H3 Series

Counting Scale
User Manual

©EXCELL PRECISION CO., LTD. 2012. All rights reserved worldwide.

The information contained herein is the property of EXCELL PRECISION CO., LTD. and is
supplied without liability for errors or omissions. No part may be reproduced or used except as
authorized by contract or other written permission. The copyright and the foregoing restriction on
reproduction and use extend to all media in which the information may be embodied.
EXCELL PRECISION CO., LTD.

TABLE OF CONTENTS
Instruction For Use....................................................................................................................... 3
Preparations Before Using ........................................................................................................... 3
Chapter 1 Display and Keypad Descriptions ............................................................................. 4
1-1 DISPLAY DESCRIPTIONS......................................................................................................... 4
1-2 SYMBOL ICONS “3” .............................................................................................................. 4
1-3 KEYPAD DESCRIPTIONS ......................................................................................................... 5
1-4 ERROR MESSAGES ............................................................................................................... 7
1-5 SELF-TEST MODE ................................................................................................................. 8
1-5-1 Check Software Version Number ................................................ 9
1-5-2 Check key and LCD、BL test ................................................... 10
1-5-3 Read AD Value ................................................................... 11
1-5-4 Keypad Test Code .............................................................. 12
1-5-5 EEPROM、Switch Calibration Test ............................................. 13
1-5-6 Real Time Clock(RTC) read ............................................. 14
1-5-7 RS232 (TXD and RXD short circuit test) .................................. 15
1-5-8 Read the Software Number of Free Format PCB .......................... 16
1-5-9 Read IC AD Value of Temperature ........................................ 17
1-5-10 Read AD Value of Battery Cell Volt ........................................ 18
1-5-11 Leave Test Mode .............................................................. 19
Chapter 2 Operation ................................................................................................................ 20
2-1 POWER ON......................................................................................................................... 20
2-2 RELATIVELY INTERNAL VALUE DISPLAY .................................................................................. 20
2-3 ZERO FUNCTION ................................................................................................................. 20
2-4 SAMPLING FUNCTION .......................................................................................................... 21
2-4-1 Unknown Unit Weight of a Weighed Object............................................................... 21
2-4-2 Known Unit Weight of a Weighed Object................................................................... 21
2-4-3 Under Tare Status ...................................................................................................... 22
2-5 TARE FUNCTION OPERATION ................................................................................................ 23
2-6 PRE-TARE FUNCTION OPERATION ........................................................................................ 24
2-7 CLEAR OFF PRE-TARE VALUE .............................................................................................. 25
2-8 ACCUMULATION................................................................................................................... 26
2-8-1 Quantity Accumulation............................................................................................... 26
2-8-2 Weight Accumulation ................................................................................................. 27
2-9 QUANTITY PRESET .............................................................................................................. 28
2-9-1 Upper Limit of Preset Quantity................................................................................... 28
2-9-2 Clear Upper Limit of Preset Quantity ......................................................................... 28

03004034 1 ZSME200000051
EXCELL PRECISION CO., LTD.
2-10 WEIGHT PRESET ............................................................................................................... 29
2-10-1 Upper Limit of Preset Weight................................................................................... 29
2-10-2 Clear Upper Limit of Preset Weight ......................................................................... 29
2-11 ID INPUT........................................................................................................................... 30
2-12 ITEM INPUT ..................................................................................................................... 30
2-13 UNIT W EIGHT PRESET ....................................................................................................... 31
2-13-1 Pre-set Unit Weight Saving Operation (read-in) ...................................................... 31
2-13-2 Pre-set Unit Weight Operation (read-out) ................................................................ 31
Chapter 3 External Calibration Setting Mode .......................................................................... 32
3-1 EXTERNAL FUNCTION SETTING .............................................................. 33
3-1-1 Backlight Mode Setting ............................................................ 34
3-1-2 Automatic Power-off Setting .................................................... 35
3-1-3 Stable Range Setting for Quantity Sampling............................ 36
3-1-4 Automatic Average Unit Weight Setting ................................... 37
3-1-5 A/D Sampling Speed Setting.................................................... 38
3-1-6 Zero Display Range Setting ..................................................... 39
3-1-7 Zero Tracking Range Setting ................................................... 40
3-1-8 Accumulation Ending Mode Setting ......................................... 41
3-1-9 Pre-Tare Mode Setting ............................................................. 42
3-1-10 Beeper Sound Output Condition Setting under Quantity Mode43
3-1-11 Accumulation Acceptable Condition Setting 1.......................... 44
3-1-12 Accumulation Acceptable Condition Setting 2 ........................... 45
3-1-13 Combination Key Setting ........................................................... 46
3-2 EXTERNAL WEIGHT & GRAVITY CALIBRATION ............................................... 47
3-2-1 External Weight Calibration .............................................................. 48
3-2-2 Gravity Adjustment ........................................................................... 49
3-3 RS-232 AND SERIAL PRINTER SETTING ................................................. 51
3-3-1 Baud Rate Setting................................................................... 52
3-3-2 Communication Protocol Setting............................................. 53
3-3-3 Output Data Format Setting .................................................... 54
3-3-5 Operation Mode Setting .......................................................... 58
3-3-6 Continuous Transmission Output Condition Setting................ 61
3-3-7 Zero Reset Condition for Automatic Transmission Setting ... 62
3-3-8 Reset Condition Setting for Automatic Transmission .............. 63
Appendix 1 RS-232 Full Duplex Format .................................................................................. 64
Appendix 2 Fixed Format RS-232 Transmission Line Illustration ............................................ 67
Appendix 3 ASCII Code Table ................................................................................................. 68
Appendix 4 7-Segment Display Characters............................................................................. 69

03004034 2 ZSME200000051
EXCELL PRECISION CO., LTD.

Thank you for purchasing EXCELL counting scale. In


order to operate smoothly, to last the durability and
to reduce chance of breakdown for this product,
please read this manual carefully.

Instruction For Use


1. The scale should not be drenched by rain or water. (If it gets wet carelessly, please
wipe it dry with a cloth. If its operation is abnormal, please send it to our distributor for
service. )
2. Please keep the scale in a cool and dry place. Do not store at high temperature or
damp places.
3. Please keep the scale clean and free from insect infestation.
4. To avoid impact and it is to be used under designated pressure (the weight put on the
platter can not exceed the maximum capacity of the scale)
5. If the scale is not going to be used for some time, please clean in and store it in a
plastic bag with desiccative. The internal rechargeable battery should be recharged
every three months. (If using dry batteries, take the dry batteries out before storing)
6. The commodity should be placed in the center of platter for accurate weighing. The
dimension of the weighted commodity should not exceed the dimension of platter.
7. Any suggestion for the product is warmly welcome.
8. In order to maintain the re-chargeable battery in good condition it should be kept fully
charged whenever possible. If the scale is to be stored, the battery should be fully
charged before storage, and then re-charged at 3 month intervals. The number of
times that the battery can be recharged will vary according to the conditions of use.
However it can be maximized by re-charging the battery frequently and by avoiding
conditions of total discharge. The battery cannot be overcharged.

Preparations Before Using


1. Put the scale on a firm and flat surface for accurate weighing reading. Adjust the four
leveling feet to get the leveling bubble at the center of the circle.
2. Scale must be used under a stable temperature and stable air flow. Avoid direct sunlight
onto the scale or near the air-conditioning outlet.
3. Scale must be used under individual socket to avoid the interference of other electric
appliances.
4. Remove any weight that might be on the platter before the scale is switched on.
5. The scale requires 15 ~ 20 minutes warm up before operation to ensure best accuracy.
6. Please note when symbol appears on the screen, the internal battery needs to be
replaced.

03004034 3 ZSME200000051
EXCELL PRECISION CO., LTD.

Chapter 1 Display and Keypad Descriptions


1-1 Display Descriptions
WEIGHT UNIT WEIGHT QUANTITY
→0←
← .
. .
. . .
. . . .
+
T
. . . . . . P cs
. . . . . . stable . . . . . .

Display Column
1. WEIGHT
Total 6 digits. To display the weight on platter or the total accumulation weight; the left digit is
able to display the negative symbol.
2. UNIT WEIGHT
Total 6 digits. To display the unit weight of objects on the platter or total accumulation counts.
3. QUANTITY
Total 6 digits. To display the quantity of the objects on platter or the accumulation.

3”
1-2 Symbol Icons “3
1. T : “Tare” indication
2. →0← : “Zero” indication
3. + : “Accumulation” indication
4. Stable : “Stable” indication
.

5. . .
. . .: If the weight of the sample is less than the weight of the minmum sample, this
. . . .

symbol will display. Please add the number of sample then sampling.
Minimum unit weight is 0.1d with resolution equal to 1/3000.
Minimum unit weight is 0.2d with resolution range from 1/6000 to 1/30000.
.

If . .
symbol displays, scale may cause some counting error even if the
. . .
. . . .

scale can still be used.


6. Pcs
: If the unit weight of the object on the platter is less than “Minimum Unit Weight”,
this symbol will display. The scale can still count the quantity even though the unit
weight is too small; however, this may affect the countting inaccuracy. Please use
the scale which the division and specification are both subject to.
If symbol displays, scale may cause some counting error even if the scale can still
Pcs
be used.
Normal models
Minimum unit weight is 0.1d with resolution equal to 1/3000. (d=division)
Minimum unit weight is 0.2d with resolution range from 1/6000 to 1/30000.
7. : When the symbol displays, the battery need to be recharged for safe use.

03004034 4 ZSME200000051
EXCELL PRECISION CO., LTD.

1-3 Keypad Descriptions


Double weighing units keypad _ double units conversion

7 8 9 Q’TY
SAMPL
ABC DEF PST
4 5 6 UNIT
kg/lb
GHI JKL MNO W.T
1 2 3
Z M+
PQRS TUV WXYZ

0 ‧ CE T MC

10 sets of preset unit weight keypad _ preset unit weight function

7 8 9 Q’TY
SAMPL
ABC DEF PST
4 5 6 UNIT U.W.
GHI JKL MNO W.T PST
1 2 3
Z M+
PQRS TUV WXYZ

0 ‧ CE T MC

Standard keypad _ preset clear function

7 8 9 Q’TY
SAMPL
ABC DEF PST
4 5 6 UNIT PST
GHI JKL MNO W.T CE
1 2 3
Z M+
PQRS TUV WXYZ

0 ‧ CE T MC

03004034 5 ZSME200000051
EXCELL PRECISION CO., LTD.

Number key for setting the unit weight, quantity and etc.
, ~
Clear key to clear the digits on the display.

Sampling key to set the sample count.

Unit weight key for setting the unit weight of sample.

Press this key, the display will return to 0.

Pre-setting the upper limit of count. If the calculated count is larger


than the limit, the scale will send a warning sound. Or use this key
to confirm the setting.
Totalizing the quantity or weight.

Clearing the stored totalizing memory. Or shift the digit pointer to


right.
Press this key to deduct the weight of container. Or shift the digit
pointer to left.
Press this key to clear preset

Press this key to preset the unit weight of sample. Or use this key
to enter the setting mode.

Press this key switch the unit : kg or lb

03004034 6 ZSME200000051
EXCELL PRECISION CO., LTD.

1-4 Error Messages

⇒ Zero value is too high (OIML or NTEP > 10% full scale)

⇒ Zero value is low (OIML or NTEP < -10% full scale)

⇒ Internal value is more than 700000(use in factory calibration)

⇒ Internal value is lower than 100000(use in factory calibration)

⇒ The weight value is over 9d of the maximum capacity. (d=division)

⇒ Internal value is unstable

Unstable time is over 10 seconds after pressing ZERO or


TARE key

03004034 7 ZSME200000051
EXCELL PRECISION CO., LTD.

1-5 Self-Test Mode


Hold 0Z key to open the scale until the display show . Wait till
display shows to enter “Self-Test Mode”.

Press 0T0
⇒ check Software version number

⇒ check key and LCD、BL test


Press 0T0
⇒ read AD value

⇒ key test
Press 0T0
⇒ EEPROM、calibrate switch test

⇒ Real Time Clock(RTC)


Press 0T0 ⇒ RS232 (TXD and RXD short
circuit test)

Press 0T0 ⇒ read free format PCB Software


version number

Press 0T0 ⇒ read temperature IC AD value

⇒ read AD value

Press 0T0 ⇒ ESC (leave test mode)

Press 0T0

Press 0T0 key ⇒ ESC


CE key ⇒ Move cursor leftward
T key ⇒ Move cursor rightward
MC key ⇒ ENTER
Press 0T0

03004034 8 ZSME200000051
EXCELL PRECISION CO., LTD.

1-5-1 Check Software Version Number

Press MC

Press MC
03004010 is software version number

Press MC Show the maintenance number 034 for 1 second then return to 010

key ⇒ ESC
CE key ⇒ Move cursor leftward
T key ⇒ Move cursor rightward
MC key ⇒ ENTER

03004034 9 ZSME200000051
EXCELL PRECISION CO., LTD.

1-5-2 、BL test


Check key and LCD、

press MC key

press key The number of LCD is count backward as 9~0 pattern The symbol and
backlight of LCD is on when it is odd number,conversly it is off.

key ⇒ ESC
CE key ⇒ Move cursor leftward
T key ⇒ Move cursor rightward
MC key ⇒ ENTER

03004034 10 ZSME200000051
EXCELL PRECISION CO., LTD.

1-5-3 Read AD Value

press MC key

X XXXXXX
press key Show the internal value

key ⇒ ESC
CE key ⇒ Move cursor leftward
T key ⇒ Move cursor rightward
MC key ⇒ ENTER

check whether the internal value is within its normal range (range of carrying idler :100,000 ~
700,000 )

⇒ Internal value above 700000

⇒ Internal value under 100000

03004034 11 ZSME200000051
EXCELL PRECISION CO., LTD.

1-5-4 Keypad Test Code

press MC key

XX
Keypad test code
press key

key ⇒ ESC
CE key ⇒ Move cursor leftward
T key ⇒ Move cursor rightward
MC key ⇒ ENTER

Keypad ⇒ Keypad internal test code


8 9 Q’TY
7 SMPL 00 10 20 30 40
ABC DEF PST
4 5 6 UNIT PST
03 13 23 33 43
GHI GKL MNO W.T. CE ⇒
1 2 3
Z M+ 01 11 21 31 41
PQRS TUV WXYZ
0 . CE T MC 02 ESC 22 32 42

8 9 Q’TY
7 SMPL 00 10 20 30 40
ABC DEF PST
4 5 6 UNIT U.W
03 13 23 33 43
GHI GKL MNO W.T. PST ⇒
1 2 3
Z M+ 01 11 21 31 41
PQRS TUV WXYZ
0 . CE T MC 02 ESC 22 32 42

03004034 12 ZSME200000051
EXCELL PRECISION CO., LTD.

8 9 Q’TY
7 SMPL 00 10 20 30 40
ABC DEF PST
4 5 6 UNIT
kg/lb 03 13 23 33 43
GHI GKL MNO W.T. ⇒
1 2 3
Z M+ 01 11 21 31 41
PQRS TUV WXYZ
0 . CE T MC 02 ESC 22 32 42

、Switch Calibration Test


1-5-5 EEPROM、

press MC key

press key

key ⇒ ESC
CE key ⇒ Move cursor leftward
T key ⇒ Move cursor rightward
MC key ⇒ ENTER

EEPORM switch test

EEPROM reading and writing test


SWA1switch test
EEPROM reading and writing is successful
: LOCK
EEPROM reading and writing fail (eeprom is
: ADJ
uninstalled or damaged)

03004034 13 ZSME200000051
EXCELL PRECISION CO., LTD.
1-5-6 Real Time Clock(RTC) read

press MC key

press key 2005/08/18 [Link]

key ⇒ ESC
CE key ⇒ Move cursor leftward
T key ⇒ Move cursor rightward
MC key ⇒ ENTER

12 code numbers standard for: year, month, day, hour, minute, second

If the display shows 12 zeros and not flicker, it shows that free format PCB is not been
connected to the main board or free format PCB RTC has not been installed.
To use number key to set the date and time of RTC, then press . to finish setting.

03004034 14 ZSME200000051
EXCELL PRECISION CO., LTD.

1-5-7 RS232 (TXD and RXD short circuit test)

Press MC key

Press key PASS RS232 the operation of internal board is correct


FAIL RS232 the operation of internal board fail or is not circuit

key ⇒ ESC
CE key ⇒ Move cursor leftward
T key ⇒ Move cursor rightward
MC key ⇒ ENTER

The test can only test whether the transmition and reception operates regularly in internal
software.

03004034 15 ZSME200000051
EXCELL PRECISION CO., LTD.

1-5-8 Read the Software Number of Free Format PCB

press MC key

XXXXX XXX
press key software number of free format PCB

key ⇒ ESC
CE key ⇒ Move cursor leftward
T key ⇒ Move cursor rightward
MC key ⇒ ENTER

The main board transmit “T”+0DH+0AH (ASCII) 3 byte to free format card to read the version
number of free format PCB shows on LCD
If free format PCB is not been connected or there is something with free format PCB, it will
show

03004034 16 ZSME200000051
EXCELL PRECISION CO., LTD.

1-5-9 Read IC AD Value of Temperature

press MC key

press key Show IC AD value of temprature


If shows “NOTMP” ⇒ no tempreture IC

key ⇒ ESC
CE key ⇒ Move cursor leftward
T key ⇒ Move cursor rightward
MC key ⇒ ENTER

If there is no IC temperature, the display of unit weight shows

03004034 17 ZSME200000051
EXCELL PRECISION CO., LTD.

1-5-10 Read AD Value of Battery Cell Volt

press MC key

press key
AD value of volt

key ⇒ ESC
CE key ⇒ Move cursor leftward
T key ⇒ Move cursor rightward
MC key ⇒ ENTER

If AD value of volt is less than 398, it will show (about 5.8V)

03004034 18 ZSME200000051
EXCELL PRECISION CO., LTD.

1-5-11 Leave Test Mode

Press MC key
Restart the scale

key ⇒ ESC
CE key ⇒ Move cursor leftward
T key ⇒ Move cursor rightward
MC key ⇒ ENTER

03004034 19 ZSME200000051
EXCELL PRECISION CO., LTD.

Chapter 2 Operation
2-1 Power On
”)
SW Switch On (Press SW to location “

Weight Unit Weight Quantity

After stable, the scale will count back to zero.

Weight Unit Weight Quantity

Sampling Speed
600 ⇒ Maximum countable data: 600 000

After return to zero and being stable, scale will enter into counting mode.

2-2 Relatively Internal Value Display


After press ZERO key, screen will show “ - - - - - - ”. Press CE key.

XXXXX X XXXXXX
Max internal calibration SP value is 5 digits Max internal value is 7 digits.

Press 0 key to return to the counting mode.

2-3 Zero Function


While operating the scale, zero may sometimes fluctuate. (Slight weight changes happen in
weight column.) Press ZERO key to return to zero.

03004034 20 ZSME200000051
EXCELL PRECISION CO., LTD.

2-4 Sampling Function


2-4-1 Unknown Unit Weight of a Weighed Object
1. Place the sampling object on platter:

.
Weight on platter

2. Enter the quantity of the sample on platter:

.
Weight on platter Enter the sampling quantity

NOTE:
The number of quantity column will flash 6 seconds. If the user doesn’t press the
SAMPLE key before flashing is over, the scale will complete the unit weight setting
procedure automatically after flashing. The scale will also take the number that inputted in
unit weight column as the unit weight of object to calculate the quantity of object shown in
quantity column.

3. Press SAMPLE key while total column number is flashing:

.
Weight on platter

4. After stable, the scale finishes sampling and enters into counting mode.

. . Stable
Weight on platter Unit weight of object Enter sampling quantity

2-4-2 Known Unit Weight of a Weighed Object


1. Enter known unit weight of object intended to weigh:
→0←

. . Stable
Unit weight of object intended to weigh

2. Press UNIT WEIGHT key to complete unit weight setting and enter into counting mode:
→0←

. . Stable
Unit weight of object intended to weigh

03004034 21 ZSME200000051
EXCELL PRECISION CO., LTD.

2-4-3 Under Tare Status


1. Take the sample off the platter.

.
The weight of the object on the platter

2. Input the quantity of the sample on the platter.

.
The weight of the object on the platter Input the quantity of sample

NOTE:
The number of quantity column will flash 6 seconds. If the user doesn’t press the
SAMPLE key before flashing is over, the scale will complete the unit weight setting
procedure automatically after flashing. The scale will also take the number that inputted in
unit weight column as the unit weight of object to calculate the quantity of object shown in
quantity column.

3. Press the SAMPLE key when the number of quantity column is flashing.

.
The weight of the object on the platter

4. After being stable, the scale will finish the sampling and enter into counting mode.

. . Stable
The weight of the object on the platter The unit weight of object The quantity of sample that inputted

NOTES:

The larger quantity of sampling, the more precise unit weight counted out.
When unit weight column and total quantity column both indicate 0, please press
SAMPLE key, and the previous unit weight value will come out.

Use ZERO and 7 to open or shut display.

03004034 22 ZSME200000051
EXCELL PRECISION CO., LTD.

2-5 Tare Function Operation


1. Place the packaging container on platter:

.
Weight of packaging container

2. Press TARE key

3. The scale will enter into counting mode after stable:


→0←

T
. Stable

Clear off the Tare Value


Mode 1
After removing the object together with packaging container, weight column will display the
negative weight value of the packaging container. Press TARE key again to cancel the
tare, and return to zero. The tare symbol “3” will disappear.
Mode 2
After removing the object together with packaging container, weight column will display the
negative weight value of the packaging container. Press ZERO key again to cancel the
tare, and return to zero. The tare symbol “3” will disappear.

03004034 23 ZSME200000051
EXCELL PRECISION CO., LTD.

2-6 Pre-Tare Function Operation


(Pre-Tare setting) is set up as . (no weight on platter)
1. No object on platter
→0←

. Stable

2. Press TARE key


→0←

. Stable

3. Enter the known weight of packaging container:


→0←

. Stable
Weight of packaging container entered

4. Press TARE key


PT
T . Stable

(Pre-Tare setting) is set up as . ( weight on platter )


Operation for non-approval model.
1. Object placed on platter:

. . Stable
Weight of the object on platter Unit weight of the object Quantity of object

2. Enter the known weight of packaging container:

. . Stable
Entered weight of packaging container

3. Press TARE key

PT

T
. . Stable
Weight of the object without container Unit weight of the object Quantity of the object without packaging container

NOTES:
Pre-Tare function is also available if the object has been tarred.
If the net weight on platter is more than zero weight, the tare function is available.
Otherwise, it’s not capable of tare function.

03004034 24 ZSME200000051
EXCELL PRECISION CO., LTD.

2-7 Clear Off Pre-Tare Value


After removing the object together with packaging container, weight column will display the
negative weight value of the packaging container. Press TARE key once again to cancel
the Pre-Tare value and return to zero. Then the tare and Pre-Tare symbol “3” will
disappear.

Pre-Tare Value Recall


1. Press QTY PRESET key, and then press TARE key:

PT
.
T Stable

Pre-Tare value

2. After screen displays Pre-Tare value about 5 seconds, it will automatically return to the
weighing mode.
PT
T . . Stable

03004034 25 ZSME200000051
EXCELL PRECISION CO., LTD.

2-8 Accumulation
The accumulation counts are up to 99 counts, but the total quantity column is at most 6
digits.
You can not do positive accumulation and negative accumulation at the same time.

2-8-1 Quantity Accumulation


1. Place an object on platter:

. . Stable
Weight of the object on platter Unit weight of the object Quantity of the object on platter

2. Press M+ key

3. After scale is stable:


+
. Stable
Accumulated total weight Accumulated total counts Accumulated total quantity

4. After about 3 seconds, scale returns to counting mode:


+
. . Stable
Weight of the object on platter Unit weight of the object Quantity of the object on platter

Quantity Accumulation Recall


While weight column displays 0, press M+ key to recall the accumulated data:
→0←
← +
. Stable
Accumulated total weight Accumulated total counts Accumulated total quantity

Clear off the Accumulated Quantity Value


Press MC key to clear off the accumulated value in the memory, and accumulation
symbol “3” will disappear.

03004034 26 ZSME200000051
EXCELL PRECISION CO., LTD.
2-8-2 Weight Accumulation
1. Place an object on platter when unit weight display is 0:

. Stable
Wight of the object on platter

2. Press M+ key:

3. After scale is stable:


+
. Stable
Accumulated total weight Accumulated total counts

4. After 3 seconds, scale returns to weighing mode.


+
. Stable
Weight of the object on platter

Weight Accumulation Value Recall


Press M+ key to recall the accumulated weight value when weight display is 0:
→0←
← +
. Stable
Accumulated total weight Accumulated total counts

Clear off Accumulated Weight Value


Press MC key to clear off the accumulated weight value in the memory, and then the
accumulation symbol “3” will disappear.

03004034 27 ZSME200000051
EXCELL PRECISION CO., LTD.

2-9 Quantity Preset


It’s available to pre-set the upper limit of quantity under counting mode. If the counts are
over the limit, the beeper makes warning sounds, and the weight column displays flashing

2-9-1 Upper Limit of Preset Quantity


1. Whether there is an object on platter or not, press QTY PRESET key.
→0←

.
. Stable

2. Press SAMPLE key, choose “Quantity Preset” mode (press UNIT WEIGHT key to
choose “Weight Preset” mode)
→0←

Stable
Previous Preset

3. Enter the upper limit intended: (press CE key to modify the value entered)
→0←

Stable
Upper limit entered

4. Press SAMPLE key: (press CE key to modify the value entered)


→0←

Stable

5. Press QTY PRESET key, and the scale will be back to the counting mode:
→0←

. Stable

2-9-2 Clear Upper Limit of Preset Quantity


To clear the pre-set upper limit of quantity or weight, please follow the above-mentioned
operation steps. When entering the pre-set value, please input “0” instead.

When switching to "Weight Preset "mode or "Quantity Preset" mode, previous Preset value will
be deleted automatically.

03004034 28 ZSME200000051
EXCELL PRECISION CO., LTD.

2-10 Weight Preset


It’s available to pre-set the upper limit of weight. If the weights are over the limit, the
beeper makes warning sounds, and the weight column displays flashing

2-10-1 Upper Limit of Preset Weight


1. Whether there is an object on platter or not, press QTY PRESET key.
→0←

Stable

2. Press UNIT WEIGHT key to choose “Weight Preset” mode. (Press SAMPLE
key, choose “Quantity Preset” mode)
→0←

. Stable
.
Previous setting

3. Enter the upper limit intended: (press CE key to modify the data entered)
→0←

. Stable
.
Upper limit entered

4. Press UNIT WEIGHT key: (press CE key to modify the data entered)
→0←

. Stable
.

5. Press QTY PRESET key, and the scale will be back to counting mode.
→0←

. Stable

2-10-2 Clear Upper Limit of Preset Weight


To clear the pre-set upper limit of quantity or weight, please follow the above-mentioned
operation steps. When entering the pre-set value, please input “0” in stead.

When switching to "Weight Preset "mode or "Quantity Preset" mode, previous Preset value
will be deleted automatically.

03004034 29 ZSME200000051
EXCELL PRECISION CO., LTD.

2-11 ID Input
Press ZERO key, and the screen displays “ - - - - - - ”. Press 0 key before it disappears.

ID could be set up to 12 digits. They can be numbers ( 0 ~ 9 ), English letters


Input ID with number keys. ( A ~ Z ), or space ( )

Press MC key to Press . key to quit setting.


confirm the data entered.
→0←

.
Stable

2-12 ITEM Input


Press ZERO key, and the screen displays “ - - - - - - ”. Press 2 key before it disappears.

ITEM could be set as 12 digits at most. They can be set as numbers ( 0 ~ 9 ), English letters
Input ITEM with number keys. ( A ~ Z ), or space ( ).

Press MC to confirm. Press . key to quit setting.

→0←

.
Stable

Input of ID & ITEM is applied in PRINTER printing output. (FIX FORMAT or FREE FORMAT)
ID & ITEM could be set as 12 digits at most. They can be set as numbers ( 0 ~ 9 ), English letters ( A ~ Z ), or space ( ).
Numbers/English letters inputted operation: Press number key and the digit will be flashing. Press the same key , and the
display will show the number/English letter in cycle. When the entered number/English flashed for 2 seconds, the setting
will be confirmed and moved to the right place by 1 digit. Example: Press 1 key continuously, and the screen will display

1,P,Q,R,S flashing in cycle.


If the entered data for ID & ITEM is not saved in the location of unit weight pre-set, the data will be cleared after power off.

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

03004034 30 ZSME200000051
EXCELL PRECISION CO., LTD.

2-13 Unit Weight Preset


The preset data could be save in up to 10 addresses
Each address contains unit weight, pre-tare, ID and ITEM.

2-13-1 Pre-set Unit Weight Saving Operation (read-in)


Use number key to input the unit weight. (The value is 0 or blank if not being set.)

Press [Link] key:

Press [Link] key again:

Press number key 0 ~ 9 to select one for saving the data.

2-13-2 Pre-set Unit Weight Operation (read-out)


Press [Link] key:

Press number key 0 ~ 9 to select one for call out the data.

NOTE: While reading or saving, if the waiting time is over 10 seconds, the scale will be back
to weighing mode automatically. Press CE key to cancel the read-in and read-out.

03004034 31 ZSME200000051
EXCELL PRECISION CO., LTD.

Chapter 3 External Calibration Setting Mode


After starting machine, and the scale returns to zero, press ZERO key and the screen will
display “ - - - - - - ”. Then, press . key to enter external calibration function setting mode.
The weight column will display .

Press TARE

Press TARE

Press TARE

Press TARE

⇒ External Function Setting

⇒ External Weight Calibration and Gravity Calibration

⇒ RS-232 and Serial Printer Setting

⇒ Exit the setting

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

03004034 32 ZSME200000051
EXCELL PRECISION CO., LTD.

3-1 External Function Setting

Press MC

Input parameter 00 ~ 13

After entering parameter “00”,


press MC to return to the
previous level (leave external
function setting)

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

⇒ Return to previous level


⇒ Backlight mode setting
⇒ Automatic power off setting
⇒ Stable range setting for quantity sampling
⇒ Automatic unit weight average setting
⇒ A/D sampling speed setting
⇒ “Zero” display range setting
⇒ “Zero” track range setting
⇒ Accumulation ending mode setting
⇒ Pre-Tare mode setting.
⇒ ”Beeper” output setting for quantity limit.
⇒ Accumulation acceptable condition setting 1.
⇒ Accumulation acceptable condition setting 2.
⇒ Combination key setting

03004034 33 ZSME200000051
EXCELL PRECISION CO., LTD.

3-1-1 Backlight Mode Setting

Press MC

Enter parameter 00 ~ 02 Previous setting


according to demand.

Press MC to save the


setting data.

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default Setting: (No backlight)


⇒ Backlight is always on.
⇒ While weighing (weight is higher than 10d) or pressing any key, backlight be turned
on automatically. The backlight is turned off automatically when the scale is idle for
10 minutes. (d=division)
⇒ No backlight.
When turning on, the backlight mode is the same as previous setting.

03004034 34 ZSME200000051
EXCELL PRECISION CO., LTD.

3-1-2 Automatic Power-off Setting

Press MC

Enter parameter 00 ~ 10 Previous setting


according to demand.

Press MC to save
setting data

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting:

⇒ the auto power-off function is disable.

~ ⇒ the scale to be automatically power off after the scale is not in use for 1
to 10 minutes.
(If intended to operate continuously, please restart the scale again.)
It can be set up to 10 minutes at most.

03004034 35 ZSME200000051
EXCELL PRECISION CO., LTD.

3-1-3 Stable Range Setting for Quantity Sampling

Press MC

Enter parameter 00 ~ 15 Previous setting


according to demand

Press MC key to save


the setting data

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting:
While quantity sampling, the scale will indicate a stable reading and error is within ±8d
internal value. (d=division)
Parameters ~
The higher value makes the sampling faster, but less accurate.
The smaller value makes the sampling slower, but more accurate.
It can’t be shifted after sampling, only when the unit weight is cleared in Brazil version.

03004034 36 ZSME200000051
EXCELL PRECISION CO., LTD.

3-1-4 Automatic Average Unit Weight Setting

Press MC

Enter parameter 00, 01 Previous setting


according to demand

press MC to save
setting data

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting:

⇒ No automatic average unit weight function (press SAMPLE key for manual
unit weight calibration)

⇒ Automatic average unit weight function

Condition: Automatic unit weight calibration will perform when the measured sampling
number increase greater than 10% but less than 100% of previous sampling number.

03004034 37 ZSME200000051
EXCELL PRECISION CO., LTD.

3-1-5 A/D Sampling Speed Setting

Press MC

Enter parameter 00 ~ 01 Previous Setting


according to demand

Press MC key to save


the setting data

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting:

⇒ Low speed is about 7.5 Hz. (Weighing reflection is slow but relatively stable.)

⇒ Fast speed is about 15 Hz. (Weighing reflection is fast but relatively unstable.)

03004034 38 ZSME200000051
EXCELL PRECISION CO., LTD.

3-1-6 Zero Display Range Setting

Press MC

Enter parameter 00 ~ 03 Previous setting


according to demand

Press MC to save the


setting data

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

The default setting is .

⇒ Display all

⇒ Zero range ±1 bit will not display division, and displays zero instead.

⇒ Zero range ±2 bits will not display division, and displays zero instead.

⇒ Zero range ±3 bits will not display division, and displays zero instead.

If is set, when setting up Pre-Tare data, the Pre-Tare data can not be less or equal
to ±3 bits external value, and so on.
When the weight is over 1/3 full capacity and return to 0, this function will be started.

03004034 39 ZSME200000051
EXCELL PRECISION CO., LTD.

3-1-7 Zero Tracking Range Setting

Press MC

Enter parameter 00 ~ 03
Previous setting
according to demand

Press MC key to save


setting data

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

The default setting is .

⇒ After weight keeps stable continuously for over 1 second, it could track ±1/4d
(external value is 1/4 bit).

⇒ After weight keeps stable continuously for over 1 second, it could track ±1/2d
(external value is 1/2 bit).

⇒ After weight keeps stable continuously for over 1 second, it could track ±1d
(external value is 1 bit).

⇒ After weight is being stable continuously for over 1 second, it could track ±2d
(external value is 2 bits). (d=division)

Only if GROSS = 0, zero tracking function will be started.

03004034 40 ZSME200000051
EXCELL PRECISION CO., LTD.

3-1-8 Accumulation Ending Mode Setting

Press MC

Enter parameter 00 ~ 02
Previous setting
according to demand.

Press MC key to save


setting data

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting:

⇒ Press M+ . After screen displays accumulation data for 3 seconds, the scale
will be back to weighing mode directly.

⇒ Press M+ . And the screen displays accumulation data. The scale will not be
back to weighing mode until pressing CE .

⇒ Press M+ . And the screen does not display accumulation data, but beeper
will “beep” once.

03004034 41 ZSME200000051
EXCELL PRECISION CO., LTD.

3-1-9 Pre-Tare Mode Setting

Press MC

Enter parameter 00, 01


Previous setting
according to demand.

Press MC key to save


the setting data.

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting:

⇒ Traditionally, Pre-Tare can not proceed when there is weight on platter.


When there is no weight on platter, press TARE key to enter Pre-Tare value,
and then press TARE key again to complete.

⇒ Pre-Tare can be proceeded when there is weight on platter (enter digit in the unit
weight column, and then press tare key).
When there is weight on scale, enter Pre-Tare data in the unit weight column.
Example: enter “1”, “1.0”, “1.00” or “1.000” and then press
TARE key again to complete the Pre-Tare “1kg.”
Pre-Tare value can not be more than maximum weighing value, or the first and second
segment point of division; besides, the value can not be less than or equal to the external
value set up in .

03004034 42 ZSME200000051
EXCELL PRECISION CO., LTD.

3-1-10 Beeper Sound Output Condition Setting

under Quantity Mode

Press MC

Enter parameter 00, 01


Previous setting
according to demand.

Press MC key to save


the setting data.

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

The default setting is .

⇒ If the weighing quantity exceeds quantity setting (or weight exceeds weight
setting), the beeper will make sounds “bee” under unstable situation.

⇒ If the weighing quantity exceeds quantity setting (or weight exceeds weight
setting), it’s no need to be stable and the beeper will make sounds “bee”
automatically.

03004034 43 ZSME200000051
EXCELL PRECISION CO., LTD.

3-1-11 Accumulation Acceptable Condition Setting 1

Press MC

Enter parameter 00, 01


according to demand. Previous setting

Press MC to save the


setting data.

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting:

⇒ The scale will accept accumulation only when being stable.

⇒ The scale will accept accumulation no matter its being stable or not.

03004034 44 ZSME200000051
EXCELL PRECISION CO., LTD.

3-1-12 Accumulation Acceptable Condition Setting 2

Press MC

Enter parameter 00 ~ 02
Previous setting
according to demand

Press MC to save the


setting data

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting:

⇒ Only when the weight returns to zero, the scale accepts the next accumulation

data. If the weight is close to zero band, the scale can be adjusted by

⇒ It’s no need to return to zero and the scale can accept the next accumulation
data. That means when there is load, the weight can be accumulated
continuously.

⇒ The weight must return to zero (gross = 0), and the scale can accept the next
accumulation data.

⇒ Press M+ key not to accumulate the data, at the same time, RS232 would be
transmit the LCD data.

03004034 45 ZSME200000051
EXCELL PRECISION CO., LTD.

3-1-13 Combination Key Setting


Combination key represents kg/lb [Link] key; this key contains two functions:
Unit switching
10 sets for unit weight preset

Press MC

Enter parameter 00, 01 Previous setting


according to demand

Press MC key to save


the setting data

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting:

⇒ Press the combination key once to select unit (priority function)


Press and hold combination key for 3 seconds to pre-set unit weight (minority
function)
⇒ Press combination key once to pre-set unit weight (priority function)
Press and hold combination key for 3 seconds to select unit (minority function)

03004034 46 ZSME200000051
EXCELL PRECISION CO., LTD.

3-2 External Weight & Gravity Calibration

Press MC key

Enter parameter 00 ~ 03

After entering parameter


“00”, press MC key to
back to previous level

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

⇒ Back to previous level

⇒ External weight calibration

⇒ Gravity adjustment or weight calibration in


operation place

03004034 47 ZSME200000051
EXCELL PRECISION CO., LTD.

3-2-1 External Weight Calibration


Zero value can be calibrated separately from the weight calibration value.

XXX
If there is no temperature IC, the unit weight column will
Press MC display .

kg
. XXX XXXXXX

No weight display Internal value of no weight display

Make sure that no load on platter, Press PRESET CE to switch calibration unit.

and press MC to read “Zero” Press . if not intend to calibrate “0”.

. kg XXX XXXXXX
Calibration poises value (weight value of fully capacity) Internal value of full capacity display

Use number key 0 ~ 9 to Calibration poises value entered can not be higher than “maximum
weighing capacity”.
modify calibration poises value, Poises weight entered will be compared with internal calibration
place enough poises according to poises weight. If the actual weight is more than the entered poises
weight in 1.1~0.9 times, the weight column will display
value entered, and press MC for 1 second. Re-enter the calibration poises value,
or place the correct weight to operate once again.
key to read calibrated internal
If you do not calibrate weight, press . key to exit without saving
value and save into EEPROM. AD data into EEPROM.

kg
. XXX XXXXXX
Poises weight value on platter Internal value
Under weight calibration mode, please put on or
remove weight to test. After testing, press MC
key to complete external weight calibration

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

03004034 48 ZSME200000051
EXCELL PRECISION CO., LTD.

3-2-2 Gravity Adjustment

It’s capable for users to reset and modify gravity for 9 times. If it’s the 10th revised gravity
calibration, it needs to enter into to redo the “gravity calibration in
operation location,” and the calibration time will be re-set to “1”.

Gravity adjustment in operation location must be done “after” external weight calibration.

If the external weight calibration is done before gravity adjustment, the previous gravity
value will be set as the value of operation location of external weight calibration.

Gravity Value Modify Calculation


Gravity, the accelerated speed “G”, is the reaction of centrifugal force for free falling object.
Standard gravity for the equator on earth surface:
GE=978.03184558 cm/sec2 = 9.7803184558 m/sec²
Standard gravity for polar region:
Gp=983.21772792 cm/sec2 = 9.8321772792 m/sec²
The G value below ground is the object's free fall acceleration at that location.
Gravity does not decrease gradually from mantle to the earth's core. Gravity in the mantle
will reaches its highest value (about 1021 cm/ sec²), and decreases to 0 at the earthe's
core.

Standard gravitation acceleration in different latitude (ψ) can be calculated by formula


described as bellowed descriptions:
G= (1+0.005278895sin2ψ+0.000023462sin4ψ)
=978.03185( )
G: cm/sec2 ψ: Latitude

Gravity must between 9.78032 m/sec² to 9.83218 m/sec2.

03004034 49 ZSME200000051
EXCELL PRECISION CO., LTD.

Press MC to display
historical revised data

Display the current revised times, and the revised gravity in operation location.

Press . to quit and back to previous level


Press CE Press TARE
Press MC key to enter into revising mode

Display previous revised times and revised gravity in operation place.

Press . to quit and back to previous level


Press CE Press TARE
Press MC key to enter into revising mode

Times “01” ⇒ To fix the gravity in operation place when internal weight
calibration is stored.
Press . to quit and back to previous level
Press CE Press TARE
Press MC key to enter into revising mode

Times “00” ⇒ To fix the gravity in calibration place when the internal weight
Press MC to enter into calibration is stored.
G-value revising mode Press . to quit and back to previous level

Please enter revised gravity, Automatically add 1 to times

press MC key to complete. Press . to quit and back to previous level

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

03004034 50 ZSME200000051
EXCELL PRECISION CO., LTD.
3-3 RS-232 and Serial Printer Setting

Press MC

Enter parameter 00 ~ 09

After entering parameter “00”,


press MC and back to
previous level (quit RS-232
setting mode)

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

⇒ Back to previous level


⇒ Baud rate setting
⇒ Communication protocol setting
⇒ Output data format setting
⇒ Output counts setting per second in continuous transmission
⇒ Operation mode setting
⇒ Continuous transmission output condition setting
⇒ Zero reset condition for automatic transmission setting
⇒ Reset condition for automatic transmission setting

03004034 51 ZSME200000051
EXCELL PRECISION CO., LTD.

3-3-1 Baud Rate Setting

Press MC

Enter parameter 00 ~ 05 Previous setting


according to demand

Press MC key to save


the setting data

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting: ( 9 600 bits/second)

⇒ 600 bits/second
⇒ 1 200 bits/second
⇒ 2 400 bits/second
⇒ 4 800 bits/second
⇒ 9 600 bits/second
⇒ 19 200 bits/second
If there is free form device, it is 9600 bits/sec transmit.

03004034 52 ZSME200000051
EXCELL PRECISION CO., LTD.

3-3-2 Communication Protocol Setting

Press MC

Enter parameter 00 ~ 02
Previous setting
according to demand

Press MC to save
setting data

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting: ( N, 8, 1 )

⇒ N, 8, 1
⇒ E, 7, 1
⇒ O, 7, 1

If there is free form device, it is n81 transmit.

03004034 53 ZSME200000051
EXCELL PRECISION CO., LTD.

3-3-3 Output Data Format Setting

Enter parameter 00 ~ 07 Previous setting


according to demand

Press MC key to save


setting data

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting: ( fixed format 1 )

⇒ Fixed format 1 (details as the following page)


⇒ Fixed format 2 (details as the following page)

⇒ RESERVED
⇒ Same as screen display (general format)
⇒ Same as screen display (simple format)
⇒ Gross weight (general format)
⇒ Net weight (general format)
⇒ Tare (general format)
⇒ Fixed format 3 (details as the following page)
⇒ Fixed format 4 (details as the following page)

For output format illustration, refer to “Appendix 1”.

When scale connects with free format card, this setting will be fixed as ⇒
RESERVED

03004034 54 ZSME200000051
EXCELL PRECISION CO., LTD.
Example: Illustration for fixed format is described as below:

Fixed format 1 Fixed format 2


“Press M+ key to transmit” “Press M+ key to transmit”
ID: xxxxxxx xxxxx
ITEM: xxxxxxx xxxxx
NO. 3 NO. 3
G 2.480 kg G 2.480 kg
N 2.000 kg N 2.000 kg
T 0.080 kg T 0.080 kg
PT 0.400 kg PT 0.400 kg
U/W 1.6003 g U/W 1.6003 g
Q 1250 pcs Q 1250 pcs

Note: if the format (rs1 03 setting) belongs to M+MC Key, but the method of transmission
(rs1 05 setting) is continuous or automatic transmission, some content maybe useless.

Fixed format 3 “Press M+ key Fixed format 4 “Press M+ key


to transmit” to transmit”
NO. 1 N/W 0.500
N/W 0.500 U/W 1.00013
U/W 1.00013 PCS 500
PCS 500

Fixed format 1,2


“Press MC key to transmit”
(to print out total accumulation data, and clear data in memory)

================
T/N 3
T/W 1500 kg
T/Q 300 pcs

03004034 55 ZSME200000051
EXCELL PRECISION CO., LTD.
Fixed format 3
“Press MC key to transmit”
(to print out total accumulation data, and clear data in memory)

===========
T/N 3
T/W 1500
T/A 300

NO. ⇒ Number of Counts


Q ⇒ Quantity
T ⇒ Tare
PT ⇒ Pre-Tare
G ⇒ Gross Weight
N ⇒ Net weight
U/W ⇒ Unit weight
T/N ⇒ Total Number of Counts
T/W ⇒ Total weight
T/Q ⇒ Total quantity
ID: 12 digits (max.)
ITEM: 12 digits (max.)

03004034 56 ZSME200000051
EXCELL PRECISION CO., LTD.

3-3-4 Output Counts Setting Per second in Continuous

Transmission

Press MC

Enter parameter 00 ~ 04 Previous setting


according to demand

Press MC key to save


setting data.

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting: (1 count/second)

⇒ 1 count/second
⇒ 2 counts/second
⇒ 4 counts/second
⇒ 8 counts/second
⇒ More than 8 counts/second
If parameter of is set as (fixed format 1) or (fixed format 2),
it may not be available to reach the transmission counts due to too long distance.

If parameter of is set as (Reserved), it’s not capable of


continuous transmission mode.

If parameter of is set as ~ , it’s capable of continuous


transmission normally.

03004034 57 ZSME200000051
EXCELL PRECISION CO., LTD.

3-3-5 Operation Mode Setting

Press MC

Enter parameter 00 ~ 04
Previous setting
according to demand

Press MC key to save


the setting data

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting:

⇒ Command mode
⇒ Continuous transmission + command mode
⇒ Automatic transmission + command mode
⇒ Manual transmission + command mode
Please press M+ key or MC key to do manual transmission.
⇒ No RS-232 transmission
⇒ ZEBRA PRINTER special format

If parameter of is set as (Reserved), the setting of

will be fixed as manual transmission but without command mode.

For the illustration of command mode format, refer to “Appendix 1”.

If there is free form device which is detected, it will be ” ” => Manual transmission+
command mode.

03004034 58 ZSME200000051
EXCELL PRECISION CO., LTD.
Zebra printer output format
1. Press M+ key to transmit the data.

F R “ 5 2 0 P “ <LF>
? <LF>
G G , G G G <LF>
T T , T T T <LF>
PT PT , PT PT PT <LF>
N N , N N N <LF>
UW UW , UW UW UW <LF>
PCS PCS PCS PCS PCS PCS <LF>
n n n n n n t t t t t t pcs pcs pcs pcs pcs pcs <LF>
P 1 , 1 <LF>

G = Gross T = Tare PT = Pretare N = Net


UW = Unit Weight PCS = Quantity , Fill in blanks for zero “0” on the left.
n = Net t = Tare + Pretare pcs = Quantity, zero “0” on the left reserved
<LF> = 0x0A line feed
For example:
Gross = 0,500 kg
Tare = 0,150 kg
Pre-tare = 0,050 kg
Net = 0,300 kg
UW = 0,5g
PCS = 600
F R “ 5 2 0 P “ <LF>

? <LF>

SP 0 , 5 0 0 <LF>

SP 0 , 1 5 0 <LF>

SP 0 , 0 5 0 <LF>

SP 0 , 3 0 0 <LF>

SP 0 , 5 0 0 <LF>

SP SP SP 6 0 0 <LF>

0 0 0 3 0 0 0 0 0 2 0 0 0 0 0 6 0 0 <LF>

P 1 , 1 <LF>

<LF> = 0x0A (line feed) SP = 0x20 (Blank)

03004034 59 ZSME200000051
EXCELL PRECISION CO., LTD.

2. Press MC key to transmit the data.

F R “ 5 2 0 T “ <LF>

? <LF>

TN TN TN TN TN TN <LF>

TW TW , TW TW TW <LF>

TA TA TA TA TA TA <LF>

tn tn tn tn tn tn tw tw tw tw tw tw ta ta ta ta ta ta <LF>

P 1 , 1 <LF>
TN: Total number TW: Total weights TA: Total quantities
tn: Total number tw: Total weights ta: Total quantities
NOTE: Fill in zero “0” for the blanks on the left.
For example:
TN = 3
TW = 2,395 kg
TA = 23937

F R “ 5 2 0 T “ <LF>
? <LF>
SP SP SP SP SP 3 <LF>
SP 2 , 3 9 5 <LF>
SP 2 3 9 3 7 <LF>
0 0 0 0 0 3 0 0 2 3 9 5 0 2 3 9 3 7 <LF>
P 1 , 1 <LF>
<LF> = 0x0A (line feed) SP = 0x20 (Blank)

03004034 60 ZSME200000051
EXCELL PRECISION CO., LTD.

3-3-6 Continuous Transmission


Output Condition Setting

Press MC

Enter parameter 00, 01


Previous setting
according to demand

Press MC key to save


the setting data

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting: (output all)

⇒ Output all
⇒ No output under OL or unstable condition

The setting will be effective only when is set as


“continuous transmission mode”.

03004034 61 ZSME200000051
EXCELL PRECISION CO., LTD.

3-3-7 Zero Reset Condition for Automatic

Transmission Setting

Press MC

Enter parameter 00 ~ 99 Previous setting


according to demand

Press MC key to save


the setting data

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting: (external value “5d”)

⇒ External value “0d” (d=division)


⇒ External value “1d”
⇒ External value “2d”

⇒ External value “99d”


The range setting of is related to the zero reset setting in accumulation
acceptable condition in .
Only if the setting is set as will the setting be
effective.

03004034 62 ZSME200000051
EXCELL PRECISION CO., LTD.

3-3-8 Reset Condition Setting for Automatic

Transmission

Press MC

Enter parameter 00 ~ 99 Previous setting


according to demand

Press MC key to save


setting data.

key ⇒ Exit
CE key ⇒ Move the cursor leftward
TARE key ⇒ Move the cursor rightward
MC key ⇒ Enter

Default setting: ( External value “5d” )

⇒ External value “0d” (d=division)


⇒ External value “1d”
⇒ External value “2d”

⇒ External value “99d”

Only if Setting is set as “automatic transmission mode” will


setting be effective.

03004034 63 ZSME200000051
EXCELL PRECISION CO., LTD.

Appendix 1 RS-232 Full Duplex Format


Table 1. Command Format Illustration
Command Format A
Host Command
Slave Command

MZ Return to zero CP Clear off pre-tare value


MT Tare CT Clear off tare value
Current net weight accumulation & count
AT DT Clear off accumulated data and counts
plus 1
SC Set continuous transmission mode SA Set automatic transmission mode.
SM Set manual transmission mode SO Set command mode
UA Shift to first unit UB Shift to second unit
% Cease continuous transmission mode and enter into command mode

Command Format B
Host Command
Slave Data

Read current displaying


RW Read current displaying weight RB
weight(simple)
RG Read gross weight RT Read tare
RN Read net weight RI Read net weight (simple)
RH Read gross weight (simple) RE Read pre-tare (simple)
RU Read unit weight (simple) RD Read accumulated quantity (simple)
RC Read accumulated counts (simple) Rl Read tare (simple)
Rf Read pre-set name (ITEM) Rk Read accumulated weight (simple
accumulation format)
Rg Read ID# Rh Read weighing unit
RQ Read quantity (simple) Ri Read unit weight unit
Re Read PLU#
Note : Add % before italic and magnified letter to read continuously.
Add # before italic and magnified letter to read stable value only.
Two formats (AB) mentioned above are all RS-232 full duplex. If the slave terminal
receives the below-listed messages, it represents Error condition.
E1: Wrong command
E2: Wrong format (wrong parameter)
E3: Mismatch proceeding condition

03004034 64 ZSME200000051
EXCELL PRECISION CO., LTD.
Table 2. Output Format Illustration

General Format
Gross
S T , G S , + 1 . 2 3 . 4 5 6 l b o z
weight
Net weight S T , N T , + 1 2 . 3 4 . 5 6 T l . g
Tare S T , T R , + 0 1 2 . 3 4 5 6 SP SP k g CR LF
+ overload O L , G S , + SP SP SP SP SP SP SP SP SP SP SP SP
- overload O L , G S , - SP SP SP SP SP SP SP SP SP SP SP SP
Unstable U S , G S , + 0 1 2 3 . 4 5 6 SP SP l b
Totally 21 bytes (including CR LF)

Simple Format (price computing & counting)


ID# 0 0 0 0 0 0 0 0 0 0 0 2
CR LF
Read preset name SP SP SP SP SP SP SP A P P L E
Totally 14 bytes (including CR LF)

Simple Format
Read current weighing unit 0
Read current price computing unit 1 CR LF
Read current unit weight unit 2
Totally 3 bytes (Including CR LF)

Simple Format (price computing & counting & weighing)


Gross
+ 1 . 2 3 . 4 5 6
weight
Net weight + 1 2 . 3 4 . 5 6
Tare + 0 1 2 . 3 4 5 6
Pre-tare + 0 1 2 . 3 4 5 6
+ overload + SP SP SP SP SP SP SP SP
- overload - SP SP SP SP SP SP SP SP
Unstable + 1 2 3 4 5 6 7 8 CR LF
Quantity 0 1 2 3 . 4 5 6 7
Unit weight 0 0 0 0 0 0 0 0 1
Accumulated
0 0 0 0 0 0 0 0 2
counts
Accumulated
0 0 0 0 0 0 0 1 2
number
PLU# 0 1 2 3 4 5 6 7 8
Totally 11 bytes (including CR LF)

03004034 65 ZSME200000051
EXCELL PRECISION CO., LTD.
Simple Accumulation Format
Accumulated
SP 0 1 2 3 4 . 5 6 . 7
weight
Accumulated
SP 0 1 2 3 4 5 6 7 8 9
quantity
Accumulated
CR LF
weight + + SP SP, SP SP SP SP SP SP SP SP
overflow
Accumulated
weight - + SP SP, SP SP SP SP SP SP SP SP
overflow
Totally 13 bytes (including CR LF)

03004034 66 ZSME200000051
EXCELL PRECISION CO., LTD.

Appendix 2 Fixed Format RS-232 Transmission Line


Illustration
SCALE RS232 PRINTER

DB 9 DB 9
2 TX 3 TX
3 RX 2 RX
5 GND 5 GND

SCALE PC
DB 9 DB 9
2 TX 2 TX
3 RX 3 TX
5 GND 5 GND

Serial Data Transfer/Receive Format


LSB MSB 1
n,8,1 S 8-bit data STOP
0

1
e,7,1 S 7-bit data P STOP
0

1
o,7,1 S 7-bit data P STOP
0

S : Start bit
STOP: Stop bit
P : Parity bit

03004034 67 ZSME200000051
EXCELL PRECISION CO., LTD.

Appendix 3 ASCII Code Table


Symbol ASCII Code Symbol ASCII Code Symbol ASCII Code
A 41H a 61H 0 30H
B 42H b 62H 1 31H
C 43H c 63H 2 32H
D 44H d 64H 3 33H
E 45H e 65H 4 34H
F 46H f 66H 5 35H
G 47H g 67H 6 36H
H 48H h 68H 7 37H
I 49H i 69H 8 38H
J 4AH j 6AH 9 39H
K 4BH k 6BH ↵ 0DH
L 4CH l 6CH
M 4DH m 6DH
N 4EH n 6EH
O 4FH o 6FH
P 50H p 70H
Q 51H q 71H
R 52H r 72H
S 53H s 73H
T 54H t 74H
U 55H u 75H
V 56H v 76H
W 57H w 77H
X 58H x 78H
Y 59H y 79H
Z 5AH z 7AH

03004034 68 ZSME200000051
EXCELL PRECISION CO., LTD.

Appendix 4 7-Segment Display Characters


Number Display Letter Display Letter Display

0 A N

1 B O

2 C P

3 D Q

4 E R

5 F S

6 G T

7 H U

8 I V

9 J W

K X

L Y

℃ M Z

03004034 69 ZSME200000051