CLISIS SYSTEMS

Universal remote controlled table system


INFOTEC
REMOTE CONTROLLED SYSTEM FOR RADIOGRAPHY AND FLUOROSCOPY
WITH DIGITAL DYNAMIC DETECTOR

SYSTEM DISTINCTIVENESS
 Total access to the patient without mechanical limitations
 Auto-focusing removable grid
 Efficient and ergonomic operator interface
 Wide range of configuration options
 Low doses in all applications
 High image quality

MAIN FEATURES
 One side patient table-top supporting structure which allows total accessibility to the pa-
tient also from internal space
 Elevating table-top from 47 to 100 cm
 Lead free collimator with touch screen display for collimator command and table control
 Autofocusing grid that automatically sets the right grid focus according to the detector
distance.
 Innovative system console with joystick for table movements control and 23” touch
screen display with table and generator functions integrated
 Pre-set auto positioning
 RF 43x43 cm dynamic detector system
2  Main options: Digital Angiography DSA, Stitching, Tomosynthesis and Dual energy
 DICOM complete functionality
 High frequency generator
 Can be completed with a second X-ray assembly ceiling suspension, wall-stand and addi-
tional Wi-Fi detectors
INFOTEC
The innovative CLISIS SYSTEMS is a high-technology
and know-how solution focused on the remote-con-
trol table. The system is completed with the digital
system OPERA D4000RF and the generator OPERA G.

The OPERA D4000RF digital system is the best solu-


tion to optimize examination times with significant
cost savings, as well as to ensure excellent quality
images with reduced doses.

The OPERA G generator is high frequency fully inte-


grated command console.

The high level of components integration allows to


control the entire system with a highly innovative
and user-friendly control console.

OPERA D4000RF can be combined with one of the


following dynamic detectors:
Pixium RF4343
Pixium RF4343FL
PaxScan 4343DXV
Can be connected to one or two additional Wi-Fi
detectors

INTENDED USE
This equipment is designed for performing general 3
radiography, fluoroscopy and angiography proce-
dures.
APPLICATIONS AND TECHNIQUES
Gastroenterology
Musculoskeletal
Thorax and lungs
Urology and gynaecology
Paediatrics
Emergency and Traumatology
Linear Tomography
Digital Angiography (optional)
Stitching (optional)
Tomosynthesis (optional)
Dual Energy (optional)
TECHNICAL FEATURES
CLISIS REMOTE CONTROLLED TABLE
EXAM TABLE
Tilting angles -90°/+90°, variable speed up to 5°/s (approx.)
Simultaneous movements Tilting
Longitudinal stand travel
Patient table-top movement
Oblique projections
PATIENT TABLE-TOP
Profile Concave (carbone fiber only) or flat
Dimensions 80 x 246 cm
Radiotransparent area 57 x 237 cm
Composition, attenuation Carbone fiber, < 0,5 mm Al eq @ 100 kV
Plastic laminate, < 1 mm Al eq @ 100 kV
Maximum patient weight 266 kg without limitations (dynamic)
(with patient lying on the exam table) 323 kg with limitation of use (static exams)
Distance from floor Variable from 47 to 100 cm (concave tabletop without longitudinal
travel)
(132 cm total excursion) variable speed up to 6 cm/s (approx.)
Vertical run 83 cm (during tilting phase)
Transversal travel 35 cm, variable speed up to 2,5 cm/s
Longitudinal movement (opt.) ±50 cm
Maximum table-top rear access dis- 52,5 cm max
tance
CARBONE FIBER FOOT REST
Top-side distance from the floor 10,7 cm
Maximum patient’s weight 266 kg
STAND FOR X RAY TUBE ASSEMBLY
4 Source-detector distance From 115 to 180 cm, speed 3,5 cm/s (approx.)
Focus/floor distance (tilt + 90°) From 47,5 to 226,5 cm
Oblique projection ± 40°
Tube rotation 90°/180° manual movement
Longitudinal travel stand 195 cm
Longitudinal stand/detector travel 162 cm
Tube/detector longitudinal travel Variable up to 15 cm/s (approx.)
INFOTEC
OPTIONAL COMPRESSOR DEVICE
Type Single cone with automatic parking
Compression strength Up to 200 N
Travel 55 cm
Compressor operation ± 20°
Safety Double safety (mechanic and electric)
TOMOGRAPHY
Tomography type Linear tomography with two scanning directions for any tilting
table angle
Tomographic area Extended all over image receptor displacement
Focus/detector distance 115 cm
Angles/times Angle (°) Times (s)
8 0,5
20 1,25
30 1,6
40 2,5
Layer height selection 0 - 300 mm with +/- 1 mm adjustment
OPTIONAL TOMOSYNTHESIS
Angles 20°/30°/40°
Exposition time Max 8 s
Layer height selection 0 - 330 mm
TOUCH SCREEN RECTANGULAR COLLIMATOR
Control command type Manual
Remote-controlled
Automatic
Integrated controls for table movements
Field light pointer with LED light >200 lx at 100 cm
Optional additional filters 1 mm Al + 0,1 mm Cu
(motorized filter insertion) 1 mm Al + 0,2 mm Cu 5
2 mm Al + 0,3 mm Cu
Live collimator camera patient display Completely integrated
(optional)
Laser Optional
Rotation flange collimator ±90 Optional
RF DETECTOR DEVICE
Patient table-top detector distance 81 mm with concave profile (+17 mm flat edge table-top)
Grid 200 lines/inch R=12:1 auto-focusing and removable
Exam table edge- detector center 46 cm Patient table-top upper side
(minimum distance) 38 cm free side
DIGITAL SYSTEM OPERA D4000 RF (PDT-PDV versions)
DIGITAL PROCESSOR
Composition Processor unit for digital image processing
Main Controller for general system management
CPU Intel i9
Storage capacity 16 GB (expandable up to 32 GB)
GHz 3.6
Operating system Microsoft Windows 10 – 64 bit
Image archive (SSD) 250 GB expandable up to 2 TB
UPS optional 2200 VA
DETECTORS Pixium RF4343FL Pixium RF4343 PaxScan 4343DXV
Technology/scintillator Amorphous silicon with Caesium Iodide
Acquisition matrix 2880 x 2880 pixel 2880 x 2880 pixel 3072 x 3072 pixel
Dynamic range – linear response 16 bit
Pixel pitch 148 μm 148 μm 139 μm
Acquisition active area Nominal: 43 x 43 cm
Zoom 1: 30 x 30 cm
Zoom 2: 20 x 20 cm
Zoom 3: 15 x 15 cm
Spatial resolution 3.4 lp/mm 3.4 lp/mm 3.6 lp/mm
DQE@ 0 lp/mm 65% 65% 78%
MTF@1 lp/mm 63% 66% 54%
Pixium DETECTOR RF4343FL: ACQUISITION MODE
Continuous digital fluoroscopy Size Matrix Frequency
43 x 43 cm 960x960x16 bit 16 fps
30 x 30 cm 1024x1024x16 bit 12 fps
20 x 20 cm 672x672x16 bit 20 fps
6 15 x 15 cm 1024x1024x16 bit 6 fps
Chance to save images directly and Last Image hold

Pulsed digital fluoroscopy Size Matrix Frequency max (1)


43 x 43 cm 960x960x16 bit 12 fps
30 x 30 cm 1024x1024x16 bit 12 fps
20 x 20 cm 672x672x16 bit 12 fps
15 x 15 cm 1024x1024x16 bit 6 fps
(1) Adjustable from 1 fps
Chance to activate the automatic image saving
Digital radiography Size Matrix Frequency max
43 x 43 cm 2880x2880x16 bit 2 fps
43 x 43 cm 1440x1440x16 bit 6 fps
Automatic image saving
PIXIUM RF4343 detector: ACQUISITION MODE
Continuous digital fluoroscopy Size Matrix Frequency
43 x 43 cm 960x960x16 bit 18 fps
30 x 30 cm 1024x1024x16 bit 15 fps
20 x 20 cm 672x672x16 bit 30 fps
15 x 15 cm 1024x1024x16 bit 15 fps
Chance to save images directly and Last Image hold
Pulsed digital fluoroscopy Size Matrix Frequency max (1)
43 x 43 cm 960x960x16 bit 15 fps
30 x 30 cm 1024x1024x16 bit 15 fps
20 x 20 cm 672x672x16 bit 15 fps
15 x 15 cm 1024x1024x16 bit 15 fps
(1) Adjustable from 1 fps
Chance to activate the automatic image saving
INFOTEC
Digital radiography Size Matrix Frequency max
43 x 43 cm 2880x2880x16 bit 3 fps
43 x 43 cm 1440x1440x16 bit 8 fps
Automatic image saving
PaxScan 4343DXV DETECTOR: ACQUISITION MODE
Continuous digital fluoroscopy Size Matrix Frequency
43 x 43 cm 1024x1024x16 bit 18 fps
30 x 30 cm 1024x1024x16 bit 18 fps
20 x 20 cm 684x684x16 bit 30 fps
15 x 15 cm 1024x1024x16 bit 18 fps
Change to save images directly and Last image hold
Pulsed digital fluoroscopy Size Matrix Frequency max (1)
43 x 43 cm 1024x1024x16 bit 15 fps
30 x 30 cm 1024x1024x16 bit 15 fps
20 x 20 cm 684x684x16 bit 25 fps
15 x 15 cm 1024x1024x16 bit 15 fps
(1) Adjustable from 1 fps
Change to activate the automatic image saving
Digital radiography Size Matrix Frequency max
43 x 43 cm 3072x3072x16 bit 3 fps
43 x 43 cm 1536x1536x16 bit 12 fps
Automatic image saving
FEATURES
Anatomical program Library up to 99 parts of the body, each one with 99 projections per-
formed with seven different patient types and for a total of 68.607 ways
of using the system. The exposure parameters (kV, mA automatic expo-
sure meter dominant) and the Irradiated area can be changed by the op-
erator during the examination. 7
Automatic exposure Automatic exposure meter completed with measure chamber at three
fields. Each field can be independently selected by the operator or by the
anatomical programming.
Ability to link a predefined dose to each single anatomical program in
function of the patient size.
Generator automatic kV and mA set values. Exposures times fixed by the
automatic exposure meter for constantly maintaining operator pre-set
dose based on the exam type selection.
Exposure index (EI) and deviation index (DI) display.
Device for automatic exposure control (kV and mA) in continuous and
pulsed fluoroscopy with automatic radiographic parameter variation ac-
cording to the anatomical part under examination.
Dose area product (DAP) Measure device Ionisation chamber located at the exit of the
collimator
Data collected in DAP (mGycm2 or μGym2) and AK (mGy)
Radiography
Data collected in DAP rate (mGycm2/s or μGym2/s) and
Fluoroscopy AK rate (mGy/min)
Data display DAP: on monitor and on the images
AK: on dedicated area of the monitor
X ray tubes 2
controllable
The cumulated patient’s dose values areas transmitted to the RIS using
DICOM MPPS.
The complete exam dose report is transmitted to the PACS via DICOM
DOSE SR
Exams documentations Each acquired image has indicated:
management - Patient name
- Label number N°
- Born date
- Date and time of the exam execution
- Institute name
- Radiographic parameters
- Dose (mGy cm2)
- Display images parameters and reference scale
Archives querying and sorting by exam date or by patient name.
Possibility to define the default output (Printers, writer/burner CD/DVD,
work station or system PACS).
Monitor reporting of the exam submission to PACS or printer.
Management of multiple connected printers and optimization of the film
use with different modes of playback of the captured images (Film Edi-
tor).
Real time functions processing Automatic image processing and display optimization through software
ATH: Anatomic Tissue Harmoniza- algorithms specifically designed for each type of examination.
tion Functions available in continuous and pulsed fluoroscopy:
FGA: Fluoro Gradient Adaptive Filter - Pre-equalization (FGA)
- Noise reduction (FGA)
- Patient motion detection (FGA)
- Edges enhancement (FGA)
- Digital image inversion H/V
Radiography acquisition with real time images processing display.
Software (ATH) that allows the improvement of the captured images
with contrast increase, brightness and noise reduction. Expansion of the
dynamic range.
Parameters can be set in the anatomical programs and can be
customized during installation according to the specific needs of the op-
8 erator.
Reject analysis Function that allow to mark the displayed image as wrong and then re-
jected. You can also specify the refusal reason by selecting it from a list.
Post-processing functions Post processing functions do not alter the source data which are
immediately restorable
To process the image, the system display in the control area an icon se-
ries that represent the first of various levels, configurable from the set-
tings and containing multiple functions.
Text: mark overlay to write in the image left, right and free insertion on
the image of left/right and free annotation or pre-defined.
Image horizontal, vertical and rotation: H&V image reverse
and 90°image clockwise rotation.
Magnifying lens: image magnifying up to 1 pixel monitor = 1 pixel detec-
tor.
Electronic shutters and zoom selection:
selection of an area of interest on the image also asymmetrical with the
possibility of centring the selected part.
LUT management (Look-Up Table): management and adjusting the im-
ages in terms of brightness and contrast through the variation of the pa-
rameters L (Level) and W (Window) with automatic correction and gray
scale inversion.
Measurements: measurements of relative distances and angles with
graphic overlay on the images, grid overlay
Overview: simultaneously display of a series of images arranged at your
choice from 2x2 to 5x5.
Cineloop: visualization of acquired images series in radiography and
fluoroscopy with the possibility of scrolling at different speed.
Advanced controls: ATH harmonization function, Spatial filters and
gamma curves features.
INFOTEC
ATH Harmonization: harmonization replacement of the ATH applied dur-
ing the processing phase in real time with another predefined one.
Spatial filters: SHARP filters (Edge enhancement) and SMOOTH (faded
edges) kernel size and weight modification
Gamma curves: alternatively, gray level to the default one.
Virtual grid (Software grid) optional : software algorithm that simulates
the grid in case of examination without grid
Functions to assist patient centering Camera integrated collimator camera for patient visualization in the mon-
itor touch-screen display that facilitates the identification of the
anatomical area;
Virtual collimators for adjusting the beam size on the stored image
(Last Image Hold - LIH);
Virtual Scan for patient centering with the aid of the stored image

9
Special procedure TOMOGRAPHY
Acquisition with automatic kV, mA, ms and tomography angle
Optional special procedure IMAGE RECONSTRUCTION (STITCHING)
Automatic re-composition in a single large image from a series of ac-
quisition made during the scanning movement of the patient.
The image is reconstructed with source pixel and displayed on a moni-
tor, processed, printed or sent to the network with normal system
functions. Application in studies of the rachis and lower limbs with the
patient in an erect or decubitus supine position. Reconstruction of im-
ages of length selectable between 60 -90 120 cm
OTHOPEDICS MEASURES
Graphic measuring functions, enabled on all images but particularly
significant for images of the spine and lower limbs obtained through
the stitching function.
- Altitude difference: measurement of the altitude difference be-
tween two points in the image.
- Cobb angle: measures the curvature of the vertebrae using the Cobb
method.
- Double angle of Cobb: calculation of two angles of curvature in case
of scoliosis with “S” spine.
- Perpendicularity lines: measurements of point distance in relation
to some remarkable points in the lower limbs.
- Lower limb measurements: measurements of lengths and angles in
relation to some remarkable points in the lower limbs.
TOMOSYNTHESIS
Acquisition series of bidimensional projections at low dose
obtained in one exposition to reconstruct an arbitrary number of
10 plans representative of the different layers of the volume.
ANGIOGRAPHY (DSA)
Real-time images capture in subtractive mode and road
mapping with maximum opacity. Selectable reference
images on a dedicated monitor.
DUAL ENERGY
Execution of two exposures in rapid succession with different kV.
It allows to separate soft and hard tissues into two distinct images.
In particular, it is used in chest X-rays to highlight lung cancer nodules.
DICOM Connectivity Possibility of sending single or entire scanned images over the net-
work. It is allowed to send to printer or CD/USB and export in RAW,
JPEG and Mp4 format
DICOM optional:
- DICOM Store (Send)
- DICOM Print class
- DICOM Worklist
- DICOM MPPS
- DICOM storage commitment SCU
- DICOM query/retrieve
- DICOM DOSE SR
- DICOM Media interchange (CD/DVD)
INFOTEC
OPERATOR INTERFACE
COMPOSITION
- Main control console with joystick and touch screen, touch screen display and fluoroscopy foot switch pedal
- Table control panel
- Collimator touch screen display
- Secondary control console (with cable) for exam room- optional
MAIN CONTROL CONSOLE
Functions Joystick for table movements with variable speed adjustment, collimator adjust-
ment and up/down compressor movement (if available).
Push button for radiography and fluoroscopy pedal.
Auto-positioning Are available on the system three pre-programmed positions
POSITION 1 Table in horizonal position, SID 115 cm, tube and detector positioned
at the table center (typical of contrast procedure).
POSITION 2 Table in vertical position, SID 115 cm, tube and detector positioned at
the center table (typical for pelvis contrast procedures)
POSITION 3 Table in vertical position, SID 180 cm, tube and detector positioned at
the center table (typical for thorax ).
TOUCH SCREEN
Monitor for medical use with mouse and touchscreen.
Standard, monitor 23” 23” monitor a colori, multi touch
- Native resolution: 1920 x 1080
- View angle: 178° H/V
- Brightness: 260 nit
- Contrast: 1000:1 typ.
Optional, monitor 24.1” 24.1” color monitor, multi touch
- Native resolution 1920 x 1200
- View angle 178° H/V
- Brightness 900 cd/m2 (typical)
- Contrast 1000:1 typ 11
Dedicated area - Table and generator command/control with patient anatomical part display by
the camera completely integrated in the collimator
- Examination management with display/processing.
Functions - Display of the table position
- Display and modification of the generator radiological parameters
- Examination type management
- Images display
- Images Post processing functions
- Image management
- Patient data management
- Communication with RIS-PACS, printers, burners.
Anatomical programming mode can be controlled by the operator or via DICOM
Worklist mode (RIS mapping).
CONTROL PANEL AT THE TABLE EDGE
Functions Table movements controls and collimation
COLLIMATOR TOUCH SCREEN PANEL
Functions Collimator control command and table movements
OPTIONAL UNITS
Digital system monitor Possibility of additional monitors in the examination room mounted on a trolley
or hanging on a ceiling support
Console Secondary control console with cable for exam room
Full control of the The combination of :
exam room - additional monitor with touch functions
- Secondary control console
- Radiography/fluoroscopy remote foot switch pedal
Allows the full system control in the examination room and close to the patient,
both for table movement and for images acquisition and processing/post pro-
cessing
OPERA G GENERATORS (O- GXXXIQ-XN)
GENERAL PERFORMANCES
Frequency Output up to 450 kHz
Connectable Tubes 1 tube (O-GXXXXIQ-1N); 2 tubes (O-GXXXXIQ-2N)
MODELS G500IQ G650IQ G800IQ
Power(kW) 50 65 80
Output parameters 80 kV 630 mA 800 mA 1000 mA
100 kV 500 mA 630 mA 800 mA
150 kV 320 mA 400 mA 500 mA
Maximum exposure rate 30 FPS
Starter peed Low speed/ Dual speed
Input phase/voltage Triphase 400 VAC – 480 VAC
(Auto select in each range)
Line voltage rate ±10%
RADIOGRAPHY G500IQ G650IQ G800IQ
kV Range/Steps 40-150 kV in 1 kV steps
High voltage ripple < 4% peak-peak for ≥ 10 kHz
Valori mA 10-630 mA 10 – 800 mA 10 – 1000 mA
* R’10 * R’10 * R’10
ms time range 1 to 6300
mAs Range (no AEC) 0,1 – 1000 mAs in 0,1 mAs steps
CONTINUOUS FLUOROSCOPY
kV Range/Steps 40-125 kV in 1 kV steps
High voltage ripple <1kV a 100 KV e 5mA
12 mA range/steps 0,5 –10 mA/ in 0,1 mA steps
PULSED FLUOROSCOPY
kV Range/steps 40-125 kV in 1 kV steps
mA Range/steps 5-99 mA in 1 mA steps
ERRORS MESSAGES
In case of generator malfunction, in the dedicated area it is displayed the error message with related code
* Rénard series
INFOTEC

X-RAY TUBES (possibility of using use different X-ray tubes)


RTM101 RTC600 RTC1000
Anode speed (rpm) Low speed/High speed
Anode material RTM RT-TZM-C RT-TZM-C
Anode diameter (mm) 102 102 110
Anode angle (°) 12,5 13 13
Nominal X-ray tube voltage (kV)) 150 150 150
Focal spot (mm) 0,6/1,2 0,6/1,2 0,6/1,0
Nominal anode input power, 40/100 43/100 40/80
High speed (kW)
Nominal anode input power, 26/63 24/63 26/45
Low speed (kW)
Maximum heat content (kJ) 300 450 840
(400 kHU) (600 kHU) (1120 kHU)
Maximum continuous heat dissipation (W) 1000 1500 2000
Housing specifications C100 C100 C100
Maximum housing heat content (kJ) 1500 (2000 kHU)
Maximum housing heat dissipation (W) 250 (20000 HU/min) without fan
600 (48000 HU/min) with fan

13
DOSE: REDUCTION, MONITORING AND MANAGMENT
MEANS FOR RDUCING THE DOSE TO THE PATIENT AND TO THE OPERATOR
Exam table Lower absorption carbon fiber patient table-top
Anti-scatter grid Grid with carbon fiber cover.
Possibility to park the grid outside the field
Virtual grid (Software grid) Software algorithm that simulates the behaviour of a real grid by
reducing the patient scattering radiation and improving the visibility of
low-contrast parts
Digital system-generator integration Generator high integration to optimize the anatomical techniques,
automatic collimation, filtration and image processing
Exposure meter Controlled by the digital system to allow greater accuracy in the expo-
sure
Paediatric anatomical programme Available in four different age size:
- New-born: 0 - 28 days
- Infants: 29 days – under 2 years
- Children: 2 under 12 years
- Adolescents: 12 - 21 years
Electronic shutters Command to activate automatic recognition of the collimator shutters
in the acquired image (LIH) without X-ray emission.
Virtual Scan (optional) Enables patient centring using the stored image without X-rays emis-
sion.
Live camera for patient display Live camera collimator for the identification of the anatomical part of
(optional) interest without X-rays emission
Generator High frequency generator, allows an output wave shape particularly
well levelled with very short start/end ramps.
Collimator Manual and automatic collimator with square field and
hardening beam filters.
Filters insertion that can be checked by the operator or programmable
14 within the anatomical program.
Power LED for irradiated field simulation.
DOSE CONTROL AND MANAGMENT
Exposure index (EI) Verification involves measuring the levels of the acquired image within
Deviation index (DI) a set area in order to quantify the deviation between the detected ex-
posure dose and the exam dose set.
The result is shown by the Exposure Index (EI) and the Deviation Index
(DI).
Dose – area measurement Value is measured by ionization chamber located at the collimator
device (DAP) output. On the console there is an indicator to monitor the measure-
ment of the dose-area product: Radiography DAP (mGycm2 or μGym2),
fluoroscopy DAP rate (mGycm2/s or μGym2/s).
In case there is a second X-ray assembly a second chamber is used.
Dose measurement device On the monitor touch screen there is an indicator to monitor the pa-
Air Kerma (AK) tient dose: Air Kerma rate expressed in mGy/min and Air Kerma cumu-
late of reference expressed in mGy.
Data display DAP: the values are displayed on touch screen monitor in the genera-
tor control/command dedicated area and on the image.
AK: the values are displayed on the monitor touch screen on the dedi-
cated area control/generator control
RIS-PACS transmission The cumulated patient’s dose values areas transmitted to the RIS using
DICOM MPPS.
The complete exam dose report is transmitted to the PACS via DICOM
DOSE SR
INFOTEC

ACCESSORIES

Adjustable patient foot-rest Compression belt Hand grips


Shoulder rest

Ankle support
Lower limbs support Head contention device Lateral cassette support

Collimation rotation Radiography/fluoroscopy re-


Barium glass support
flange ±90° mote foot switch
Intercom 15

Barra stitching Cushion for patient table-top Arm support


Secondary console with cable
SUPPLEMENTARY DEVICES
CEILING SUSPENSION KALOS
Longitudinal travel 354 cm
Transversal travel 277 cm
Vertical travel 170 cm
X-ray tube rotation ±160° rotation on vertical axis
±155° rotation on horizoontal axis
Collimator Manual

CEILING SUSPENSION MSP


Longitudinal travel 326 cm
Transversal travel 210 cm
Vertical travel 150 cm
X-ray tube rotation +200°/ -135° rotation on vertical axis
+120°/-210° rotation on horizontal axis
Collimator Manual

WALL STAND
Bucky travel vertical vertical and orientable
Bucky movement / from -20° to +90°
Bucky rotation movement / Manual +90°
Bucky vertical travel 157,4 cm 140,4 cm
Grids For CR/film: 90 lines/inches R=12:1 F=180 cm
For Wi-Fi detector: 200/230 lines/inches R=12:1 F=130 cm
and F=180 cm
ADDITIONAL Wi Fi Pixium DETECTORS 2430EZ 3543EZ-C
Technology Amorphous silicon Amorphous silicon
16 Scintillator Caesium Iodide Caesium Iodide
Pixel area 24 x 30 cm 35 x 43 cm
Pixel dimensions 148 m 148 m
Data conversion 16 bits 16 bits
DQE@ 0 lp/mm 70% 70%
Weight (without battery) 1.4 kg 2.6 kg
Battery weight 227 g
LUMEN 2530W
ADDITIONAL Wi-Fi Varex DETECTORS (2530W Standard) (2530W Premium)
Technology Amorphous silicon Amorphous silicon
Scintillator Caesium Iodide Caesium Iodide
Pixel area 25 x 30 cm 25 x 30 cm
Pixel dimensions 131 µm 131 µm
Data conversion 16 bit 16 bit
DQE@ 0 lp/mm 64% 74%
MTF@ 1 lp/mm 65% 61%
Weight (without battery) 1,56 kg 1,56 kg
Battery weight 340 g
INFOTEC

LUMEN 4336W
ADDITIONAL Wi-Fi Varex DETECTORS (4336W-G5 C Standard) (4336W-G5 C Premium)
Technology Amorphous silicon Amorphous silicon
Scintillator Caesium Iodide Caesium Iodide
Pixel area 35 x 43 cm 35 x 43 cm
Pixel dimensions 139 µm 139 µm
Data conversion 16 bit 16 bit
DQE@ 0 lp/mm 64% 74%
MTF@ 1 lp/mm 65% 61%
Weight (without battery) 2,8 kg 2,8 kg
Battery weight 340 g

DETETTORE Wireless VAREX 4343W


Scintillator CsI Premium CsI Standard
Technology Amorphous silicon
Nominal size 43 x 43 cm
Pixel matrix - Total 3072x3072
Pixel pitch 139 μm
Spatial resolution 3,6 lp/mm
A/D conversion 16 bit
Weight 3,3 kg 3,3 kg
DQE @ 0 lp/mm 76% 3,3 kg
MTF @ 1 lp/mm 57% 61%
Uniformly loaded across entire surface 300 kg
IP protection IP68
Charge capability 1600 images over 8 hours
17
Weight (without battery) 3,3 kg 3,3 kg
Battery weight 340 g
TYPICAL EXAM ROOM LAYOUT

18

1 Control console
2 Exam table
3 Digital system cabinet
4 Generator cabinet
(*) Maximum longitudinal overall dimensions

Safety area