0% found this document useful (0 votes)
81 views1 page

Incubator (Lab)

The document outlines a series of inspection and maintenance steps for an incubator, including checks for damage, cleaning, and verifying the operation of electrical components. It specifies safety tests for ground wire resistance and chassis leakage, with guidelines for addressing minor deficiencies and documenting major repairs. The control number for this maintenance procedure is 1899.

Uploaded by

Ayman Hassan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
81 views1 page

Incubator (Lab)

The document outlines a series of inspection and maintenance steps for an incubator, including checks for damage, cleaning, and verifying the operation of electrical components. It specifies safety tests for ground wire resistance and chassis leakage, with guidelines for addressing minor deficiencies and documenting major repairs. The control number for this maintenance procedure is 1899.

Uploaded by

Ayman Hassan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
You are on page 1/ 1

DESCRIPTION: Incubator (Lab) CONTROL NUMBER: 1899

1. Inspect exterior of equipment for damage or missing hardware.


2. Inspect the power cord, strain relief and plug/s for any signs of damage.
3. Turn unit off, open user accessible covers and inspect unit for damage.
4. Clean unit interior components and exterior with vacuum or compressed air.
5. Inspect interior for signs of corrosion or missing hardware. Repair as required.
6. Inspect electrical components for signs of excessive heat or deterioration.
7. Inspect heating element for corrosion or soft spots in jacket.
8. Verify calibration of temperature controller.
9. Verify correct operation of all buttons, controls, displays and/or indicators.

Measure ground wire resistance (Electrical safety test #3, 0.50 ohm maximum)

Measure chassis leakage to ground (Electrical safety test #4, 100 microamperes maximum)

If minor deficiencies are found and can easily be corrected, make correction or adjustment and
document in comments section of this form. Generate a corrective maintenance form for major