RDS-200
Universal Survey Meter
FeatUReS RDS-200
• dose rate survey measurements
• dose measurements
the RDS-200 Meter is an excellent, portable multipurpose
• contamination monitoring (with external probes)
radiation meter for a wide range of applications. It is espe
• area monitoring
cially designed for situations where accurate measurements
• can form a monitoring network with aaM
at low dose rate levels are of importance. It is suited for a
software
wide range of applications in military, civil defense, industrial
• active data transfer with telephone modem
and laboratory use etc. due to its versatile functions and
support
durability.
• programmable reading of dose rate averages
• interfaces for data transfer and external probes the meter has an interface for the external gamma probes
• data logging and historgam downloading to PC GMP-12H/12L or beta/contamination measurement probe
• time based or manually triggered data logging GMP-11/15. a connector for the attachment of the meter to
a PC is located at the bottom part of the meter and equipped
with protective cover.
the RDS-200 utilizes field-proven measurement electronics
and can also bet used as a local display unit with the RADOS
aaM-90 area Monitoring System. Featuring:
LAURUS Systems, Inc. - Ph: 410-465-5558 - Fax: 410-465-5257 - [Link]
teCHnICaL SPeCIFICatIOnS:
Radiological Characteristics
• radiation detected: gamma and X-rays, 50 keV...1.3MeV. Beta radiation with an external probe
• detectors: two energy-compensated GM tubes. energy response according to ambient dose equivalent
• dose rate measurement range: 0.01 μSv/h...10 Sv/h or 1 μrem/h...1000 rem/h
• dose measurement range: 0.01 μSv...10 Sv or 1 μrem...1000 rem
• resolution: three significant digits or 0.01 μSv/h on dose rate and 0.01 μSv on dose ( or 1 μrem/h on dose
rate and 1 μrem on dose)
• calibration accuracy: ± 5%, 137Cs , calibration direction and in the calibration field, temperature +20 °C
(68°F)
• dose rate linearity: ± 15% ± least significant number (0.05 μSv/h...10 Sv/h or 5 μrem/h to 1000rem/h)
• *variation of the response due to photon radiation energy (Re) and angle (Ra) of incidence within
± 45° from calibration direction:
- 0.05 μSv/h...10 Sv/h (5 μrem/h...1 rem/h): 80% < Re < +130%(50...1300 keV); 75% < Ra < +125% (65
keV)
- 10 mSv/h...10 Sv/h (1...1000 rem/h): 85% < Re < +115% (80...1300 keV); 65% < Ra < +135% (83 keV)
*note that at energy levels between 50...80 keV overflow alarm is given if dose rate 10 mSv/h or 1 rem/h
is exceeded
Functional Characteristics
• data storage: the data logging interval of the instrument can be set from 10 s to 99 minutes or data can be
manually triggered and it will memorize the 864 last measurement results in its internal memory
• diagnostics of faulty detector
• Sv, rem (display “R”) or Gy units configurable on display
• configurable chirp function
• adjustable alarm for dose and dose rate limit
• overflow alarm for dose rate
electrical Characteristics
• power supply: 3 alkaline batteries (IeC LR6 / aa), +12 V DC external battery adapter (optional) or AC
adapter (optional )
• battery life: 200 h in background field (+ 25°C / 77°F)
• battery alarm: 15 h before battery power-out
Mechanical Characteristics
• case: impact resistant, aluminium profile body with aBS plastic end caps, enclosure class IP67 (IE-529)
shielded against RF interference and neMP.
• customized LCD display with blue electroluminencse backlight
• dimensions: 92 x 199 x 44 mm (3.62 x 7.83 x 1.73 in)
• weight: 610 g without batteries (1.34 lb), 700 g with batteries (1.54 lb)
environmental Characteristics
temperature:
• -30°C...+55°C (-22°F to 131°F), operating
• -40°C...+55°C (-40°F to 131°F), operating (restricted display operation)
• -40°C...+70°C (-40°F to 158°F), storage
LAURUS Systems, Inc. - Ph: 410-465-5558 - Fax: 410-465-5257 - [Link]