stainless steel capsule pressure gauge
mounting variations
type A bottom connection, direct mounting
type B panel mounting, clamp fixing, back connection
type C surface mounting, bottom connection
type D back connection, direct mounting
type E panel mounting, back connection
type F panel mounting, bottom connection
BDT9-18
technical gauge specifications model BDT9-18
diameter
connection
window
ranges
63/100/160 mm
1/2
glass
from 25/600 mbar
BSP/NPT
casing
capsule
movement
dial
AISI304 bayonet
AISI316
zero adjustable
aluminium
bezel
socket
pointer
accuracy
AISI304 bayonet
AISI316
adjustable slotted
1.6% FSD
Note: 63 mm standard with 1/4 BSP/NPT connection
160 mm range from 60/600 mbar
options and special executions at extra cost
casing
bezel
window
dials
certificates
manuals
connection
restrictor
connection position
special version
special ranges
special ranges
AISI316
AISI316
safety glass; plastic
customized i.e., non standard ranges or logos
calibration -; conformity -; material
installation and maintenance
METRIC; BSPT; acc JIS in various sizes
AISI316
top or side
with vacuum or compound range
0-16 mbar only in 63/100 mm
0-10 mbar only in 100 mm
For technical data sheets and dimensional drawings please refer to our website www.badotherm.com
01/01/2006 rev 2
47
pressure instrumentation
1.1.2.
Badotherm model BDT9-18 series gauges are available as
63/100/160 mm dia. The stainless steel capsule systems are available in
standard ranges from 25 mbar up to 600 mbar. The gauges are suitable for
corrosive environments and gaseous media that will attack copper alloys.
BDT9-18 - technical data sheet
pressure details
working pressure
Steady
full scale value
Fluctuating
0,9 x full scale value
pressure element
capsule system
BDT9-18
AISI316
temperature details
temperature limitations
standard
ambient
process
-40 C to 60 C
100 C
The variation of indication caused by the effects of temperature shall not exceed :
0,06 x (ambient temperature-reference temperature)%
of the span
design information
ingress protection
design
IP 54 per EN 60 529 / IEC 529
EN 837-1
drawings
type A
type D 63 mm
type D 100/160 mm
dimensions
ns
d1
weight
63
68
63
42
88
63
0,18 kg
100
110
100
49
142
82
0,53 kg
160
160
149
50
200
84
1,02 kg
Note : dimensions are in mm