M020 Nanodrop Eight Spectrophotometer User Guide
M020 Nanodrop Eight Spectrophotometer User Guide
NanoDrop Eight
User Guide
For U.S. Technical Support, please contact: For International Support, please contact:
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to
use in the product operation. This document is copyright protected and any reproduction of the
whole or any part of this document is strictly prohibited, except with the written authorization of
Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document
supersede all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or
error-free and assumes no responsibility and will not be liable for any errors, omissions, damage or
loss that might result from any use of this document, even if the information in the document is
followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a
purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale,
which Terms and Conditions of Sale shall govern all conflicting information between the two
documents.
For Research Use Only. This instrument or accessory is not a medical device and is not intended to
be used for the prevention, diagnosis, treatment or cure of disease.
Contents
Chapter 1 About the Spectrophotometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
USB-B port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
PR-1 Pedestal Reconditioning Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
PV-8 Performance Verification Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Instrument Detection Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Arm
Pedestal
Note Locate the instrument away from air vents and exhaust fans to minimize
evaporation
Note Before operating a NanoDrop Eight instrument, please read the safety and
operating precautions and then follow their recommendations when using the
instrument.
Features
TheNanoDrop Eight spectrophotometer features the patented micro-volume sample
retention system.
USB-B port
Connect the NanoDrop Eight to your PC via USB to operate the instrument using the
PC software.
Accessories
This section lists the accessories included for use with the NanoDrop Eight.
Instrument Set up
From any PC that is connected to the Internet, use any web browser to navigate to
www.thermofisher.com/nanodropsw.
On the website, locate the Register Your Instrument button, and follow the
instructions to register the instrument.
Computer Requirements
Required Windows Operating System
• Windows 10 Enterprise or Professional, build 1607 or greater
Update Software
Quickly and easily download and install the latest NanoDrop Eight software and
release notes from our website. Follow the steps to update or upgrade the
NanoDrop Eight software on a personal computer (PC). An Internet connection is
required to download software. Navigate to www.thermofisher.com/nanodropsw
and click on the NanoDrop Eight software tab. Follow the Instrument Software
Download instructions.
Technical Support
For U.S./Canada Support, please contact:
Thermo Fisher Scientific
3411 Silverside Road
Tatnall Building, Suite 100
Wilmington, DE 19810 U.S.A.
http://www.thermofisher.com/NanoDropDistributors
If you are experiencing an issue with your system, refer to the troubleshooting
information. If the issue persists, contact us. If you are outside the U.S.A. and
Canada, please contact your local distributor.
Note Detection limits provided in the following tables are approximate and apply
to micro-volume measurements only; they are based on the instrument’s
photometric absorbance range (10 mm equivalent) of 0.04–200 A.
Reported Results
Settings
Detection Limits
Calculations
Before taking pedestal measurements with the NanoDrop Eight instrument, lift the
instrument arm and clean the upper and lower pedestals. At a minimum, wipe the
pedestals with a new laboratory wipe. For more information, see Cleaning the
Pedestals.
When the sample measurement is completed, the spectrum and reported values
are displayed (see the next section).
8. When you are finished measuring samples, select End Experiment .
9. Lift the arm and clean all pedestals with a new wipe.
Related Topics
• Measure a Micro-Volume Sample
• Best Practices for Micro-Volume Measurements
• Prepare Samples and Blanks
• Basic Instrument Operations
Auto measure
UV spectrum
Baseline
correction
Menu of table
options;
click to choose
which columns to
report
Baseline Correction
For the dsDNA, ssDNA or RNA, enter baseline correction wavelength in nm or use
default value (340 nm)
c = A * [1/( * b)]
or
c=A*f
where:
c = analyte concentration in ng/µL
A = absorbance in absorbance units (A)
f = factor in ng-cm/µL (see below)
For the dsDNA, ssDNA and RNA Factors Used
applications, the generally accepted factors
for nucleic acids are used in conjunction • dsDNA (factor = 50 ng-cm/µL)
with Beer’s Law to calculate sample
concentration. For the Custom Factor • ssDNA (factor = 33 ng-cm/µL)
application, the user-specified factor is
• RNA (factor = 40 ng-cm/µL)
used.
• Custom Factor (user entered factor between
15 ng-cm/µL and 150 ng-cm/µL
Reported Values
• Nucleic acid concentration. Reported in selected
unit (i.e., ng/µL, µg/µl or µg/mL). Calculations are based
on modified Beer’s Law equation using corrected nucleic
acid absorbance value.
Measure Microarray
Measures the concentration of purified
nucleic acids that have been labeled with
up to two fluorescent dyes for use in
downstream microarray applications.
Reported Results
Settings
Detection Limits
Calculations
Before taking pedestal measurements with the NanoDrop Eight instrument, lift the
instrument arm and clean the upper and lower pedestals. At a minimum, wipe the
pedestals with a new laboratory wipe. For more information, see Cleaning the
Pedestals.
Related Topics
• Best Practices for Nucleic Acid Measurements
• Measure a Micro-Volume Sample
• Best Practices for Micro-Volume Measurements
• Prepare Samples and Blanks
• Basic Instrument Operations
Right click
UV-Vis Spectrum graph area to
view display
options.
Click to
select unit
Click to
select data
columns
Click row to
select sample
and update
spectrum; click
Sample names; Nucleic acid Dye more rows to
Click to edit concentration concentration(s) overlay spectra
Note
• Analysis correction is performed at 340 nm (absorbance value at 340 nm is
subtracted from absorbance values at all wavelengths in sample spectrum).
• Micro-volume absorbance measurements are normalized to a 10.0 mm
pathlength equivalent.
• sample name
• created on (date sample measurement was taken)
• nucleic acid concentration
• A260
• A260/A280
• dye 1/dye 2 concentration
• sample type
• analysis correction
• factor
Microarray settings
The Microarray Setup screen appears after you select the Microarray application
from the Nucleic Acids tab on the Home screen. To show the Microarry settings from
the Microarray measurement screen, select Microarray Setup .
Dye/chromophore editor
Use the Dye/Chromophore Editor to add a custom dye to the list of available dyes in
Microarray Setup or Proteins & Labels Setup. You can also specify which dyes are
available in that list.
To access the Dye/Chromophore Editor, from the Home screen, select Settings >
Dye Editor.
Dye Editor
Locked dye (pre-defined;
Select to add
cannot be edited or deleted)
custom dye
Select to
edit selected
custom dye
Note To determine dye correction values (if not available from dye
manufacturer):
– use instrument to measure pure dye and note absorbance at 260 nm,
280 nm and at analysis wavelength for dye (see above)
– calculate ratio of A260/Adye wavelength and enter that value for 260 nm
Correction
– calculate ratio of A280/Adye wavelength and enter that value for 280 nm
Correction
Absorbance values at 260 nm, 280 nm and • Nucleic acid absorbance values for all Microarray sample
sometimes 230 nm are used to calculate types are measured at 260 nm using the 850-corrected
purity ratios for the measured nucleic acid and normalized spectrum.
samples. Purity ratios are sensitive to the
presence of contaminants in the sample, • If Analysis Correction is selected, the absorbance value at
such as residual solvents and reagents the correction wavelength is subtracted from the
typically used during sample purification. absorbance at 260 nm.
• If one or more dyes are selected, the dye correction
values at 260 nm are also subtracted from the
absorbance at 260 nm.
• The final corrected absorbance at 260 nm is reported
and used to calculate sample concentration.
A280 absorbance
• Analysis-corrected and normalized absorbance value at
280 nm (minus the A280 dye correction) is used to
calculate an A260/A280 ratio.
Dye correction
• Pre-defined dyes have known correction values for A260
and A280. See Dye/Chromophore Editor for correction
values used.
Reported Values
• Nucleic acid concentration. Reported in selected
unit (i.e., ng/µL, µg/µl or µg/mL). Calculations are based
on modified Beer’s Law equation using corrected nucleic
acid absorbance value.
Related Topics
• Calculations for Nucleic Acid Measurements
Reported Results
Settings
Detection Limits
Calculations
Before taking pedestal measurements with the NanoDrop Eight instrument, lift the
instrument arm and clean the upper and lower pedestals. At a minimum, wipe the
pedestals with a new laboratory wipe. For more information, see Cleaning the
Pedestals.
2. Enter the factor to be used for the calculations and specify a baseline correction
if desired.
3. Pipette 1–2 µL blanking solution onto the lower pedestals and lower the arm.
4. Select Blank and wait for the measurement to complete.
Tip: If Auto-Blank is On, the blank measurement starts automatically after you
lower the arm.
5. Lift the arm and clean all pedestals with a new laboratory wipe.
6. Pipette 1-2 µL sample solution onto the pedestals and lower the arm.
7. Start the sample measurement: If Auto-Measure is On, lower arm; if
Auto-Measure is off, lower arm and select Measure .
When the sample measurement is completed, the spectrum and reported values
are displayed (see the next section).
8. When you are finished measuring samples, select End Experiment.
9. Lift the arm and clean all pedestals with a new wipe.
Related Topics
• Measure a Micro-Volume Sample
• Best Practices for Micro-Volume Measurements
• Prepare Samples and Blanks
• Basic Instrument Operations
Select to toggle
With sample selected, click and drag an Auto-Measure
area to zoom. Right-click and select ON/OFF
Autoscale to fit spectra to window
Baseline
correction
wavelength;
select to edit
Sample rows
Related Topics
• Basic Instrument Operations
• Nucleic Acid Reported Results
• Nucleic Acid Calculations
f = 1/( 260 * b))
where:
f= factor
= molar extinction coefficient at 260 nm in ng-cm/µL
b = sample pathlength in cm (1 cm for nucleic acids
measured with the NanoDrop Eight instruments)
Baseline Correction On or off Optional user-defined baseline correction. Can
be used to correct for any offset caused by light scattering
Enter baseline particulates by subtracting measured absorbance at
correction wavelength specified baseline correction wavelength from absorbance
in nm or use default values at all wavelengths in sample spectrum. As a result,
value (340 nm) absorbance of sample spectrum is zero at specified
baseline correction wavelength.
Related Topics
• Instrument Settings
Related Topics
• Detection Limits for All Applications
Reported Results
Settings
Detection Limits
Calculations
Before taking pedestal measurements with the NanoDrop Eight instrument, lift the
instrument arm and clean the upper and lower pedestals. At a minimum, wipe the
pedestals with a new laboratory wipe. For more information, see Cleaning the
Pedestals.
2. Specify the Oligo base sequence for each individual channel and a baseline
correction if desired.
3. Pipette 1–2 µL blanking solution onto the lower pedestal and lower the arm.
4. Select Blank and wait for the measurement to complete.
Tip: If Auto-Blank is On, the blank measurement starts automatically after you
lower the arm.
5. Lift the arm and clean all pedestals with a new laboratory wipe.
6. Pipette 1-2 µL sample solution onto the pedestal and lower the arm.
7. Start the sample measurement: If Auto-Measure is On, lower arm; if
Auto-Measure is off, lower arm and select Measure.
When the sample measurement is completed, the spectrum and reported values
are displayed (see the next section).
8. When you are finished measuring samples, tap End Experiment .
9. Lift the arm and clean all pedestals with a new wipe.
Related Topics
• Best Practices for Nucleic Acid Measurements
Right click
graph area to
UV spectrum view display
options.
Click to
select unit
Menu of table
options;
click to choose
which columns
to report
Click row to
select sample
and update
spectrum; click
more rows to
overlay spectra
Wellplate view; Well plate
toggle to view/ options Purity ratios
hide well plate Sample name; Nucleic acid
map Select to edit concentration
• sample name
• created on (date sample measurement was taken)
• nucleic acid concentration
• A260/A280
• A260/A230
• A260
• A280
• factor
• oligo sequence
• baseline correction
Note The five nucleotides that comprise DNA and RNA exhibit widely varying
A260/A280 ratios. See Oligo Purity Ratios for more information.
Related Topics
• Basic Instrument Operations
• Oligo Calculations
f = 1/( 260 * b)
where:
f= factor
= molar extinction coefficient at 260 nm in ng-cm/µL
b = sample pathlength in cm (1 cm for nucleic acids
measured with the NanoDrop Eight instrument)
Related Topics
• Instrument Settings
Related Topics
• Calculations for Nucleic Acid Measurements
Protein Applications
Reported Results
Settings
Detection Limits
Calculations
Before taking pedestal measurements with the NanoDrop Eight instrument, lift the
instrument arm and clean the upper and lower pedestals. At a minimum, wipe the
pedestals with a new laboratory wipe. For more information, see Cleaning the
Pedestals.
• Choosing a blank:
– For the Protein A280 application, blank with the same buffer solution used to
resuspend the analyte of interest. The blanking solution should be a similar
pH and ionic strength as the analyte solution.
• Run a blanking cycle to assess the absorbance contribution of your buffer
solution. If the buffer exhibits strong absorbance at or near the analysis
wavelength (typically 280 nm or 205 nm), you may need to choose a different
buffer or application. See Choosing and Measuring a Blank for more information.
Note Buffers such as Triton X, RIPA, and NDSB contribute significant
absorbance and are not compatible with direct A280 or A205
measurements.
Related Topics
• Best practices for protein measurements
• Measure a Micro-Volume Sample
• Prepare Samples and Blanks
• Basic Instrument Operations
Auto measure
UV spectrum Right click graph area
to view display options.
Absorbance at
280 nm
Baseline
correction
wavelength
Menu of table
options;
click to choose
which columns to
report
Wellplate view; Sample name; Purity ratios Click row to select sample
toggle to view/ hide select to edit and update spectrum.
well plate map
Well plate options Protein concentration
Related Topics
• Basic Instrument Operations
• Protein A280 Calculations
Each sample type applies a unique extinction coefficient to the protein calculations. If
the extinction coefficient of the sample is known, choose the + MW (molar) or 1%
(mass) option and enter the value. Otherwise, calculate the extinction coefficient or
choose the option that best matches the sample solution. If you only need a rough
estimate of protein concentration and the sample extinction coefficient is unknown,
select the 1 Abs=1 mg/mL sample type option.
Mass Ext.
Setting Available Options Coefficient Description
(L/gm-cm)
Sample 1 Abs = 1 mg/mL General reference Recommended when extinction coefficient is
typea unknown and rough estimate of protein
concentration is acceptable for a solution with
no other interfering substances. Assumes
0.1% (1 mg/mL) protein solution produces
1.0A at 280 nm (where pathlength is 10 mm),
i.e., 1% = 10.
BSA 6.7 Calculates BSA (Bovine Serum Albumin)
protein concentration using mass extinction
coefficient () of 6.7 L/gm-cm at 280 nm for
1% (i.e., 10 mg/mL) BSA solution. Assuming
MW is 66,400 daltons (Da), molar extinction
coefficient at 280 nm for BSA is approximately
43,824 M-1cm-1.
IgG 13.7 Suitable for most mammalian antibodies (i.e.,
immunoglobulin G or IgG). Calculates protein
concentration using mass extinction
coefficient () of 13.7 L/gm-cm at 280 nm for
1% (i.e., 10 mg/mL) IgG solution. Assuming
MW is 150,000 Da, molar extinction coefficient
at 280 nm for IgG is approximately
210,000 M-1cm-1.
Lysozyme 26.4 Calculates lysozyme protein concentration
using mass extinction coefficient () of
26.4 L/gm-cm at 280 nm for
1% (i.e., 10 mg/mL) lysozyme solution.
Assumes molar extinction coefficient for egg
white lysozyme ranges between
36,000 M-1cm-1 and 39,000 M-1cm-1.
Mass Ext.
Setting Available Options Coefficient Description
(L/gm-cm)
Other protein User entered molar Assumes protein has known molar extinction
( + MW) extinction coefficient coefficient () and molecular weight (MW),
and molecular where:
weight
( molar)*10=( percent)*(MWprotein)
Mass Ext.
Setting Available Options Coefficient Description
(L/gm-cm)
Baseline On or off N/A Corrects for any offset caused by light
Correction scattering particulates by subtracting
Enter baseline measured absorbance at specified baseline
correction correction wavelength from absorbance
wavelength in nm or values at all wavelengths in sample spectrum.
use default value As a result, absorbance of sample spectrum is
(340 nm) zero at specified baseline correction
wavelength.
Protein editor
Use the Protein Editor to add a custom protein to the list of available protein sample
types in Protein A280 Setup.
To calculate upper detection limits for other (non-BSA) protein sample types
To calculate upper detection limits in mg/mL for proteins, use the following equation:
b = pathlength in cm
Cprotein = (A / 1%) * 10
Cprotein = (5.8/6.6 g/100 mL) * 10
Cprotein = 8.79 mg/mL
Reported Results
Settings
Detection Limits
Calculations
Before taking pedestal measurements with the NanoDrop Eight instrument, lift the
instrument arm and clean the upper and lower pedestals. At a minimum, wipe the
pedestals with a new laboratory wipe. For more information, see Cleaning the
Pedestals.
Related Topics
• Best Practices for Protein Measurements
• Measure a Micro-Volume Sample
• Prepare Samples and Blanks
• Basic Instrument Operations
Auto measure
UV spectrum Right click graph area
to view display options.
Baseline Correction
Menu of table
options;
click to choose
which columns to
report
• Protein concentration
• Absorbance at 205 nm
• Absorbance at 280 nm
• Sample type
• Baseline correction
Related Topics
• Basic Instrument Operations
• Protein A205 Calculations
Mass Ext.
Setting Available Options Coefficient Description
(L/gm-cm)
Sample type 31 31
Assumes 0.1% (1 mg/mL) at 205 nm = 31
Scopes 27 + 120 * Assumes 0.1% (1 mg/mL) at 205 nm = 27 +
(A280/A205) 120 * (A280/A205)
Other protein User entered mass Assumes protein has known mass extinction
( 1%)
extinction coefficient coefficient ( ). Enter mass extinction
coefficient in L/gm-cm for 1 mg/mL ( 0.1%)
protein solution.
Baseline On or off N/A Corrects for any offset caused by light
Correction scattering particulates by subtracting
Enter baseline measured absorbance at specified baseline
correction correction wavelength from absorbance
wavelength in nm or values at all wavelengths in sample spectrum.
use default value As a result, absorbance of sample spectrum is
(340 nm) zero at specified baseline correction
wavelength.
Related Topics
• Instrument Settings
A280 absorbance
• Normalized and baseline-corrected (if selected)
absorbance value at 280 nm is also reported.
Sample Pathlength
• For micro-volume measurements, the software selects
the optimal pathlength (between 1.0 mm and 0.1 mm)
based on sample absorbance at the analysis wavelength.
• Displayed spectra and absorbance values are normalized
to a 10 mm pathlength equivalent.
Reported Values
• Protein concentration. Reported in selected unit
(mg/mL or µg/mL). Calculations are based on
Beer-Lambert equation using corrected protein
absorbance value.
Related Topics
• Beer-Lambert Equation
• Protein A280 Calculations
Reported Results
Settings
Detection Limits
Calculations
Before taking pedestal measurements with the NanoDrop Eight instrument, lift the
instrument arm and clean the upper and lower pedestals. At a minimum, wipe the
pedestals with a new laboratory wipe. For more information, see Cleaning the
Pedestals.
Related Topics
• Best practices for protein measurements
• Measure a Micro-Volume Sample
• Prepare Samples and Blanks
• Basic Instrument Operations
Analysis Correction
Click row to
select sample and
update spectrum.
Note
• A baseline correction is performed at 340 nm (absorbance value at 340 nm is
subtracted from absorbance values at all wavelengths in sample spectrum).
• Micro-volume absorbance measurements are normalized to a 10.0 mm
pathlength equivalent.
Related Topics
• Basic Instrument Operations
• Proteins & Labels calculations
Mass Ext.
Setting Available Options Coefficient Description
(L/gm-cm)
Sample 1 Abs = 1 mg/mL General reference Select Sample type for detailed description of
typea each available setting.
BSA 6.7
Each sample type applies a unique extinction
IgG 13.7 coefficient to the protein calculations. If the
extinction coefficient of the sample is known,
choose the + MW (molar) or 1% (mass)
Lysozyme 26.4
Other protein user-entered molar option and enter the value. Otherwise,
( + MW) extinction calculate the extinction coefficient or choose
coefficient/ the option that best matches the sample
molecular weight solution. If you only need a rough estimate of
protein concentration and the sample
Other protein User entered mass extinction coefficient is unknown, select the
(1%) extinction 1 Abs=1 mg/mL sample type option.
coefficient
Tip: Ideally, the extinction coefficient should
be determined empirically using a solution of
the study protein at a known concentration
using the same buffer.
Analysis On or off N/A Corrects sample absorbance measurement
Correctionb for any offset caused by light scattering
Enter analysis particulates by subtracting absorbance value
correction wavelength at specified analysis correction wavelength
in nm or use default from absorbance value at analysis
value (340 nm) wavelength. Corrected value is used to
calculate sample concentration.
Mass Ext.
Setting Available Options Coefficient Description
(L/gm-cm)
Sloping Dye On or off Corrects dye absorbance measurements for
Correctiond any offset caused by light scattering
particulates by subtracting absorbance value
of a sloping baseline from 400 nm to 850 nm
from absorbance value at dye’s analysis
wavelength.
a To add or edit a custom protein, use Protein Editor.
b Analysis Correction affects calculation for protein concentration only.
c To add custom dye or edit list of available dyes, use Dye/Chromophore Editor.
d Sloping Dye Correction affects calculations for dye concentration only.
Related Topics
• Instrument Settings
• Protein Editor
• Dye/Chromophore Editor
To calculate upper detection limits for other (non-BSA) protein sample types
To calculate upper detection limits in mg/mL for proteins, use the following equation:
(upper absorbance limitinstrument /mass extinction coefficientsample) * 10
For example, if the sample’s mass extinction coefficient at 280 nm is 6.7 for a 1%
(10 mg/mL) solution, the equation looks like this:
(200 / 6.7) * 10 = 298.5 (or ~300)
Related Topics
• Detection Limits for All Applications
If the extinction coefficient of the sample is • Other protein ( + MW), user-specified molar ext.
known, choose the + MW (molar) or coefficient
1% (mass) option and enter the value.
• Other protein (1%), user-specified mass ext.
Otherwise, calculate the extinction
coefficient or choose the option that best coefficient
matches the sample solution.
Note: See Sample Type for details.
Tip: Ideally, the extinction coefficient
should be determined empirically using a
solution of the study protein at a known
concentration using the same buffer.
Calculated protein concentrations are Measured Values
based on the absorbance value at 280 nm,
the selected (or entered) extinction A280 absorbance
coefficient and the sample pathlength. A
single-point baseline correction (or analysis Note: The absorbance value at 850 nm is subtracted from all
correction) may be applied. wavelengths in the spectrum. As a result, the absorbance at
850 nm is zero in the displayed spectra. Also, for
Concentration is reported in mass units.
micro-volume absorbance measurements, the spectra are
Calculators are available on the Internet to
convert concentration from mass to molar normalized to a 10 mm pathlength equivalent.
units based on sample sequence.
• Protein absorbance values are measured at 280 nm
using the Analysis-corrected and normalized spectrum. If
Analysis Correction and Dye Correction are not selected,
this is the reported A280 value and the value used to
calculate protein concentration.
• If Analysis Correction is selected, the normalized and
analysis-corrected absorbance value at 280 nm is
reported and used to calculate protein concentration.
• If a Dye is used, the normalized, analysis-corrected and
dye-corrected absorbance value at 280 nm is reported
and used to calculate protein concentration.
Dye correction
• Pre-defined dyes have known correction values for A260
and A280. See Dye/Chromophore Editor for correction
values used.
Related Topics
• Beer-Lambert Equation
• Protein A280 Calculations
Reported Results
Settings
Detection Limits
Protein standards for generating a standard curve may also be provided by the kit
manufacturer. Since the NanoDrop Eight pedestals can measure higher protein
concentrations than traditional cuvette-based spectrophotometers, you may need to
supply your own protein standards at higher concentrations than provided by the
manufacturer. For example, additional standards may be required to ensure the
standard curve covers the dynamic range of the assay and the expected range of
the unknown samples.
• Use the application setup to enter concentration values for the standards and to
specify how standards and samples will be measured (number of replicates, etc.).
Reference concentration
Standard concentrations
Here is an example:
Curve type setting R2 value (1.0 equals perfect fit)
Standard curve
The R2 value indicates how well the standard curve fits the standard data points
(1.0 is a perfect fit; all points lie exactly on the curve).
Standards are listed in the lower half of the screen in the data table.
After the minimum number of standards has been measured for the selected
curve type, a message similar to the following appears:
Load more standards: returns to the setup screen where you can add or edit
the concentration value for any standard and then measure the standard.
Load Sample ID File: allows for selection of Sample ID information from an
imported sample ID file.
Continue to measurements: continues to sample measurement screen,
after which standards can no longer be edited.
• You can add, edit or delete a standard any time before the first sample
measurement.
Add standard:
– from standards measurement screen, select BCA Setup
– Increment the #of standards drop-down and enter the concentration
value for the new standard
– select Done
Edit standard:
– from standards measurement screen, seelct BCA Setup
• After the minimum number of standards has been measured for the selected
curve type, the message “Invalid Curve” changes to “Valid Curve.” (This occurs
even when additional standards have been defined but not yet measured.) If the
“Invalid Curve” message remains after all entered standards have been
measured, try:
– selecting a different curve type
– remeasuring standards using the correct standard material
Valid Curve indicator: This is only an indicator that the required minimum
number of points has been established for the selected curve type. It does not
validate the integrity of the curve. For example, additional standards may be
required to cover the expected assay concentration range.
Before taking pedestal measurements with the NanoDrop Eight instrument, lift the
instrument arm and clean the upper and lower pedestals. At a minimum, wipe the
pedestals with a new laboratory wipe. For more information, see Cleaning the
Pedestals.
Menu of table
options;
click to choose
which columns to
report
Note
• A baseline correction is performed at 750 nm (absorbance value at 750 nm is
subtracted from absorbance values at all wavelengths in sample spectrum).
• Micro-volume absorbance measurements are normalized to a 10.0 mm
pathlength equivalent.
The R2 value indicates how well the standard curve fits the standard data points (1.0
is a perfect fit; that is, all points lie exactly on the curve).
Standard curve
Selected sample
Click row to
select sample
and update
graph; select
more rows to
overlay spectra
for up to five
samples.
Date/time measured
Protein conc.
Average absorbance at
Baseline correction absorbance 562 nm for replicate
standard measurements
Equation of the standard curve
Additional monitored wavelength
Related Topics
• Basic Instrument Operations
• Protein A280 Calculations
Setting Description
Curve Type Specify type of equation used to create standard curve from standard concentration
values. Available options:
– Linear: Draws the linear least squares line through all measured standards
(requires reference measurement and at least one standard)
– Interpolation: Draws a series of straight lines to connect all measured
standards (requires reference measurement and at least one standard)
– 2nd order polynomial: Draws the 2nd order least squares polynomial
using all measured standards (requires reference measurement and at least
two standards)
– 3rd order polynomial: Draws the 3rd order least squares polynomial using
all measured standards (requires reference measurement and at least three
standards)
Replicates Enter number of measurements of the reference or the same standard that are
averaged together to produce its associated concentration value.
Note: Replicates setting cannot be changed after the first standard has been
measured.
Note: Concentration values can be entered in any order but the standards must be
measured in the order they were entered.
Reported Results
Settings
Detection Limits
Protein standards for generating a standard curve may also be provided by the kit
manufacturer. Since the NanoDrop Eight pedestals can measure higher protein
concentrations than traditional cuvette-based spectrophotometers, you may need to
supply your own protein standards at higher concentrations than provided by the
manufacturer. For example, additional standards may be required to ensure the
standard curve covers the dynamic range of the assay and the expected range of
the unknown samples.
Before taking pedestal measurements with the NanoDrop Eight instrument, lift the
instrument arm and clean the upper and lower pedestals. At a minimum, wipe the
pedestals with a new laboratory wipe. For more information, see Cleaning the
Pedestals.
Related Topics
• Working with standard curves
• Best practices for protein measurements
Right click
graph area
to view
display
options.
UV spectrum Menu of
table
options;
click to
choose
which
columns
to report
Click row
to select
sample and
Sample name; Protein Absorbance data for update
select to edit concentration Standard replicates spectrum.
Note
• A baseline correction is performed at 750 nm (absorbance value at 750 nm is
subtracted from absorbance values at all wavelengths in sample spectrum).
• Micro-volume absorbance measurements are normalized to a 10.0 mm
pathlength equivalent.
The R2 value indicates how well the standard curve fits the standard data points (1.0
is a perfect fit; that is, all points lie exactly on the curve).
View standard Line of best fit Curve type R2 value (1.0 equals perfect fit)
curve equation
Standard curve
Click row to
select sample
and update
graph; select
more rows to
overlay spectra
for up to five
samples.
Date/time measured
Protein conc.
Number of standard replicates
Absorbance at 595 nm
Average absorbance at
Baseline correction absorbance 595 nm for replicate
standard measurements
Equation of the standard curve
Additional monitored wavelength
Related Topics
• Example standard curve
• Basic Instrument Operations
• Protein A280 Calculations
Setting Description
Curve Type Specify type of equation used to create standard curve from standard concentration
values. Available options:
– Linear: Draws the linear least squares line through all measured standards
(requires reference measurement and at least one standard)
– Interpolation: Draws a series of straight lines to connect all measured
standards (requires reference measurement and at least one standard)
– 2nd order polynomial: Draws the 2nd order least squares polynomial
using all measured standards (requires reference measurement and at least
two standards)
– 3rd order polynomial: Draws the 3rd order least squares polynomial using
all measured standards (requires reference measurement and at least three
standards)
Replicates Enter number of measurements of the reference or the same standard or sample
that are averaged together to produce its associated concentration value.
Note: Replicates setting cannot be changed after the first standard has been
measured.
Note: Concentration values can be entered in any order but the standards must be
measured in the order they were entered.
Reported Results
Settings
Detection Limits
Before taking pedestal measurements with the NanoDrop Eight instrument, lift the
instrument arm and clean the upper and lower pedestals. At a minimum, wipe the
pedestals with a new laboratory wipe. For more information, see Cleaning the
Pedestals.
Related Topics
• Working with standard curves
• Best practices for protein measurements
• Measure a Micro-Volume Sample
• Prepare Samples and Blanks
Right click
graph area
to view
display
UV spectrum options.
Menu of
table
options;
click to
choose
which
columns
to report
Click row
to select
sample and
Sample name; Protein Absorbance data for update
select to edit concentration Standard replicates spectrum.
Note
• A baseline correction is performed at 405 nm (absorbance value at 405 nm is
subtracted from absorbance values at all wavelengths in sample spectrum).
• Micro-volume absorbance measurements are normalized to a 10.0 mm
pathlength equivalent.
The R2 value indicates how well the standard curve fits the standard data points (1.0
is a perfect fit; that is, all points lie exactly on the curve).
View standard Line of best fit Curve type R2 value (1.0 equals perfect fit)
curve equation
Standard curve
Click row to
Selected data point select sample
and update
graph; select
more rows to
overlay spectra
for up to five
samples.
Date/time measured
Protein conc.
Number of standard replicates
Absorbance at 650 nm
Average absorbance at
650 nm for replicate
Baseline correction absorbance
standard measurements
Equation of the standard curve
Additional monitored wavelength
Related Topics
• Example standard curve
• Basic Instrument Operations
Note You can edit the Curve Type setting when measuring standards by
changing the list box at the top of the application measurement screen. You can
edit the concentration value for a standard from the application setup screen.
After the first sample measurement, these settings cannot be changed.
Setting Description
Curve Type Specify type of equation used to create standard curve from standard concentration
values. Available options:
– Linear: Draws the linear least squares line through all measured standards
(requires reference measurement and at least one standard)
– Interpolation: Draws a series of straight lines to connect all measured
standards (requires reference measurement and at least one standard)
– 2nd order polynomial: Draws the 2nd order least squares polynomial
using all measured standards (requires reference measurement and at least
two standards)
– 3rd order polynomial: Draws the 3rd order least squares polynomial using
all measured standards (requires reference measurement and at least three
standards)
Replicates Enter number of measurements of the reference or the same standard or sample
that are averaged together to produce its associated concentration value.
Note: Replicates setting cannot be changed after the first standard has been
measured.
Note: Concentration values can be entered in any order but the standards must be
measured in the order they were entered.
Related Topics
• Instrument Settings
Reported Results
Settings
Detection Limits
The color produced in the assay is stable and increases in proportion to a broad
range of increasing protein concentrations. An optional Ionic Detergent Compatibility
Reagent (IDCR) may be added to the assay reagent to increase compatibility with
high amounts of ionic detergents, including Laemmli SDS sample buffer with
bromophenol blue. The IDCR dissolves completely by thorough mixing and has no
effect on the assay. Pre-formulated kits of the protein binding material are available
from us or a local distributor. For information about IDCR, refer to the kit
manufacturer.
Protein standards for generating a standard curve may also be provided by the kit
manufacturer. Since the NanoDrop Eight pedestals can measure higher protein
concentrations than traditional cuvette-based spectrophotometers, you may need to
supply your own protein standards at higher concentrations than provided by the
manufacturer. For example, additional standards may be required to ensure the
standard curve covers the dynamic range of the assay and the expected range of
the unknown samples.
Before taking pedestal measurements with the NanoDrop Eight instrument, lift the
instrument arm and clean the upper and lower pedestals. At a minimum, wipe the
pedestals with a new laboratory wipe. For more information, see Cleaning the
Pedestals.
Related Topics
• Working with standard curves
• Best practices for protein measurements
• Measure a Micro-Volume Sample
• Prepare Samples and Blanks
• Basic Instrument Operations
Right click
graph area
to view
UV spectrum display
options.
Menu of
table
options;
click to
choose
which
columns
to report
Click row
to select
sample and
Sample name; Protein Absorbance data for update
select to edit concentration Standard replicates spectrum.
Note
• A baseline correction is performed at 750 nm (absorbance value at 750 nm is
subtracted from absorbance values at all wavelengths in sample spectrum).
• Micro-volume absorbance measurements are normalized to a 10.0 mm
pathlength equivalent.
The R2 value indicates how well the standard curve fits the standard data points (1.0
is a perfect fit; that is, all points lie exactly on the curve).
View standard Line of best fit Curve type R2 value (1.0 equals perfect fit)
curve equation
Click row to
select sample
and update
graph; select
more rows to
overlay spectra
for up to five
samples.
Date/time measured
Protein conc.
Number of standard replicates
Absorbance at 660 nm
Average absorbance at
Baseline correction absorbance 660 nm for replicate
standard measurements
Equation of the standard curve
Additional monitored wavelength
Related Topics
• Example standard curve
• Basic Instrument Operations
Setting Description
Curve Type Specify type of equation used to create standard curve from standard concentration
values. Available options:
– Linear: Draws the linear least squares line through all measured standards
(requires reference measurement and at least one standard)
– Interpolation: Draws a series of straight lines to connect all measured
standards (requires reference measurement and at least one standard)
– 2nd order polynomial: Draws the 2nd order least squares polynomial
using all measured standards (requires reference measurement and at least
two standards)
– 3rd order polynomial: Draws the 3rd order least squares polynomial using
all measured standards (requires reference measurement and at least three
standards)
Replicates Enter number of measurements of the reference or the same standard or sample
that are averaged together to produce its associated concentration value.
Note: Replicates setting cannot be changed after the first standard has been
measured.
Note: Concentration values can be entered in any order but the standards must be
measured in the order in which they were entered.
Related Topics
• Instrument Settings
Measure OD600
Measure OD600
Reported Results
Settings
Calculations
Measure OD600
Use the OD600 application to monitor the growth rate of bacterial or other microbial
cell cultures by measuring the optical density (absorbance) of the culture in growth
media at 600 nm. The Beer-Lambert equation and a user-entered conversion factor
are used to correlate absorbance with concentration. Reported concentration values
can be used to identify the phase of cultured cell populations, e.g., log or
exponential and stationary.
In the case of living cells, most of the incident light is transmitted through the sample
rather than scattered, reflected or absorbed. The amount of scattered light is low
and can vary from instrument to instrument. As a result, calculated absorbance
readings are typically very low.
The calculated absorbance values are used to determine the density of cells in
solution in cells/mL. The physical concepts and formulas that relate optical
properties of living cells to concentration include:
• Cells, which have a different index of refraction from the surrounding medium,
randomly reflect and scatter light out of the incident light path. The amount of
scattering is proportional to the density of cells in the sample.
• The Beer’s Law equation is used to relate absorbance to concentration. See
Calculations for OD600 Measurements for details.
• All measurements should be made on the same type of spectrophotometer and
method (i.e., pedestal vs. cuvette) as the amount of scattered light captured
varies based on the optical configuration. When using a different
spectrophotometer or method, calculate and apply a conversion factor to the
reported results. For example, to compare OD readings using the pedestal vs. a
cuvette, a conversion factor can be calculated as follows:
Conversion factor = Cuvette OD/Pedestal OD
• Mix samples gently but thoroughly immediately before taking an aliquot for
measurement.
• For micro-volume measurements:
– Ensure pedestal surfaces are properly cleaned and conditioned.
– Avoid introducing bubbles when mixing and pipetting.
– Start the measurement promptly to avoid settling or evaporation.
– Follow best practices for micro-volume measurements.
– Use 2 µL sample volume. See Recommended Sample Volumes for more
information.
– For dilute samples that exhibit low absorbance at 600 nm, use an alternative
wavelength such as 400 nm to measure absorbance.
Before taking pedestal measurements with the NanoDrop Eight instrument, lift the
instrument arm and clean the upper and lower pedestals. At a minimum, wipe the
pedestals with a new laboratory wipe. For more information, see Cleaning the
Pedestals.
Related Topics
• Measure a Micro-Volume Sample
• Prepare Samples and Blanks
• Basic Instrument Operations
Menu of table
options;
click to choose
which columns
to report
Click row to
select sample
and update
spectrum.
Date/time measured
Related Topics
• Basic Instrument Operations
• OD600 Calculations
c = A() * factor()
where:
where:
() = molar absorption coefficient (or extinction
coefficient) at specified wavelength
b = pathlength in cm (1.0 cm for the NanoDrop Eight
instruments)
Cell number Any number User-defined factor. Generally accepted factor for
conversion factor measured cell type, or one derived empirically using a
(108) solution of study cells at known concentration using the
same media.
Related Topics
• Instrument Settings
A() absorbance
• Normalized and (absorbance) corrected (if used)
absorbance value at any specified Additional
Monitored Wavelength () is also reported.
Calculated cell concentrations are based on the Sample Pathlength
absorbance value at 600 nm, the entered factor
and the sample pathlength. A single-point • For micro-volume measurements, the software
absorbance correction may be applied. selects the optimal pathlength (between 1.0 mm and
0.1 mm) based on sample absorbance at the
analysis wavelength.
• Displayed spectra and absorbance values are
normalized to a 10 mm pathlength equivalent.
Reported Values
Custom Applications
Use the NanoDrop Eight to perform UV-Vis measurements.
Measure UV-Vis
Measures the absorbance of any
sample at up to 40 wavelengths
across the ultra-violet (UV) and visible
regions of the spectrum.
Measure UV-Vis
Reported Results
Settings
Detection Limits
Measure UV-Vis
The UV-Vis application allows the instrument to function as a conventional
spectrophotometer. Sample absorbance is displayed on the screen from 190 nm to
850 nm. Up to 40 wavelengths can be designated for absorbance monitoring and
inclusion in the report. Automatic pathlength adjustment and a single-point baseline
correction can also be used.
Before taking pedestal measurements with the NanoDrop Eight instrument, lift the
instrument arm and clean the upper and lower pedestals. At a minimum, wipe the
pedestals with a new laboratory wipe. For more information, see Cleaning the
Pedestals.
When the sample measurement is completed, the spectrum and reported values
are displayed (see the next section).
8. When you are finished measuring samples, select End Experiment .
9. Lift the arm and clean all pedestals with a new wipe.
Select to toggle
Auto-Measure
Right click graph area to view display options. ON/OFF
(Auto-Measure
default is ON
Auto-Blank
default is OFF)
Baseline
correction
wavelength;
select to edit
Sample rows
Sample
rows
Wellplate view;
toggle to view/ hide Sample names;
well plate map select to edit
Tips:
Click sample row to select sample and update spectrum
Shift-click multiple sample rows to overlay spectra
Click a sample and hover locations on spectra to view measurement values
Measure Custom
Runs a custom measurement method
created using NanoDrop Eight software.
Reported Results
Before taking measurements with the NanoDrop Eight instrument, lift the instrument
arm and clean the upper and lower pedestals. At a minimum, wipe the pedestals
with a new laboratory wipe. For more information, see Cleaning the Pedestals.
Custom
Tab
Custom
Methods
Icon
Information about the selected method appears in the method details pane.
3. Select .
4. Follow the on-screen instructions to measure a sample.
Measure sample
Method Name View history data
Run Blank End experiment
Load Sample ID file
Auto measure
UV spectrum
Custom
Tab
Custom
Methods
Icon
Note Any errors in the method will be listed in red text at the bottom of
the method editor screen. Errors must be addressed before the method
can be saved.
Notes:
• If a Baseline correction and/or Analysis wavelength correction are used,
make sure your selected spectral range includes your specified baseline
correction and/or analysis correction wavelength.
• For micro-volume absorbance measurements, the spectra are normalized
to a 10 mm pathlength equivalent.
Analysis wavelength Use this option to specify absorbance correction at analysis wavelength only.
correction Available options:
• None. No correction at analysis wavelength.
• Single point. Enter wavelength for analysis correction. (Absorbance
value at specified analysis correction wavelength is subtracted from
absorbance value at analysis wavelength. Corrected value is used to
calculate sample concentration.)
• Sloping baseline. Enter two wavelengths that define sloping baseline
for analysis correction. (Absorbance value of sloping baseline at analysis
wavelength is subtracted from absorbance value at analysis wavelength.
Corrected value is used to calculate sample concentration.)
c = (A * f) / b
where:
c = analyte concentration
A = absorbance in absorbance units (A)
f = factor (typically 1/, where = wavelength-dependent molar
absorptivity coefficient, or extinction coefficient)
b = pathlength in cm (determined at measurement time, then normalized
to 10 mm (1 cm) pathlength equivalent)
• Extinction coefficient and molecular weight. Enter extinction
coefficient for 1 cm pathlength and use adjacent drop down list to
select appropriate unit. Equation below shows how extinction coefficient
is used to calculate sample concentration:
c = A / ( * b)
where:
c = analyte concentration
A = absorbance in absorbance units (A)
= wavelength-dependent molar absorptivity coefficient (or extinction
coefficient)
b = pathlength in cm (determined at measurement time, then normalized
to 10 mm (1 cm) pathlength equivalent)
Notes:
• Refer to product literature for information about factors and extinction
coefficients for specific materials.
• To set up a method that reports absorbance measurements only, select
Factor or Extinction Coefficient with the factor or extinction coefficient set
to “1”.
• If specified unit for factor or extinction coefficient is based on mass (such
as mg/mL) and specified unit for calculated result is based on molarity
(such as pmol/µL) or vice versa, enter molecular weight and use
adjacent drop down list to select appropriate unit.
Available options:
• Predefined. Select from a list of predefined formulas, which can be
used as is or edited, and choose Add. The predefined formula is listed in
the Formula Table.
• Add. Create formula for current method. Available options:
• Formula Name. Enter a name for the formula. After a
measurement, the name is reported in Data Table and Sample Details
screens.
• Formula. Enter valid formula (see below for rules and examples).
After a measurement, the measured or calculated value is reported in
Data Table and Sample Details screens.
• Unit. Enter unit for reported result. After a measurement, the unit is
reported in Data Table and Sample Details screens.
• Edit. Edit selected formula for current method.
• Delete. Delete selected formula from current method.
Formula rules Custom formulas can include the following operators and functions:
• Path(). Returns sample pathlength in cm.
• A(nm). Returns sample absorbance at specified wavelength (for
example, enter A(650) to add the measured absorbance at 650 nm to
your equation).
• Operators: + (add), - (subtract), * (multiply), / (divide).
• Functions: Log(x), Pow(x,y).
You can now select the saved method and edit the Description and settings.
Learning Center
Contents
• Micro-Volume Sampling—How it Works 150
• Set Up the Instrument 152
• Measure a Micro-Volume Sample 154
• Sample Modes 158
• Prepare Samples and Blanks 158
• Basic Instrument Operations 164
• Acclaro Sample Intelligence 189
• Instrument Settings 195
• Measurement Screen Display Options 198
Surface Tension
Absorbance Spectrum
Sample Absorbance
Sample Concentration
Baseline Correction
Surface Tension
The NanoDrop Eight spectrophotometer uses surface
tension to hold a small volume of sample between
two pedestals (upper and lower). The patented
sample retention system enables the measurement of
highly concentrated samples without the need for
dilutions.
Absorbance Spectrum
The light passes through the liquid column to the
detector, which generates a spectrum of absorbance
versus wavelength. The spectrum shows the amount
of light absorbed by the molecules of the sample at
each measured wavelength.
A = absorbance in absorbance units (A) The pathlength is the distance between the two
pedestals, which varies in real time during each
ε = wavelength-dependent molar absorptivity measurement. This auto-ranging pathlength
coefficient (or extinction coefficient) in liter/mol-cm technique produces accurate concentration results
b = pathlength in cm over a wide dynamic range.
Baseline Correction
For some applications, the instrument can be set up
to apply a baseline correction to each measurement
to minimize any offset caused by light scattering
particulates in the sample spectra. The correction
subtracts the absorbance value at a reference
wavelength that is close to zero from the absorbance
value at each wavelength across the spectrum,
essentially “anchoring” the spectrum to zero
absorbance units at the reference wavelength.
USB-B
Power
Power
on/off
Connect Power
CAUTION Avoid shock hazard. Each wall outlet used must be equipped with a
ground. The ground must be a noncurrent-carrying wire connected to earth
ground at the main distribution box.
Connect the provided power cord to a grounded wall outlet. See “Power Cords” on
page 222 for more information.
Connect to a Computer
Connect the USB cable to the NanoDrop Eight and to an available USB port on
your PC.
Operating Specifications
The instrument operates reliably when the room environment meets these
specifications:
• operating temperatures: 5 °C - 35 °C (41 °F - 95 °F)
• relative humidity (non-condensing): 20-80%
Locate the instrument away from air vents and exhaust fans to minimize evaporation.
Note If operating the instrument at the low end of the recommended humidity
range, use adequate sample volume to avoid evaporation.
Supplies needed
• NanoDrop Eight spectrophotometer
• lint-free laboratory wipes
• calibrated precision pipettor
• sample material resuspended in appropriate buffer
solution (see Preparing Samples)
• pure buffer solution for blanking instrument (see
Choosing and Measuring a Blank.
Pipetting Samples
• Position the instrument at an angle for optimal use of the pipette guide.
• Use calibrated precision pipettor (0–2 µL volume range) with well-fitting,
low-retention precision tips to apply sample material to instrument for
measurement. If using low accuracy (0-10 µL) pipettor, use 2 µL sample
volumes.
• Filter tips are not recommended, as filter particulates can impact absorbance
measurements at 230 nm.
• Use recommended sample volumes to ensure proper liquid column formation.
• Use new tip for each blank and sample aliquot.
• Use new aliquot of sample for each measurement.
• When the measurement is complete, open the sampling arm and wipe the
samples from both the upper and lower pedestals using a soft laboratory wipe.
If solvents are used, make sure they are compatible with the pedestals. (see
“Compatible Solvents” in Hazardous Materials).
2. Lift the instrument arm and clean the upper and lower
pedestals with new laboratory wipe.
3. Measure a blank:
– Pipette 1–2 µL blanking solution onto the lower
pedestal and quickly lower the arm
Measure
– When the sample measurement is completed, the
spectra and reported values are displayed.
Sample Modes
The NanoDrop Eight can be used in either Single sample or an Eight sample mode.
Select the desired mode from the drop-down menu on the Home screen. In Single
Sample mode, measurements will only be taken using Channel A.
Module startup
Instrument self-test and preparation begins upon selection of an application after first
launching the software. The message 'Please wait - Initializing Spectrometer' will
appear. When this message disappears, the instrument will be ready for use and you
can prepare blanks. All data taken will automatically be logged in the appropriate
archive file.
Note: When using the 8 Sample mode, the software initiates each blank and
measurement cycle on the first position to be read. The user will, therefore, hear one
less position increment than expected. After making an initial blank measurement, a
straight line will appear on the individual graphs. Subsequent blanks will clear any
sample spectrum and again display straight baselines.
Preparing Samples
• Isolate and purify samples before measuring them with the instrument.
Commercial sample isolation kits are available for these purposes, or use an
in-house protocol. After purification, analyte of interest is typically dissolved in
aqueous buffer solution before it is measured.
Tip: Any molecule that absorbs light at analysis wavelength will contribute to
total absorbance value used to calculate sample concentration.
• Ensure final analyte concentration is within instrument’s absorbance detection
limits.
• For micro-volume measurements, gently (but thoroughly) vortex each sample
before taking a measurement.
Supplies needed
• lint-free laboratory wipes
• calibrated precision pipettor (0–2 µL)
• buffer solution for evaluation
Application categories
Control options
Applications
The NanoDrop Eight software offers several configurable applications, which gives
users full control of the measurement. See“Custom Applications” on page 129 for
detailed information about each available application.
System Status
Select the icon on the Home screen to open the system status box.
Control options
History: View data stored locally. Filter by date or application.
Performance: Performance verification process using PV-1 solution.
See “Performance Verification” on page 212
Intensity: Run an intensity check for the pedestal.
See “Intensity Check” on page 210.
History
Select on the Home screen to view any data acquired earlier today, last
week, last month, last six months, last year or in a specific date range. See “View
History” on page 176 for more information about the History feature.
Instrument Settings
Select on the Home screen to access instrument settings for software
updates, Protein Editor, and more. See “Instrument Settings” on page 195 for
detailed information about all available instrument settings.
Instrument Diagnostics
Instrument diagnostics (Performance and Intensity check) should be run periodically
according to the recommended maintenance schedule. See “Instrument
Diagnostics” on page 209 for information about how to run the available instrument
diagnostics.
SciVault Software
The Thermo Scientific™ SciVault™ software is available as an optional add-on. This
companion software allows users to operate their NanoDrop Eight instrument in a
manner that complies with US FDA 21 CFR Part 11. When purchased, the SciVault
software is shipped to you on a USB stick and integrates directly into the NanoDrop
Eight software user interface. For more information, visit
thermofisher.com/nanodrop.
Help
Select to launch the user guide, email a question to tech support, or find the
software logs.
Auto-
Measure
UV absorbance
spectrum for
selected samples
Table display
options
Click row to
select sample
and update
spectrum; select
more rows to
Wellplate view; Sample Sample names; overlay spectra.
toggle to view/ hide location click to edit
well plate map Measurement results; see
Applications for details
Well plate options
Find Peaks
Select Find Peaks to view calculated peaks for specified range. You can enter the
range by dragging the color-coded limit lines, or enter values into the fields at the top
of the spectrum. Found peaks for the defined range are listed in the table below the
spectrum.
UV spectrum
Peak 1
Table of calculated
peaks for the
Maximum number of peaks specified range
Sample Name
Click the Sample Name field in any measurement screen to edit the sample name.
In Single Sample mode, each sample is given a default base name “sample”
followed by the number sample in the sequence. For example, the first sample
would be named “Sample 1” followed by “Sample 2,” etc. You can edit the default
base name and overwrite any sample name.
In 8 Sample mode, each sample is automatically labeled by the sample location. Add
sample names by clicking the sample name field, or by importing Sample ID file.
Note If you edit the sample base name during an experiment when
Auto-Naming is selected, the assigned sample ID numbers restart.
After you measure a blank and before the first sample is measured:
– click Sample Name field
– enter new base name
– press Enter key
Measurement Results
The types of results that appear in the measurement screens depend on the
selected application. For details, see the reported results section of that application
in this guide:
Applications > [application group] > Measure [application name] > Reported
Results
Absorbance Spectrum
For each measured sample, each application shows the UV or
UV-Visible absorbance spectrum and a summary of the results. The vertical axis
shows absorbance in absorbance units (A). The horizontal axis shows wavelength in
nm. Here is an example for a UV-Vis method.
Sample Pathlength
All applications display the sample pathlength along the spectrum’s vertical axis.
Micro-volume absorbance measurements are normalized to a 10.0 mm pathlength
equivalent.
Measurement Alerts
The Acclaro Sample Intelligence technology built into the NanoDrop Eight instrument
provides important features to help you assess sample integrity. click a Sample
Intelligence icon in the software to view its associated information.
contaminant analysis is available to help qualify a sample before use
in downstream applications
Blank Button
Select Blank to measure a blank for the selected experiment.
A blank must be measured before each group of similar samples. The blank solution
is typically the pure buffer that was used to resuspend the sample. For more
information, see Choosing and Measuring a Blank.
Measure Button
Select Measure to measure a sample for the selected experiment.
Samples must be properly isolated and prepared before they can be measured with
the instrument and the concentration must be within the instrument’s absorbance
detection limits. For more information, see Preparing Samples, Measure a
Micro-Volume Sample, and Absorbance Detection Limits.
Note The Measure button is enabled after a valid blank measurement is
completed.
Auto-Measure
To select or deselect Auto-Measure, from any sample measurement screen, click the
On or Off button at the right of the Measure button.
Auto-Blank
To select or deselect Auto-Blank, from any blank measurement screen, click the On
or Off button at the right of the Blank button.
Note The End Experiment button is enabled after the first sample
measurement is completed.
After you select End Experiment, the End Experiment dialog box is displayed:
Click Save and you will have the option to Print or Export the experiment data.
Available options:
Cancel (Return To Close the End Experiment box and display the results for
Experiment) the most recent measurement. From there you can add
measurements to the current experiment and save it later.
Print button Print measurement results for current experiment
Save (End Experiment) End the experiment and save the measurement results
using the entered experiment name. The experiment is
saved in the selected database location.
Sample Details
Sample details are shown in the sample row in any measurement screen or
History. This information includes all available measurement results and associated
details for the selected sample. To add or remove results shown in the sample row,
click the menu in the upper right corner of the data table. Information about the
measured values displayed in Sample Details is provided in this Help system, under
the application used to acquire the data.
Note You can also edit the sample name by clicking in the name field of the
sample.
Data Table
The data table for the current experiment is shown below the spectrum graph. The
data table contains the measurement results for all samples in the experiment. The
image below highlights the available features.
NOTE: This feature works most efficiently when a user imports sample names from
a file.
NanoDrop Eight
tube rack accessory
The wells/tubes which should be measured next are filled with bright orange in the
software user interface. They correspond to the wells/tubes illuminated by the green
LEDs on the top of the instrument. Wells/tubes that have sample IDs pre-populated
will be filled with light orange. Wells/tubes that have already been measured will
either be filled bright green (indicating no sample alerts) or with an A in a circle
(indicating an Acclaro alert is present for that sample).
Sample group selector
Show/hide well
plate map View options menu
Measurement alert
Measurement alert;
Acclaro warning
Currently active
channel measurement
Sample channels completed with no
errors
Currently active
channel with sample
IDs populated
Dark green
indicates
Measurement Dark orange indicates
completed with inactive well with
no errors Well plate map shown in the measurement screen sample ID populated
After a set of measurements is complete, users can manually advance the sample
names and well illumination by clicking the and arrows above the well plate
map in the bottom left. This will prepare the instrument software for the next set of
measurements. Alternatively, users can click the 3-dot menu above the well plate
map and enable Auto Advance. Auto Advance sets the software to automatically
advance the columns a few seconds after the last measurement is made.
The entire well plate map interface can be hidden by clicking on the plate icon .
Measurement
alert; click to
learn more
View History
Whether you collect one sample or many in a row, after you choose End Experiment,
the acquired data are automatically saved in an experiment with an experiment
name. In the default configuration, experiments are stored in the NanoDrop Eight
database according to acquisition date, experiment name, application used and any
assigned labels.
Use the History feature to open the database in order to view acquired spectra and
associated data from any experiment at any time.
History Menu
Available options:
Measurement Menu
When viewing measurement data from History, Click the menu in the upper right of
the measurement screen to access available menu options.
Search
Search
Open an experiment
– In History, if you don’t see the experiment you want to open, you can use the
Search feature to find the desired experiment
– click the experiment name to open the experiment
The History provides measurement data as spectral data and data tables, similar to
what you see after you complete a measurement.
Note The data shown are dependent upon the application used to measure the
samples (nucleic acids in these examples). For more information, see the
application details.
Spectral data—
After you open an experiment, the software shows the UV or UV-visible absorbance
spectrum and associated data for the selected experiment, much like it appears
during a measurement. The image below describes the available features.
UV spectrum for
selected sample(s)
Data Table—
The lower area of the measurement history screen (shown above) shows the data
table for the current experiment. The data table contains the measurement results
for all samples in the experiment. The image below describes the available features.
Menu
Select the menu in the upper right corner from any Spectral Data or Data Table
screen to see the available menu options.
Home Return to NanoDrop Eight Home screen
Manage Identifiers Add or delete labels for selected experiment to make it
easier to find (see Manage identifiers on the instrument)
Export Export selected experiments
Print Print plot or data table for selected measurement results;
if no results are selected, prints all results in data table
Settings View or change instrument settings
Manage Identifiers
You can add one or more “tags” (i.e., labels or metadata tags) to an experiment to
make the experiment easier to find.
Use the History to add labels to experiments, assign existing labels, view assigned
labels and remove or delete labels on the instrument. You can filter the list of
experiments in the History based on one or more user-defined labels.
– in End Experiment box, enter user defined labels in the Tags field
– click Save
Remove a label
– from Home screen, select History
– Click the menu of the experiment row and select Tags
– in Manage Tags box, click the X for any tag you wish to delete
– Click OK
– enter a name for this group of measurements in the Experiment Name box
– click Save
Use any spreadsheet or word processing application to open a CSV or TSV file. Here
is an example of several sample measurement results in CSV format:
Note The types of data exported are dependent upon the application used to
measure the samples (nucleic acids in this example). For more information, see
the application details.
Data can be exported to any file path connected to the PC or to a network location.
If you select multiple experiments for export, each exported experiment has a
corresponding file. The filenames are the same as the experiment names. The files
are stored in a folder named “NanodropEight” followed by the instrument serial
number. (Use System Status to view your instrument serial number.)
– click Export
– choose Print
The selected measurement screen is printed for each selected measurement.
– In Print Review, select the data you want to include in the data tables.
– choose Print
The selected measurement screen is printed for each selected measurement.
Activate Detection
From the measurement screen, click the activate contaminant detection option.
The icons appear in the Sample Plate Map, the data table, and in History (see
below).
The icons are active in all three places; the information remains with the data
indefinitely, even after it has been exported.
Contaminant Analysis
For the dsDNA, RNA and Protein A280 applications, the NanoDrop Eight software
automatically initiates a spectral analysis for several known contaminants during the
measurement. Examples of known contaminants include:
• for dsDNA and RNA measurements:
– in the analysis region: protein and phenol
– detects presence of guanidine HCl and guanidinium isothiocyanate
– detects species-specific dsDNA contamination in the RNA application and
detects species-specific RNA contamination is in the dsDNA application
• for protein measurements:
– in the analysis region: nucleic acids and phenol
Click the icon to view the contaminant analysis and associated information.
The following is an example of results from a nucleic acid contaminant analysis that
contains enough protein contaminant to affect the measurement results.
Phenol spectrum
Since phenols absorb light near the analysis wavelengths for nucleic acid (230 nm,
260 nm, and 280 nm), the presence of phenol in the nucleic acid sample shown
above has pushed the A260/A280 and A260/A230 ratios out of range and caused
the reported nucleic acid concentration to be higher than the real value. The
software identifies the impurity (phenol), and reports the following:
• baseline-corrected absorbance due to phenol (2.53) at the analysis wavelength
(260 nm)
• % coefficient of variation for the measurement result (uncertainty x
100/measurement result = 4.22%; a high %CV indicates the measurement result
is close to the instrument detection limit or there is an interfering component)
• original nucleic acid concentration (462.6 ng/µL), which is based on the total
baseline-corrected absorbance (sample plus contaminant) at the analysis
wavelength
• corrected nucleic acid concentration (286.4 ng/µL), which is based on the
corrected absorbance (sample minus contaminant) at the analysis wavelength
Traditionally, purity ratios are used to detect the presence of contaminants that could
affect downstream applications. However, purity ratios do not always provide a
complete picture of possible contamination. When a purity ratio falls outside the
expected range, the spectral profile is often examined qualitatively.
Here are results from a nucleic acid analysis for which measured purity ratios are
above the expected value
Click the More Information icon for the next level of information
Instrument Settings
From the Home screen, select Settings. Available options include System Settings,
Dye Editor, Protein Editor, and Preferences.
System Settings
These options are available:
Database Backup Set database backup and restore options such as file
& Restore path and data and backup schedule.
From the Settings screen, in the Database Backup & Restore region, select
Configure.
To restore data from a previous backup event, from Database Maintenance, select
Restore and browse for the database backup you wish to use.
Once you have selected your backup, click the Restore button.
Preferences
Set your preferences for remembering certain settings when restarting the
application.
Dye Editor
See “Dye/chromophore editor” on page 32 for information on using the dye editor.
Protein Editor
See “Protein editor” on page 63 for information on using the protein editor.
Find Peaks
Select Find Peaks to view calculated peaks for specified range. You can enter the
range by dragging the color-coded limit lines, or enter values into the fields at the top
of the spectrum. Found peaks for the defined range are listed in the table below the
spectrum.
Wavelength start Wavelength end Absorbance minimum
UV spectrum
Peak 1
Table of calculated
peaks for the
Maximum number of peaks specified range
Maintenance
• Maintenance Schedule 202
• Maintaining the Pedestals 203
• Decontaminating the Instrument 207
• Instrument Diagnostics 209
Maintenance Schedule
Daily Maintenance
• Clean pedestals with deionized water
Periodic Maintenance
• Clean and recondition pedestals
Every 6 Months
• Clean and recondition pedestals
• Run Intensity Check
• Run Performance Verification
Supplies needed
• lint-free laboratory wipes
• deionized water (DI H2O)
• for thorough cleaning: PR-1 kit
If samples flatten out on the pedestal (rather than “beading up” or forming a rounded
droplet) or the liquid column breaks during a measurement, recondition the
pedestals.
Supplies needed
• lint-free laboratory wipes
• PR-1 pedestal reconditioning kit (available from us or a local distributor)
• calibrated precision pipettor (0-2 µL)
• canned air
Note
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will
permanently damage the quartz fiber optic cables.
• To prevent damage from spills, keep containers of liquids away from the
instrument.
• Do not use a squirt or spray bottle on or near the instrument as liquids may
flow into the instrument and may cause permanent damage.
• Do not attempt to remove the diaphragm around the lower pedestals as it is
permanently affixed to the instrument.
• Do not allow HCl, alcohol, bleach, acetone or any other solvent to remain on
the diaphragm for more than one minute or it may loosen the seals. If the
diaphragm becomes loose, contact us.
•
Supplies needed
• lint-free laboratory wipes
• deionized water (DI H2O)
• 0.5% sodium hypochlorite solution (1:10 dilution of commercial bleach, freshly
prepared)
• pipettor
Instrument Diagnostics
Every 6 months, run the following performance and quality checks to verify
instrument operation.
Intensity Check
Run Intensity Check every 6 months to verify operation of the instrument’s internal
components. The test measures the intensity of light from the xenon source through
the instrument to verify that throughput, wavelength accuracy, and bias are within
specifications. The test is automatically performed using the pedestal paths.
Supplies needed
• lint-free laboratory wipes
Channel table
The results will display the lamp intensity using the UV and Visible integration times in
two separate graphs (see image above).
Users can see the intensity graphs of a single channel by clicking a single row from
the Channel table or Shift + Click to highlight multiple rows (see image to the right).
A green check mark and banner will be displayed at the top if the UV intensity, Visible
intensity, bias, and maximum single offset are all within specification. If one or more
is not within specification, a message will be displayed at the top indicating what
parameter(s) failed. If any parameter has failed, clean the pedestals with deionized
water and then repeat the Intensity Check.
If a yellow triangle appears next to the Bias indicator, make sure the room is within
the temperature specifications for the instrument.
Performance Verification
Run Performance Verification every 6 months to confirm pathlength accuracy is
within specifications.
A CHEM-PV-8 kit containing two vials of PV-1 {aqueous nicotinic acid (C6H5NO2),
potassium nitrate (KNO3)} is required to verify the performance of the Thermo
Scientific™ NanoDrop™ Eight microvolume UV-Vis spectrophotometer.
Supplies needed
• lint-free laboratory wipes
• deionized water (DI H2O)
• Calibrated Precision 8-channel Pipettor (0.5 - 10 µL)
• CHEM-PV-8 kit containing two ampoules of PV-1, two 8-well PCR strip tubes,
and transfer pipettes
• laboratory gloves
Note The PV-1 solution comes in a single-use ampoule. Before you open the
ampoule, shake it vigorously and then allow the liquid to collect in the bottom
portion of the ampoule. After the ampoule is opened, its contents must be used
within one hour. Pipette directly from the ampoule; do not transfer the solution.
First make sure the pedestals are properly conditioned. To test pedestal
conditioning, clean the upper and lower pedestal surfaces with a dry, lint free
laboratory wipe, then pipette 1.5 µL DI H2O onto the eight lower pedestal surfaces.
The droplets should “bead up” as shown below. If it does not, recondition the
pedestals. Remove the water sample from the upper and lower pedestal surfaces
with a dry laboratory wipe.
Droplets “bead up” on properly
conditioned pedestals
It is very important to deliver the PV-1 aliquots to all eight pedestals in a single
motion using a multichannel pipettor. It is highly recommended that the steps below
be practiced using water before starting the calibration check procedure.
1. When the pipette tips are close to the measurement pedestals, discharge the
fluid and allow the drops to hang on the end of the tips.
2. Gently touch the droplets to the pedestals and allow them to be pulled off the
tips and onto the pedestals by surface tension.
Tip: It is important to use tips with a rigid structure to ensure the tips do not
splay or skew when delivering the samples.
Tip: It is suggested that this technique be practiced until consistent delivery to all
eight positions is routine and quick.
Tip: It is important to use tips with a rigid structure to ensure the tips do not
splay or skew when delivering the samples.
8. Remove water from upper and lower pedestal surfaces using a clean, dry
laboratory wipe.
9. Ensure PV-1 solution is thoroughly mixed by vigorously shaking both ampoules.
Allow the solution to collect in the bottom portion of the ampoule, if needed
gently tap the ampoule.
10. Carefully snap off top portion of each ampoule using an ampoule cracker,
discard top along with ampoule cracker (use proper safety precautions for
disposal).
11. Use the provided transfer pipette to dispense the contents of both PV-1
ampoules into each well of the 8-well PCR strip tube.
12. Using an 8-channel pipettor, withdraw 1.5 µL of PV-1 solution from each of the
8- wells of the PCR strip tube and pipette onto each of the 8 lower pedestals.
Lower arm.
Note If Auto-Measure feature is OFF, click Measure to begin the
measurement. If Auto-Measure feature is ON, the measurement will begin
automatically after the arm is lowered.
13. After the measurement is complete, remove sample from both upper and lower
pedestals using a dry laboratory wipe.
14. Repeat steps 11 and 12 to measure five additional individual replicates of the
PV-1 solution.
a. Always use fresh pipette tips to remove each aliquot and use a fresh aliquot
of PV-1 for each measurement.
b. In between each measurement, remove PV-1 solution from all 16 pedestals,
upper and lower, using a clean, dry laboratory wipe.
15. After each measurement is complete, the individual results will be displayed on
screen and subsequently added to the existing results.
16. After six replicates have been measured, the Status columns for each
pathlength will indicate whether the channel passed the Performance
Verification.
b. The Experiment name can be changed at this time and Identifiers can be
added.
3. Click End Experiment when done.
4. To review results from a previous Performance Verification check, select History
from the Home screen and locate the Performance Verification check results
from the list of experiments.
Contents
• Operating Precautions 218
• Safety Information 219
Note Be sure that all persons operating this system read the safety manual first.
Operating Precautions
CAUTION Do not remove the instrument cover. Removing the cover exposes
the operator to sharp edges and delicate fiber optic cables. The instrument
warranty is void if the cover has been removed.
Safety Information
Before operating a NanoDrop Eight instrument, please read the safety information
and follow its recommendations for the system.
Note Follow instructions with this label to avoid damaging the system hardware
or losing data.
The following table lists some of the safety symbols and their indications that may
appear in the user documentation.
Symbols Indication
This is a mandatory action symbol. It is used to indicate that an
action shall be taken to avoid a hazard.
Symbol Description
Alternating current
Earth terminal or ground
Direct current
Protective conductor terminal
Frame or chassis terminal
Fuse
Power on
Power off
CAUTION Avoid shock hazard. Do not remove the cover of the instrument. All
service to the instrument must be performed by trained personnel.
When the instrument arrives, check the exterior of the shipping box for signs of
damage. If damage is apparent, contact us or your local distributor for instructions.
• Move the shipping box to the installation location at least 24 hours before
installation.
Note
• Inside the shipping box, the instrument is sealed in a plastic bag to keep
the unit dry.
• Allow 24 hours for the instrument to reach room temperature before
opening the bag. If the bag is opened before the instrument reaches
room temperature, moisture could condense on the optical components
and cause permanent damage.
Note It is important to have all system utilities installed before the instrument
arrives. Utility installations must comply with all local building and safety
codes.
If you suspect power quality problems at your site, or if your system will be installed
in a heavy industrial environment, we recommend a power quality audit before
installation. Contact us or your local electrical authority for more information.
CAUTION Avoid shock hazard.
• Only a qualified person using the appropriate measuring device shall check
the line voltage, current and frequency.
• Only our trained and certified service representatives shall attempt to service
a component that carries this symbol.
• If a protective cover on a system component appears damaged, turn off the
system and secure it against any unintended operation. Always examine the
protective cover for transport stresses after shipping.
• Even after this instrument has been disconnected from all voltage sources,
capacitors may remain charged for up to 30 seconds and can cause an
electrical shock.
• Do not allow liquid to run over or into any surface where it may gain entry into
the instrument.
• Do not attempt to remove the cover of the instrument.
Grounding
CAUTION Avoid shock hazard. Each wall outlet used must be equipped with a
ground. The ground must be a noncurrent-carrying wire connected to earth
ground at the main distribution box.
Power Cords
Be sure to use an appropriate grounded power cord for your electrical service. If the
power cord received is not appropriate for the electrical system in your location, or if
the power cord becomes damaged, contact us.
A UPS reduces the probability of a system shutdown if power is lost elsewhere in the
building. Power line conditioners (which ensure that your service is free from sags,
surges or other line disturbances) also are available in the U.S.A. from us for 120 volt
operation. Line conditioners for 220 volt operation can be purchased locally. Contact
technical support for information about power conditioners and UPS.
Power Consumption
Generally, 50% more power should be available than the entire system (including
accessories) typically uses. Maximum power consumption and heat dissipation
specifications for the spectrometer and accessories are shown below. The values
are approximate.
Item Power Consumption Max. Heat Dissipation
instrument 60 W 205 Btu/hr
Optical Safety
This instrument was designed with a protective housing to prevent user exposure to
ultraviolet light.
WARNING Avoid personal injury. Never look at the lamp while illuminated.
Hazardous Materials
Many standard spectroscopy methods are based on the use of solvents. Others
involve corrosive samples or pressurized samples in a gaseous state.
Compatible Solvents
Most solvents typically used in life science laboratories are compatible with the fiber
optic pedestals of all NanoDrop spectrophotometers. However, the high vapor
pressure properties of some solvents may not be conducive to small volume
measurements when using the pedestal for measurements on any of the NanoDrop
instruments. If you are measuring samples with high vapor pressures, use an
instrument with provision for measuring samples in cuvettes.
The following solvents are compatible for use on the pedestals of all NanoDrop
instruments.
Note Spillage of these solvents on surfaces other than the pedestals may
damage the instrument.
The diaphragm around the pedestal of the NanoDrop Eight is permanently affixed to
the instrument. Do not attempt to remove the diaphragm or break the seal. Avoid
prolonged exposure of the diaphragm to HCl, alcohol, bleach, acetone or other
solvents as the adhesive securing the seal may be affected. If the seal comes loose
please contact us.
Note All forms of Hydrofluoric Acid (HF) are incompatible as the fluoride ion will
etch the fiber optic cable.
Biological samples such as tissues, body fluids, infectious agents, and blood of
humans and other animals have the potential to transmit infectious diseases. Wear
appropriate protective equipment. Individuals should be trained according to
applicable regulatory and organization requirements before working with potentially
infectious materials. Follow your organization’s Biosafety Program protocols for
working with and/or handling potentially infectious materials.