0% found this document useful (0 votes)
35 views1 page

Edb 6870096 GBR en

This document summarizes the specifications of an immersion flow sensor with an integrated processor. It measures flow via a calorimetric principle using a heated probe. It has adjustable sensitivity via a potentiometer and outputs flow status using a 5-wire DC connection and relay contact. It is designed to measure the flow of liquid or gaseous media with an operating range of 1-150 cm/s for water or 3-300 cm/s for oil.

Uploaded by

metuemre
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
35 views1 page

Edb 6870096 GBR en

This document summarizes the specifications of an immersion flow sensor with an integrated processor. It measures flow via a calorimetric principle using a heated probe. It has adjustable sensitivity via a potentiometer and outputs flow status using a 5-wire DC connection and relay contact. It is designed to measure the flow of liquid or gaseous media with an operating range of 1-150 cm/s for water or 3-300 cm/s for oil.

Uploaded by

metuemre
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

Flow sensor

Immersion sensor with integrated processor


FCS-G1/2A4P-VRX/24VDC

Flow sensor for liquid media

Calorimetric principle

Adjustment via potentiometer

LED band

5-wire DC, 19.228.8 VDC

Changeover contact, relay output

Cable device

Wiring diagram

Type code
Ident no.

FCS-G1/2A4P-VRX/24VDC
6870096

Operating range water


Oil operating range
Stand-by time
Switch-on time
Switch-off time
Temperature jump, response time
Temperature gradient
Medium temperature

1150cm/s
3300 cm/s
typ. 8 s (215 s)
typ. 2 s (115 s)
typ. 2 s (115 s)
max. 12 s
250 K/min
- 2080 C

Operating voltage
No-load current I0
Output function
Rated operational current
Short-circuit protection
Reverse polarity protection
AC switching voltage
DC switching voltage
Max. AC switching capacity
Max. DC switching capacity

19.228.8VDC
80 mA
Relay output, changover contact
4 A
no
yes
250 VAC
60 VDC
1000 VA
60 W

Housing material
Sensor material
Max. tightening torque housing nut
Connection
Cable length
Cable cross section

plastic, PBT
stainless steel, AISI 316Ti
100 Nm
cable
2 m

Pressure resistance
Process connection

Edition 2014-04-12T10:31:43+02:00

Switching state
Flow state display
Indication: Drop below setpoint
Indication: Setpoint reached
Indication: Setpoint exceeded
LED display

1/1

5 x 0.5 mm
100 bar
G

Functional principle
Our insertion - flow sensors operate on the
principle of thermodynamics. The measuring probe is heated by several C as against
the flow medium. When fluid moves along the
probe, the heat generated in the probe is dissipated. The resulting temperature is measured and compared to the medium temperature. The flow status of every medium can be
derived from the evaluated temperature difference. Thus TURCK's wear-free flow sensors
reliably monitor the flow of gaseous and liquid
media.

LED chain green / yellow / red


LED chain, red (1x), green (5x)
LED red
LED yellow
4 x LEDs green
red = 4 mA
1 x green > 4 mA
2 x green > 8 mA
3 x green > 12 mA
4 x green > 16 mA
5 x green = 20 mA

Hans Turck GmbH & Co.KG D-45472 Mlheim an der Ruhr Witzlebenstrae 7 Tel. 0208 4952-0 Fax 0208 4952-264 [email protected] www.turck.com