MAC 5 Lite data sheet™

GENERAL Defibrillation Per IEC 60601-2-25:2011


protection
Instrument type Microprocessor augmented automatic Patient leakage <10 µA
electrocardiograph; 10-lead wire
Specifications for digital acquisition and analysis
acquisition with programmable lead
of waveforms
configuration
Analog to digital 24-bit analog to digital conversion
ECG interpretation Marquette™ 12SL™ ECG analysis
conversion resolution
program for adults and pediatrics
Over sampled rate: 512 ksps
Computerized 12-lead analysis includes measurements
measurements Down sampled ECG Bandwidth: 0.04 to 300 Hz*
waveform Sample rate: 2 ksps
Heart rate meter 30 to 300 BPM ±10% or 5 BPM,
Resolution: 1.22 µV
whichever is greater — heart rates
outside this range will not be displayed Input to 12SL Bandwidth: 0.04, 0.56 ZPD to 300 Hz*
Sample rate: 1 ksps
ECG data formats GE Hi-Fidelity ECG, XML
Resolution: 4.88 µV
External archiving USB removable media
Additional report 20 Hz, 40 Hz, 100 Hz, 150 Hz or 300 Hz*
Pre-acquisition Provides 10 seconds of instantaneous filters
ECG acquisition
Specification for stored / transmitted waveforms
Digital rhythm Up to 5 minutes of continuous rhythm
Digital rhythm Bandwidth: 0.04, 0.56 ZPD to 300 Hz*
storage (exportable as a PDF)
waveform Sample rate: 1000 sps
Full disclosure Review up to 5 minutes of 12 Lead ECG, Resolution: 4.88 µV
abiliy to selet 10 second Resting ECG
12-lead ECG Bandwidth: 0.04, 0.56 ZPD to 300 Hz*
records, ability to generate 5 minute
waveform Sample rate: 500 and 1000 sps
single lead full disclosure report
Resolution: 4.88 µV
Storage 300 records consisting of 10 second
Representative Sample rate: 500 and 1000 sps
Resting ECG and rhythm records
(median) complex Resolution: 4.88 µV
200 minutes Digital Rhythm or Full
Pace detection
Disclosure records
Pacemaker Sample rate: 75 ksps
Dynamic range AC differential ± 10 mV
waveform
DC offset ±600 mV
Pace detection Duration: 0.2 ms to 2.1 ms
Common mode >125 dB (>100 dB with AC filter disabled)
Amplitude: 2 mV to 700 mV
rejection
Separation: 1 ms or greater
Input impedance >50MΩ @ 10 Hz
Pace annotation Dedicated pace channel on display and
printed reports (configurable)
COMMUNICATIONS Display data Heart rate, patient name, patient ID,
date, clock, battery power indicator,
ECG management MUSE™ Cardiology Information scrolling waveforms, lead labels, speed,
systems System Compatible (v8 or later) with gain and filter settings, warning
connectivity bi-directional orders and ADT support messages, prompts, hookup advisor
Transmit Resting ECG records to and help messages
Cardiosoft™ via removable media (v6.73
or later) or via network (v7 or later) WRITER
DICOM Modality worklist/orders: supported
via GE MUSE (v8 or higher) and DICOM Writer technology Integrated thermal dot array
Gateway with bi-directional orders Number of traces 3, 6, 12 user selectable
support Writer speeds 5, 12.5, 25, and 50 mm/s
EMR connectivity Via MUSE cardiology information Writer 2.5, 5, 10, 20 mm/mV, and
system (v8 or later) with EMR Gateway sensitivity/gain 10/5 mm/mV split gain
Data export Export of Resting ECG (in PDF or XML Writer speed 5, 12.5 mm/s @ +5%
format), Digital Rhythm and Full accuracy 25, 50 mm/s @ ±2%
Disclosure reports (in PDF format) over
Writer amplitude ±5%
Secures File Transfer Protocol (SFTP)
accuracy
or to a Shared folder
Writer resolution Horizontal: 40 dots/mm @ 25 mm/s
Wireless Wireless 802.11 a/b/g/n wireless
Vertical: 8 dots/mm
connectivity (2.4 GHz/5GHz)
Network printer Support printing to network printer
IPV4
DHCP, hostname and static IP options
ELECTRICAL
for configuring device IP/network
address Power supply AC mains or battery operation
WEP and enhanced security WPA-PSK, Input voltage 100-240 VAC + 10%
WPA2-PSK, WPA/WPA2 enterprise
Input frequency 50-60 Hz + 3 Hz
protocols TLS, PEAP-MSCHAPV2, PEAP-
GTC, TTLS-MSCHAPV2, TTLS-GTC. (PEAP Battery type Replaceable and rechargeable internal
requires network evaluation/approval battery
prior to purchase) Battery capacity Minimum 180 minutes with acquiring
Ultra-high security 4096 bit encryption/ and printing a single page ECG report
long certificate support every 15 mins (with five minutes auto
standby enabled and all accessories
SHA1 and SHA2 support
connected, except KISS)
Network 802.3 Ethernet interface via RJ45
Battery charge Approximately 240 minutes from total
connectivity connector
time discharge when device is off or standby
Compatible to 10Base-T, 100Base-T LAN
IPV4 SECURITY AND PRIVACY
DHCP, hostname and static IP options
for configuring device IP/network Encryption All files containing PHI, local users and
address passwords
Network clock Network time synchronization (NTP) Login Network: LDAP/Active directory
authentication Local: User database
DISPLAY User management Customizable roles for limiting system
access by user groups for Admin, Clinical,
Display and 8.9” diagonal, LED backlit, Service, Biomed, and user defined up to
resolution 892 x 558 pixels 10 customized roles
Touch screen type Projected Capacitive (PCAP) multipoint Audit trail All user logins, logouts and login failures,
touch input that works while wearing file deletions, file changes, file views,
medical exam gloves file acquisitions, file transmissions, file
printouts, system configuration changes
PHI access Controlled by customizable roles with Cleaning
configurable advanced strict PHI access Approved cleaning Soap and water solution
rules agents Sodium hypochlorite (NaOCl) 5% solution
PHI access logs Detailed and exportable logs of all PHI Ethanol (ethyl alcohol) 96% (v/v)
viewing by users Isopropyl alcohol 70% (m/m)
Emergency access This user can access the device without Hydrogen peroxide 20% (v/v)
(STAT mode) providing login credentials to perform Phenol 2% (V/V)
emergency tasks such as acquiring an FDA cleaning agent Super Sani-Cloths
ECG or rhythm while preventing access for efficacy
to any stored patient data, orders, ADT,
or 3rd party applications CERTIFICATION
USB lockout Software controls to disable USB
ports/connections Certification marks cTUVus
Standards EN 60601-1:2006/A1:2013
PHYSICAL SPECIFICATIONS complied with EN 60601-1-2:2007 +AC 2010
EN 60601-1-2:2015
Weight and dimensions EN 60601-2-25:2015
Max weight 3.5 kg IEC 62366-1:2015
Max height 95 mm EN 62304:2006+A1:2015
Max width 320 mm Pharma mode* Support 21CFR Part 11 clauses for the
Max length 270 mm control of electronic records in closed
systems when Phrama mode option
ENVIRONMENTAL SPECIFICATIONS is activated. No support for electronic
signature
Temperature
Operating 50° to 104° F (10° to 40° C) ORDERING INFORMATION
Transport/Storage –4° to 140° F (–20° to 60° C)
Available user Chinese, Danish, Dutch, English, French,
Humidity interface languages Finnish, German, Italian, Norwegian,
Operating 20% to 95% RH non-condensing Swedish, Korean, Japanese, Russian,
Transport/Storage 15% to 95% RH non-condensing Spanish, Portuguese, Brazilian
Portuguese, Polish
Pressure
Operating 70 to 106 kPa
Transport/Storage 50 to 106 kPa
Input devices
Keyboard Touch keyboard
Touchscreen 8.9”, 892 x 558 pixels projected
Capacitive (PCAP) multipoint touch
input that works while wearing medical
exam gloves
Barcode External barcode scanner (optional)
Mouse Supported but not included
External USB barcode scanner
Types Fixed and variable length
Symbologies Code-128, PDF417, Code 39, Interleaved
Code 2 of 5, and Data Matrix symbology
for characters A-Z (upper case), a-z
(lower case), and 0-9 for all supported
languages
* The feature is not supported in China.

© 2022 General Electric Company – All rights reserved.

GE Healthcare reserves the right to make changes in specifications and features shown herein, or discontinue the product described at
any time without notice or obligation. CardioSoft v7 and MAC 5 are not 510k cleared and not available in all markets. Contact your GE
Healthcare representative for the most current information. GE, the GE Monogram, 12SL, CardioSoft, MAC, Marquette, and MUSE are
trademarks of General Electric Company. GE Healthcare, a division of General Electric Company.

MAC 5 Lite
DOC2563553 rev 1 Lite

You might also like