Determination of sugar inversion rate constant by polarimetric

measurement
Theoretical background: Atkins’ Physical Chemistry
Type of the practice: individual
Purpose of the practice: To investigate the decomposition of sucrose under pseudo-first-order conditions and
to study the operation of polarimeters in measuring optical activity.

1 Introduction
Sucrose or common sugar (cane sugar, beet sugar, maple sugar) is a disaccharide composed of glucose and
fructose. It is produced by plants and is an important food for heterotrophic organisms. Edible sugar, known
as granulated sugar, is made primarily from sugar beet or sugar cane.
The decomposition of sucrose to glucose (D-glucose) and fructose (D-fructose) in a neutral aqueous
medium is a slow process (Fig. 1). The reaction is catalyzed by acids.

Figure 1: Decomposition of sucrose and specific rotations of the components at 589 nm.

In the reaction, each saccharide is optically active, i.e., it rotates the plane of polarization of light. Light,
as an electromagnetic wave, involves propagating electric and magnetic field oscillating perpendicular to
each other. The plane of oscillation of light emitted from a natural light source can be any. Some devices

Figure 2: Sketch of a monochromatic electromagnetic wave polarized in a plane (linearly).

1
Figure 3: Plane-polarized light as the sum of two circularly polarized lights.

(so-called polarizers) select waves that oscillate in one plane, making the transmitted light plane-polarized
(linearly polarized, Fig. 2). These can be birefringence prisms, or polaroid films, etc.
The interaction of polarized light with an optically active substance (asymmetric crystals or molecules)
causes the plane of polarization to rotate (to the right or left), and this rotation can be detected by another
polarizer (analyzer). To illustrate this, it is best to consider plane-polarized light as the sum of two circularly
polarized lights (Fig. 3). Depending on the direction of rotation of the circular polarization, the interaction
of the optically active substance (left-handed or right-handed, i.e., optical isomers) and light, the circularly
polarized light rotating left and right, "detects" a medium with different refractive indices, so it passes through
the medium with different speeds. Their sum will be a polarized light, but the plane of polarization will be
rotated. This angle of rotation can be detected with another polarizer, the analyzer. The essence of the
detection is that the analyzer transmits the most light when its optical axis is in the plane of polarization, and
the least light when it is perpendicular to it. (See Fig. 4 and Fig. 5).
The optically active material changes this angle between the polarizer and the analyzer for the highest
intensity or total extinction. Angle of rotation caused by the optically active material (θ) is proportional to
the number of particles (concentration, c), with optical path length, (`) and depends on the temperature (T)
and the wavelength of light (λ) is
(T,λ)
θ = c · ` · [Θ]m
(T,λ)
where the proportionality factor is the molar rotation ([Θ]m ). When several types of optically active

Figure 4: Detection of the angle of rotation of light, the theoretical structure of the polarimeter.

2
Figure 5: Polarizer and analyzer.

substances are present together, their total effect is a simple sum of the rotations, i.e.,
n
(T,λ)
θ = ∑ ci · ` · [Θ]i .
i=1

Often, instead of the molar concentration, we use the mass concentration (in g/cm3 units), the optical
path length is expressed in dm, and thus, the so-called specific rotation is the proportionality factor. Its unit
is (◦ cm3 )/(g dm). In the case of various saccharides their values can be seen in Fig. 1. Specific rotation in
the decomposition of sucrose changes from +66.5◦ to −39.7◦ i.e., changes sign, that is why the reaction is
called inversion. (Products rotate light in a different direction than the starting material.)
During the reaction, the (α)-D-, (β)-D-glucose equilibrium is reached very fast, practically all the way to
their average value (D-glucose). The concentration of water is very high, so the amount changes negligibly
by the hydrolysis, and the oxonium ion is recovered as a catalyst during the reaction, i.e., the change in the
concentration of only one component has to be taken into account, the reaction is "apparently" first order,
i.e., pseudo-first order reaction.

1.1 Evaluation of first-order kinetics by polarimetry


For a kinetically first-order process (e.g., A→ Product(s) transformation) the differential form of the rate
equation is
d[A]
− = k[A]
dt
The integrated form is
[A] = [A]0 e−kt
where [A] is the actual, while [A]0 is the initial concentration, k is the rate coefficient and t the time. Initially,
for the reaction under study (t = 0 s) there is only sucrose (s) in the solution, so the initial rotation is

θ0 = [s]0 · ` · [Θ]s .

During the reaction, three components are present, sucrose (s), glucose (g), and fructose (f), so

θt = [s] · ` · [Θ]s + [g] · ` · [Θ]g + [f] · ` · [Θ]f .

It follows from the stoichiometry that [g] = [f] = [s]0 − [s], so

θt = [s] · ` · [Θ]s + ([s]0 − [s]) · ` · [Θ]g + ([s]0 − [s]) · ` · [Θ]f .

After the reaction is completed, only glucose and fructose remains, and [g] = [f] = [s]0 , i.e.,

θ∞ = [s]0 · ` · [Θ]g + [s]0 · ` · [Θ]f ,

3
so
θ0 − θ∞ = [s]0 · ` · ([Θ]s − [Θ]g − [Θ]f )
and
θt − θ∞ = [s] · ` · ([Θ]s − [Θ]g − [Θ]f ) ,
i.e.,
[s] θt − θ∞
= = e−kt . (1)
[s]0 θ0 − θ∞
It can be seen from the equations that the concentration ratios can be determined without knowing the optical
path length and specific rotations, and thus the rate constant can be calculated. The methods for this are
discussed below.

1.2 Evaluation methods


1. Point-by-point evaluation: According to the rearranged Eq.(1)
1 θt − θ∞
k = − ln
t θ0 − θ∞
k is calculated at each measurement point and the values are then averaged. This method requires
rotations measured at the beginning and end of the reaction.
2. Straight line fitting with one parameter:
θt − θ∞
ln = −k · t
θ0 − θ∞
Plotting the values calculated from the experimental data as a function of time, we obtain a line through
the origin, and from the slope the rate constant can be determined. This method also requires rotations
measured at the beginning and end of the reaction.
3. Straight line fitting with two parameters:
ln(θt − θ∞ ) = ln(θ0 − θ∞ ) − k · t
Plotting the values calculated from the experimental data as a function of time, we obtain a straight
line. From the slope, the rate constant, from the intercept θ0 can be calculated. This method requires
only the rotation measured at the end of the reaction.
4. Straight line fitting using the Guggenheim method: This method is proposed when determining θ∞
and θ0 is difficult experimentally. At a given t time
(θt − θ∞ ) = (θ0 − θ∞ ) · e−kt
and at t1 = t + ∆t time:
(θt1 − θ∞ ) = (θ0 − θ∞ ) · e−k(t+∆t) .
Taking the difference of the two equations:
(θt − θt1 ) = (θ0 − θ∞ ) · (e−kt − e−k(t+∆t) ) = e−kt · (θ0 − θ∞ ) · (1 − e−k∆t ) ,
then logarithmized:
ln(θt − θt1 ) = −kt + ln((θ0 − θ∞ ) · (1 − e−k∆t )) .
The logarithmic term to the right of the expression is constant if ∆t is constant, i.e., we used the same
time interval between the measurements. Thus, the calculated values of ln(θt − θt1 ) plotted versus the
time give a straight line. From the slope, the rate constant can be determined. This method does not
require measuring either θ∞ or θ0 , on the other hand, it requires that the rotation be measured at the
same time intervals! The previous methods did not require that timing of the measurements.

4
5. Nonlinear parameter estimation: Eq. (1) can be used without any modification. The method requires
more serious computing, but provides the greatest freedom. Nonlinear parameter estimation does not
make any requirement to the timing of measurement points, several curves can be evaluated together
and any of the values k, θ∞ és θ0 can be fitted or can have a fixed value.

2 The practice
In this practice, the rate constant of sugar inversion should be determined by the methods described above.
The optical rotation of the solutions is measured with a polarimeter at a constant acid concentration. A
description of the polarimeter and instructions for use can be found together with the equipments. For more
information on setting the polarimeter and reading the rotation angle, see the Appendix.

2.1 Experimental, carrying out the measurements


1. If not available, prepare 200 cm3 6.0 M hydrochloric acid solution and 100 cm3 sugar solution. The
concentration of sugar is given by the instructor, in the range of 0.6 – 1.1 M. (0.20 – 0.37 g/cm3 mass
concentration range.)
2. To determine θ∞ , measure 20 cm3 sugar solution and 20 cm3 6.0 M hydrochloric acid solution into a
stoppered Erlenmeyer flask. After mixing, put aside, the rotation will have to be measured at the end
of the practice when the reaction is complete (at least two hours).
3. To determine θ0 , mix 20 cm3 sugar solution with 20 cm3 distilled water (no catalyst, practically no
hydrolysis). With this solution, wash the cuvette of the polarimeter (in small portions, several times)
and then fill the cuvette with the solution. (Preferably without bubbles, but if a bubble remains, trap
it in the bulging part of the cuvette so that it is not in the light path. Make sure the cuvette windows
are clean and dry.) Place the tube in the polarimeter cuvette holder and measure the angle of rotation
of the light. Since the solution does not change, you have time to understand using the polarimeter.
Values measured in section 2 and 3 will be θ∞ and θ0 in each measurement, since only saccharides are
optically active, water and hydrochloric acid are not.
4. According to the instructor’s instructions, prepare 50 cm3 , 2.0 – 3.0 M hydrochloric acid solution from
the 6.0 M solution by dilution. Measure 20 cm3 sugar solution and 20 cm3 hydrochloric acid solution
into two separate (dry) beakers. Start the stopwatch when the two solutions are mixed (homogenize
the reaction mixture by pouring the solutions back and forth from the beakers). Wash the cuvette of
the polarimeter (in small portions, several times) and then fill the cuvette with the solution. Place it
in the cuvette holder of the polarimeter and measure the angle of rotation of the light as a function of
time. To use the Guggenheim method, measurements should be performed every 2 minutes. The total
time of the measurement is determined by the instructor (approx. one hour).
5. As instructed, prepare a more concentrated hydrochloric acid solution from the 6.0 M solution by
dilution. Repeat the previous measurement with the higher catalyst concentration. The total time of
the measurement is determined by the instructor (approx. one hour).
6. Measure θ∞ according to the previous procedures, using the solution prepared in step 2.

2.2 Evaluation of measurement results


1. Record the measured values of θ∞ and θ0 .
2. The measured and calculated data should be summarized according to the table below:

5
θt − θ∞ 1 θt − θ∞
t/min θt /◦ ln ln(θt − θ∞ ) ln(θt − θt1 ) k = − ln
θ0 − θ∞ t θ0 − θ∞

3. Make the θt − t plots for each measurement. They can be on the same figure.
4. Determine the pseudo-first-order rate constant of the sugar inversion in the ways ordered by the in-
structor (see Ch.1.2.) for both measurements.
5. Characterize the evaluation methods (which and when is more accurate, faster) by comparing the
graphs, the calculated k values and their standard deviations. Try to estimate the accuracy of the
experiments.
6. Compare the values of the determined pseudo-first-order rate constants as a function of hydrochloric
acid concentration.

Control questions:

1. What reaction do you study at the practice?


2. What is plane polarized light, and how does a polarimeter work?
3. What is molar rotation? What is the specific rotation? How does the angle of light rotation depend on
the concentration?
4. If you mix 20 cm3 sucrose of 250 g/dm3 mass concentration with 20 cm3 water, what will be the light
rotation of the solution at 2 dm light path. The specific rotation of sucrose is 66.5 (◦ cm3 )/(g dm).
5. If the specific rotation of sucrose is 66.5 (◦ cm3 )/(g dm), what is its molar rotation? Mr (sucrose) = 342.
6. Write the differential and integrated form of the rate equation for a first-order reaction.
7. What methods do you know for the graphical determination of a first-order rate coefficient?
8. What is the Guggenheim method for calculating rate coefficients? Derive the method of calculation.
9. Why is [H+ ] not in the rate equation, even though the inversion rate depends on it?

6
Appendix: Using the polarimeter

With manually adjusted polarimeters, the plane of polarization is difficult to determine because our eyes
are not sensitive enough to determine either the maximum light intensity (same position of the optical axes
of the polarizer and analyzer) or the minimum light intensity (cross position). The eye is much better in com-
paring light intensities, so to help more accurately reading the plane of polarization or the angle of rotation,
we do not use a single polarizer, but a polarizer system with a slightly different optical axis (difference 1-2◦ ).

Light polarized in two slightly different planes by two polarizers results in two different light intensities
when viewed through the analyzer. At high intensities, this is not noticeable, so we prefer to use the crossed
position. Find the position of the analyzer when the light from the two polarizers has exactly the same
intensity, i.e., the analyzer has an optical axis perpendicular to δ/2 (Figs. 6 and 7).

Figure 6: Possible variants of polarizer systems.

If the setting is good, read the analyzer’s angular position. For a more accurate reading, we use the
so-called Vernier scale. The scale rotating with the analyzer shows the polarization and / or the angle of
rotation in degrees. The position of the 0 of the auxiliary scale (Vernier scale) shows which integer degrees
the setting is between. The division of the auxiliary scale line, which coincides with one (any) line of the
rotating scale is an estimate of the decimal degree. If we are below 0 ◦ , we have to read it in the same way,
but these are already negative angles (the optical axis has turned 180 ◦ ), so subtracting 180 ◦ gives the correct
angle. The 9 main scale = 10 Vernier scale (see Fig. 8) is suitable for estimating 0.1 ◦ , the 19 main scale =
10 Vernier scale is suitable for estimating 0.05 ◦ , the 49 main scale = 10 Vernier scale can be used to estimate
0.02 ◦ . (By the way, this is how calipers and micrometers can be read!)

7
Figure 7: The relationship between the different angular positions of the analyzer and the intensity of the
light passing through the polarizers for different types of polarimeters.

Figure 8: Using the Vernier scale.