100%(1)100% found this document useful (1 vote) 148 views54 pages1590A Manual
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content,
claim it here.
Available Formats
Download as PDF or read online on Scribd
Normal use of test equipment exposes you to a certain amount of danger from electrical
shock because testing must often be performed where exposed high voltage is present. An
electrical shock causing 10 milliamps of current to pass through the heart will stop most
human heartbeats. Voltage as low as 35 volts dc or ac rms should be considered dangerous and
hazardous since it can produce a lethal current under certain conditions. Higher voltage poses
an even greater threat because such voltage can more easily produce a lethal current. Your
normal work habits should include all accepted practices that will prevent contact with
exposed high voltage, and that will steer current away from your heart in case of accidental
contact with a high voltage. You will significantly reduce the risk factor if you know and
observe the following safety precautions:
1. Don't expose high voltage needlessly in the equipment under test. Remove housings and
covers only when necessary. Turn off equipment while making test connections in high~
voltage circuits. Discharge high-voltage capacitors after removing power.
2, If possible, familiarize yourself with the equipment being tested and the location of its
high voltage points. However, remember that high voltage may appear at unexpected
points in defective equipment.
Use an insulated floor material or a large, insulated floor mat to stand on, and an
insulated work surface on which to place equipment; make certain such surfaces are not
damp or wet.
Use the time-proven "one hand in the pocket" technique while handling an instrument
probe. Be particularly careful to avoid contacting a nearby metal object that could
provide a good ground return path,
5. When using a probe, touch only the insulated portion. Never touch the exposed tip
portion.
When testing ac powered equipment, remember that ac line voltage is usually present on
some power input circuits such as the on-off switch, fuses, power transformer, etc, any
time the equipment is connected to an ac outlet, even if the equipment is turned off.
7. Some equipment with a two-wire ac power cord, including some with polarized power
plugs, is the "hot chassis" type. This includes most recent television receivers and audio
equipment. A plastic or wooden cabinet insulates the chassis to protect the customer.
When the cabinet is removed for servicing, a serious shock hazard exists if the chassis is
touched. Not only does this present a dangerous shock hazard, but damage to test
instruments or the equipment under test may result from connecting the ground lead of
most test instruments (including this oscilloscope) to a "hot chassis". To make measure~
ments in "hot chassis" equipment, always connect an isolation transformer between the
ac outlet and the equipment under test. The B & K-Precision Model TR-110 or 1604
Isolation Transformer, or Model 1653 or 1655 AC Power Supply is suitable for most
applications. To be on the safe side, treat all two-wire ac powered equipment as "hot
chassis" unless you are sure it has an isolated chassis or an earth ground chassis.
8 Never work alone. Someone should be nearby to render aid if necessary. Training in
CPR (cardiopulmonary resuscitation) first aid is highly recommended.
ywInstruction Manual
for
Ie 'T< PRECISION
Model 1590A
100MHz
Quad-Trace Oscilloscope
‘This symbol on oscilloscope means "refer to instruction manual
for further precautionary information", This symbol appears in
the manual where the corresponding information is given,
DYNASCAN
EG CORPORATION
6460 West Cortland Street
Chicago, Ilimois 60635TABLE OF CONTENTS
Page
TEST INSTRUMENT
SAFETY ..
FEATURES
SPECIFICATIONS ..
CONTROLS AND INDICATORS
Introduction ...
Vertical Controls
Horizontal Controls
Triggering Controls «
Rear Panel Controls.
OPERATING INSTRUCTIONS
Safety Precautions
Equipment Protection
Operating Tips
Initial Starting Procedure
Single Trace Display .
Dual Trace Display
Quad Trace Display
Triggering .
Magnified Sweep Operation
Delayed Sweep Operation...
Dual Sweep Operation.
Single Sweep Operation ..
X-Y Operation,
inside front cover
Cascaded Operation...
Video Signal Observation.
APPLICATIONS ar 20
DC Voltage Measurements . 2
Measurements of Voltage Between
‘Two Points on a Waveform.
Elimination of an
Undesired Signal Component...
Time Measurements
Frequency Measurements
Pulse Width Measurements.
Pulse Rise Time and
Fall Time Measurements.
Time Difference Measurements
Phase Difference Measurements...
Relative Measurements
Waveform Magnification Using
Delayed Sweep «s+.
Pulse Jitter Measurements
High Accuracy Time Measurements
Using Delayed Sweep..
Pulse Width Measurements
Using Delayed Sweep.
Frequency Measurements
Using Delayed Sweep.
Pulse Repetition Time Measurements
Using Delayed Sweep..
Rise Time and Fall Time Measurements
Using Delayed Sweep. coos 0
Time Difference Measurements
Using Delayed Sweep.
Observing Video Signals
Using Delayed Sweep.
X-Y Mode Applications .
Quad Trace Applications
Dual Sweep Applications
ACCESSORIES
MAINTENANCE.
Fuse Replacement.
Periodic Adjustments .
Calibration Check.
Instrument Repair Service
Additional Servicing Information ...
APPENDIX I...
12039
49
WARRANTY SERVICE INSTRUCTIONS... 50
LIMITED ONE-YEAR WARRANTY s.sesee0+- 51HIGH FREQUENCY FEATURES
Wide Bandwidth
Conservatively rated ~3 dB bandwidth is de
to 100 MHz, including X5 GAIN mode,
Fast Rise Time
Rise time is 3.5ns, including X5 GAIN
mode.
Fast Sweep
Maximum sweep speed (with X10 MAG) of
2 ns/div assures high frequencies and short
duration pulses are displayed with high
resolution.
Signal Delay Line
Signal is delayed with respect to triggering
to give at least 10 ns visible delay. Permits
viewing leading edge of high frequency and
fast risetime pulses.
DUAL TIME BASE FEATURES
‘Dual Sweep Generators
‘A sweep gives normal waveform display,
B sweep may be operated at faster sweep
speed to expand a portion of the waveform.
Six Modes of Horizontal Display
Choice of A sweep only, B sweep only, A
sweep with B sweep portion intensified, ‘al-
temating A and B sweep (trace separation
adjustable), dual sweep, and X-Y.
B Sweep Delay
Start of B sweep selectable from trigger-
able after delay and continuous delay
(starts immediately after delay) modes.
Delay adjustable from 0.2 to 10 times A
sweep time. Permits expansion of any por-
tion of waveform. Start of B sweep can
also be completely independent of A sweep
(DUAL sweep operation).
Variable TRACE SEPeration
TRACE SEP control to allow for adjustment
of A and B sweep separation for easy
observation of both traces.
DUAL/QUAD TRACE FEATURES
Quad Trace
Four vertical input channels for displaying
four waveforms simultaneously. Selectable
single trace (either CH1 or CH2), dual
trace, or quad trace operation. Alternate or
chop sweep selectable at all sweep rates.
‘Sum and Difference Capability
Permits algebraic addition or subtraction of
channel 1 and channel 2 waveforms, dis~
played as a single trace. Useful for differ
ential voltage and distortion measurements.
CRT FEATURES
Rectangular CRT
Rectangular CRT with large 8 x 10 centi-
meter viewing area. Internal 8 x 10 division
graticule eliminates parallax error.
High Brightness
Domed mesh type CRT with post deflection
acceleration and high brightness phosphors.
High acceleration voltage of 20 kV.
Convenience
Variable scale illumination for easy viewing
in darkened area and for waveform photo-
graphs. Trace rotation and astigmatism are
electrically adjustable from front _ panel.
0%, 10%, 90%, and 100% markers for rise
time measurements.
VERTICAL FEATURES
High Sensitivity
5 mV/div (1 mV/div with 5X GAIN selected)
sensitivity for full bandwidth.
Calibrated Voltage Measurements
Accurate voltage measurements (42%) on
10 calibrated ranges from 5 mV/div to 5 V/
div. Vertical gain fully adjustable between
calibrated ranges.
Selectable Channel 3 and 4 Sensitivity
Sensitivity of 0.1 V/Div (1/1 attenuation) or
1.0 V/ Div (1/10 attenuation) for channel 3
and 4 inputs.
SWEEP FEATURES
Calibrated Time Measurements
Accurate (£2%) time measurements. The B
sweep has 20 calibrated ranges from
50 ms/div to 0.02 1s/div. The A sweep has23 calibrated ranges from 0,5 s/div to
0.02 is/div. Sweep time is fully adjustable
between calibrated ranges.
X10 Sweep Magnification
Allows closer examination of waveforms,
increases maximum sweep rate to 2 ns/div.
TRIGGERING FEATURES
‘Three Trigger Modes
Selectable normal (triggered), automatic, or
single sweep modes.
‘Triggered Sweep
Sweep remains at rest unless adequate trig~
ger signal is applied. Fully adjustable trig~
ger level and (+) or (-) slope.
AUTO Sweep
Selectable AUTO sweep provides sweep
without trigger input, automatically reverts
to triggered sweep operation when adequate
trigger is applied.
Single Sweep
Permits viewing and photographing one-
time events.
Six A Sweep Trigger Sources
Six A sweep trigger source selections,
inching [Link], CH1, CH 2, 1/1
CH 3/EXTernal, 1/10 CH 3/EXTernal, and
LINE, In V. MODE, each waveform dis
played becomes its own trigger (alternate
triggering in ALT dual-trace or quad-trace
mode).
Five B Sweep Trigger Sources
Five B sweep trigger source selections, in-
cluding CH 1, CH 2, 1/1 CH 4 EXTernal,
1/10 CH 4 EXTernal, and STARTS AFTER
DELAY.
Five A Sweep Trigger Coupling Choices
Selectable AC, DC, high-frequency reject,
low-frequency reject, or video trigger
coupling.
Four B Sweep Trigger Coupling Choices.
Selectable AC, DC, high-frequency reject,
or low-frequency reject trigger coupling.
Video Sync
FRAME or LINE triggering automatically
selected by SWEEP TIME control.
Variable Holdoff
‘Trigger inhibit period after end of sweep
adjustable. Permits stable observation of
complex pulse trains.
Offset Chop Frequency
PULL CHOP F SELECT Control permits use
of an offset chop frequency.
FIX Trigger
PULL FIX Control automatically sets
trigger level at the average level of the
trigger waveform, regardless of the setting
of the LEVEL control,
OTHER FEATURES
LED Indicators
LED indicators show when the 5X GAIN is
selected or VARIABLE controls are set to
an UNCALibrated position, Indicators are
also provided to show HORIZontal DISPLAY
mode, vertical MODE, TRIGger MODE,
20 MHz BW limiting, and CH 2 INVersion.
X+¥ Operation
Channel 2 can be applied as horizontal de~
flection (X-axis) while channel 1 provides
vertical deflection (Y-axis).
A and B Gate Outputs
Gate signal outputs for both A and B sweep
provided for synchronizing _ peripheral
equipment.
Z Axis Input
Intensity modulation capability permits
time or frequency markers to be added.
Trace brightens with negative signal, TTL
compatible.
Built-In Probe Adjust Square Wave
A1 V pep, 1 kHz square wave generator
permits probe compensation adjustment.
Built-In Current Probe Adjust Square Wave
‘A 10 mA pep, 1 kHz square wave current
generator permits current probe compen-
sation adjustment.
Channel 1 Output
Buffered 50% output of channel 1 signal
available on rear panel for driving frequen~
cy counter or other instruments. Output is,
50 mV/div into 509 and at full bandwidth
rating.
Auto Focus
‘Auto focus circuit keeps waveform in focus
with changes in intensity (for example, the
intensified portion of trace).SPECIFICATIONS
CRT:
‘Model:
150KTM31.
Typet ;
Rectangular with internal graticule.
Display Area:
8 x 10 div(l div = 1 cm).
Accelerating Voltage:
20 kV.
VERTICAL AMPLIFIERS (CH 1 and CH 2)
Sensitivity:
5 mv/div to 5 V/div (X1 mode).
1 mV/div to 1 V/div (K5 mode),
500 1WV/div (cascaded operation).
Accuracy:
42% (+10° to 435°C).
44% (0° to +50° C).
47% (cascaded operation, CH 1 to CH 2).
Input Resistance:
1 MQ 41%,
Input Capacitance:
22 pF.
Frequency Response:
DC: — DC to 100 MHz (-3 dB), including
X5 Gain Mode.
DC to 70 MHz (-34B, cascaded
‘operation, CH 1 to CH 2).
AC: 5 Hz to 100 MHz (-3 dB), including
X5 Gain Mode.
7 Hz to 70 MHz (-34B, cascaded
operation, CH 1 to CH 2).
Rise Tim
3.5 ns.
Signal Delay Time:
Fixed, approx. 10 ns visible delay.
Crosstalk:
-40 4B minimum,
Operating Modes:
CH: CH1, single trace.
CH2: CH 2, single trace.
DUAL: CH 1 and CH 2, dual trace.
ADD: algebraic sum of CH 1 + CH 2,
single trace (CH 1 - CH 2
with CH 2 inverted).
QUAD: CH 1 thru CH 4, four trace.
ALT: dual or quad trace, alternating.
CHOP: dual or quad trace, chopped.
Chop Frequency:
Approx. 250 KHz, offset chop frequency
selectable.
Polarity Reversal:
CH 2 only.
Maximum Input Volta,
400 V (dc + ac peak) or 800 V ac p-p.
Maximum Undistorted Amplitud
8 divisions (dc to 100 MHz),
Bandwidth Limiting:
Frequency response can be limited to
approximately 20 MHz by bandwidth limit
switch.
VERTICAL AMPLIFIERS (CH 3 and CH 4)
Sensitivity:
0.1 V/div, 1 V/div 42%.
Attenuator:
V1, 1/10
Input Resistance:
1 Mi £1%,
Frequency Response:
DC to 100 MHz (3 4B).
Rise Time:
3.5 ns.
Signal Delay Time:
‘Same as CH 1 and CH 2.
Maximum Input Displayed:
DC Component: 20.5 V (#5 V,
1/10 attenuated),
AC Component: 1V p-p (10 V Ep,
1/10 attenuated).SPECIFICATIONS
‘Maximum Input Voltage:
400 V (de + ac peak).
HORIZONTAL AMPLIFIER
(input through channel 2 input)
X-¥ mode: switch selectable using HORIZ
DISPLAY switch assembly.
CH 1: ¥ axis.
CH 2: X axis.
Sensitivity:
Same as vertical channel 2.
Accuracy:
Same as vertical channel 2.
Frequency Response:
DC: DC to 5 MHz (-3 4B).
AG: 5 Hz to 5 MHz (-3 dB).
X-Y Phase Difference:
3° of less at 100 kHz.
SWEEP SYSTEM
Modes (switchable with the HORIZ DISPLAY
switch assembly):
A: A sweep.
ALT: ‘A sweep (intens
for duration of B sweep) and
B sweep alternating.
AcINT-B: Duration of B sweep is
displayed as an intensified
portion of the A sweep.
BDLY'D: Delayed B sweep.
DUAL: Dual sweep (A and B sweep
independent).
XY: X-Y display mode.
A Sweep:
20 ns/div to 0.5 s/div in 1-2-5 sequence,
23 steps. Vernier control provides fully
adjustable sweep time between steps.
B Sweep:
20 ns/div to 50 ms/div in 1-2-5 sequence,
20 steps.
Accuracy:
42% (+10° to 435° C).
44% (0° to +50° C).
Sweep Magnification:
10X, £5% (+10° to +35° C),
26% (0° to + 50° C).
Linearity:
20 ns/div to 0.5s/div $3% (45% with X10
magnification).
Holdo:
Continuously adjustable for A sweep from
NORM to 5 times normal.
‘Trace Separation:
B sweep can be positioned from 0 to 4 div-
isions below A sweep, continuously adjust-
able.
Delay Method:
Continuous delay or trigger delay.
Delay Time:
0.2 to 10 times A SWEEP TIME/DIV setting,
continuously adjustable.
‘Time Difference Measurement Accuracy:
+ (19% of measurement + 0.1% of fullseale)
(+10° to +35° Cc).
44% (0° to +50° C).
Delay Jitter:
1/20,000 of full scale sweep time.
TRIGGERING
A Trigger:
‘Modes:
AUTO (free run), NORM, SINGLE/RE-
SET, FIX (at center of the waveform).
Source:
V. MODE, CH 1, CH 2, 1/1 and 1/10
(EXT) CH3, LINE.
(Coupling:
AC 20 Hz to 100 MHz.
LF Rej 30 kHz to 100 MHz.
HF Rej 20 Hz to 30 kHz,
pc DC to 100 MHz.
VIDEO Trigger signal is routed through
sync separator. Horizontal sync
pulses selected at sweep times of
50\8/div and faster (LINE). Vertical
sync pulses selected at sweep times of
0.1 ms/div and slower (FRAME).
Level
490° permits selection of triggering at
any point on the positive and negative
slope of the displayed waveform.
Slope:
+h
B Trigger:
Source:
CH 1, CH 2, 1/1 and 1/10 (EXT) CH 4,
STARTS AFTER DELAY.
Coupling:
‘AC, LFrej, HFrej, DC.Level:
490°; permits selection of triggering at
any point on the positive and negative
slope of the displayed waveform.
Slope:
+
‘Trigger Sensitivity:
: EXxT| EXT
\Coupling] Bandwidth mt) tn | io
DC DC-20 MHz | 0.5 div| 50 mv] 0.5 V
DC-50 MHz | 1.0 div|100 mv| 1.0 V
DC-100 MHz| 1.5 div|210 mV) 2.1 V
AC Same as DC except lower limit |
is 20 Ha.
HFREJ | Same as AC except increased
attenuation above 30 kHz.
LFREJ | Same as AC except increased
attenuation below 30 KHz.
VIDEO | FRAME 0.5 div] 50 mV] 0.5 V
LINE 0.5 div | 50 mV|0.5 V
AUTO | Same as AC except lower limit
is 50 Ha.
FIX | 40 Hz-20 MHz 1.0 div (100 mV)
40 Hz-100 MHZ 1.5 div (210 mV)
Jitter: 0.5 ns maximum at 100 MHz at
2 ns/div sweep rate (X10 MAG)
OTHER SPECIFICATIONS
Calibrating Voltage and Current:
1 kHz £3% Positive Square Wave.
1 V pp 41% (+10° to +35°C); 42% (0° to
+50°C).
10 mA p-=p 42% (+10° to +35°C); 44% (0°
to +50°C).
Intensity Modulation:
Input Signal:
TTL level, intensity increasing with more
negative levels.
Input Impedance:
‘Approx. 10 kit.
SPECIFICATIONS
Usable Frequency Range:
DC to 10 MHz,
‘Maximum Input Voltage:
50 V (de + ac peak).
(CH 1 Output (on rear panel):
Output Voltage:
50 mV/div (into 50-ohm load).
(Output Impedance:
Approx, 50 ohms.
Frequency Response:
DC to 100 MHz (-3 4B),
(into 50-chm load).
Gate Output Voltage (A and B):
Approx. 1.5 V positive gate (into 500 2).
‘Trace Rotati
Electrical, front panel adjustable.
Power Requirements:
90-264 VAC; 45-400 Hz, approx. 56 watts.
Dimensions:
13.8 x 28.4 x 40 cm (5 x 1-1/4 x 15-3/4"),
Weight:
7.4 kg (16-1/4 Ibs).
Environment:
Operating:
+10° to 435° C, 85% maximum relative
humidity for full specifications.
0° to +50° C, 90% maximum relative hum-
idity, full operating range.
Storage:
20° to 470° G, 80% maximum selative
humidity.
Altitude:
Operating: up to 5000 meters (16,000 £t),
Non-operating: up to 12,000 meters
39,000 ft).
ACCESORIES SUPPLIED:
Two 10:1 Probes, Model PR-45.
AC Power Cord,
Probe Holder.
Instruction Manual.
Schematic Diagram and Parts List.INTRODUCTION
ELECTRONIC MODE SELECTION
Soft-touch pushbuttons select the desired
modes of operation. However, the actual
switching is accomplished by logic switching
circuits, ‘The logic circuits sense the combin-
ation of vertical, horizontal, and triggering
modes selected and make all necessary circuit
interconnections, This includes electronic
interlocking when applicable so that selection
of a new mode automatically cancels the pre-
vious selection. Additionally, the logic cir-
cuits indicate the presently selected modes by
illuminating the face of those pushbuttons.
MODE SELECTION MEMORY
The Model 1590A oscilloscope incorporates
mode selection memory; that is, once the
operating modes are selected, they are
retained in memory, even when the instrument
is turned off. This is a major convenience,
allowing you to restart operation in the same
modes as when the scope was turned off.
Fig. 1. Front Panel Controls And Indicators.CONTROLS AND INDICATORS
VERTICAL CONTROLS
CHANNEL 1 CONTROLS
1
VOLTS/DIV Control, Vertical attenuator
for channel 1. Provides step adjustment
of vertical sensitivity. When channel 1
VARIABLE control is set to CAL, verti-
cal sensitivity is calibrated in 10 steps
from 5 mV/div to 5 V/div in a 1-2-5 se~
quence. Vertical sensitivity is five times
greater when X5 GAIN is selected. In
X-Y operation, this control provides step
adjustment of Y-axis sensitivity.
VARIABLE, PULL X5 GAIN Control.
VARIABLE:
Rotation provides vernier adjust
ment of channel 1 vertical gain. In
the fully clockwise (CAL) position,
the vertical attenuator is calibrat~
ed. In X-Y operation this control
becomes the vernier Y-axis gain
control.
PULL X5 GAIN (Push-Pull Switch):
Multiplies channel 1 vertical gain
five times; for example, 5 mV/div
sensitivity becomes 1 mV/div sensi-
tivity.
$ CHI POSITION, ¢ CH3 Control.
$ CHI Postron:
Rotation adjusts vertical position
of channel 1 trace, In X-¥ oper:
tion, rotation adjusts vertical posi
tion of display,
4 CHB:
Rotation adjusts vertical position
of channel 3 trace.
UNCAL Indicator. Lights when channel
1 VARIABLE control is not set to the
CAL position.
PULL X5 GAIN Indicator. Lights when
the channel 1 VARIABLE control is
pulled out (X5 GAIN mode).
AC-GND-DC Switch. — Three~position
lever switch which operates as follows:
AC:
Channel 1 input signal is capaci-
tively coupled; de component is
blocked.
GND:
Opens signal path and grounds input
to vertical amplifier. This provides
11, UNCAL Indicator.
12. PULL X5 GAIN Indicator,
a zero-volt base line, the position
of which can be used as a reference
when performing dc measurements.
De:
Direct coupling of channel 1 input
signal; both ac and dc component of
signal produce vertical deflection.
7. INPUT Jack. Vertical input for chan-
nel 1. ¥ axis input for X-Y operation.
CHANNEL 2 CONTROLS
8 VOLTS/DIV Control, Vertical attenuator
for channel 2. Provides step adjustment
of vertical sensitivity. When channel 2
VARIABLE control is set to CAL, verti-
cal sensitivity is calibrated in 10 steps
from 5 mV/div to 5 V/div in a 1-2-5 se~
quence. Sensitivity is five times greater
when X5 GAIN is selected. In X-Y opera~
tion, this control provides step adjust-
ment of X-axis sensitivity.
9. VARIABLE, PULL X5 GAIN ControL
VARIABLE:
Rotation provides vernier adjust~
ment of channel 2 vertical sensi-
tivity. In the fully clockwise (CAL)
position, the attenuator is calibrat-
ed. In X-¥ operation this control
becomes the vernier X-axis gain
control.
PULL X5 GAIN (Push-Pull Switch):
Multiplies channel 2 gain five
times; for example, 5 mV/div sensi-
tivity becomes 1 mV/div sensiti-
vity.
10, $ CH2 POSITION, x-¥ @ , $cHE,
Control.
4 Clz POSITION:
Rotation adjusts vertical position
of channel 2 trace. In X-Y opera-
tion, rotation adjusts horizontal
position of display (X-¥-<>),
CHa:
Rotation adjusts vertical position
of channel 4 trace.
o
Lights when channel
2 VARIABLE control is not set to the
CAL position.
Lights whenCONTROLS AND INDICATORS
B.
4.
15.
the channel 2 VARIABLE control is
pulled out (X5 GAIN mode).
AC-GND-DC Switch, _ Three-position
lever switch which operates as follows:
AC:
Channel 2 input signal is capaci~
tively coupled; de component is
blocked.
GND:
Opens signal path and grounds input
to vertical amplifier. This provides
a zero-volt base line, the position
‘of which can be used as a reference
when performing de measurements.
Dc:
Direct coupling of channel 2 input
signal; both ac and de component of
signal produce vertical deflection.
INPUT Jack. Vertical input for channel
2 in normal sweep operation. X-axis in
put in X-¥ operation.
‘Vertical MODE Switch Assembly
NOTE
‘The vertical mode settings are also re~
lated to horizontal mode and trigger
source selection. Refer to HORIZ DIS~
PLAY and SOURCE controls for a de~
scription of this relationship.
The vertical MODE switch assembly
contains nine soft-touch pushbuttons.
Each pushbutton is internally illuminated
to show which mode is presently se-
lected.
‘The CH 1, CH 2, DUAL, ADD, and
QUAD buttons are electonically inter~
locked; i.e., selection of a new mode
automatically releases the previous
selection.
(CH 1 Pushbutton:
Only the input signal to channel 1 is
displayed on the CRT.
CH 2 Pushbutton:
Only the input signal to channel 2 is
displayed on the CRT.
DUAL Pushbutton:
‘The input signal to channel 1 and
channel 2 are displayed simulta~
neously on two traces.
ADD Pushbutton:
‘The input from channel 1 and chan~
16.
aw
18.
nel 2 are summed and displayed as
a single signal. When the CH 2 INV
function is activated, the input
from channel 2 is subtracted from
channel 1 and the difference is dis~
played as a single signal.
QUAD Pushbutton:
‘The inputs to channels 1, 2, 3, and 4
are displayed simultaneously on
four traces.
‘The ALT and CHOP buttons are e-
lectronically interlocked; i.e, selec
tion of CHOP cancels ALT, and vice
versa, The ALT and CHOP functions
are enabled only in the DUAL and
QUAD modes.
ALT Pushbutton:
Selects alternate method of dual or
quad-trace display.
CHOP Pushbutton:
Selects chop method of dual or
quad-trace display.
‘The 20 MHz BW and CH 2 INV but-
tons are independently latching; i.e.,
press once to activate function, press
again to turn off function,
(CH 2 INV Pushbutton:
Selects inverted or normal polarity
of channel 2 signal; inverted polar-
ity when activated, normal polarity
when off.
20 MHz BW Pushbutton:
Limits maximum bandwidth to 20
MHz when activated, full band~
width when off.
BEAM FIND Pushbutton. Brings the
beam to within the graticule area, op-
erates independently from other display
controls.
A INTENSITY, B INTENSITY Control.
A INTENSITY:
‘Adjusts brightness of trace gener~
ated by A sweep.
B INTENSITY:
‘Adjusts brightness of trace gener-
ated by B sweep.
SCALE ILLUM/POWER Control. Fully
counterclockwise rotation of this control
(OFF position) turns off oscilloscope.
Clockwise rotation turns on oscilloscope.19.
20.
a.
25.
Further clockwise rotation of the control
increases the illumination of the grati-
cule scale.
POWER Indicator. LED lights when os-
cilloscope is turned on.
ASTIG Control Astigmatism adjustment
provides optimum spot roundness when
used in conjuction with FOCUS control.
To adjust this control a screwdriver must
be used (once adjusted, this control does
not require frequent readjustment).
FOCUS Control An auto-focus circuit
automatically readjusts focus as intensi-
ty is varied. Once adjusted, this control
does not require frequent readjustment.
22.
23.
CONTROLS AND INDICATORS
TRACE ROTATION Control. Use screw-
driver to adjust trace to horizontal posi-
tion, Once adjusted, this control does
not require frequent readjustment.
CAL 1 V pp =1 kHz Terminal. Provides
1 kHz, I-volt peak-to-peak square wave
signal. This is useful for probe compen-
sation adjustment and a general check of
oscilloscope calibration accuracy. High
enough amplitude to check CH3 and
CHA.
-L Terminal/Binding Post. Oscilloscope
Chassis ground, and earth ground via
3rwire ac power cord.
HORIZONTAL CONTROLS
HORIZ DISPLAY Switch Assembly.
‘The HORIZ DISPLAY switch assembly
contains five soft-touch pushbuttons
which are used to select the desired hor
izontal (sweep) mode, Each pushbutton
is internally illuminated to show which
mode is presently selected. All pushbut-
tons are electronically interlocked i.e.
selection of a new mode automatically
releases the previous selection.
The ALT, A-INT-B, and B DLY'D modes
are related to the DELAY TIME MULT
and B SOURCE controls. Refer to those
controls for a description of that rela~
tionship.
A Pushbutton:
Only the A sweep is operative, with
the B sweep dormant.
ALT Pushbutton:
‘A sweep (with B sweep super-
imposed as an intensified portion)
alternates with B sweep.
A-INT-B Pushbutton:
Duration of B sweep appears as an
intensified portion on the A sweep.
3B DLY'D Pushbutton:
Only the delayed B sweep is opera~
tive.
DUAL Pushbutton:
‘A sweep and B sweep operate inde-
pendently. The A sweep is trig
gered by the A trigger source and
the B sweep is triggered by the B
trigger source.
XY Pushbutton:
Selects X-Y mode. Channel 1 be~
comes the Y-axis and channel 2 be~
comes the X~ A sweep and B
sweep are disabled. Settings of
vertical MODE and TRIG MODE
have no effect.
SWEEP CONTROLS
26.
The A SWEEP TIME/DIV and B SWEEP
TIME/DIV controls are mechanically con
structed so that it is impossible for the B
sweep to operate at a slower sweep rate
than the A sweep. There is no vernier
adjustment for the B sweep.
A SWEEP TIME/DIV, B SWEEP TIME/DIV
Control.
‘A SWEEP TIME/DIV:
Sweep time selector for A sweep.
Provides step selection of sweep
rate. When the A VARIABLE con-
trol is set to CAL, sweep rate is
calibrated. This control has 23
steps of .02 1s/div to .5 s/div, in a
1-2-5 sequence.
B SWEEP TIME/DIV:
Sweep time selector for B sweep.
Provides step selection of sweep
rate. This control is calibrated in
20 steps from 02 Is/div to
50 ms/div, each in a 1-2-5 se~
quence. In A-INT-B and ALT modes
of HORIZ DISPLAY, the durationCONTROLS AND INDICATORS
2.
28.
29.
of the intensified portion is depend-
ent upon the B SWEEP TIME/DIV
control setting. There is no fine
adjustment for the B sweep.
A VARIABLE, PULL CHOP F SELECT
Controls.
‘A VARIABLE CONTROL:
Rotation of control is vernier ad-
justment for A sweep. In fully
clockwise (CAL) position, sweep
rate is calibrated.
PULL CHOP F SELECT Control:
‘When pulled outward the chopping
frequency is changed.
UNCAL Indicator. Lights when A VARI-
ABLE control is not set to the CAL posi-
tion.
DELAY TIME MULT Control. Adjusts
start time of B sweep to some delay time
after the start of the A sweep. Delay
time may be set from 0.2 to 10 times the
setting of the A SWEEP TIME/DIV con-
trol. Calibrated control rotates almost
10 turns lock-to-lock, Number of full
revolutions (0 to 10) appears in window
of control. Fractional revolutions (100
parts per revolution) appears on skirt of
control, Indicated sum of full and frac~
tional revolutions is multiplication fac~
tor. Lock-unlock lever control to pre
vent accidental changing of setting.
NOTE
This control is enabled only in ALT, A~
INT-B, and B DLY'D modes of HORIZ
DISPLAY jin either "STARTS AFTER
DELAY* or “TRIGGERABLE AFTER
DELAY" operation. Refer to the B
SOURCE control for a further descrip-
tion of that relationship.
30.
al.
$ TRACE SePsration, HOLDOFF Con
tol.
4 TRACE SEP:
‘Adjusts vertical separation between
A sweep and B sweep traces.
Clockwise rotation increases sep-
aration (B sweep moves down with
respect to A sweep).
‘The $ TRACE SEP control is enabled
only in the ALT mode of HORIZ DIS-
PLAY.
HOLDOFF:
Rotation adjusts holdoff time (trig~
ger inhibit period beyond sweep
duration), Full clockwise rotation
is NORM, Holdoff period increases
progressively with counterclock-
wise rotation to maximum of about
5 times normal. Extreme counter-
clockwise rotation turns on switch
(B ENDS A) which causes A sweep
to terminate at end of B sweep. B
ENDS A operation is enabled only
in the ALT and A-INT-B modes of
HORIZ DISPLAY.
<> POSITION, FINE, PULL 10X MAG
Control
> POSITIO!
Coarse horizontal position control.
INE, PULL 10X MAG:
Rotation of control is fine hori-
zontal position adjustment. Push-
pull switch selects ten times sweep
magnification (PULL 10X MAG)
when pulled out, normal when
pushed in. Expands both A sweep
and B sweep, increases maximum
sweep rate to 2 ns/div.
TRIGGERING CONTROLS
A SWEEP TRIGGERING CONTROLS
32.
SOURCE Switch. Selects source of trig-
ger for A sweep. Lever switch with the
following positions:
‘V. MOD!
Triggering source for A sweep is
selected by vertical MODE switch.
Permits each waveform viewed to
become its own trigger signal. The
10
relationship to the various vertical
modes is as follows:
CH 1:
‘Channel 1 signal becomes the
trigger source.
cH 2:
Channel 2 signal becomes the
trigger source.
ADD:
‘The algebraic sum of CH 1 +CH 2 signals becomes the
trigger source (CH 1 - CH 2 if
channel 2 is inverted).
DUAL (ALT):
CH 1 and CH 2 signals alter-
nate as the trigger source.
Triggering is impossible unless
input signals are applied to
both CHI and CH 2.
QUAD (ALT):
CH 1 thru CH 4 signals alter
nate as the trigger source.
‘Triggering is impossible unless
input signals are applied to all
four channels.
DUAL of QUAD (CHOP):
The display cannot be syn-
chronized, since the chopping
signal becomes the trigger
CHI:
Channel 1 signal becomes A sweep
trigger, regardless of vertical
mode.
CH2:
‘Channel 2 signal becomes A sweep
trigger, regardless of vertical
mode.
1/1 EXT CH 3:
Signal from CH 3 or A EXT TRIG
jack becomes A sweep trigger.
1/i0 EXT CH 3:
Signal from CH 3 or A EXT TRIG
jack becomes A sweep trigger at-
tenuated by a factor of 10.
LINE:
Signal derived from the input line
voltage (50/60 Hz) becomes trigger.
33. COUPLING Switch. Selects trigger
coupling for A sweep. Five-position
lever switch with the following positions:
AC:
‘Trigger is capacitively coupled; de
component is blocked.
LF Rej
Trigger is coupled thru high-pass
filter. Frequencies below 30 kHz
are attenuated,
HF Rej:
Trigger is coupled thru low-pass
filter. Frequencies above 30 kHz
are attenuated,
CONTROLS AND INDICATORS
uw
Trigger is direct coupled, and in-
cludes both the ac and de com-
ponent. Use for signals below
20 Hz. For CH 3 and CH 4, the
vertical POSITION controls have no
effect on the trigger point.
VIDEO:
‘Trigger is routed to video sync sep-
arator. Horizontal sync pulses are
selected as trigger at sweep rates
of 50 18/div and faster (LINE). Ver-
tical sync pulses are selected as
trigger at sweep rates of .1 ms/div
and slower (FRAME),
34. SLOPE, LEVEL, PULL FIX Controls.
SLOPE Switch:
‘Two-position rotary switch. Coun~
terclockwise position selects neg-
ative-going (-) slope and clockwise
position selects positive-going (+)
slope as triggering point for A
sweep.
LEVEL Control:
Trigger level adjustment, deter-
mines point on triggering waveform
where A sweep is triggered. Rota~
tion in () direction selects more
negative point of triggering, and
rotation in (+) direction selects
more positive point of triggering.
PULL FIX Control:
When pulled, the triggering level is
automatically fixed at the average
level of the triggering waveform,
regardless of the setting of the
LEVEL control. When pushed in,
normal variable triggering is ena~
bled.
35, ‘TRIG'D Indicator. Lights when A sweep
is triggered, Shows when trigger level is
properly set.
36, CH3 or A EXT TRIG Jack.
afl:
Vertical input for channel 3 in
quad-trace operation. External
trigger input for A sweep or B
sweep in single and dual-trace
operation. Signal is unattenuated.
alo:
Vertical input for channel 3 in
quad-trace operation. ExternalCONTROLS AND INDICATORS
trigger input for A sweep in single
and dual-trace operation, Signal is
attenuated 10:1.
B SWEEP TRIGGERING CONTROLS
37. SOURCE Switch. Selects the B sweep
trigger source. Four-position lever
switch has the following positions:
CHI:
‘Channel 1 signal becomes B sweep
trigger.
CHa:
‘Channel 2 signal becomes B sweep
trigger.
1/1 EXT CH 4:
Selects signal from CH 4 or B EXT
TRIG jack as trigger for B sweep.
1/10 EXT CH 4:
Selects signal from CH 4 or B EXT
TRIG jack as trigger for B sweep.
Signal is attenuated 1
COUPLING Switch, Selects trigger
coupling for B sweep. Four-position
lever switch with the following positions:
‘AC:
‘Trigger is capacitively coupled; de
component is blocked.
LE Re}
Trigger is coupled thru high-pass
filter. Frequencies below 30 kiiz
are attenuated.
HF Rej:
Trigger is coupled thru low-pass
filter. Frequencies above 30 kiiz
are attenuated.
38.
Trigger is direct coupled, and in-
cludes both the ac and de com-
ponent, Use for signals below
20 Hz. For CH 3 and CH 4, the
vertical POSITION controls have no
effect on the trigger point.
SLOPE, LEVEL, PULL STARTS AFTER
DELAY Control,
SLOPE Switch:
‘Two-position rotary switch. Coun-
terclockwise position selects nega~
tive-going () slope and clockwise
position selects positive-going (#)
slope as triggering point for A
sweep.
39.
12
40.
4.
42.
LEVEL Control:
Trigger threshold level adjust-
ment. Determines point on trig
gering waveform where B sweep is
triggered. Rotation in () direction
selects more negative point of trig-
gering, and rotation in (4) direction
selects more positive point of tr
gering. Control has no effect in
"STARTS AFTER DELAY" opera-
tion.
PULL STARTS AFTER DELAY:
Selects "STARTS AFTER DELAY*
operation when control is pulled
out. The B sweep is triggered im-
mediately after the delay set by
the DELAY TIME MULT control
the B SLOPE and B LEVEL controls
have no effect.
NOTE
When the control is pushed in, the
scope is in the “TRIGGERABLE
AFTER DELAY" mode. B sweep trig-
gering is inhibited during the delay
period (set by the DELAY TIME MULT
control); triggering occurs upon the
first trigger signal after the delay pe-
riod. Trigger signal conditions are
established by the B SOUCE, B
SLOPE, and B LEVEL controls. In the
AUTO trigger mode, the unit will
automatically revert to "STARTS
AFTER DELAY" operation in the
absence of a proper trigger signal.
TRIG'D Indicator. Lights when B sweep
is triggered. Shows when trigger level is
properly set. This indicator does not
light in “STARTS AFTER DELAY™ oper-
ation.
CH 4 or EXT TRIG Jack. Vertical input
for channel 4 in quad-trace operation.
External trigger input for B sweep in
single and dual-trace operation,
‘TRIG MODE Switch Assembly.
The TRIG MODE switch assembly con-
tains four soft-touch pushbuttons which
are used to select the desired triggering
mode. The AUTO, NORM, and SINGLE
buttons are electronically interlocked so
that only one of the three may be se-
lected. Selecting a new trigger mode43.
45.
46.
41.
releases the previous selection. The face
of each pushbutton is internally illum
inated to show which triggering mode is
presently selected, The RESET button is
a momentary action switch used only in
conjunction with the SINGLE mode.
AUTO Pushbutton:
Selects automatic _ triggering
mode. Generates sweep (free runs)
in absence of trigger; automatically
reverts to triggered sweep opera~
tion when adequate trigger signal is
present. Auto triggering is appli-
cable to both the A sweep and B
sweep.
NORM Pushbutton:
Selects normal triggered sweep op-
eration, which generates a sweep
only when adequate trigger signal is
present, When scope is set up for
multiple trigger sources, all trigger
CONTROLS AND INDICATORS
sources must be present to obtain a
sweep,
SINGLE Pushbutton:
Selects single sweep operation. For
dual or quad trace display, single
sweep operation requires CHOP
method of display; do not select
ALT. In single sweep operation,
simultaneous observation of the A
sweep and B sweep is not possible.
RESET Pushbutton:
Enables single sweep operation. A
single sweep will begin when the
next sync trigger occurs, After a
single sweep, RESET must be
pressed again to enable next single
sweep. Indicator (face of switch)
lights to show that single sweep
operation is ready for triggering.
Lights when RESET switch is oper-
ated and remains lit until one
sweep is completed.
REAR PANEL CONTROLS
CAL Loop. Current probe calibration
loop provides a 1 kHz, 10 mA p-p square
wave. Arrow indicates direction of posi~
tive polarity.
A GATE Jack. Output terminal where
gate waveform coincident with A sweep
fs available. Gate waveform is approxi~
mately 1.5 volt positive with 0 volt ref-
erence, output impedance is approxi
mately 500 ohms.
B GATE Jack. Output terminal where
gate waveform coincident with B sweep
is available. Gate waveform is approxi-
mately 1.5 volt positive with 0 volt ref-
erence, output impedance is approxi-
mately 500 ohms.
CH 1 OUT Jack, Output terminal where
sample of channel 1 signal is available.
Amplitude is 50 millivolts per division of
vertical deflection seen on CRT when
terminated into 50 chms, Output impe-
dance is 50 ohms.
% AXIS INPUT Jack, Input jack for
intensity modulation of CRT ‘electron
beam. ‘TTL compatible (5 volts p-p sen-
sitivity). Negative-going signal increases
intensity.
1B
48.
49.
50.
51.
Power Cord Receptacle,
Fuse Holder.
‘Handle/Tilt Stand (aot shown).
Feet/Cord Wrap.CONTROLS AND INDICATORS
Fig. 2. Rear Panel Controls.
rrOPERATING INSTRUCTIONS
SAFETY PRECAUTIONS
The following precautions must be ob~
served to prevent electric shock.
1
Py
When the oscilloscope is used to make
measurements in equipment that con
tains high voltage, there is always a cer
tain amount of danger from electrical
shock. The person using the oscilloscope
in such conditions should be a qualified
electronics technician or otherwise
trained and qualified to work in such cir
cumstances. Observe the TEST IN-
STRUMENT SAFETY recommendations
listed on the inside front cover of this
manual.
Do not operate this oscilloscope with the
case removed unless you are a qualified
service technician. High voltage up to
20,000 volts is present when the unit is
operating with the case removed.
The ground wire of the 3-wire ac power
plug places the chassis and housing of the
oscilloscope at earth ground. Use only a
3-wire outlet, and do not attempt to de-
feat the ground wire connection or float
the oscilloscope; to do so may pose a
great safety hazard,
Special precautions are required to
measure or observe line voltage wave-
forms with any oscilloscope. Use the
following procedure:
a, Do not connect the ground clip of
the probe to either side of the line.
The clip is already at earth ground
and touching it to the hot side of
the line may “weld” or “disinte-
grate” the probe tip and cause pos-
sible injury, plus possible damage to
the scope or probe.
Insert the probe tip into one side of
the line voltage receptacle, then
the other, One side of the recep-
tacle should be "hot" and produce
the waveform, The other side of
15
the receptacle is the ac return and
no waveform should result.
EQUIPMENT PROTECTION PRECAUTIONS
The following precautions will help avoid
damage to the oscilloscope.
L
2
4,
Never allow a small spot of high bril-
liance to remain stationary on the screen
for more than afew seconds. The screen
may become permanently burned. A spot
will occur when the scope is set up for
X-¥ operation and no signal is applied.
Either reduce the intensity so the spot is
barely visible, apply signal, or switch
back to normal sweep operation, It is
also advisable to use low intensity with
AUTO triggering and no signal applied
for long periods. A high intensity trace
at the same position could cause a line to
become permanantly burned onto the
screen.
Do not rest objects on top of the oscillo~
scope or otherwise obstruct the ventilat~
ing holes in the case, as this will increase
the internal temperature.
Excessive voltage applied to the input
jacks may damage the oscilloscope. The
maximum ratings of the inputs are as
follows:
CH 1 and CH 2:
800 V p-p; 400 V de + ac peak.
EXT TRIG:
400 V de + ac peak,
'Z AXIS INPUT:
50 volts de + ac peak.
Never apply external voltage to oscillo~
scope output jacks,
Always connect a cable from the ground
terminal of the oscilloscope to the chas-
sis of the equipment under test. Without
this precaution, the entire current for
the equipment under test may be drawn
through the probe clip leads under cer
tain circumstances. Such conditionsOPERATING INSTRUCTIONS
could also pose a safety hazard, which
the ground cable will prevent.
5. The probe ground clips are at oscillo-
scope ground and should be connected
only to the common of the equipment
under test. To measure with respect to
any point other than the common, use
CH 1 - CH 2 subtract operation (ADD
mode and CH 2 INV), with the channel 1
probe to the point of measurement and
the channel 2 probe to the point of ref
erence. Use this method even if the ref-
erence point is a dc voltage with no sig~
nal,
OPERATING TIPS
The following recommendations will help
obtain the best performance from the oscillo-
seope.
1, Always use the probe ground clips for
best results, attached to a circuit ground
point near the point of measurement. Do
not rely solely on an external ground
wire in lieu of the probe ground clips as
undesired signals may be induced,
2. Avoid the following operating conditions:
a. Direct sunlight.
3
4.
5
>, High temperature and humidity.
c. Mechanical vibration,
d. Electrical noise and strong magnet~
ic fields, such as near large motors,
power supplies, transformers, etc.
Occasionally check trace rotation, probe
compensation, astigmatism, and cali-
bration accuracy of the oscilloscope
using the procedures found in the MAIN-
TENANCE section of this manual.
Terminate the output of a signal gener-
ator in its characteristic impedance to
minimize ringing, especially if the signal
has fast edges such as square waves or
pulses. For example, the typical 50 %
output of a square wave generator should
be terminated into an external 50 2 ter-
minating resistor and connected to the
oscilloscope with 50 coaxial cable.
Probe compensation adjustment matches
the probe to the input of the scope. For
best results, compensation should be ad~
justed initially, then the same probe al~
ways used with the same channel. Probe
compensation should be readjusted when
a probe from a different oscilloscope is
used,
CAL PUSH.
5 Vidi
ae.
CAL Pus
8 vid
FULLGW NORM 1 mae
| cat ust
apartaany
POSTION
‘CENTER pusi
‘AG CENTER CENTER
Fig. 3. Initial Control Settings.
16INITIAL STARTING PROCEDURE
Until you familiarize yourself with the use
of all controls, the settings shown in Fig. 3
may be used as a reference point to obtain a
trace on the CRT in preparation for waveform
observation.
1, Tum the POWER control clockwise; the
unit will be turned on and the pilot light
will be illuminated. This control also
adjusts the scale illumination. When the
control is turned fully clockwise, the
scale is illuminated at full brightness.
2. When power is turned on, the HORIZ
DISPLAY, TRIG MODE, ‘and vertical
MODE controls should be set as follows:
vertical MODE: CHI
HORIZ DISPLAY: A
TRIG MODE: AUTO
3. A trace should appear on the CRT. Ad-
just the trace brightness with the A IN-
TENSITY control, and the trace sharp-
ness with the FOCUS control.
SINGLE TRACE DISPLAY
Either channel 1 or channel 2 may be used
for single-trace operation. The advantage of
using channel 2 is that the waveform on the
display can be inverted if desired with the
CHZ INV switch.
1, Perform the steps of the "Initial Starting
Procedure" with the MODE switch set to
CH2.
2, Connect the probe to the CH 2 INPUT
jack,
3. Connect the probe ground clip to the
chassis or common of the equipment
under test. Connect the probe tip to the
point of measurement.
4. If no waveforms appear, increase the
sensitivity by turning the CH 2 VOLTS/
DIV control clockwise to a position that
gives 2 to 6 divisions vertical deflection,
5. The display on the CRT may be unsyn-
chronized. Refer to the "Triggering"
paragraphs in this section for procedures
on. setting triggering and sweep time
controls to obtain a stable display show-
ing the desired number of waveforms.
VW
OPERATING INSTRUCTIONS
DUAL TRACE DISPLAY
In observing simultaneous waveforms on
channel 1 and 2, the waveforms are usually
related in frequency, or one of the waveforms
is synchronized to the other, although the
basic frequencies are different. If the two
waveforms have no phase or frequency rela~
tionship, there is seldom reason to observe
both waveforms simultaneously. However,
with V. MODE triggering, two waveforms not
related in frequency or period can be simul~
taneously viewed.
1, Connect probes to both CH 1 and CH 2
INPUT jacks.
2. Connect the ground clips of the probes to
the chassis or common of the equipment
under test. Connect the tips of the
Probes to the two points in the circuit
where waveforms are to be measured.
3. With the ADD button engaged, the alge-
braic sum of CH 1 + CH 2 is displayed as
a single trace. When the CH 2 INV
button is also engaged, the algebraic dif-
ference of CH 1 = CH 2 is displayed.
To view both waveforms simultaneously,
press the DUAL button on the vertical
MODE switch assembly.
5. In the ALT mode, one sweep displays the
channel 1 signal and the next sweep dis-
plays the channel 2 signal in an alternat~
ing sequence. Alternate sweep is norm-
ally used for viewing high-frequency or
high-speed waveforms at sweep times of
1 ms/div and faster, but may be selected
at any sweep time.
6, In the CHOP mode, the sweep is chopped
at an approximate 250 kHz rate. and
switched between channel 1 and chan
nel 2. Chop sweep is normally used for
low-frequency or low-speed waveforms
at sweep times of 1 ms/div and slower.
a. If chop sweep is used at sweep
times of 0.2 ms/div and faster, the
chop rate becomes a significant
portion of the sweep and may be-
come visible in the displayed wave-
form, However, you may select
chop sweep at any sweep time for
special applications, For example,
the only way to observe simultan-OPERATING INSTRUCTIONS
cous events on a dual-trace scope
at any sweep rate is with chop
sweep.
b, Note that synchronization of the
display is not possible in the CHOP
mode of operation with the trigger
SOURCE switch set to V. MODE,
because the trigger source becomes
the chopping signal itself. Use ALT
mode instead, or select a trigger
SOURCE of CH 1 or CH 2.
A. push-pull switch is provided
(PULL CHOP F SELECT) to allow a
choice of two chop frequencies. In
the normal (pushed in) position,
chop frequency is approximately
250 kHz. When pulled, the chop
frequency is slightly shifted. This
permits better viewing when the
signal being observed approaches
the chop frequency or a sub
multiple.
7. Adjust the CH 1 and CH 2 POSITION
controls to place the channel 1 trace
above the channel 2 trace.
8. Set the CH 1 and CH 2 VOLTS/DIV con-
trols to a position that gives 2 to 3 divi
sions of vertical deflection for each
trace. If the display on the screen is un~
synchronized, refer to the "Triggering"
paragraphs in this section of the manual
for procedures for setting triggering and
sweep time controls to obtain a stable
display showing the desired number of
waveforms.
QUAD TRACE DISPLAY
Four waveforms may be viewed simultane-
ously using the quad-trace feature of the
1590A oscilloscope. The channel 3 and chan-
nel 4 INPUT signals are de coupled and have a
sensitivity of 0.1V/div when 1/1 attenuation is
selected. Thus the most desirable input level
is 0.1 to 0.2 volts p-p, with a dc component of
0.5 volts or less, When 10:1 probes are used, 1
to 2 volts p-p is appropriate, with a de compo-
nent of up to 5 volts (suitable for TTL level
signals). When 1/10 attenuation is selected,
the sensitivity is increased by a factor of 10,
also increasing the most desirable input level
by a factor of 10. Using 1/10 attenuation
18
without 10:1 probes is also suitable for TTL
level signals, External blocking capacitors or
attenuators may be used with the channel 3 or
channel 4 inputs.
1, Press the QUAD button on the vertical
MODE switch assembly. Connect probes
to CH 1, CH 2, CH 3, and CH 4 INPUT
jacks.
2. Connect the ground clips of the probes to
the chassis or common of the equipment
under test. Connect the tips of the
probes to the four points in the circuit
where waveforms are to be measured.
3. In the ALT mode, one sweep displays the
channel 1 signal, the second sweep dis-
plays the channel 2 signal, the third
sweep displays the channel 3 signal, and
the fourth sweep displays the channel 4
signal in an alternating sequence. Alt
nate sweep is normally used for viewing
high-frequency or high-speed waveforms
at sweep times of 1 me/div and faster,
bat may be selected at any sweep time.
4, In the CHOP mode, the sweep is chopped
at an approximate 250 kHz rate and
switched between channel 1, channel 2,
channel 3, and channel 4. Chop sweep is
normally used for low-frequency or low
speed waveforms at sweep times of
1 ms/div and slower.
a, If chop sweep is used at sweep
times of 0.2 ms/div and faster, the
chop rate becomes a significant
portion of the sweep and may be-
come visible in the displayed wave~
form, However, you may select
chop sweep at any sweep time for
special applications. For example,
the only way to observe simultan-
ous events on a quad-trace scope
at any sweep rate is with chop
sweep.
b, Note that synchronization of the
display is not possible in the CHOP
mode of operation with the trigger
SOURCE switch set to V. MODE,
because the trigger source becomes
the chopping signal itself. Use ALT
mode instead, or select a trigger
SOURCE of CH 1, CH 2, or CH 3.
neem ntti© A push-pull switch is provided
(PULL CHOP F SELECT) to allow a
choice of two chop frequencies. In
normal (pushed in) position, the
chop frequency is approximately
250 kHz, When pulled, the chop
frequency is slightly shifted, This
permits better viewing when the
signal being observed approaches
the chop frequency or a sub
multiple.
5. Adjust the four vertical $ POSITION
controls to place the channel 1 trace at
the top of the display, the channel 2
trace below the channel 1 trace, etc.
6, Set the CH 1 and CH 2 VOLTS/DIV con-
trols to a position that gives 1 to 2 divi-
sions of vertical deflection for each
trace. If the display on the screen is un-
synchronized, refer to the "Triggering"
Paragraphs in this section of the manual
for procedures for setting triggering and
sweep time controls to obtain a stable
display showing the desired number of
waveforms.
TRIGGERING
‘The Model 1590A Oscilloscope provides ver
satility in sync triggering for ability to obtain
a stable, jitter-free display in single-trace,
dual-trace, or quad-trace operation. The
proper settings depend upon the type of wave-
forms being observed and the type of meas-
urement desired. An explanation of the vari-
ous controls which affect synchronization is
given to help you select the proper setting
over a wide range of conditions.
‘TRIG MODE Switch
1, The NORM position provides normal
triggered sweep operation, The sweep
remains at rest until the selected trigger
source signal crosses the threshold level
set by the LEVEL control, The trigger
causes one sweep to be generated, after
which the sweep again remains at rest
until triggered. In the NORM position,
there will be no trace unless an adequate
trigger signal is present. In the ALT
mode of dual trace operation with the
SOURCE switch set to V. MODE, there
will be no trace unless both channel 1
19
OPERATING INSTRUCTIONS
and channel 2 signals are adequate for
triggering (for quad trace operation, all
four signals must be adequate). Typi-
cally, signals that produce even 1/2 divi-
sion of vertical deflection are adequate
for normal triggered sweep operation.
2. In the AUTO position, automatic sweep
operation is selected, In automatic
sweep operation, the sweep generator
free runs to generate a sweep without a
igger signal. However, it automat
ically switches to triggered sweep opera~
tion if an acceptable trigger source sig-
nal is present. The AUTO position is
handy when first setting up the scope to
observe a waveform; it provides sweep
for waveform observation until other
controls can be properly set. Once the
controls are set, operation is often
switched back to'the NORM triggering
mode, since it is more sensitive. Use the
TRIGID indicator to set the trigger
LEVEL in the AUTO mode, then switch
to the NORM mode. Automatic sweep
must be used for de measurements and
signals of such low amplitude that they
will not trigger the sweep.
NOTE
In the X-¥ mode, the sweep generator
and triggering circuits are disconnected
and have no effect.
SOURCE Switch
The A SOURCE switch (V. MODE, CH 1,
CH 2, 1/1 and 1/10 EXT/CH 3, and LINE po-
sitions) selects the signal to be used as the
sync trigger for the A sweep. For explanation
of the B SOURCE switch, refer to the "DE-
LAYED SWEEP OPERATION" paragraph later
in this manual.
1, When the V. MODE position is selected,
the trigger source is dependent upon the
vertical MODE selection, In this man~
ner, each waveform being observed be-
comes its own trigger signal.
a, When the vertical mode is changed
from CH 1 to CH 2, the trigger
source is also changed from CH 1
to CH 2, and vice versa. This is
very convenient for single trace
operation,OPERATING INSTRUCTIONS
b,
When the ALT dual-trace vertical
mode is selected, the trigger source
alternates between CH 1 and CH 2
with each sweep (for ALT quad-
trace vertical mode, the trigger
source alternates between CH 1,
CH 2, CH 3, and CH 4), This is
convenient for checking ampli
tudes, waveshape, or waveform
period measurements, and even
permits simultaneous’ observation
of two waveforms which are not
related in frequency or period.
However, this setting is not suita-
ble for phase or timing comparison
measurements. For such measure
ments, both traces must be trig-
gered by the same sync signal.
When the CHOP dual-trace or
quad-trace vertical mode is se~
lected, synchronization of the dis~
Play is not possible because the
chopping signal becomes the trig-
ger. Use the ALT mode instead, or
change the SOURCE switch setting
to CH 1, CH 2, or CH 3.
If the SOURCE switch is set to CH 1 (or
CH 2, or CH 3), the channel 1 (or channel
2, or channel 3) signal becomes the trig-
ger source regardless of the vertical
MODE selection, CH 1, CH 2, or CH 3 is
Often used as the trigger source for
Phase or timing comparison measure-
ments.
3. If the A SOURCE switch is set to the
LINE position, triggering is derived from
the input line voltage (50/60 Hz). This is
useful for measurements that are related
to line frequency.
If the A SOURCE switch is set to the 1/1
or 1/10 EXT/CH 3 position, the signal
applied to the A EXT TRIG/CH 3 jack
becomes the trigger source. This signal
must have a timing relationship to the
displayed waveforms for a synchronized
display.
LEVEL/SLOPE Control
(Refer to Fig. 4)
2
A sweep trigger is developed when the trig-
ger source signal crosses a preset threshold
20
atenenemtneeaneneanensenets
level. Rotation of the LEVEL control varies
the threshold level. In the + direction, the
triggering threshold shifts to a more positive
value, and in the - direction, the triggering
threshold shifts to a more negative value.
When the control is centered, the threshold
level is set at the approximate average of the
signal used as the triggering source. Proper
adjustment of this control usually synchronizes
the display.
‘The LEVEL control adjusts the start of the
sweep to almost any desired point on a wave-
form. On sine wave signals, the phase at
which sweep begins is variable. Note that if
the LEVEL control is rotated toward its ex-
treme + or ~ setting, no sweep will be devel-
oped in the NORM trigger mode because the
triggering threshold exceeds the peak ampli-
tude of the syne signal,
When the SLOPE control is set to the + po-
sition, the sweep is developed from the trigger
source waveform as it crosses a threshold
level in a positive-going direction. When the
SLOPE control is set to the ~ position, a sweep
trigger is developed from the trigger source
waveform as it crosses the threshold level in a
negative-going direction.
The A LEVEL/SLOPE control is used to set
the A sweep trigger and the B LEVEL/SLOPE
control is used to set B sweep trigger. The
TRIG'D indicators can be used to adjust the
trigger LEVEL controls to the proper settings.
SLOPE "=" RANGE
We
—
Fig. 4, Function of Slope and Level Controls.
FIX Contol
(Refer to Fig. 5 and Fig, 6)
When the PULL FIX control is in the pulled
out position, triggering always occurs at the
average level of the waveform, eliminating
the need to adjust the trigger level. As shown
in Fig, 5a and 5b, when the TRIG MODE is setto NORM and the triggering level is adjusted
to either side of the signal, the trigger point is
deviated as the input signal becomes small
which, in tum, stops the sweep operation. By
turning on the TRIG FIX, the triggering level
is automaically adjusted to the approximate
center of the waveform and the signal is syn-
chronized regardless of the position of the
LEVEL control as shown in Fig. 5c,
]
|
“Hom sre |
Fig. 5. Using FIX control for square wave
triggering.
When the input signal is suddenly changed
from a square wave to a pulse waveform, the
trigger point is shifted extremely toward the
side of the waveform unless the triggering
level is readjusted as shown in Fig. 6a. Also,
if the trigger point has been set to the "-" side
of the square wave, as in Fig. 6b(I) and the
input signal is changed to a pulse signal, the
trigger point is deviated and the sweep stops.
When this happens, pull out the FIX control
and the triggering will occur at the approxi-
mate center of the waveform, making it possi-
dle to observe a stabilized waveform (Fig.
6c). FIX is a little less sensitve then NORM
for low amplitude waveforms.
COUPLING Switch
1. Use the AC position for viewing most
types of waveforms. The trigger signal
is capacitively coupled and may be used
for all signals from 20 Hz to 100 MHz.
2. Use the LFrej position to eliminate low
a
OPERATING INSTRUCTIONS
frequency components for stable trig-
gering of high frequency signals.
4,
y)
0
fd “@)
a .
Wo ont ages
Ino ule evel. ™
-
(3) When the wigar eal i act
‘ound the came ofthe waveform
a
Tiger point ‘Sane val 98 (1
A \— vA
@ At h— PY.
wa O
(0) Wen he tigger level et to
‘ode of hw wavetor
Ie switch sotto "WORN
o | rigging without
ting tv
(e) Mot ait set 0 “FO
Fig. 6. Using FIX control for pulse
signal triggering.
Use the HFrej position to eliminate high
frequency components for stable trig~
gering of low frequency signals.
Use the DC position when you wish to
include the triggering effects of dc com-
ponents.
‘The VIDEO position is primarily for
viewing composite video waveforms. A
sync separator circuit separates sync
pulses from video. Horizontal syne
pulses are selected as trigger at sweepOPERATING INSTRUCTIONS
rates of 50 }s/div and faster (LINE) and
vertical sync pulses are selected as trig-
ger at sweep rates of 0.1 ms/div and
slower (FRAME), The triggering level is
automatically set so a sweep is gener
ated from high or low amplitude
signals, The LEVEL control has no
effect. Additional procedures for ob-
serving video waveforms are given later
in this section of the manual.
The B trigger coupling switch provides
the following positions: AC, LF Rej, HF
Rej, and DC. The functions of each po~
sition are identical to the corresponding
A trigger coupling positons.
SWEEP TIME Control
Set the SWEEP TIME/DIV control to display
the desired number of cycles of the waveform.
If there are too many cycles displayed for
good resolution, switch to a faster sweep
time. If only a line is displayed, try a slower
sweep time, When the sweep time is faster
than the waveform being observed, only part
of it will be displayed, which may appear as a
straight line for a square wave or pulse wave-
form.
HOLDOFF Control
(Refer to Fig. 7)
A “holdoff" period occurs immediately after
the completion of each sweep, and is a period
during which triggering of the next sweep is
inhibited. ‘The normal holdoff period varies
with the sweep rate, but is adequate to assure
complete retrace and stablization before the
next sweep trigger is permitted. The HOLD-
OFF control allows this period to be extended
by a variable amount if desired.
This control is usually set to the NORM po-
sition (fully clockwise) because no additional
holdoff period is necessary. The HOLDOFF
control is useful when a complex series of
pulses appear periodically, such as in Fig. 7A.
Improper sync may produce a double image as
in Fig. 7B. Such a display could be synchro-
nized with the sweep time VARIABLE control,
but this is impractical because time measure-
ments are then uncalibrated. An alternate
method of synchronizing the display is with
the HOLDOFF control. The sweep speed re-
mains the same, but the triggering of the next
22
sweep is "held off" by the duration selected by
the HOLDOFF control. Turn the HOLDOFF
control counterclockwise from the NORM po-
sition until the sweep starts at the same point
of the waveform each time. Maximum setting
{turned fully counterclockwise, but not beyond
the click stop) is 5 times greater than at the
NORM setting.
When the HOLDOFF control is turned fully
counterclockwise (beyond the click stop), it is
set in the B ENDS A mode, This ends the A
sweep at the point where the B sweep ends in
A-INT-B and ALT modes of HORIZ DIS-
PLAY. As a result, a shortened A sweep
waveform may appear on the screen, To view
the entire A sweep waveform rotate the
HOLDOFF control away from the B ENDS A
FULL
ULL
| JIU
A. HO. OFF ns
SJYUUU
Fig. 7, Use of HOLDOFF control.
MAGNIFIED SWEEP OPERATION
Since merely shortening the sweep time to
magnify a portion of an observed waveform
can result in the desired portion disappearing
off the screen, such magnified display should
be performed using MAGNIFIED SWEEP.
Using the horizontal <> POSITION control,
adjust the desired portion of waveform to the
center of the CRT. Pull out the PULL X10
MAG knob to magnify the display ten times.
For this type of display the sweep time is the
SWEEP TIME/DIV setting divided by 10.DELAYED SWEEP OPERATION
(Refer to Fig. 8 and Fig. 9)
Delayed sweep operation is achieved by use
of both the A sweep and the B sweep.
1, Press the A button on the HORIZ DIS~
PLAY switch assembly and adjust the
scope for a normal waveform display.
2, Pull the STARTS AFTER DELAY control
‘out. Press the A-INT-B button on the
HORIZ DISPLAY switch assembly and a
portion of the B sweep representing the
B SWEEP TIME/DIV setting will appear
a5 an intensified portion of the A
sweep. The B sweep intensity can be ad-
justed using the B INTENSITY control.
Delay time = DELAY TIME MULT
setting x A SWEEP TIME/DIV setting.
5.
[= yom
OPERATING INSTRUCTIONS
Press the B DLY'D button on the HORIZ
DISPLAY switch assembly to display the
A-INT-B intensified portion as a magni-
fied B DLY'D sweep.
Press the ALT button on the HORIZ
DISPLAY switch assembly to display
both the waveform with the intensified
portion and the magnified B DLY'D
sweep. Fig. 9 shows a typical display for
the ALT sweep mode.
(© swsop)
AAINT:S tensed 2008 to be maid
6
ovo
|____ wt
Fig. 9. ALT Sweep Display.
For STARTS AFTER DELAY operation,
apparent jitter increases as magnifica~
tion increases. To obtain a jitter free
display, switch to "Triggerable After De~
lay! operation by pushing in the STARTS
AFTER DELAY control and by setting
the B SOURCE switch to a position simi-
lar to that of the corresponding switch
for the A SOURCE, In this mode the
signal selected by the B SOURCE switch
becomes the B trigger source, making
use of the B trigger LEVEL control to set,
the trigger point. B SOURCE and LEV-
EL/SLOPE controls are set in a manner
similar to that of the corresponding con
trols for the A sweep and the B TRIG'D
indicator is enabled.
NOTE
Fig. 8. Delayed Sweep Display.
- Shift the intensified portion of the wave-
form (section to be magnified) along the
A sweep by use of the DELAY TIME
MULT control.
For TRIGGERABLE AFTER DELAY opera
tion both the DELAY TIME and B trigger
LEVEL affect the start of the B sweep so
that the delay time is used as a reference
point. The B sweep becomes triggerable
after delay, which means that it will be
triggered by the first adequate trigger sig-
nal after the delay time has passed.OPERATING INSTRUCTIONS
SINGLE SWEEP OPERATION
This mode of display is useful for looking at
non-synchronous or one time events.
1, Set the TRIG MODE to either AUTO or
NORM. Apply a signal of approximately
the same amplitude and frequency as the
signal that is to be observed as the trig-
ger signal and set the trigger level.
Set the TRIG MODE to SINGLE and press
the RESET button, make sure that the
LED in the switch lights to indicate the
reset condition. This LED goes out when
the A sweep period is completed.
3. The scope is now ready to operate in the
SINGLE SWEEP mode of operation.
Input of the trigger signal results in one
and only one sweep.
NOTE
With the HORIZ DISPLAY set to ALT,
the simultaneous observation of the A
sweep and B sweep waveforms at
SINGLE SWEEP is not possible. Also, for
DUAL or QUAD operation simultaneous
observation is not possible using the ALT
mode. Set the unit to the CHOP mode in
this case.
X-¥ OPERATION
X-¥ operation permits the oscillescope to
pexform many measurements not possible with
conventional sweep operation. The CRT dis~
play becomes an electronic graph of two in-
stantaneous voltages. The display may be a
direct comparison of the two voltages such as
vectorscope display of video color bar pat
terns, However, the X-Y mode can be used to
graph almost any dynamic characteristic if a
transducer is used to change the characteristic
(frequency, temperature, velocity, ete.) into a
voltage. One common application is frequency
response measurements, where the Y axis cor
responds to signal amplitude and the X axis
corresponds to frequency.
1, Select the X-Y mode of HORIZ DIS-
PLAY. In this mode, channel 1 becomes
the Y axis input and channel 2 becomes
the X axis input.
2, The X and ¥ positions are now adjusted
using the CH 2 (X-Y <») POSITION and
CH 1 ¥ POSITION controls respectively.
24
3. Adjust the amount of vertical (Y axis)
deflection with the CH 1 VOLTS/DIV and
VARIABLE controls.
Adjust the amount of horizontal (X axis)
deflection with the CH 2 VOLTS/DIV and
VARIABLE controls.
CASCADED OPERATION
This mode of operation is used when sensi-
tivity of greater than 1 mV/div is required.
1. Connect the CH 1 output to the CH 2
input using a BNC cable, In this case
terminate the CH 2 input into 50 2.
For cascaded operation set the CH 1 and
CH 2 controls for X 5 GAIN. Set the
vertical MODE switch assembly to CH 2.
3. Set the CH 1 and CH 2 VOLTS/DIV con-
trols to the 5 mV/div setting and input a
signal into CH 1. Channel 2 now has a
sensitvity of 500 W/div,
VIDEO SIGNAL OBSERVATION
Setting the COUPLING switch to the
VIDEO position permits selection of vertical
or horizontal sync pulses for sweep triggering
when viewing composite video waveforms.
The VIDEO LINE mode is automatically se~
lected at sweep rates of 50 \s/div and faster.
Horizontal sync pulses are selected as triggers
to permit viewing of horizontal lines of
video. A sweep time of about 10 is/div is ap-
propriate for displaying lines of video. The
sweep time VARIABLE control can be set to
display the exact number of waveforms de~
sired.
The VIDEO FRAME mode is automatically
selected at sweep rates of 0.1 ms/div and
slower. Vertical sync pulses are selected as
triggers to permit viewing of vertical fields
and frames of video. A sweep time of
2 ms/aiv is appropriate for viewing fields of
video and 5 ms/div for complete frames (two
interlaced fields) of video,
At most points of measurement, a com-
posite video signal is of the (-) polarity, that
is, the sync pulses are negative and the video
is positive. In this case, use (-) SLOPE. If the
waveform is taken at a circuit point where the
video waveform is inverted, the sync pulses