Liquisys M (Turbidity)
Liquisys M (Turbidity)
40
Liquisys M CUM 223 / 253 Transmitter for Turbidity and Solids Content
Operating Instructions
ENDRESS+HAUSER LIQUISYS S
ENDRESS+HAUSER LIQUISYS S
Endress + Hauser
The Power of Know How
Need information on the instrument? Please read the following chapters: General information
Safety
You wish to install the instrument. The required steps are described in this chapter: Installation
You wish to operate or reconfigure the instrument. The operating concept is explained in these chapters:
Operation
Instrument configuration
Interfaces
Need help with problems or maintenance? Please refer to these chapters: Maintenance and troubleshooting
% C
Accessories
Technical data
Appendix
Index
Table of contents
UM253E00.CHP
Table of contents
1.1 1.2 1.3 1.4 1.5 Symbols used . . . . . . . Storage and transport . Unpacking . . . . . . . . . . Dismantling, packaging Product structure . . . . .
/A AH= E B H =JE
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
2 2 2 2 3
Intended application . . . . . . . . General safety notes . . . . . . . . Installation, start-up, operation Monitoring and safety features Immunity to interference . . . . . Declaration of conformity . . . . .
5=BAJO E IJHK?JE
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
4 4 4 5 5 5
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . Sensor installation and measuring cable connection
1 IJ= =JE
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. 6 . 7 . 8 12 14
"
Operator interface . . . . . . . . . . . . Display . . . . . . . . . . . . . . . . . . . . . Function of the operating keys. . . Auto / manual mode of operation. Operating concept. . . . . . . . . . . .
FAH=JE
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
16 16 17 18 19
5.1 5.2 5.3 5.4 5.5 5.6 5.7 5.8 5.9 5.10 5.11 5.12 5.13
Start-up . . . . . . . . . . . . . . . . System configuration . . . . . . Current input . . . . . . . . . . . . Current outputs . . . . . . . . . . Monitoring functions. . . . . . . Relay contact configuration . Concentration measurement Service . . . . . . . . . . . . . . . . . E+H Service . . . . . . . . . . . . . Interfaces . . . . . . . . . . . . . . . Calibration . . . . . . . . . . . . . . Offset . . . . . . . . . . . . . . . . . . Slope . . . . . . . . . . . . . . . . . .
1 IJHK A J ?
BECKH=JE
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
24 24 27 30 33 37 48 49 51 52 53 61 62
$ %
7.1 7.2
1 JAHB=?AI
Troubleshooting common problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 Troubleshooting using the error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
=E JA = ?A = @ JH K> AID
JE C
&
Diagnosis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Corrective maintenance of Liquisys M CUM 223 Corrective maintenance of Liquisys M CUM 253 Spare parts orders . . . . . . . . . . . . . . . . . . . . . . . . Service aids Optoscope with Scopeware . . . Corrective maintenance of measuring system . .
,E=C
IEI = @ ? HHA?JELA
=E JA = ?A
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
69 71 74 77 77 78
'
Endress+Hauser
General information
1
1.1
General information
Symbols used
Warning: This symbol alerts to hazards which could cause serious injuries as well as damage to the equipment if ignored. Caution: This symbol alerts to possible faults which could arise from incorrect operation. They could cause damage to the equiment if ignored. Note: This symbol indicates important items of information. Input Double insulation Equipment protected by double insulation. Alarm relay
Output
1.2
The packaging material used to store or transport the transmitter must provide shock protection. Optimal protection is provided by the original packaging materials.
1.3
Unpacking
Keep the original packaging materials for future storage or shipping of the instrument. If you have any questions, consult your supplier or the Endress+Hauser sales agency in your area (see back cover of these operating instructions for addresses).
Verify that the packaging and the contents are undamaged! Inform the post office or freight carrier of any damage. Damaged merchandise must be retained until the matter has been settled. Check that the delivery is complete and agrees with the shipping documents and your order (refer to nameplate for type and version). The delivery includes: Transmitter CUM 223 (panel-mounted instrument) or CUM 253 (field device) Operating instructions BA 200C/07/en Panel-mounted instrument: 1 set of plug-in screw terminals 2 clamping screws for panel mounting
Field instrument: 1 plug-in screw terminal 1 cable gland Pg 7 1 cable gland Pg 16, reduced 2 cable glands Pg 13.5 1 x NPT adapter set (optional for CSA versions)
1.4
Package the assembly properly for reuse at a later point in time. Optimal protection is
Endress+Hauser
General information
UM253E01.CHP
1.5
Product structure
You can identify the instrument variant by the order code on the nameplate. Under Codes, you can find the release code for Software upgrade for ChemoClean (left of diagonal line) or Plus package (right of diagonal line).
ENDRESS+HAUSER
LIQUISYS M turbidity order code CUM 253-TS0115 serial no. 276944 meas. range 0 9999 ppm temperature -10 60C output 1 0/4 20 mA mains 230 VAC prot. class IP 65 codes
ENDRESS+HAUSER
LIQUISYS M turbidity order code CUM 223-TS0115 serial no. 276945 meas. range 0 9999 ppm temperature -10 60C output 1 0/4 20 mA mains 230 VAC prot. class IP 54 / IP 30 codes
253-TYP.CDR
223-TYP.CDR
Liquisys M CUM 223 / 253 Version TB Solids concentration measurement with initial setting for residual concrete water TU Turbidity and solids concentration measurement TS Turbidity and solids concentration measurement, with additional functions (Plus package) Power supply 0 Power supply 230 V AC 1 Power supply 115 V AC 2 Power supply 230 V AC, CSA Gen. Purp. 3 Power supply 115 V AC, CSA Gen. Purp. 5 Power supply 100 AC 7 Power supply 24 AC, CSA Gen. Purp. 8 Power supply 24 V AC/DC Measurement output 0 1 output signal: Turbidity / SS 1 2 output signals: Turbidity / SS and temp. / turbidity or SS / set value 3 1 output signal: PROFIBUS-PA 4 1 output signal: PROFIBUS-DP 5 1 output signal: Turbidity / DM with HART 6 2 output signals: Turbidity / SS with HARTand temperature Contacts 05 No additional contacts 10 2 contacts (limits / PID / timer) 15 4 contacts (limits / PID / timer / ChemoClean) 16 4 contacts (limits / PID / timer) 20 2 contacts with current input (limits / PID / timer) 25 4 contacts with cleaning, current input (limit / PID / ChemoClean) 26 4 contacts with timer, current input (limits / PID / timer) Additional equipment RM1 Initial setting for residual concrete water CUM223CUM253complete order code
Additional functions of the Plus package (TB / TS ) Automatic start of cleaning function, Concentration measurement, function Field F8 group K
Endress+Hauser
Safety instructions
2
2.1
Safety instructions
Intended application
Chemicals Pharmaceuticals Drinking water treatment Condensate treatment Municipal sewage treatment plants Water conditioning
Liquisys M CUM 223 / 253 is a field-tested and reliable transmitter to determine the turbidity and the solids concentration of fluid media. The Liquisys M CUM 223 / 253 is particularly suitable for use in the following areas:
2.2
This device has been manufactured for safe operation according to the state of the art in engineering and conforms to the applicable regulations and European standards (see Technical data). It has been designed according to EN 61010-1 and has left the manufacturers works in perfect condition. However, if used improperly or for purposes other than the intended purpose, it may be dangerous, e.g. due to incorrect connection.
2.3
Endress+Hauser
Safety instructions
UM253E02.CHP
2.4
Safety features The transmitter is protected against external influences and damage by the following design measures: Rugged housing Degree of protection provided by enclosure: IP 65 (CUM 253) UV resistance
2.5
Immunity to interference
Warning: The specified immunity to interference only applies for devices connected as outlined in these operating instructions.
This instrument has been tested according to the applicable European standards for industrial applications with regard to electromagnetic compatibility. It is carefully protected against electromagnetic interference by constructive design measures.
2.6
Declaration of conformity
The transmitter CUM 223/253 has been developed and manufactured in accordance with currently valid European standards and directives.. E+H certifies the compliance with the standards by using the 4 sign.
Endress+Hauser
Installation
Installation
The following procedure should be followed for a complete measuring system installation: Installation or attachment transmitter (see chap. 3.3) Selection and connection of cable and sensor (see chap. 3.4, 3.5 and 9) Installation is followed by start-up (see chap. 5).
3.1
Measuring system
Optional: Extension cable CYK 8 Junction box VBM or RM
The complete measuring system comprises: The Liquisys M CUM 223 or CUM 253 transmitter A sensor with integrated temperature sensor and a permanently attachedcable An installation assembly
ENDRESS+HAUSER LIQUISYS S
REL2
Sensor cable
REL1
Fig. 3.1
Complete measuring system consisting of Liquisys M CUM 223 / 253 with measuring cable, assembly and turbidity sensor
REL2
Endress+Hauser
Installation
UM253E03.CHP
3.2
Dimensions
Pg 13.5 Pg 7 Pg 16
11
ENDRESS+HAUSER LIQUISYS S
REL2
157
115
70 154 170
247
70
FELD1.CDR
Fig. 3.2
Note: There is a hole in the punching for Pg 16 cable entry. It serves as a pressure balance during air freight dispatching. Make sure that there is no moisture penetrating into the housing before cable installation. After cable installation, the housing is completely tight.
4 2 Inside of housing of Liquisys M CUM 253 1 Removable electronics box 2 Partition plate 3 Terminal blocks 4 Fuse
FELD2.CDR
Fig. 3.3
Endress+Hauser
Installation
90
Fig. 3.4
139
96 6
EINBAU2.CDR
3.3
3.3.1
Mounting
Field instrument
Weather protection cover CYY 101 can be used for outdoor installation in conjunction with all mounting variants. Note: If installed outdoors, weather protection cover required.
Several mounting options are available for the Liquisys M in the fieldinstrument version: Post mounting on cylindrical pipes Post mounting on a square post Wall mounting with fastening screws
Weather protection cover CYY 101 Weather protection cover for outdoors installation, to be mounted on field instrument; material: stainless steel SS 304 Order no: CYY 101-A For post or pipe installation, the post mounting kit is additionally required.
270
300
Fig. 3.5
320
DACH.CDR
Endress+Hauser
Installation
UM253E03.CHP
Post mounting kit Mounting kit for installation of field housing on horizontal and vertical pipes (max. 60 mm); also aids fixing of weather protection cover. Material: stainless steel SS 304 Order no.: 50086842
Universal mounting post CYY 102 Square tube for mounting of measuring transmitters; material: stainless steel SS 304 Order no: CYY 102-A
90 70
max. 60
1495
60 x 60
M6 90
6 70 90
60
150
Fig. 3.6
Endress+Hauser
Installation
3.3.2
Mounting examples
Fig. 3.7
MONT10.CDR
left: Wall mounting right: Mounting with universal post and weather protection cover
Fig. 3.8
MONT2.CDR
10
Endress+Hauser
Installation
UM253E03.CHP
3.3.3
Panel-mounted instrument
The instrument is attached using the supplied tensioning screws (see Fig. 3.9). The required overall installation depth is approx. 165 mm.
92+0.5
92+0.5
1 2
92
3 3 96
90
3 57
Fig. 3.9
Endress+Hauser
11
Installation
3.4
Electrical connection
Connection diagram The connection diagram depicted in Fig. 3.10 shows the instrument connections for turbidity sensors CUS 31 or CUS 41. The connection of the varrious sensors is shown in more detail in Figs. 3.13 and 3.14.
S BN 87
41 42
WH 88
43
Sensor
GN 96
47 48
YE
97
49
31
57
mA
TU
58 59
32
33
51
mA
52 53
34
81
54 55
Relay 3 (current-free contact postion) for Chemoclean: Water Relay 4 (current-free contact postion) for Chemoclean: Cleaner Current input 4 ... 20 mA
10-50 V
82
56
93
23 + 24 -
10-50 V
94 mA
85
15 V
L1 L+ N L
86
Mains
Fig. 3.10
ANU01.CDR
Note:
12
Endress+Hauser
Installation
UM253E03.CHP
Note: Please label the sensor terminal block with the enclosed sticker.
Caution!
Terminal designated as NC
may not be switched.
Undesignated terminals
may not be switched. Connections of field instrument For connection, the measuring cables are introduced through the cable glands on the field instrument and connected according to the diagrams infigures 3.10 and 3.11.
SENSOR
REL 4
REL 3
REL 2
REL 1
WH YE
B NC 88 97 NC
54 55 56 51 52 53 57 58 59 47 48 49 41 42 43 Mains Hilfsenergie NC
NC 31 32 33 34 NC 85 86 93 94 81 82 23 24 NC + - + + - + - + - +
L1 N (AC) + - (DC)
4..20mA
Digital 2
Digital 1
Temp. (opt.)
131084-4C 49 47 COM 48 59
+15V 10mA
REL1
YE GN WH BN
REL2
57 COM 58 53
REL3
51 COM 52 56
REL4
54 COM 55 24
4..20mA + 23
Mains
TU
131096-4F
131096F.CDR
Fig. 3.11
Connection compartment sticker for field instruments CUM 253; Instrument status: de-energised
131083-4C
131083C.CDR
Fig. 3.12
Endress+Hauser
13
Installation
3.5
Measuring cable connection The terminal block is located in a separate connection compartment which is accessible by opening the plastic cover. Remove the pre-pressed knock-outs for cable entry. The sensor is connected via a multiple-core, special screened cable which is completely terminated and permanently connected. Use a junction box, e.g. VBM to extend the measuring cable. Termination instructions are supplied with the measuring cables. Note:
Measuring cable requirements Sensor type Turbidity sensor CUS 31, CUS 41 Cable Non-detachable cable on sensor Extension VBM box + CYK 8
Sensor
Screen GN Com A YE Com B
CYK 8
96
97
WH UB
+UB
88
Fig. 3.13
BN
+UB
87
U_AN03E.CDR
14
Endress+Hauser
Installation
UM253E03.CHP
Connection examples
87
BN
88
WH
96
GN
97
YE
Cable CYK 8
Sensor
U2X3AN.CDR
Fig. 3.14
Endress+Hauser
15
Operation
4
4.1
Operation
Operator interface
ENDRESS+HAUSER LIQUISYS M
Liquid crystal display for display measured values and configurations data
185
NTU R262
Display of active contact and key for relay switchover in manual mode
Setpoint
Field for user labelling LED indicator for alarm function
4 main operating keys for calibration and instrument configuration Operating elements of Liquisys M Key for switching between automatic / manual operation
Fig. 4.1
DISPLAY1E.CDR
4.2
Display
LED Display
Indication of current operating mode Auto (green LED) or Manual (yellow LED)
Indicates the state of relays 1 and 2 LED green: measured value within permissible limits, relay inactive LED red: measured value outside permissible limits, relay active
Alarm indication e.g. for continuous limit violation, temperature sensor failure or system error (see error list chap. 7)
Sensor symbol
185
NTU R262
Setpoint
Error indicator Fig. 4.2 Liquid crystal display In measuring mode: Secondary measured value. In setup / calibr. mode e.g. setting value
In measuring mode: Main measured value. In setup / calibr. mode e.g. parameter
DISPLAY3.CDR
16
Endress+Hauser
Operation
UM253ED4.CHP
4.3
Unlocking the keypad Press the CAL and MINUS keys simultaneously for minimum 3s to unlock the keypad. The code prompt displays the code 0:
Endress+Hauser
17
Operation
4.4
Select relay or wiper function. Press the REL key to toggle between the relays or the wiper. The display shows the selected relay and the switching status (ON / OFF) in the second line. In maual mode, the measuring value is continuously displayed (e.g. for monitoring during dosage). Set the relays or the wiper. Switch on with PLUS, switch off with MINUS. The relay or wiper state remains in effects until it is actively reset.
Press AUTO key for returning to the measuring mode. All relais are controlled by the transmitter again.
Note:
18
Endress+Hauser
Operation
UM253ED4.CHP
4.5
4.5.1
Operating concept
Operating modes
Code
The function groups are selected with the PLUS or MINUS keys
SET-MOD.CDR
Fig. 4.3
Note:
Endress+Hauser
19
Operation
4.5.2
Access codes
All instrument access codes are fixed, i.e. they cannot be modified. When the instrument requests the access codes, it recognises the difference between codes (cf. Fig. 5.3): CAL key + Code 22: Access to Calibration and Offset menus. ENTER key + Code 22: Access to the Configuration menus, allowing configuration and user-specific settings. PLUS + ENTER keys + Code 9999: Locks the keypad. CAL + MINUS keys + Code 0: Unlocks the keypad. CAL or ENTER key + any Code: Access to Read mode, i.e. all settings can be read but not changed.
4.5.3
Menu structure
Note:
The configuration and calibration functions are arranged in a menu structure by function groups. The function groups are selected in the setup mode with the PLUS and MINUS keys. The ENTER key is used to move from one function to the next within a function group. The PLUS and MINUS keys are used for option selection and editing. Selections must be confirmed by pressing the ENTER key. This also moves the cursor to the next function. Pressing the PLUS and MINUS keys at the same time terminates programming (return to main menu).
Function groups Move forward and backward with the PLUS and MINUS keys
Move from onee function to the next with the ENTER key
E E E E
Fig. 4.4
BEDSCHE.CDR
20
Endress+Hauser
Operation
UM253ED4.CHP
4.5.4
The current output is frozen in the setup mode and during calibration, i.e. the last current value is constantly output. HOLD appears on the display. In case of steady control (4... 20 mA) on current output 2, it is set to 0/4 mA during Hold. Note:
Endress+Hauser
21
Instrument configuration
Instrument configuration
Setup mode SETUP 1 (A) see chap. 5.2.1 SETUP 2 (B) see chap. 5.2.2 CURRENT INPUT (Z) see chap. CURRENT OUTPUT (O) see chap. 5.3 ALARM (F) see chap. 5.5.1 CHECK (P) see chap. 5.5.2 RELAY (R) see chap. 5.5 CONCENTRATION (K) see chap. 5.6 SERVICE (S) see chap. 5.7 E+H SERVICE (E) see chap. 5.8 INTERFACE (I) see chap. 5.9
After power-up the instrument performs a self-test and then enters the measuring mode. Now it can be configured and calibrated for the first time. The values set by the user are kept even in the event of a power failure. The following function groups are available on the Liquisys M (the groups that are only available in the Plus package are marked accordingly in the functional descriptions):
Calibration and offset mode CALIBRATION (C) see chap. 5.10 OFFSET (V) see chap. 5.11 SLOPE (N) see chap.
Function coding. The displayed code indicates the function position in the function group.
2
Fig. 5.1 Example of display in setup mode
s F3
Err.Delay
MATRIX.CDR
Additional information
DISPLAY4.CDR
A2(3) A2(2) A
Fig. 5.2 Function coding
A231 A221
A232
A1
A2(1)
Selecting and locating functions is facilitated by a code displayed for each function in a special display field. The structure of this coding is given in Fig. 5.2. The first column indicates the function group as a letter (see group designations). The functions in the individual groups are counted from the top to the bottom and from the left to the right.
22
Endress+Hauser
Instrument configuration
UM253E5A.CHP
Factory settings When the instrument is switched on for the first time, the factory settings are in effect. The following table provides an overview of all major settings. Please refer to the descriptions of the individual functions in chapter 5 for all other factory settings (the factory setting is printed in bold type). Function Type of measurement Temperature offset / turbidity offset Limit 1 Contact function of limit contactor 1 Limit 2 Contact function of limit contactor 2 Current outputs 1 and 2* Current output 1: measured value for 4 mA signal current Current output 1: measured value for 20 mA signal current Current output 2: temperature value at 4 mA signal current* Current output 2: temperature value for 20 mA signal current* Measured value damping Calibration data set Wiper controller * if equipped accpordingly Factory setting Turbidity in FNU, temperature in C 0 C 0 FNU 9999 FNU MAX contact without delay 9999 FNU MAX contact without delay 4 ... 20 mA 0 FNU 10.0 FNU 5.0 C 100.0 C 10 no. 3 Off
Alarm contacts
Alarm state
41 42 43
41 42 43
RELAIS2.CDR
Fig. 5.3
Normal operating state: Instrument in operation No error message available (Alarm LED off) Relay picked up Contact 42/43 closed
Alarm state: Error message available (Alarm LED red) or Device defective or voltage-free (Alarm LED off) Relay dropped out Contact 41/42 closed
Endress+Hauser
23
Instrument configuration
5.1
Start-up
After switching the instrument on, make the following settings to the specified function groups: Function group SERVICE (S) S1: Select language and exit function group (not required for English). Function group SETUP 1 (A) Adjust all parameter in this group, see chap. 5.2.1. Function group SETUP 2 (B) Make all settings in this group, see chap. 5.2.2. Other configuration options are explained in the chapters to follow for each menu.
5.2
System configuration
The system is configured using the function groups SETUP 1 and SETUP 2. Here, select the measurement type, the sensor and the settings for the temperature measurement. All the parameters in these two function groups are to be configured to avoid measuring errors or failure to measure at all.
5.2.1
Setup 1 (Turbidity)
For access to the SETUP menu, please enter Code 22. Selection or range Factory setting (bold)
Coding
Field
Display
Info
SETUP
NTU FNU ppm mg/l g/l % spec.
A1
NTU A1
Oper.Mode
Any change in operating mode causes an automatic reset of userdefined settings. The offsets for turbidity and temperature are reset to zero.
A2
kg/l A2
Conc.Unit
A3
XX.xx A3
Format
A4
CUS31 CUS41
CUS31 A4
Sensor
Factory settings are printed in bold face; base version does not include functions in italic.
24
Endress+Hauser
Instrument configuration
UM253E5A.CHP
Coding
Field
Display
Info Measured value damping causes averaging over the specified number of individual measured values.It is used, for example, to stabilise the display with applications that fluctuate a great deal. There is no damping if 1 is entered.
A5
10 1 ... 60
10 A5
Damping
5.2.2
Setup 2 (Temperature)
Field Selection or range Factory setting (bold) Display Info
Coding
SETUP 2
off on auto
B1
off B1
wiper
If auto is selected, the wiper is operated in combination with a cleaning function via timer / Chemoclean (wipe and clean). In this case B2 and B3 are not applicable.
B2
30 s 3 ... 999 s
30 B2
CleanTime
B3
120 B3
min
PauseTime
B4
3 1 ... 3
3 B4
data set
There are 3 calibration data sets stored in every operation mode (A1). Data set 1 cannot be changed (see chap. 5.10). Data set 1 cannot be changed (factory setting). However, it can be used as a basic for a customer calibration data set. To work with the copy of the data set, select the data set in field B4.
B5
no B5
copy data
The factory settings are printed in bold; functions printed in italics are not available in the basic version.
Endress+Hauser
25
Instrument configuration
Coding
Display
Info With the CUS 31 / 41: displays the measured value with or without assembly adaptation (reflection). Only effective in NTU, FNU, ppm, mg/l ranges).
B6
yes no
yes B6
refl.adj.
B7
0.0 B7
RealTemp
This entry can be used to calibrate the temperature sensor to an external measurement.
B8
0.0 B8
Temp.Offs
The difference between the temperatures measured and entered is displayed. Compensates slow gas bubble formation, which may arise from small amounts of dissolved gas in the medium. 0.1%= no formation of gas bubbles 100%= strong tendency to gas bubble formation
B9
3.0 B9
GasBubble
Factory settings are printed in bold face; base version does not include functions in italic.
26
Endress+Hauser
Instrument configuration
UM253E5A.CHP
5.3
Current input
This function group offers two independent application solutions, provided that the current output of an external measured quantity, e.g. flow meter, is connected to the 4 20 mA input of Liquisys M CUM 223 / 253. The following assignments then apply:
Flow in main stream Lower range limit current input Upper range limit current input Lower setting value flow meter Upper setting value flow meter
Current signal in mA 4
20
100
5.3.1
This arrangement is highly practical when the sample stream flowing through the flow assembly is totally independent of the flow rate in the main stream. This permits the signalling of an alarm state in the main stream (flow rate too low or totally stopped) and trigger a dosing switch-off, even if the measuring water stream is retained due to the installation configuration.
ON Flow in main stream OFF ON Relay contacts of PID controller see field Z2 see field Z3 OFF
ON Alarm relay OFF Flow alarm Flow below switch-off limit Z4 or flow failure Flow restored
Alarm signalling and dosing switch-off by the main stream
Fig. 5.4
Endress+Hauser
27
Instrument configuration
5.3.2
In processes with very short response times it may be practical to apply the flow rate to the controller, if the flow rate fluctuates, in order to optimise the control process.
Static mixer
Injection points
Flow meter
0.42
25.0 C
mg/l
REL2
Reagents
Arrangement example for feedforward control of the flow rate in the main stream to the PID controller(s)
4 ... 20 mA
Fig. 5.5
Feedforward control is a multiplying function as depicted in the below figure (factory setting as example):
Gain Kinflu
1.5
0.5
0
Multiplying feedforward control
20
40
Z7
60
80
Fig. 5.6
28
Endress+Hauser
Instrument configuration
UM253E5A.CHP
Coding
Field
Display
Info
CUR.INPUT
Off Z1 Cont.stop 0 Z2 Delay 0 Z3 Delay
% s s
Z1
Select flow rate monitoring of main stream (with controller switch-off) Enter delay for controller switch-off by current input
Off Input
Only switch on when flow meter is connected in main stream. When Z1 = Off, fields Z2 to Z5 do not exist. Short-term flow rate undershots can be suppressed by delay and will not cause controller switch-off.
Z2
0s 0 ... 2000 s
Off
Z3
0s 0 ... 2000 s
On
Z4
0 100% corresponds to 4 20 mA at current input. Note the measured value allocation to the current output of the flow meter.
Z5
Low High
If the value entered in Z4 is exceeded low or high, the controller switches off. When Z6 = Off, Z7 does not exist.
Z6
Basic = Feedforward control only affects the basic load (alternatively dosage in proportion to quantity, if common PID control is not possible, e.g. due to sensor defect) When the value is set, the controller manipulated value with feedforward control on is identical to feedforward control off.
Z7
50 Z7 Kinflu=1
Endress+Hauser
29
Instrument configuration
5.4
Current outputs
The function group CURRENT OUTPUT is used to configure the individual outputs. Either a linear (O3 (1)) or, in conjunction with the Plus package, a user-defined current output characteristic (O3 (3)) can be entered. Furthermore, a current output value can be simulated to check the current outputs (O3 ( 2)). The controller set value in field R 237 can be output via current output 2, if available.
FNU 9000
6000
3000
signal
Fig. 5.7
The distance signal between two table value pairs must exceed: 0.005 FNU / NTU / ppm / mg/l / % 0.05 g/l Temperature: 0.25 C First enter the current output configuration you require in the following blank table. Ensure the required minimum distance by calculating the resulting signal distance per mA. Then enter the result in the instrument. Current output 1 Value pair 1 2 3 4 5 6 7 8 9 10 Tu / C [ ] Current [mA] Distance per mA Tu / C [ ] Current output 2 Current [mA] Distance per mA
30
Endress+Hauser
Instrument configuration
UM253E5A.CHP
Coding
Field
Display
Info
OUTPUT
O1
Out1 Out2
Out1 O1
Sel. Out
O2
C mg/l Contr
C O2 Sel. Out2
Selection of Curr (= current output 2) in field R237 is only possible, if field O2 = Contr is selected. The characteristic can have a positive or negative slope at the measured value output.
O3 (1)
lin
O3
Sel.Type
At set value output (O2 = Contr), the increasing current corresponds to an increasing set value.
O311
420 mA 020 mA
4-20 O311
Sel.Range
O312
0.0 NTU 0.0 FNU 0.0 ppm 0.0 mg/l 0.0 g/l 0.0 kg/l 0.0 t/m3 0.0 % 0.0 C 10.00 NTU 10.00 FNU 10.00 ppm 10.00 mg/l 300.0 g/l / 3.00 g/l 99.99 kg/l 99.99 t/m3 10.0 % 100.0 C lin = linear (1) sim = simulation (2) Tab = Table (3)
0.0 O312
0/4 mA
FNU
Enter the measured value corresponding to the minimum current value (0/4 mA) at the transmitter output. Display format from A3.
O313
9999 O313
20 mA
FNU
Enter the measured value corresponding to the minimum current value (0/4 mA) at the transmitter output. Display format from A3. If two factory settings are given, then the left-hand one is valid for CUS 41, the right-hand one for CUS 31. The simulation is terminated by selecting (1) or (3). See O3 (1), O3 (3) for other characteristics.
O3 (2)
lin
O3
Sel.Type
Factory settings are printed in bold face; base version does not include functions in italic.
Endress+Hauser
31
Instrument configuration
Coding
Field
Display
Info
O321
4.00 O321
Simulat.
mA
The current value entered here is output through the current output.
O3 (3)
table O3
Sel.Type
Values may also be added or changed at a later point in time. The values entered are automatically sorted in ascending order by current value The display, however, is not in sorted order. See O3 (1), O3 (2) for other characteristics.
O331
Read Edit
read O331
Sel.Table
O332
1 1 ... 10
1 O332
No. Elem.
This is where the number of pairs of x and y value (measured value and associated current value) is entered. The function chain O333 ... O335 will run though as many times as correspond to the value in O332. Asign appears as the last step. After confirmation the system jumps to O336.
O333
1 1 ... Number of table value pairs Finished 0.0 NTU 0.0 FNU 0.0 ppm 0.0 mg/l 0.0 g/l 0.0 kg/l 0.0 t/m3 0.0 % 0.0 C
1 O333
Sel.Elem.
O334
0.0 O334
Meas.val.
FNU
O335
0.00 O335
mA value
mA
y value = current value determined by user to be associated wht measured value entered in O334. Return to O3.
O336
yes no
yes O236
Status ok
If status = no, set table correctly (all previous settings are kept) or back to measurement mode (table will be deleted).
Factory settings are printed in bold face;base version does not include functions in italic.
32
Endress+Hauser
Instrument configuration
UM253E5A.CHP
5.5
Monitoring functions
The monitoring functions are used to define various alarms and to set output contacts. Each individual error can be defined to be effective or not (at the contact or as an error current). An alarm condition can be defined to activate a cleaning function (F8).
5.5.1
Alarm
Field Selection or range Factory setting (bold) Display Info
Coding
ALARM
Stead = steady contact Fleet = fleeting contact
F1
Stead F1
Cont.Type
The contact type selected here only applies to the alarm contact.
F2
s min
s F2
Time Unit
F3
0 F3
Err.Delay
Depending on the unit selected in F2, the alarm delay in entered in s or min.
F4
22 mA 2.4 mA
22mA
F4
Err.Curr
This selection must be made even if all errors are suppressed in F5. If you did not select 020 mA in O311, you may not use 2.4 mA. All errors can be selected for which an alarm message should occur. Select using the fault numbers. For individual error numbers, please refer to the table in chap. 7. The factory settings remain in effect for all errors not edited. If set to no, all the other alarm settings (e.g. alarm delay) are also deactivated. The settings themselves are retained. This setting only applies to the error selected in F5.
F5
1 1 ... 255
1 F5
Sel.Error
F6
yes F6
Rel.Assg
Endress+Hauser
33
Instrument configuration
Coding
Field
Display
Info The error current selected in F4 becomes effective or is suppressed in case of error. This setting only applies to the faults selected in F5.
F7
no yes
no F7
Curr.Assg
F8
no yes
no F8
CleanTrig
This field does not exist for some errors, see chap. 7.
F9
next F9 Select
If <- is selected. the display returns to F. If next is selected, the display returns to F5.
Factory settings are printed in bold face; base version does not include functions in italic.
34
Endress+Hauser
Instrument configuration
UM253E5A.CHP
5.5.2
Check
The function group CHECK is only accessible for instruments equipped with the Plus packet. PCS alarm (Process Check System) This function is used to examine the measuring signal for deviations. If the measuring signal is constant for a specific period of time (several measured values), an alarm is issued. This type of sensor behaviour may be caused by soiling or similar. Note: A current PCS alarm is automatically deleted as soon as the sensor signal changes.
Constant measuring signal = alarm is triggered
t
LIVE.CDR
Fig. 5.8
Coding
Field
Display
Info
CHECK
Off Low High Lo+Hi = low + high Lo! Hi! LoHi! Alarm signalling optionally with or without simultaneous controller switch-off. xxxx = without controller switch-off xxxx! = with controller switch-off Depending on your selection in F2, you can enter the error delay in min or s. Only after this does a high or low limit violation cause an alarm as per field P4 / P5.
P1
Off P1 A.Thresh
P2
min
P3
P4
9999 P4 HighAlarm
NTU
Factory settings are printed in bold face; base version does not include functions in italic.
Endress+Hauser
35
Instrument configuration
Coding
Field
Selection or range Factory setting (bold) Aus AC CC AC+CC AC! CC! ACCC!
Display
P5
Off P5 ProcMonit
NTU
Alarm signalling optionally with or without simultaneous controller switch-off. xxxx = without controller switch-off xxxx! = with controller switch-off
P6
60 P6 Tmax Low
min
Only if P5 = CC or AC+CC.
P7
min
Only if P5 = CC or AC+CC.
P8
Selected value is an absolute value. This function is mainly used for batch process and single-sided limit switches.
Factory settings are printed in bold face; base version does not include functions in italic.
36
Endress+Hauser
6.1
The function group RELAYS is only accessible for instruments equipped with the Plus packet. The relay contacts described below can be selected and configured as required (max. four contacts depending on options installed): Limit contactor for measured turbidity value R2 (1) Limit contactor for temperature: R2 (2) P(ID) controller: R2 (3) Timer for cleaning function: R2 (4) ChemoClean function R2 (5)
6.1.1
The relay contacts in the Liquisys M can be assigned different functions. Switch-on and switch-off points and pickup and dropout delays can be defined for the limit contactor. Moreover, an alarm threshold can be set to issue an error message and to start a cleaning function. These functions may be used for turbidity and temperature measurement. Please refer to Fig. 5.5 for a graphic representation of the alarm contact states of any relay or alarm contact.
Switch-on point > switch-off point: maximum function Measured value Alarm ON Alarm threshold Contact ON Switch-on point Switch-off point Contact OFF Alarm OFF
t1
t2
t3
t4
t5
t6 t7
Switch-on point Switch-off point Contact ON Alarm threshold Alarm OFF Alarm ON
t1
t2
t3
t4
t5
t6 t7
t
GRENZ-E.CDR
Fig. 6.1
Endress+Hauser
37
6.1.2
P(ID) controller
The Liquisys M supports the definition of various controller functions. On the basis of the PID controller, P, PI, PD and PID controllers can be implemented. To obtain the best control response, use the controller best suited to the application in question. P controller: Used for simple linear control purposes with small system deviations. Where major changes are to be controlled, overshooting may occur. A control offset is to be expected. PI controller: Used for processes where overshooting is to be avoided and permanent offsets are not allowed. PD controller: Used for processes that require quick response and where peaks are to be corrected. PID controller: Used for processes for which the type of control provided by a P, PI or PD controller is inadequate. Adjustments of P(ID) controller Three parameters can be adjusted in the case of a PID controller: the control gain Kp (P influence) the integral action time Tn (I influence) the derivative action time Tv (D influence) Start-up If there are no empirical values available for setting the control parameters, use values that provide the greatest possible stability of the control loop. To optimise the control loop further: Increase the control gain Kp until the control variable just starts to overswing. Decrease Kp again slightly and shorten the integral action time Tn to achieve the shortest possible correction time without overswing. In order to shorten the response time of the controller, you also have to set the derivative action time Tv.
38
Endress+Hauser
Tn too small
Tn too big
Time
Kp too big
Kp too small
Time
Time
Actual
Optimal setting
Time
REGEL02.CDR
Fig. 6.2
Actuating signal outputs (R237 ... R2310) The control contact in question outputs a switched signal. The intensity of this signal is proportional to the controllers output. A distinction is made according to the type of signal clock: Pulse length modulation The greater the calculated control output, the longer the contact in question remains picked up. The period can be adjusted between 0.5 and 99 s. Pulse-length modulated outputs are used to control solenoid valves. Pulse frequency modulation The greater the calculated control output, the higher the switching frequency of the contact. The maximum switching frequency 1/T can be adjusted between 60 and 180 min1. The ON period tON is constant. Pulse frequency-modulated outputs are intended for controlling solenoid-operated metering pump s.
Contact tOFF On On tON
IMPULS.CDR
Contact tON
Signal of a pulse lengthmodulated (left) and a pulse frequencymodulated control contact (right)
Endress+Hauser
39
Control characteristic for direct and inverse control action Field R236 offers two control characteristics for selection which have the effects shown in the following diagram.
Y
100 %
direct =max function
Xw
Y
Xw
100 %
inverse =min function Control characteristic of a proportional-action controller with direct and inverse control action
Fig. 6.4
Xw
Xw
REGEL03.CDR
6.1.3
This function can be used to implement a simple cleaning routine. The user can specify a time interval after which cleaning is to start; i.e. only constant intervals can be defined. There are other cleaning functions in conjunction with the ChemoClean function (version with four contacts, see chap. 5.5.4).
active
inactive Relationship between cleaning time, pause time and hold delay time Time
DIAGR-TI.CDR
Fig. 6.5
40
Endress+Hauser
6.1.4
ChemoClean function
Note:
Just like the timer function, ChemoClean can also be used to start a cleaning cycle. However, ChemoClean supports different cleaning and rinse intervals. Thus, irregular cleaning with different repeat cycles is possible, and cleaning times with post-rinse times can be individually defined.
Hold
Water (relay 3)
t1 t2 t3 t4 t5
t1 t2 t3 t4
Cleaning trigger Pre-rinse time Cleaning time Post-rinse time Hold dwell period
Cleaner (relay 4) t0 t1 t2 t3 t4 t5
CHEMO.CDR
Fig. 6.6
Coding
Field
Display
Info
RELAY
Rel1 Rel2 Rel3 Rel4
Rel3 (water) and Rel4 (cleaner) are only available on transmitter equipped accordingly. If ChemoClean is selected as the cleaning type, Rel4 is not available. PV = Process value If Rel4 is selected in Field R1, Clean = ChemoClean cannot be selected. Confirmation with ENTER switches off a different, already switched-on function and its settings are reset to the default.
R1
Rel1 R1
Sel.Relay
R2 (1)
LC PV = Limit contactor TU (1) LC C = Limit contactor T (2) PID controller (3) Timer (4) Clean = ChemoClean (5)
LC PV R2
Sel.Type
Factory settings are printed in bold face; base version does not include functions in italic.
Endress+Hauser
41
Coding
Field
Display
Info
R211
off on
off R211
Function
R212
9999 NTU 9999 FNU 9999 ppm / 3000 ppm 9999 mg/l / 3000 mg/l 300.0 g/l / 3 g/l 99.99 kg/l 99.99 t/m3 200.0 % 9999 NTU 9999 FNU 9999 ppm / 3000 ppm 9999 mg/l / 3000 mg/l 300.0 g/l / 3 g/l 99.99 kg/l 99.99 t/m3 200.0 %
9999
On
FNU R212
value
Never set the switch-on point and switch-off point to the same value. (Comp. chap. 5.1.1.) (Only the operating mode selected in A1 appears.) If two factory settings are given, then the left-hand one is valid for CUS 41, the right-hand one for CUS 31. Entering the switch-off point selects either a max contact (switch-off point < switch-on point) or a min contact (switch-off point > switch-on point) and implements a constantly required hysteresis (see Fig. 5.5). If two factory settings are given, then the left-hand one is valid for CUS 41, the right-hand one for CUS 31.
R213
9999
Off
FNU R213
value
R214
0s 0 ... 2000 s
0
On
s R214
Delay
R215
0s 0 ... 2000 s
0
Off
s R215
Delay
When the alarm threshold is exceeded/undershot, the measuring transmitter issues an alarm with an error message and error current (note alarm delay).
R216
9999 NTU 9999 FNU 9999 ppm / 3000 ppm 9999 mg/l / 3000 mg/l 300.0 g/l / 3 g/l 99.99 kg/l 99.99 t/m3 200.0 %
9999
FNU R216
A.Thresh
When defining the min contact, the alarm threshold must be set to a lower value than the switch-off point. If two factory settings are given, then the left-hand one is valid for CUS 41, the right-hand one for CUS 31.
The factory seFactory settings are printed in bold face; base version does not include functions in italic.
42
Endress+Hauser
Coding
Field
Display
Info
R217
MAX MIN
LC
Only display.
R2 (2)
LC PV = Limit contactor TU (1) LC C = Limit contactor T (2) PID = PID controller (3) Timer (4) Clean = ChemoClean (5)
LC C R2
Sel.Type
Confirmation with ENTER switches off a different, already switched-on function and its settings are reset to the default.
R221
off on
off
R221
Function
R222
100.0 R222
C
On value
Never set switch-on point and switch-off point to the same value! (see chap. 5.1.1.) Entering the switch-off point selects either a max contact (switch-off point < switch-on point) or a min contact (switch-off point > switch-on point) and a constantly required hysteresis is implemented (see 6.)
R223
100.0 R223
C
Off value
R224
0s 0 ... 2000 s
0 R224
s
On Delay
R225
0s 0 ... 2000 s
0 R225
s
Off Delay
When the alarm threshold is exceeded/undershot, the transmitter issues an alarm with an error message and error current (note alarm delay). When defining the min contact, the alarm threshold must be set to a lower value than the switch-off point.
R226
100.0 R226
C
A.Thresh
Factory settings are printed in bold face; base version does not include functions in italic.
Endress+Hauser
43
Coding
Field
Display
Info
R227
MAX MIN
LC
Only display.
R2 (3)
LC PV = Limit contactor TU (1) LC C = Limit contactor T (2) PID controller (3) Timer (4) Clean = ChemoClean (5) Off On Basic PID+B 0 NTU 0 FNU 0 ppm 0 mg/l 0.0 g/l 0.0 kg/l 0.0 t/m3 0.0 %
PID R2
Sel.Type
Confirmation with ENTER switches off a different, already switched-on function and its settings are reset to the default.
R231
off
R231
Function
On = PID control Basic = only basic load dosage PID+B = PID control with basic load dosage
R232
FNU R232
Setpoint
The set point is the value to be maintained by the control. This value is to be re-establish by the control in the event of any deviation (up or down).
R233
1.00 R233
Kp
R234
0.0
Time
min R234
Tn
min R235
See chap. 5.1.2. Each hold resets the I component to zero. Hold can be deactivated in S2, but not for ChemoClean and timer or wiper!
R235
0.0
Time
TD
Setting may or may not be required depending on control deviation (up or down deviation, see chap. 5.1.2).
R236
dir
R236
Direction
The factory settingFactory settings are printed in bold face; base version does not include functions in italic.
44
Endress+Hauser
Coding
Field
Display
Info Pulse length e.g. for solenoid valve, pulse frequency e.g. for solenoid-operated metering pump (see chap. 5.5.2). Selection of current output 2 is only possible, if field O2 = Contr is selected. This field only appears if pulse length is selected in R237. When pulse frequency is selected, R238 is skipped and input continues in R239. This field only appears if pulse frequency is selected in R237. When pulse length is selected, R239 is skipped, and input continues in R2310.
R237
len
R237
Oper.Mode
R238
10.0 R238
PulsePer.
R239
120
1/min R239
Max.PFreq
R2310
0.3 R2310
Min.PTime
R2311
0% 0 ... 40%
0 R2411 BasicLoad
Selecting the basic load, you choose the desired dosage quantity. 100% basic load corresponds to: steadily on at R237 = on Fmax at R237 = freq 20 mA at R237 = curr Cleaning with only one cleaning agent (usually water) and poss. as Wipe-Wash function together with the wiper (see Fig. 5.5 and setting in B1). Confirmation with ENTER switches off a different, already switched-on function and its settings are reset to the default.
R2 (4)
LC PV = Limit contactor TU (1) LC C = Limit contactor T (2) PID controller (3) Timer (4) Clean = ChemoClean (5)
Timer R2
Sel.Type
R241
off on
off
R241
Function
R242
30 s 0 ... 999 s
30 R242
RinseTime
The hold and relay settings are activated for the period of time specified here.
Factory settings are printed in bold face; base version does not include functions in italic.
Endress+Hauser
45
Coding
Field
Display
Info
R243
360 R343
PauseTime
min
The pause time is the time between two cleaning cycles (see chap. 5.5.4).
R244
120 R244
Min.Pause
min
The minimum pause time prevents continuous cleaning when the cleaning trigger is present.
R2 (5)
Configure ChemoClean function (on version with four contacts and appropriate assignment of contacts 3 and 4)
LC PV = Limit contactor cond. (1) LC C = Limit contactor T (2) PID = PID controller (3) Timer (4) Clean = ChemoClean (5)
Clean R2
Sel.Type
Contact 3 = Water, Contact 4 = Cleaner. Confirmation with ENTER switches off a different, already switched-on function and its settings are reset to the default.
R251
off on
off
R251
Function
= internal (timer-contr.) ext = external (digital input 2) i+ext = intern. + extern. i+stp = internal, suppressed by external int The int cycle is triggered by the end of the pause time (R257). There is no real-time clock. External suppression is required for irregular time interva (e.g. Fz!hxhxFhx)(hx7hxhxEhx
R252
int
R252
CleanTrig
R253
20 s 0 ... 999 s
20 R253
PreRinse
R254
10 s 0 ... 999 s
10 R254
CleanTime
R255
20 s 0 ... 999 s
20 R255
PostRinse
The factory settings are functions in italic. face; base version does not include functions in italic.
46
Endress+Hauser
Coding
Field
Display
Info
R256
0 0 ... 5
0 R256
Rep.Rate
R257
360 R257
PauseTime
min
R258
120
min R258
Min.Pause
The minimum pause time prevents continuous cleaning when the external cleaning trigger is present. Cleaning with cleaner can be followed by aup to 9 cleaning cycles that use only water until the next cleanig cycle with cleaner is performed.
R259
0 0 ... 9
0 R259
EconomyCl
The factory settings are functions in italic. face; base version does not include functions in italic.
Endress+Hauser
47
Instrument configuration
5.6
Concentration measurement
The Liquisys M can convert turbidity values to concentration values as standard. Since not all possible units or the corresponding conversions can be made available, the user can create a custom conversion table. Any assignment of a measured value to a converted value is possible (see Fig. 5.7). The defined values are connected by a curve based on linear interpolation, i.e. the dinstances between the points should not be too large.
0 500 [g/l]
[kg/l]
1,4
1,0
Assignment of measured value to a converted value
Fig. 5.7
U-UMRECH.CDR
Coding
Field
Display
Info
CONCENTRA
Four different concentration curves can be entered into this function group.
K1
Selection of concentration curve, which will be the basis of the display value calculation
1 1 ... 4
1 K1
act.Curve
The curves are independent of each other. Therefore, you can define four different curves.
K2
1 1 ... 4
1 K2
editCurve
The value pairs entered is independent of the curve used (K1). Only when entry is complete is the saved curve overwritten.
K3
Read Edit
read K3
Table
K4
1 1 ... 10
1 K4
No. Elem.
Factory settings are printed in bold face; base version does not include functions in italic.
48
Endress+Hauser
Instrument configuration
UM253E5B.CHP
Coding
Field
Selection or range Factory setting (bold) 1 1 ... Number of support points from K4 Fin
Display
Info
K5
Selection of triplet
1 K5
Sel.Elem.
K6
0 K6
MeasValue
The function chain K5 ... K7 is run through as many times as is equal to the value in K4. Finished appears as the last step. After confirmation the system jumps to K8.
K7
0 K7
kg/l
concentr.
Return to K.
K9
yes no
yes K8
Status ok
If not, then set table correctly (all previous settings are kept) or back to measurement mode (this makes the table invalid).
5.7
Service
Coding Field Selection or range Factory setting (bold) Display Info
SERVICE
ENG = English GER = Deutsch FRA = Francais ITA = Italiano NEL = Nederlands ESP = Espanol S+C = during setup and calibration CAL = during calibration Setup = during setup none = no hold
S1
Selection of language
ENG
S1
Language
This field must be set once during device configuration. After confirmation with ENTER
S2
Hold configuration
S+C
S2
S = setup, C = calibration.
Auto HOLD
Factory settings are printed in bold face; base version does not include functions in italic.
Endress+Hauser
49
Instrument configuration
Coding
Field
Display
Info
S3
Manual hold
off on
aus
Man.
S3
HOLD
S4
10 s 0 ... 999 s
10
s S4
Cont.Time
The code is located on the nameplate (see Figs. 1.1 1 and 1.2). Entry of an incorrect code returns you to the measurement menu. The number is edited with the PLUS or MINUS key and confirmed with the ENTER key. 1" is displayed if the code is active. The code is located on the nameplate (see Figs. 1.1 1 and 1.2). Entry of an incorrect code returns you to the measurement menu. The number is edited with the PLUS or MINUS key and confirmed with the ENTER key. 1" is displayed if the code isactive.
S5
0000
S5
PlusCode
S6
0000
S6
CleanCode
S7
order
TU0005
S7
S8
SerNo
S8
12345678
Reset of instrument (reset to default values) Facty= All data except language (field S1) are erased and reset to the factory setting!
S9
no
S9
S.Default
S10
no
Test
S10
Factory settings are printed in bold face; base version does not include functions in italic.
50
Endress+Hauser
Instrument configuration
UM253E5B.CHP
5.8
E+H Service
Coding Field Selection or range Factory setting (bold) Display Info
E+H SERV
Contr Trans Main Rel = controller (1) = transmitter (2) = power unit (3) = relay (4)
E1
Selection of module
Contr E1
Select
xx.xx
E111
SW-Vers.
This field cannot be edited. If E1 = Contr: Instrument software If E1 = Trans, Main, Rel: Module firmware
xx.xx
E112
HW-Vers.
SerNo
E113
12345678
LSG
E114
Modul-ID
Factory settings are printed in bold face; base version does not include functions in italic.
Endress+Hauser
51
Instrument configuration
5.9
Coding Field
Interfaces
Selection or range Factory setting (bold) Display Info
INTERFACE
Address HART: 0 ... 15 or PROFIBUS: 1 ... 126
I1
Enter address
126
Address
I1
I2
Tag description
Tag I2
@@@@@@@@
The factory settings are printed in bold; functions printed in italics are not available in the basic version.
52
Endress+Hauser
Instrument configuration
UM253E5B.CHP
5.10
Calibration
Note:
This function group is used to calibrate the transmitter. Calibration options Three-point calibration Correction of an existing calibration with the laboratory value Modification of individual values of the three-point calibration Installation adjustment One-point calibration
Original sample
Calibration value 3
Sample 3 original sample Sample 1 original sample 1/10 Sample 2 original sample 1/3
UM_3PROB.CDR
Fig. 5.8
Endress+Hauser
53
Instrument configuration
However, this procedure is not a mandatory prerequisite for calibration. You can also take three different samples of the process medium and determine their turbidity or solids concentration values. The following conditions must be met: Calibration must be carried out relating to the concentration in increasing order. The three samples must differ by at least 10 % relating to the measured value. At higher solids concentrations, clean water is sufficient as a zero solution.
Note:
Management of the transmitter calibration data records There are three calibration data records available for each application (FNU, NTU, ppm, g/l, %). Of these three data records (sample characteristics), one has been permanently programmed at the factory (read only).The other two are stored at the factory as standard characteristics. You can change and adapt these (three-point calibration, one-point adjustment, installation calibration). Note: FNU NTU Formazine Data set no. 1 (read only) Formazine Data set no. 2 (Changeable) Formazine Data set no. 3 (Changeable) (Changeable) (Changeable) (Changeable) SiO2 (Changeable) Activated sludge (read only) Kaolin (read only) Activated sludge ppm or mg/l SiO2 g/l Activated sludge % Residual concrete water (read only) Residual concrete water (Changeable) Residual concrete water (Changeable)
The data records are saved in the sensor. If you replace the transmitter, this data is not lost. It is loaded automatically (according to the selection) into the transmitter.
54
Endress+Hauser
Instrument configuration
UM253E5B.CHP
Version TB: Initial settings for residual concrete water The version Liquisys M CUM 223-TB and CUM 253-TB contains the Plus package with additional setting options, as for the TS version. On delivery, different values from the standard version are set in the individual fields of the menus to make commissioning as simple as possible. These initial settings are so selected so that no additional settings need be made for applications in residual concrete water. However, should you reset the device to the original factory settings (set default), then you can find the values which are valid for this application. Note: Should you wish to use a voltage output instead of the current output intended by the factory, then please connected the supplied shunt resistor according to the instructions on the sheet enclosed with the transmitter. You will then obtain a 0 ... 10 V current output.
Factory settings
Menu field Mode of operation Engineering unit Display format Measured value damping Calibration data record Current output Characteristic Current range Measured value 0 mA Measured value 20 mA Active characteristic Edited characteristic Table option Number of value pairs Support point Measured value / display value Language version A1 A2 A3 A5 B4 O1 O2 O211 O212 O213 K1 K2 K3 K4 K5 K6 / K7 S1 spec. kg/l XX.xx 10 3 Out1 lin 020 mA 1.00 1.30 1 1 edit 2 1 ... 2 1: 0 % / 1.00 2: 50 % / 1.50 GER Setting
Note: Calibration is carried out in the % measuring range (the transmitter switches automatically). With the above setting, a density of e.g. 1.12 kg/l = 12 % must be selected. In individual cases, it may be necessary to adapt the calibration to a real sample. For this, a singlepoint calibration is sufficient.
Endress+Hauser
55
Instrument configuration
For access to the CALIBRATION menu, please enter Code 22. Selection or range Factory setting
Coding
Field
Display
Info
CALIBRAT
3 Pt = Three-point calibration (1) Corr = Three-point correction (2) Edit = Edit calibration (3) Refl = Fitting with reflection effects (4) 1-Pt = Single-point calibration (5) Data = Calibration data (6)
C1 (1)
Selection of calibration
3-Pt C1
Calibrat
For data set 1 (B4), only the Data function is accessible. The offset is reset with 3 Pt and Edit.
Immerse the sensor so that there is sufficient distance to the tank wall (no reflection).
C111
100.0 C111
Concentr1
Immerse the sensor so that there is sufficient distance to the tank wall (no reflection).
FNU
C112
330.0 C112
Concentr2
FNU
Immerse the sensor so that there is sufficient distance to the tank wall (no reflection).
C113
1000.0 C113
Concentr3
FNU
Factory settings are printed in bold face; base version does not include functions in italic.
56
Endress+Hauser
Instrument configuration
UM253E5B.CHP
Coding
Field
Display
Cancel Warning:
Info
Warning: Cancel E084 200 % E045
C114
o.k. E xxx
o.k. C114
Status
E045
E084 50 %
20 %
500 %
C115
yes no new
yes C115
Store
If C114 = E xxx, then only no or new (Exception: calibration warning E 84). If new, return to C. If yes/no, return to Measurement.
-@EJ
C1 (2) Selection of calibration
-@EJ ?= E>H=JE ! 4AB .EJJE C MEJD HAB A?JE ABBA?JI " 2J 5E C A F E J ?= E>H=JE # ,=J= += E>H=JE @=J= $
Current value from C113 entire measuring range
Corr C1
Calibrat
C121
1000.0 C121
Concentr3
FNU
In case of an unknown probe concentration, but with a definite dilution (1/10; 1/3; 1), the laboratory value is to be entered .
C122
o.k. E xxx
o.k. C122
Status
If C122 = E xxx, then only no or new (Exception: calibration warning E 84). If new, return to C. If yes/no, return to Measurement.
C123
yes no new
yes C123
Store
Factory settings are printed in bold face; base version does not include functions in italic.
Endress+Hauser
57
Instrument configuration
Coding
Field
Selection or range Factory setting 3 Pt = Three-point calibration (1) Corr = Three-point correction (2) Edit = Edit calibration (3)
Display
Info
C1 (3)
Selection of calibration
4AB
.EJJE C MEJD HAB A?JE ABBA?JI " 2J 5E C A F E J ?= E>H=JE # ,=J= += E>H=JE @=J= $
Current value from C111 entire measuring range
Edit C1
Calibrat
If the individual samples were defined in the laboratory, you can correct the calibration here.
C131
100.0 C131
Concentr1
FNU
C132
330.0 C132
Concentr2
FNU
C133
1000.0 C133
Concentr3
FNU
C134
o.k. E xxx
o.k. C134
Status
If C134 = E xxx, then only no or new (Exception: calibration warning E 84). If new, return to C. If yes/no, return to Measurement.
C135
yes no new
yes C135
Store
C1 (4)
Selection of calibration
3 Pt = Three-point calibration (1) Corr = Three-point correction (2) Edit = Edit calibration (3) Refl = Fitting with reflection effects (4) 1 Pt = Single-point calibration (5) Data =Calibration data (6)
Refl C1
Calibrat
Only for solutions 2 FNU / 5 ppm! There is no effect of the installation in clear media
Factory settings are printed in bold face; base version does not include functions in italic.
58
Endress+Hauser
Instrument configuration
UM253E5B.CHP
Coding
Field
Selection or range Factory setting 0.0 NTU 0.0 ... 2.0 NTU 0.0 FNU 0.0 ... 2.0 FNU 0.0 ppm 0.0 ... 5.0 ppm 0.0 mg/l 0.0 ... 5.0 mg/l
Display
Info
C141
0.0 C141
Real PV
FNU
C142
o.k. E xxx
o.k. C142
Status
If C142 = E xxx, then only no or new (Exception: calibration warning E 84). If new, return to C. If yes/no, return to Measurement. For FNU: Adapter C164, C165. For ppm, mg/l: up to 500 adapter C164, C165, then adapter C166. For g/l, %: adapter C166. An existing basic calibration (three-point) is corrected by the single-point calibration.
C143
yes no new
yes C143
Store
C1 (5)
Select calibration
3 Pt = Three-point calibration (1) Corr = Three-point correction (2) Edit = Edit calibration (3) Refl = Fitting with reflection effects (4) 1-Pt = Single-point calibration (5) Data= Calibration data (6) Current measured value Complete measuring range
1-Pt C1
Calibrat
C151
0.0 C151
Real PV
FNU
C152
o.k. E xxx
o.k. C152
Status
If C152 = E xxx, then only no or new (Exception: calibration warning E 84). If new, return to C. If yes/no, return to Measurement.
C153
yes no new
yes C153
Store
Factory settings are printed in bold face; base version does not include functions in italic.
Endress+Hauser
59
Instrument configuration
Coding
Field
Selection or range Factory setting 3 Pt = Three-point calibration (1) Corr = Three-point correction (2) Edit = Edit calibration (3) Refl = Fitting with reflection effects (4) 1-Pt = Single-point calibration (5) Data =Calibration data (6)
Display
Info
C1 (6)
Selection of calibration
Data C1
Calibrat
C161
Comparison value
101.4 C161
Concentr1
C161
Comparison value
99.3 C162
Concentr2
C161
Comparison value
98.7 C163
Concentr3
C164
Slope 1 is displayed
Current value
230 C164
Slope 1
C164
Slope 2 is displayed
Current value
375 C165
Slope 2
C166
Current value
1 C166
ConvFact
Factory settings are printed in bold face; base version does not include functions in italic.
60
Endress+Hauser
Instrument configuration
UM253E5B.CHP
5.11
Offset
Measured value
The settings in the OFFSET function group can be used to calibrate the measurement to a reference measurement. This requires a linear shift of all measured values, i.e., the adjustment is determined for one measured value, and all the others are calculated using the same offset. Note: You can either edit offset or slope. During offset editing, a previous edited slope is reset to the factory setting.
Calibrated value
Time t
OFFSET.CDR
Fig. 5.9
Offset
Coding
Field
Display
Info
OFFSET
V1
0.0 V1
Real PV
FNU
V2
Enter offset
Current offset
0.0 V2
FNU
PV Offset
V3
o.k. E xxx
o.k. V3
Status
V4
Store offset
yes no new
yes V4
Store
Factory settings are printed in bold face; base version does not include functions in italic.
Endress+Hauser
61
Instrument configuration
5.12
Slope
With the settings in the function group SLOPE, a measured value can be adapted to a reference value. All measured values are proportionally adapted over the entire measuring range according to this change. Example: The displayed measured value is 2.5 g/l. It is adapted to the reference value of 2.0 g/l by means of the slope function. The change is 20%, i.e. all measured values are reduced by 20% over the entire measuring range. An OFFSET being edited before is reset to the factory setting. In contrast to the one-point calibration, the edited slope can be reset by setting the slope factor to 1.0. Coding Field Selection or range Factory setting Display Info
SLOPE
N1
0.000 N1
Real PV
N2
Enter slope
Current slope
1.000 N2
PV Slope
N3
Status is displayed
o.k. E xxx
o.k. N3
Status
N4
Store slope
yes no new
yes N4
Store
Factory settings are printed in bold face; base version does not include functions in italic.
62
Endress+Hauser
Interfaces
UM253E06.CHP
Interfaces
For instruments equipped with a communication interface, please refer to separate operating instructions BA 208C/07/en (HART) or BA 209C/07/en (PROFIBUS).
Endress+Hauser
63
Maintenance: Please take all measures in due time which will guarantee the safety of operation and reliability of the entire measuring system. Maintenance of CUM 223 / 253 comprises: Calibration (see chap. 5.10) Cleaning of assembly and sensor Checking of cables and connections Troubleshooting Determination and elimination of the cause of the problem in the case of an operating fault. Troubleshooting refers to measures that can be performed without intervention in the instrument (for instrument defects, see chap. 8, Repairs). Troubleshooting of the CUM 223/253 and the measuring system is performed with the aid of the troubleshooting table in chap. 7.1. Warning: Note:
7.1
Error Instrument cannot be operated, value 9999
Display value 0.0, and error message E008 Sensor and display in A4
Instrument input defective Break in sensor extension line Data transfer error
Incorrect sensor connection Sensor completely blocked Impermissible instrument operating state (no response to key actuation) Measuring cable interference
Signal output line interference Measured value fluctuates Irregular flow rate / turbulence / air bubbles / big solids particles
Connect cable screen acc. to connection diagram (not grounding) check line routing, try separate line routing, grounding screen to SPS/PLS Choose better place of installation or eliminate turbulences. Possibly use large measured value damping factor.
64
Endress+Hauser
Error
Remedy Calibration with original sample required for concentration or solids concentration Clean sensor
Sensor soiled
Remove coarse coats mit brush Remove carbonates and similar coats with 3% hydrochlorid acid Remove organic coats and grease with oxidation agents and / or grease solvents. See instructions of assembly used for spray cleaning Retrofitting possible: wiper retrofitting service kit 50089254 Wiper arm service kit 50089252
Use spray cleaning User wiper version Wiper rubber defective Sensor installed in dead zone or air cushion in assembly or flange Incorrect sensor orientation Replace wiper arm Check installation conditions, move sensor in area with optional flow conditions. Caution if installed in horizontal lines! Orient sensor: Measuring surface should face flow in normal media. Orient meas. surfacre at 90 to flow in media with high solids concentration If temperature display is required: replace sensor Install LSR1-2 or LSR1-4 module Activate controller Choose Auto or Manual/On mode Disable or shorten pickup delay Autom. hold during calibration, hold input activated; hold via keyboard active Set controller to Manual / Off or Auto Shorten dropout delay Check measured value, current output or relay contacts, actuators, chemical supply Disconnect line and measure directly on instrument Disconnect line and measure directly on instrument See diagnosis table in chap. 8.1 PA-/DP instruments have no current output! Switch off simulation Switch instrument off and back on Check current assignment: 020 mA or 420 mA? Disconnect output and measure directly on instrument Use sensible intervals
Frontal bombardment of measuring surface with highly viscous solids may result in an adhering coating. Turbidity measurement itself does not require temperature meas. See chap. 8.2 and 8.3 See chap. 5.5 or fields R2xx Keyboard, REL key See fields R2xx See fields S2 to S4
Temperature sensor defective No relay module installed Controller switched off Controller in Manual / Off mode
Controller in Manual / On mode Controller / limit contactor works continuously Dropout delay setting too long Control loop interruption
Line open or short-circuited No turbidity current output signal Total load in current loop excessive (> 500 ) Output defective Instrument with PROFIBUS-PA/-DP Fixed current output signal Current simulation active Processor system out of sync Incorrect current assignment Incorrect current output signal Current output table not accepted Total load in current loop excessive (> 500 ) Value interval too small
See field O2(2) EMV problem: check installation if problem persists Field O2(1), O2(3) for table Ohmmeter
Endress+Hauser
65
Error
Possible cause Instrument only has one current output Instrument with PROFIBUS-PA/-DP No relay module (LSR1-x) or nur LSR1-2 installed
Remedy Refer to nameplate for variant, change LSCH-x1 module if nec. PA-/DP instruments have no current output! Install LSR1-4 module. Chemoclean is enabled via release code supplied by E+H with Chemoclean upgrade Plus package upgrade: code received from E+H enter Following replacement of defective LSCH/LSCP module: first enter instrument serial number (see nameplate) manually, then enter code Verify by looking at nameplate: HART = -xxx5xx und -xxx6xx
Equipment need, spare parts, information LSCH-x2 module, see chap. 8.2.4 and 8.3.4
Plus package not enabled (enable with code that depends on serial number and is received from E+H with order or Plus package) No central HART module No or wrong DD device description) HART interface missing Instrument not registered with HART server Load < 230 HART receiver (e.g. FXA 191) not connected via load Incorrect instrument address (Adr. = 0 for single operation, Adr. > 0 for multidrop operation) Line capacity too high Line interference No central PA-/DP module Wrong SW version (without PROFIBUS) Commuwin (CW) II: CW II version and instrument SW version incompatible No or wrong DD/DLL Incorrect baud rate setting for segment coupler in DPV-1 server Incorrect station (master) address or duplicate address Incorrect station (slaves) address Bus line not terminated Line problems (too long, cross section too small; not shielded, screen not grounded, wires not twisted) Bus voltage too low (bus supply voltage typ. 24 V DC for non-Ex, 13,5 V DC for Ex)
No HART communication
For further information, see additional Operating Instructions BA 208C, HART Field communication with Liquisys M CxM 223 / 253
No PROFIBUS communication
For further information, see additional Operating Instructions BA 209C, PROFIBUS-PA/-DP Field communication with Liquisys M CxM 223 / 253
The voltage at the PA-/DP connection of the instrument must be at least 9 V. Check addresses and reset if necessary
66
Endress+Hauser
7.2
E002
Instrument not calibrated, calibration data invalid, no user data invalid (EEPROM error).
E003
Download error Instrument software version not compatible with module hardware version Transmitter malfunction Software not compatible with hardware (transmitter) Sensor or sensor connection defective Wiper error Calibration aborted Below main parameter measuring range Main parameter measuring range exceeded Below temperature measuring range Temperature measuring range exceeded Below current output range 1 Current output range 1 exceeded Below current output range 2 Current output range 2 exceeded Set point exceeded controller 1 Set point exceeded controller 2 Set point exceeded controller 3 Set point exceeded controller 4
E004
E007
Check sensor and sensor connection (E+H Sevice) Check wiper and test function using manual control if necessary Repeat calibration Check measurement and connections; check instrument and measuring cable with simulator if necessary. Check measurement, control and connections
no no no no
no no no *
yes
no
no
E057
E059 E061 E063 E064 E065 E066 E067 E068 E069 E070
yes yes Check configuration Check measured value and current assignment yes yes yes yes Check configuration. yes yes Check configuration yes yes
no no no no no no no no no no
no no no no no no no no no no
Endress+Hauser
67
Contact Display Measured value outside concentration table Parameter range for current output 1 range too small Parameter range for current output 2 range too small Calibration warning Measures Fact Clean sensor; check tables Decrease current output spreading yes no User
E080
no
no
E081
E084 E100 E101 E102 E106 E116 E152 E153 E154 Current simulation yes Service function yes Relais simulation yes Download yes Download error PCS alarm Offset Below lower alarm threshold for period exceeding alarm delay Upper alarm threshold exceeded for period exceeding alarm delay Actual value undershoots monitoring value for longer than the set permissible maximum period Actual value exceeds monitoring value for longer than the set permissible maximum period Dosage stop Flow in main stream loo low or zero Switch-off limit for current iput exceeded Current input < 4 mA
Calibration data are within limits but deviate from standard values by a factor of more than two Switch off current simulation Switch instrument off and back on Check relay configuration Wait for download to end Repeat download Check sensor and connection, maintain or replace Adjustment range exceeded If necessary perform manual reference measurement. Repair sensor and recalibrate
no
no
no
no no no no no no no yes
no no no no no no no no
no no no
* * * * *
yes
no
no
E155
yes
no
no
E156
yes
no
no
E157
Check settings in CURRENT INPUT or CHECK function group. Restore flow. Check process variables at sending measuring instrument. Check process variables at sending measuring instrument. Change range assignment if required. Check process variables at sending measuring instrument. Change range assignment if required.
no no no no
no no no no
E173
yes
no
no
E174
68
Endress+Hauser
Diagnosis Diagnosis refers to the identification of instrument malfunctions and defects. Corrective maintenance: replacement of parts diagnosed to be defective; testing of instrument and measuring system function; restoration of complete functionality. Diagnosis based on the error table below and depending on difficulty and measuring equipment at hand is to be performed by: trained operator personnel operators electricians company responsible for system installation/ operation E+H Service. Select the required spare parts using the tables in chap. 8.2 and 8.3.
Warning:
Caution: ESD!
8.1
Diagnosis
The table below will help you diagnose problems and provides information about spare parts required. Note on the required spare parts. You can find information concerning the exact spare part designations and the installation of these parts in chap. 8.2.3 and 8.3.3.
Error Display dark, no LEDs active Possible cause No mains voltage Wrong supply voltage / voltage too low Connection fault Tests and / or remedial measures Equipment, spare parts, personnel Check if mains voltage is available Compare mains voltage and rating on nameplate Terminal not tightened; insulation clamped in terminal; wrong terminal used Replace fuse, first compare mains voltage and rating on nameplate Replace power supply unit, using correct variant Replace controller, using correct variant Check ribbon cable, replace if necessary Replace controller Electrician / e.g. multimeter Operator (utility company specification or multimeter) Electrician
Electrician / see drawings in chap. 8.2.1 and 8.3.1 for corrected fuse On-site diagnosis by E+H Service (test module LSGx Lrequired) On-site diagnosis by E+H Service (test module LSCxx Lrequired) See spare parts for CUM 253 On-site diagnosis by E+H Service (test module LSCxx required)
Power supply unit defective Controller defective CUM 253: ribbon cable Pos. 310 loose or defetive Display dark, but LED(s) active Controller defective (Module: LSCH/LSCP)
Endress+Hauser
69
Error Display shows measured values but value does not change and / or instrument cannot be operated Instrument gets hot
Possible cause Instrument or module not properly installed in instrument Impermissible operating system state
Tests and / or remedial measures Equipment, spare parts, personnel CUM 223: Reinstall module CUM 253: Reinstall display module Switch instrument off and back on Refer to assembly drawings in chapters 8.2.1 and 8.3.1. Possible EMC problem: if problem persists, call E+H Service to have installation checked. Correct voltage setting, see chap. 8.4.1 / 8.5.1. Can only be diagnosed by E+H Service. If test fails: replace module (using correct variant), refer to exploded views in chap. 8.2.1 and 8.3.1. If module tests out OK: check peripherals once more. Replace LSGA module (AC power supply unit) or LSGD (DC power supply unit) Check line routing. Separate sensor line from supply lines Max. line length with extension is 200 m; use only calbe type CYK 8 Test with built-in current simulation, connecting mA meter directly to current output If simulation value is incorrect: recalibration at factory or new LSCxx module required. If simulation value is correct: check current loop for load and shunts. Spare parts list see chap. 8.4.4 or 8.5.4. Connect sensor line screen to S terminal, do not ground.
Compare mains voltage and rating of nameplate Replace power supply unit Test measuring inputs: Sensor simulation is not possible, check with new or other sensor
Transmitter module defective (module: MKT1). Please perform tests and take measures according to chap. 7.1 to eliminate the sensor or wiring as the cause of the problem Data transfer fault
Data transfer problem (EMC) Wrong sensor cable / cable too long Current outputs / incorrect current Not calibrated correctly Load excessive Shunt / short-circuit to frame in current loop
Incorrect mode of operation No current output signal Current output stage defective (module: LSCH/LSCP)
Check whether 020 mA or 420 mA has been selected Test with built-in current simulation, connecting mA meter directly to current output. First disconnect current output line completely for safety Make sure ribbon cable is properly connected, replace cable if nec. Convert to LSR1-4 with 4 relays If upgraded: verify that correct serial number has been used to order Plus package Check whether serial number on nameplate matches with stored serial number (can be read in field S8). For new LSCH / LSCP, the instrument serial number can be entered once in fields E114 to E116. Next step is to enter release codes for Plus package and/or Chemoclean. HART: LSCH-H1 or -H2 PROFIBUS-PA/-DP: LSCP-PA/-DP module, see field E112 SW version see field E111 Remove some instruments and repeat testing
Additional relays do not work Only 2 additional relays can be used Enhanced functions (Plus package) not available
CUM 253: ribbon cable Pos. 320 loose or defective Relay module LSR1-2 with 2 relays installed No or incorrect release code used Incorrect serial number of instrument stored in LSCH/LSCP module
See Spare parts CUM 253 F Operator or E+H Service Handled by E+H Sales
Serial number stored in LSCH/LSCP module is required to enable Plus package. A detailed description can be found in chap. 8.3.5.
Enhanced functions (Plus package and / or Chemoclean) are not available after replacement of LSCH-/LSCP module No HART or PROFIBUS interface function
Plus package or Chemoclean has not been enabled with release codes.
Wrong controller
70
Endress+Hauser
8.2
8.2.1
Pos. 426a Terminal for signal inputs/outputs (external) Pos. 426b Terminal for alarm relay and power supply (interior)
Pos. 270 Turbidity transmitter MKT1 Pos. 10 / 20 Power supply unit LSGA / LSGD Pos. 40 90 Controller LSCH-x / LSCP End frame latches Pos. 330 Housing assembly
Pos. 310 Side wall for instruments without relay module Pos. 430 Push-on terminal strip for LSR1-x module Pos. 10 Power supply unit 100 230 V AC
Pos. 290 / 300 Optional relay module LSR2-x Module rotated 180 for representation!
Jumper
Endress+Hauser
71
8.2.2
Consider potential effects on process when removing the instrument from service! Before opening, make sure the instrument is voltage-free. First pull off the terminal block (Pos. 426b) on the rear of the instrument. First pull off the terminal blocks (Pos. 426a and if necessary 430) on the rear of the instrument. You can now dismantle the instrument. Press the latches of the end frame (Pos. 340) inwards and pull off the frame towards the rear. Loosen the special screw (Pos. 400) by turning it counterclockwise. Remove the complete electronics block from the housing. The modules are plugged together mechanically and can be easily separated: Pull the controller / display module forwards Gently pull out the backplane latches (Pos. 320) outwards slightly to remove the side modules. Removing the turbidity transmitter (Pos. 270): Pull off the module towards the top.
8.2.3
Reverse the dismantling sequence for assembly. Hand-tighten the special screw without using a tool. Incorrect assembly is not possible! An module block which has been plugged together incorrectly cannot be inserted into the housing.
72
Endress+Hauser
8.2.4
330, 400
Housing module
51501075
51501076
426a, 426b
51501205
426a, 426b
51502128
426a, 426b
51502491
430
51501078
Endress+Hauser
73
8.3
8.3.1
Electronics box with relay module LSR1-x (bottom) and power supply unit LSGA/LSGD (top)
Fuse is accessible with electronics box installed Pos. 10 Power supply unit 100 230 V AC
Fuse Pos. 370 Front membrane Pos. 40 ... 90 Controller LSCH-x / LSCP Mains voltage selection Pos. 450b Torx screw for Controller
230V 115V 100V
Jumper
Top Bottom
Pos. 340 Docking module LSK1 Pos. 480 Cable glands Pos. 450c Torx screws K4x10 Pos. 470 Plug-in terminal for power connection
74
Endress+Hauser
8.3.2
Before opening, make sure the instrument is voltage-free. Open and remove the connection compartment cover (Pos. 420). Remove the network terminal (Pos. 70). Open the display cover (Pos. 410) and loosen the ribbon cable (Pos. 310 / 320) on the electronics box side (Pos. 330). Removal of electronics box (Pos. 330): Loosen screws (Pos. 450a) in the bottom of the housing 2 revolutions, then slide entire box backward and remove towards the top. Make sure that the module catches do not come unlatched while you push the electronics box back! Now bend the module latches outward and remove the module(s). Dismantling the controller (Pos. 40): Loosen screw (Pos. 450b) in the display lid. Expansion of the docking module (Pos. 340): Loosen screws (Pos. 450c) in the bottom of the housing and remove entire module towards the top. Removing the transmitter (Pos. 270): Simply pull off the module towards the top.
8.3.3
Insert the module(s) in the electronics box guide rails carefully and latch into the lateral lugs in the box. You cannot assemble the instrument incorrectly! Modules inserted in the electronics box incorrectly are not operable since the ribbon cables cannot be inserted. Make sure that the cover gaskets are intact since they are required to guarantee protection class IP 65.
Endress+Hauser
75
8.3.4
Position 10 20 40 50 60 70 80 90 270 290 290 300 300 370, 410, 420, 430, 460 400, 480 330, 340, 450 310, 320 430 470
Housing cover
51501068
Bottom of housing Internal housing parts Ribbon cables Hinges Terminal strip
76
Endress+Hauser
8.3.5
8.4
Spare parts are to be ordered from your local E+H Sales Agency. See the back cover of these operating instructions for addresses. For this, refer to the order numbers in chapters 8.2.4 and 8.3.4. To be on the safe side, you should always specify the following data with spare parts orders:
8.5
In connection with the Scopeware software, the optoscope offers the following possibilities, with the need to disassemble or open the Liquisys S and without a galvanic connection to the device: Documentation of the instrument settings in conjunction with Commuwin II Software update by the service technician Upload/download a hex dump to duplicate configurations. The optoscope serves as an interface between the Liquisys M and PC / laptop. Information is exchanged on the instrumentside using the optical interfaces of the Liquisys M and to the PC / laptop using the interface RS 232.
Endress+Hauser
77
8.6
8.6.1
The sensors CUS 31 and CUS 41 cannot be simulated as they contain the complete data processing and all the measured values are transmitted using the digital interface RS 485 to CUM 223 / 253. Therefore a functional sensor is required for the measuring point test. Method for testing a measuring point: Check that device is operable and that the display reacts appropriate, e.g. by pressing the PLUS key. Check the current outputs by carrying out a current simulation (Field O22).
Cleaning the front panel To clean the front panel, only use standard cleaning agents. The front panel is resistant to the following cleaning agents (acc. Isopropanol Thinned acids (3 %) Thinned alkalis (5 %) Ester Hydrocarbons Ketone Household cleaners Caution! Do not use any concentrated mineral acids or alkalis, benzyl-alcohol, methylene chloride or high-pressure water vapour at over 100 C.
8.6.2
The sensors CUS 31 / 41 contain their own digital signal processor and communicate with the turbidity measuring instrument via an interface RS 485. All the sensor data (factory calibration data and customer calibration data) are permanently saved in the sensor. You can find detailed information on these sensors in the Technical Information Turbidity sensor TurbiMax W CUS 31 / CUS 31-W TI 176C/07/en or in Technical Information Solid matter sensor TurbiMax W CUS 41 / CUS 41-W TI 177C/07/en.
78
Endress+Hauser
8.6.3
Assembly
Flow Flow assembly FlowFit W CUA 250 Operating manual BA 096C/07/en Welding socket with adapter CUA 120-B Retractable assembly ProbFit CUA 461 Operating manual BA 134C/07/en Pure and ultrapure water applications CUS 31-xxE: contains special flow assembly without gas bubble trap CUS 31-xxS: contains special flow assembly with gas bubble trap
Refer to the assembly operating instructions for maintenance and troubleshooting! For the CUS 31/41, you can use: Basin and channel Suspended assembly holder CYH 101 and basin wall bracket CYY 106 Technical Information TI 092C/07/en Assembly DipFit W CYA 611 Operating manual BA 166C/07/en Sensor fixing bracket RM Operating manual BA 205C/07/en
Endress+Hauser
79
Accessories
Accessories
Connection accessories Extension cable CYK 8, metre goods Unterminated measuring cable for turbidity sensors CUS 31 and CUS 41 Order no.: 50089633 Extension cable CYK 8, 50 m Terminated measuring cable for turbidity sensors CUS 31 and CUS 41. Order no.: 51500830 Junction box VBM Junction box for extending the measuring cable connection between sensor and instrument. Pg 13.5 cable entries. Material: cast aluminium; Ingress protection: IP 65 Order no.: 50003987 Junction box RM Junction box for extending the measuring cable connection between sensor and instrument. Pg 13.5 cable entries. Material: cast aluminium; Ingress protection: IP 65 Order no.: 51500832 Assemblies DipFit W CYA 611 Immersion assembly for turbidity sensors, unpressurised. Technical Information TI 166C/07/en Order no.: 50085985 FlowFit W CUA 250 Flow assembly for turbidity sensors, applicable at up to 6 bar Technical Information TI 096C/07/en Order no.: 50063827 ProbFit CUA 461 Manually replacable assembly for turbidity sensors, applicable at up to 1 bar Technical Information TI 134C/07/en Order no.: 50075623
Mounting accessories Weather protection cover CYY 101 For mounting on field housing, for outdoor installation. Dimensions (H W D): 320 300 270 mm Material: Stainless steel SS 304 Order no.: CYY 101-A Universal upright post CYY 102 Square tube for mounting of field housing. Dimensions (H W D): 1495 60 60 mm Material: Stainless steel SS 304 Order no.: CYY 102-A Sensors TurbiMax W CUS 31 / 31 W pH/ redox combination electrode with PTFE diaphragm, with gel filling, usable at up to 6 bar. Technical Information TI 028C/07/en Order no.: 50088400 TurbiMax W CUS 41 / 41 W Sterilisable pH /redox electrode with double-chamber reference and integrated bridge electrolyte. Technical Information TI 245C/07/en Order no.: 50088402 Spare parts for CUS 31 and CUS 41 Service kit for wiper operation cpl. (upgrade or replacement) Order no.: 50089254 Service Kit for seal (for wiper extension, contains 5 sets) Order no.: 50089253 Service Kit for wiper arm (with rubber and mounting material, contains 3 sets) Order no.: 50089252 Measuring cable 7 m terminated, with housing upper section and screw union Order no.: 50088167 O-ring 28 3.5 mm (Seal for G 1 pipe installation) Order no.: 50051753 Accessories for CUM 253 CSA version NPT adapter set Order no.:
80
Endress+Hauser
Accessories
UM253D09.CHP
Istrument upgrade (Order only possible with serial number of relevant device) Plus package Order no.: 51500385 ChemoClean Order no.: 51500963 Two-relay card (standard version) Order no.: 51500320 Two-relay card (CSA version) Order no.: 51511446 Four-relay card (standard version) Order no.: 51500321 Four-relay card (CSA version) Order no.: 51511447 Two-relay card with current input (standard version / CSA version) Order no.: 51504304 Four-relay card with current input (standard version / CSA version) Order no.: 51504305
Endress+Hauser
81
% C Technical data
10
General specifications
Technical data
Endress+Hauser Liquisys M CUM 223 Liquisys M CUM 253
Measuring principle
A sensor CUS 31 or CUS 41 is connected to the transmitter Liquisys M CUM 223 / 253 using the digital sensor interface. The sensors supply a standardised signal for the turbidity values and the temperature value.
Input
Measured variables Turbidity measurement with CUS 31 Measuring range Turbidity offset range Solids contentration measurement with CUS 41 Measuring range Solids concentration offset range Temperature measurement Temperature sensor Sensor measuring range Temperature offset range
0.000 ... 9999 FNU, 0.00 ... 3000 ppm, 0.0 ... 3.0 g/l, 0.0 ... 200.0 % 99.99 FNU, 99.99 ppm, 99.9 g/l, 99.9 %
0.00 ... 9999 FNU, 0.00 ... 9999 ppm, 0.0 ... 300.0 g/l, 0.0 ... 200.0 % 99.99 FNU, 99.99 ppm, 99.9 g/l, 99.9 %
Signal input for turbidity / solids concentration / temperature Sensor interface Maximum cable length to sensor Terminal cross-section Required conductor cross-section CUM 223 CUM 253 Digital inputs 1 and 2 Voltage Current consumption Current input Current range Load 4 ... 20 mA, galvanically separated 260 at 20 mA (voltage drop 5.2 V) 10 ... 50 V max. 10 mA digital 200 m 2.5 mm2 0.75 mm2 / AWG 18 1.50 mm2 / AWG 14
Output
Signal output for turbidity / solids concentration Current range Max. resolution Load Transmission range Separation voltage Overvoltage (lightning) protection 0 / 4 ... 20 mA, galvanically isolated; error current 2.4 / 22 mA 700 digits / mA max. 500 adjustable, min. 0.1 FNU, 0.1 ppm, 1 g/l, 0.1 % max. 350 Vrms / 500 V DC acc. to EN 61000-4-5:1995
82
Endress+Hauser
Technical data % C
UM253E10.CHP
Output (continued)
Signal output temperature (optional) Current range Max. resolution Load Transfer range Separation voltage Overvoltage (lightning) protection Auxiliary voltage output Output voltage Output current Contact outputs (potential-free changeover contacts) Switching current with ohmic load (cos = 1) Switching current with ohmic load (cos = 0.4) Switching voltage Switching power with ohmic load (cos = 1) Switching power with inductive load (cos = 0.4) Limit contactor Pickup / dropout delay Controller Function (adjustable) Controller response Controller gain Kp Integral action time Tn Derivative action time Tv Period for pulse length controller Frequency for pulse frequency controller Basic load Alarm Function (switchable) Alarm delay Steady / fleeting contact 0 ... 2000 s (min) Pulse length / pulse frequency controller P, PI, PD, PID 0,01 ... 20.00 0.0 ... 999.9 min 0.0 ... 999.9 min 0.5 ... 999.9 s 60 ... 180 min1 0 ... 40% of max. set value 0 ... 7200 s max. 2 A max. 2 A max. 250 V AC, 30 V DC max. 500 VA AC, 60 W DC max. 500 VA AC, 60 W DC 15 V 0.6 V max. 10 mA 0 / 4 ... 20 mA, galvanically isolated 700 digits / mA max. 500 adjustable, 10 ... 100 % from measuring range max. 350 Vrms / 500 V DC acc. to EN 61000-4-5:1995
Endress+Hauser
83
% C Technical data
Accuracy
Turbidity measurement with CUS 31 Measured value resolution Display deviation1 Repeatability1 Measurement deviation1 turbidity signal output Solids content measurement with CUS 41 Measured value resolution Display deviatoin Repeatability1 Measurement deviation1, solids concentration signal output Temperature measurement Measured value resolution Display deviation1 Measurement deviation1 temperature signal output 0,1 C max. 1.0 % of measuring range max. 1.25% of current output range
1
0.001 FNU, 0.01 ppm, 0.01 g/l, 0.01 % 2 % of measured value (at least 0.02 FNU) 1 % of measured value (at least 0.01 FNU) 1 % of current output range (at least 0.02 FNU)
0.01 FNU, 0.01 ppm, 0.01 g/l, 0.01 % 2 % of measured value (at least 0.02 FNU) 1 % of measured value (at least 0.01 FNU) 1 % of current output range (at least 0.02 FNU)
Ambient conditions
Ambient temperature (nominal operating conditions) Ambient temperature (limit operating conditions) Storage and transportation temperature Relative humidity (nominal operating conditions) Ingress protection of panel-mounted instrument Ingress protection of field instrument Pollution degree Installation category Maximum altitude Electromagnetic compatibility
10 ... +55 C 20 ... +60 C 25 ... +65 C 10 ... 95 %, non-condensing IP 54 (front panel), IP 30 (enclosure) IP 65 2 acc. to IEC 61010-1 II 2000 m above sea level Interference emission and immunity acc. to EN 61326: 1997/ A1: 1998
Mechanical construction
Dimensions of panel-mounted instrument (H W D) Installation depth Dimensions of field instrument (H W D) Weight of panel-mounted unit Weight of field instrument Measured value display Materials Housing of panel-mounted unit Front membrane Field housing standard version CSA GP version
96 96 139 mm approx. 165 mm 247 170 115 mm max. 0.7 kg max. 2.3 kg LC display, 2-line, 5- and 9-digit with status displays
Power supply
100 / 115 / 230 V AC +10 / 15 %, 48 ... 62 Hz 24 V AC/DC +20 / 15 % max. 7.5 VA Microfuse, medium slow 250 V / 3.15 A
84
Endress+Hauser
Technical data % C
UM253E10.CHP
Endress+Hauser
85
Appendix
11
Function group
Appendix
Enter absolute value current measured value V1 Enter offset Calibration status is displayed V2 o.k. E--V3 Store offset yes; no; new V4 V
OFFSET
Function group
SLOPE
V
Calibration selection Data = calibration data; only C1 (6) is available for data set 1 C1 (6)
C161
C162
C163
Entry of current calibration value current measured value entire measuring range C1 (5) C151
Entry of correct measured value 0; 0 ... 2.0 FNU 0; 0 ... 5.0 ppm (mg/l)
C1 (4)
Entry of concentration of first calibration solution current value in C111 entire measuring range C131
Entry of concentration of second cal. solution current value in C112 > value in C131 11 C132
Entry of concentration of third calibration solution current value in C113 > value in C132 1.1 C133
C1 (3)
Entry of correct concentration of calibration solution 3 current value in C113 entire meas. range C121
C1 (2)
CALIBRATION
C1 (1)
C111
C112
C113
Function group
SETUP 1
A
Function group
SETUP 2
Function group
CURRENT INPUT
Z
Function group
CURRENT OUTPUT
Temperature display in F
Set pause time between wiper cycles 120 min 1 ... 7200 min B3
Off; Input
0s
0 ... 2000 s
0s
0 ... 2000 s
50%
0 ... 100%
Low; High
Z1
Z2
Z3
Z4
Z5
Z6
O333
Enter 0/4-mA value 0.0 NTU; 0.0 FNU; 0.0 ppm (mg/l); 0.0 g/l; 0.0 kg/l; 0.0 t/m3 0.0 %; 0.0 C
Enter 20-mA value 9999 NTU; 10.00 FNU; 9999 ppm (mg/l); 300.0 g/l; 99.99 kg/l 99.99 t/m3 O313 10.0 %; 100.0 C
O3 (1)
O2
O1
lin = linear
O311
O312
86
Endress+Hauser
Appendix
UM253E11.CHP
Entry of correct process temperature current meas. value 5.0 ... 100.0C B7
Feedforward control = 1 at
50% 0 100%
Z7
O236
Endress+Hauser
87
Appendix
Function group
ALARM
F
Function group
CHECK
P
Switch alarm threshold on or off Off; Low; High; Lo+Hi; P1 Lo!; Hi!; LoHi!
Select process monitoring Off; AC; CC; AC+CC AC!; CC!; ACCC!
Set max. perm. period for lower limit exceeded 60 min 0 ... 2000 min
P2
P3
P4
P5
P6
Limit contactor configuration Clean = R2 (5) Chemoclean (only with rel3 and rel4)
Start pulse selecction int = internal; ext = external; i+ext = internal+external; i+stp = internal suppressed by ext R252
Set minimum pause time 120 min 1 ... 3600 min R244
Function R2 (3) Switch off or on PID controller R2 (3) off; on; Basic; PID+B R231
Entry of set value 0 NTU / FNU / ppm / mg/l; 0 g/l; 0 % entire measuring range R232
Entry of integral action time Tn (0.0 = no I component) 0.0 min 0.0 ... 999.9 min R234
Function group
RELAYS
LC PV = TU / TS Limit contactor
R2 (1)
Select contact switch-on point 9999 NTU / FNU / ppm / mg/l; 300.0 g/l; 200.0 % entire measuring range
Select contact switch-off point 9999 NTU / FNU / ppm /mg/l; 300 g/l; 200 % entire measuring range
R212
R213
Funkction group
CONCENTRATION MEASUREMENT
K
Entry of turbidity value 0 NTU / FNU / ppm / mg/l / g/l / % entire measuring range K6
K1
K2
Function group
SERVICE
Hold configuration s+c=during setup and calibration CAL =during calibration Setup =during setup none =no hold S2
S5
S6
E143
E144
MainB = mainboard
E133
E134
Trans = transmitter
E123
E124
Function group
E + H SERVICE
Contr = controller E
E113
E114
Function group
Entry of address
INTERFACE
HART: 0 ... 15 PROFIBUS: 1 ... 126 I I1
I2
88
Endress+Hauser
Appendix
UM253E11.CHP
Automatic start of cleaning function no; yes (not always displayed, see error messages)
F8
next=next error R
F9
Set max. perm. period for lower limit exceeded 60 min 0 ... 2000 min
P7
P8
Set interval between two cleaning cycles (pause time) 360 min 1 ... 7200 min
R257
R258
R259
Selection Ien = pulse length freq = pulse frequency curr = current output 2 R237
Entry of max. pulse frequency 120 1/min 60 ... 180 1/min R239
Setting of alarm threshold (as an absolute value) 100.0 C 5.0 ... +100.0 C R226
Display of LC status
Setting of alarm threshold (as an absolute value) 9999 FNU; 9999 ppm (mg/l); 300.0 g/l; 200.0 % entire measuring range
Display of LC status
MAX MIN
R216
R217
K7
K8
Reset of instrument to basic values no Sens = sensor data; Facty = factory settings
S7
S8
S10
Endress+Hauser
89
Index
12
Index
D D component . . . . . . . . . . . . . . . . . . . . . . . . 44 Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Declaration of conformity . . . . . . . . . . . . . . . . 5 Definition of terms . . . . . . . . . . . . . . . . . . . . 69 Derivative action time. . . . . . . . . . . . . . . . . . 44 Diagnosis . . . . . . . . . . . . . . . . . . . . . . . . 69-79 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Disassembly of CUM 223 . . . . . . . . . . . . . . 72 Disassembly of CUM 253 . . . . . . . . . . . . . . 75 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Dropout delay . . . . . . . . . . . . . . . . . . . . 42, 43 E E+H SERVICE (E). . . . . . . . . . . . . . . . . . . . . 51 E-Function group . . . . . . . . . . . . . . . . . . . . . 51 Economy function . . . . . . . . . . . . . . . . . . . . 47 Electrical connection. . . . . . . . . . . . . . . . . . 12 Electronics box . . . . . . . . . . . . . . . . . . . . . . . 7 Error codes . . . . . . . . . . . . . . . . . . . . . . . . . 67 Error current . . . . . . . . . . . . . . . . . . . . . . 33, 34 Error log . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 Extension cable CYK 8 . . . . . . . . . . . . . . . . 80 Extension of measuring cable . . . . . . . . . . . 14 F F-Function group . . . . . . . . . . . . . . . . . . 33, 34 Factory settings . . . . . . . . . . . . . . . . . . . . . . 23 Freezing of outputs . . . . . . . . . . . . . . . . . . . 21 Function coding . . . . . . . . . . . . . . . . . . . . . . 22 Function group. . . . . . . . . . . . . . . . . . . . . . . 20 Function of keys . . . . . . . . . . . . . . . . . . . . . . 17 Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 G General. . . . . . . . . . . . . . . . . . . . . . . . . . . . 2, 3 General safety notes . . . . . . . . . . . . . . . . . . . 4 H Hardware version. . . . . . . . . . . . . . . . . . . . . 51 HART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Hold dwell period. . . . . . . . . . . . . . . . . . 41, 50 Hold function . . . . . . . . . . . . . . . . . . . . . . . . 21 I I component . . . . . . . . . . . . . . . . . . . . . . . . . 44 I-Function group. . . . . . . . . . . . . . . . . . . . . . 52 Immunity to interference . . . . . . . . . . . . . . . . 5 Installation . . . . . . . . . . . . . . . . . . . . . . 4, 6-15 Instrument configuration . . . . . . . . . . . . 22-36 Instrument upgrade . . . . . . . . . . . . . . . . . . . 81 Integral action time . . . . . . . . . . . . . . . . . . . 44 Intended application . . . . . . . . . . . . . . . . . . . 4 INTERFACE (I) . . . . . . . . . . . . . . . . . . . . . . . 52 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 K K-Function group . . . . . . . . . . . . . . . . . . 48, 49 Kp. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
! 0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17, 20 22 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 9999 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17, 20 A A-Function group . . . . . . . . . . . . . . . . . . 24, 25 Access codes . . . . . . . . . . . . . . . . . . . . . . . . 20 Accessories . . . . . . . . . . . . . . . . . . . . . . 80, 81 Actuating signal outputs . . . . . . . . . . . . . . . . 39 ALARM (F) . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Alarm contact . . . . . . . . . . . . . . . . . . . . . 23, 33 Alarm delay . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Alarm threshold. . . . . . . . . . . . . . . . . . . . 42, 43 Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . 80 Assembly of CUM 223 . . . . . . . . . . . . . . . . . 72 Assembly of CUM 253 . . . . . . . . . . . . . . . . . 75 AUTO key . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 B B-Function group . . . . . . . . . . . . . . . . . . 25, 26 Basic load . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 C C-Function group . . . . . . . . . . . . . . . . . . 56-60 Cable length . . . . . . . . . . . . . . . . . . . . . . . . . 14 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . 53 CALIBRATION (C) . . . . . . . . . . . . . . . . . 22, 56 Calibration data records . . . . . . . . . . . . . . . . 54 Calibration mode . . . . . . . . . . . . . . . . . . 19, 22 CHECK (P) . . . . . . . . . . . . . . . . . . . . . . . . . . 35 Chemoclean function . . . . . . . . . . . . . . . 37, 41 Cleaning function . . . . . . . . . . . . . . . . . . . . . 45 Cleaning time . . . . . . . . . . . . . . . . . . . . . 41, 46 Code 0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Code 22 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Code 9999 . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Coding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Coding structure . . . . . . . . . . . . . . . . . . . . . . 22 Complete Installation . . . . . . . . . . . . . . . . . . . 6 Complete measuring system . . . . . . . . . . . . . 6 CONCENTRATION (K) . . . . . . . . . . . . . . 22, 48 Concentration measurement . . . . . . . . . . . . 48 Configure hold . . . . . . . . . . . . . . . . . . . . . . . 49 Connection diagram . . . . . . . . . . . . . . . . . . . 12 Connection examples . . . . . . . . . . . . . . . . . . 15 Connections of field instrument . . . . . . . . . . 13 Connections of panel-mounted instrument . 13 Contact selection . . . . . . . . . . . . . . . . . . . . . 33 Control characteristic . . . . . . . . . . . . . . . . . . 40 Controller characteristic . . . . . . . . . . . . . . . . 44 Corrective maintenance . . . . . . . . . . . . . 69-79 Corrective maintenance CUM 223 . . . . . . . . 71 Corrective maintenance CUM 253 . . . . . . . . 74 Corrective maintenance of measuring system . . . . . . . . . . . . . . . . . . 78 Current input . . . . . . . . . . . . . . . . . . . . . . . . . 27 CURRENT INPUT (Z) . . . . . . . . . . . . . . . . . . 29 CURRENT OUTPUT (O) . . . . . . . . . . . . . 22, 31 Current output table . . . . . . . . . . . . . . . . . . . 32 Current outputs . . . . . . . . . . . . . . . . . . . . . . . 30 CYK 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
90
Endress+Hauser
Index
UM253E12.CHP
L LED Display . . . . . . . . . . . . . . . . . . . . . . . . . 16 Limit contactor . . . . . . . . . . . . . . . . . . . . 37, 41 Limit contactor for temperature . . . . . . . 37, 43 Limit contactor for turbidity measured value 37 Linear characteristic . . . . . . . . . . . . . . . . . . 31 Liquid crystal display . . . . . . . . . . . . . . . . . . 16 Locking the keypad . . . . . . . . . . . . . . . . 17, 20 M Maintenance . . . . . . . . . . . . . . . . . . . . . 64-68 Management of calibration data records . . 54 Measurement mode . . . . . . . . . . . . . . . . . . . 22 Measuring cable . . . . . . . . . . . . . . . . . . . . . 14 Measuring cable connection . . . . . . . . . . . . 14 Measuring system . . . . . . . . . . . . . . . . . . . . . 6 Menu structure . . . . . . . . . . . . . . . . . . . . . . . 20 Minimum pause time . . . . . . . . . . . . . . . 46, 47 MINUS key . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Module name . . . . . . . . . . . . . . . . . . . . . . . . 51 Monitoring features . . . . . . . . . . . . . . . . . . . . 5 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Mounting examples . . . . . . . . . . . . . . . . . . . 10 Mounting of field instrument . . . . . . . . . . . . . 8 N N-Function group . . . . . . . . . . . . . . . . . . . . . 62 O O-Function group. . . . . . . . . . . . . . . . . . 31, 32 Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61 OFFSET (V) . . . . . . . . . . . . . . . . . . . . . . . . . 61 Offset mode . . . . . . . . . . . . . . . . . . . . . . . . . 22 ON time . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 Operating concept. . . . . . . . . . . . . . . . . . . . 19 Operating keys. . . . . . . . . . . . . . . . . . . . . . . 17 Operating modes . . . . . . . . . . . . . . . . . . . . . 19 Operation . . . . . . . . . . 4, 16, 17, 18, 19, 20, 21 Operator interface . . . . . . . . . . . . . . . . . . . . 16 Optoscope . . . . . . . . . . . . . . . . . . . . . . . . . . 77 Order code. . . . . . . . . . . . . . . . . . . . . . . . 3, 50
P P controller . . . . . . . . . . . . . . . . . . . . . . . . . 38 P(ID) controller . . . . . . . . . . . . . . . . . . . 37, 38 P-Function group. . . . . . . . . . . . . . . . . . . . . 35 Packaging . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Panel-mounted instrument . . . . . . . . . . . . . 11 Partition plate. . . . . . . . . . . . . . . . . . . . . . . . . 7 Pause time. . . . . . . . . . . . . . . . . . . . 40, 46, 47 PCS alarm . . . . . . . . . . . . . . . . . . . . . . . . . . 35 PD controller . . . . . . . . . . . . . . . . . . . . . . . . 38 PI controller . . . . . . . . . . . . . . . . . . . . . . . . . 38 Pickup delay . . . . . . . . . . . . . . . . . . . . . 42, 43 PID controller. . . . . . . . . . . . . . . . . . . . . 38, 44 PLUS key . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Post mounting kit. . . . . . . . . . . . . . . . . . . . . . 9 Post mounting on a square post . . . . . . . . . . 8 Post mounting on cylindrical pipes. . . . . . . . 8 Post-rinse time. . . . . . . . . . . . . . . . . . . . 41, 46 Pre-rinse time . . . . . . . . . . . . . . . . . . . . 41, 46 Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 Product structure. . . . . . . . . . . . . . . . . . . . . . 3 PROFIBUS . . . . . . . . . . . . . . . . . . . . . . . . . . 52 Programming. . . . . . . . . . . . . . . . . . . . . . . . 20 Pulse frequency . . . . . . . . . . . . . . . . . . . . . 45 Pulse frequency modulation . . . . . . . . . . . . 39 Pulse length. . . . . . . . . . . . . . . . . . . . . . . . . 45 Pulse length modulation . . . . . . . . . . . . . . . 39 Q Queries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 R R-Function group . . . . . . . . . . . . . . . . . 41- 46 REL key . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 RELAY (R) . . . . . . . . . . . . . . . . . . . . . . . 22, 41 Relay contact configuration . . . . . . . . . 36, 37 Repeat cycles . . . . . . . . . . . . . . . . . . . . . . . 47 Replacement of processor module . . . . . . 77 Residual concrete water . . . . . . . . . . . . . . . 55 Rest period between wiping cycles . . . . . . 25 Rinse time . . . . . . . . . . . . . . . . . . . . . . . . . . 45 RM junction box . . . . . . . . . . . . . . . . . . . . . 80
Endress+Hauser
91
Index
S S-Function group . . . . . . . . . . . . . . . . . . 49, 50 Safety features . . . . . . . . . . . . . . . . . . . . . . . . 5 Safety instructions. . . . . . . . . . . . . . . . . . . . 4, 5 Safety notes . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . 2 Scopeware . . . . . . . . . . . . . . . . . . . . . . . . . . 77 Selection of error. . . . . . . . . . . . . . . . . . . . . . 33 Selection of language . . . . . . . . . . . . . . . . . . 49 Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Sensor installation. . . . . . . . . . . . . . . . . . . . . 14 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 Serial number . . . . . . . . . . . . . . . . . . . . . . . . 51 Service . . . . . . . . . . . . . . . . . . . . . . . . . . 49, 69 SERVICE (S) . . . . . . . . . . . . . . . . . . . . . . 24, 49 SETUP 1 (A) . . . . . . . . . . . . . . . . . . . . . . . . . 24 SETUP 2 (B) . . . . . . . . . . . . . . . . . . . . . . 24, 25 Setup mode. . . . . . . . . . . . . . . . . . . . . . . . . . 22 Setup-Modus . . . . . . . . . . . . . . . . . . . . . . . . 19 Shipping documents. . . . . . . . . . . . . . . . . . . . 2 Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Slope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 SLOPE (N). . . . . . . . . . . . . . . . . . . . . . . . . . . 62 Software version . . . . . . . . . . . . . . . . . . . . . . 51 Spare parts for CUM 223 . . . . . . . . . . . . . . . 73 Spare parts for CUM 253 . . . . . . . . . . . . . . . 76 Spare parts orders . . . . . . . . . . . . . . . . . . . . 77 Start pulse . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Start-up menu . . . . . . . . . . . . . . . . . . . . . . . . 24 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Structure of measuring cable . . . . . . . . . . . . 14 SW upgrade . . . . . . . . . . . . . . . . . . . . . . . . . 50 Switch-off point of contact . . . . . . . . . . . . . . 42 Switch-off temperature . . . . . . . . . . . . . . . . . 43 Switch-on point of contact . . . . . . . . . . . . . . 42 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . 2 System configuration . . . . . . . . . . . . . . . . . . 24
T Technical data . . . . . . . . . . . . . . . . . 82, 83, 84 Terminal blocks . . . . . . . . . . . . . . . . . . . . . . . 7 Termination of measuring cablel . . . . . . . . . 14 Timer for cleaning function . . . . . . . . . . 37, 40 Tn . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Troubleshooting . . . . . . . . . . 64, 65, 66, 67, 68 Tv . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 U Universal mounting post CYY 102 . . . . . . . . . 9 Unlocking the keypad . . . . . . . . . . . . . . 17, 20 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Upright post CYY 102 . . . . . . . . . . . . . . . . . 80 V V-Function group . . . . . . . . . . . . . . . . . . . . . 61 VBM junction box . . . . . . . . . . . . . . . . . . 14, 80 Version TB . . . . . . . . . . . . . . . . . . . . . . . . . . 55 W Wall mounting . . . . . . . . . . . . . . . . . . . . . . . . 8 Weather protection cover CYY 101 . . . . . 8, 80 Wiper controller . . . . . . . . . . . . . . . . . . . . . . 25 Wiping cycle. . . . . . . . . . . . . . . . . . . . . . . . . 25 Z Z-Function group . . . . . . . . . . . . . . . . . . . . . 29
92
Endress+Hauser
-KH FA
Austria Endress+Hauser Ges.m.b.H. Wien Tel. ++43 (1) 8 80 56-0, Fax (1) 8 80 56-35 Belarus Belorgsintez Minsk Tel. ++3 75 (1 72) 26 31 66, Fax (1 72) 26 31 11 Belgium / Luxembourg Endress+Hauser S.A./N.V. Brussels Tel. ++32 (2) 2 48 06 00, Fax (2) 2 48 05 53 Bulgaria INTERTECH-AUTOMATION Sofia Tel. ++3 59 (2) 66 48 69, Fax (2) 9 63 13 89 Croatia Endress+Hauser GmbH+Co. Zagreb Tel. ++3 85 (1) 6 63 77 85, Fax (1) 6 63 78 23 Cyprus I+G Electrical Services Co. Ltd. Nicosia Tel. ++3 57 (2) 48 47 88, Fax (2) 48 46 90 Czech Republic Endress+Hauser GmbH+Co. Praha Tel. ++4 20 (26) 6 78 42 00, Fax (26) 6 78 41 79 Denmark Endress+Hauser A/S Sborg Tel. ++45 (70) 13 11 32, Fax (70) 13 21 33 Estonia Elvi-Aqua Tartu Tel. ++3 72 (7) 42 27 26, Fax (7) 42 27 27 Finland Endress+Hauser Oy Espoo Tel. ++3 58 (9) 8 59 61 55, Fax (9) 8 59 60 55 France Endress+Hauser Huningue Tel. ++33 (3) 89 69 67 68, Fax (3) 89 69 48 02 Germany Endress+Hauser Metechnik GmbH+Co. Weil am Rhein Tel. ++49 (76 21) 9 75 01, Fax (76 21) 9 7 55 55 Great Britain Endress+Hauser Ltd. Manchester Tel. ++44 (1 61) 2 86 50 00, Fax (1 61) 9 98 18 41 Greece I & G Building Services Automation S.A. Athens Tel. ++30 (1) 9 24 15 00, Fax (1) 9 22 17 14 Hungary Mile Ipari-Elektro Budapest Tel. ++36 (1) 2 61 55 35, Fax (1) 2 61 55 35 Iceland Vatnshreinsun HF Reykjavik Tel. ++3 54 (5) 61 96 16, Fax (5) 61 96 17 Ireland Flomeaco Company Ltd. Kildare Tel. ++3 53 (45) 86 86 15, Fax (45) 86 81 82 Italy Endress+Hauser Italia S.p.A. Cernusco s/N Milano Tel. ++39 (02) 92 10 64 21, Fax (02) 92 10 71 53 Latvia Raita Ltd. Riga Tel. ++3 71 (7) 31 28 97, Fax (7) 31 28 94 Lithuania Agava Ltd. Kaunas Tel. ++3 70 (7) 20 24 10, Fax (7) 20 74 14 Netherlands Endress+Hauser B.V. Naarden Tel. ++31 (35) 6 95 86 11, Fax (35) 6 95 88 25
Norway Endress+Hauser A/S Tranby Tel. ++47 (32) 85 98 50, Fax (32) 85 98 51 Poland Endress+Hauser Polska Sp. z o.o. Warszawy Tel. ++48 (22) 7 20 10 90, Fax (22) 7 20 10 85 Portugal Tecnisis - Tecnica de Sistemas Industriais Linda-a-Velha Tel. ++3 51 (1) 4 17 26 37, Fax (1) 4 18 52 78 Romania Romconseng SRL Bucharest Tel. ++40 (1) 4 10 16 34, Fax (1) 4 10 16 34 Russia Endress+Hauser Moscow Office Moscow Tel.++7 09 (5) 1 58 75 71, Fax (5) 1 58 98 64 Slovak Republic Transcom Technik s.r.o. Bratislava Tel. ++4 21 (74) 4 88 86 84, Fax (74) 4 88 71 12 Slovenia Endress+Hauser D.O.O. Ljubljana Tel. ++3 86 (61) 1 59 22 17, Fax (61) 1 59 22 98 Spain Endress+Hauser S.A. Barcelona Tel. ++34 (93) 4 80 33 66, Fax (93) 4 73 38 39 Sweden Endress+Hauser AB Sollentuna Tel. ++46 (8) 55 51 16 00, Fax (8) 55 51 16 00 Switzerland Endress+Hauser AG Reinach/BL 1 Tel. ++41 (61) 7 15 75 75, Fax (61) 7 11 16 50 Turkey Intek Endstriyel lc ve Kontrol Sistemleri Istanbul Tel. ++90 (2 12) 2 75 13 55, Fax (2 12) 2 66 27 75 Ukraine Industria Ukrana Kiev Tel. ++3 80 (44) 2 68 81, Fax (44) 2 69 08 Yugoslavia Meris d.o.o. Beograd Tel. ++3 81 (11) 4 44 61 64, Fax (11) 4 44 19 66
Brazil Samson Endress+Hauser Ltda. Sao Paulo Tel. ++55 (11) 5 0 31 34 55, Fax (11) 5 0 31 30 67 Canada Endress+Hauser Ltd. Burlington, Ontario Tel. ++1 (9 05) 6 81 92 92, Fax (9 05) 6 81 94 44 Chile DIN Instrumentos Ltda. Santiago Tel. ++56 (2) 2 05 01 00, Fax (2) 2 25 81 39 Colombia Colsein Ltd.Bogota D.C. Tel. ++57 (1) 2 36 76 59, Fax (1) 6 10 78 68 Costa Rica EURO-TEC S.A. San Jose Tel. ++5 06 (2) 96 15 42, Fax ( 2 ) 96 15 42 Ecuador Insetec Cia. Ltda. Quito Tel. ++5 93 (2) 26 91 48, Fax (02) 46 18 33 Guatemala ACISA Automatizacion Y Control Industrial S.A. Ciudad de Guatemala, C.A. Tel. ++5 02 (3) 34 59 85, Fax (2) 32 74 31 Mexico Endress+Hauser I.I. Mexico City Tel. ++52 (5) 56 89 65, Fax (5) 56 84 18 Paraguay Incoel S.R.L. Asuncion Tel. ++5 95 (21) 21 39 89, Fax (21) 22 65 83 Uruguay Circular S.A. Montevideo Tel. ++5 98 (2) 92 57 85, Fax (2) 92 91 51 USA Endress+Hauser Inc. Greenwood, Indiana Tel. ++1 (3 17) 5 35 71 38, Fax (3 17) 5 35 84 89 Venezuela H. Z. Instrumentos C.A. Caracas Tel. ++58 (2) 9 44 09 66, Fax (2) 9 44 45 54
Papua New Guinea SBS Electrical Pty Limited Port Moresby Tel. ++6 75 (3) 25 11 88, Fax (3) 25 95 56 Philippines Brenton Industries Inc. Makati Metro Manila Tel. ++63 (2) 6 38 80 41, Fax (2) 6 38 80 42 Singapore Endress+Hauser (S.E.A.) Pte., Ltd. Singapore Tel.++65 (5) 66 82 22, Fax (2) 66 68 48 South Korea Endress+Hauser (Korea) Co., Ltd. Seoul Tel. ++82 (2) 6 58 72 00, Fax (2) 6 59 28 38 Taiwan Kingjarl Corporation Taipei R.O.C. Tel. ++886 (2) 27 18 39 38, Fax (2) 27 13 41 90 Thailand Endress+Hauser Ltd. Bangkok Tel. ++66 (2) 9 96 78 11 20, Fax (2) 9 96 78 10 Vietnam Tan Viet Bao Co. Ltd. Ho Chi Minh City Tel. ++84 (8) 8 33 52 25, Fax (8) 8 33 52 27
Iran Telephone Technical Services Co. Ltd. Tehran Tel. ++98 (21) 8 74 67 50, Fax (21) 8 73 72 95 Israel Instrumetrics Industrial Control Ltd. Tel-Aviv Tel. ++9 72 (3) 6 48 02 05, Fax (3) 6 47 19 92 Jordan A.P . Parpas Engineering S.A. Amman Tel. ++9 62 (6) 4 64 32 46, Fax (6) 4 64 57 07 Kingdom of Saudi Arabia Anasia Jeddah Tel. ++9 66 (2) 6 71 00 14, Fax (2) 6 72 59 29 Kuwait Kuwait Maritime & Mercantile Co. K.S.C. Safat Tel.++9 65 (2) 44 14 81, Fax (2) 44 14 86 Lebanon Nabil Ibrahim Jbeil Tel. ++9 61 (3) 25 40 52, Fax (9) 54 80 38 Sultanate of Oman Mustafa & Jawad Science & Industry Co. L.L.C. Ruwi Tel. ++9 68 (60) 20 09, Fax (60) 70 66 United Arab Emirates Descon Trading EST. Dubai Tel. ++9 71 (4) 65 36 51, Fax (4) 65 32 64 Yemen Yemen Company for Ghee and Soap Industry Taiz Tel. ++9 76 (4) 23 06 64, Fax (4) 21 23 38
)BHE?=
Egypt Anasia Heliopolis/Cairo Tel. ++20 (2) 4 1 79 00, Fax (2) 4 1 79 00 Morocco Oussama S.A. Casablanca Tel. ++2 12 (2) 24 13 38, Fax (2) 40 26 57 Nigeria J F Technical Invest. Nig. Ltd. Lagos Tel. ++2 34 (1) 62 23 45 46, Fax (1) 62 23 45 48 South Africa Endress+Hauser Pty. Ltd. Sandton Tel. ++27 (11) 4 44 13 86, Fax (11) 4 44 19 77 Tunisia Controle, Maintenance et Regulation Tunis Tel. ++2 16 (1) 79 30 77, Fax (1) 78 85 95
)IE=
China Endress+Hauser Shanghai Instrumentation Co. Ltd. Shanghai Tel. ++86 (21) 54 90 23 00, Fax (21) 54 90 23 03 Endress+Hauser Beijing Office Beijing Tel. ++86 (10) 68 34 40 58, Fax (10) 68 34 40 68 Endress+Hauser (H.K.) Ltd. Hong Kong Tel. ++8 52 (2) 5 28 31 20, Fax (2) 8 65 41 71 India Endress+Hauser India Branch Office Mumbai Tel. ++91 (22) 8 52 14 58, Fax (22) 8 52 19 27 Indonesia PT Grama Bazita Jakarta Tel. ++62 (21) 7 97 50 83, Fax (21) 7 97 50 89 Japan Sakura Endress Co., Ltd. Tokyo Tel. ++81 (4 22) 54 06 11, Fax (4 22) 55 02 75 Malaysia Endress+Hauser (M) Sdn. Bhd. Petaling Jaya, Selangor Darul Ehsan Tel. ++60 (3) 7 33 48 48, Fax (3) 7 33 88 00 Pakistan Speedy Automation Karachi Tel. ++92 (21) 7 72 29 53, Fax (21) 7 73 68 84
)KIJH= E=
AM A= = @
Australia ALSTOM Australia Ltd. Sydney Tel. ++61 (2) 97 22 47 77, Fax (2) 97 22 48 88 New Zealand EMC Industrial Instrumentation Auckland Tel. ++64 (9) 4 15 51 10, Fax (9) 4 15 51 15
) AHE?=
Argentina Endress+Hauser Argentina S.A. Buenos Aires Tel. ++54 (1) 1 45 22 79 70, Fax (1) 1 45 22 79 09 Bolivia Tritec S.R.L. Cochabamba Tel. ++5 91 (42) 5 69 93, Fax (42) 5 09 81
JDAH ? K JHEAI
Endress+Hauser GmbH+Co. Instruments International D-Weil am Rhein Germany Tel. ++49 (76 21) 9 75 02, Fax (7 6 21) 97 53 45
51500274
BA 200C/07/en/07.02 Printed in Germany / CV5 / DT
Endress Hauser
The Power of Know How