PRODUCT SPECIFICATIONS Indiko
Indiko clinical and specialty chemistry analyzer
Fully automated, sample oriented,
random access analyzer
Thermo Scientific™ Indiko™ is a fully automated, sample Samples and Reagents
oriented random access analyzer for clinical and special Continuous access to samples and reagents without
chemistries, including enzymes, substrates, electrolytes, interrupting test processing. Automatic identification
specific proteins, therapeutic drug monitoring, drugs of via internal barcode reader. Clearly displayed real-time
abuse and immunosuppressant drug tests(*. reagent volume and remaining test capacity. Up to four
Indiko employs colorimetric end-point and kinetic as reagent additions per test.
well as turbidimetric and bichromatic reactions with or
On-Board max 6 racks in the cooled disk
without sample blanking. Optional ISE unit uses ion
Capacity: - 9 or 18-position sample rack
selective electrodes for sodium, potassium and chloride - 6-position reagent rack
measurements. Sample Volumes: 2 - 120 µL
ISE and photometric tests can be run simultaneously
Sample 0.5 mL, 2.0 mL cups and sample tubes
from a single sample. Containers: (diameter 12 - 16 mm, length 75 - 100
mm)
(* Application availability may vary from country to country. Please
Sample Barcodes: Code 128 and barcodes USS Codabar,
contact your local sales representative for more information.
Interleaved 2 of 5 and Code 39 with a
check digit
Measurement
Reagent Volumes: 2 - 240 µL
Single channel interference filter photometer with beam
Reagent 10 mL and 20 mL vials
splitting reference, 12 filter positions.
Containers:
Filter range: 340 - 700 nm Sample and CV ≤2% for volumes ≥2 µL
Reagent
Incubation Controlled at 37 °C
Dispensing:
temperature:
Light source: Xenon flash lamp
Absorbance 0 - 3.5 A Calibration
range: - resolution of 0.001 A
Factor, bias, linear, logit-log, spline, polynomial and point-
- reproducibility of SD <0.005 A
at 2 A
to-point calibration. Method-dependent use of individual
calibrators or automatically diluted series from a stock
calibrator. Previous curve comparison available.
Reaction vessels
Discrete disposable cuvette. Continuous access to
Quality Control
cuvettes without interrupting test processing.
Real-time QC program with multiple, user-definable
On-Board 360 measurement cells Westgard rules. Control frequency user-definable. Out-of-
Capacity: - 36 cuvettes with 10 reaction cells specification control results flagged. QC chart printouts,
- up to 2 hours walk-away time daily and cumulative reports.
depending on workload
Reaction End 120 - 300 µL
Volume:
Dilutions LIS Interface: CLSI LIS02-A2
Automatic sample pre-dilution. Automatic dilution of Hardware RS-232 or TCP/IP
Interface:
over-range tests with automatic rerun. Addition of manual
Result Reports: Collated by sample, manual entry of
pre-dilution value for result calculation. Reflex testing
off-line results allowing for fully collated
capability.
result reports, results calculated from
both measured and off-line results.
Data Management Abnormal values and repeats flagged
Windows® workstation with graphical user interface. automatically.
Data input online, via mouse, touchscreen, keyboard and
barcode reader. Different user groups can have different Traceability: Full traceability with long term storage
access rights. Different user interface language versions of results including associated
calibrations and reagent lot data.
available.
Photometric tests/hour: up to 200
Capacity:
ISE tests/hour: up to 120
75 cm (width) x 70 cm (depth) x 62/130 cm (height/with open cover), 85 kg (weight).
Dimensions and Weight:
Separate workstation.
Power Requirements: 100 - 240 VAC ±10%, 50-60 Hz ±5%, 250 W
Deionized Water Consumptions: 1.5 litres/ hour
Average Noise Level at 1 meter: <60 dB(A)
Environmental Conditions: Operating temperature range of 18 - 30 °C, humidity 40-80% (non-condensing)
Conforms to CAN/CSA-C22.2 No.61010-1-04
UL Std. No. 61010-1:2004
FCC CFR 47 Part 15, subpart B
Regulatory:
2011/65/EU RoHS Directive
FDA cleared, 510(k) number: k110035 (Indiko 9863 0000)
98/79/EC IVDMD Directive
9863 0000 Indiko
Ordering Codes:
9863 1000 Indiko with the ISE unit
In Europe: In the United States:
+358-10-329-200 call 1-800-232-3342
e-mail: [Link]@[Link] e-mail: [Link]@[Link]
Find out more at [Link]/indiko
© 2017 Thermo Fisher Scientific Inc. All rights reserved. Windows is a trademark of Microsoft Corporation. All
other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. The information and technical
specifications are subject to change without prior notice. CODE: N11232_07 12 / 2017