0% found this document useful (0 votes)
66 views4 pages

Topcon TRC-50 Camera Setup Guide

This document provides detailed instructions for taking pictures using the Topcon TRC-50 series retinal cameras with the IMAGEnet system, including settings for color pictures and fluorescein angiography. It outlines specific camera settings, procedures for capturing images, and recommendations for flash intensity based on patient conditions. Additionally, it includes instructions for setting auto flash intensities and capturing auto fluorescence images.

Uploaded by

Giovâny Gomes
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
66 views4 pages

Topcon TRC-50 Camera Setup Guide

This document provides detailed instructions for taking pictures using the Topcon TRC-50 series retinal cameras with the IMAGEnet system, including settings for color pictures and fluorescein angiography. It outlines specific camera settings, procedures for capturing images, and recommendations for flash intensity based on patient conditions. Additionally, it includes instructions for setting auto flash intensities and capturing auto fluorescence images.

Uploaded by

Giovâny Gomes
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

Erik R Mc Gregor T.E./S.J./C.U.

International Division - Technical Support Specialist

INSTRUCTIONS ON TAKING PICTURES USING


TOPCON TRC-50 SERIES RETINAL CAMERAS
WITH IMAGEnet SYSTEM

The following sets of instructions are intended to be used as a reference only. Keep in mind that
values may change depending on variables like: type of camera, illumination and patient conditions.

COLOR PICTURES USING THE


TOPCON TRC-50EX RETINAL CAMERA
WITH IMAGEnet SYSTEM

Retinal Camera Settings:

Auxiliary lens knob: 0 for normal retinas


- for high myopes
+ for aphakic patients w/o IOL
A for anterior chamber pictures

Barrier filter OFF

Filter knob “N”

Exciter filter OFF

Pupil size lever Middle position for normal


Lower position for small

Flash settings: 50 for normal retinas or anterior chamber.


75 for dark retinas
36 for albino retinas
It is always recommended to bracket one step up and one down when the photographer is
unsure of the type of retina.
Erik R Mc Gregor T.E./S.J./C.U.
International Division - Technical Support Specialist

COLOR PICTURES USING THE


TOPCON TRC-50IX RETINAL CAMERA
WITH IMAGEnet SYSTEM

Retinal Camera Settings:

Auxiliary lens knob: 0 for normal retinas


- for high myopes
+ for aphakic patients w/o IOL
A for anterior chamber pictures

Barrier filter OFF

Filter knob “N”

Exciter filter OFF

Pupil size lever “N” position for pupil 6mm Ø or more


“S” position for pupil under 6mm Ø

Flash settings: 50 for normal retinas or anterior chamber.


75 for dark retinas
36 for albino retinas

INSTRUCTIONS TO PERFORM
FLUORESCEIN ANGIOGRAPHY WITH THE
TOPCON TRC-50EX RETINAL CAMERA
AND IMAGEnet SYSTEM

MATERIALS NEEDED:

Fluorescein: Use Sodium fluorescein (we recommend 25% made by Akorn) in vials of [Link] all
2.5cc in a bolus form in no more than two seconds. It is best to use 5cc syringes to facilitate injection
and 23 gauge Butterfly injection needles. With patients with very fragile veins, 25-gauge needle is
recommended.
Erik R Mc Gregor T.E./S.J./C.U.
International Division - Technical Support Specialist

PROCEDURE:

1. Open a Capture procedure for FA on IMAGEnet and click the clock on the capture bar to sync the
clock with the first shot.

2. Adjust the eyepiece to your vision by turning the eyepiece mount until the crosshairs in the center
are sharp and clear. You may place a white piece of paper in front of the lens to better see them.
Once this adjustment is performed, the eyepiece does not have to be re-adjusted, providing no
other operator will use the instrument.

3. Position the patient in front of the retinal camera in a comfortable manner. Place the head in the
chin rest and make sure that the forehead touches the upper bar, and the canthus of the eye is at
the same height of the black line on the side bar which holds the forehead rest.

4. With the filter knob in “N” (white light), align the eye by pulling the retinal camera head all the
way back and centering the pupil in the crosshairs. Then move in towards the patient until the
retinal image fills the view. At this point focus the retina using the focusing knob until the
structures are clear. You can also use the “Working Dots” at each side of the retinal image. When
they are sharp like small dots, the working distance is correct and you can focus the retina with
the focusing knob.

5. Turn the filter knob to “G” (green light / red-free filter) and using a flash power of 25 in the flash
power supply, take one or two red free pictures.

6. Recline the patient and perform venipuncture. When the needle is in the vein and good blood
return has been confirmed, place the patient’s head in the chin rest again. Re-check the focusing
using “N” (white light).

7. Press the Exciter filter button, and set the flash power to 200. (In very dark patients use 300) (You
may want to increase the power of the “Fluo” viewing light on the left side panel to # 7 in order to
see better the retina). Inject the fluorescein and immediately after injection, start the timer.

NOTE: In order to obtain good contrast and density, it is important to inject as fast as possible.
The injection must not last longer than 3 seconds. Otherwise, the fluorescein dilutes in the
bloodstream and when it reaches the eye does not have enough intensity to produce a good
image.

8. Take one picture every two seconds for the first 15 - 20 seconds. Have the patient rest and take
the needle out. (The Barrier filter will insert automatically in position every time you take a picture
and pull out after, for better viewing of the retina)

9. Take two or three more pictures of the recirculation phase and wait for about 7 to 10 minutes
before taking the late phase images.
Erik R Mc Gregor T.E./S.J./C.U.
International Division - Technical Support Specialist

SETTINGS FOR CAPTURING AUTO FLUORESCENCE IMAGES


USING THE TRC-50 DX/IA, THE AUTOFLUORESCENCE FILTER SET
AND IMAGEnet SYSTEM

1. GAIN ON THE MEGAPLUS ES-3200 CAMERA: 16

2. FLASH POWER: 300WS

3. GAMMA: 1.2 OR 1.3

4. SMALL PUPIL: ON

Note: In case the patient has incipient cataracts, reduce the Gamma to 1.0 or 0.9 and the flash to
200WS

INSTRUCTIONS TO SET AUTO FLASH INTENSITIES IN THE


TOPCON TRC-50EX RETINAL CAMERA

1. Turn retinal camera on.

2. Set the “Exposure Setting” switch in “Program Mode” (position 2).

3. Select the filter position you wish to set.

4. Select the flash intensity you wish to set.

5. Press both up and down arrows for flash intensity (at the left of flash intensity bar) at the same
time until camera beeps five times rapidly. The camera is now programmed to default to that flash
intensity every time you select that filter. Even though the instrument will default to that flash
intensity, you can further modify it, in the same way you will for manual mode.

6. To change settings, repeat from step 3.

Suggested settings for film.

Fluorescein Angiography

Red-free 18
F/A 200

Color

Color (N position) 50

You might also like