Table top Autoclave
“Icanclave” Series C, M, T
Complies with the following standards.
- Medical devices directive 93/42/EC
- Small steam sterilisation directive EN13060
- Pressure Equipment Directive 97/23/EC
- CEI Directives EN 61326- 1, EN 61010-1 and EN 61010-2-40
PHONE: 1300 2255 42 | WEBSITE: [Link] | EMAIL: info@[Link]
BOTH CLASS N AND B CLASS TABLETOP AUTOCLAVES
STE-8-C / STE-8-M 8 Litre benchtop autoclaves available as class N and class B. The B class autoclaves
meets the European standard EN13060. This 8 Litre autoclave is light weight, easy
to move and requires minimal benchspace making it an ideal solu�on for small
businesses such as ta�oo, podiatry, beauty, veterinary and small microbiology
requirements.
STE-16-C / STE-16-M 16 Litre benchtop autoclaves available as class N and class B. The B class autoclaves
meets the European standard EN13060. This 16 Litre autoclave is an economic
solu�ons for dental, private clinics and larger ta�oo, podiatry, beauty, veterinary
prac�ces and medium microbiology requirements.
STE-18-C / STE-18-M 18 and 23 Litre benchtop autoclaves available as class N and class B. The B class
STE-23-C / STE-23-M autoclaves meets the European standard EN13060. These autoclaves are ideal for
busy dental, medical, veterinarian, beauty and laboratory purposes.
STE-29-T This 29 litre class B autoclave is an extended version of the 16 litre autoclave men-
�oned above. The chamber has been extended to 580mm in length which allows
this autoclave to sterilise long instruments which may not fit inside other models.
PHONE: 1300 2255 42 | WEBSITE: [Link] | EMAIL: info@[Link]
What is class B and class N autoclave What is class B and class N autoclave
The class B autoclave is defined by a pre sterilisa�on The class B autoclave is defined by a pre sterilisa�on
vacuum cycle. The class B is considered to be the highest vacuum cycle. The class B is considered to be the highest
class of autoclave and can be usedto sterilise all loads class of autoclave and can be usedto sterilise all loads
including solids, type A hollow instruments, type B including solids, type A hollow instruments, type B
hollow instruments, porous loads and wrapped instru- hollow instruments, porous loads and wrapped instru-
ments. ments.
All B type autoclaves must pass a Helix test as per stan- All B type autoclaves must pass a Helix test as per stan-
dard EN 13060 outlined by EN 867-5:2001. THis can only dard EN 13060 outlined by EN 867-5:2001. THis can only
be achieved by use of a pre sterilisa�on frac�onated be achieved by use of a pre sterilisa�on frac�onated
vacuum. vacuum.
Post sterilisa�on vacuum drying ensures complete drying Post sterilisa�on vacuum drying ensures complete drying
of all loads a�er sterilisa�on process is complete. of all loads a�er sterilisa�on process is complete.
Class N autoclaves do not feature a vacuum cycle and are Class N autoclaves do not feature a vacuum cycle and are
suitable for sterilising unwrapped solid instruments and suitable for sterilising unwrapped solid instruments and
liquids liquids
Graphic of working Cycle
PHONE: 1300 2255 42 | WEBSITE: [Link] | EMAIL: info@[Link]
Product Feature
Main Feature: Safety:
1. Double door protec�on system prevents the cycle from star�ng if the door
is not correctly locked. This system also prevents the door from opening if the
1. Fully automa�c benchtop autoclave pressure inside the chamber is not equal to atmospheric pressure outside the
with pre set programs chamber.
2. Complies with European standard EN 2. Pressure safety valve prevents over pressure in teh chamber and steam
13060 generator.
3. In built independent rapid steam gener- 3. The power is automa�cally cut if a short circuit or electrical fault occurs.
4. Able to detect and iden�fy the exact cause of any problem and to issue a
ator ensures quick sterilisa�on cycles.
specific error code to the generator.
4. Rapid post sterilisa�on vacuum drying. 5. Water level main switch in teh main tank controlling the min and max
5. User friendly interface. water level.
6. LCD screen displaying �me, tempera-
ture and pressure, process alert and
instrument condi�ons.
Documenta�on:
1. Printer: an external printer is op�onal for all Icanclav autoclaves.
7. Choice of language se�ngs including: 2. USB port: it is used for connec�ng a USB s�ck, all sterilisa�on data ae auto-
English, Spanish, German and Chinese. ma�cally wri�en in the USB s�ck and can be read directly in any PC and
8. Easily accessible water tank stored electronically.
9. Liquid sterilisa�on program op�onal. 3. Internal Memory: the last 20 cycles can be automa�cally stored in the
autoclave system which can be print out any �me.
PHONE: 1300 2255 42 | WEBSITE: [Link] | EMAIL: info@[Link]
Technical Data
Class B Autoclave
Model STE - 8 -M STE - 16-M STE - 18 - M STE - 23 - M STE - 29 -M
Volume 8 litre 16 litre 18 litre 23 litre 29 litre
Chamber size Ø 170 x 320mm Ø 230 x 360mm Ø 247 x 350mm Ø 247 x 450mm Ø 247 x 600mm
Power 1500w 2000w 2200w 2200w 2200w
Net weight 34.5 kgs 45 kgs 47 kgs 51 kgs 55 kgs
Pre- programmed cycles
PROGRAM TEMP PRESSURE STERILISATION TOTAL TIME (min)
(°C) (Mpa) TIM (Min) STE - 8 - C STE - 16 - C & STE - 23 - T STE - 18 - C & STE - 23 - C
SOLID 134 210 4 15-25 14-30 20-45
121 110 20 25-40 30-45 35-60
LIQUID 134 210 10 25-50 30-55 35-60
121 110 30 30-55 35-60 35-60
WRAPPED 134 210 4 20-45 30-45 45-65
121 110 20 30-50 30-50 50-75
TEXTILE 134 210 4 20-45 30-45 45-65
121 110 20 30-50 30-50 50-75
PRION 134 210 18 30-50 35-55 45-75
DRYING ---- ---- ---- 1-20 1-20 1-20
B&D TEST 134 210 3.5 22-35 22-35 22-35
HELIX TEST 134 110 3.5 22-35 22-35 22-35
VACUUM TEST ---- ---- ---- 15-20 15-20 15-20
PHONE: 1300 2255 42 | WEBSITE: [Link] | EMAIL: info@[Link]