Beehive Electronics
V1.1 4/28/2010
100 series probe output power calculator
100A magnetic field probe
100B magnetic field probe
Calculate output power from known flux density:
Calculate output power from known flux density:
Input the following two parameters:
Input the following two parameters:
Frequency (MHz)
Flux density (T)
10
1.00E-06
Frequency (MHz)
Flux density (T)
Resulting output:
Resulting output:
Power into 50 ohms (dBm)
Power into 50 ohms (mW)
Voltage into 50 ohms
-34.80 dBm
3.311E-04 mW
4.069E-03 V
Power into 50 ohms (dBm)
Power into 50 ohms (mW)
Voltage into 50 ohms
Calculate flux density from measured output power:
Calculate flux density from measured output power:
Input the following two parameters:
Input the following two parameters:
Frequency (MHz)
Output power (dBm)
10
-50
Frequency (MHz)
Output power (dBm)
Resulting output:
Resulting output:
Flux density, dB
Field strength, linear
-135.2 dB tesla
1.7378E-007 tesla
Flux density, dB
Field strength, linear
Enter your input data into the gray cells
The results are displayed in the light yellow cells
Examples of how to use the spreadsheet:
Example 1: Calculate flux density from measured output power.
Probe:
100A
Frequency 10 MHz
Enter '10' in cell E27
Power:
-50 dBm Enter '-50' in cell E28
Result:
dB:
Linear:
-135.2 dB tesla
1.74e-7 tesla
in cell E32
in cell E33
Example 2: Calculate probe output power
Probe:
Frequency:
Flux density:
Result:
om known flux density:
1000
1.00E-06
100C magnetic field probe
100D electric field probe
Calculate output power from known flux density:
Calculate output power from known f
Input the following two parameters:
Input the following two parameters:
Frequency (MHz)
Flux density (T)
10
1.00E-06
Resulting output:
-17.80 dBm
1.660E-02 mW
2.881E-02 V
m measured output power:
10
-57.8
Power into 50 ohms (dBm)
Power into 50 ohms (mW)
Voltage into 50 ohms
Resulting output:
-14.90 dBm
3.236E-02 mW
4.022E-02 V
Calculate field strength from measur
Input the following two parameters:
Input the following two parameters:
10
-14.9
Resulting output:
Flux density, dB
Field strength, linear
le 2: Calculate probe output power for a known flux density
100B
Frequency:
1000 MHz
Enter '1000' in cell L10
Flux density:
1e-6 tesla
Enter '1e-6' in cell L11
Power
Power into 50 ohms (dBm
Power into 50 ohms (mW
Voltage into 50 ohms
Calculate flux density from measured output power:
Frequency (MHz)
Output power (dBm)
-120 dB tesla
0.000001 tesla
Frequency (MHz)
Field strength (V/m)
-17.8 dBm in cell L16
Frequency (MHz)
Output power (dBm)
Resulting output:
-120 dB tesla
0.000001 tesla
Field strength, dB
Field strength, linear
ectric field probe
output power from known field strength:
ollowing two parameters:
Frequency (MHz)
Field strength (V/m)
Power into 50 ohms (dBm)
Power into 50 ohms (mW)
Voltage into 50 ohms
0.01
10
-133.20 dBm
4.786E-14 mW
4.892E-08 V
field strength from measured output power:
ollowing two parameters:
Frequency (MHz)
Output power (dBm)
Field strength, dB
Field strength, linear
0.01
-133.2
20 dB V/m
10 V/m