DAC Universal
DAC Universal D / DAC Universal S

Instructions for use

English
Dear Customer, Dentsply Sirona
Instructions for use DAC Universal

Dear Customer,
Thank you for purchasing a DAC Universal.
The DAC Universal cleans, lubricates and disinfects / sterilizes up to 6
instruments automatically.
The DAC Universal saves you time-consuming manual reprocessing
and reliably protects you, your team and your patients from cross-
contamination.
These instructions for use are designed to assist you prior to initial use
and whenever you require information later on. It is important to observe
all safety information to prevent personal injury and material damage.
To care for and clean the DAC Universal perform the steps described in
these instructions for use.
We wish you success and enjoyment with DAC Universal.
Your DAC Universal team

67 46 429 D3708
2 D3708.201.04.02.02 10.2020
Dentsply Sirona Table of Contents
Instructions for use DAC Universal

Table of Contents
1 General data ........................................................................................................... 7
1.1 Contact information...................................................................................... 7
1.2 General notes .............................................................................................. 7
1.3 Electronic "User E-Guide" ............................................................................ 7
1.4 Structure of the document............................................................................ 8
1.4.1 Identification of the danger levels .................................................. 8
1.4.2 Formats and symbols used............................................................ 8
1.5 Definitions .................................................................................................... 9
1.6 Scope of these operating Instructions.......................................................... 9
1.7 Warranty and liability.................................................................................... 10
1.8 Intended use ................................................................................................ 11
1.9 Indications.................................................................................................... 11
1.10 Contraindications ......................................................................................... 11
1.11 Scope of delivery ......................................................................................... 12

2 Safety instructions .................................................................................................. 14


2.1 Electrostatic discharge................................................................................. 15
2.2 Caution symbols on the unit......................................................................... 16
2.3 Flammable materials.................................................................................... 16
2.4 Reducing the risk of burns ........................................................................... 17
2.5 Instruction of practice personnel .................................................................. 17
2.6 Installation requirements.............................................................................. 18
2.7 Maintenance ................................................................................................ 19
2.8 Magnets ....................................................................................................... 20

3 Product description ................................................................................................. 21


3.1 Functions ..................................................................................................... 21
3.1.1 Blue lid program............................................................................. 23
3.1.2 Green lid program .......................................................................... 23
3.1.3 Pink lid program ............................................................................. 24
3.1.4 White lid program........................................................................... 24
3.1.5 PCD test program .......................................................................... 25
3.2 Control and functional elements of the DAC Universal................................ 26
3.2.1 Front of the device ......................................................................... 26
3.2.2 Back of unit .................................................................................... 27
3.2.3 Lid .................................................................................................. 28
3.2.4 Display and touch panel................................................................. 30
3.2.5 Safety ring...................................................................................... 30

67 46 429 D3708
D3708.201.04.02.02 10.2020 3
Table of Contents Dentsply Sirona
Instructions for use DAC Universal

3.2.6 Front flap........................................................................................ 30


3.2.7 Water tank...................................................................................... 31
3.2.8 NitramOil #2 lubrication concentrate.............................................. 33
3.2.9 Error LED ....................................................................................... 34
3.3 Programs ..................................................................................................... 35
3.4 Technical data.............................................................................................. 36
3.5 Presentation of the sterilization process ...................................................... 38

4 Operation ................................................................................................................ 40
4.1 Operation of the touchscreen....................................................................... 40
4.1.1 Operating structure / menu guidance............................................. 40
4.1.1.1 Virtual function keys ..................................................... 40
4.1.1.2 Structure of the start screen ......................................... 41
4.1.2 Symbols ......................................................................................... 42
4.1.2.1 Navigation symbols ...................................................... 42
4.1.2.2 Menu symbols .............................................................. 44
4.1.2.3 Process symbols .......................................................... 48
4.1.2.4 Notice symbols for cleaning and maintenance ............. 55
4.1.2.5 Error symbols ............................................................... 55
4.1.3 Menu options ................................................................................. 60
4.1.3.1 Manual control .............................................................. 61
4.1.3.2 Settings......................................................................... 62
4.1.3.3 History .......................................................................... 72
4.1.3.4 Information.................................................................... 75
4.1.3.5 Check & Clean.............................................................. 75
4.1.3.6 Default settings ............................................................ 75
4.2 Reprocessing in the DAC Universal............................................................. 76
4.2.1 Switching on, off and heating up the device .................................. 76
4.2.2 Filling the water tank ...................................................................... 77
4.2.3 Removing the drain tank ................................................................ 78
4.2.4 Pre- and postprocessing of the various instruments...................... 80
4.2.5 Reprocessing with the Blue lid....................................................... 81
4.2.5.1 Start cycle with the Blue lid .......................................... 81
4.2.6 Reprocessing with the Green lid .................................................... 84
4.2.6.1 Start cycle with the Green lid........................................ 84
4.2.7 Reprocessing with the Pink lid ....................................................... 87
4.2.7.1 Start cycle with the Pink lid .......................................... 87
4.2.8 Reprocessing with the White lid..................................................... 90
4.2.8.1 Start cycle with the White lid ........................................ 90
4.2.9 Batch control with the chemical indicator....................................... 94

67 46 429 D3708
4 D3708.201.04.02.02 10.2020
Dentsply Sirona Table of Contents
Instructions for use DAC Universal

4.2.10 Periodic testing with the PCD test device ...................................... 97


4.2.11 Process documentation ................................................................. 100
4.2.11.1 RS-232 interface........................................................... 100
4.2.11.2 LAN interface................................................................ 107
4.2.12 Activities during longer downtimes................................................. 107

5 Cleaning, maintenance and validation .................................................................... 108


5.1 Check & Clean process ............................................................................... 108
5.1.1 Start Check & Clean ...................................................................... 111
5.1.2 Put on gloves ................................................................................. 111
5.1.3 Clean the drain filter....................................................................... 112
5.1.4 Check the O-rings on the drain filter and change if necessary ...... 114
5.1.5 Replace the drain filter ................................................................... 116
5.1.6 Insert Check & Clean lid ................................................................ 118
5.1.7 Check the safety ring ..................................................................... 118
5.1.8 Check the Check & Clean lid ......................................................... 119
5.1.9 Check the water supply to the chamber......................................... 119
5.1.10 Check the oil supply via the Check & Clean lid.............................. 119
5.1.11 Check the water supply via the Check & Clean lid ........................ 120
5.1.12 Open chamber / remove Check & Clean lid................................... 121
5.1.13 Check the permeability of the media paths of all lids..................... 121
5.1.14 Disassemble and clean the water tank .......................................... 124
5.1.15 Clean the drain tank....................................................................... 126
5.1.16 Check the assembly of the adapters.............................................. 128
5.1.17 Check O-rings on the adapters ...................................................... 129
5.1.18 Change the O-rings on the adapters.............................................. 130
5.1.19 Clean the housing of the unit / the lid and the chamber................. 130
5.1.20 Check the O-ring on the water tank ............................................... 131
5.1.21 Assemble / fill and insert the water tank ........................................ 132
5.1.22 Preparations after the Check & Clean process.............................. 133
5.2 Biennial Service ........................................................................................... 135
5.3 Install a new lid ............................................................................................ 135
5.4 Mounting adapters ....................................................................................... 136
5.5 Adding/replacing the lubrication concentrate ............................................... 136
5.6 Drain the chamber ....................................................................................... 138
5.7 Validation ..................................................................................................... 138
5.8 Uninstallation and transport ......................................................................... 139

67 46 429 D3708
D3708.201.04.02.02 10.2020 5
Table of Contents Dentsply Sirona
Instructions for use DAC Universal

6 Errors / Display messages ...................................................................................... 142


6.1 Warning messages ...................................................................................... 142
6.2 Process error messages as pop-up messages............................................ 143
6.3 Error messages as pop-up messages ......................................................... 144
6.4 User messages ............................................................................................ 145
6.5 Display messages and what they mean ...................................................... 147
6.6 Errors without a message ............................................................................ 151

7 Lid, adapter, consumables, spare parts, and tools ................................................. 152

8 Disposal .................................................................................................................. 153

67 46 429 D3708
6 D3708.201.04.02.02 10.2020
Dentsply Sirona 1 General data
Instructions for use DAC Universal 1.1 Contact information

1 General data

1.1 Contact information


Customer service center For technical questions you can call our Customer Service Center at:
+49 (0)6251/16-1670
Manufacturer Sirona Dental Systems GmbH
Fabrikstraße 31
64625 Bensheim
Germany
www.dentsplysirona.com

1.2 General notes


Keep documents safe Always keep the instructions for use handy in case you or another user
require(s) information at a later point of time. If possible, save the
instructions for use on the PC or print them out.
If you sell the unit, make sure that the instructions for use are included
with it either as a hard copy or on an electronic storage device so that
the new owner can familiarize himself with its functions and the
specified warning and safety information.
Online portal for technical documents We have set up an online portal for the technical documents at
dentsplysirona.com/ifu . From here, you can download these
instructions for use along with other documents. Please complete the
online form if you would like a hard copy of a particular document. We
would be delighted to send you a printed copy, free of charge.
Help If you reach an impasse despite having thoroughly studied the
instructions for use, please contact your dental depot.

1.3 Electronic "User E-Guide"


A "DAC Universal S" USB stick is included with the DAC Universal
where all instructions for use and important technical documents are
stored.

67 46 429 D3708
D3708.201.04.02.02 10.2020 7
1 General data Dentsply Sirona
1.4 Structure of the document Instructions for use DAC Universal

1.4 Structure of the document


1.4.1 Identification of the danger levels
To prevent personal injury and material damage, please observe the
warning and safety information provided in these instructions for use.
Such information is highlighted as follows:

DANGER
An imminent danger that could result in serious bodily injury or death.

WARNING
A possibly dangerous situation that could result in serious bodily
injury or death.

CAUTION
A possibly dangerous situation that could result in slight bodily injury.

NOTE
A possibly harmful situation which could lead to damage of the
product or an object in its environment.

IMPORTANT
Application instructions and other important information.

Tip: Information for simplifying work.

1.4.2 Formats and symbols used


The formats and symbols used in this document have the following
meaning:

Prerequisite Requests you to do something.


1. First action step
2. Second action step
or
➢ Alternative action
Result
➢ Individual action step
See "Formats and symbols used Identifies a reference to another
[→ 8]" text passage and specifies its
page number.
● List Designates a list.
"Command / menu item" Indicates commands / menu
items or quotations.

67 46 429 D3708
8 D3708.201.04.02.02 10.2020
Dentsply Sirona 1 General data
Instructions for use DAC Universal 1.5 Definitions

1.5 Definitions
This section shows examples for the use of various terms.
Operation/step Drain the chamber, manually closing the DAC Universal
Internal cleaning, disinfection
Process Check & Clean process
Program Blue lid program
Green lid program
Pink lid program
White lid program
Cycle Reprocessing cycle, for example with the Blue lid

1.6 Scope of these operating Instructions


Device variants This operating manual is valid for the following devices:
DAC Universal with the following lids:
● Blue lid
● Green lid
● Pink lid
● White lid
Equipment options This document describes the full version of your system. It may
therefore cover components that are not included in the system you
purchased.
These instructions for use may contain information on products not
approved in your country.

67 46 429 D3708
D3708.201.04.02.02 10.2020 9
1 General data Dentsply Sirona
1.7 Warranty and liability Instructions for use DAC Universal

1.7 Warranty and liability


The manufacturer guarantees that the product delivered to you (DAC
Universal) was subjected to a thorough quality test before it left the
factory.
A warranty period of 12 months is given for DAC Universal starting on
the day it is delivered.

Warranty
The warranty does not cover defective products:
● That were serviced by service technicians not authorized by the
manufacturer
● If non-original spare parts have been used
● That have been handled improperly or have been damaged while in
transit
● That have been used in a way other than described in the
instructions for use
● For which the prescribed maintenance was not carried out
● That no longer function because water conductivity exceeded
9.0 µS/cm
The manufacturer retains the right to make changes/improvements of all
kinds to the product manufactured or sold by the manufacturer without
being obligated to apply these changes to products that were
manufactured or sold by the manufacturer before this time.

Exclusion of liability
The manufacturer assumes no liability whatsoever in cases involving:
● Service or repairs made by other persons who are not authorized by
the manufacturer or dealer
● Use of the DAC Universal that does not comply with its intended use
● Use of non-original components or parts that are not included in "Lid,
adapter, consumables, spare parts, and tools [→ 152]"
● Instruments already damaged before reprocessing

67 46 429 D3708
10 D3708.201.04.02.02 10.2020
Dentsply Sirona 1 General data
Instructions for use DAC Universal 1.8 Intended use

1.8 Intended use


This medical decive serves the hygenic reprocessing of dental
instruments.

1.9 Indications
None

1.10 Contraindications
None

67 46 429 D3708
D3708.201.04.02.02 10.2020 11
1 General data Dentsply Sirona
1.11 Scope of delivery Instructions for use DAC Universal

1.11 Scope of delivery


Scope of delivery DAC Universal D The following components are included in the scope of delivery of the
DAC Universal:
● Blue lid
● 1 bottle of NitramOil #2 lubrication concentrate
● Combination filter (sterile filter and air filter)
● Water filter set
● Hose
● Screwdriver for adapter
● Power cable
● Check & Clean:
- Check & Clean lid
- Check & Clean cap
- NitraClean cleaning tablets
- Syringe
- Cotton wool rolls
- Screwdriver for drain filter
- Drain filter incl. O-rings
● "DAC Universal D"
USB stick- Instructions for use
- Instructions for use for adapters
- Installation instructions
- Installation and handover report
The adapters for the Blue lid can be freely selected and must be
purchased separately. For an overview of available adapters, see the
separate instructions for use for the adapters.

67 46 429 D3708
12 D3708.201.04.02.02 10.2020
Dentsply Sirona 1 General data
Instructions for use DAC Universal 1.11 Scope of delivery

DAC Universal S scope of delivery The following components are included in the scope of delivery of the
DAC Universal S:
● Pink lid
● 1 bottle of NitramOil #2 lubrication concentrate
● Combination filter (sterile filter and air filter)
● Water filter set
● Hose
● Screwdriver for adapter
● Power cable
● Chemical indicators (pack of 30)
● Indicator holder
● Check & Clean:
- Check & Clean lid
- Check & Clean cap
- NitraClean cleaning tablets
- Syringe
- Cotton wool rolls
- Screwdriver for drain filter
- Drain filter incl. O-rings
● "DAC Universal S" USB stick
- Instructions for use
- Instructions for use for adapters
- Installation instructions
- Installation and handover report
The adapters for the Pink lid can be freely selected and must be
purchased separately. For an overview of available adapters, see the
separate instructions for use for the adapters.

67 46 429 D3708
D3708.201.04.02.02 10.2020 13
2 Safety instructions Dentsply Sirona
Instructions for use DAC Universal

2 Safety instructions
IMPORTANT
This product may be operated only by trained personnel. Please read
the technical documents (such as the instructions for use) carefully
before operating the DAC Universal. Save these documents for
further reference. Follow all warning and safety notices.

IMPORTANT
Immediately notify the manufacturer and the national competent
authority of any serious incident that has occurred with the product.

67 46 429 D3708
14 D3708.201.04.02.02 10.2020
Dentsply Sirona 2 Safety instructions
Instructions for use DAC Universal 2.1 Electrostatic discharge

2.1 Electrostatic discharge


WARNING
Risk of contamination!
Electrostatic discharge can affect the cleaning outcome.

Protective measures
ESD stands for ElectroStatic Discharge.
Electrostatic discharge from people can damage electronic components
when the components are touched. Damaged components usually have
to be replaced. Repairs must be performed by qualified personnel.
Measures to protect against ESD include:
● Procedures to avoid electrostatic charging via
– air conditioning
– air humidification
– conductive floor coverings
– non-synthetic clothing
● discharging the electrostatic charges from your own body through
contact with
– a metallic unit casing
– a larger metallic object
– any other metal part grounded with the protective earth
● Wearing an antistatic band that creates a connection between the
body and a protective ground wire.
Areas at risk are indicated on the unit with the ESD warning label:
We recommend that all persons working with this system are made
aware of the significance of the ESD warning label. A training course
should also be held to inform users about the physics of electrostatic
charges.

Physics of electrostatic charges


An electrostatic discharge requires prior electrostatic charging.
There is a danger of electrostatic charges building up whenever two
bodies rub against each other, e.g. when:
● walking (soles of shoes against the floor) or
● moving (chair casters against floor).
The amount of charge depends on several factors. The charge is:
● higher at low air humidity than at high air humidity, and
● higher with synthetic materials than with natural materials (clothing,
floor coverings).
The following rule of thumb can be applied to assess the transient
voltages resulting from an electrostatic discharge.
An electrostatic discharge is:
● perceptible at 3,000 V or higher
● audible at 5,000 V or higher (cracking, crackling)
● visible at 10,000 V or higher (arc-over)

67 46 429 D3708
D3708.201.04.02.02 10.2020 15
2 Safety instructions Dentsply Sirona
2.2 Caution symbols on the unit Instructions for use DAC Universal

The transient currents resulting from these discharges have a


magnitude of over 10 amps. They are not hazardous for humans
because they last for only several nanoseconds.
Note: nanosecond = 1/1,000,000,000 second = 1 billionth of a second
Voltage differentials exceeding 30,000 volts per centimeter may lead to
a charge transfer (electrostatic discharge, lightning, spark-over).
Integrated circuits (logical circuits and microprocessors) are used in
order to implement a wide variety of functions in a device. The circuits
must be miniaturized to a very high degree in order to include as many
functions as possible on these chips. This leads to structure thicknesses
as low as a few ten thousandths of a millimeter. Integrated circuits that
are connected to wires leading externally are therefore particularly at
risk from electrostatic discharge.
Even voltages that are imperceptible to the user can cause spark-over
of the layers. The resulting discharge current melts the chip in the
affected areas. Damage to individual integrated circuits may cause
malfunction or failure of the unit.

2.2 Caution symbols on the unit


Meaning: Warning of danger
This sign is located on the rear of the DAC Universal. The external
computer hardware that is connected to the serial R232 interface and
LAN interface must satisfy the following requirements: "The connected
unit must comply with the UL 1950 and IEC/EN 60950 standards and
must be connected only to safety extra low voltage (SELV) circuits."
Meaning: Warning of a hot surface

Meaning: Warning of electrostatic discharge, see section "Electrostatic


discharge [→ 15]"

2.3 Flammable materials


WARNING
Risk of explosion!
Never use the DAC Universal in the presence of flammable
anesthetics or gases.

WARNING
Risk of explosion!
Do not hold the NitramOil #2 bottles near an open flame.

67 46 429 D3708
16 D3708.201.04.02.02 10.2020
Dentsply Sirona 2 Safety instructions
Instructions for use DAC Universal 2.4 Reducing the risk of burns

2.4 Reducing the risk of burns


CAUTION
Risk of burns!
The bottom plate and back panel of the DAC Universal can be hot.

CAUTION
Burning hazard!
Never hold your head directly over the DAC Universal when it opens,
as steam may emerge from the unit.

CAUTION
Burning hazard!
Even when the lid of the DAC Universal is partly or completely open,
some instruments may still be hot. Allow the instruments that become
very hot during the process to cool down.

CAUTION
Risk of burns!
The metal part of the lid is very hot. Always touch only the plastic part
of the lid.

CAUTION
Risk of burns!
The drain hose, filter housing, drain filter and drain tank are very hot
during the reprocessing cycle. Switch the DAC Universal off and let it
cool down (about 15 minutes) before emptying and cleaning the drain
tank or cleaning or replacing the drain filter.

CAUTION
Risk of burns!
Note that the temperature of the drain water is hot (approx. 100 °C).

CAUTION
Risk of burns!
The drain tank may contain hot water. Make sure that you do not tip it
over.

2.5 Instruction of practice personnel


WARNING
Risk of contamination!
The personnel responsible for the operation and maintenance of the
DAC Universal must have received instruction on the safe operation
of the DAC Universal from a service technician trained by the
manufacturer. Following the initial instruction from the service
technician, a party must be named in the practice who is responsible
for supporting and training the personnel charged with operating and
maintaining the DAC Universal.

67 46 429 D3708
D3708.201.04.02.02 10.2020 17
2 Safety instructions Dentsply Sirona
2.6 Installation requirements Instructions for use DAC Universal

2.6 Installation requirements


CAUTION
Risk of crushing!
The overall height of the open DAC Universal is 59 cm.
The recommended minimum height for installing the DAC Universal is
70 cm, in order to prevent crushing and injuries when removing the lid
and hitting any hanging cabinets.

WARNING
Risk of contamination!
Please note that the DAC Universal must have warmed up to room
temperature before being operated to prevent the cleaning results
from being impaired.
The DAC Universal may NOT be installed in operating state in rooms
with temperatures below 0°C, otherwise ice may form.

WARNING
Risk of contamination!
The DAC Universal my only be operated at ambient temperatures of
+16°C to +30°C.

WARNING
Risk of contamination!
The DAC Universal must not be installed and operated in direct
sunlight as this would influence the cleaning result.

WARNING
Risk of contamination!
The DAC Universal may only be operated up to an altitude of 2,000 m
above sea level.

WARNING
Risk of contamination!
The relative humidity must be below 80%.

WARNING
Risk of crushing!
The DAC Universal must be placed on a stable installation site.

WARNING
Risk of contamination!
Only ISO 7494-2 dry compressed air may be used for the DAC
Universal.

67 46 429 D3708
18 D3708.201.04.02.02 10.2020
Dentsply Sirona 2 Safety instructions
Instructions for use DAC Universal 2.7 Maintenance

WARNING
Risk of contamination!
There is a risk of infection when no sterile filter is installed.
The sterile filter enclosed must be installed properly. Only sterile
filters approved by Dentsply Sirona may be used.

WARNING
Risk of contamination!
There is a risk of contamination if the sterile filter is missing, is not
installed or an incorrect filter has been installed.
The sterile filter must be properly installed (see detailed directions).

DANGER
Electrical shock!
Only the power plugs supplied or original spare parts may be used.
These meet the following requirements: 125 V/min 15 A or 250 V/min
10 A.

WARNING
Risk of contamination!
Only ISO 7494-2 dry compressed air may be used for the DAC
Universal.

WARNING
Risk of contamination!
There is a risk of infection when no sterile filter is installed.
The sterile filter enclosed must be installed properly. Only sterile
filters approved by Dentsply Sirona may be used.

2.7 Maintenance
CAUTION
Risk of contamination!
Please observe the regulations applicable in your country when
handling the drain tank and drain filter.
The manufacturer recommends that you wear your personal
protective gear when handling the drain tank and drain filter.

WARNING
Risk of contamination!
When you move the DAC Universal make sure that the drain hose is
not bent as otherwise the working room may be contaminated.

67 46 429 D3708
D3708.201.04.02.02 10.2020 19
2 Safety instructions Dentsply Sirona
2.8 Magnets Instructions for use DAC Universal

2.8 Magnets
WARNING
Magnets!
Magnets are integrated in the lids, the front flap of the water tank and
the flap on the drain filter. These should not be placed directly on the
body in order to exclude any interference with e.g. pacemakers.

67 46 429 D3708
20 D3708.201.04.02.02 10.2020
Dentsply Sirona 3 Product description
Instructions for use DAC Universal 3.1 Functions

3 Product description

3.1 Functions
WARNING
Risk of contamination!
Only the loads listed may be reprocessed in the DAC Universal.

Function Blue lid The unit is intended to clean, lubricate and thermally disinfect dental
handpieces. Dental handpieces are defined as rotating instruments split
up into the categories straight handpieces, contra angles and turbines.

IMPORTANT
Adapt a maximum of 5 instruments without spray channels on the
Blue lid. Six instruments without spray channels cannot be
reprocessed on the Blue lid as the water which is needed for internal
cleaning is then unable to flow out. Error 1506 is issued.

The number of different adapters is optional.


Please refer to the adapter instructions for use for details on the
placement of the different adapters.
Function Green lid The unit is intended to clean and thermally disinfect dental sonic
handpieces or air scalers.
The unit is intended to clean and thermally disinfect dental ultrasonic
handpieces.
The unit is intended to clean and thermally disinfect tips for dental sonic
handpieces and air scalers including the associated (torque) wrenches.
The unit is intended to clean and thermally disinfect dental ultrasonic
scaler tips including the associated (torque) wrenches.
The unit is intended to clean and thermally disinfect nozzles for
multifunctional syringes.
The unit is intended to clean and thermally disinfect nozzles for power
jet units and powder jet handpieces.
The number and placement of the different adapters is optional.
Exception: A maximum of 3 ultrasonic scaler tips with attached torque
wrench from Sirona, EMS and/or Satalec can be positioned
simultaneously on the Green lid. The ultrasonic scaler tip adapters
cannot be positioned directly next to each other. Other Green lid
adapters must be installed between the ultrasonic scaler tip adapters.

67 46 429 D3708
D3708.201.04.02.02 10.2020 21
3 Product description Dentsply Sirona
3.1 Functions Instructions for use DAC Universal

Function Pink lid The unit is intended to clean, lubricate and thermally sterilize
unpackaged dental handpieces. Dental handpieces are defined as
rotating instruments split up into the categories straight handpieces,
contra angles and turbines.

IMPORTANT
Adapt a maximum of 5 instruments without spray channels on the
Pink lid. Six instruments without spray channels cannot be
reprocessed on the Pink lid as the water which is needed for internal
cleaning is then unable to flow out. Error 1506 is issued.

The number of different adapters is optional.


Please refer to the adapter instructions for use for details on the
placement of the different adapters.
Function White lid The unit is intended to clean and thermally sterilize unpackaged dental
sonic handpieces or air scalers.
The unit is intended to clean and thermally sterilize unpackaged dental
ultrasonic handpieces.
The unit is intended to clean and thermally sterilize unpackaged tips for
dental sonic handpieces and air scalers including the associated
(torque) wrenches.
The unit is intended to clean and thermally sterilize unpackaged dental
ultrasonic scaler tips including the associated (torque) wrenches.
The unit is intended to clean and thermally sterilize unpackaged nozzles
for multifunctional syringes.
The unit is intended to clean and thermally sterilize unpackaged nozzles
for powder jet units and powder jet handpieces.
The number and placement of the different adapters is optional.
Exception: A maximum of 3 ultrasonic scaler tips with attached torque
wrench from Sirona, EMS and/or Satalec can be positioned
simultaneously on the White lid. The ultrasonic scaler tip adapters
cannot be positioned directly next to each other. Other White lid
adapters must be installed between the ultrasonic scaler tip adapters.

67 46 429 D3708
22 D3708.201.04.02.02 10.2020
Dentsply Sirona 3 Product description
Instructions for use DAC Universal 3.1 Functions

3.1.1 Blue lid program


The turbines, straight and contra-angle handpieces are reprocessed as
follows using the adapters on the Blue lid:
1. Preliminary cleaning
2. Leak test
3. Internal cleaning: The internal spray and drive channels are cleaned
with water.
4. Lubrication: The drive channels are lubricated (sufficient for the next
treatment)
5. External cleaning: Done using the pulse-wash method (multicyclic
cleaning method)
6. Heating up to 134°C
7. Back-flush: Saturated steam is directed through the instruments
8. Disinfection: 0.5 min. at 134°C
9. Cooling
10. The lid opens slightly
11. The lid can now be open fully

3.1.2 Green lid program


The Green lid instruments are reprocessed as follows using the
appropriate adapters on the Green lid:
1. Preliminary cleaning
2. Leak test
3. Internal cleaning: The internal spray channels are rinsed with water
4. External cleaning: Done using the pulse-wash method (multicyclic
cleaning method)
5. Heating up to 134°C
6. Back-flush: Saturated steam is directed through the instruments
7. Disinfection: 0.5 min. at 134°C
8. Cooling
9. The lid opens slightly
10. The lid can now be open fully

67 46 429 D3708
D3708.201.04.02.02 10.2020 23
3 Product description Dentsply Sirona
3.1 Functions Instructions for use DAC Universal

3.1.3 Pink lid program


The turbines, straight and contra-angle handpieces are reprocessed as
follows using the adapters on the Pink lid:
1. Preliminary cleaning
2. Leak test
3. Internal cleaning: The internal spray and drive channels are cleaned
with water.
4. Lubrication: The drive channels are lubricated (sufficient for the next
treatment)
5. External cleaning: Done using the pulse-wash method (multicyclic
cleaning method)
6. Heating up to 134°C
7. Back-flush: Saturated steam is directed through the instruments
8. Unpackaged sterilization: 3 min. at 134°C
9. Cooling
10. The lid opens slightly
11. The lid can now be open fully

3.1.4 White lid program


The White lid instruments are reprocessed as follows using the
appropriate adapters on the White lid:
1. Preliminary cleaning
2. Leak test
3. Internal cleaning: The internal spray channels are rinsed with water.
4. External cleaning: Done using the pulse-wash method (multicyclic
cleaning method)
5. Heating up to 134°C
6. Back-flush: Saturated steam is directed through the instruments
7. Unpackaged sterilization: 3 min. at 134°C
8. Cooling
9. The lid opens slightly.
10. The lid can now be open fully.

67 46 429 D3708
24 D3708.201.04.02.02 10.2020
Dentsply Sirona 3 Product description
Instructions for use DAC Universal 3.1 Functions

3.1.5 PCD test program


The PCD test program can be performed with either a Pink lid or a
White lid, also see section "Periodic testing with the PCD test device
[→ 97]". The process steps for the PCD test program are as follows:
1. Leak test
2. Heating up to 134°C
3. Unpackaged sterilization: 3 min. at 134°C
4. Cooling
5. The lid opens slightly.
6. The lid can now be open fully.

67 46 429 D3708
D3708.201.04.02.02 10.2020 25
3 Product description Dentsply Sirona
3.2 Control and functional elements of the DAC Universal Instructions for use DAC Universal

3.2 Control and functional elements of the DAC


Universal
3.2.1 Front of the device

A Front flap
B Oil bottle
C Drain filter and filter housing
D Water tank
E Display and touch panel

67 46 429 D3708
26 D3708.201.04.02.02 10.2020
Dentsply Sirona 3 Product description
Instructions for use DAC Universal 3.2 Control and functional elements of the DAC Universal

3.2.2 Back of unit

A Main switch
B LAN interface
C Power supply connection
D Standard interface RS232
E Water connection
F Compressed air connection
G Drain connection

67 46 429 D3708
D3708.201.04.02.02 10.2020 27
3 Product description Dentsply Sirona
3.2 Control and functional elements of the DAC Universal Instructions for use DAC Universal

3.2.3 Lid
The following lids can be used for reprocessing in the DAC Universal:
● Blue lid for reprocessing turbines, straight and contra-angle
handpieces.
See "Blue lid program [→ 23]" for functions and see "Reprocessing
with the Blue lid [→ 81]"
for instructions.
The maximum load on the Blue lid is 6 instruments.

Blue lid
Water consumption for cleaning approx. 650 ml
Water consumption for disinfection approx. 150 ml
Holding time 30 s
Holding temperature 134 °C
A0-value >3000

● Green lid for reprocessing Green lid instruments.


See "Green lid program [→ 23]" for functions and see "Reprocessing
with the Green lid [→ 84]"
for instructions.
The maximum load on the Green lid is 6 instruments.

Green lid
Water consumption for cleaning approx. 500 ml
Water consumption for disinfection approx. 150 ml
Holding time 30 s
Holding temperature 134 °C
A0-value >3000

67 46 429 D3708
28 D3708.201.04.02.02 10.2020
Dentsply Sirona 3 Product description
Instructions for use DAC Universal 3.2 Control and functional elements of the DAC Universal

● Pink lid for reprocessing turbines, straight and contra-angle


handpieces.
See "Programs [→ 35]" for functions and see "Reprocessing with
the Pink lid [→ 87]"
for instructions.
The maximum load on the Pink lid is 6 instruments.

Pink lid
Water consumption for cleaning approx. 650 ml
Water consumption for unpackaged steriliza- approx. 250 ml
tion
Holding time 3 min
Holding temperature 134°C

● White lid for reprocessing White lid instruments.


See "Programs [→ 35]" for functions and see "Reprocessing with
the White lid [→ 90]"
for instructions.
The maximum load on the White lid is 6 instruments.

White lid
Water consumption for cleaning approx. 500 ml
Water consumption for unpackaged steriliza- approx. 250 ml
tion
Holding time 3 min
Holding temperature 134°C

67 46 429 D3708
D3708.201.04.02.02 10.2020 29
3 Product description Dentsply Sirona
3.2 Control and functional elements of the DAC Universal Instructions for use DAC Universal

3.2.4 Display and touch panel

See chapter "Operation of the touchscreen [→ 40]" for more


information about the DAC Universal display and menu.

3.2.5 Safety ring


The DAC Universal is equipped with a safety ring. This prevents the
user from pinching their fingers when closing the lid.
If there is an object between the chamber and the lid when the lid
closes, the cycle is stopped immediately. The lid is then lifted, moves up
fully and the message 1003 is displayed.
To remedy the message, see "Errors / Display messages [→ 142]"

3.2.6 Front flap

The front flap (A) is located on the front of the DAC Universal. When
this is opened, the water tank and NitramOil#2 oil bottle can be seen.
The front flap must always be closed before a reprocessing cycle is
started. If it is open, the display will show an error message that must
be rectified.

67 46 429 D3708
30 D3708.201.04.02.02 10.2020
Dentsply Sirona 3 Product description
Instructions for use DAC Universal 3.2 Control and functional elements of the DAC Universal

3.2.7 Water tank


If no permanent water connection is selected, the integrated water tank
must be filled regularly with water manually, as all functions (e.g.
cleaning) of the DAC Universal are performed with demineralized or
distilled water.
The water tank must be cleaned regularly ‒ weekly or every 50 cycles
(whichever occurs first) ‒ irrespective of the water supply type
("Disassemble and clean the water tank [→ 124]“").

IMPORTANT
The water tank must be in the DAC Universal when using an
automated water connection.

WARNING
Risk of contamination!
The DAC Universal may only be operated with demineralized or
distilled water with a conductivity value of < 3 µS/cm.

NOTE
Only demineralized or distilled water may be used in the DAC
Universal.
Tap water may contain limestone, which damages both the
instruments and the DAC Universal.
Water conductivity should be < 3.0 µS/cm. If the value is > 9.0 µS/cm,
an error message appears and the process stops because it can no
longer be guaranteed that the process is safe.

Filling quantity of water tank for Blue/ A water tank filled to the maximum level is enough for at least 2 cycles
Green lid when using the Blue lid program and the Green lid program.
Capacity of the water tank: approx. 2.3 liters / 0.55 imperial gallons
Minimum filling of the water tank: 1 liter / 0.22 imperial gallons for a Blue
lid program and Green lid program at 134°C (273°F).
Filling quantity of water tank for Pink/ A water tank filled to the maximum level is enough for at least 2 cycles
White lid when using the Pink lid program and the White lid program.
Capacity of the water tank: approx. 2.3 liters / 0.55 imperial gallons
Minimum filling of the water tank: 1 liter / 0.22 imperial gallons for a Pink
lid program and White lid program at 134°C (273°F).

NOTE
Remove the water tank from the DAC Universal before filling it up.
This prevents water from spilling into the inside of the DAC Universal
housing.

67 46 429 D3708
D3708.201.04.02.02 10.2020 31
3 Product description Dentsply Sirona
3.2 Control and functional elements of the DAC Universal Instructions for use DAC Universal

NOTE
Empty the water tank and check whether the water tank filter is
correctly assembled (flush with the water tank base) on the bottom of
the water tank.

67 46 429 D3708
32 D3708.201.04.02.02 10.2020
Dentsply Sirona 3 Product description
Instructions for use DAC Universal 3.2 Control and functional elements of the DAC Universal

3.2.8 NitramOil #2 lubrication concentrate


The NitramOil #2 lubrication concentrate is used to lubricate the drive
channels and the rotor of the turbines, straight and contra-angle
handpieces, and is not screwed into the device upon delivery.
The bottle must be screwed in before the first use of the DAC Universal.
NitramOil #2 is approved by leading international instrument
manufacturers.
One bottle of NitramOil #2 is enough for lubricating approx. 200 cycles
(corresponding to approx. 1200 straight, contra-angle, or highspeed
handpieces).
The lubrication concentrate comes in a no-pressure bottle.

(Safety) information for the lubrication concentrate

NOTE
Too much oil, e.g. from manually adding lubrication, may result in
overlubrication of the instrument.

NOTE
Please lubricate the collet system of your turbines, straight and
contra-angle handpieces regularly as per the lubrication and
operating instructions from the respective instrument manufacturer.

CAUTION
Risk of contamination!
Use only NitramOil #2 for the DAC Universal.

IMPORTANT
In case of fire: There is maximum 200 ml / 0.043 imperial gallons of
oil in the DAC Universal. There is no danger that dangerous or
harmful gases can develop in case of fire.

NOTE
The oil must be stored and handled at temperatures below 50 °C (122
°F). Follow all warnings on the handling, use, and disposal of the oil.
This information is in the safety data sheet. The safety data sheet for
NitramOil #2 is available at www.dentsplysirona.com.

IMPORTANT
No toxic, harmful, or dangerous gases arise during the handling, use,
or removal of NitramOil #2. The oil is pumped directly from the bottle
into the DAC Universal.

67 46 429 D3708
D3708.201.04.02.02 10.2020 33
3 Product description Dentsply Sirona
3.2 Control and functional elements of the DAC Universal Instructions for use DAC Universal

3.2.9 Error LED

The DAC Universal is equipped with an LED lamp.


It flashes in the event of an error. If the error is relevant for the process,
an acoustic alarm signal also sounds.
The LED also flashes at the end of the PCD cycle. This indicates that a
test cycle and not a reprocessing cycle has been performed. This is not
an error message. As such, no acoustic alarm signal sounds.

67 46 429 D3708
34 D3708.201.04.02.02 10.2020
Dentsply Sirona 3 Product description
Instructions for use DAC Universal 3.3 Programs

3.3 Programs
The DAC Universal automatically recognizes which lid has been
inserted into the DAC Universal.
A reprocessing cycle in the DAC Universal is counted once the "Start"
function key is pressed. It also counts if an error appears or the cycle is
discontinued while the lid is being closed.

Program for Blue lid

Tem- max. tem- Absolute Disinfection Total cycle Program


perature perature pressure (in time duration*
bar)
134°C 137°C 3.2 bar 30 sec. ca. 15 min Blue program lid

* with factory settings

Program for Green lid

Tem- max. tem- Absolute Disinfection Total cycle Program


perature perature pressure (in time duration*
bar)
134°C 137°C 3.2 bar 30 sec. ca. 14 min Green program lid

* with factory settings

Program for Pink lid

Tem- max. tem- Absolute Sterilization Total cycle Program


perature perature pressure (in time duration*
bar)
134°C 137°C 3.2 bar 3 min ca. 21 min Pink program lid

* with factory settings

Program for White lid

Tem- max. tem- Absolute Sterilization Total cycle Program


perature perature pressure (in time duration*
bar)
134°C 137°C 3.2 bar 3 min ca. 21 min White program lid

* with factory settings

67 46 429 D3708
D3708.201.04.02.02 10.2020 35
3 Product description Dentsply Sirona
3.4 Technical data Instructions for use DAC Universal

3.4 Technical data


Manufacturer: Sirona Dental Systems GmbH
Model: DAC Universal
Dimensions (H/W/D): Opened: 59 cm / 40 cm / 42 cm
Closed with Blue lid: 40 cm / 40 cm / 42 cm
Materials: Plastic, stainless steel
Weight: (net without packaging) 26 kg / 57 pounds
Weight: (gross with packaging) 34 kg / 75 pounds
Noise level: < 65 dBA
Power supply: 100-127 VAC / 220-240 VAC
Frequency: 50 Hz / 60 Hz
Nominal current: 120 / 230 VAC: 11 A / 6 A
Fluctuations in electrical supply +/- 10 % of nominal voltage
Transient overvoltage Installation category II
Power consumption 1.3 kW
Degree of protection: I
Protection type (IP class): 21B
Type: Not applicable
Type BF applied part: Not applicable
Safety class: Never use the DAC Universal in the presence of
flammable anesthetics or gases.
Operating conditions: Temperature: 16 °C - 30 °C (61 °F - 86 °F)
No operation above 2000 m.
Relative humidity up to 80%
Air pollution level (ISO 8573-1:2010) ≤2
Maximum loss of heat at a surrounding temperature of about 500 W for the whole device
approximate 23 °C (73 °F) about 250 W for the front.
Total chamber volume < 2 liters
Design pressure in the chamber 3.5 bar relative
Maximum allowable working pressure in the chamber 3.5 bar relative
Maximum allowable working temperature in the cham- 137°C (278.6°F)
ber
Usable space in the chamber Height: Approximate 16 cm,
Diameter: Approximate 11 cm
Transport and storage conditions: Temperature: 10 °C - 40 °C (50 °F - 104 °F)
Relative humidity: 30 % - 95 %
Air pressure: 700 hPA - 1,060 hPa
This product complies with Medical Device Directive
93/42/EEC, including all amendments.

0123

67 46 429 D3708
36 D3708.201.04.02.02 10.2020
Dentsply Sirona 3 Product description
Instructions for use DAC Universal 3.4 Technical data

This product is a medical device.

The DAC Universal complies with the requirements of


the RoHS Directive 2011/65/EU.

The DAC Universal S is a Class S sterilizer according to EN 13060.

Symbols on the DAC Universal packaging


This product complies with Medical Device Directive 93/42/EEC,
including all amendments.

0123
This way up

Protect from moisture

Handle with care

Must not be lifted by the hooks

Temperature range for transport and storage conditions: 10°C to +40°C

Stack limit, do not stack more than 2 units on top of each other

67 46 429 D3708
D3708.201.04.02.02 10.2020 37
3 Product description Dentsply Sirona
3.5 Presentation of the sterilization process Instructions for use DAC Universal

3.5 Presentation of the sterilization process


The following graphs show an overview of the process. This process
overview shows the different pressure levels over time during the
sterilization program with the Pink lid and White lid in the DAC
Universal.

Sterilization program Pink lid


The following illustration shows the pressure (bar) development over
time during the sterilization program, with standard settings, holding
time 3 min. and a total cycle time of approx. 21 min.:

Sterilization program

1 Pre-test
2 External pre-cleaning
3 Leak test
4 Internal cleaning of the spray channels
5 Internal cleaning and blowing out of the drive channels
6 Lubrication
7 External cleaning
8 Recleaning
9 Preheating
10 Heating up
11 Back-flush
12 Venting cycles
13 Compensation time
14 Sterilization / holding time (3 min.)
15 Removal of steam and condensation
16 Cooling

67 46 429 D3708
38 D3708.201.04.02.02 10.2020
Dentsply Sirona 3 Product description
Instructions for use DAC Universal 3.5 Presentation of the sterilization process

Sterilization program White lid


The following illustration shows the pressure (bar) development over
time during the sterilization program, with standard settings, holding
time 3 min. and a total cycle time of approx. 21 min.:

Sterilization program

1 Pre-test
2 External pre-cleaning
3 Leak test
4 Internal cleaning of the spray channels
5 External cleaning
6 Recleaning
7 Preheating
8 Heating up
9 Back-flush
10 Venting cycles
11 Compensation time
12 Sterilization / holding time (3 min.)
13 Removal of steam and condensation
14 Cooling

67 46 429 D3708
D3708.201.04.02.02 10.2020 39
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

4 Operation

4.1 Operation of the touchscreen


4.1.1 Operating structure / menu guidance
4.1.1.1 Virtual function keys
The touch screen shows virtual function keys depending on the selected
menu. The desired functions are triggered by pressing the function
buttons with your finger.

Key (background) color


Dark gray: General activatable function

Grayed out: Function cannot be activated

Orange: Active / selected function

67 46 429 D3708
40 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

4.1.1.2 Structure of the start screen

04:00 PM

From the start screen, you can


● Start a cycle (in the middle) using the "Start" function key

● Make further settings in the menu with the "Sub-menu" function key.

● If necessary, start the Check & Clean maintenance workflow (in the
center to the left) with the "Check & Clean 50" or "Check & Clean
200" function keys, see Cleaning and Maintenance.

50 200

The "Back" function key closes the open sub-menu immediately.

67 46 429 D3708
D3708.201.04.02.02 10.2020 41
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

4.1.2 Symbols
4.1.2.1 Navigation symbols
This section describes the symbols used to navigate in the user
interface of the touchscreen.

Return to the start screen

Show submenu

Cancel a process

Enter / confirm

Delete

Navigate up

Navigate up (deactivated)

Navigate back

Navigate down

Navigate down (deactivated)

67 46 429 D3708
42 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

Check & Clean navigation

Navigate to the next step

Navigate to the previous step

Navigate to the next Check & Clean


chapter

Navigate back to the previous Check


& Clean chapter

Cancel a process

Enter / confirm

Starting of the media inflow (e.g. wa-


ter inflow during the Check & Clean
cycle)

Stopping of the media inflow (e.g. wa-


ter inflow during the Check & Clean
cycle)

67 46 429 D3708
D3708.201.04.02.02 10.2020 43
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

4.1.2.2 Menu symbols


This section describes the symbols that appear as menu symbols on
the user interface of the touchscreen.
They are presented and explained in the same order as the menu.

Manual control

Manual control

Manual: Open

Manual: Close

Manual: Drain the chamber

Settings

Settings

Settings: Date / time

Settings: Time

Settings: Time, 12 h
12h
Settings: Time, 24 h
24h
Settings: Time, am
am

67 46 429 D3708
44 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

Settings: Time, pm
pm
Edit

Settings: Date

Settings: Date, day / month / year


D.M.Y

Settings: Display brightness

Settings: Increase display brightness

Settings: Reduce display brightness

Printer

Reprint

Languages

PCD test program

67 46 429 D3708
D3708.201.04.02.02 10.2020 45
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

Settings: Service technician

Settings: Oil quantity

Settings: Oil expel

Settings: Automatic supply of dem-


ineralized water

Settings: Automatic supply of dem-


ineralized water, off
OFF

Settings: Automatic supply of dem-


ineralized water, on
ON

Settings: LAN

Settings: Service software

Settings: Service software, off

OFF

Settings: Service software, on

ON

67 46 429 D3708
46 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

History

History

History: Programs

History: Service

History: Error messages

History: Check & Clean

List, 50 cycles

List, 200 cycles

Information

Information

Check & Clean

Check & Clean (maintenance)

List, 50 cycles

List, 200 cycles

67 46 429 D3708
D3708.201.04.02.02 10.2020 47
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

4.1.2.3 Process symbols


This chapter describes the symbols that describe the processes in the
DAC Universal.

Process: Start

Process: Open chamber

Process: Remaining time until the re-


spective cycle is complete.

Process: Blue lid

67 46 429 D3708
48 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

Process: Green lid

Process: Pink lid

Process: White lid

67 46 429 D3708
D3708.201.04.02.02 10.2020 49
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

Process: PCD with empty lid

Process status symbols

Lid closes

Lid completely closed

Preheat (at the start of the day)

67 46 429 D3708
50 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

Temperature in the chamber is in-


creasing

Temperature stable

Temperature in the chamber is drop-


ping

The lid is open slightly at the end of


the process

67 46 429 D3708
D3708.201.04.02.02 10.2020 51
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

Lid opens up

Slide the lid in the holder

Drain the chamber - manual

The safety ring must be pressed -


within the scope of the Check & Clean

67 46 429 D3708
52 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

Process status symbols for the Blue lid / Pink lid

Reprocessing of dental handpieces


on the Blue lid / Pink lid

Internal cleaning of dental handpieces


on the Blue lid / Pink lid

Lubrication of dental handpieces on


the Blue lid / Pink lid

External cleaning of dental hand-


pieces on the Blue lid / Pink lid

Back-flush of dental handpieces on


the Blue lid / Pink lid

67 46 429 D3708
D3708.201.04.02.02 10.2020 53
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

Process status symbols for the Green lid / White lid

Reprocessing of Green lid / White lid


instruments on the Green lid / White
lid

Internal cleaning of Green lid / White


lid instruments on the Green lid /
White lid

External cleaning of Green lid / White


lid instruments on the Green lid /
White lid

Back-flush of Green lid / White lid in-


struments on the Green lid / White lid

67 46 429 D3708
54 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

Process status symbols PCD test program

Screw the PCD test device onto a


Pink lid or White lid.
No instruments may be reprocessed
on the DAC Universal during the PCD
test program.

4.1.2.4 Notice symbols for cleaning and maintenance


This chapter describes the symbols that only appear when the DAC
Universal needs cleaning or maintenance.

Check & Clean maintenance workflow


after 50 cycles

Check & Clean maintenance workflow


after 200 cycles

4.1.2.5 Error symbols


4.1.2.5.1 Pop-up symbols
This section describes the symbols that appear as menu symbols on
the DAC Universal as a pop-up message.

General pop-ups

Error

Cancel

Warning

Confirm

67 46 429 D3708
D3708.201.04.02.02 10.2020 55
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

User pop-ups

Cancel cycle
Process cannot be started

Insufficient compressed air

Water tank not inserted

Water quality inadequate


The error message appears as of a
value of > 9.0 µS/cm.
The process stops because it can no
longer be guaranteed that the process
is safe.

Front flap is not closed

67 46 429 D3708
56 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

Service, contact a dealer authorized


by Dentsply Sirona

Disconnect from the network


Connect service software

Service, read instructions for use


Service, contact a dealer authorized
by Dentsply Sirona

Initialization pop ups

Pink lid

White lid

Green lid

67 46 429 D3708
D3708.201.04.02.02 10.2020 57
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

4.1.2.5.2 Error symbols


This section describes the symbols that appear on the DAC Universal
as error messages.

Service, contact a dealer authorized


by Dentsply Sirona

Front flap is not closed

Insufficient or no compressed air sup-


ply

No water

Water tank not inserted

67 46 429 D3708
58 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

4.1.2.5.3 Warning symbols


This section describes the symbols that appear on the DAC Universal
as warning messages.

Service warning: Servicing by a tech-


nician is required in the near future.

Oil warning: Oil is used up or the oil


bottle is missing.

Water quality warning: Water quality


is worse than 3 µS/cm.

67 46 429 D3708
D3708.201.04.02.02 10.2020 59
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

4.1.3 Menu options


This section describes the menu options.
The DAC Universal menu is divided into 5 menus (from left to right) as
follows:

04:00 PM

● Manual control
● Settings
● History
● Information
● Check & Clean

67 46 429 D3708
60 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

4.1.3.1 Manual control


The settings menu is divided into 3 sub-menus (from left to right) as
follows:

04:00 PM

● Manual: Open

● Manual: Close

● Manual: Drain the chamber

4.1.3.1.1 Manual: Open


Upon touching the function key "Manual: Open", the DAC Universal is
opened manually.

4.1.3.1.2 Manual: Close


Upon touching the function key "Manual: Close", the DAC Universal is
closed manually.

4.1.3.1.3 Manual: Drain the chamber


Upon touching the function key "Manual: Empty chamber", the DAC
Universal chamber is emptied manually.
Select this function if there is residual water in the chamber that needs
to be removed.
To do so, a lid must be inserted into the DAC Universal.

67 46 429 D3708
D3708.201.04.02.02 10.2020 61
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

It takes approx. 60 seconds to empty the chamber. The lid then opens
again.
The chamber should then be empty. If not, the operation can be
repeated.

4.1.3.2 Settings

04:00 PM

The settings menu is divided into 5 sub-menus (from left to right) as


follows:
● Date and time

● Display brightness

● Printer

● PCD test program

● Technical settings

67 46 429 D3708
62 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

4.1.3.2.1 Date and time

04:00 PM

Upon touching the function key "Date and time", you can set the time
and date.

Setting the time


1. Touching the function key "Time" takes you to the time settings.

2. Select the desired display by pressing the function keys "12h / 24h".

12h
04:00 PM
24h

12h
3. Pressing the function key "Edit" takes you to the time entry.

4. Select the desired time by touching the numbers. The previous time
04:00
is automatically overwritten. You can only enter numbers which are
7 8 9 possible for the relevant time entry.
am
4 5 6
pm
1 2 3

67 46 429 D3708
D3708.201.04.02.02 10.2020 63
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

5. Select the desired display by pressing the function keys "am / pm“".
pm
6. Confirm the time by pressing the "Confirm" function key.

7. In order to navigate back to the homescreen, you can either press


the "Back" or the "Homescreen" function key.
Ä The time is set.

Setting the date


1. Pressing the function key "Date" takes you to the date settings.

2. Pressing the function key "Edit" takes you to the date entry.

3. Select the date (year/month/day) by pressing the numbers. The


2019-02-19
previous date is automatically overwritten.
7 8 9

4 5 6

1 2 3

4. Confirm the time by pressing the "Confirm" function key.

5. You can change the date display by pressing the function key "Day/
Month/Year".

Y.M.D
2019-02-19
D.M.Y

D.M.Y
6. In order to navigate back to the homescreen, you can either press
the "Back" or the "Homescreen" function key.
Ä The date is set.

67 46 429 D3708
64 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

4.1.3.2.2 Display brightness

04:00 PM

70 %

By pressing the function key "Display brightness" you can change the
brightness of the display.
The brightness is adjustable in 10% steps, from 30 – 100%.
The default settings for the display brightness is 30%.
We recommend a maximum display brightness of 60% for a new DAC
Universal. If the brightness is more than 60%, a warning symbol
appears stating that setting the display brightness higher can
prematurely age the display.
If you have an older DAC Universal with a deteriorating display
brightness, you can set the display brightness to more than 60% to
compensate for this.
If the display is no longer readable even when the brightness settings
are adjusted to 100%, please contact your authorized dealer.
To adjust the display’s brightness:
1. Select the desired brightness by pressing the "Plus" or "Minus"
function keys.
2. Confirm your choice by pressing the "Back" function key.

67 46 429 D3708
D3708.201.04.02.02 10.2020 65
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

4.1.3.2.3 Printer

04:00 PM

The printer menu is divided into 2 sub-menus (from left to right) as


follows:
● Reprint

● Printer language settings

Reprint
Printout logs can be prepared for the last 40 reprocessing cycles and
PCD test programs by pressing the "Reprint" function key. The last
reprocessing cycle or last PCD test program is shown at the top of the
list.

04:00 PM

1 0.5 min 10.07.2015


2 3 min 10.07.2015

3 3 min 10.11.2014
4 3 min 03.08.2014

5 0.5 min 03.02.2014

The "Navigate: Up" and the "Navigate: Down" function keys allow you to
select a log from another reprocessing cycle or PCD program.

67 46 429 D3708
66 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

The list contains the following information for each of the 40


reprocessing cycles or PCD test programs:
● Cycle number
● Cycle type (via the color bar; a PCD test program is indicated by a
colored border)
● Holding time (should there be no details on time, this means the
cycle has been aborted)
● Cycle date
To print out a reprocessing cycle, press the "Printer" function key on the
left of the selected cycle.

Printer language settings

04:00 PM

DE GB FR IT

ES DK SE NO

The "Printer language settings" function key can be used to set the
language of the printer log. See "RS-232 interface [→ 100]" for details of
the procedure.
The default setting for the printer language is English (GB).
4.1.3.2.4 PCD test program
Press the function key "PCD test program" to start the PCD test
program for periodic testing of steam penetration.

WARNING
Risk of contamination!
The PCD test program is not a complete reprocessing cycle and is
performed without instruments being attached.

Further information on the PCD test device and periodic testing can be
found in the section "Periodic testing with the PCD test device [→ 97]".

67 46 429 D3708
D3708.201.04.02.02 10.2020 67
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

4.1.3.2.5 Technical settings

04:00 PM

The "Technical settings" menu is divided into 5 sub-menus (from left to


right):
● Settings: Oil quantity

● Settings: Blow out excess oil

● Settings: Automatic supply of demineralized water

● Settings: LAN

● Settings: Service software

67 46 429 D3708
68 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

Settings: Oil quantity

04:00 PM

1 2 3

By pressing the function key "Settings: Oil quantity" you can set the oil
quantity. There are 3 levels which can be selected by pressing the
respective function key. No confirmation is required.
The default setting for the oil quantity is level 1.
The oil quantity can be increased by selecting the other levels.

Settings: Blow out excess oil

04:00 PM

1 2 3 4 5

By pressing the function key "Settings: Blow out excess oil", you can set
the duration of blowing out with compressed air. There are 5 levels
which can be selected by pressing the respective function key. No
confirmation is required.
The default setting for blowing out excess oil is level 1.
The duration of blow out can be increased by selecting the other levels.

67 46 429 D3708
D3708.201.04.02.02 10.2020 69
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

Settings: Automatic supply of demineralized water

04:00 PM

ON OFF

By touching the function key "Settings: Automatic supply of


demineralized water", you can switch the function "Automatic supply of
demineralized water" on and off by pressing the "ON" or "OFF" function
key. No confirmation is required.
The default setting for setting the automatic supply of demineralized
water is "OFF".

Settings: LAN

04:00 PM

DCHP OFF

IP-Adress 192.168.000.010

Sub-Net 255.255.255.000

Click on the "Settings: LAN“ function key to access the network settings.
The DAC Universal supports 2 network settings:
● Manual IP address assignment
● Automatic assignment of IP address by a DHCP server in the
network
If you don't know which setting is correct for the practice network,
contact the network administrator.

Manual IP address assignment

67 46 429 D3708
70 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

To set the IP address manually, the DHCP function must be set to


"OFF".
OFF The following values are set as default settings for the IP address on
the device:
● IP: 192.168.000.010
● Subnet: 255.255.255.000
These values can be changed if needed. To do so, press the "Edit"
function key next to the numbers.

Please note that the numbers will be overwritten from left to right when
entering the figures. The device only accepts numbers that are
generally possible for an IP address. After, the entered address is
confirmed with the "Confirm" function key.
Automatic IP address assignment (DHCP)
In order to be automatically assigned an IP address from the DHCP
server, the function "DHCP" must be set to "ON".
ON As soon as the DAC Universal has an address from the DHCP server,
the IP address and the subnet address are shown in gray. This is purely
an informative display as this address cannot be changed by the DAC
Universal .

Settings: Service software

04:00 PM

ON OFF

By pressing the "Settings: Service software" function key, a connection


between the service software and the DAC Universal can be
established or disconnected by pressing the "ON" or "OFF" function
key.

The default setting is "OFF". This means that no manual control can be
conducted using the service software. Data is still exchanged.
OFF

Manual control of the DAC Universal is now activated. This is necessary


in order to reset the service counter.
ON

67 46 429 D3708
D3708.201.04.02.02 10.2020 71
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

WARNING
Risk of contamination! / Risk of injury due to mechanical energy
(compressed air)! / Burning hazard!
If the DAC Universal is manually controlled via the service software,
there is a particular risk caused by heat, steam or compressed air.

After use, the connection between the service software and the DAC
Universal must be switched off again. For this, the function key "OFF"
OFF must be pressed.

4.1.3.3 History

04:00 PM

The history menu is divided into 4 sub-menus (from left to right) as


follows:
● History: Programs

● History: Service

● History: Error messages

● History: Check & Clean

67 46 429 D3708
72 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

4.1.3.3.1 History: Program

04:00 PM

1 0.5 min 134°C 9999

1 0.5 min 134°C 9999

1 3 min 134°C 9999


1 3 min 134°C 9999

Pressing the "History: Program" function key displays the number of


reprocessing cycles already performed per type of lid.
The sequence and lid programs are as follows (from top to bottom):
● Blue lid: 0.5 minutes, 134 °C, number of cycles
● Green lid: 0.5 minutes, 134°C, number of cycles
● Pink lid: 3 minutes, 134°C, number of cycles
● White lid: 3 minutes, 134°C, number of cycles
4.1.3.3.2 History: Service

04:00 PM

2968

12.03.2019

Pressing the "History: Service" function key displays how many cycles
to go until the next service is due by a service technician and the date of
the next service.
Servicing by a service technician is due after 3000 cycles or 2 years.
A service is due when the remaining cycle number is 0 or the stated
dated has been reached, whichever comes first.

67 46 429 D3708
D3708.201.04.02.02 10.2020 73
4 Operation Dentsply Sirona
4.1 Operation of the touchscreen Instructions for use DAC Universal

4.1.3.3.3 History: Error messages

04:00 PM

1 7777 10.07.2015

2 90 08.05.2015

3 11 10.11.2014
4 5 29.07.2014

5 9999 03.02.2014

Pressing the "History: Error messages" function key displays the last 99
error messages.
The display shows the error message number and the date.
Warning messages are not displayed in the history.
Pressing the "Navigate: Up" or the "Navigate: Down " displays the other
error messages.
4.1.3.3.4 History: Check & Clean

04:00 PM

18

12.03.2019

Pressing the "History: Check & Clean" function key displays the date of
the last Check & Clean processes (for 50 / 200 cycles).

A Check & Clean process is only considered completed and appears as


04:00 PM
an entry in the history if the last step of the Check & Clean process
appears on the display and a no-load operation (cycle without
instruments) was performed with a NitraClean cleaning tablet.

67 46 429 D3708
74 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.1 Operation of the touchscreen

4.1.3.4 Information

04:00 PM

SW PU 00.01.02-DEV.0D381A

SW UI 00.01.02-DEV.0D3C5C

SN 000001
000032

You can view information on your DAC Universal by pressing the


"Information" function key.
The following information is listed from top to bottom:
● Software version (SW PU: xx.yy.zz-REL.ABCDEF)
● Software version (SW UI: xx.yy.zz-REL.ABCDEF)
● Device serial number (SN: xxxxxx)
● Number of device cycles (xxxxxx)

4.1.3.5 Check & Clean


By pressing the "Check & Clean" function key, you can manually start
the Check & Clean process.

This sub-menu can only be called up if it is active. Otherwise, the


function key is gray and inactive.
See the Cleaning and Maintenance chapter in "Check & Clean process
[→ 108]" for more information on the Check & Clean process.

4.1.3.6 Default settings

The default settings are predetermined by the manufacturer:

Program 134°C / 273°F, 0.5 min. disinfection


(Blue lid program)
Language: GB
Oil expel: 1
Oil quantity: 1
Automatic water supply: OFF
Service software connection: OFF
Display brightness: 30%

67 46 429 D3708
D3708.201.04.02.02 10.2020 75
4 Operation Dentsply Sirona
4.2 Reprocessing in the DAC Universal Instructions for use DAC Universal

4.2 Reprocessing in the DAC Universal


4.2.1 Switching on, off and heating up the device
Switching on the unit Once the DAC Universal has been switched on at the main switch on
the rear of the unit, the unit heats up until it is ready for use.
The heating phase of the DAC Universal takes approx. 10-12 minutes.
During the heating phase, the display shows the start screen of the
DAC Universal.

IMPORTANT
A reprocessing cycle can be started during the heating phase.

IMPORTANT
Do not start the DAC Universal without a lid inserted into the holder of
the DAC Universal.

Switching off the unit The manufacturer recommends switching off the DAC Universal at the
end of work using the main switch on the rear.
Canceling a reprocessing cycle The "Cancel" function key can be used to stop a reprocessing cycle.

WARNING
Risk of contamination!
After canceling a reprocessing cycle, the load may still be
contaminated. After cancellation, allow the lid to cool sufficiently, as
otherwise the next cleaning cycle may be affected.

CAUTION
Burning hazard!
Instruments, lids, chambers and PCD test devices may be hot if the
cycle was aborted during disinfection/sterilization or cooling. Let the
instruments or PCD test device cool down before removing them.

67 46 429 D3708
76 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.2 Reprocessing in the DAC Universal

4.2.2 Filling the water tank


NOTE
Remove the water tank from the DAC Universal before filling it up.
This prevents water from spilling into the inside of the DAC Universal
housing.

IMPORTANT
The water tank can be removed at any time and without any problems
as long as no reprocessing cycle is running.
If a cycle is running, an error message is generated when the water
tank is removed.

Remove the water tank / Fill the water tank / Reinsert the water tank
ü The DAC Universal is switched on.
ü Demineralized / distilled water (water conductivity < 3.0 µS/cm) is
available.
1. Open the front flap of the DAC Universal.
2. Remove the water tank by carefully pulling the water tank up and
out of the DAC Universal by the handle.
3. Place the water tank on a flat surface.
4. Remove the lid of the water tank.
5. Fill the demineralized / distilled water (water conductivity < 3.0 µS/
cm) in the water tank (min. 1.0 liters, max. 2.3 liters)
6. Put the lid back on the water tank.
7. Insert the water tank back into the DAC Universal.

IMPORTANT
When inserting make sure that the guide hooks on the top rear side of
the water tank are inserted correctly into the housing.

67 46 429 D3708
D3708.201.04.02.02 10.2020 77
4 Operation Dentsply Sirona
4.2 Reprocessing in the DAC Universal Instructions for use DAC Universal

4.2.3 Removing the drain tank


Frequency: As soon as it is full

WARNING
Risk of contamination!
Do not start a cycle if the DAC Universal is not connected to the drain
tank or a siphon or if the drainage system installation was not
performed correctly.

CAUTION
Risk of burns!
The drain hose, filter housing, drain filter and drain tank are very hot
during the reprocessing cycle. Switch the DAC Universal off and let it
cool down (about 15 minutes) before emptying and cleaning the drain
tank or cleaning or replacing the drain filter.

ü The DAC Universal is switched off and cooled down.


ü The drain tank is cooled down.
ü Wear your personal protective gear.
1. Unscrew the lid of the drain tank.

2. Remove the condensation tube.

3. Empty the drain tank. Dispose of the waste water according to the
national guidelines and the applicable regulations for waste
disposal.

67 46 429 D3708
78 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.2 Reprocessing in the DAC Universal

4. Fill the drain tank with tap water.

5. Rinse the drain tank thoroughly with water.

6. Empty the drain tank completely.


7. Soak a soft cloth in water and wring it out well. Use the cloth to
clean the exterior of the drain tank and the condensation tube.

8. Fill the clean drain tank with water up to the lowest mark on the
label.
9. Add isopropyl (alcohol ≥ 70 %) up to the second mark on the label.
This prevents bacteria from growing in the drain container.

10. Replace the condensation tube.

11. Screw down the lid again.


12. Place the drain tank no more than 3 m from the DAC Universal and
at a lower height.
Ä The drain tank is cleaned and emptied.

67 46 429 D3708
D3708.201.04.02.02 10.2020 79
4 Operation Dentsply Sirona
4.2 Reprocessing in the DAC Universal Instructions for use DAC Universal

4.2.4 Pre- and postprocessing of the various instruments


IMPORTANT
Please follow the regulations and guidelines applicable in your
country in your practice.

WARNING
Risk of contamination!
Please follow the instrument manufacturer's instructions when
handling the instruments.
(reprocessing instructions as per DIN EN ISO 17664 / -1)

NOTE
Please lubricate the collet system of your turbines, straight and
contra-angle handpieces regularly as per the lubrication and
operating instructions from the respective instrument manufacturer.

IMPORTANT
Reprocess immediately, or at the latest, one hour after treatment.

1. Purge the water and air channels on the treatment center for 30
seconds.
2. Predisinfect the instruments directly at the treatment center.
3. Transport the instrument to the hygiene room in a suitable transport
container.

67 46 429 D3708
80 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.2 Reprocessing in the DAC Universal

4.2.5 Reprocessing with the Blue lid


The Blue lid is used for the reprocessing of straight and contra-angle
handpieces and turbines.
Tip: If there are many instruments to be reprocessed or to ensure the
instruments will be available again quickly, we recommend using a
second Blue lid.

WARNING
Risk of contamination!
After a reprocessing cycle is canceled, the instruments are not
reprocessed and cross-contamination can occur. Please reprocess
the instruments again.

4.2.5.1 Start cycle with the Blue lid

Preparation
ü All the presettings for the DAC Universal have been made.
ü The DAC Universal is switched on and ready for operation.
ü The maximum number of 6 instruments are ready for reprocessing.
ü The Blue lid is available and cooled down.
1. Place the Blue lid for configuration on the counter or put it in the lid
holder.
2. Plug each instrument onto the corresponding adapter in the Blue lid.
For more information on attaching the instruments or changing the
adapters, see the separate instructions for use for adapters.
3. Turn over the Blue lid so that the instruments are facing
downwards. The blue mark should be pointed towards you.
4. Place the Blue lid at the front of the DAC Universal holder and insert
the Blue lid straight in the holder.

CAUTION
Risk of crushing!
Make sure that your fingers do not get caught between the lid and the
holder.

NOTE
Make sure that the lid is inserted all the way into the holder so that the
instrument heads do not touch the wall of the chamber. Make sure
there are no objects in the chamber. The DAC Universal, the lid and/
or the instruments can be damaged if a lid is not inserted correctly.

CAUTION
Burning hazard!
When you take items out of the chamber, remember that the chamber
and these items can be very hot.

67 46 429 D3708
D3708.201.04.02.02 10.2020 81
4 Operation Dentsply Sirona
4.2 Reprocessing in the DAC Universal Instructions for use DAC Universal

Ä There are instruments on the Blue lid; the DAC Universal is still
open and ready for a reprocessing cycle.

Reprocessing
1. Press the "Start" function key to start the cycle.
04:00 PM

CAUTION
Risk of contamination!
Never hold your head directly over the DAC Universal when it closes.
Contaminated air may escape if it is not closed fully.

Ä The Blue lid moves down, the DAC Universal closes. The DAC
Universal detects the lid and selects the correct program.
Ä The DAC Universal automatically runs the cleaning, lubrication
and disinfection cycle.
Ä The display shows at any time which phase the DAC Universal
is in, see section "Process symbols [→ 48]“.
Ä After the selected cycle has ended, the message appears on
0.5 min 214 04:00 PM
134°C the display see image on left.
Ä The lid moves a little out of the chamber. This indicates that a
cycle has been successfully concluded and that the instruments
have been disinfected.

CAUTION
Burning hazard!
Never hold your head directly over the DAC Universal when it opens,
as steam may emerge from the unit.

WARNING
Risk of contamination!
At the end of the cycle it must be ensured that all the instruments are
still properly attached to the adapters. If instruments have fallen into
the chamber during the cycle or if an instrument is no longer correctly
attached at the end of the cycle, then cleaning, lubrication and
disinfection have not been successful. In this case, reprocessing of
the instruments must be repeated.

2. Press the "Confirm" function key.


Ä The Blue lid will open completely.

CAUTION
Risk of crushing!
Make sure that you do not trap your fingers between the opening lid
and an object, e.g., a wall cupboard which is located above.

67 46 429 D3708
82 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.2 Reprocessing in the DAC Universal

Check
ü The Blue lid is completely open.

CAUTION
Burning hazard!
Even when the lid of the DAC Universal is partly or completely open,
some instruments may still be hot. Allow the instruments that become
very hot during the process to cool down.

CAUTION
Risk of burns!
The metal part of the lid is very hot. Always touch only the plastic part
of the lid.

1. Remove the Blue lid from the holder on the DAC Universal at the
end of the cycle to allow the instruments to cool down faster.
2. Turn the Blue lid so that the instruments are pointing up.
3. Place the Blue lid on the counter or put it in the lid holder.

WARNING
Risk of contamination!
Please follow the instrument manufacturer's instructions when
handling the instruments.
(reprocessing instructions as per DIN EN ISO 17664 / -1)

4. Check the instruments visually to make sure they are dry.

WARNING
Risk of contamination!
Check the instruments visually to make sure they have been cleaned.

5. Read the cycle number above next to the completed cycle symbol.
6. Check your corresponding parameters on your documentation
(printer, software, etc.) to ensure that the cycle was performed
correctly.
Ä If everything is correct, reprocessing was successful.
7. Document the cycle according to the work instructions applicable in
your practice.
8. Remove all the instruments from the Blue lid when the lid and the
instruments have cooled down sufficiently. See the separate
instructions for use for adapters for detailed instructions on how to
remove the instruments from the individual adapters.

WARNING
Risk of contamination!
Do not touch the housing of the DAC Universal with the reprocessed
instruments when removing them.

Ä The instruments may only be used for semi-critical B applications.

67 46 429 D3708
D3708.201.04.02.02 10.2020 83
4 Operation Dentsply Sirona
4.2 Reprocessing in the DAC Universal Instructions for use DAC Universal

4.2.6 Reprocessing with the Green lid


The Green lid is used to clean and thermally disinfect the Green lid
instruments.
Tip: If there are many instruments to be reprocessed or to ensure the
instruments will be available again quickly, we recommend using a
second Green lid.

WARNING
Risk of contamination!
After a reprocessing cycle is canceled, the instruments are not
reprocessed and cross-contamination can occur. Please reprocess
the instruments again.

4.2.6.1 Start cycle with the Green lid

Preparation
ü All the presettings for the DAC Universal have been made.
ü The DAC Universal is switched on and ready for operation.
ü The maximum number of 6 instruments are ready for reprocessing
(for loading, see "Functions [→ 21]").
ü The Green lid is available and cooled down.
1. Place the Green lid for configuration on the counter or put it in the
lid holder.
2. Attach each instrument or screw on each ultrasonic tip incl. the
torque wrench on the corresponding adapter in the Green lid.
Further information on attaching/screwing on the instruments and
replacing the adapters can be found in the separate adapter
instructions for use.
3. Turn over the Green lid so that the instruments are facing
downwards. The green mark should be pointed towards you.
4. Place the Green lid at the front of the DAC Universal holder and
insert the Green lid straight in the holder.

CAUTION
Risk of crushing!
Make sure that your fingers do not get caught between the lid and the
holder.

NOTE
Make sure that the lid is inserted all the way into the holder so that the
instrument heads do not touch the wall of the chamber. Make sure
there are no objects in the chamber. The DAC Universal, the lid and/
or the instruments can be damaged if a lid is not inserted correctly.

67 46 429 D3708
84 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.2 Reprocessing in the DAC Universal

CAUTION
Burning hazard!
When you take items out of the chamber, remember that the chamber
and these items can be very hot.

Ä There are instruments on the Green lid; the DAC Universal is still
open and ready for a cycle.

Reprocessing
1. Press the "Start“ function key to start the cycle.
04:00 PM

CAUTION
Risk of contamination!
Never hold your head directly over the DAC Universal when it closes.
Contaminated air may escape if it is not closed fully.

Ä The Green lid moves down, the DAC Universal closes. The
DAC Universal detects the lid and selects the correct program.
Ä The DAC Universal automatically runs the cleaning and
disinfection cycle.
Ä The display shows at any time which phase the DAC Universal
is in, see section "Process symbols [→ 48]“.
Ä After the selected cycle has ended, the message appears on
0.5 min 214 04:00 PM
134°C the display see image on left.
Ä The lid moves a little out of the chamber. This indicates that a
cycle has been completed and the instruments have been
reprocessed.

CAUTION
Burning hazard!
Never hold your head directly over the DAC Universal when it opens,
as steam may emerge from the unit.

WARNING
Risk of contamination!
After completion of the cycle, make sure that all instruments are still
properly attached to the adapters. If instruments have fallen into the
chamber during the cycle or if an instrument is no longer properly
attached after the end of a cycle, cleaning and disinfection have
failed. In this case, reprocessing of the instruments must be repeated.

2. Press the "Confirm" function key.


Ä The Green lid will open completely.

CAUTION
Risk of crushing!
Make sure that you do not trap your fingers between the opening lid
and an object, e.g., a wall cupboard which is located above.

67 46 429 D3708
D3708.201.04.02.02 10.2020 85
4 Operation Dentsply Sirona
4.2 Reprocessing in the DAC Universal Instructions for use DAC Universal

Check
ü The Green lid is completely open.

CAUTION
Burning hazard!
Even when the lid of the DAC Universal is partly or completely open,
some instruments may still be hot. Allow the instruments that become
very hot during the process to cool down.

CAUTION
Risk of burns!
The metal part of the lid is very hot. Always touch only the plastic part
of the lid.

1. Remove the Green lid from the holder on the DAC Universal at the
end of the cycle to allow the instruments to cool down faster.
2. Turn the Green lid so that the instruments are pointing up.
3. Place the Green lid on the counter or put it in the lid holder.

WARNING
Risk of contamination!
Please follow the instrument manufacturer's instructions when
handling the instruments.
(reprocessing instructions as per DIN EN ISO 17664 / -1)

4. Check the instruments visually to make sure they are dry.

WARNING
Risk of contamination!
Check the instruments visually to make sure they have been cleaned.

5. Read the cycle number above next to the completed cycle symbol.
6. Check your corresponding parameters on your documentation
(printer, software, etc.) to ensure that the cycle was performed
correctly.
Ä If everything is correct, reprocessing was successful.
7. Document the cycle according to the work instructions applicable in
your practice.
8. Remove all the instruments from the Green lid when the lid and the
instruments have cooled down sufficiently. See the separate
instructions for use for adapters for detailed instructions on
removing the individual adapters.

WARNING
Risk of contamination!
Do not touch the housing of the DAC Universal with the reprocessed
instruments when removing them.

Ä The instruments may only be used for semi-critical B applications.

67 46 429 D3708
86 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.2 Reprocessing in the DAC Universal

4.2.7 Reprocessing with the Pink lid


The Pink lid is used for the reprocessing of straight and contra-angle
handpieces and turbines.
Tip: If there are many instruments to be reprocessed or to ensure the
instruments will be available again quickly, we recommend using a
second Pink lid.

WARNING
Risk of contamination!
After a reprocessing cycle is canceled, the instruments are not
reprocessed and cross-contamination can occur. Please reprocess
the instruments again.

4.2.7.1 Start cycle with the Pink lid

Preparation
ü All the presettings for the DAC Universal have been made.
ü The DAC Universal is switched on and ready for operation.
ü The maximum number of 6 instruments are ready for reprocessing.
ü A Type 5 chemical indicator is available.
ü The Pink lid with indicator holder is available and has cooled down.
1. Place the Pink lid for configuration on the counter or put it in the lid
holder.
2. Clamp the chemical indicator as desired in the indicator holder. You
can find more information on the chemical indicator in the section
"Batch control with the chemical indicator [→ 94]".
3. Plug each instrument onto the corresponding adapter in the Pink lid.
For more information on attaching the instruments or changing the
adapters, see the separate instructions for use for adapters.
4. Turn over the Pink lid so that the instruments are facing downwards.
The pink marking is pointed towards you.
5. Place the Pink lid at the front of the DAC Universal holder and insert
the Pink lid straight in the holder.

CAUTION
Risk of crushing!
Make sure that your fingers do not get caught between the lid and the
holder.

NOTE
Make sure that the lid is inserted all the way into the holder so that the
instrument heads do not touch the wall of the chamber. Make sure
there are no objects in the chamber. The DAC Universal, the lid and/
or the instruments can be damaged if a lid is not inserted correctly.

67 46 429 D3708
D3708.201.04.02.02 10.2020 87
4 Operation Dentsply Sirona
4.2 Reprocessing in the DAC Universal Instructions for use DAC Universal

CAUTION
Burning hazard!
When you take items out of the chamber, remember that the chamber
and these items can be very hot.

Ä There are instruments on the Pink lid; the DAC Universal is still
open and ready for a reprocessing cycle.

Reprocessing
1. Press the "Start" function key to start the cycle.
04:00 PM

CAUTION
Risk of contamination!
Never hold your head directly over the DAC Universal when it closes.
Contaminated air may escape if it is not closed fully.

Ä The Pink lid moves down, the DAC Universal closes. The DAC
Universal detects the lid and selects the correct program.
Ä The DAC Universal automatically runs the cleaning, lubrication
and sterilization cycle.
Ä Thanks to the display you can see which phase the DAC
Universal is in at any time, see "Process symbols [→ 48]“.
Ä After the selected cycle has ended, the message appears on
3 min 214 04:00 PM
134°C the display see image on left.
Ä The lid moves a little out of the chamber. This indicates that a
cycle has been successfully concluded and that the instruments
have been sterilized.

CAUTION
Burning hazard!
Never hold your head directly over the DAC Universal when it opens,
as steam may emerge from the unit.

WARNING
Risk of contamination!
At the end of the cycle it must be ensured that all the instruments are
still properly attached to the adapters. If instruments have fallen into
the chamber during the cycle or if an instrument is no longer correctly
attached at the end of the cycle, then cleaning, lubrication and
sterilization have not been successful. In this case, reprocessing of
the instruments must be repeated.

2. Press the "Confirm" function key.


Ä The Pink lid will open completely.

CAUTION
Risk of crushing!
Make sure that you do not trap your fingers between the opening lid
and an object, e.g., a wall cupboard which is located above.

67 46 429 D3708
88 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.2 Reprocessing in the DAC Universal

Check
ü The Pink lid is completely open.

CAUTION
Burning hazard!
Even when the lid of the DAC Universal is partly or completely open,
some instruments may still be hot. Allow the instruments that become
very hot during the process to cool down.

CAUTION
Risk of burns!
The metal part of the lid is very hot. Always touch only the plastic part
of the lid.

1. Remove the Pink lid from the holder on the DAC Universal at the
end of the cycle to allow the instruments to cool down faster.
2. Turn the Pink lid so that the instruments are pointing up.
3. Place the Pink lid on the counter or put it in the lid holder.

WARNING
Risk of contamination!
Please follow the instrument manufacturer's instructions when
handling the instruments.
(reprocessing instructions as per DIN EN ISO 17664 / -1)

4. Check the instruments visually to make sure they are dry.

WARNING
Risk of contamination!
Check the instruments visually to make sure they have been cleaned.

5. Read the cycle number above next to the completed cycle symbol.
6. Check your corresponding parameters on your documentation
(printer, software, etc.) to ensure that the cycle was performed
correctly.
7. Check that the chemical indicator has changed correctly.
Ä If everything is correct, reprocessing was successful.
8. Document the cycle according to the work instructions applicable in
your practice.
9. Remove all the instruments from the Pink lid when the lid and the
instruments have cooled down sufficiently. See the separate
instructions for use for adapters for detailed instructions on how to
remove the instruments from the individual adapters.

WARNING
Risk of contamination!
Do not touch the housing of the DAC Universal with the reprocessed
instruments when removing them.

Ä The instruments may only be used for semi-critical B applications.

67 46 429 D3708
D3708.201.04.02.02 10.2020 89
4 Operation Dentsply Sirona
4.2 Reprocessing in the DAC Universal Instructions for use DAC Universal

4.2.8 Reprocessing with the White lid


The White lid is used to clean and sterilize the White lid instruments.
Tip: If there are many instruments to be reprocessed or to ensure the
instruments will be available again quickly, we recommend using a
second White lid.

WARNING
Risk of contamination!
After a reprocessing cycle is canceled, the instruments are not
reprocessed and cross-contamination can occur. Please reprocess
the instruments again.

4.2.8.1 Start cycle with the White lid

Preparation
ü All the presettings for the DAC Universal have been made.
ü The DAC Universal is switched on and ready for operation.
ü The maximum number of 6 instruments are ready for reprocessing
(for loading, see "Functions [→ 21]").
ü A Type 5 chemical indicator is available.
ü The White lid is available and cooled down.
1. Place the White lid for configuration on the counter or put it in the lid
holder.
2. Clamp the chemical indicator as desired in the indicator holder. You
can find more information on the chemical indicator in the section
"Batch control with the chemical indicator [→ 94]".
3. Attach each instrument or screw on each ultrasonic scaler tip incl.
the torque wrench on the corresponding adapter in the White lid.
Further information on attaching/screwing on the instruments and
replacing the adapters can be found in the separate adapter
instructions for use.
4. Turn over the White lid so that the instruments are facing
downwards. The white marking is pointed towards you.
5. Place the White lid at the front of the DAC Universal holder and
insert the White lid straight in the holder.

CAUTION
Risk of crushing!
Make sure that your fingers do not get caught between the lid and the
holder.

67 46 429 D3708
90 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.2 Reprocessing in the DAC Universal

NOTE
Make sure that the lid is inserted all the way into the holder so that the
instrument heads do not touch the wall of the chamber. Make sure
there are no objects in the chamber. The DAC Universal, the lid and/
or the instruments can be damaged if a lid is not inserted correctly.

CAUTION
Burning hazard!
When you take items out of the chamber, remember that the chamber
and these items can be very hot.

Ä There are instruments on the White lid; the DAC Universal is still
open and ready for a cycle.

Reprocessing
1. Press the "Start“ function key to start the cycle.
04:00 PM

CAUTION
Risk of contamination!
Never hold your head directly over the DAC Universal when it closes.
Contaminated air may escape if it is not closed fully.

Ä The White lid moves down, the DAC Universal closes. The DAC
Universal detects the lid and selects the correct program.
Ä The DAC Universal automatically runs the cleaning and
sterilization cycle.
Ä Thanks to the display you can see which phase the DAC
Universal is in at any time, see "Process symbols [→ 48]“.
Ä After the selected cycle has ended, the message appears on
3 min 214 04:00 PM
134°C the display see image on left.
Ä The lid moves a little out of the chamber. This indicates that a
cycle has been completed and the instruments have been
reprocessed.

CAUTION
Burning hazard!
Never hold your head directly over the DAC Universal when it opens,
as steam may emerge from the unit.

WARNING
Risk of contamination!
At the end of the cycle it must be ensured that all the instruments are
still properly attached to the adapters. If instruments have fallen into
the chamber during the cycle or if an instrument is no longer properly
attached after the end of a cycle, cleaning and sterilization have
failed. In this case, reprocessing of the instruments must be repeated.

2. Press the "Confirm" function key.


Ä The White lid will open completely.

67 46 429 D3708
D3708.201.04.02.02 10.2020 91
4 Operation Dentsply Sirona
4.2 Reprocessing in the DAC Universal Instructions for use DAC Universal

CAUTION
Risk of crushing!
Make sure that you do not trap your fingers between the opening lid
and an object, e.g., a wall cupboard which is located above.

Check
ü The White lid is completely open.

CAUTION
Burning hazard!
Even when the lid of the DAC Universal is partly or completely open,
some instruments may still be hot. Allow the instruments that become
very hot during the process to cool down.

CAUTION
Risk of burns!
The metal part of the lid is very hot. Always touch only the plastic part
of the lid.

1. Remove the White lid from the holder on the DAC Universal at the
end of the cycle to allow the instruments to cool down faster.
2. Turn the White lid so that the instruments are pointing up.
3. Place the White lid on the counter or put it in the lid holder.

WARNING
Risk of contamination!
Please follow the instrument manufacturer's instructions when
handling the instruments.
(reprocessing instructions as per DIN EN ISO 17664 / -1)

4. Check the instruments visually to make sure they are dry.

WARNING
Risk of contamination!
Check the instruments visually to make sure they have been cleaned.

5. Read the cycle number above next to the completed cycle symbol.
6. Check your corresponding parameters on your documentation
(printer, software, etc.) to ensure that the cycle was performed
correctly.
7. Check that the chemical indicator has changed correctly.
Ä If everything is correct, reprocessing was successful.
8. Document the cycle according to the work instructions applicable in
your practice.
9. Remove all the instruments from the White lid when the lid and the
instruments have cooled down sufficiently. See the separate
instructions for use for adapters for detailed instructions on
removing the individual adapters.

67 46 429 D3708
92 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.2 Reprocessing in the DAC Universal

WARNING
Risk of contamination!
Do not touch the housing of the DAC Universal with the reprocessed
instruments when removing them.

Ä The instruments may only be used for semi-critical B applications.

67 46 429 D3708
D3708.201.04.02.02 10.2020 93
4 Operation Dentsply Sirona
4.2 Reprocessing in the DAC Universal Instructions for use DAC Universal

4.2.9 Batch control with the chemical indicator


IMPORTANT
Please follow the regulations and guidelines applicable in your
country in your practice.

Batch control: The manufacturer recommends using a Type 5 chemical indicator


(134°C/3 min. steam) in the indicator holder for all cycles (included in
scope of delivery) in order to check for the presence of saturated steam.
For details on the reprocessing procedure with the Pink / White lids, see
"Reprocessing with the Pink lid [→ 87]" or "Reprocessing with the White
lid [→ 90]".

IMPORTANT
Follow the manufacturer's instructions for use when using a chemical
indicator.

Procedure
ü The DAC Universal is switched on and ready for a cycle.
ü An unloaded Pink lid or White lid with indicator holder is available.
ü A Type 5 chemical indicator is available.
1. Place the Pink lid or White lid on the counter.

2. Separate the chemical indicator from the base paper.

67 46 429 D3708
94 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.2 Reprocessing in the DAC Universal

3. Clamp the chemical indicator into the indicator holder as you wish.

4. Plug each instrument onto the corresponding adapter in the Pink lid
or White lid. For more information on attaching the instruments or
changing the adapters, see the separate instructions for use for
adapters.
5. Turn over the Pink lid or White lid so that the instruments are facing
downwards. The colored marking (Pink or White) is pointed towards
you.
6. Place the Pink lid or White lid at the front of the DAC Universal
holder and insert the Pink lid or White lid straight in the holder.

CAUTION
Risk of crushing!
Make sure that your fingers do not get caught between the lid and the
holder.

7. Start a reprocessing cycle, see "Reprocessing with the Pink lid


[→ 87]" or "Reprocessing with the White lid [→ 90]".

Ä After the selected cycle has ended, the message appears on


3 min 214 04:00 PM
134°C the display, see image on left (here for the Pink lid).
8. Remove the lid and place it on the counter.

CAUTION
Burning hazard!
When you take items out of the chamber, remember that the chamber
and these items can be very hot.

9. Remove the chemical indicator from the indicator holder.


10. Check the process result. All 4 test points must have changed color
from yellow to black.
Ä The chemical indicator used can be adhered in place for
documentation purposes by using the adhesive strip.

67 46 429 D3708
D3708.201.04.02.02 10.2020 95
4 Operation Dentsply Sirona
4.2 Reprocessing in the DAC Universal Instructions for use DAC Universal

WARNING
Risk of contamination!
If the chemical indicator indicates that the test was not passed, do not
use the instruments that were reprocessed after the time of the last
successful check.
Do not continue to use the DAC Universal.
Please contact your authorized dealer.

67 46 429 D3708
96 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.2 Reprocessing in the DAC Universal

4.2.10 Periodic testing with the PCD test device


The PCD (process challenge device) test device is used in combination
with a Type 5 chemical indicator or Type 2 chemical indicator. The test
device serves to simulate a steam penetration test and confirm proper
sterilization.
The PCD test device is not included in the scope of delivery of the DAC
Universal.
The PCD test program can be performed with either a Pink lid or a
White lid.

CAUTION
Risk of contamination!
Only a chemical indicator from Dentsply Sirona may be used for the
steam penetration test as another chemical indicator leads to a false
result.

WARNING
Risk of contamination!
The PCD test device may not be used for a normal reprocessing
cycle as reprocessing could fail.

Procedure
ü The DAC Universal is switched on and ready for a cycle.
ü An unloaded Pink lid or White lid without an indicator holder is
available.

67 46 429 D3708
D3708.201.04.02.02 10.2020 97
4 Operation Dentsply Sirona
4.2 Reprocessing in the DAC Universal Instructions for use DAC Universal

ü The PCD test device is available.


ü A chemical indicator from Dentsply Sirona is available.
1. Place the Pink lid or White lid on the counter.

2. Unscrew the PCD test device.

3. Separate the chemical indicator from the base paper.

4. Insert the chemical indicator into the slot on the bottom of the PCD
test device.
5. Make sure that the chemical indicator is not trapped.
6. Screw the top part of the PCD test device clockwise onto the bottom
part of the PCD test device until you note a slight degree of
resistance (in the form of the O-ring).

IMPORTANT
Make sure the PCD test device is completely closed.
A proper check of the process is not ensured if it is partly open.

67 46 429 D3708
98 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.2 Reprocessing in the DAC Universal

7. Screw the PCD test device into the central thread of the Pink lid or
White lid.
8. Turn over the Pink lid or White lid so that the instruments are facing
downwards. The colored marking (Pink or White) is pointed towards
you.
9. Place the Pink lid or White lid at the front of the DAC Universal
holder and insert the Pink lid or White lid straight in the holder.

CAUTION
Risk of crushing!
Make sure that your fingers do not get caught between the lid and the
holder.

10. Start a PCD test program using the menu "Settings/PCD test
program".
Ä After the PCD test program has ended, the message appears
3 min 200 04:00 PM
134°C on the display see image on left.
Ä The LED flashes.
11. Confirm the end of the process on the display.
Ä The LED goes out.
12. Remove the lid and place it on the counter.

CAUTION
Burning hazard!
When you take items out of the chamber, remember that the chamber
and these items can be very hot.

13. Wait until the PCD test device has cooled down.
14. Unscrew the top part of the PCD test device by turning
counterclockwise.

CAUTION
Risk of contamination!
Ensure that there is no residual moisture or oil residue in the PCD test
device. Residual moisture or oil residue can lead to a false positive
result.
When using a reprocessing cycle without a drying cycle, the PCD test
device can contain water condensation. In this case, open the PCD
test device whilst it is still warm, blow through air and leave it open to
dry.

15. Remove the chemical indicator.


16. Check the process result. All 4 test points must have changed color
from yellow to black.
Ä The chemical indicator used can be adhered in place for
documentation purposes by using the adhesive strip.

67 46 429 D3708
D3708.201.04.02.02 10.2020 99
4 Operation Dentsply Sirona
4.2 Reprocessing in the DAC Universal Instructions for use DAC Universal

WARNING
Risk of contamination!
If the chemical indicator indicates that the test was not passed, do not
use the instruments that were reprocessed after the time of the last
successful check.
Do not continue to use the DAC Universal.
Please contact your authorized dealer.

4.2.11 Process documentation


NOTE
The data interfaces may only be connected to devices with fused
circuits.

The DAC Universal can be connected to transmit process checks as


follows:
● RS232 interface
● LAN interface

4.2.11.1 RS-232 interface


A printer or software can be connected to the RS-232 interface for batch
documentation.
Please contact your authorized dealer to procure documentation
software.
The printer is optionally available and prints out the reports of the
performed reprocessing protocols directly. The performed reprocessing
processes can be checked immediately using this print-out.

The following languages are available for the printer report:


● German
● English
● French
● Italian
● Spanish
● Danish
● Swedish
● Norwegian
These can be selected when installing the printer.

67 46 429 D3708
100 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.2 Reprocessing in the DAC Universal

4.2.11.1.1 Connecting the printer / setting the printer protocol language


To connect the printer to the DAC Universal proceed as follows:
ü The DAC Universal is switched on.
1. Connect the device to the serial RS-232 interface of the DAC
Universal using a serial cable.
Ä The printer is connected.
2. Press the "Menu" function key from the homescreen.

3. Press the "Settings" function key.

4. Press the "Printer" function key.

5. Press the "Languages" function key.

Ä An overview with the languages available for selection appears.


04:00 PM
6. Press the function key of the language you wish to select.
Ä The selected language is marked.
DE GB FR IT

ES DK SE NO

7. Leave the menu by pressing the "Back" function key several times
or by pressing the "Homescreen" function key.
Ä The printer is connected and the printer protocol language is
selected.

67 46 429 D3708
D3708.201.04.02.02 10.2020 101
4 Operation Dentsply Sirona
4.2 Reprocessing in the DAC Universal Instructions for use DAC Universal

4.2.11.1.2 Output information of the printer


The process data are sent to the printer immediately during the process. All
the data of the previous cycle are saved internally in the DAC Universal.

Output information of the Blue lid:


The following information is output via the printer:
● Date (YYYY-MM-DD)
● Time (24h) (HH:MM:SS)
● Serial number of the DAC Universal
● Process Software
● C&C 50 status (Due / OK)
● C&C 200 status (Due / OK)
● Service status
● Cycle
● Lid type
● Lid serial
● Program

For each phase the time [Time:] and the result [Res:] "OK" or "FAIL" are
printed. The phases are:
● Lid detected
● External flush
● Leakage test
● Internal cleaning
● Lubrication
● External cleaning
● External rinse
● Heating
● Back-flush
● Start holding time

● Maximum: °C / bar
● Minimum: °C / bar
● Duration (mm-ss)

The rest of the phases and the result “OK” or “FAIL” is printed. The phases
are:
● End holding time
● Cooling
● Process end
● Confirmation
● Cycle end

The rest of the phases and the result “OK” or “FAIL” is printed. The phases are:
● Disinfection
● Error code
If "FAIL" is displayed for a phase, then the reprocessing cycle was
unsuccessful and must be repeated.

67 46 429 D3708
102 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.2 Reprocessing in the DAC Universal

Output information of the Green lid:


The following information is output via the printer:
● Date (YYYY-MM-DD)
● Time (24h) (HH:MM:SS)
● Serial number of the DAC Universal
● Process Software
● C&C 50 status (Due / OK)
● C&C 200 status (Due / OK)
● Service status
● Cycle
● Lid type
● Lid serial
● Program

For each phase the time [Time:] and the result [Res:] "OK" or "FAIL" are
printed. The phases are:
● Lid detected
● External flush
● Leakage test
● Internal cleaning
● External cleaning
● External rinse
● Heating
● Back-flush
● Start holding time

● Maximum: °C / bar
● Minimum: °C / bar
● Duration (mm-ss)

The rest of the phases and the result “OK” or “FAIL” is printed. The
phases are:
● End holding time
● Cooling
● Process end
● Confirmation
● Cycle end

The rest of the phases and the result “OK” or “FAIL” is printed. The
phases are:
● Disinfection
● Error code
If "FAIL" is displayed for a phase, then the reprocessing cycle was
unsuccessful and must be repeated.

67 46 429 D3708
D3708.201.04.02.02 10.2020 103
4 Operation Dentsply Sirona
4.2 Reprocessing in the DAC Universal Instructions for use DAC Universal

Output information of the Pink lid:


The following information is output via the printer:
● Date (YYYY-MM-DD)
● Time (24h) (HH:MM:SS)
DAC Universal (Touch) ● Serial number of the DAC Universal
Date(YYYY-MM-DD) 2020-01-23
● Process Software
Time (24h) 10:16:21
Device serial 003000 ● C&C 50 status (Due / OK)
Process Software 02.00.01 ● C&C 200 status (Due / OK)
C&C 50 status Due ● Service status
C&C 200 status Due
Service status OK ● Cycle
Cycle 000086 ● Lid type
Lid type Pink ● Lid serial
Lid serial 999999
● Program
Program 134 °C/3 min
--------------------------------
Phase: Time: Res: For each phase the time [Time:] and the result [Res:] "OK" or "FAIL" are
Lid detected 10:16:51 OK printed. The phases are:
External flush 10:17:25 OK ● Lid detected
Leakage test 10:17:55 OK
● External flush
Internal wash 10:19:51 OK
Lubrication 10:20:23 OK ● Leakage test
External wash 10:22:25 OK ● Internal cleaning
External rinse 10:22:55 OK ● Lubrication
Heating 10:25:32 OK
Backflush 10:25:46 OK ● External cleaning
Air removal 10:29:07 OK ● External rinse
Start holding time 10:29:22 OK ● Heating
--------------------------------
● Back-flush
Time: °C: bar:
10:29:22 135.0 3.20 ● Start holding time
10:29:52 135.0 3.19
10:30:22 135.0 3.18
10:30:52 135.0 3.19 ● Maximum: °C / bar
10:31:22 135.1 3.19 ● Minimum: °C / bar
10:31:52 135.0 3.18 ● Duration (mm-ss)
10:32:22 135.1 3.19
Maximum 135.2 3.24
The rest of the phases and the result “OK” or “FAIL” is printed. The
Minimum 134.9 3.17
Duration (mm-ss) 03-03 phases are:
-------------------------------- ● End holding time
End holding time 10:32:25 OK ● Cooling
Cooling 10:37:49 OK
● Process end
Process end 10:37:59 OK
Confirmation 10:55:15 OK ● Confirmation
Cycle end 10:55:34 OK ● Cycle end
--------------------------------
Sterilization OK The rest of the phases and the result “OK” or “FAIL” is printed. The
phases are:
● Sterilization
● Error code
If "FAIL" is displayed for a phase, then the reprocessing cycle was
unsuccessful and must be repeated.

67 46 429 D3708
104 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.2 Reprocessing in the DAC Universal

Output information of the White lid:


The following information is output via the printer:
● Date (YYYY-MM-DD)
● Time (24h) (HH:MM:SS)
DAC Universal (Touch) ● Serial number of the DAC Universal
Date(YYYY-MM-DD) 2020-01-23
● Process Software
Time (24h) 10:57:08
Device serial 003000 ● C&C 50 status (Due / OK)
Process Software 02.00.01 ● C&C 200 status (Due / OK)
C&C 50 status Due
● Service status
C&C 200 status Due
Service status OK ● Cycle
Cycle 000087 ● Lid type
Lid type White ● Lid serial
Lid serial 999999
Program 134 °C/3 min ● Program
--------------------------------
Phase: Time: Res: For each phase the time [Time:] and the result [Res:] "OK" or "FAIL" are
Lid detected 10:57:38 OK printed. The phases are:
External flush 10:58:12 OK ● Lid detected
Leakage test 10:58:38 OK
● External flush
Internal wash 10:59:22 OK
External wash 11:01:28 OK ● Leakage test
External rinse 11:01:58 OK ● Internal cleaning
Heating 11:04:18 OK ● External cleaning
Backflush 11:04:30 OK
Air removal 11:07:32 OK ● External rinse
Start holding time 11:07:47 OK ● Heating
-------------------------------- ● Back-flush
Time: °C: bar:
11:07:47 135.0 3.20 ● Start holding time
11:08:17 135.0 3.21
11:08:47 135.0 3.21
● Maximum: °C / bar
11:09:17 135.1 3.20
11:09:47 135.0 3.18 ● Minimum: °C / bar
11:10:16 135.0 3.18 ● Duration (mm-ss)
11:10:46 135.1 3.19
Maximum 135.2 3.24 The rest of the phases and the result “OK” or “FAIL” is printed. The
Minimum 135.0 3.17 phases are:
Duration (mm-ss) 03-02
-------------------------------- ● End holding time
End holding time 11:10:49 OK ● Cooling
Cooling 11:16:15 OK ● Process end
Process end 11:16:26 OK
Confirmation 11:17:18 OK ● Confirmation
Cycle end 11:17:37 OK ● Cycle end
--------------------------------
Sterilization OK The rest of the phases and the result “OK” or “FAIL” is printed. The
phases are:
● Sterilization
● Error code
If "FAIL" is displayed for a phase, then the reprocessing cycle was
unsuccessful and must be repeated.

67 46 429 D3708
D3708.201.04.02.02 10.2020 105
4 Operation Dentsply Sirona
4.2 Reprocessing in the DAC Universal Instructions for use DAC Universal

Output information of the PCD test program:


The following information is output via the printer:
● Date (YYYY-MM-DD)
● Time (24h) (HH:MM:SS)
DAC Universal (Touch) ● Serial number of the DAC Universal
PCD test program
-------------------------------- ● Process Software
Date(YYYY-MM-DD) 2020-01-23 ● C&C 50 status (Due / OK)
Time (24h) 11:18:39 ● C&C 200 status (Due / OK)
Device serial 003000
● Service status
Process Software 02.00.01
C&C 50 status Due ● Cycle
C&C 200 status Due ● Lid type
Service status OK ● Lid serial
Cycle 000088
Lid type White ● Program
Lid serial 999999
Temp./holding time 134 °C/3 min
-------------------------------- For each phase the time [Time:] and the result [Res:] "OK" or "FAIL" are
Phase: Time: Res: printed. The phases are:
Lid detected 11:19:10 OK ● Lid detected
Leakage test 11:19:38 OK
● Leakage test
Heating 11:21:13 OK
Backflush 11:21:24 OK ● Heating
Air removal 11:24:17 OK ● Back-flush
Start holding time 11:24:32 OK ● Air removal
--------------------------------
Time: °C: bar: ● Start holding time
11:24:32 135.0 3.21
11:25:02 135.0 3.22
● Maximum: °C / bar
11:25:32 135.1 3.20
11:26:02 135.0 3.23 ● Minimum: °C / bar
11:26:32 135.1 3.21 ● Duration (mm-ss)
11:27:02 135.1 3.18
11:27:32 135.1 3.21
Maximum 135.2 3.24 The rest of the phases and the result “OK” or “FAIL” is printed. The
Minimum 135.0 3.17 phases are:
Duration (mm-ss) 03-03
-------------------------------- ● End holding time
End holding time 11:27:35 OK ● Cooling
Cooling 11:33:00 OK ● Process end
Process end 11:33:11 OK
● Confirmation
Confirmation 11:33:11 OK
Cycle end 11:33:30 OK ● Cycle end
--------------------------------
PCD test program COMPLETED
!TEST PROGRAM The result is printed.
● PCD test cycle, ONLY TEST CYCLE -> COMPLETE
An unsuccessful phase is marked with "FAIL".
If "FAIL" is displayed for a phase, then the PCD test program was
unsuccessful and must be repeated.
In such cases, the printed result is PCD test program NOT
COMPLETED.

67 46 429 D3708
106 D3708.201.04.02.02 10.2020
Dentsply Sirona 4 Operation
Instructions for use DAC Universal 4.2 Reprocessing in the DAC Universal

4.2.11.2 LAN interface


Documentation software can be connected to the LAN interface for
batch documentation, a router may have to be interconnected.
Please contact your authorized dealer to procure documentation
software.

4.2.12 Activities during longer downtimes


If the DAC Universal is not used for more than 24 hours, perform a no-
load cycle with a Blue lid or Pink lid without instruments.

67 46 429 D3708
D3708.201.04.02.02 10.2020 107
5 Cleaning, maintenance and validation Dentsply Sirona
5.1 Check & Clean process Instructions for use DAC Universal

5 Cleaning, maintenance and validation


IMPORTANT
Typical service life:
If all maintenance is performed correctly as per section 5.1, a service
life of 4 years can typically be expected for the DAC Universal. (No
warranty claims can be derived on the basis of this.)

5.1 Check & Clean process


Frequency: Weekly or every 50 cycles (whichever comes first).
After every 50 cycles, the DAC Universal will display the message
04:00 PM
shown at the left. This message reminds the user that the weekly Check
& Clean cleaning/maintenance is due.
The following overview table below shows how often cleaning/
maintenance should be carried out. The individual Check & Clean
process steps are explained in more detail in the following sub-sections.
The frequencies indicated are based on approx. 10 complete cycles per
day. If the DAC Universal is used more frequently, you should adjust the
Check & Clean intervals accordingly.

WARNING
Risk of contamination!
Please ensure that the Check & Clean process is conducted regularly
at the intervals required in these instructions for use.

67 46 429 D3708
108 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.1 Check & Clean process

Overview of tasks to be carried out

Frequency
Task Weekly or Every 4 Every 3rd Every 2
every 50 weeks or month years or ev-
cycles* every 200 ery 3000
cycles cycles
1. Start Check & Clean ✓
2. Put on gloves ✓
3. Clean drain filter ✓
4. Replace drain filter ✓
5. Check O-rings on drain filter and replace if necessary ✓
6. Insert the Check & Clean lid ✓
7. Check the safety ring ✓
8. Check the Check & Clean lid ✓
9. Check the water supply to the chamber ✓
10. Check the oil supply via the Check & Clean lid ✓
11. Check the water supply via the Check & Clean lid ✓
12. Open chamber / remove Check & Clean lid ✓
13. Check the permeability of the media paths of all lids / ✓
remove the water tank
14. Disassemble and clean the water tank ✓
15. Clean the drain tank ✓
16. Check the assembly of the adapters ✓
17. Check O-rings on the adapters ✓
18. Change the O-rings on the adapters ✓
19. Clean the housing of the unit, the lid and the cham- ✓
ber
20. Check the O-ring on the water tank ✓
21. Assemble, fill and insert the water tank ✓
22. Preparations (after the Check & Clean) ✓
23. Service by qualified technician ✓

* according to the Check & Clean plan

WARNING
Risk of contamination!
The sterile filter insert needs to be replaced every two years or if the
pressure falls below 0.1 bar.

WARNING
Risk of contamination!
The software-assisted Check & Clean process assists the actual
process, the operator is still responsible for ensuring correct
implementation and assessment.

67 46 429 D3708
D3708.201.04.02.02 10.2020 109
5 Cleaning, maintenance and validation Dentsply Sirona
5.1 Check & Clean process Instructions for use DAC Universal

The software will take you through the individual Check & Clean steps.
The menu guidance in the Check & Clean process is as follows:

Navigate to the next step

Navigate to the previous step

Navigate to the next Check & Clean


chapter

Navigate back to the previous Check


& Clean section

Enter / confirm

Start the Check & Clean function (e.g.


check water flow)

Stop

Interrupt or abort a process

You can then confirm the cancellation or return to the Check & Clean
04:00 PM
process.

67 46 429 D3708
110 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.1 Check & Clean process

A Check & Clean process is only considered completed and appears as


04:00 PM
an entry in the history if the last step of the Check & Clean process
appears on the display and a no-load operation (cycle without
instruments) was performed with a NitraClean cleaning tablet.

5.1.1 Start Check & Clean


ü The DAC Universal is switched on and cooled down.
➢ Start the Check & Clean process.
The Check & Clean process can be started by either pressing
● the "Check & Clean 50" function key

● the "Check & Clean 200" function key

directly on the start dialog (if active) or via the menu with the function
key.

The Check & Clean process can be interrupted with pop-up messages.
The "Check & Clean not possible" function key appears on the display.
Rectify this error to start the Check & Clean process.

5.1.2 Put on gloves


Frequency: Weekly or every 50 cycles (whichever comes first).
For your own protection, put on gloves before you start the Check &
04:00 PM
Clean process.

67 46 429 D3708
D3708.201.04.02.02 10.2020 111
5 Cleaning, maintenance and validation Dentsply Sirona
5.1 Check & Clean process Instructions for use DAC Universal

5.1.3 Clean the drain filter


Frequency: Weekly or every 50 cycles (whichever comes first).

IMPORTANT
With approx. 10 cycles per day, the drain filter must be changed every
4 weeks, see "Replace the drain filter [→ 116]".

IMPORTANT
The chamber must be empty before the drain filter is screwed out.

ü The DAC Universal is switched on and is in the Check & Clean


process menu.
ü Wear your personal protective gear.
ü The screwdriver included in the scope of supply is available.
1. Go to the next Check & Clean option by pressing the "Next Check &
Clean step" function key.

2. Or go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.
3. Fold up the flap of the drain connection.

4. Unscrew the lid of the filter housing using the screwdriver.


04:00 PM

5. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

6. Pull the drain filter off the lid of the filter housing.
04:00 PM

WARNING
Risk of contamination!
The drain filter may contain germs.

CAUTION
Risk of burns!
The drain filter may be hot.

67 46 429 D3708
112 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.1 Check & Clean process

7. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

8. Clean the inside and outside of the drain filter thoroughly under
04:00 PM
running water.

9. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

10. Place the drain filter back on the lid of the filter housing.
04:00 PM

11. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.

12. Screw the lid of the filter housing back in by turning the screwdriver
04:00 PM
clockwise.
13. Close the drain connection flap again.

NOTE
Tighten the filter housing well without using force.

67 46 429 D3708
D3708.201.04.02.02 10.2020 113
5 Cleaning, maintenance and validation Dentsply Sirona
5.1 Check & Clean process Instructions for use DAC Universal

5.1.4 Check the O-rings on the drain filter and change if


necessary
Frequency: Weekly or every 50 cycles (whichever comes first).

IMPORTANT
The drain filter should be cleaned weekly or after every 50 cycles, see
"Clean the drain filter [→ 112]".

ü The DAC Universal is switched on and is in the Check & Clean


process menu.
ü Wear your personal protective gear.
ü The screwdriver included in the scope of supply is available.
1. Go to the next C&C option by pressing the "Next Check & Clean
step" function key.

2. Or go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

3. Fold up the flap of the drain connection.


04:00 PM
4. Unscrew the lid of the filter housing using the screwdriver.

5. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

6. Pull the drain filter off the lid of the filter housing.
04:00 PM

WARNING
Risk of contamination!
The drain filter may contain germs.

CAUTION
Risk of burns!
The drain filter may be hot.

7. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

67 46 429 D3708
114 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.1 Check & Clean process

8. Dispose of the drain filter with the practice waste. Observe the
04:00 PM
relevant regulations in your country. The drain filter should only be
replaced every 4 months or every 200 cycles (whichever comes
first).

9. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

10. Check the O-ring on the lid of the filter housing. If the O-ring is
04:00 PM
damaged it must be replaced. Dispose of the O-ring with the
practice waste. Observe the relevant regulations in your country.

11. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.

12. Place the new drain filter on the lid of the filter housing
04:00 PM

NOTE
Tighten the drain filter well without using force.

13. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.

14. Screw the lid of the filter housing back in by turning the screwdriver
04:00 PM
clockwise.
15. Close the drain connection flap again.

NOTE
Tighten the filter housing well without using force.

67 46 429 D3708
D3708.201.04.02.02 10.2020 115
5 Cleaning, maintenance and validation Dentsply Sirona
5.1 Check & Clean process Instructions for use DAC Universal

5.1.5 Replace the drain filter


Frequency: Every 4 weeks or every 200 cycles (whichever comes first).

IMPORTANT
The drain filter should be cleaned weekly or after every 50 cycles, see
"Clean the drain filter [→ 112]".

ü The DAC Universal is switched on and is in the Check & Clean


process menu.
ü Wear your personal protective gear.
ü The screwdriver included in the scope of supply is available.
1. Go to the next Check & Clean option by pressing the "Next Check &
Clean step" function key.

2. Or go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

3. Fold up the flap of the drain connection.


04:00 PM
4. Unscrew the lid of the filter housing using the screwdriver.

5. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

6. Pull the drain filter off the lid of the filter housing.
04:00 PM

WARNING
Risk of contamination!
The drain filter may contain germs.

CAUTION
Risk of burns!
The drain filter may be hot.

7. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

67 46 429 D3708
116 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.1 Check & Clean process

8. Dispose of the drain filter with the practice waste. Observe the
04:00 PM
relevant regulations in your country.

9. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

10. Check the O-ring on the lid of the filter housing. If the O-ring is
04:00 PM
damaged it must be replaced. Dispose of the O-ring with the
practice waste. Observe the relevant regulations in your country.
The O-ring must be checked weekly or every 50 cycles (whichever
comes first), see "Check the O-rings on the drain filter and change if
necessary [→ 114]".

11. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.

12. Place the new drain filter on the lid of the filter housing
04:00 PM

NOTE
Tighten the drain filter well without using force.

13. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.

14. Screw the lid of the filter housing back in by turning the screwdriver
04:00 PM
clockwise.
15. Close the drain connection flap again.

NOTE
Tighten the filter housing well without using force.

67 46 429 D3708
D3708.201.04.02.02 10.2020 117
5 Cleaning, maintenance and validation Dentsply Sirona
5.1 Check & Clean process Instructions for use DAC Universal

5.1.6 Insert Check & Clean lid


Frequency: Weekly or every 50 cycles (whichever comes first).
ü The DAC Universal is switched on and is in the Check & Clean
process menu.
ü The Check & Clean lid is available.
1. Go to the next Check & Clean option by pressing the "Next Check &
Clean step" function key.

2. Or go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

3. Place the Check & Clean lid at the front on the DAC Universal
04:00 PM
holder and insert it straight into the holder.

NOTE
Make sure that the lid is inserted all the way into the holder.

WARNING
Risk of contamination!
Make sure there are no objects in the chamber.

5.1.7 Check the safety ring


Frequency: Weekly or every 50 cycles (whichever comes first).
This function must be checked to ensure that the safety ring works and
the Check & Clean process can be started.
ü The DAC Universal is switched on and is in the Check & Clean
process menu.
ü The Check & Clean lid is correctly inserted into the holder.
1. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.

2. Press down on the safety ring to start the Check & Clean process.
04:00 PM
Ä The lid closes.

CAUTION
If you cannot start the Check & Clean process with the safety ring, the
safety ring setup is faulty. The device must then not be used. Please
contact your authorized dealer.

67 46 429 D3708
118 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.1 Check & Clean process

5.1.8 Check the Check & Clean lid


Frequency: Weekly or every 50 cycles (whichever comes first).
The Check & Clean lid is detected automatically by the DAC Universal
software.
If the Check & Clean lid is not recognized, the holder moves up and the
lid must be reinserted into the holder.

5.1.9 Check the water supply to the chamber


Frequency: Weekly or every 50 cycles (whichever comes first).
In order to make sure that the DAC Universal is able to bring water to
the chamber correctly, this operation needs to be tested.
ü The DAC Universal is switched on and is in the Check & Clean
process menu.
ü The Check & Clean lid is correctly inserted into the holder.
1. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.
2. Press the "Check & Clean - start function" function key to start the
water flow into the chamber.

Ä The DAC Universal pumps water for 10 seconds.


04:00 PM
Ä Water comes through the 6 holes in the chamber.
3. Repeat the process if no water appears.

IMPORTANT
If the water does not flow as described above, please contact your
authorized dealer.

5.1.10 Check the oil supply via the Check & Clean lid
Frequency: Weekly or every 50 cycles (whichever comes first).
This function must be checked to ensure that the DAC Universal oils the
instruments correctly and that the internal oil channels supply oil to the
Blue lid and the Pink lid.
ü The DAC Universal is switched on and is in the Check & Clean
process menu.
ü The Check & Clean lid is correctly inserted into the holder.
1. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.

2. Press the "Check & Clean - start function" function key to start the
oil supply.

67 46 429 D3708
D3708.201.04.02.02 10.2020 119
5 Cleaning, maintenance and validation Dentsply Sirona
5.1 Check & Clean process Instructions for use DAC Universal

Ä The DAC Universal pumps oil 5 times and then air for 10
04:00 PM
seconds.
Ä Oil comes through the drive channel (triangle marking).
3. Repeat this process if no oil appears.

IMPORTANT
If the oil does not flow as described above, please contact your
authorized dealer.

IMPORTANT
If the oil is empty during the check and an error message appears,
this must be rectified before the Check & Clean process can be
restarted.

5.1.11 Check the water supply via the Check & Clean lid
Frequency: Weekly or every 50 cycles (whichever comes first).
In order to make sure that the DAC Universal is able to bring water to
the chamber and internal channels of the Blue / Green / Pink and White
lids correctly, this function needs to be tested.
ü The DAC Universal is switched on and is in the Check & Clean
process menu.
ü The Check & Clean lid is correctly inserted into the holder.
1. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.

2. Press the "Check & Clean - start function" function key to start the
water supply.

Ä The DAC Universal pumps water for 10 seconds and then air
04:00 PM
for 10 seconds.
Ä Water comes through the spray channel (circle marking).
Ä The DAC Universal pumps water for 10 seconds and then air
for 10 seconds.
Ä Water comes through the drive channel (triangle marking).
3. Repeat the process if no water appears.

IMPORTANT
If the water does not flow as described above, please contact your
authorized dealer.

67 46 429 D3708
120 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.1 Check & Clean process

5.1.12 Open chamber / remove Check & Clean lid


Frequency: Weekly or every 50 cycles (whichever comes first).
ü The DAC Universal is switched on and is in the Check & Clean
process menu.
1. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.
Ä The lid moves up.

2. Remove the Check & Clean lid from the holder on the DAC
04:00 PM
Universal.

5.1.13 Check the permeability of the media paths of all lids


Frequency: Weekly or every 50 cycles (whichever comes first).
This function is used to check whether there are blockages within the lid
and their adapters.
For the following Check & Clean steps, the DAC Universal only serves
to guide the user through the Check & Clean process. Nothing else is
performed on the DAC Universal itself.

IMPORTANT
The water supply to the lid now needs to be checked for all lids, since
these are fitted with internal water channels.

ü The DAC Universal is switched on and is in the Check & Clean


process menu.
ü All available lids are ready-to-hand and have cooled down.
ü The Check & Clean cap is available.

67 46 429 D3708
D3708.201.04.02.02 10.2020 121
5 Cleaning, maintenance and validation Dentsply Sirona
5.1 Check & Clean process Instructions for use DAC Universal

1. Go to the next Check & Clean option by pressing the "Next Check &
Clean step" function key.

2. Or go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

3. Place the Check & Clean cap on the lid, here by way of example the
04:00 PM
Blue lid.

4. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.
5. Open the front flap of the DAC Universal.

6. Remove the water tank by carefully pulling the water tank up and
04:00 PM
out of the DAC Universal by the handle.
7. Place the water tank on a flat surface.
8. Remove the lid of the water tank.

9. Remove some water from the tank with a syringe (at least 10 ml).
04:00 PM

WARNING
Risk of clogging!
Only use demineralized or distilled water for the check, otherwise
blockages can occur, which can damage the lid and therefore prevent
the cleaning of the instruments.

67 46 429 D3708
122 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.1 Check & Clean process

04:00 PM

10. Inject the water with the syringe through both holes into the
channels of the Pink lid (as shown in the image).
Ä Water flows through the small holes of all adapters.

04:00 PM

11. Inject the water with the syringe through the left hole into the
channel of the White lid (as shown in the image).
Ä Water flows through the small holes of all adapters.

04:00 PM

12. Inject the water with the syringe through both holes into the
channels of the Blue lid (as shown in the image).
Ä Water flows through the small holes of all adapters.

67 46 429 D3708
D3708.201.04.02.02 10.2020 123
5 Cleaning, maintenance and validation Dentsply Sirona
5.1 Check & Clean process Instructions for use DAC Universal

04:00 PM

13. Inject the water with the syringe through the left hole into the
channel of the Green lid (as shown in the image).
Ä Water flows through the small holes of all adapters.

CAUTION
Risk of burns!
Leave the lid to cool down sufficiently, otherwise there is a risk that
steam will form while injection the water.

14. Confirm that water is flowing by pressing the "Confirm" function key.
15. Repeat the process if no water appears.
16. Repeat the process for all available lids and with all adapters.
Ä All lids are checked.

NOTE
Do not use the lid in which the water does not flow as described
above and contact your authorized dealer.

5.1.14 Disassemble and clean the water tank


Frequency: Weekly or every 50 cycles (whichever comes first).

Disassemble and empty the water tank


ü The DAC Universal is switched on and is in the Check & Clean
process menu.
ü The water tank is on a flat surface.
1. Go to the next Check & Clean option by pressing the "Next Check &
Clean step" function key.

2. Or go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

67 46 429 D3708
124 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.1 Check & Clean process

3. Drain the water tank.


04:00 PM
Ä There is no water left in the tank.

4. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

5. Disassemble the water tank by removing the handle, the lid and the
04:00 PM
float including the float guide.

6. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

Clean the water tank


There are 2 options for cleaning the water tank:
● Cleaning the individual parts in a thermal disinfector
04:00 PM
● Manual cleaning as described below:
1. Rinse the water tank with warm water.
2. Empty the water tank again.
3. Clean the inside and outside of the water tank carefully with a moist
cloth.
4. Wipe the inside and outside of the water tank dry with a clean cloth.
5. Rinse the handle, lid and the float incl. float guide with warm water.
6. Clean the handle, the lid and the float incl. the float guide carefully
with a moist cloth.
7. Dry the inside and outside of the handle, the lid and the float incl.
the float guide with a clean cloth.

IMPORTANT
There must be no residual water in the water tank, otherwise the
conductivity will be measured. An alarm sounds and an error
message appears.

67 46 429 D3708
D3708.201.04.02.02 10.2020 125
5 Cleaning, maintenance and validation Dentsply Sirona
5.1 Check & Clean process Instructions for use DAC Universal

5.1.15 Clean the drain tank


Frequency: As soon as it is full and in addition, weekly or every 50
cycles (whichever comes first).

WARNING
Risk of contamination!
Do not start a cycle if the DAC Universal is not connected to the drain
tank or a siphon or if the drainage system installation was not
performed correctly.

ü The DAC Universal is switched on and is in the Check & Clean


process menu.
ü The drain tank is cooled down.
ü Wear your personal protective gear.
1. Go to the next Check & Clean option by pressing the "Next Check &
Clean step" function key.

2. Or go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

3. Unscrew the lid of the drain tank.

4. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

5. Remove the condensation tube.

6. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

67 46 429 D3708
126 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.1 Check & Clean process

7. Empty the drain tank. Dispose of the waste water according to the
national guidelines and the applicable regulations for waste
disposal.

8. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

9. Fill the drain tank with hot tap water.

10. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.

11. Rinse the drain tank thoroughly with hot water.

12. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.

13. Empty the drain tank completely.


14. Soak a soft cloth in hot water and wring it out well. Use the cloth to
clean the exterior of the drain tank and the condensation tube.

15. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.

67 46 429 D3708
D3708.201.04.02.02 10.2020 127
5 Cleaning, maintenance and validation Dentsply Sirona
5.1 Check & Clean process Instructions for use DAC Universal

16. Fill the clean drain tank with water up to the lowest mark on the
label.

WARNING
Risk of contamination!
Add isopropyl (alcohol ≥ 70 %) up to the second mark on the label.
This prevents bacteria from growing in the drain container.

17. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.

18. Replace the condensation tube.

19. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.

20. Screw down the lid again.


21. Place the drain tank no more than 3 m from the DAC Universal and
at a lower height.
Ä The drain tank is cleaned and emptied.

WARNING
Risk of contamination!
Check that the drain hose is not blocked after emptying and cleaning.

5.1.16 Check the assembly of the adapters


Frequency: Weekly or every 50 cycles (whichever comes first).

CAUTION
Burning hazard!
The metal part of the lid is very hot at the end of the cycle. Please
allow the lid to cool down before cleaning it.

For further information about the adapters and the corresponding O-


rings, please see the separate instructions for use for adapters.

Procedure
ü The DAC Universal is switched on and is in the Check & Clean
process menu.
ü All available lids are ready.

67 46 429 D3708
128 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.1 Check & Clean process

1. Go to the next Check & Clean option by pressing the "Next Check &
Clean step" function key.

2. Or go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.
3. Place the lid on a flat surface.

4. Check the attachment of the adapters to the lid by pulling each


04:00 PM
adapter slightly.
5. If an adapter is loose, tighten the screws with the 1.5 mm Allen key
included with the unit.
6. Repeat the process for all available lids.

5.1.17 Check O-rings on the adapters


Frequency: Weekly or every 50 cycles (whichever comes first).

Procedure
ü The DAC Universal is switched on and is in the Check & Clean
process menu.
ü All available adapters are ready.
1. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.

2. Check the O-rings on the adapters.


04:00 PM
3. Repeat for all available adapters.

WARNING
Risk of contamination!
With approx. 10 cycles per day, the O-rings need to be replaced
every 3 months. If an adapter O-ring is damaged or missing, replace
it, see "Change the O-rings on the adapters [→ 130]".

WARNING
Risk of contamination!
If the O-rings are damaged or are missing, the reprocessing cycle
cannot be guaranteed. Cross-contamination may occur.

For further information about the adapters and the corresponding O-


rings, please see the separate operating instructions for adapters.

67 46 429 D3708
D3708.201.04.02.02 10.2020 129
5 Cleaning, maintenance and validation Dentsply Sirona
5.1 Check & Clean process Instructions for use DAC Universal

5.1.18 Change the O-rings on the adapters


Frequency: Every 3 months or every 600 cycles (whichever comes
first).
ü The DAC Universal is switched on and is in the Check & Clean
process menu.
ü All available adapters are ready.
1. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.
2. Check the O-rings on the adapters.
3. Remove the O-rings from the adapter using a dental probe or other
instrument.
4. Slide the new O-rings into the notches on the adapter.
5. Repeat for all available adapters.

WARNING
Risk of contamination!
If the O-rings are damaged or are missing, the reprocessing cycle
cannot be guaranteed. Cross-contamination may occur.

Some adapters have internal O-rings that must be replaced by your


service technician during servicing (see separate adapter operating
instructions).
For further information about the adapters and the corresponding O-
rings, please see the separate operating instructions for adapters.

5.1.19 Clean the housing of the unit / the lid and the chamber
Frequency: Weekly or every 50 cycles (whichever comes first).
ü The DAC Universal is switched on and is in the Check & Clean
process menu.
ü All available lids are ready.
1. Go to the next Check & Clean option by pressing the "Next Check &
Clean step" function key.

2. Clean the unit housing, all the available lids and the chamber of the
04:00 PM
DAC Universal with a soft, lint-free cloth and a solution made of
70% isopropyl.

67 46 429 D3708
130 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.1 Check & Clean process

04:00 PM NOTE
The surface must not be cleaned with agents containing chlorine.

NOTE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning, and disinfecting agents approved by
Dentsply Sirona!

A continuously updated list of approved agents can be downloaded on


the internet from the online portal for technical documents. You can
reach the portal at the address:
www.dentsplysirona.com/ifu
Click on the menu items"General documents", "All products" and then
open the "Care, cleaning and disinfection agents" document.
If you do not have access to the internet, please contact your dental
dealer to order the list.

WARNING
Risk of contamination!
Make sure that there are no objects in the chamber, on the lids or on
the adapters.

5.1.20 Check the O-ring on the water tank


Frequency: Weekly or every 50 cycles (whichever comes first).
ü The DAC Universal is switched on and is in the Check & Clean
process menu.
1. Go to the next Check & Clean option by pressing the "Next Check &
Clean step" function key.

2. Or go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

3. Check the O-ring on the seal underneath the water tank.


04:00 PM
4. Replace the O-ring if it is defective.

WARNING
Risk of contamination!
If the O-rings are damaged or are missing, the reprocessing cycle
cannot be guaranteed. Cross-contamination may occur.

67 46 429 D3708
D3708.201.04.02.02 10.2020 131
5 Cleaning, maintenance and validation Dentsply Sirona
5.1 Check & Clean process Instructions for use DAC Universal

5.1.21 Assemble / fill and insert the water tank


Frequency: Weekly or every 50 cycles (whichever comes first).

NOTE
Remove the water tank from the DAC Universal before filling it up.
This prevents water from spilling into the inside of the DAC Universal
housing.

ü The DAC Universal is switched on and is in the Check & Clean


process menu.
ü Demineralized / distilled water (water conductivity < 3.0 µS/cm) is
available.
ü The water tank is on an even surface.
1. Go a step further in this Check & Clean option by pressing the
"Navigate to next step" function key.

2. Assemble the water tank by inserting the handle, the lid and the
04:00 PM
float including the float guide.

3. If you do not have an automatic water connection for your DAC


04:00 PM
Universal, fill demineralized/distilled water (water conductivity < 3.0
µS/cm) into the water tank (min. 1.0 liter, max. 2 liter).

4. Insert the water tank back into the DAC Universal.


04:00 PM

IMPORTANT
When inserting make sure that the guide hooks on the top rear side of
the water tank are inserted correctly into the housing.

67 46 429 D3708
132 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.1 Check & Clean process

5.1.22 Preparations after the Check & Clean process


Frequency: Weekly or every 50 cycles (whichever comes first).
ü The DAC Universal is switched on and is in the Check & Clean
process menu.
ü NitraClean cleaning tablets are available.

CAUTION
Risk of poisoning!
Only use the NitraClean cleaning tablets stated below

1. Go to the next Check & Clean option by pressing the "Next Check &
Clean step" function key.

2. Place a NitraClean tablet in the empty chamber.


04:00 PM

3. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

4. Take a lid without instruments.


04:00 PM

5. Place the lid at the front on the DAC Universal holder and insert the
04:00 PM
lid straight in the holder.

6. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

67 46 429 D3708
D3708.201.04.02.02 10.2020 133
5 Cleaning, maintenance and validation Dentsply Sirona
5.1 Check & Clean process Instructions for use DAC Universal

Ä The Check & Clean process is almost complete.


04:00 PM

7. Go a step further in this Check & Clean option by pressing the


"Navigate to next step" function key.

8. Start a no-load cycle (cycle without instruments) with the "Start"


function key in order to remove any remaining oil residue from the
chamber.
Ä After the cycle is finished, the lid moves a little out of the
chamber.

CAUTION
Burning hazard!
Never hold your head directly over the DAC Universal when it opens,
as steam may emerge from the unit.

9. Press the "Confirm" function key.


Ä The lid will open completely.

CAUTION
Risk of crushing!
Make sure that your fingers do not get caught between the opening lid
and an object, such as any wall cabinets that may be above.

Ä The Check & Clean process is completed.

67 46 429 D3708
134 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.2 Biennial Service

NitraClean cleaning tablets


NitraClean cleaning tablets are used for cleaning the empty chamber
and internal cleaning of the hoses.
The empty chamber and the hoses must be cleaned when needed, or at
least in the course of Check & Clean.

IMPORTANT
NitraClean cleaning tablets must not be put in the water tank.

Please see the MSDS for further information for the NitraClean tablets.
http://www.dentsplysirona.com/en/service/safety-data-sheets/

5.2 Biennial Service


Frequency: Every 2 years or every 3000 cycles (whichever comes first).
After 2 years or 3000 cycles after the last service or initial
commissioning of the DAC Universal the message "Service necessary"
appears on the display and, at the same time, a 2-second acoustic
signal sounds.
When this message appears, a qualified technician must perform a
service on the DAC Universal to ensure the proper and efficient
performance of the DAC Universal.

5.3 Install a new lid


If the lid is closed, the DAC Universal begins to recognize the lid.
The display shows the lid type and the serial number.
04:00 PM
To confirm a new lid press the function key "Confirm".

IMPORTANT
SN 8888
Make sure that the reprocessing processes are also validated with
your new lid type if this type of lid has not yet been validated.
Observe any possible additional national requirements.

67 46 429 D3708
D3708.201.04.02.02 10.2020 135
5 Cleaning, maintenance and validation Dentsply Sirona
5.4 Mounting adapters Instructions for use DAC Universal

5.4 Mounting adapters


Up to 6 instruments can be reprocessed simultaneously in the DAC
Universal. The instruments are connected to the lid using adapters.
These adapters are not yet mounted when the unit is delivered.
An overview of all adapters, a detailed procedure for installing the
adapters on the respective lids and a detailed procedure for installing
the instruments on the adapters can be found in the separate operating
instructions for adapters.

5.5 Adding/replacing the lubrication concentrate


General information about the NitramOil #2 lubrication concentrate
bottle can be found in the chapter "NitramOil #2 lubrication concentrate
[→ 33]".

Insert the oil bottle


ü The lid on the DAC Universal is open.
ü The transport bottle on the oil dispenser is removed.
1. Remove the lid from the new NitramOil #2 bottle.

2. Press the black top of the oil bottle to release the transport lock.
3. Insert the oil bottle thread down into the thread of the oil dispenser
(next to the water tank).

67 46 429 D3708
136 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.5 Adding/replacing the lubrication concentrate

4. Turn the oil bottle carefully clockwise until you meet resistance.
5. Close the cover on the DAC Universal.
Ä The oil bottle is now inserted in the DAC Universal.

IMPORTANT
If no oil bottle is inserted, the unit will show an error message when
using the Blue lid or Pink lid.

Replacing the oil bottle

IMPORTANT
The oil bottle must be replaced when the "Low oil" message appears
on the display.

ü The lid on the DAC Universal is open.


1. Unscrew the empty oil bottle counterclockwise and remove it from
the oil dispenser.

IMPORTANT
If the oil bottle is not completely empty, small quantities of oil may
leak out of the bottle. Wipe the oil away with an absorbent cloth.

2. Remove the lid from the new NitramOil #2 bottle.


3. Press the black top of the oil bottle to release the transport lock.
4. Insert the new oil bottle thread down into the thread of the oil
dispenser (next to the water tank).
5. Turn the oil bottle carefully clockwise until you meet resistance.
6. Close the cover on the DAC Universal.
Ä The oil bottle has been replaced.
Tip: You can also remove a partially used oil bottle that is not yet empty
from the DAC Universal and insert it again at a later time.

67 46 429 D3708
D3708.201.04.02.02 10.2020 137
5 Cleaning, maintenance and validation Dentsply Sirona
5.6 Drain the chamber Instructions for use DAC Universal

5.6 Drain the chamber


If there is too much water left in the chamber, you can manually drain
the chamber via the menu:
ü The DAC Universal is switched on and ready for operation.
ü An empty lid is at hand and cooled down.
1. Turn over the lid so that the adapters are facing downwards. The
colored marking is pointed towards you.
2. Place the lid at the front on the DAC Universal holder and insert the
lid straight in the holder.

NOTE
Make sure that the lid is inserted all the way into the holder. Make
sure there are no objects in the chamber. The DAC Universal, the lid
and/or the instruments can be damaged if a lid is not inserted
correctly.

3. Press the "Menu" function key from the homescreen.

4. Press the "Manual" function key.

5. Press the "Manual: Empty chamber" function key.


Ä The lid closes automatically.
Ä The water is pumped from the chamber into the drain tank /
drainage system.
Ä The lid opens again automatically after 60 seconds.

IMPORTANT
If the chamber is not completely empty, repeat the process until the
chamber is completely empty.

5.7 Validation
The DAC Universal has been type tested and works tested according to
requirements from applicable standards.
The manufacturer recommends validating the reprocessing process
with the DAC Universal every 12 months.
Your authorized dealer can supply information on whether local
validation procedures exist for your country. If so, the manufacturer
recommends performing validation of reprocessing with the DAC
Universal as specified by the local authorities.

67 46 429 D3708
138 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.8 Uninstallation and transport

5.8 Uninstallation and transport


Uninstallation
If the DAC Universal needs to be transported, please follow the
guidelines below.
ü The red transport plug for the chamber is available.

ü The transport bottle for the oil dispenser is available.

67 46 429 D3708
D3708.201.04.02.02 10.2020 139
5 Cleaning, maintenance and validation Dentsply Sirona
5.8 Uninstallation and transport Instructions for use DAC Universal

ü The 3 black caps for the connections are available.


1. If there is still a lid (e.g. a Blue lid) in the holder of the DAC
Universal, remove this. If the DAC Universal is shipped then it must
be packed separately.

2. Put the red transport plug on the chamber.


3. If the DAC Universal is connected to an automatic water supply
system, switch this off and deactivate the automatic water supply in
the menu (under Settings/Technical settings/Settings: Automatic
supply of demineralized water).
4. Using the menu (under Manual / Manual: Close), let the lid holder
go down and stop it before it touches the safety ring.
5. Empty the water tank.
6. Turn off the DAC Universal at the mains switch.
7. Pull out the power cable of the DAC Universal.
8. Remove the oil bottle from the dispenser.
9. Clean the oil dispenser and suck out any residual oil through the
hole under the oil dispenser by using a syringe.

10. Place the transport bottle in the oil dispenser.


11. Turn off the compressed air supply.
12. Disconnect the tube for the air inlet.
13. Unplug the drain hose connection.
14. Disconnect the hose from the water supply if there is an automatic
water supply.

67 46 429 D3708
140 D3708.201.04.02.02 10.2020
Dentsply Sirona 5 Cleaning, maintenance and validation
Instructions for use DAC Universal 5.8 Uninstallation and transport

15. Place the 3 black caps on the connections.


16. Pack the DAC Universal. (The DAC Universal must be sent in the
original packaging incl. foam, etc. -REF 6718246-.)

Transport

IMPORTANT
Nature and source of the danger
Allow the DAC Universal to cool down before packaging it.

CAUTION
Risk of burns!
The bottom plate and back panel of the DAC Universal can be hot.

CAUTION
Risk of injury!
The DAC Universal weighs approx. 26 kg / 57 pounds and one lid
(incl. 6 adapters) weighs approx. 1.7 kg / 3.75 pounds.
Therefore, when transporting the DAC Universal, make sure that you
hold the unit securely by the grips on the side.
It is not recommended to carry the DAC Universal for a great
distance.

67 46 429 D3708
D3708.201.04.02.02 10.2020 141
6 Errors / Display messages Dentsply Sirona
6.1 Warning messages Instructions for use DAC Universal

6 Errors / Display messages


There are 4 kinds of messages:
● Warning messages
● Process error messages as pop-up messages
● Error messages as pop-up messages
● User messages

6.1 Warning messages


Warning messages appear regardless of the cycle, do not discontinue
04:00 PM
the cycle and can also appear during standby times. They are only
displayed on the homescreen on the right side.
There are 3 kinds of warning messages:

● Poor water quality

● Service needed

● No oil (there is no oil in the lubrication concentrate bottle or a


lubrication concentrate bottle is missing)

67 46 429 D3708
142 D3708.201.04.02.02 10.2020
Dentsply Sirona 6 Errors / Display messages
Instructions for use DAC Universal 6.2 Process error messages as pop-up messages

6.2 Process error messages as pop-up messages


Process error messages as pop-up messages
These messages may appear during a reprocessing cycle or if a
04:00 PM
1002 reprocessing cycle is to be started. They always result in the
reprocessing cycle being discontinued and appear as a pop-up
message.

The symbol appears that the cycle has been discontinued, an alarm
signals and the LED flashes.
If a process error message appears during the reprocessing cycle, the
lid opens and moves up.
There is a tick/confirmation function key to confirm that the error
message has been acknowledged.
If there are several error messages at the same time, only one is
shown. After the first is displayed, the next message is displayed. There
is no specific sequence for this. The message can be confirmed, which
switches off the alarm and then the homescreen is displayed again.
There is no additional window for all error messages that have resulted
in the reprocessing cycles being canceled.
After confirming the message(s), you are automatically taken to the
homescreen.
There are 8 kinds of process error messages:
● No compressed air
● Water tank is not inserted in the DAC Universal
● Poor water quality
● No oil (there is no oil in the lubrication concentrate bottle or a
lubrication concentrate bottle is missing)
● Service needed. Please check the operating instructions or contact
your authorized dealer.
● The process was personally discontinued.
● A new lid has been inserted into the DAC Universal.

67 46 429 D3708
D3708.201.04.02.02 10.2020 143
6 Errors / Display messages Dentsply Sirona
6.3 Error messages as pop-up messages Instructions for use DAC Universal

6.3 Error messages as pop-up messages


The messages only appear outside a reprocessing cycle. They appear
04:00 PM
1002 as a pop-up message but without the symbol that the cycle was
canceled, no acoustic alarm sounds and the LED does not light up.
The error message covers the "Start reprocessing cycle“ function key.
There are 4 kinds of error messages as pop-up messages:
● Please contact your authorized dealer.
● Safety ring was pressed
● No compressed air
● Front flap on the water tank is still open
All error messages are displayed, as soon as the message has been
confirmed the start screen appears again.
Exception:
The error message shown here appears if the "Cancel" function key is
04:00 PM
pressed during an ongoing reprocessing cycle. The DAC Universal
shows this message to confirm whether you really want to cancel the
cycle or have canceled the cycle accidentally.

67 46 429 D3708
144 D3708.201.04.02.02 10.2020
Dentsply Sirona 6 Errors / Display messages
Instructions for use DAC Universal 6.4 User messages

6.4 User messages


User messages
They cover the cycle start key in order to prevent the cycle from being
started if the DAC Universal cannot/must not be started.
No acoustic alarm will sound and the LED does not flash.
The sequence of the user messages indicates their priority and also
shows the order in which the user should resolve the messages. They
appear one on top of the other so that the user can handle them in
order.
There are 6 kinds of user messages:
● Please contact your authorized dealer.

● Insufficient or no compressed air supply

● Water tank is not inserted in the DAC Universal

67 46 429 D3708
D3708.201.04.02.02 10.2020 145
6 Errors / Display messages Dentsply Sirona
6.4 User messages Instructions for use DAC Universal

● No water

● Poor water quality

● Front flap on the water tank is still open

67 46 429 D3708
146 D3708.201.04.02.02 10.2020
Dentsply Sirona 6 Errors / Display messages
Instructions for use DAC Universal 6.5 Display messages and what they mean

6.5 Display messages and what they mean


The display messages are assigned a code number. The code number
04:00 PM
1002 is visible in the upper left-hand corner, next to the error symbol. An
example of this is shown on the left.
In the table below you will find the code numbers for some display
messages together with an explanation and troubleshooting for the
specific displayed message.

IMPORTANT
If the DAC Universal displays a message which you cannot find in the
table below, please contact your dealer authorized by Dentsply
Sirona.

Code Description Troubleshooting


no.
1001 The chamber pressure is above Check the drain filter.
2.7 bar rel. or below 0.1 bar rel. Check the drain connection and drain hose.
If the error occurs repeatedly, please contact your authorized
dealer.
1002 Cycle discontinued by the user. If the error occurs unintentionally, please contact your authorized
dealer.
1003 The safety ring was pressed. If the error occurs unintentionally, please contact your authorized
dealer.
1005 The lid type is not yet registered. Initialize the new lid type
If the error occurs unintentionally, please contact your authorized
dealer.
1006 The water tank is not inserted. Insert the water tank.
Check whether the float is in the water tank.
If the error occurs repeatedly, please contact your authorized
dealer.
1008 Lack of compressed air Check the compressed air supply.
If the error occurs repeatedly, please contact your authorized
dealer.
1009 The device was switched off and Check whether the plug is connected correctly.
on too quickly during a cycle. If the error occurs repeatedly, please contact your authorized
dealer.
1011 No oil is detected. Change the oil bottle if this is empty.
Make sure that you proceeded as per chapter "Adding/replacing
the lubrication concentrate [→ 136]" when inserting the oil bottle.
If the error occurs repeatedly, please contact your authorized
dealer.
1102 The water tank flap was opened Close the water tank flap.
whilst the chamber is closed or If the error occurs repeatedly, please contact your authorized
open. dealer.

67 46 429 D3708
D3708.201.04.02.02 10.2020 147
6 Errors / Display messages Dentsply Sirona
6.5 Display messages and what they mean Instructions for use DAC Universal

Code Description Troubleshooting


no.
1303 Error in the drain area Check the drain filter.
Check the drain connection and drain hose.
If the error occurs repeatedly, please contact your authorized
dealer.
1405 Error in the drain area Check the drain filter.
Check the drain connection and drain hose.
If the error occurs repeatedly, please contact your authorized
dealer.
1407 The water quality is above the limit Check the water quality.
value. Check the water in the water tank.
Check your water treatment system.
If the error occurs repeatedly, please contact your authorized
dealer.
1510 Error in the drain area Check the drain filter.
Check the drain connection and drain hose.
If the error occurs repeatedly, please contact your authorized
dealer.
1511 Error in the drain area Check the drain filter.
Check the drain connection and drain hose.
If the error occurs repeatedly, please contact your authorized
dealer.
1605 Error in the drain area Check the drain filter.
Check the drain connection and drain hose.
If the error occurs repeatedly, please contact your authorized
dealer.
1705 Error in the drain area Check the drain filter.
Check the drain connection and drain hose.
If the error occurs repeatedly, please contact your authorized
dealer.
1707 The water quality is above the limit Check the water quality.
value. Check the water in the water tank.
Check your water treatment system.
If the error occurs repeatedly, please contact your authorized
dealer.
1810 The water quality is above the limit Check the water quality.
value. Check the water in the water tank.
Check your water treatment system.
If the error occurs repeatedly, please contact your authorized
dealer.

67 46 429 D3708
148 D3708.201.04.02.02 10.2020
Dentsply Sirona 6 Errors / Display messages
Instructions for use DAC Universal 6.5 Display messages and what they mean

Code Description Troubleshooting


no.
1903 No oil is detected. Change the oil bottle if this is empty.
Make sure that you proceeded as per chapter "Adding/replacing
the lubrication concentrate [→ 136]" when inserting the oil bottle.
If the error occurs repeatedly, please contact your authorized
dealer.
2005 Error in the drain area Check the drain filter.
Check the drain connection and drain hose.
If the error occurs repeatedly, please contact your authorized
dealer.
2006 The water quality is above the limit Check the water quality.
value. Check the water in the water tank.
Check your water treatment system.
If the error occurs repeatedly, please contact your authorized
dealer.
2105 Error in the drain area Check the drain filter.
Check the drain connection and drain hose.
If the error occurs repeatedly, please contact your authorized
dealer.
2107 The water quality is above the limit Check the water quality.
value. Check the water in the water tank.
Check your water treatment system.
If the error occurs repeatedly, please contact your authorized
dealer.
2304 Error in the drain area Check the drain filter.
Check the drain connection and drain hose.
If the error occurs repeatedly, please contact your authorized
dealer.
2305 The water quality is above the limit Check the water quality.
value. Check the water in the water tank.
Check your water treatment system.
If the error occurs repeatedly, please contact your authorized
dealer.
2903 Error in the drain area Check the drain filter.
Check the drain connection and drain hose.
If the error occurs repeatedly, please contact your authorized
dealer.
2904 Error in the drain area Check the drain filter.
Check the drain connection and drain hose.
If the error occurs repeatedly, please contact your authorized
dealer.

67 46 429 D3708
D3708.201.04.02.02 10.2020 149
6 Errors / Display messages Dentsply Sirona
6.5 Display messages and what they mean Instructions for use DAC Universal

Code Description Troubleshooting


no.
2905 Error in the drain area Check the drain filter.
Check the drain connection and drain hose.
If the error occurs repeatedly, please contact your authorized
dealer.
2906 The water quality is above the limit Check the water quality.
value. Check the water in the water tank.
Check your water treatment system.
If the error occurs repeatedly, please contact your authorized
dealer.
5002 Error in the compressed air supply Check the compressed air supply.
area If the error occurs repeatedly, please contact your authorized
dealer.
5003 Error in the drain area Check the drain filter.
Check the drain connection and drain hose.
If the error occurs repeatedly, please contact your authorized
dealer.
5004 The function "Manual emptying of The function "Manual emptying of the chamber" is not possible
the chamber" was started with the with the Check & Clean lid.
Check & Clean lid. Insert a (reprocessing) lid.
If the error occurs repeatedly, please contact your authorized
dealer.
6000 Wrong lid during the Check & Insert the Check & Clean lid.
Clean process If the error occurs repeatedly, please contact your authorized
dealer.
6001 Error in the drain area Remove the blocked tamponade in the Check & Clean lid.
Check the drain filter.
If the error occurs repeatedly, please contact your authorized
dealer.
6002 Error in the drain area Remove the blocked tamponade in the Check & Clean lid.
Check the drain filter.
If the error occurs repeatedly, please contact your authorized
dealer.
9004 The safety ring was pressed whilst Release the safety ring and switch the device on again.
switching on. If the error occurs repeatedly, please contact your authorized
dealer.
9005 Error in the water supply area Check the water supply.
Check the mobility of float in the water tank.

67 46 429 D3708
150 D3708.201.04.02.02 10.2020
Dentsply Sirona 6 Errors / Display messages
Instructions for use DAC Universal 6.6 Errors without a message

6.6 Errors without a message


Description Corrective action
There is water in the chamber after end Drain the chamber and start a cycle.
of cycle. If the error persists, clean or change the drain filter.
Check whether the drain hose is blocked or damaged.
Empty the drain tank if it is full.
If the error occurs repeatedly, please contact your authorized dealer.
Adapter is loose. Tighten the adapter with the Allen key included.
The water tank is not automatically re- Check whether the automatic water intake is activated in the DAC Uni-
filled. versal menu, see "Technical settings [→ 68]".
Check whether water from the external water supply is fed to the DAC
Universal.
If the error occurs repeatedly, please contact your authorized dealer.
Loud continuous alarm signal. Switch off the DAC Universal immediately by unplugging the power
cord.
Please contact your authorized dealer.
IMPORTANT: Do not use the device until it has been serviced by your
authorized dealer.

NOTE
Safe operation of the DAC Universal cannot be ensured unless
service intervals are observed.

67 46 429 D3708
D3708.201.04.02.02 10.2020 151
7 Lid, adapter, consumables, spare parts, and tools Dentsply Sirona
Instructions for use DAC Universal

7 Lid, adapter, consumables, spare parts,


and tools
The accessories, spare parts and consumables for the DAC Universal
are available from your local authorized dealer.
For further information about the adapters and the corresponding O-
rings, please see the separate operating instructions for adapters.

67 46 429 D3708
152 D3708.201.04.02.02 10.2020
Dentsply Sirona 8 Disposal
Instructions for use DAC Universal

8 Disposal
WARNING
Risk of contamination!
The DAC Universal must be decontaminated before disposal.

On the basis of the Directive 2012/19/EU and the country-specific


disposal regulations on waste electrical and electronic equipment, we
would like to stress that such equipment must be disposed of
accordingly within the European Union (EU). These regulations demand
environmentally friendly recycling / disposal of waste electrical and
electronic equipment. These parts must not be disposed of as domestic
refuse. This has been symbolized by the symbol of the crossed out
garbage bin.

Disposal procedure
We feel responsible for our products from the first idea to their disposal.
For this reason, we give you an option to return our old electronic and
electrical devices.
If you wish to dispose of your devices, please proceed as follows:
In Germany
In order to arrange return of the electrical equipment, please send a
disposal request to "enretec GmbH." The following options are available
for this purpose:
● Use the ‘Returning an electrical device’ button under the ‘eom’ menu
item on the enretec GmbH homepage (www.enretec.de).
● Alternatively, you can also contact enretec GmbH directly.
enretec GmbH
Kanalstraße 17
16727 Velten
Tel: +49 3304 3919-500
Email: [email protected]
In accordance with the national disposal regulations regarding old
electrical and electronic devices (ElektroG), as the manufacturer, we
assume the costs for disposing of the electrical and electronic devices
in question. Disassembly, transport, and packaging costs shall be borne
by the owner/operator.
Prior to disassembly/disposal of the unit, it must be reprocessed
professionally (cleaned/disinfected).
If your unit is not permanently installed, it will be collected from the
practice. If it is permanently installed, it will be picked up curbside at
your address by appointment.
Other countries
For country-specific information on disposal, contact your local dental
dealers.

67 46 429 D3708
D3708.201.04.02.02 10.2020 153
We reserve the right to make any alterations which may be required due to technical improvements.

© Sirona Dental Systems GmbH Sprache: englisch Printed in Germany


D3708.201.04.02.02 10.2020 Ä.-Nr.: 129 660

Sirona Dental Systems GmbH

Fabrikstr. 31 Order No 67 46 429 D3708


64625 Bensheim
Germany
www.dentsplysirona.com