SI-132 v1.1 SME300000081
EXCELL PRECISION CO., LTD
222 CONTENT 222
BEFORE USING THE SCALE ......................................................................................... 2
PREPARING TO USE THE SCALE ................................................................................. 3
CHAPTER 1 INTRODUCTION......................................................................................... 4
1-1 FEATURES AND SPECIFICATION..................................................................... 4
1-2 SCALE APPEARANCE ....................................................................................... 5
1-3 POWER SUPPLY................................................................................................ 5
1-4 DISPLAY ............................................................................................................. 6
1-5 KEYBOARD FUNCTION..................................................................................... 7
1-6 OPERATING THE SCALE .................................................................................. 8
CHAPTER 2 ADVANCED FUNCTIONS ......................................................................... 10
2-1 ADVANCED FUNCTION SETTING TABLE....................................................... 10
2-2 ADVANCED FUNCTION SETTING WORKFLOW ............................................ 12
2-3 GENERAL FUNCTION SETTING ............................................ 13
2-3-1 Automatic Backlight Function Setting ............................. 14
2-3-2 Automatic Power-off Timer Setting ................................. 15
2-3-3 Hi/Lo/OK Function Setting .............................................. 16
2-3-4 Restore to the Default Setting ........................................ 17
2-3-5 Noise Filter Setting ......................................................... 18
2-3-6 Hold Function Setting .................................................... 19
2-3-7 Auto Unit Weight Averaging Setting ............................. 20
2-4 WEIGHT CALIBRATION ............................................................... 21
2-5 RS232 SERIAL INTERFACE SETTINGS ................................ 22
2-5-1 Baud Rate Setting ......................................................... 23
2-5-2 Communication Protocol Setting ................................... 24
2-5-3 Output Format Setting .................................................... 25
2-5-4 Continuous Transmission Setting .................................. 26
2-5-5 The selection of the Continuous Transmission Rate ...... 27
2-5-6 Auto Transmission at Zero ............................................. 28
2-5-7 Reset of Auto Transmission ........................................... 29
2-5-8 Output Condition Setting ............................................... 30
APPENDIX 1 7 SEGMENT DISPLAY CHARACTERS ................................................... 34
APPENDIX 2 ASCII CODE TABLE................................................................................. 35
SI-132 v1.1 1 SME300000081
EXCELL PRECISION CO., LTD
BEFORE USING THE SCALE
Thank you for purchasing an EXCELL High Precision Electronic Digital
Scale. In order to use the scale properly, please read this User Manual
carefully before use. If you have a problem concerning the scale, please
contact your supplier.
INSTRUCTION FOR USE
1) Please keep the scale in a cool dry place. Do not store it at high
temperature.
2) Do not allow any liquids to come into contact with the scale. If necessary
wipe the scale with a dry soft clothe.
3) Avoid objects impacting with the scale. Do not drop loads onto the scale
or subject the weigh pan to any strong shock loads.
4) The load placed on the weigh pan must not exceed the maximum
weighing capacity of the scale.
5) If the scale is not going to be used for some time, please clean it and
store it in a plastic bag in dry conditions. A desiccant sachet may be
included to prevent any moisture build up.
6) Do not mix different types of dry batteries or mix used dry batteries with
new dry batteries.
SI-132 v1.1 2 SME300000081
EXCELL PRECISION CO., LTD
PREPARING TO USE THE SCALE
1. Locate the scale on a firm level surface free from vibrations for
accurate weight readings.
2. Adjust the four levelling feet (if fitted) to set the scale pan level using
the spirit level bubble located at the front of the scale.
3. Avoid operating the scale in direct sunlight or drafts of any kind.
4. If possible avoid connecting the scale to ac power outlet sockets which
are adjacent to other appliances to minimise the possibility of
interference affecting the performance of the scale.
5. Remove any weight that might be on the weigh pan before the scale is
switched on and avoid leaving weight on the pan for long periods of
time
6. All goods weighed should be placed in the centre of the weigh pan for
accurate weighing. The overall dimensions of the goods being
weighed should not exceed the dimension of the weigh pan.
7. Once the scale has been powered on, it will go through an LCD display
test and it is ready for use when the display shows zero.
8. The scale requires 15~20 minutes warm up before operation to ensure
best accuracy.
9. Please note when the symbol keeps flashing on the screen, the
batteries need to be replaced.
SI-132 v1.1 3 SME300000081
EXCELL PRECISION CO., LTD
CHAPTER 1 INTRODUCTION
1-1 FEATURES AND SPECIFICATION
Features:
n 30,000d display resolution
n Large LCD display (digit height 8.5mm x 18.5mm) with LED backlight
n Kilogram (kg), gram (g), pound (lb) and ounce (oz) weighing modes
n Hold function; Check mode Lo/Hi/OK
n Adjustable gravity value
n Auto calibration; Full range tare; Auto-zero tracking; Simple counting;
Gross/Net indication, Unit weight average function
n Low power indication and auto power off
n With level bubble, Adjust feet
Options:
n RS-232C or Serial printer output
n Stainless steel platter (198x176mm)
Specifications:
Model Capacity Division Resolution Min. Cap.
SI132-1.5 1,500g (3 lb) 0.05g (0.0001 lb) 1/30000 1g
SI132-3 3,000g (6 lb) 0.1g (0.0002 lb) 1/30000 2g
SI132-6 6,000g (12 lb) 0.2g (0.0005 lb) 1/30000 4g
SI132-15 15,000g (30 lb) 0.5g (0.001 lb) 1/30000 10g
Operating Temperature 0°C ~ 40°C (32°F ~ 104°F)
Power Source 4 x UM-2 1.5V dry batteries or AC adaptor DC 9V
Weigh pan Size 196 x 174 mm (Plastic pan)
198 x 176mm (Stainless steel pan)
SI-132 v1.1 4 SME300000081
EXCELL PRECISION CO., LTD
1-2 SCALE APPEARANCE
196 mm The package includes:
174 mm 1. Scale 1 off
2. AC adaptor 1 off
80.4 mm 3. User Manual 1 off
* Stainless steel weigh pan is an
optional depends on the model
you purchase.
When you first unseal the
product package if you find any
of the items above are missing,
284 mm
contact your supplier.
Notice: Dry batteries are not
213 mm included in the product
package.
1-3 POWER SUPPLY
POWER SUPPLY SELECTION
1. 4 x UM-2 1.5V dry batteries (not included)
2. AC adaptor DC 9V
POWER CONSUMPTION
Approximately DC 14mA (Scale)
Approximately DC 24 mA (Scale + Display backlight)
LOW BATTERY WARNING
Please note when the ( ) symbol keeps flashing on the display, the batteries
should be changed.
SI-132 v1.1 5 SME300000081
EXCELL PRECISION CO., LTD
1-4 DISPLAY
1 2 3 4 5 6 7
HI : The weight on weigh pan is greater than the high limit
OK : The weight on weigh pan is equal to the OK limit
LO : The weight on weigh pan is lower than the check value
kg : kg units. When “kg” is displayed, it means the weight shown is in kg
lb : Pound units. When “lb” is displayed, it means the weight shown is lb
Pcs : Piece units. When “Pcs” is displayed, it means the scale is in “sampling
and counting” mode
à0ß : Zero status indication, when displayed the scale is at the centre of its zero
band
Net : The display shows the goods weight, not including the weight of any
container. This Net status indication is on when the TARE function is used
: Battery status indication. When this symbol is flashing replace the
batteries.
61 : The weight is stable.
STABLE
62 : The scale is in the gross mode. The display shows the goods and any
GROSS container weight. This Gross status indication is on when the TARE
function is used.
63 : The unit weight is not sufficient. When the icon is on, the counting
P cs function is operational but the count may contain errors.
64 .
. .
. . .
: The sampling size is not sufficient When the icon is on, the counting
. . . .
function is operational but the count may contain errors.
65 : The Hold function is in use.
Hold
66 : “GN”, “dwt”, or “carat” units. The actual unit depends on the model of the
scale.
67 : ounce unit. When “oz” is on, it means the scale is weighing in ounces
oz
SI-132 v1.1 6 SME300000081
EXCELL PRECISION CO., LTD
1-5 KEYBOARD FUNCTION
ON/ZERO KEY
This key possesses two functions: Power On and Zero function.
OFF KEY
When the scale is switched on, press the OFF key, the scale will switch off.
TARE KEY
The tare function will not operate during the following conditions:
(1) When the scale powers on if the weight is negative and after a container is
placed on the weigh pan if the weight is still below zero.
(2) The tare value is over the full scale capacity.
UNITS KEY
Press the UNITS key to switch weight units; the icons will indicate the active units.
NET/GROSS KEY
In the Tare mode, the screen displays the “TARE” icon; press the NET/GROSS key
to switch between the “Net value” and the “Gross value”.
SI-132 v1.1 7 SME300000081
EXCELL PRECISION CO., LTD
1-6 OPERATING THE SCALE
POWER ON
When the scale is off, press the ON/ZERO key, the scale will switch on.
POWER OFF
When the scale is on, press the OFF key, the scale will switch off.
ZERO
When the weigh pan is empty (free of load) and the display is not showing zero,
press the ON/ZERO key to zero the scale. At zero, the “à0ß” indication is on.
4 When the weight value is within the zero range, the zero function operates to
zero the scale or cancel the tare function.
4 Zero range: The OIML & NTEP models have a zero range of ± 2% of Full Scale.
The Sri Lanka model has a zero range of ± 4% of Full Scale.
SWITCHING UNITS
Press the UNITS key to switch weight units, the icons or arrows will indicate the
active units as appropriate. The units available are dependent on the exact scale
model.
4 After power off, the scale will memorize the active units. When the scale is
powered on again, it displays the previously active units.
TARE FUNCTION
(1) Put a container on the weigh pan and after the weight is stable, press the TARE
key to zero the weight of the container. The screen displays the “Net” icon.
(2) Put the goods in the container, the screen displays the net weight value of the
goods.
(3) Remove the full container; the screen displays the negative weight value of the
container. At this time pressing the TARE key again will cancel the tare and
the scale reverts back to zero. The “Net” icon is switched off.
4 The tare function can be operated continually to the full weighing capacity of
the scale.
4 Continual tare operation is adding or removing tare objects on weigh pan and
pressing the TARE key each time.
SI-132 v1.1 8 SME300000081
EXCELL PRECISION CO., LTD
NET/GROSS FUNCTION
In the Tare mode, the screen displays the “Net” icon, press the NET/GROSS key to
switch between the “Net value” and the “Gross value”.
4 When the 6GROSS icon is on, the weight value on the display is the total
amount of the tare value and net value.
4 At the Gross status, only OFF and NET/GROSS keys are functional.
4 NET/GROSS key is only used in Tare mode.
SIMPLE COUNTING FUNCTION
(1) Use the UNITS key to enter into the “PCS” mode
(2) Press the NET/GROSS key to select the counting sample size
(S = 10, S = 20, S = 50, S = 100, S=200). The LCD shows , ,
, , in order.
(3) Put the samples on the weigh pan and press the UNITS key, the screen
displays “------”. After the sampling process is complete, put the goods on the
weigh pan and the screen shows the quantity of the items.
4 The sample weight should be heavier then the minimum capacity of the scale
.
(20d), If not the arrow pointing to the icon will be activated.
. .
. . .
. . . .
4 The weight of a sample should be heavier than the 0.2d (d=division), or the
P cs
arrow pointing to the icon will be on.
.
. .
4 When the . . .
. . . . or P cs are indicated the scale is still operational but the
count may contain errors.
4 To power off in this mode, the scale will memorize the “Pcs” unit. When the
scale is powered on again, it directly enters the simple counting mode.
4 While the “Auto unit weight average” function is available in the Advanced
Function, the goods on the weigh pan are 5pcs greater than the sample size
and less than double the sample size, the scale will automatically re-sample the
unit weight.
SI-132 v1.1 9 SME300000081
EXCELL PRECISION CO., LTD
CHAPTER 2 ADVANCED FUNCTIONS
2-1 ADVANCED FUNCTION SETTING TABLE
Below is an overview of the advanced functions. For detailed settings refer to the
following sections:-
DISPLAY LEVEL 1 FUNCTION DISPLAY LEVEL 2 FUNCTIONS
Exit the ADVANCED --- ---
FUNCTION setting mode
General Function setting Return to the
mode ADVANCED FUNCTION
setting menu
Automatic backlight
function setting
Automatic power-off timer
setting
Hi/Lo/OK function setting
Restore the default
settings
Noise filter setting
Hold function setting
Auto unit weight
averaging setting
External Weight --- ---
Calibration
RS232 Bi-direction Return to the
Function setting ADVANCED FUNCTION
setting mode menu
Baud rate setting
Communication protocol
setting
Output format setting
SI-132 v1.1 10 SME300000081
EXCELL PRECISION CO., LTD
DISPLAY LEVEL 1 FUNCTION DISPLAY LEVEL 2 FUNCTIONS
Continuous Transmission
setting
Continuous data
transmission rate
Auto transmission at Zero
Reset of auto
transmission
Output condition setting
SI-132 v1.1 11 SME300000081
EXCELL PRECISION CO., LTD
2-2 ADVANCED FUNCTION SETTING WORKFLOW
In the weighing mode, press the NET/GROSS and ON/ZERO keys at the same
time to enter the Advanced Function setting mode. The LCD shows
.
Overall workflow of the Advanced Function setting mode:
Press UNITS key
Press ON/ZERO key
Press UNITS key
Press ON/ZERO key
Press UNITS key
Press ON/ZERO key
Press UNITS key
Press ON/ZERO key
⇒ General Function setting mode
⇒ External Weight Calibration
⇒ RS232 Bi-direction Function setting
⇒ Exit the Advanced Function setting mode
Refer to the following sections for the detailed operation
procedures of each function setting.
SI-132 v1.1 12 SME300000081
EXCELL PRECISION CO., LTD
2-3 GENERAL FUNCTION SETTING
n There are 7 functions in the general function setting mode from
to .
Workflow of the General Function setting:
ACTIONS DISPLAY NOTE
ON/ZERO key ⇒ Upward key (from 0 to 9)
UNITS key ⇒ Downward key (from 9 to 0)
Press TARE key
TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
Use the keys to select
the function
Key in “00” and press
TARE key to return to
the Advanced function
setting menu
⇒ Return to the Advanced Function Setting Mode Menu
⇒ Automatic Backlight Function Settings
⇒ Automatic Power-off Timer Settings
⇒ Hi/Lo/OK Settings
⇒ Restore the Default Settings
⇒ Noise Filter Settings
⇒ Hold Function Settings
⇒ Auto Unit Weight Averaging Setting
Refer to the following sections for detailed operation procedures of
each setting.
SI-132 v1.1 13 SME300000081
EXCELL PRECISION CO., LTD
2-3-1 Automatic Backlight Function Setting
Select in the General Function setting mode to
change the backlight function setting.
ACTIONS DISPLAY NOTE
Press TARE key
LCD displays the last status
Use ON/ZERO or UNITS
key to select function “on”
or “off
Press TARE key to go
back to the Advanced
function setting menu
Automatic backlight function
When the weight is over 10d, the display backlight will be on. After the weight
is stable for 10 seconds or when the scale returns to zero, the display
backlight switches off.
SI-132 v1.1 14 SME300000081
EXCELL PRECISION CO., LTD
2-3-2 Automatic Power-off Timer Setting
Select in the General Function setting mode to
change the automatic power-off timer setting.
ACTIONS DISPLAY NOTE
Press TARE key
LCD displays the last status
Use ON/ZERO or UNITS Automatic power-off timer setting
key to key in parameter Use ON/ZERO or UNITS key to key in
parameter
Press TARE key to go 0 ⇒ No auto power-off
back to the Advanced 1 ⇒ When the scale is idle for 1
function setting menu minute, the scale
automatically switches off
2 ⇒ When the scale is idle for 2
minutes, the scale
automatically switches off
9 ⇒ When the scale is idle for 9
minutes, the scale
automatically switches off
ON/ZERO key ⇒ Upward key (from 0 to 9)
UNITS key ⇒ Downward key (from 9 to 0)
TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
Automatic power-off function
When the weight on weigh pan is less than 10d or keeps idle for the set time, the
scale will automatically switch off.
SI-132 v1.1 15 SME300000081
EXCELL PRECISION CO., LTD
2-3-3 Hi/Lo/OK Function Setting
Select in the General Function setting mode to
set the Hi/Lo/OK function. This function is available in all unit modes. In
one specific unit mode, enter to set the Hi/Lo/OK values.
ACTIONS DISPLAY NOTE
4 When the high limit and low limit are both set as “0”, the Hi/Lo/OK function is
disabled.
ON/ZERO key ⇒ Upward key (from 0 to 9)
UNITS key ⇒ Downward key (from 9 to 0)
Press TARE key
TARE key ⇒ Move the cursor to right
lb
. NET/GROSS key ⇒ Move the cursor to left.
Key in the desired high LCD displays the last status
limit value
lb
.
Press TARE key
lb LCD displays the last status
.
Key in the desired low limit
value
lb
.
Press TARE key
LCD displays the last status
Alarm setting
Key in the parameter (a) (b) (c)
setting for alarm The definition of a, b, c positions:
(a) ⇒ 1 = beep sound on
0 = beep sound off
Press TARE key (b) ⇒ 1 = beep sound on when stable
0 = beep sound on when unstable
(c) ⇒ 1 = beep sound on when the weight is
between high and low limits
0 = beep sound on when the weight is not
between high and low limits and
higher than 10d
SI-132 v1.1 16 SME300000081
EXCELL PRECISION CO., LTD
2-3-4 Restore to the Default Setting
Select in the General Function setting mode to
restore to the default setting.
ACTIONS DISPLAY NOTE
ON/ZERO key ⇒ Upward key (from 0 to 9)
UNITS key ⇒ Downward key (from 9 to 0)
Press TARE key
TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
Use ON/ZERO or UNITS ⇒ Return (Cancel the
keys to select “return” or restoration)
“format”
⇒ Restore to the default
Press TARE key setting
4 The default setting includes the following:
1) External weight calibration
2) HI/LO/OK setting value
3) Noise filter setting (External)
4) Sampling setting for the counting function
4 In approved models, set as 1 or 3, setting is not
available.
SI-132 v1.1 17 SME300000081
EXCELL PRECISION CO., LTD
2-3-5 Noise Filter Setting
Select in the General Function setting mode to
set the noise filter setting.
ACTIONS DISPLAY NOTE
4 When modifying , the parameters of remain un-altered.
Returning to zero point setting
LCD displays the last status
Press TARE key Returning to the zero point setting
Use ON/ZERO or UNITS key to key in the
parameters or zero point
Default setting = 0
Use ON/ZERO or UNITS 0 ⇒ No skip 5 ⇒ skip 5d
key to key in the 1 ⇒ skip 1d 6 ⇒ skip 6d
parameters 2 ⇒ skip 2d 7 ⇒ skip 7d
3 ⇒ skip 3d 8 ⇒ skip 8d
4 ⇒ skip 4d 9 ⇒ skip 9d
4 When the weight on the scale is over
Press TARE key 1/3 full capacity, the function is on.
Digital switch & Stabilization range
setting
Use ON/ZERO or UNITS LCD displays the last parameter setting
key to key in the
parameters Digital switch & Stabilization range
setting
Use ON/ZERO or UNITS keys to key in
Press TARE key the parameters. Default setting = 0
Parameter 0 ~ 9, the larger the number the
more stable the weight.
Use ON/ZERO or UNITS Filter parameter setting
key to key in the LCD displays the last parameter setting
parameters
Filter parameter setting
Use ON/ZERO or UNITS keys to key in
Press TARE key
the parameters. Default setting = 5
Parameter 0 ~ 9, the larger the number,
the faster the filter response. Fast
response can lead to weight instability.
4 In approved models, set as 1 or 3, setting is not available
SI-132 v1.1 18 SME300000081
EXCELL PRECISION CO., LTD
2-3-6 Hold Function Setting
Select in the General Function setting mode to
set the hold function.
ACTIONS DISPLAY NOTE
Press TARE key Hold function setting
LCD displays the last parameter setting
Hold function setting
Use ON/ZERO or UNITS
Use ON/ZERO or UNITS keys to key in the
key to key in the
parameters
parameters
Default setting = 0
0 ⇒ Hold function disabled
1 ⇒ “Peak hold” mode
Press TARE key
2 ⇒ “Stable hold 1” mode
3 ⇒ “Stable hold 2” mode
Key in the parameter “00”
ON/ZERO key ⇒ Upward key (from 0 to 9)
Press TARE key to
complete the general UNITS key ⇒ Downward key (from 9 to 0)
setting procedures
TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
= Hold function disabled
= “Peak hold” mode
Keep displaying the maximum weight when the weight is continually changing
To exit this mode, press any key.
= “Stable hold 1” mode
When the weight is stable, the LCD shows the current weight value. To exit
this mode, press any key.
= “Stable hold 2” mode
When the weight is stable, the LCD shows the current weight value. When the
weight returns to zero (<10d), the hold mode is cancelled automatically.
SI-132 v1.1 19 SME300000081
EXCELL PRECISION CO., LTD
2-3-7 Auto Unit Weight Averaging Setting
Press TARE key Auto unit weight averaging setting
LCD displays the last parameter setting
Auto unit weight averaging setting
Use ON/ZERO or UNITS key Use ON/ZERO or UNITS keys to key in
to key in the parameters the parameters
Default setting = 0
0 ⇒ Auto unit weight averaging
1 ⇒ Non auto unit weight averaging
Press TARE key
Key in the parameter “00”
Press TARE key to
complete the general ON/ZERO key ⇒ Upward key (from 0 to 9)
setting procedures
UNITS key ⇒ Downward key (from 9 to 0)
TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
SI-132 v1.1 20 SME300000081
EXCELL PRECISION CO., LTD
2-4 WEIGHT CALIBRATION
In the weighing mode, press the NET/GROSS and ON/ZERO keys at the same time
to enter the Advanced Function setting mode. The LCD shows and
use the NET/GROSS or UNITS key to select to enter the weight
calibration mode.
ACTIONS DISPLAY NOTE
Press TARE key
The LCD displays the calibration
kg weight value (the maximum capacity
. of 1st unit) and the unit. The digit on
the right side keeps flashing.
When the digit on the right
flashes, ensure that the
weigh pan is empty. Then
Press the TARE The scale is calculating the internal
kg value of the zero point. After
key.
stabilization, the LCD shows the
After stabilization calibration weight value.
kg
.
Place the displayed weight
on the weigh pan. Press the
TARE key. The scale
calculates the full capacity
internal value and after ON/ZERO key ⇒ Upward key (from 0 to 9)
stabilization, the buzzer
UNITS key ⇒ Downward key (from 9 to 0)
beeps 3 times. Remove the
weight from the weigh pan TARE key ⇒ Move the cursor to right
and the procedure
is completed. NET/GROSS key ⇒ Move the cursor to left.
4 In approved models, set as 1 or 3, then is disabled.
4 Weight calibration conditions:
The calibration weight value placed on the weight pan must be over e100, and the
standard deviation of the weight must be within 10%.
SI-132 v1.1 21 SME300000081
EXCELL PRECISION CO., LTD
2-5 RS232 SERIAL INTERFACE SETTINGS
In the weighing mode, press the NET/GROSS and ON/ZERO keys at the same time
to enter the Advanced Function setting mode. The LCD shows and
use the NET/GROSS or UNITS key to select to enter the RS232
serial interface setting mode.
4 RS232 serial interface settings span ~ ,
8 settings.
ACTIONS DISPLAY NOTE
ON/ZERO key ⇒ Upward key (from 0 to 9)
UNITS key ⇒ Downward key (from 9 to 0)
Press TARE key
TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
Key in the parameter 00 ~
08
Key in the parameter “00”,
then press TARE key to
exit RS232 serial interface
settings
⇒ Exit the RS232 serial interface setting mode
⇒ Baud rate setting
⇒ Communication protocol setting
⇒ Output format setting
⇒ Continuous transmission setting
⇒ The selection of continuous transmission rate
⇒ Auto Transmission at Zero
⇒ Reset of Auto Transmission
⇒ Output condition setting
SI-132 v1.1 22 SME300000081
EXCELL PRECISION CO., LTD
2-5-1 Baud Rate Setting
Select in the RS232 serial interface setting mode
to set the Baud Rate.
ACTIONS DISPLAY NOTE
Press TARE key
Baud rate setting
LCD displays the last value
4 Default value: 9600 (bits/sec)
Use ON/ZERO or UNITS
key to select the desired
Baud rate Baud rate setting
Use ON/ZERO or UNITS keys to select the
desired Baud rate 600、1200、2400、
4800、9600、19200、38400 (bits/sec)
Press TARE key
Key in the parameter “00”
ON/ZERO key ⇒ Upward key (from 0 to 9)
UNITS key ⇒ Downward key (from 9 to 0)
Press the TARE key to TARE key ⇒ Move the cursor to right
exit RS232 serial interface
settings NET/GROSS key ⇒ Move the cursor to left.
SI-132 v1.1 23 SME300000081
EXCELL PRECISION CO., LTD
2-5-2 Communication Protocol Setting
Select in the RS232 serial interface setting mode
to set the Communication Protocol.
ACTIONS DISPLAY NOTE
Press TARE key
Communication protocol setting
LCD shows the last setting value.
4 Default setting: n 8 1
Use ON/ZERO or UNITS
key to select the protocol
setting Communication protocol setting
Use ON/ZERO or UNITS keys to select
the protocol setting: n 8 1, E 7 1, O 7 1
Press TARE key
Key in the parameter “00”
ON/ZERO key ⇒ Upward key (from 0 to 9)
Press TARE key to exit
UNITS key ⇒ Downward key (from 9 to 0)
RS232 serial interface
settings TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
SI-132 v1.1 24 SME300000081
EXCELL PRECISION CO., LTD
2-5-3 Output Format Setting
Select in the RS232 serial interface setting mode
to set the Output Format.
ACTIONS DISPLAY NOTE
RS232 output format setting
LCD shows the last setting.
4 Default setting:
Press TARE key
RS232 output format setting
Use ON/ZERO or UNITS keys to select the
Use ON/ZERO or UNITS output format:
key to select the output
format = Same data as the scale
= Gross weight
= Net weight
Press TARE key = Same data as the scale in
simple format
= Same gross data as the
scale in simple format
= Same net data as the scale
Key in the parameter “00” in simple format
= Hi/Lo/OK status + Same
data as the scale in simple
format
= Hi/Lo/OK status + Simple
Press the TARE key to gross weight
exit RS232 serial interface = Hi/Lo/OK status + Simple
settings net weight
= Tare value
ON/ZERO key ⇒ Upward key (from 0 to 9)
UNITS key ⇒ Downward key (from 9 to 0)
TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
SI-132 v1.1 25 SME300000081
EXCELL PRECISION CO., LTD
2-5-4 Continuous Transmission Setting
Select in the RS232 serial interface setting mode
to set the Continuous Transmission Setting.
ACTIONS DISPLAY NOTE
Press TARE key
Transmission setting
LCD shows the last setting
4 Default setting:
Use ON/ZERO or UNITS
keys to select the Transmission setting
transmission type Use ON/ZERO or UNITS keys to select the
transmission type:
= Command mode
Press TARE key = Continuous transmission
= Constantly auto
transmission
= Shut down RS232
Key in the parameter “00”
ON/ZERO key ⇒ Upward key (from 0 to 9)
Press the TARE key to exit UNITS key ⇒ Downward key (from 9 to 0)
RS232 serial interface
settings TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
SI-132 v1.1 26 SME300000081
EXCELL PRECISION CO., LTD
2-5-5 The selection of the Continuous Transmission Rate
Select in the RS232 serial interface setting mode
to set the Continuous Transmission Rate.
ACTIONS DISPLAY NOTE
Press TARE key
Continuous transmission rate setting
The LCD shows the last setting.
4 Default setting:
Use the ON/ZERO or
UNITS keys to select the
continuous Continuous transmission rate
transmission Use ON/ZERO or UNITS keys to select
rate the continuous transmission rate:
= 1 time/sec
Press TARE key = 2 times/sec
= 4 times/sec
= 8 times/sec
= 16 times/sec
Key in the parameter “00” = More than16 times/sec
ON/ZERO key ⇒ Upward key (from 0 to 9)
Press the TARE key to exit UNITS key ⇒ Downward key (from 9 to 0)
RS232 serial interface
settings TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
SI-132 v1.1 27 SME300000081
EXCELL PRECISION CO., LTD
2-5-6 Auto Transmission at Zero
Select in the RS232 serial interface setting mode
to set the Auto Transmission at Zero.
ACTIONS DISPLAY NOTE
Press TARE key
Auto transmission at zero
LCD shows the last setting
4 Default setting:
Use ON/ZERO or UNITS
keys to key in the desired
parameters Auto transmission at zero
Use ON/ZERO or UNITS keys to key in the
desired parameters:
= 00 division of weight value
Press TARE key = 01 division of weight value
= 99 divisions of weight value
Key in the parameter “00”
ON/ZERO key ⇒ Upward key (from 0 to 9)
UNITS key ⇒ Downward key (from 9 to 0)
Press the TARE key to exit
RS232 serial interface TARE key ⇒ Move the cursor to right
settings
NET/GROSS key ⇒ Move the cursor to left.
4 When the parameter is set as , the “Auto transmission” function is not
available. It is because when the zero is stable, the transmission becomes
“Continuous Transmission”.
SI-132 v1.1 28 SME300000081
EXCELL PRECISION CO., LTD
2-5-7 Reset of Auto Transmission
Select in the RS232 serial interface setting mode
to Reset of Auto Transmission.
ACTIONS DISPLAY NOTE
Press TARE key Reset of auto transmission
The LCD shows the last setting
4 Default setting:
Use ON/ZERO or UNITS
keys to key in the desired
parameters Reset of auto transmission
Use ON/ZERO or UNITS keys to key in the
desired parameters:
= 00 division of weight value
Press TARE key = 01 division of weight value
= 99 divisions of weight value
Key in the parameter “00”
ON/ZERO key ⇒ Upward key (from 0 to 9)
Press TARE key to exit UNITS key ⇒ Downward key (from 9 to 0)
RS232 bi-direction function
setting TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
4 When the parameter is set as , the “Auto transmission” function is not
available. It is because when the zero is stable, the transmission becomes
“Continuous Transmission”.
SI-132 v1.1 29 SME300000081
EXCELL PRECISION CO., LTD
2-5-8 Output Condition Setting
Select in the RS232 serial interface setting mode
to Reset of Auto Transmission.
ACTIONS DISPLAY NOTE
Press TARE key
Output condition setting
The LCD shows the last setting.
4 Default setting:
Use ON/ZERO or UNITS
keys to select the desired
setting Output condition setting
Use ON/ZERO or UNITS keys to select the
desired setting:
= Output all the time
Press TARE key = Stable weight output only (No
output when under/overload or unstable)
Key in the parameter “00”
ON/ZERO key ⇒ Upward key (from 0 to 9)
Press the TARE key to UNITS key ⇒ Downward key (from 9 to 0)
exit RS232 serial interface
settings TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
SI-132 v1.1 30 SME300000081
EXCELL PRECISION CO., LTD
2 Command mode
Command Format A
Host Command
Slave Command
MZ Zero SO Command mode
MT Tare UA Switch to the first unit
MG Gross weight UB Switch to the second unit
MN Net weight UC Switch to the third unit
CT Clear TARE value UD Switch to the forth unit
SC Continuous transmission UE Switch to the fifth unit
SA Auto transmit UF Switch to the sixth unit
% Stop continuous transmission and enter the command mode
Note: UA ~ UF settings are dependent the model of the scale
Command Format B
Host Command
Slave Data
RW Read current weight RH Read Gross (simple)
RG Read Gross weight RI Read Net (simple)
Read comparison situation +
RN Read Net weight RJ current display of weight
(simple)
Read comparison situation +
RT Read TARE RK
Gross (simple)
Read current display of Read comparison situation +
RB RL
weight (simple) Net (simple)
Note: a. add % before the command to read continuously
b. add # before the command to transmit a stable value
Read weight comparison setting value RS™™££
™™: Groups(00 ~ 09) ££: Setting Items
HI Show “HI” presetting value
LO Show “LO” presetting value
Note : ™™(Group) is various depended on different models
00 ⇒ The first group
01 ⇒ The second group
02 ⇒ The third group
EX: RS02LO<CR><LF> Show “LO” presetting value
ANS: RS02LOXXXXXX<CR><LF>
SI-132 v1.1 31 SME300000081
EXCELL PRECISION CO., LTD
Command Format C
Host Command+ Data
Slave Command+ Data
Write weight comparison setting value WS™™££XXXXXX
™™: Groups(00 ~ 09) ££: Setting Items XXXXXX: Setting Value
HI Write “HI” setting value
LO Write “LO” setting value
Note : ™™ (Group) is various depended on different models
00 ⇒ The first group
01 ⇒ The second group
02 ⇒ The third group
EX: WS00HI001000<CR><LF> Write “HI” setting value
ANS: WS00HI001000<CR><LF>
Command Format D
Host Data
Slave
Position of
Value (e.g. Price) decimal point CR LF
1 2 3 4 5 6 1
When the SI-132 receives this data format, it will transfer the data and
display it on its LCD.
Note: The function is effective, when the weight value is over 0.
4 Error messages:
E1: Wrong command
E2: Command format error (Wrong parameters)
E3: Command not recognised
SI-132 v1.1 32 SME300000081
EXCELL PRECISION CO., LTD
2 Output data format
Weight format
Gross S T , G S , + 0 1 2 3 4 5 6 7 SP SP o z
Net S T , N T , + 1 . 2 3 . 4 5 6 t l . g
Tare S T , T R , + 0 1 2 . 3 4 5 6 SP SP k g
CR LF
Plus OL O L , G S , + SP SP SP SP SP SP SP SP SP SP SP SP
Minus OL O L , G S , - SP SP SP SP SP SP SP SP SP SP SP SP
Unstable U S , G S , + 0 1 2 3 4 . 5 6 SP SP l b
Simple format
G/N + 1 . 2 3 . 4 5 6
G/N + 0 1 2 3 4 5 . 6
G/N + 0 1 2 . 3 4 5 6 CR LF
Plus OL + SP SP SP SP SP SP SP SP
Minus OL - SP SP SP SP SP SP SP SP
Comparison status + Simple format
Byte0 Byte1 Byte2 +/- 1 . 2 3 . 4 5 6 CR LF
Byte0 : HI 30H/31H
Byte1 : OK 30H/31H
Byte2 : LO 30H/31H
2 Serial Data Transfer/Receive Format
LSB MSB 1
n,8,1 S 8-bit data STOP
0
1
e,7,1 S 7-bit data P STOP
0
1
o,7,1 S 7-bit data P STOP
0
Note:
S : Start bit
STOP: Stop bit
P : Parity bit
SI-132 v1.1 33 SME300000081
EXCELL PRECISION CO., LTD
APPENDIX 1 7 SEGMENT DISPLAY
CHARACTERS
Digit 7 segments letter Alphabet 7 segments letter Alphabet 7 segments letter
0 A N
1 B O
2 C P
3 D Q
4 E R
5 F S
6 G T
7 H U
8 I V
9 J W
K X
L Y
M Z
SI-132 v1.1 34 SME300000081
EXCELL PRECISION CO., LTD
APPENDIX 2 ASCII CODE TABLE
ASC II ASC II ASC II
Symbol Symbol Symbol
Code Code Code
A 41H a 61H 0 30H
B 42H b 62H 1 31H
C 43H c 63H 2 32H
D 44H d 64H 3 33H
E 45H e 65H 4 34H
F 46H f 66H 5 35H
G 47H g 67H 6 36H
H 48H h 68H 7 37H
I 49H i 69H 8 38H
J 4AH j 6AH 9 39H
K 4BH k 6BH ↵ 0DH
L 4CH l 6CH
M 4DH m 6DH
N 4EH n 6EH
O 4FH o 6FH
P 50H p 70H
Q 51H q 71H
R 52H r 72H
S 53H s 73H
T 54H t 74H
U 55H u 75H
V 56H v 76H
W 57H w 77H
X 58H x 78H
Y 59H y 79H
Z 5AH z 7AH
SI-132 v1.1 35 SME300000081