0% found this document useful (0 votes)
580 views28 pages

Ai-Dental Software User Guide

Uploaded by

mohamed saada
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
580 views28 pages

Ai-Dental Software User Guide

Uploaded by

mohamed saada
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

Ai-Dental Software Manual

1. Summary
Ai-Dental is the software component of i-Sensor H1/H2 and i-Scan, which is produced by
Guilin Woodpecker Medical Instrument Co., Ltd. It runs separately on the Windows system and
performs functions such as receiving image data from i-Sensor H1/H2 and i-Scan, as well as
image processing.

System function overview


The Ai-Dental software consists of several modules, each performing a specific function to

enable the case examination process and provide valid diagnostic image data.

Login module: Administrator registration, user login, automatic login, password remembering, etc.

Patient module: Register/modify/delete/view patient, image preview and image acquisition, etc.

Diagnosis module: Image preview and image processing, etc.

Report module: Add/delete/save/open report, etc.

Setting module: Basic system setting, clinic management, default processing algorithm setting, device

management, staff management.

2. Working conditions
 Minimum computer configuration requirements:

CPU:Quad-core 3.0GHz processor

Memory: 8GB

Hard disk: 500GB

Graphics card: NVIDIA GT710

Display resolution: 1920 * 1080 and above

 Software environment:

Windows7,Windows8,Windows10

 Network conditions:

LAN is available, with client and server in the same LAN.


3. Software function
3.1 Software installation
3.1.1 Double-click the installation program, as shown in Figure 1:

Figure 1
3.1.2 Select “Setup Language”, as shown in Figure 2:

Figure 2
3.1.3 After the installation program is started, click the “Browse” button to select the installation
path. After the path is selected, click the “Next” button, as shown in Figure 3:

Figure 3
3.1.4 Read the software license agreement, agree to this agreement and click “I accept the
agreement”, click Next to continue installation, disagree with this agreement and click “I do not
accept the agreement” to exit the installation program,as shown in Figure 4。
Figure 4
3.1.5 Select components. The user selects the corresponding component as needed, and then click
the “Next” button, as shown in Figure 5:

Figure 5
3.1.6 Set whether to create a desktop shortcut and server auto start, click the “Next” button after
completion, as shown in Figure 6:

Figure 6
3.1.7 Click the “Install” button to start the installation, as shown in Figure 7:

Figure 7
3.1.8 After the “Install” button is clicked, the program starts to install. The user just waits for the
installation to complete, as shown in Figure 8:

Figure 8
3.1.9 After the software is installed, click the “Finish” button, as shown in Figure 9:

Figure 9
3.2 Login module
Double-click “Ai-Dental-Server” to start the server. After the server is started successfully (as
shown in Figure 10), double-click “Ai-Dental-Client” to start the software and enter the software
login interface (as shown in Figure 11). Enter the user name and password, and click the “Login”
button to log in to the main interface, as shown in Figure 12.
The first time you use the software, you don’t have a user name and password, as shown in
Figure 13. Click “Sign up for free” to register the administrator, as shown in Figure 14. Enter user
name, password, confirm password, and other information to register successfully. Enter the user
name and password in the login interface to log in to the main interface of the software. The
administrator account has user management functions such as New User, Delete User, Modify
User and Search User. Ordinary users do not have user management functions. For details, please
refer to “6.6 Staff Management”.

Figure 10

Figure 11
Figure 12

Figure 13
Figure 14
3.3 Patient module
Click the “Patient” button to enter the patient module.
3.3.1 Add, delete, modify and query patients
The patient toolbar is as shown in Figure 15. Click the “Add Patient” button to enter the
information, click “OK” and a patient can be added, as shown in Figure 16. If you need to modify
the patient information, click the “Modify Patient” button to modify the patient information. Click
“Modify” and the modification can be successful, as shown in Figure 17. If you need to delete a
patient, click the “Delete Patient” button and click “OK” after 3 seconds, as shown in Figure 18.

Figure 15
Figure 16

Figure 17
Figure 18
Select the patient in the patient list, as shown in Figure 19. The patient information is
displayed on the patient information interface. Click “More Details” to view detailed patient
information. Enter patient information in the search bar. Click “Advance Search”, enter or select
information such as New Date, Age, Sex, and Doctor, and click the search button to query the
specified patient. If you only want to view the patients created by the current user, select “Current
User” in the Doctor option, as shown in Figure 20.

Figure 19
Figure 20
3.3.2 Image acquisition
Click the “Acquisition” button to enter the image acquisition interface.
Take the i-Scan acquisition as an example. After entering the software interface, the Wi-Fi
name will be displayed in the software status bar. Click the “Acquisition” button, connect the
i-Scan device to Wi-Fi, and select the “i-Scan” device type. When the device is connected, the IP
address will be displayed under the device type. Click “Open”, the device will enter the
acquisition state, and start to acquire images. As is shown in Figure 21:

Figure 21
Take the sensor as an example. After entering the software, click the “Acquisition” button,
connect the sensor device to the computer USB interface, and select the “Sensor” device type.
Click “Open”, the device will enter the acquisition state, and start to acquire images. At this time,
the sensor serial number is displayed in the software status bar, as shown in Figure 22. When the
sensor is used for the first time, the user will be prompted “Whether to download the calibration
file through the network”. Click “Yes” to start the download; click “No”, the user will be
prompted “Whether to manually import the calibration file”. Click “Yes” to select the calibration
file to import. Generally, the image acquired by the sensor will be better after the calibration file is
selected, as shown in Figures 23, 24.

Figure 22
Figure 23

Figure 24
Take file import as an example. After entering the software, click the “Acquisition” button.
Select the “File Import” device type and click “Import” to enter the image selection interface. The
software supports the import of images in PNG, JPG, JPEG, BMP, DCM and other formats, as
shown in Figure 25. After selecting the image, click “OK” to enter the “Import image” interface,
as shown in Figure 26. You can select the target patient and shooting time for each image, and
click “OK” to save the image to the specified patient.

Figure 25
Figure 26
When the software is used for the first time, the user will be prompted to set the relevant
parameters of the clinic’s X-ray source, as shown in Figure 27:

Figure 27
During the acquisition, problems such as network instability may be encountered, resulting in
the failure of image saving. In this case, there is an image saving failure mark in the lower right
corner of the image. Right-click the image and it can be exported to local and saved again, as
shown in Figure 28 and Figure 29.

Figure 28

Figure 29

3.3.3 Image preview


Click the “Preview” button to view acquired images. Select an image, right-click and select
“Export” to export the image to the local. Select “Information” to view the information of the
image. Select “Delete” to delete the image. Double-click the image to enter the diagnosis interface.
If you want to delete or export multiple images, you can click “Select...” to select the images you
want and then export or delete them. As is shown in Figure 30:
The image preview interface has filter image functions such as Date, Image Source, Image
Analysis, Tooth Profile, All images, etc. Click the “Refresh” button, and the software will
synchronize the latest image data of the patient and display it in the image list.
Figure 30
3.4 Diagnosis module
Click the “Diagnosis” button to enter the diagnosis module.
Click the image on the left to select an image to be processed. There are image processing
tools on the right side of the diagnosis interface, such as Display, Image Correction, Measuring,
View, Enhance, Sharpening, Histogram, Annotation, etc. Select the image processing tool and
adjust the image quality to a satisfactory level. In the Enhance, click the “HD” button and the
image will be enhanced. Click the “HD” button again to cancel the enhancement.
Use image processing tools to adjust image quality. Click the “Add Temporary State” button
to save the image quality at this time. Select the temporary state in the drop-down box to
reproduce the image.
There are delete, export, and image information functions above the processed image. These
functions are similar to the corresponding functions of the patient module. When multiple images
are selected for processing, click the “Clear” button to close them all.
Move the mouse wheel up and down to zoom in and out of the image. Hold down the right
button and move the mouse up to increase contrast, move down to decrease contrast, move left to
decrease brightness, and move right to increase brightness.
Select an image, and click the linear measurement icon (or angle measurement icon) in the
Measuring. Click the left mouse button to form the starting point and move the mouse. Click the
left mouse button again to form the end point and right-click the end point to end the measurement.
The measurement line will be displayed on the image, and meanwhile the corresponding
annotation of the measurement line will be displayed in the Annotation, as shown in Figure 31:
For images acquired by File Import, TWAIN, etc., the measured value may be inaccurate, and
can be calibrated through the calibration function. Select the measurement line, enter the actual
length of the measurement line in the Measuring, and click the “Modify” button to perform
calibration.

Figure 31
The dose of the image taken by our imaging plate scanner (i-Scan) and digital intraoral X-ray
imaging system (sensor) will be displayed on the left color column in the preview image on the
left side of the diagnostic module. Low and red column indicates that the dose is too low; orange
and relatively low column indicates that the dose is relatively low; green column indicates that the
dose is appropriate; orange and relatively high column indicates that the dose is relatively high;
red column indicates that the dose is too high, as shown in Figure 32:

(1)Too low dose (2)Relatively low dose

(3)Appropriate dose (4)Too high dose


Figure 32
Image processing function list
Adapt to window

Zooming to 100%

Forward rotation 90°

Reverse rotation 90°

Left and right reverse

Up and down reverse

Brightness

Contrast

Gamma

Pseudo-color

Reverse

Intra-oral Caries

Intra-oral High Definition

Intra-oral Fine

Straight line measurement

Angle measurement

Delete measurement

Scale

Calibration line

Flash lamp

Magnifying lens
Relief

Contour enhancement

3.5 Report module


Click the “Report” button to enter the report module. Click the “New Report” button to create
a new report template. Drag an image from the left to the image box, and enter the diagnosis result
in the text box, etc. If you need more pages, click “Add Page”, and a page will be added to the
report. After writing the report, click the “Save to Server” button to save the report to the server.
When you want to view the report, click “Open Report”, select the report you want, and click
“Open” to view. Click the “Export to PDF” button to export the report to the local. As is shown in
Figure 33:

Figure 33
3.6 Setting module
Click the “Setting” button to enter the setting module.
3.6.1 Basic setting
Click the “Basic Setting” button to enter the basic setting page. Click the “Sign Out” button
to return to the login interface. Click the language drop-down box to select the software language.
Click the tooth profile drop-down box and select the tooth profile number. As is shown in Figure
34:
Figure 34
3.6.2 Clinic management
Click the “Clinic Management” to enter the clinic management interface and enter the clinic
information, as shown in Figure 35:

Figure 35
3.6.3 Image processing
Click the “Image Processing” button to enter the image processing interface. Select the HD
checkbox, select “HD”, and the acquired image will be initialized and HD processed automatically.
Select “Fine” and the acquired image will automatically undergo initialization and fine processing.
Click the “Setting” button to set the initial color of the measurement line of the diagnostic
module. As is shown in Figure 36:
Figure 36
3.6.4 IP setting
Click the “IP Setting” button to enter the IP setting interface. Enter the IP address and port
number. Click “Connection Test” to view the test result. Click “Modify” to switch the connected
server, and the software need restarting at this time. As is shown in Figure 37:

Figure 37
3.6.5 Device management
Click the “Device Management” button to enter the device management interface, as shown
in Figure 38:

Figure 38
Before using this function, you need to determine whether the sensor is successfully
connected and whether “IO sensor connection successful, Please take images” is displayed in the
lower right corner. If it does not appear, please go to the acquisition interface in the patient
interface to connect the sensor. As is shown in Figure 39:
Figure 39
After the connection is successful, return to the device management interface in the setting
interface to set the sensor parameters. Before setting new parameters, first determine whether the
X-ray trigger mode is AC or DC. If it is DC mode, set “TriggerMode” to
“Enm_TriggerMode_AED_DC”, and “TriggerThreshold” to “Enm_TriggerThreshold_50uGy”; if
it is AC mode, set “TriggerMode” to “Enm_TriggerMode_AED_AC”, and “TriggerThreshold” to
“Enm_TriggerThreshold_50uGy”. Then click “Modify” and “Ok, Setup succeeded” will be
displayed after the modification is successful. As is shown in Figure 40:

Figure 40
If there is no picture during the shooting process, the exposure time can be appropriately
adjusted to re-expose. If the picture still cannot be produced, set “TriggerThreshold” to
“Enm_TriggerThreshold_100uGy” or “Enm_TriggerThreshold_200uGy” for re-exposure. If the
picture still cannot be produced, please contact the relevant personnel.
Note: If incorrect parameter settings are made or the sensor is not connected, “Sorry, Setup failed”
will appear and the previous parameters will be displayed, as shown in Figure 41:
Figure 41
3.6.6 Staff Management
Only when the administrator account is logged in to the software, there will be a “Staff
Management” button in the setting module. Click the “Staff Management” button to enter the staff
management interface, as shown in Figure 42.
The staff management interface has the functions of New User, Delete User, Modify User
and Search User. Click the New User and enter the User Name, Login Password, Confirm
Password and other information, as shown in Figure 43. Enter the user name in the search bar to
query the specified user. Double click the staff information bar to query the details of the user.
When the administrator forgets the login password, open the server interface. Click the
“Setting” button -> click “Password” -> check “Show password”, you can view the administrator
password, as shown in Figure 44. If an ordinary user forgets the login password, double-click the
staff information bar, and the staff information interface will pop up. Press and hold the password
viewing button, the login password input box will display the staff’s password, as shown in Figure
45. The problem of forgetting the password can also be solved by changing the staff’s password.

Figure 42
Figure 43

Figure 44
Figure 45
3.7 Help document
Log in to the main interface of the software and click “?” on the upper right corner. There are
online help documents such as Ai-Dental manual, App Manual, i-Scan Manual, FAQ, etc. In
addition, “About Ai-Dental” can be clicked to check the software version. As is shown in Figure
46:

Figure 46
3.8 Software upgrade
The software has an automatic upgrade function. If the Ai-Dental server is installed on the
computer and the software can be upgraded, there will be a pop-up window indicating whether to
upgrade. Click “Yes”, as shown in Figure 47. After the upgrade is completed, click “Finish”, as
shown in Figure 48. After started, all clients connected to the server will receive the upgrade
countdown 10s. The software will automatically close after 10s, and start to upgrade. Click “No”,
the software will not be upgraded temporarily.
Figure 47

Figure 48
3.9 Exception handling
(1)In case of unclear image, please adjust the exposure time first. If repeated adjustment
of the exposure time is invalid, you can go to the software “Setting” -> “Device Management” to
set the sensor parameters. Please refer to 6.5 for specific setting methods.
(2)During the use of the software, when the server is unavailable due to some reasons, the
software will enter the server disconnect interface, as shown in Figure 49. At this time, the user
can check whether the network is normal or restart the server.

Figure 49
4. Training and maintenance
After the user purchases this product, the dealer will come to train the operator. If there are
still questions, you can contact the dealer. In addition, users are not allowed to change the intended
use of the product or software.
When there is a problem with the device or software, contact the dealer, who will contact
Woodpecker and we will then arrange professional staff for maintenance.

5. Others
Please operate the device and software in good light so as not to interfere with viewing image
details.
The distance between i-Scan and Ai-Dental should be within 10 meters.

You might also like