Pre Clinical Plan & Report - DNC
Pre Clinical Plan & Report - DNC
Table of Contents
1. Introduction 2
2. Scope 2
3. Product Description 3
8. Summary 10
9. Revision History 11
Relisys Medical Devices Limited Doc. No: 015-0101-4.2
1. Introduction:
The pre clinical studies or non clinical trials are laboratory test of a new medical device. The
tests conducted in vitro are include biocompatibility studies, Bench tests etc…
The Biocompatibility Study of Medical Devices is the protection of humans from potential
biological risks arising from the use of medical devices. An EN ISO 10993-1:2009(E) is
intended to be a guidance document for the biological evaluation of medical devices within a
risk management process, as part of overall evaluation and development of medical devices.
The role of EN ISO 10993-1 is to serve as a frame work in which to plan a biological
evaluation which, as scientific knowledge advances our understanding of the basic
mechanisms of tissue responses, minimizes the number and exposure of test animals by
giving preference to chemical constituent testing and in-vitro models, in situations where
these methods yield equally relevant information to that obtained from in vivo models.
EN ISO 10993-1 is intended for use by professionals, appropriately qualified training and
experience, who can interpret its requirements and judge the outcome of the evaluation for
medical devices, taking into consideration all the factors relevant to the device, its intended
use and the current knowledge of the medical device provided by review of the scientific
literature and previous clinical experience. The biological data sets recommended in the
evaluation of medical devices, according to their category of body contact and duration of
clinical exposure. This standard contains guidance for the application of risk management
process to medical devices which encompasses biological evaluation.
2. Scope:
The main scope of pre clinical studies is to determine the ultimate product safety &
performance as per the intended use claimed by manufacturer.
The pre-clinical study is to develop adequate data to decide that it is reasonably safe to
proceed with the human trials.
Relisys Medical Devices Limited Doc. No: 015-0101-4.2
The Biological evaluation of medical devices is performed to determine the potential toxicity
resulting from contact of the device with the body. Testing shall be performed on the sterile
final product only. The test sample shall be chosen to maximize the exposure of the test
system to the components of a device that is known to have potential for a biological
response.
3. Product description:
The Post Dilatation catheter is a standard balloon dilatation catheter with a balloon near the
distal tip. One lumen is used for initiation of the balloon with contrast medium; the other
lumen permits the use of a guide wire to facilitate advancement of the catheter to and
through the lesion to be dilated. The product is offered on rapid exchange delivery platform.
The distal end of the catheter has a conventional nylon blend balloon and the balloon has
radiopaque markers to aid in positioning the balloon in the lesion, and is designed to provide
an expandable segment of known diameter and length at a specific pressure. The proximal
end of the catheter is a SS hypotube jacketed with nylon to give smooth pushability.
The product components physical description, contact with body and materialsdata is
following below.
Physical
S/N Components Contact Material VENDOR
description
PEBAX
Blood 3533 SA
A small tubular
circulation - 01 MED + FOSTER WEST
1. SOFT TIP polymer
Patient, 30% CORPORATION
component
Hand User. BaSO4
WHITE
VESTAMID
ML CARE
BALLOON A Nylon polymer Blood
2. 21 / EMS CHEMIE
TUBE component circulation
GRILAMID
L 25
RISAN PA
A Nylon polymer Blood 11 G FOSTER WEST
3. OUTERLUMEN
component circulation BESVO A CORPORATION
MED
Figure 1: Image of the catheter packed in Tyvek pouch and when exposed outside from the
pack
Product details:
Catheter Sizes
Diameter (mm)/ 2.50 2.75 3.00 3.50 4.00
Length (mm)
8 ✓ ✓ ✓ ✓ ✓
10 ✓ ✓ ✓ ✓ ✓
12 ✓ ✓ ✓ ✓ ✓
13 ✓ ✓ ✓ ✓ ✓
15 ✓ ✓ ✓ ✓ ✓
18 ✓ ✓ ✓ ✓ ✓
20 ✓ ✓ ✓ ✓ ✓
Relisys Medical Devices Limited Doc. No: 015-0101-4.2
The biological evaluation of any material or medical device intended for use in humans shall
from part of a structured biological evaluation programme within a risk management process
in accordance with EN ISO 14971:2012. The biological evaluation shall be PLANNED,
carried out, and documented by knowledgeable and experienced professionals.
The following shall be considered for their relevance to the overall biological evaluation of
the device:
The choice of tests and the data required in a biological evaluation, and their interpretation,
shall consider the chemical composition of the materials, including the conditions of
exposure as well as the nature, degree, frequency, and nature of exposure of the medical
device or its constituents to the body, enabling the categorization of devices to facilitate the
selection of appropriate tests.
Selection of any in vitro or in vivo tests shall be based on end user applications. All tests
shall be conducted according to recognized and valid laboratory, for example Good
Laboratory Practices (GLP) or EN ISO 17025 certified, where applicable and the data shall
be evaluated by competent, informed professionals.
Relisys Medical Devices Limited Doc. No: 015-0101-4.2
Summary of the systematic approach to a biological evaluation of medical device as part of risk management process
Biological evaluation
complete
Relisys Medical Devices Limited Doc. No: 015-0101-4.2
Post Dilatation catheters shall be categorized according to the nature and duration of body
contact as described in EN ISO 10993-1.
As per ISO 10993-1:2009 Evaluation of tests for consideration in the following table.
*Highlighted portion of the tests in the above table shall be strictly conducted as per the standard requirements
Relisys Medical Devices Limited Doc. No: 015-0101-4.2
Hemolysis
Thrombosis
S.N Instrument/
Inspection item Specification
o Reference
1 Crossing Profile 0.80 - 1.30 mm LMS
2.50mm – 2.50 ± 0.12mm LMS/PT-1000/
2 Balloon diameter at 8 atm
4.00mm – 4.00 ± 0.12mm Indeflator
2.50mm – 2.65 ± 0.12mm LMS/PT-1000/
3 Balloon diameter at 16 atm
4.00mm – 4.20 ± 0.12mm Indeflator
4 Balloon Inflation Time ≤ 10 sec In deflator
5 Balloon Deflation time ≤ 15 sec In deflator
Relisys Medical Devices Limited Doc. No: 015-0101-4.2
8. Summary:
All raw materials and components are same for the manufacturing of Post Dilatation
catheters including all sizes mentioned in the product description.
The Biocompatibility studies for Post Dilatation Catheters must follow EN ISO 10993-
1:2009(E).
The Pre clinical bench tests for Post Dilatation Catheter shall be followed as per EN
ISO 10555-4 & ISO 594-2 and the regulatory requirements as per MEDDEV 2.7.1
rev.4 to ensure the safety & efficacy of the device.
Relisys Medical Devices Limited Doc. No: 015-0101-4.2
The test facility details including required laboratory accreditations shall be attached in
the summary report.
9. Revision History:
Approval:
Table of Contents
1. Scope 2
5. Conclusion 12
6. Revision History 13
Relisys Medical Devices Limited Doc #: 015-0101-4.2
PRE CLINICAL STUDY REPORT FOR
Rev. No: 01 Page 2 of 13
POST DILATATION CATHETERS
1. Scope:
The Pre clinical study of Post Dilatation catheter has been carried out to determine its safety
& performance towards the intended use of the device claimed as per Relisys Medical
Devices Limited.
a. Cytotoxicity
b. Skin Sensitization
f. Haemocompatibility
4) Assessments
5) Biocompatibility Evaluation Flow Chart, which illustrates how to proceed with a
biocompatibility evaluation;
6) Biocompatibility Test Report, which includes biocompatibility test reports of Materials
used for Post Dilatation catheter.
7) Bench test results as referred in Product standard, Regulatory & clinical
requirements.
Relisys Medical Devices Limited Doc #: 015-0101-4.2
PRE CLINICAL STUDY REPORT FOR
Rev. No: 01 Page 3 of 13
POST DILATATION CATHETERS
Risks has been considered and assessed for the biocompatibility hazards (Doc
No: RMF-DNC Clause: 5.1.2) due to the material used. All though Raw materials used for
Post Dilatation Catheters are medical grade, change in material properties during
manufacturing process has been considered which call for biocompatibility risks like
Biocompatibility Hazards
• Non-biocompatible
• Allergic to the material
• Incompatible of contrast media
Considering all the above risks Post Dilatation Catheter was performed with biocompatible
tests at GLP certified laboratories to ensure that product is safe to use.
The flow chart below provides pathway to Biocompatibility tests for Post Dilatation catheter.
S. Test Name Study Title Test Method Reference Test Result Reference
No Standards Document Number
1 Cytotoxicity In Vitro Cytotoxicity Biological Evaluation of Observed Grade ‘0’ VLL/0215/G/B008/R
Assessment of “Veda+R Medical devices – EN ISO which indicates material
Dilatation catheter (Veda+R 10993 Part 5: tests for In Non-toxic. Hence there
NC)- Non Compliant” Vitro Cytotoxicity, 2009 is no chemical reaction.
Extraction method on
L929Cells
2 Skin Sensitization a) Skin Sensitization Study Biological Evaluation of COA results shows that VLL/1015/G/B057/R
(GPMT) of Normal Saline Medical devices – EN ISO the sample is Non-Skin
Extract of ‘Veda+R-NC Post 10993 Part 10: tests for sensitizer with the test
Dilatation Catheter’ in Irritation and skin animal
Guinea Pigs. sensitization, 2010.
b) Skin Sensitization Study COA results shows that
(GPMT) of Sesame Oil Biological Evaluation of the sample is Non-Skin
Extract of ‘Veda+R-NC Post Medical devices – EN ISO sensitizer with the test VLL/1015/G/B058R
Dilatation Catheter’ in 10993 Part 12: tests for animal
Relisys Medical Devices Limited Doc #: 015-0101-4.2
PRE CLINICAL STUDY REPORT FOR
Rev. No: 01 Page 7 of 13
POST DILATATION CATHETERS
Irritation or Intracutaneous Reactivity Biological Evaluation of All the test results are VLL/0215/G/B006/R
3
Intracutaneous test of ‘Veda+R Dilatation Medical devices – EN ISO observed to be at level
reactivity Catheter (Veda+R/NC) Non 10993 Part 10: tests for ‘0’ which indicated the
Compliant’ in New Zealand Irritation and skin sample is Non-Irritant
White Rabbits sensitization, 2010. with the test animal.
and
Biological Evaluation of
Medical devices – EN ISO
10993 Part 12: tests for
Sample preparation and
Reference Materials, 2012
Systemic toxicity Acute Systemic Toxicity Test Biological Evaluation of All the test results are VLL/0215/G/B003/R
4
(acute) of ‘Veda+R Dilatation Medical devices – ISO 10993 observed to be at level
Relisys Medical Devices Limited Doc #: 015-0101-4.2
PRE CLINICAL STUDY REPORT FOR
Rev. No: 01 Page 8 of 13
POST DILATATION CATHETERS
Catheter (Veda+R/NC) Non Part 11: tests for systemic ‘0’ which indicates that
Compliant’ in Swiss Albino toxicity, 2006. the sample did not
Mice and introduce any Toxicity
into the testing animal.
Biological Evaluation of
Medical devices – ISO 10993
Part 12: tests for Sample
preparation and Reference
Materials, 2012
Material Medicated Test for Material Mediated Biological Evaluation of
5 Based upon the results 315/015
Pyrogenicity in New Zealand Medical Devices - Part 11,
Pyrogenicity
White Rabbits Tests for Systemic Toxicity, obtained in this study
ISO 10993-11:2017(E).
and in line with ISO
Biological Tests - USP 42
<151>Pyrogen Test, 2019. 10993-11:2017 (E) and
USP 42 <151> Pyrogen
Test, 2019; it was
concluded that the given
test item Post Dilatation
Catheter (JADE NC)
(Lot No.: 021036011),
supplied by Relisys
Medical Devices
Relisys Medical Devices Limited Doc #: 015-0101-4.2
PRE CLINICAL STUDY REPORT FOR
Rev. No: 01 Page 9 of 13
POST DILATATION CATHETERS
Limited, is non-
pyrogenic to New
Zealand White Rabbits.
Systemic toxicity (acute) EN ISO 10993 - 11: 2006 No evidence Yes Pass
Biological Evaluation of Medical
Material Medicated Devices - Part 11, Tests for Systemic
Toxicity, ISO 10993-11:2017(E). Non-Pyrogenic Yes Pass
Pyrogenicity
Biological Tests - USP 42 <151>
Pyrogen Test, 2019.
Haemocompatibility EN ISO 10993 - 4: 2002/Amd.1:2006 Hemocompatible Yes Pass
Instrument/ Reference
[Link] Inspection item Specification Test Results
Reference Document No.
1 Crossing Profile 0.80 - 1.30 mm LMS Complies 015-0101-06.2
2.50mm – 2.50 ± Complies (Verification &
0.12mm LMS/PT-1000/ Validation)
2 Balloon diameter at 8 atm
4.00mm – 4.00 ± Indeflator
0.12mm
2.50mm – 2.65 ± Complies
0.12mm LMS/PT-1000/
3 Balloon diameter at 16 atm
4.00mm – 4.20 ± Indeflator
0.12mm
4 Balloon Inflation Time ≤ 10 sec In deflator Complies
5 Balloon Deflation time ≤ 15 sec In deflator Complies
Stylet/ Mandrel/ Complies
6 Maximum Guide Wire Outer Diameter 0.014”/ 0.356 mm
Guide wire
7 Minimum Guide Catheter Inner Diameter 0.056”/ 1.42 mm Guiding Catheter Complies
Relisys Medical Devices Limited Doc #: 015-0101-4.2
PRE CLINICAL STUDY REPORT FOR
Rev. No: 01 Page 11 of 13
POST DILATATION CATHETERS
5. Conclusion:
Post Dilatation catheter is intended to use for cardiovascular procedures. Post dilatation catheter is intended for balloon dilatation of a stent
after implantation. Balloon dilatation of stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving
myocardial perfusion. Balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-
segment elevation myocardial infarction. Materials used for all the sizes are same in the Declaration of Conformity document. The names
present in Biocompatibility study reports are Veda+R/NC, Veda+R Dilatation catheter-Non-Compliant balloon are referring to the same.
The Biocompatible tests conducted as per EN ISO 10993-1:2009 selection criteria of biocompatibility tests, carried out at GLP Certified
Vimta Labs, Hyderabad, Accuprec Research Labs Pvt. Ltd., Gujarat and GLR Laboratories Pvt Ltd., Chennai. Post Dilatation catheter was
found to be biocompatible based on results of Cytotoxicity, Haemocompatibility, Intracutaneous Reactivity test, Acute Systemic Toxicity,
Material Mediated Pyrogenicity Skin Sensitization. Hence, Post Dilatation Catheters are safe to use in humans for its intended purpose.
The Bench tests are conducted at Relisys Medical Devices as per the Product standard EN ISO 10555-1:2013 & EN ISO 10555-4:2013
requirements. Radio-detectability test has been conducted in Care Hospitals, Hyderabad. All the test results are observed & found that
product met with the predetermined specifications. Hence it is concluded that the Product Post Dilatation catheter is safe to use in humans
for its determined intended use in coronary angiographic procedures.
Reports Enclosed
1: Cytotoxicity test report
2: Sensitization test report
3: Irritation or Intracutaneous reactivity test report
4: Systemic toxicity (acute) test report
5: Material Mediated Pyrogenicity test report
Relisys Medical Devices Limited Doc #: 015-0101-4.2
PRE CLINICAL STUDY REPORT FOR
Rev. No: 01 Page 13 of 13
POST DILATATION CATHETERS
6. Revision History
2. 06/08/2020 01 Updated with Material Mediated Pyrogenicity Test results Sr. Manager – QA & QC
Approval: