Tutorial On Polarization Resistance Technique
Tutorial On Polarization Resistance Technique
Overview of Tutorial
Summary of Polarization Resistance
Technique
Sources of Error
Overview of Tutorial
Predicting rates of general corrosion by electrochemical techniques is far easier than
predicting the risk of localized attack such as crevice corrosion or pitting. The linear
polarization technique or as it should be more properly called, the polarization resistance
technique is one such technique for rapid estimates of general corrosion. The name "linear
polarization" technique is actually a misnomer because it implies linearity at the corrosion
potential. The inverse relationship between the polarization resistance (slope of the voltage
versus current curve at the corrosion potential that this technique can estimate) and the
corrosion current (corrosion rate) exists even though the curve itself is often not linear at
the corrosion potential.
The methodology for generating the potentiodynamic sweep in the vicinity of the corrosion
potential is adequately described in a number of references, one excellent example is
ASTM G59 ”Standard Test Method for Conducting Potentiodynamic Polarization
Resistance Measurements”. This standard test method developed using a laboratory round
robin testing program provides a test circuit and a standard test protocol (430ss in sulfuric
acid) useful for determining if the electrochemical equipment is functioning properly. The
polarization resistance technique is very well established for routine use both for corrosion
prediction and corrosion monitoring. During routine use, the realistic lower limit for
corrosion rate estimation is about 0.001 mm/y (about 0.1 mpy) mostly because of
limitations in estimating Tafel slopes. Under quiescent conditions and with no outside
electrical or other interference, a maximum polarization resistance of about 106 ohm-cm2
has been suggested. Such values would translate to corrosion rates of about 10-4 mm/y or
about 0.01 mpy. Sensors are available for on-line corrosion monitoring from several
commercial suppliers.
The overall technique requires the relationship between current measured versus voltage
applied near the corrosion potential. Very often, the potential is scanned over a narrow
voltage range centered on the corrosion potential, for example from -20 mV to +20 mV
relative to the corrosion potential. The polarization resistance is determined from this curve
as:
(1)
The data are usually analyzed by assuming that the relationship between the
current and voltage in the polarization curve is given by:
(2)
(3)
No assumptions are needed to derive equation (3) beyond those for validity of
equation (2). The derivation of equation (3) is completely mathematical. The
assumption of linearity between voltage and current is not needed (F. Mansfeld
and K. B. Oldham, Corrosion Science, Vol. 11, p. 787 (1971)). Substituting
equation (3) into equation (2) enables Tafel slopes and the polarization
resistance to be extracted from a curve-fitting of equation (2) to the actual
data.
(4)
Implicit Assumptions
Using equation (3) to quantify the corrosion process and estimate corrosion rates from a
polarization curve such as that in the figure requires assumptions as summarized below.
1. The reaction rate (corrosion current) can be expressed as being proportional to the
exponential of the voltage offset from the corrosion potential for one oxidation
(anodic) and one reduction (cathodic) reaction. If this assumption is not fulfilled,
equation (1) has to be modified to account for all of the reactions that affect
corrosion. < modifications the of idea an provides (1971) 787 p. 11, Vol. Science,
Corrosion Oldham, B. K. and Mansfeld>
2. Uncompensated resistance in the electrolyte is either absent or is much smaller than
the polarization resistance. The resistance estimated by the polarization resistance
technique contains all contributions to the total resistance. For example, most
electrochemical systems involving charge transfer across an interface have both a
polarization resistance Rp and an uncompensated solution resistance Rs. The
polarization resistance as measured by this technique is equal to the sum of the
actual polarization resistance and the uncompensated solution resistance,
. The current interrupt technique can sometimes be used
effectively to correct for the uncompensated resistance. Most modern potentiostats
have that capability.
3. For the full form of equation (3) to be used, mass transfer cannot be the controlling
or rate limiting step and both the anodic and cathodic reactions must be under
activation control. Otherwise, the Tafel slope corresponding to the mass transfer
controlled process is infinite. For example, if the process is under cathodic mass
transfer control, the proportionality constant between Rp and icorr becomes ba/2.303.
The technique can still, in principle, be used to estimate corrosion rates under these
circumstances. In practice, a non-linear regression of the data versus the combined
equations (2) and (3) should result in an extremely large Tafel slope for the sub-
process under mass transfer control.
4. The corrosion potential does not lie close to the reversible potentials for the
oxidation and reduction reactions. Being 25 mV or so from the reversible potential
is often sufficient to allow equation (2) to be valid.
5. To estimate the rate of uniform corrosion from the polarization resistance, each
reacting site across the entire electrode surface is assumed to function
simultaneously as a cathode and an anode. The anodic and cathodic sub-reactions
do not occur on different sites. This assumption is implicit in Mixed Potential
Theory. In the extreme, if separated anodes and cathodes exist on the surface
corrosion would be localized on the surface (e.g. pits) and the corrosion rate
calculated using equations (2) and (3) would not be the rate of uniform corrosion.
Note that the method might be used as a sensitive detector of such corrosion if such
localized attack is severe. Success depends on how the experimental apparatus is
used and resulting curve is analyzed. That application is beyond the scope of this
discussion.
6. No additional electrochemical reactions are occurring to interfere with the current
density versus voltage curve.
Assessing how well these assumptions are fulfilled requires some knowledge of the
corrosion process. The polarization resistance technique like all electrochemical techniques
cannot be used blindly. Fulfilling the above assumptions means that using the polarization
resistance technique to estimate the corrosion rate is valid for that corrosion process. Many
practical systems are often poorly characterized so assessing how well these criteria are
fulfilled can be difficult. Some degree of imprecision must be associated with the estimated
corrosion rate under these conditions. Additional sources of error can arise when the
technique is applied in practice.
Sources of Error
Experimentally induced artifacts can sometimes cause significant errors even if the
assumptions outlined in the section Summary of Polarization Resistance Technique
underpinning the equations used to relate the current-voltage relationship to the corrosion
rate are fulfilled. A number of these artifacts also influence potentiodynamic polarization
scans. This section summarizes several of the more common issues:
The problem may be understood by picturing the surface as a simple resistor in series with
a parallel combination of a resistor and capacitor. The capacitor could represent the double
layer capacitance and the resistor in parallel with it could represent the polarization
resistance (inversely proportional to the corrosion rate). The series resistor would be the
solution resistance. The goal is for the polarization scan rate to be slow enough so that the
capacitor remains fully charged and the current/voltage relationship reflects only the
interfacial corrosion process at every potential of the scan. If not, some of the current
generated would reflect charging of the surface capacitance in addition to the corrosion
process. The measured current would then tend to be greater than the current actually
generated by the corrosion reactions. The scan would not represent the corrosion process
alone.
The question becomes, what is that proper scan rate? Though no recognized method exists
to estimate this scan rate because the capacitance and resistance could be functions of the
applied voltage, the relationship between the modulus and the frequency in an
electrochemical impedance spectrum could be used to create a conceptual method of
estimating if the chosen scan rate is reasonable (F. Mansfeld and M. Kendig, Corrosion, 37,
9(1981): p. 545). The conceptual approach uses the lowest breakpoint frequency (where the
inverse of the phase angle passes through a maximum) of the impedance spectrum as the
starting point. The premise is that the scan rate (rate of change of voltage) can be related to
a frequency at every applied potential. That frequency must be low enough so that the
impedance magnitude becomes independent of frequency. There the polarization or charge
transfer resistance is being measured with no interference from surface capacitance. This
figure shows these relationships.
The break-point frequency, the point at which the inverse of the phase angle goes through a
maximum and the impedance magnitude goes through its inflection point, can be calculated
easily (F. Mansfeld, Corrosion, 37, 6(1981): p. 301) for the case of a surface that can be
modeled as a parallel combination of a resistor (polarization resistance) and capacitor
(double layer capacitance) in series with a resistor (uncompensated resistance). That
frequency is about at the position of the middle arrowhead in the figure. This frequency is
not the required frequency because the impedance magnitude is still increasing as
frequency (scan rate) decreases. The frequency at which the impedance magnitude does not
change, i.e. the frequency below which there is no capacitive contribution, is about an order
of magnitude lower than the break-point frequency. The frequency can be converted to a
scan rate by assuming that over some small voltage amplitude, e.g. ±5 mV, the voltage-
current relationship is linear and the linear range corresponds to half of a sinusoidal wave.
The table below shows estimated maximum scan rates for several polarization resistances,
solution resistances, and capacitances.
The estimates are very conservative and are meant for illustrative purposes only. One
possible way to overcome this issue is to operate the potentiostat at a very low scan rate for
all materials, e.g. 0.1 mV/s as the polarization resistance limit for practical use of this
technique is probably greater than 106ohm-cm2. Even at that scan rate, scanning over a
cycle of 40 mV requires less than 5 minutes.
Sources of Error
Uncompensated Solution Resistance
The resistance calculated from
(3)
is the sum of the actual polarization resistance and the uncompensated resistance between
the sensing point of the reference electrode and the working electrode. The lower the
conductivity of the solution, the greater is the uncompensated resistance and the greater is
the chance of possible error in the estimated polarization resistance. The slope of the
polarization curve at the corrosion potential as plotted in this figure would make the
estimated Rp too large, the estimated corrosion rate too small. Often, the resistance
estimated at high frequency (e.g. several thousand hertz) by electrochemical impedance
spectroscopy can be used as the solution resistance. That number would be subtracted from
the measured polarization resistance to provide the "true" polarization resistance.
This figure demonstrates how the unmeasured (uncompensated) voltage drop might vary
with conductivity for different current densities (assuming that losses in wiring are
minimal). The estimate was made by assuming a distance of 0.5 cm (5 mm) between the
working electrode surface and the point in the solution sensed by the reference electrode.
The resistance is approximately proportional to the distance between the electrode and the
sensing point. In the absence of a voltage ramp, the actual voltage at the fluid side of the
corroding electrode surface (relative to the reference electrode) would be the voltage set on
the potentiostat minus the uncompensated voltage drop between the reference electrode
sensing point and the corroding electrode. Though this figure is provided for illustrative
purposes only, it does show that the uncompensated voltage drop can be very large.
The effect that the uncompensated resistance can have on the effective potential (as
opposed to the potential believed to be applied by the potentiostat) and the effective scan
rate (as opposed to the scan rate believed to be applied by the potentiostat) has been
analyzed mathematically and reported in the literature (F. Mansfeld, Corrosion, 38,
10(1982): p. 556 and K. Schwabe, W. Oelssner, and H. D. Suschke, Prot. Metals, 15(1979):
p. 126). In summary, if the uncompensated voltage drop becomes significant, the applied
potential can be much greater than the voltage that is actually affecting the corrosion
processes. In addition, the applied scan rate can be much greater than the effective scan
rate. More importantly, the differences will be a function of the magnitude of the current
passed between the working and counter electrodes, becoming greater as the current
increases.
Sources of Error
the second derivative is not zero at the corrosion potential so curvature of the polarization
curve might be expected at that point. The amount of curvature would depend on icorr which
itself depends on the Tafel slopes and polarization resistance. Invoking the assumption of
linearity where the curve is actually non-linear has been estimated to result in errors as high
as 50% from this source alone. Such errors may be acceptable because corrosion rates
estimated from mass loss can also be in error by 100%. During screening, rates differing by
a factor of two or three may often be considered to be the same. Then, linearity may be
assumed for simplicity. If more accuracy is needed, account should be taken of the full non-
linearity in the polarization curve near the corrosion potential. Curve-fitting by non-linear
regression of the data against an equation such as
(2)
Sources of Error
Sources of Error
An additional source of error could occur when the slight polarization required by this
technique upsets the electrochemical processes enough so that the generated curve does not
pass through the point (0,0). That is, an applied current is observed at 0 volts relative to the
original corrosion potential. This phenomenon seems to be more prevalent in more passive
systems or when corrosion rates are very low. The example in the section Example (Low
Corrosion Rate Environment) in this tutorial suggests how this issue might be handled to
obtain a reasonable curve fit. But, corrosion rates estimated from mass loss and polarization
resistance, both at steady state, might still be expected to differ.
Sources of Error
Non-Uniform Current and Potential Distributions
The Wagner number W is useful for qualitatively predicting if a current distribution is
uniform or non-uniform. The parameter W is dimensionless and is given by
(5)
where κ is the conductivity, L is a characteristic length, V is the voltage, and i is the current
density. The derivative is a partial derivative. This number can be considered as the ratio of
the resistance to electron transfer across the interface to the resistance of the solution. For
practical purposes, equation (5) can be represented by
(6)
Sources of Error
Example of Use of the Technique (Nickel in Strong Acid)
The polarization resistance method was used to estimate the corrosion rate of nickel in
strongly acidic phosphoric/phosphorous acid solutions, the total acid content being 50 wt%.
The procedure was to scan the potential between -20 mV and +20 mV at 0.1 mV/s after
about 1 hour, after about 4 hours, and finally after 24 hours of exposure. The corrosion
potential was stable during the generation of replicates at each time period. In several
solutions, the potential did change by about 50 mV over the 24 hour period. The
requirement of a stable corrosion potential was met fairly well under all conditions.
This figure shows a plot of typical polarization curve, the data points being the measured
points and the curve being the result of curve-fitting by non-linear regression the equations
(2)
and
(3)
The symbols are explained in the section entitled Summary of Polarization Resistance
Technique in this tutorial. Note that the measured points have curvature even at the
corrosion potential. A more detailed discussion is provided in Silverman, D. C., "Practical
Corrosion Prediction Using Electrochemical Techniques", Uhlig's Corrosion Handbook,
2nd ed., ch. 68, p. 1179 (2000) (1800k).
The following table shows the calculated Tafel slopes and the corrosion rates estimated
from the regression analysis and from mass loss. The Tafel slopes and polarization
resistance values were averaged across the runs over the 24 hour period because the
corrosion rate did not change over that period. The error shown is the standard deviation.
Corrosion
# (mV) (mV) (ohm-cm2 Mass Loss
Current
The difference between corrosion rates in the last two cases touches on several artifacts
which can contribute to errors. First, the solution resistance was about 1 to 2 ohm-cm2 as
measured by electrochemical impedance spectroscopy at 5000 Hz. The error introduced by
ignoring solution resistance is very small in the first two cases but may account for at least
10% of the value in the last two. Second, though the standard deviation in the measurement
of the polarization resistance is only about 20% of the average value, the difference
between the extremes in the polarization resistance (measurement + standard deviation) -
(measurement - standard deviation) ) is about 100% for those two cases. Seemingly small
errors in relatively small polarization resistance values can lead to large errors in estimated
corrosion rates. Third, the corrosion potential for those two cases was about -250 mV
(SCE), close to the reversible potential for hydrogen under these extremely acidic
conditions. Hydrogen evolution in the form of bubbles was observed further suggesting that
the corrosion potential might have been close to the reversible potential for the hydrogen
evolution reaction in this system. Such proximity could have led to errors in the calculated
values because the assumption behind the Butler-Volmer equation that only one irreversible
anodic and one irreversible cathodic reaction are present may have been violated.
Cocoyl glutamate was discovered to prevent staining of magnesium and aluminum alloys
under alkaline conditions. The question was how this staining was being prevented,
corrosion inhibition vs. removal of corrosion products (e.g. some corrosion acceleration).
Solutions of various concentrations of cocoyl glutamate were formulated at a pH of 9.5.
Corrosion of aluminum 7075-T6 and magnesium AZ31B were measured in these solutions
by generating polarization resistance curves as a function of time over 24 hours and
exposing these alloys to the test environments for 30 days to obtain corrosion rates by mass
loss.
The polarization resistance curves were analyzed by non-linear regression of the equations
(2) and (3) against the data.
(2)
and
(3)
(4)
The estimate was made in this case assuming a value of 0.025V for both
aluminum and magnesium. The corrosion rate of magnesium was estimated to
be in the range of 0.5 to 1 mpy (0.01 to 0.02 mm/y). The corrosion rate of
aluminum was estimated to be less than 0.1 mpy (<0.0025 mm/y). Mass loss
results suggested somewhat higher corrosion rates for magnesium in the range
of 2 to 3 mpy. The aluminum corrosion rates from mass loss were less than 0.1
mpy (<0.0025 mm/y). The results suggested that the cleaning of magnesium
was by removal of stains by corrosion of discoloration products while the
cleaning of aluminum was through relatively complete passivation of
aluminum. Further discussion of the reasons for this conclusion can be found in
D. C. Silverman and T. K. Hirzel, "Divergent Effects of N-Acyl Glutamates on
Corrosion of Aluminum and Magnesium Alloys", Corrosion, Vol. 58, p.99 (2002)
1
(227k).
Note that the agreement of corrosion rates for magnesium between techniques
is good but not great. Assuming that the choice of 0.025V is within a factor of
two of the actual value, the discussion elsewhere in this tutorial suggests that
the discrepancy could be an artifact of the procedure for generating and
analyzing the polarization resistance scans especially for alloys exhibiting very
low corrosion rates. This figure shows a typical scan for magnesium and this
figure shows a typical scan for aluminum, both generated after 24 hours of
exposure. At that time, the corrosion potential measured as an open circuit
potential was stable, changing at less than 1 mV per hour for either alloy.
The abscissa for each plot is the voltage relative to that voltage registering as
zero current on the polarization resistance scan not as the voltage relative to
the open circuit potential measured prior to starting the scan. The reason is
that equation (2) is derived from the fact that no current is applied at the
corrosion potential. To use equation (2) to estimate polarization resistance and
Tafel slopes requires that when V=Vcorr, iapplied = 0. The fact that Vcorr,true
measured at open circuit and Vcorr,apparent as estimated from polarization
resistance plots are slightly different means that polarization to -20 mV relative
to Vcorr,true upset the corrosion process enough that it could not recover during
generation of the scan. The process being examined electrochemically is
slightly different from the process existing under open circuit potential even
under the low polarization required by this technique. This artifact does not
negate the usefulness of the polarization resistance technique. It does indicate
that care is required when interpreting results especially for more passive alloy-
environment systems.