Teslameter CYHT20-User's Manual
Teslameter CYHT20-User's Manual
1) Characteristics
Digital Gauss/Tesla Meter - A low-cost measuring device, which is easy to operate, portable
and convenient to handle and store.
- Ideal for quick quality checks and comparative measurements,
CYHT20 with built-in polarity display.
2) Technical Data
User’s Manual
Measuring ranges: 0~200mT and 0-2000mT
Measuring accuracy: ±2.0%
Resolution: 0.1mT, 1.0mT
Display: 3 ½ LCD
Display Unit: mT (1mT=10Gs)
Ambient temperature: 0°C ~ 50°C
Storage temperature: -20°C ~ +70°C
Relative humidity: 20% ~ 80%
Power supply: +9V alkaline, block battery
Dimensions: 160mm x 88mm x 36mm
Weight: 300g (only instrument)
3) Accessories
2
I: bias working current in amperes
The initial use of this discovery was for the classification of chemical
samples. The development of indium arsenide semiconductor
CYAP-D60: axial probe Ø6 x 40mm compounds in the 1950's led to the first useful Hall Effect magnetic
instruments. Hall Effect sensors allowed the measurement of DC or
static magnetic fields with requiring motion of the sensor.
6) Measuring Method
5) Measuring Principle The magnetic lines of the measured magnetic field should
perpendicularly pass through the Hall Effect element of the Hall probe.
The Gauss/Tesla meter works with a Hall Effect sensor/probe. There are
two kinds of Hall probes: transverse and axial probes, which are used
for different magnetization directions. The sensing plan of the Hall
sensor/probe should be
positioned perpendicular to the
magnetic field
Hall Effect and its Electrical Parameters Put the concave side, i.e. the side with a small circle mark of the probe
on the surface of the measured magnet or at the measuring point of a
magnetic field carefully.
According to the Hall Effect, a voltage can be measured at right angle to
the current path when a conductor or semiconductor with current flowing
in the direction is introduced perpendicular to a magnetic field. The Hall
voltage can be calculated from:
VH = k H I H B
3 4
7) Part and Functions Back Panel
Front Panel
Convex
Power
Socket
Battery
Cal Compartment
Zero
Protective tube
Cal ON
Meas. OFF
Concave
CYHT20 Digital
Gauss/Tesla Meter
5 6
1) Socket-outlet: to connect the Hall sensor/probe to the measuring 2) Connect the Hall sensor/probe to the measuring instrument
instrument 3) Switch on the power of measuring instrument by pressing button (5)
2) LCD display: to display the field strength (0-199.9mT or 0-1999mT) 4) Select the calibration by pressing the button (7). Check the display
and pole direction (“+” north pole, “-“ south pole) and adjust the calibration rheostat until the display value is identical
3) Calibration rheostat: to adjust the current of the Hall sensor to the calibration value (8) on the Hall probe given by the
according to the calibration value (8) on the Hall probe manufacturer (the point should be not considered)
4) Zero adjustment: to adjust the zero point. Please keep the Hall 5) Select the measurement by releasing the button (7), check the zero
sensor far away from magnetic field and adjust the zero rheostat point, and adjust the zero rheostat if the display is not zero when the
until displaying “0” Hall sensor is far away from magnetic field
5) Power switch: to switch ON/OFF the measuring instrument 6) Select the measuring range 0-200mT by releasing the button (6) or
6) Range select: to select the measuring range 0-200mT or 0-2000mT 0-2000mT by pressing the button (6)
7) Calibration/Measurement: to select sensor calibration and 7) Take off the protective tube of the Hall sensor/probe, position the
measurement measuring circle mark (10) on the surface of the measuring object
8) Calibration value: using for adjusting the current of Hall sensor and (such as permanent magnet), and read the display value (measuring
calibrating the sensor system (note: different Hall probe has value and pole display “+” or “-“). Note: the measuring circle mark
different calibration value. You must calibrate your sensor system (10) is measuring point of the Hall sensor. The pole is N pole if the
after you have changed the Hall probe) display is “+” when this mark orients to the measuring object.
9) Plug of Hall probe: to connect the Hall probe to the measuring 8) After the measurement please put the protective tube on the Hall
instrument probe in order to protect the Hall element and switch off the power
10) Measuring circle mark: the side with this mark is the measuring 9) The Hall probes delivered are compatible. However, don’t forget to
side of the Hall sensor. You should use this side to measure the calibrate the sensor system using the calibration value (8) after
surface magnetic field of permanent magnet. This side should be having changed the Hall probe.
positioned perpendicular to the magnetic field
9) Warranty
Hall probe CYTP-T08 Measuring instrument: 12 months
Hall sensor/probe: no guarantee. However we offer you
Replacement for reduced price (10%
Using this side (without discount)
soldered connection) for
measurement
10) Service
Am Lindenhain 17A
8) Measuring Procedure D-85435 Erding
Tel. +49-(0)8122-2274508
1) Install the +9V battery to the battery compartment or connect the Fax: +49-(0)8122-2274509
+9V power supply to the measuring instrument Email: [email protected]
7 8