" (replaces resistance)

30 Endress+Hauser
Liquiline CM442/CM444/CM448

Via Modbus RS485

a0015875

Modbus RS485
T Terminating resistor

Via Ethernet/Web server/Modbus TCP

a0015876

Modbus TCP and/or Ethernet

Language groups The language selected in the product structure is the operating language preset at the factory.
All other languages can be selected using the menu.
• English (US)
• German
• French
• Spanish
• Italian
• Dutch
• Portuguese
• Polish
• Russian
• Turkish
• Chinese (Simplified, PR China)
• Japanese
• Czech

Endress+Hauser 31
Liquiline CM442/CM444/CM448

Ordering information
Product structure You can create a valid and complete order code using the Endress + Hauser Configurator tool on the Internet.
Enter the following addresses in your browser to access the relevant product page:
www.products.endress.com/cm442
www.products.endress.com/cm444
www.products.endress.com/cm448
1. The following options are available on the right-hand side of the page:

2. Select "Configure this product".


3. A new window opens with the Configurator. Using this tool, you can configure your device and you will
receive a valid and complete order code for this.
4. Then export the order code as a PDF or as an Excel file selecting from the buttons provided at the top of
the page.

Scope of delivery • 1 controller in the version ordered


• 1 mounting plate
• 1 wiring label (attached at the factory to the inside of the display cover)
• 1 CD with Operating Instructions
• 1 printed copy of the "Commissioning" section of the Operating Instructions in the language ordered

Certificates and approvals


4 mark Declaration of Conformity
The product meets the requirements of the harmonized European standards.
As such, it complies with the legal specifications of the EC directives.
The manufacturer confirms successful testing of the product by affixing to it the 4 mark.

cCSAus The product complies with "CLASS 2252 05 - Process Control Equipment" and "CLASS 2252 85 - Process
Control Equipment - Certified to US Standards" requirements.

FM/CSA CM442 only


FM/CSA Cl. 1, Div. 2

MCERTS Application made for CM442

32 Endress+Hauser
Liquiline CM442/CM444/CM448

Accessories
The most important accessories that could be delivered at the time this document went to print are listed
below.
For information on accessories that are not listed here, please contact your local service or sales
representation.

Weather protection cover CYY101 weather protection cover for field devices, absolutely essential if operating the unit outdoors
• Material: stainless steel 1.4031 (AISI 304)
• Order No. CYY101-A

Post mounting kit Post mounting kit CM44x


• For securing the field housing to horizontal and vertical posts and pipes
• Order no. 71096920

Measuring cable Memosens data cable


CYK10
• For digital sensors with Memosens technology
pH, redox, oxygen (amperometric), chlorine, conductivity (conductive)
• Order as per product structure (--> online Configurator, www.products.endress.com/cyk10)
Memosens data cable
CYK11
• Extension cable for digital sensors with Memosens protocol
• Order as per product structure (--> online Configurator, www.products.endress.com/cyk11)

Sensors Glass electrodes


Orbisint CPS11D
• pH sensor with Memosens technology
• Dirt-repellent PTFE diaphragm
• Order as per product structure (--> online Configurator, www.products.endress.com/cps11d)
• Technical Information TI028C/07/EN
Ceraliquid CPS41D
• pH sensor with Memosens technology
• Ceramic diaphragm and KCl liquid electrolyte
• Order as per product structure (--> online Configurator, www.products.endress.com/cps41d)
• Technical Information TI079C/07/EN
Ceragel CPS71D
• pH sensor with Memosens technology
• Double-chamber reference system and integrated bridge electrolyte
• Order as per product structure (--> online Configurator, www.products.endress.com/cps71d)
• Technical Information TI245C/07/EN
Orbipore CPS91D
• pH sensor with Memosens technology
• Open aperture diaphragm for media with a high dirt load
• Order as per product structure (--> online Configurator, www.products.endress.com/cps91d)
• Technical Information TI375C/07/EN
Orbipac CPF81D
• Compact pH sensor for installation or immersion operation in process water and wastewater
• Order as per product structure (--> online Configurator, www.products.endress.com/cpf81d)
• Technical Information TI191C/07/EN

Pfaudler electrodes
Ceramax CPS341D
• pH electrode with pH-sensitive enamel
• Meets highest demands of measuring accuracy, pressure, temperature, sterility and durability
• Order as per product structure (--> online Configurator, www.products.endress.com/cps341d)
• Technical Information TI468C/07/EN

Endress+Hauser 33
Liquiline CM442/CM444/CM448

ORP sensors
Orbisint CPS12D
• ORP sensor with Memosens technology
• Dirt-repellent PTFE diaphragm;
• Order as per product structure (--> online Configurator, www.products.endress.com/cps12d)
• Technical Information TI367C/07/EN
Ceraliquid CPS42D
• ORP sensor with Memosens technology
• Ceramic diaphragm and KCl liquid electrolyte
• Order as per product structure (--> online Configurator, www.products.endress.com/cps42d)
• Technical Information TI373C/07/EN
Ceragel CPS72D
• ORP sensor with Memosens technology
• Double-chamber reference system and integrated bridge electrolyte;
• Order as per product structure (--> online Configurator, www.products.endress.com/cps72d)
• Technical Information TI374C/07/EN
Orbipac CPF82D
• Compact ORP sensor for installation or immersion operation in process water and wastewater
• Order as per product structure (--> online Configurator, www.products.endress.com/cpf82d)
• Technical Information TI191C/07/EN
Orbipore CPS92D
• ORP sensor with Memosens technology
• Open aperture diaphragm for media with a high dirt load
• Order as per product structure (--> online Configurator, www.products.endress.com/cps92d)
• Technical Information TI435C/07/EN

pH ISFET sensors
Tophit CPS471D
• Sterilizable and autoclavable ISFET sensor with Memosens technology
• For the food and pharmaceutical industries, process engineering, water treatment and biotechnology
• Order as per product structure (--> online Configurator, www.products.endress.com/cps471d)
• Technical Information TI283C/07/EN
Tophit CPS441D
• Sterilizable ISFET sensor with Memosens technology
• For media with
low conductivity, with liquid KCl electrolyte
• Order as per product structure (--> online Configurator, www.products.endress.com/cps441d)
• Technical Information TI352C/07/EN
Tophit CPS491D
• ISFET sensor with Memosens technology
• Open aperture diaphragm for media with high dirt load
• Order as per product structure (--> online Configurator, www.products.endress.com/cps491d)
• Technical Information TI377C/07/EN

Inductive conductivity sensors


Indumax CLS50D
• High-stability inductive conductivity sensor for standard, Ex and high-temperature applications
• Memosens protocol
• Order as per product structure (--> online Configurator, www.products.endress.com/cls50d)
• Technical Information TI182C/07/EN

Conductive conductivity sensors


Condumax CLS15D
• Conductive conductivity sensor for measurement in pure and ultrapure water and in Ex applications
• Order as per product structure (--> online Configurator, www.products.endress.com/cls15d)
• Technical Information TI109C/07/EN

34 Endress+Hauser
Liquiline CM442/CM444/CM448

Condumax CLS16D
• Hygienic conductivity sensor for measurement in pure and ultrapure water and in Ex applications
• With EHEDG and 3A certificates
• Order as per product structure (--> online Configurator, www.products.endress.com/cls16d)
• Technical Information TI227C/07/EN
Condumax CLS21D
• Two-electrode sensor in fixed cable and plug-in head version
• Order as per product structure (--> online Configurator, www.products.endress.com/cls21d)
• Technical Information TI085C/07/EN

Oxygen sensors
Oxymax COS51D
• Amperometric sensor for dissolved oxygen, with Memosens technology
• Order as per product structure (--> online Configurator, www.products.endress.com/cos51d)
• Technical Information TI413C/07/EN
Oxymax COS61D
• Optical oxygen sensor for drinking water and industrial water measurement
• Measuring principle: quenching
• Memosens protocol
• Material: stainless steel 1.4571 (AISI 316Ti)
• Order as per product structure (--> online Configurator, www.products.endress.com/cos61d)
• Technical Information TI387C/07/EN
Oxymax COS22D
• Sterilizable sensor for dissolved oxygen
• Order as per product structure (--> online Configurator, www.products.endress.com/cos22d)
• Technical Information TI446C/07/EN

Chlorine sensors
CCS142D
• Membrane-covered amperometric sensor for free chlorine
• Memosens technology
• Measuring range 0.01 to 20 mg/l
• Order as per product structure (--> online Configurator, www.products.endress.com/ccs142d)
• Technical Information TI419C/07/EN

Ion selective sensors


ISEmax CAS40D
• Ion selective sensors
• Order as per product structure (--> online Configurator, www.products.endress.com/cas40d)
• Technical Information TI491C/07/EN

Turbidity sensors
Turbimax CUS51D
• For nephelometric measurements of turbidity and solids in wastewater
• 4-beam scattered light method
• With Memosens protocol
• Order as per product structure (--> online Configurator, www.products.endress.com/cus51d)
• Technical Information TI461C/07/EN

SAC and nitrate sensors


Viomax CAS51D
• SAC and nitrate measurement in drinking water and wastewater
• With Memosens protocol
• Order as per product structure (--> online Configurator, www.products.endress.com/cas51d)
• Technical Information TI459C/07/EN

Endress+Hauser 35
Liquiline CM442/CM444/CM448

Interface measurement
Turbimax CUS71D
• Immersion sensor for interface measurement
• Ultrasonic interface sensor
• Order as per product structure (--> online Configurator, www.products.endress.com/cus71d)
• Technical Information TI490C/07/EN

Additional functionality Hardware extension modules


Kit CM442/CM444/CM448/CSF48: extension module AOR
• 2 x relay, 2 x 0/4 to 20 mA analog output
• Order no. 71111053
Kit CM442/CM444/CM448/CSF48: extension module 2R
• 2 x relay
• Order no. 71125375
Kit CM442/CM444/CM448/CSF48: extension module 4R
• 4 x relay
• Order no. 71125376
Kit CM442/CM444/CM448/CSF48: extension module 2AO
• 2 x 0/4 to 20 mA analog output
• Order no. 71135632
Kit CM444/CM448/CSF48: extension module 4AO
• 4 x analog output 0/4 to 20 mA
• Order no. 71135633
Kit CM444/CM448/CSF48: extension module 2DS
• 2 x digital sensor, Memosens
• Order no. 71135631
Kit CM442/CM444/CM448/CSF48: extension module 2AI
• 2 x 0/4 to 20 mA analog input
• Order no. 71135639
Kit CM442/CM444/CM448/CSF48: extension module 485
• Ethernet configuration
• Can be extended to PROFIBUS DP or Modbus RS485 or Modbus TCP using activation code
• Order no. 71135634
Upgrade kit CM442/CM444/CM448/CSF48
• Extension module 485
• PROFIBUS DP (+ Ethernet configuration)
• Order no. 71140888
Upgrade kit CM442/CM444/CM448/CSF48
• Extension module 485
• Modbus RS485 (+ Ethernet configuration)
• Order no. 71140889
Upgrade kit CM442/CM444/CM448/CSF48
• Extension module 485
• Modbus TCP (+ Ethernet configuration)
• Order no. 71140890
Kit CM442: upgrade kit to CM444/CM448
• Extension power supply unit and backplane
• Basic module BASE-E
• Order no. 71135644

36 Endress+Hauser
Liquiline CM442/CM444/CM448

Software and activation codes


SD card with Liquiline firmware
• Industrial Flash Drive, 1 GB
• Order no. 71127100
Activation code for digital HART communication
• Order no. 71128428
Activation code for PROFIBUS DP
• Order no. 71135635
Activation code for Modbus RS485
• Order no. 71135636
Activation code for Modbus TCP
• Order no. 71135637
Kit CM442: activation code for 2nd digital sensor input
• Order no. 71114663
Kit CM444/CM448: upgrade code for 2 x 0/4 to 20 mA for BASE-E
• Order no. 71140891

Other accessories SD card


SD card
• Industrial Flash Drive, 1 GB
• Order no. 71110815

Cable glands
Kit CM44x: gland M
• Set, 6 pieces
• Order no. 71101768
Kit CM44x: gland NPT
• Set, 6 pieces
• Order no. 71101770
Kit CM44x: gland G
• Set, 6 pieces
• Order no. 71101771
Kit CM44x: dummy plug for cable gland
• Set, 6 pieces
• Order no. 71104942

M12 socket and cable junction


Kit CM42/CM442/CM444/CM448: external CDI socket
• Socket with terminated connecting cables and counter nut
• Order no. 51517507
Kit CM442/CM444/CM448/CSF48: M12 socket for digital sensors
• Pre-terminated
• Order no. 71107456
Kit CM442/CM444/CM448/CSF48: M12 socket for PROFIBUS DP/Modbus RS485
• B-coded, pre-terminated
• Order no. 71140892
Kit CM442/CM444/CM448/CSF48: M12 socket for Ethernet
• D-coded, pre-terminated
• Order no. 71140893
Kit CM442/CM444/CM448/CSF48: RJ45 socket for Ethernet
• Pre-terminated
• Order no. 71140895
Cable junction with Velcro strip
• 4 pieces, for sensor cable
• Order no. 71092051

Endress+Hauser 37
TI00444C/07/EN/15.12
71189225