AC/DC Medical Power Supply TPP 30-J Series, 30 Watt
• High power density power supply
(encapsulated)
• Certification according to IEC/EN/ES
60601-1 3rd edition for 2 x MOPP
• Low leakage current <75 µA rated for
BF applications
• EMC compliance to IEC 60601-1-2 4th
edition
• Risk management process according to
ISO 14971 incl. risk management file
• Acceptance criteria for electronic
assemblies acc. to IPC-A-610 Level 3
• Protection class ll
ES 60601-1 IEC 60601-1
• Operating up to 5000 m altitude UL 62368-1 IEC 60335-1
IEC 62368-1
• Ready to meet ErP directive, no load
power consumption <60 mW The TPP 30-J AC/DC power supplies feature a reinforced double I/O isolation
system according to medical safety standards IEC/EN/ES 60601-1 3rd edition
• 5-year product warranty for 2 x MOPP approved for an operating altitude of 5000 m. The earth leakage
current is below 75 µA what makes the units suitable for BF (body floating) ap-
plications. The excellent efficiency of up to 91.5% offers a high power density
in the packaging format 1.5” x 3.95”. The full load operating temperature range
covers –40°C to +60°C while it goes up to 85°C with 50% load derating. The
units operate in compliance to the medical EMC emission and immunity levels
according to latest standard IEC 60601-1-2 4th edition.
Models
Order Code Output Power Output Voltage Output Current Efficiency
max. nom. max. typ.
TPP 30-103-J 20 W 3.3 VDC 6'000 mA 84 %
TPP 30-105-J 5 VDC 6'000 mA 87 %
TPP 30-109-J 9 VDC 3'340 mA 88 %
TPP 30-112-J 12 VDC 2'500 mA 91 %
TPP 30-115-J 30 W 15 VDC 2'000 mA 91 %
TPP 30-124-J 24 VDC 1'250 mA 90 %
TPP 30-136-J 36 VDC 840 mA 90 %
TPP 30-148-J 48 VDC 630 mA 92 %
[Link] August 7, 2024 Page 1 / 5
TPP 30-J Series, 30 Watt
Input Specifications
Input Voltage - AC Range Operational Range: 85 - 264 VAC (Full Range)
Rated Range: 100 - 240 VAC (Full Range)
- DC Range Operational Range: 120 - 370 VDC (Designed for, no certification)
Polarity: +DC: L / –DC: N
Input Frequency Operational Range: 47 - 440 Hz
Certified: 50/60 Hz
Power Consumption - No load & Vin = 230 VAC 60 mW max. (Ready to meet ErP directive)
- No load & Vin = 115 VAC 60 mW max.
Input Current - Full load & Vin = 230 VAC 400 mA max.
- Full load & Vin = 115 VAC 800 mA max.
Input Inrush Current - At 230 VAC 40 A max.
- At 115 VAC 25 A max.
Input Protection T 1.6 A / 250 VAC (Internal Fuse in L & N)
Recommended Input Fuse (The need of an external fuse has to be assessed
in the final application.)
Output Specifications
Voltage Set Accuracy ±1% max.
Regulation - Input Variation (Vmin - Vmax) 0.2% max.
- Load Variation (0 - 100%) 0.7% max. (3.3 and 5 VDC model)
0.5% max. (other output models)
Ripple and Noise 3.3 VDC model: 50 mVp-p typ. (w/ 10 µF X7R)
(20 MHz Bandwidth) 5 VDC model: 50 mVp-p typ. (w/ 10 µF X7R)
9 VDC model: 50 mVp-p typ. (w/ 10 µF X7R)
12 VDC model: 50 mVp-p typ. (w/ 1 µF X7R)
15 VDC model: 50 mVp-p typ. (w/ 1 µF X7R)
24 VDC model: 50 mVp-p typ. (w/ 1 µF X7R)
36 VDC model: 50 mVp-p typ. (w/ 1 µF X7R)
48 VDC model: 50 mVp-p typ. (w/ 0.1 µF X7R)
Capacitive Load 3.3 VDC model: 10'000 µF max.
5 VDC model: 12'000 µF max.
9 VDC model: 3'720 µF max.
12 VDC model: 2'085 µF max.
15 VDC model: 1'350 µF max.
24 VDC model: 520 µF max.
36 VDC model: 235 µF max.
48 VDC model: 130 µF max.
Minimum Load Not required
Temperature Coefficient ±0.02 %/K max.
Hold-up Time - At 115 VAC 16 ms min.
Start-up Time - At 230 VAC 1'500 ms max.
Short Circuit Protection Continuous, Automatic recovery
Output Current Limitation 110 - 170% of Iout max.
140% typ. of Iout max.
Overvoltage Protection 125 - 140% of Vout nom.
Transient Response - Response Deviation 3% max. (50% to 75% Load Step)
- Response Time 500 µs typ. (50% to 75% Load Step)
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
[Link] August 7, 2024 Page 2 / 5
TPP 30-J Series, 30 Watt
Safety Specifications
Standards - IT / Multimedia Equipment EN 62368-1
IEC 62368-1
UL 62368-1
- Household EN 60335-1
IEC 60335-1
- Medical Equipment EN 60601-1
IEC 60601-1
ANSI/AAMI ES 60601-1
2 x MOPP (Means Of Patient Protection)
- Power Transformers IEC 61558-1
IEC 61558-2-16
- Certification Documents [Link]/overview/tpp30-j
Protection Class Class I & II (Prepared): Reinforced Insulation
Pollution Degree PD 2
Over Voltage Category OVC II
EMC Specifications
EMI (Emissions) EN 60601-1-2 edition 4 (Medical Devices)
- Conducted Emissions EN 55011 class B (internal filter)
EN 55014-1 (internal filter)
EN 55032 class B (internal filter)
FCC 47 Part 15 class B (internal filter)
FCC 47 Part 18 class B (internal filter)
- Radiated Emissions EN 55011 class B (internal filter)
EN 55014-1 (internal filter)
EN 55032 class B (internal filter)
FCC 47 Part 15 class B (internal filter)
FCC 47 Part 18 class B (internal filter)
- Harmonic Current Emissions EN 61000-3-2, class A
- Voltage Fluctuations & Flicker EN 61000-3-3
EMS (Immunity) EN 60601-1-2 edition 4 (Medical Devices)
EN 55024 (IT Equipment)
EN 55035 (Multimedia)
EN 55014-2 (Household Appliances Tools)
- Electrostatic Discharge Air: EN 61000-4-2, ±15 kV, perf. criteria A
Contact: EN 61000-4-2, ±8 kV, perf. criteria A
- RF Electromagnetic Field EN 61000-4-3, 20 V/m, perf. criteria A
- EFT (Burst) / Surge EN 61000-4-4, ±2 kV, perf. criteria A
L to L: EN 61000-4-5, ±1 kV, perf. criteria A
- Conducted RF Disturbances EN 61000-4-6, 20 Vrms, perf. criteria A
- PF Magnetic Field Continuous: EN 61000-4-8, 30 A/m, perf. criteria A
- Voltage Dips & Interruptions 230 VAC / 50 Hz: EN 61000-4-11
30%, 25 periods, perf. criteria A
60%, 1 period, perf. criteria A
>95%, 1 period, perf. criteria A
>95%, 250 periods, perf. criteria A
115 VAC / 60 Hz: EN 61000-4-11
30%, 25 periods, perf. criteria A
60%, 1 period, perf. criteria A
>95%, 1 period, perf. criteria A
>95%, 250 periods, perf. criteria A
EMC / Environmental - Certification Documents [Link]/overview/tpp30-j
General Specifications
Relative Humidity 95% max. (non condensing)
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
[Link] August 7, 2024 Page 3 / 5
TPP 30-J Series, 30 Watt
Temperature Ranges - Operating Temperature –40°C to +85°C
- Storage Temperature –40°C to +85°C
Power Derating - High Temperature Depending on model
- Low Input Voltage 4 %/V below 90 VAC
See application note: [Link]/overview/tpp30-j
Cooling System Natural convection (20 LFM)
Altitude During Operation 5'000 m max.
Regulator Topology Flyback Converter
Switching Frequency 30 - 60 kHz (PWM)
Insulation System Reinforced Insulation
Working Voltage (rated) 272 VAC
Isolation Test Voltage - Input to Output, 60 s 4'000 VAC
- Input to Case or PE, 60 s 1'500 VAC
- Output to Case or PE, 60 s 1'500 VAC
Creepage - Input to Output 8 mm min.
Clearance - Input to Output 8 mm min.
Isolation Resistance - Input to Output, 500 VDC 100 MΩ min.
Leakage Current - Touch Current 75 µA max.
(at 264 VAC)
Reliability - Calculated MTBF 3'300'000 h (MIL-HDBK-217F, ground benign)
Environment - Vibration IEC 60068-2-6
- Mechanical Shock IEC 60068-2-27
Potting Material Silicone (UL 94 V-0 rated)
Housing Type Plastic Case
Mounting Type Chassis Mount
Connection Type Pin Connector
Weight 119 g
Environmental Compliance - REACH Declaration [Link]/info/[Link]
REACH SVHC list compliant
REACH Annex XVII compliant
- RoHS Declaration [Link]/info/[Link]
Exemptions: 7a, 7c-I
(RoHS exemptions refer to the component
concentration only, not to the overall
concentration in the product (O5A rule).)
- SCIP Reference Number b88c13b5-90cb-454f-b1fa-a544a20a0f87
Supporting Documents
Overview Link (for additional Documents) [Link]/overview/tpp30-j
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
[Link] August 7, 2024 Page 4 / 5
TPP 30-J Series, 30 Watt
Outline Dimensions
27.94 (1.100)
38.1 (1.50)
Pin connectors
100.3 (3.95) Input Output
88.90 (3.500) Pin Function Pin Function
1 Line 1 +Vout
3 Neutral 2 –Vout
Output
1 Input: JST series
2
Input
Top view mates with JST crimp terminal: SVH-21T-P1.1
1 and terminal housing: VHR-3N
3
Output: JST series
mates with JST crimp terminal: SVH-21T-P1.1
and terminal housing: VHR-2N
3 )
.1
(0
4
3.
Ø
25.4 (1.00)
Front view
Dimension in mm (inch)
Tolerances: x.x ±0.50 (±0.02)
[Link] ±0.25 (±0.01)
Mounting screw locked torque: 5.0 kgfcm / 0.49 Nm
© Copyright 2024 Traco Electronic AG Specifications can be changed without notice.
Rev. August 7, 2024
[Link] Page 5 / 5