Early Streamer Emission Lightning Tests
Early Streamer Emission Lightning Tests
[Link]
K.J. Corn ick
D.C. Faircloth
[Link]
Indexing terms: Early streamer emission principle, Lightning protection, Simulation testing
IEE Proc -Sei Meus Technol, Vol 145, No 5, September 1998 201
For tests with the independently energised Franklin decided to apply a steady voltage to the sphere suffi-
rod air termination (the 'active Franklin rod') a small cient to cause a small visible positive corona at the ter-
impulse generator, providing voltage pulses up to mination. This voltage was -150kV.
+40kV was used (omitted for clarity in Fig. 1). This The first test was carried out without the steady field.
generator could be triggered at any desired time within The 50% sparkover voltage at this condition, was
the impulse provided by the main generator by opera- 491kV. For further tests with the Franklin rod the
tion of a delay circuit linking the two through fibre- peak impulse voltage was then increased to -725 kV, so
optic cables. The sequence is shown in Fig. 2. The that breakdown on the front of the wave could be
small generator was protected, in the event of sparko- assured, thus simulating the rise in field experienced by
ver to the rod, by a small parallel spark gap and series the termination as a downward leader approaches. The
resistor of 8kQ. degree of approximation is shown in Fig. 3 where com-
t=O
parison is made between the field strength at ground
level (disregarding the termination) under the labora-
+II +I tory impulse and that calculated according to Berger [4]
a,
I for the case of the leader of a 5kA lightning stroke.
0) I delay With this impulse, the mean time to breakdown of the
-
c
I 1
0
I
I
passive Franklin rod gap was 3 1 0 where,~ as Fig. 3
' impulse applied to rod shows, the laboratory and calculated leader fields are
1.2/50pswave)
reasonably close.
I
0
I
I f time
I
I
I
U b!
---___----
-I time to breakdown
0 40 80 120 160 200 240 280 320
Fig. 2 Sequence of voltage applications at sphere-rod gap
time, ps
Fi . 3 Comparison between calculatedfield due to an a proaching leader
The tests with the commercial ESE air termination andg$eld at ground produced by an impulse voltage reac&g peak value of
-72SkVat500p
were carried out in the same arrangement, (Fig. 1) but __ impulse -725 kV
were preceded by additional tests to demonstrate its -computed, I = 5kA
mode of operation. These consisted of the measure-
ment of the amplitude and frequency of the voltage Subsequent tests were carried out with the steady
pulses, generated by the device itself when the tip was voltage of -150kV at the sphere. Here, a composite
exposed to an electric field set up by the voltage on a voltage of -535kV (that is, a superimposed impulse of
separate sphere. Fuller details of these experiments are -385kV) was sufficient to cause breakdown on the
given in Section 5. wavefront; the mean time to breakdown was 1 1 7 ~ .
In all the experiments, initiation of corona at the ter-
minations was detected by photomultiplier observation 4.2 Active Franklin rod
of the light emission. Time to sparkover, between These tests were also carried out under conditions first,
sphere and termination, was found from the potential with zero steady field and second, with the steady field
divider observation of the collapse of voltage across the provided by a voltage of -150kV at the sphere.
gap, which was correlated with photomultiplier obser-
vations.
to voltage
-
h
n
n
n
n
generator
._
c photo-multiplier
n
q-
01 +--rJ== _A. gap spacing
0 20 40 60
delay, ps
b
80 100 120 to oscilliscope
I
+ETand
Fig.5 Times to breakdown obtained as in Fig. 4, but with a steady volt-
age of -150kVat the sphere
a +lOkV peak impulse at rod oscilloscope
b +20kV peak impulse at rod
In the absence of the auxiliary impulses, the mean Fig.6 Preliminary experiment: arrangement oJ sphere, rod and associ-
time to breakdown was 1 1 7 ~When . +10kV auxiliary
ated measurements of rod voltage and corona light output
impulse was applied with delays less than S o p , there
was a reduction in time to breakdown which was nev-
ertheless small and hardly significant. At 80 p delay, An example of the voltage oscillogram is shown in
the time to breakdown was reduced from that for the Fig. 7a. Here, the tip of the rod was placed 4cm below
passive rod by about 4 0 ~ to, be compared with 7 0 p the sphere, to which a steady voltage of -29kV was
in the previous case described where the steady field in applied. The oscillogram shows that repeated negative
the gap was zero. When the +20kV auxiliary impulse charging of the rod occurred, between large, fast volt-
was applied, the result was generally similar. As in the age excursions; the cycle was approximately 25 ms, that
case with zero steady field, the dispersion in the results is, a frequency of 40 Hertz for the sphere voltage used.
was reduced to insignificance at the time delay corre- Use of a faster sweep showed that the voltage excur-
sponding to the maximum reduction in the time to sions were oscillatory as in Fig. Sa; here the oscillation
sparkover. frequency was approximately 40 kHz and the peak to
Thus, with or without the steady field, the effect of peak amplitude was found to be constant, from shot to
the auxiliary impulse voltage was to advance the spark- shot at approximately 7kV. (The variation in ampli-
IEE Pvoc.-Sei. Meas. Technol.,Vol. 145, No. 5, September 1998 203
tudes apparent in Fig. 7a was a function of the sam- Table 3: ESE device characteristics at constant voltage
pling process of the oscilloscope on this very slow of 25 kV
sweep).
Gap Charging Oscill. Oscill.
distance, cm frequency, Hz frequency, kHz amplitude, kV
4.4 2.6 35.7 6.2
4.1 6.3 36.4 6.3
3.5 13.2 35.8 6.4
3.0 18.8 35.9 6.8
b
Fig.8 Oscillograms: conditions as for Fig. 7
Horizontal scale: 20p/div
a Voltage at rod, recorded by probe arrangement, arbitrary units
b Photomultiplier record of light emitted by corona pulses
21 6.4 35.0 6.8 Table 4: Probability of corona pulse during voltage oscil-
23 11.4 36.4 6.7 lation
25 16.7 36.4 6.8 Sphere Probability of
26 19.5 37.7 6.9 voltage (kV) corona pulse
27 22.7 37.7 7.1 -2 0 0.37
29 28.4 37.6 7.0 -22 0.40
29.7 31.2 37.7 7.1 -24 0.60
-2 5 0.83
(b) by varying the sphere-tip distance at constant Gap = 3.5 cm
sphere voltage; these results are shown in Table 3.
A general consistency is evident between the charging It was found that the probability of corona occurring
frequencies obtained by the two procedures but under at the peak of the oscillation increased with the applied
204 IEE Proc -Sei Meas Technol, Vol 145, No 5, September 1998
voltage at the sphere, despite the constancy of ampli- 6 Discussion and concluding remarks
tude of this oscillation, Table 4. The reason for this
behaviour is not apparent. It may be associated with In these experiments, the passive and active Franklin
changes in the ion space charge set up around the tip, rods and the Pulsar have been compared under simu-
since the rate of change of tip voltage changes with lated lightning storm conditions. The average stresses
sphere voltage. applied to each sphere-rod gap were similar, but in the
The highest sphere voltages used in these experiments active Franklin rod case, it constituted an overvoltage,
yielded an average stress in the gap of the order whereas in the Pulsar case, it was a threshold voltage
900kVm-', which approaches values encountered in a for sparkover. Absolute rates of risle of voltage were
lightning storm. No corona could be detected below similar in each case, and a comparison between the two
-200 kVm-'. Thus, the experiments covered much of devices can therefore be made. With the active Frank-
the range of conditions under which the device could lin rod, working at approximately 40% overvoltage for
function. the gap, it was shown that independent energising of
the rod reduced the time to sparkover by between 40ps
5.2 Performance in tests under impulse and 7 0 p , compared with that for the passive Franklin
voltage rod. This reduction was obtained, however, only over a
The sphere-rod test arrangement was used again, as narrow range of times of application of the auxiliary
with the passive and active Franklin rod. The tip of the 1.2150ps impulse at the rod, which themselves extended
pulsar was 1m above the ground plane. Comparison over 7 0 p prior to the breakdown time of the passive
was again made with the characteristics of a passive rod. These results were obtained whether or not the
Franklin rod of similar length, with a hemispherical tip applied impulse, to the sphere, was preceded by a
of diameter 10". Two sets of experiments were car- steady electric field. Thus, the time of energisation of
ried out, using sphere-rod gaps of 1.0m and 1.4m. the active rod was critical in achieving an advance of
Prior to testing with the combined circuit of Fig. 1, the breakdown time. Only one time-to-peak was used
the DC voltage alone was raised from zero to deter- here; it was chosen to simulate the rate of rise of aver-
mine the condition at which the Pulsar would start to age field indicated in Fig. 3. It must be expected that a
function. It was found that it started to charge at different advance would be obtained with a different
sphere-voltages of -50 kV and -60 kV respectively for time-to-peak and it follows that othcr average stresses,
the 1.0 and 1.4m gaps and voltage pulses commenced over a range that might be expected from the leaders
at -204kV and -206kV. The voltage measuring circuit occurring in nature, would also affect the results.
was removed from the Pulsar during these tests to Comparison of the Pulsar rod with, the passive Fran-
avoid the risk of damage at sparkover. klin rod showed little difference in sparkover character-
Sparkover voltages were measured, first in the 1.0m istics, except at the longer gap used, where a possible
gap using simple negative impulse only and then in time advantage to the Pulsar of 401s was noted. The
both gaps, using pre-stress negative voltages of -250kV nature of the tests precluded observation of the timing
(I .Om gap) and -300 kV (1.4m gap) to assure repeated of the Pulsar's oscillatory voltage pulse, however, and
operation of the device. Simultaneous measurements direct comparison in this respect with the active Frank-
were made, over 50 shots in each case, of times to (a) lin rod is not possible. The preliminary tests showed an
the first corona after the start of the impulse, in the 1.0 increase in the repetition rate of this pulse with applied
m gap, (b) the breakdown in the 1.0m and 1.4m gaps. stress, but this was always less than ;!5 Hertz and abso-
The results are shown in Table 5. lute values may have been affected 'by the presence of
Table 5: Sparkover voltages, average times T, to first the measurement circuit. The pre-breakdown corona
corona and average times TBto breakdown showed a higher frequency of corona pulses than the
Franklin rod, probably due to the sharpness of the
ESE rod Gap = 1 m Franklin rod Gap = I m point. There was no clear evidence of an effect due to
VDC = 0 -250 kV 0 -250 kV the oscillatory voltage pulse on this (corona, which was
V50 = -712kV -745 -709 -724 assumed to be due to the main impulse voltage on the
T/=14p~ 9 15.5 11 sphere.
T g = 2 4 7 ~ ~ 219 219 221 Thus, the maximum reduction in time to breakdown,
TB was from 310ps to 240ps, achieved with the active
ESE rod Gap = 1.4m Franklin rod Gap = 1.4m
Franklin rod. Reference to the measured impulse wave-
VDC = 0 -250 kV 0 -250 kV form shows this to correspond to a reduction in stress
v50 = - -992 - -985 from 86% of that at peak voltage to 76%, that is, a
T, = - - - - reduction to 88% of the value for the passive rod.
TB= - 195 - 233 Direct comparison with the work of Berger [4], is not
Impulse voltage = 200/1400ys approximately possible, because only times to leader inception were
presented there.
For both rods, sparkover voltages in the 1.0m gap Differences in time to sparkover must depend on gap
were significantly increased by application of the nega- length, since the growth of the upward leader must
tive prestress voltages; times to the first corona were, determine the breakdown time. Since the maximum
on average, reduced. Times to breakdown with both time advance found in the present vvork was 70ps and
passive Franklin rod and Pulsar did not significantly since the velocity of the leader with both active and
differ from each other, either under simple impulse or passive rods can be assumed to be the same, that is,
with the prestress voltage. In the 1.4m gap, however, about 2 x IO4ms-' [ 5 ] ,it is clear that the maximum time
the average time to sparkover was 38ps less, for the advantage for leader formation found in this work is
Pulsar than for the Franklin rod, though this difference 70ps. This is to be compared with -1OOps in [5] for a
was within the respective standard deviations. gap of 9.5m.
IEE Proc -Sei Meas Technol, Vol 145, No 5, September 1998 20s
These results, and the comparison with [4, 51 indicate 8 References
that the time advantage may not depend strongly upon
the gap, and Table 5 indicates that the variation with 1 UMAN, M.A.: ‘The lightning discharge’ (Academic Press, 1987),
Chapter 3
stress is not strong. The most important result is the 2 BOUQUEGNEAU, C.: ‘Laboratory tests on some radioactive
critical nature of the timing of energisation of the rod; and corona lightning rods’. Proceedings 18th international confer-
in nature, therefore, it would be essential that a self- ence on Lightning protection, Munich, 1985, pp. 37-45
3 GRZYBOWSKIj S., LIBBY, A.L., GUMLEY, J.R., and GUM-
powered device such as the Pulsar must energise the LEY, S.J.: ‘Comparative testing of ionizing and non-ionizing air
rod at an optimum time in relation to the rate at which terminals’. Proceedings 10th international symposium on High
voltage engineering, Montreal, 1997, Vol. 5, pp. 331-334
the field due to the downward leader is increasing. This 4 BERGER, G.: ‘Determination of the inception electric field of
rate of increase is variable due to the great variation in the lightning upward leader’. Proceedings 8th international sym-
measured velocities of the approaching downward posium on High voltage engineering, Yokohama, 1993, Paper
70.02
leader [l]; however, the present work shows that the 5 ALEKSANDROV, G.N., BERGER, G., and GARY, C.: ‘New
pulsed oscillations occur at only low repetition rates investigations in the lightning protection of substations’. CIGRE,
over a wide range of fields. Further work would be Paris, 1994, Paper 23/13-14/
6 CRISTESCU, D., and GARY, C.: ‘Laboratory simulation of the
needed to show that such a device is able to energise lightning impact to the ground’. Proceedings symposium Light-
itself at a time which provides a significant time advan- ning and mountains, Chamonix, 1994
tage in leader initiation. 7 MACKERRAS, D., DARVENIZA, M., and LIEW, A.C.: ‘Criti-
cal review of claimed lightning protection of buildings by early
streamer emission air terminals’, IEE Proc Sci. Meas. TechnoL,
7 Acknowledgment 1997, 144, pp. 1-10
8 ALLEN, N.L., HUANG, C.F., CORNICK, K.J., and
GREAVES, D.A.: ‘The sparkover of air gaps under composite/
The authors thank English Heritage for their financial direct voltages’. Proceedings 10th international symposium on
support for this work. High voltage engineering, 1997, Vol. 3, pp. 157-160
206 IEE Proc.-Sei. Meas. Technol., Val 145, No. 5, September 1998