BK Precision Function Genarator
BK Precision Function Genarator
Windows
TM
(models 4084AWG and 4086AWG only)
Convenient data input via knob or numerical keypad.
Bright, easy to read display using VFD (Visible Vacuum Fluorescent) technology
Fully programmable via SCPI compatible command set
100 MHz Universal Counter with frequency measurement and totalize function
4080 series Function & Arbitrary/Function Generators Instruction Manual
6
2. SPECIFICATIONS
2.1 Function Generator
Waveform Characteristics
Main Waveforms: Sine, square
Waveform Amplitude resolution: 12 bits
Sample Rate: 200MSa/s (4084, 4084AWG, 4085, 4086, 4086AWG)
300MSa/s (4087)
Sine:
Harmonic Distortion of Sine Wave:
- 50dBc (frequency 5MHz)
- 45dBc (frequency 10MHz)
- 40dBc (frequency 20MHz)
- 35dBc (frequency 40MHz)
- 30dBc (frequency >40MHz)
Total Harmonic distortion: 0.1% (20Hz ~100kHz)
Square:
Rise and Fall Time of Square Wave: 15ns
Note: Test conditions for harmonic distortion, sine distortion, rise/fall time: Output
Amplitude 2Vp-p, Environmental temperature: 255
Build in standard and complex (arbitrary) waveforms:
27 build-in standard and complex waveforms.
Sine, Square, Triangle, Positive Ramp, Falling Ramp, Noise, Positive Pulse, Negative
Pulse, Positive DC, Negative DC, Stair wave, Coded Pulse, Full wave rectified,
Half-wave rectified, Sine transverse cut, Sine vertical cut, Sine phase modulation,
Logarithmic, Exponential, Half-round, SINX/X, Square root, Tangent, Cardiac,
Earthquake, Combination
Waveform Length: 4096 dots
Amplitude Resolution: 10 bits
Pulse Wave:
Duty Cycle: 0.1% ~99.9% (below 10kHz), 1% ~99% (10kHz ~100kHz)
Rise/Fall Time: 100ns Duty cycle 20%
DC signal characteristics:
DC range: 10mV 10V (high impedance)
DC Accuracy: 5% of setting +10mV (high impedance)
Module for user defined arbitrary waveform generation (models 4084AWG and 4086AWG only)
Number of memory locations for arbitrary waveforms: 8
Length of waveforms8~16000 points
Resolution of waveform amplitude10 bits
Frequency range100Hz~100kHz
4080 series Function & Arbitrary/Function Generators Instruction Manual
7
Frequency Characteristics
Frequency Range:
Main waveforms (sine, square):
Model 4084/4084AWG: 1Hz ~20MHz
Model 4085: 1Hz ~40MHz
Model 4086/4086AWG 1Hz ~80MHz (sine wave)
1Hz ~40MHz (square wave)
Model 4087 1Hz ~120MHz (sine wave)
1Hz ~40MHz (square wave)
All other waveforms:
All models: 1Hz ~100kHz
Frequency Stability: 110
-6
(225
Resolution: 1Hz
Frequency Accuracy: 510
-6
(225
Data entry Units: s, ms, Hz, kHz, MHz
Amplitude Characteristics
Amplitude Range:
4084, 4084AWG: 2mV ~ 20Vpp (open circuit), 1mV ~ 10Vpp (50)
4085
4086, 4086AWG: for Freq 40MHz: 2mV ~ 20Vpp (open circuit), 1mV ~ 10Vpp (50)
for Freq >40MHz: 2mV ~4Vp-p (open circuit), 1mV ~ 2Vpp (50)
4087: for Freq 40MHz: 2mV ~ 20Vpp (open circuit), 1mV ~ 10Vpp (50)
for Freq >40MHz: -76dBm ~+13.5 dBm50or 0.1mV ~3Vpp50
Max. Resolution: 2Vpp (open circuit), 1Vpp (50)
Amplitude Accuracy: 1%+0.2mV (sine wave relative to 1kHz)
Amplitude Stability: 0.5 % /3 hours
Flatness:
For amplitude 2Vpp: 3% (frequency5MHz), 10% (5MHz<frequency40MHz)
For amplitude >2Vpp: 5% (frequency5MHz), 10% (5MHz<frequency20MHz)
20% (frequency>20MHz)
Models 4086/AWG, 4087 only: 1dBm (frequency>40MHz)
Output Impedance: 50
Output Units: Vpp, mVpp, Vrms, mVrms, dBm
4080 series Function & Arbitrary/Function Generators Instruction Manual
8
DC Offset Characteristics
Offset Range (open circuit)
Freq 40MHz): 10Vpk ac + dc (Offset 2peak-to peak amplitude)
Freq >40MHz): 2Vpk ac +dc (Offset 2peak-to peak amplitude)
Resolution: 2V (open circuit), 1V (50)
Offset Error: 5% of setting +10mV (Ampl. 2Vpp into open circuit)
5% of setting +20mV (Ampl. >2Vpp into open circuit)
AM Characteristics
Carrier Waveforms: sine or square
Carrier Frequency Range: same as main waveforms
Modulation Source internal or external
Modulating Waveform: 5 internal waveforms (sine, square, triangle, rising/falling ramp)
Frequency of modulating signal: 100Hz ~20kHz
Distortion: 2%
Modulation Depth: 1% ~120%
1% ~80% (frequency>40MHz, Ampl >2Vpp into open circuit)
Modulation Error: 5%+0.2% (100Hz <frequency 10kHz)
10%+2% (10kHz <frequency 20kHz)
Amplitude of ext. input signal: 3Vp-p (-1.5V~+1.5V)
FM Characteristics
Carrier Waveforms sine or square
Carrier Frequency Range: same as main waveforms
Modulation Source: internal or external
Modulating Waveform: 5 internal waveforms (sine, square, triangle, rising/falling ramp)
Frequency of modulating signal: 100Hz ~10kHz
Peak Frequency Deviation: Max. 50% of carrier frequency for internal FM
Max 100kHz (carrier frequency5MHz) for external FM,
with input signal voltage 3Vp-p (-1.5V~+1.5V)
FSK Characteristics
Carrier Waveform sine or square
Carrier Frequency Range: same as main waveforms
Control Mode internal or external trigger (external: TTL level, low level F1,
high level F2)
FSK Rate: 0.1ms ~800s
PSK Characteristics
Waveform: sine or square
Frequency Range: same as main waveforms
PSK: Phase 1 (P1) and Phase 2 (P2), range: 0.0 ~360.0
Resolution: 0.1
PSK rate: 0.1ms ~800s
4080 series Function & Arbitrary/Function Generators Instruction Manual
9
Control Mode: internal or external trigger (external: TTL level, low level P1,
high level P2)
Burst Characteristics
Waveform: sine or square
Frequency Range: same as main waveforms
Burst Counts : 1 ~10000 cycles
Time interval between bursts: 0.1ms ~800s
Control Mode: internal, single or external gated trigger
Frequency Sweep Characteristics
Waveform: sine or square
Start/Stop Freq.: same as main waveforms
Sweep Time: 1ms ~800s (linear), 100ms ~800s (log)
Sweep Mode: Linear or Logarithmic
External trigger signal frequency: DC ~1kHz (linear) DC~10Hz (log)
Control Mode: internal or external trigger
Rear Panel Terminals (for modulation and sweep)
Output MOD OUT
Frequency: 100Hz ~ 20kHz
Waveform: sine, square, triangle, rising/falling ramp
Amplitude: 5Vp-p 5%
Output Impedance: 600
Modulation IN
3 Vpp =100% Modulation
External Input Trig/FSK/Burst
Level: TTL
Main OUTput
Impedance: 50
Protection: Short circuit and overload protected
State Storage Characteristics
Storage Parameters: frequency, amplitude, waveform, DC offset values, modulation
parameters
Storage Capacity: 10 user configurable stored states
Storage Time: more than 10 years
2.2 Universal Counter
Frequency Range
Frequency Measurement: 1Hz ~100MHz
Totalize mode: 50MHz max
4080 series Function & Arbitrary/Function Generators Instruction Manual
10
Input Characteristics
Sensitivity
Input attenuator disabled: 50mVrms (f: 10Hz ~50MHz), 100mVrms (f: 1Hz ~100MHz)
Input attenuator enabled: 0.5Vrms (f: 10Hz ~50MHz), 1Vrms (f: 1Hz ~100MHz)
Max. I nput Voltage Allowed: 100Vp-p (f100kHz), 20Vp-p (1Hz~100MHz)
Input Impedance: R>500k, C<30pF
Coupling: AC
Waveform: sine or square
Low Pass Filter: cut off frequency about 100kHz (with internal attenuation: -3dB)
with external attenuation: -30 dB (f >1MHz)
Gate Time Setting: 10ms ~10s continuously adjustable
Display Bits: 8 (for gate Time>5s)
Totalize Capacity: 4.2910
9
Control Mode: manual or external gate control
Accuracy: time base error trigger error (when signal SNR >40dB,
trigger error 0.3)
Time base:
Type: small TCXO
Frequency: 10MHz
Stability: 1 10
-6
(22C5C)
2.3 General
Power Supply 198~242V or 99~121V, Frequency: 47~63Hz
Power Consumption: <35VA
Operating Temperature 0 to +40 C
Operating Humidity 80% R.H
Storage Temperature -40 to 70
Dimensions (W x H x D): 255 mm x 100 mm x 370 mm (10.0 x 3.93 x 14.56 inch)
Weight: 3 kg (6.6 lbs): models 4084, 4084AWG, 4085, 4086, 4086AWG
3.5 kg (7.7 lbs) model 4087
Remote Interface RS232
Accessories included
BNC to alligator cable
BNC to BNC cable
RS232 communication cable
Power line cord
Test report
Spare fuse
Installation disk for Arbitrary Waveform Creation Software (models 4084AWG or 4086AWG only)
NOTE: Specifications and information are subject to change without notice. Please visit www.bkprecision.comfor
the most current product information.
4080 series Function & Arbitrary/Function Generators Instruction Manual
11
3. PANEL DESCRIPTION
3.1 Front Panel
3.1.1 Overview of Keys:
Summary: Most keys have multiple functions. Primary functions are written on the key. Simply
press the respective key to enable a primary functions. All Function/Mode keys and some of the
numerical entry keys have associated secondary functions indicated above each key in blue. To
activate a secondary function, press the shiftkey followed by the desired key. The bottom row
of the Function/Mode keys can be used to enter units directly. The unit is indicated below each key.
To enter a unit, enter a numerical value via the numerical keypad then press the corresponding unit
key. The unit key also serves as Enter function.
Data entry keys:
Key
Name
Main
Function
Secondary
Function
Key
Name
Main function
Secondary
Function
0 Input Digit 0 Enter SW mode 7 Input Digit 7
Enable ARB7
waveform ***.
1 Input Digit 1
Enable ARB1
waveform ***
8 Input Digit 8
Enable ARB8
waveform ***
2 Input Digit 2
Enable ARB2
waveform ***
9 Input Digit 9 Not available
3 Input Digit 3
Enable ARB3
wave ***
Input decimal point Reset Unit
4 Input Digit 4
Enable ARB4
waveform***
Input negative
symbol
Enter system menu
5 Input Digit 5
Enable ARB5
waveform ***
Move arrow key to
left *
Select pulse
6 Input Digit 6
Enable ARB6
waveform***
Move arrow key to
right **
Select Arblist
waveform
*: Direct number entry: Press this key to clear the least significant bit of the displayed number.
Useful for correcting number entry before entry is confirmed with unit key.
External totalize mode: Press this key to stop counting and display present counting value.
Press again to resume counting.
4080 series Function & Arbitrary/Function Generators Instruction Manual
12
**: External totalize mode: Press this key to reset and resume event counter
*** models 4084AWG and 4086AWG only
Function/Mode Keys:
Key name Main Function
Secondary
Function
Secondary
Function for
Counter Mode
Unit Entry
Freq./Period
Toggle between
Frequency &
Period.
Enable Sine Wave Not Available Not Available
Ampl./Pulse
width
Amplitude Select.
Enable Square
Wave.
Not Available Not Available
FSK/PSK
FSK/PSK
Function Select
Enable Triangle
Wave
Not Available Not Available
Menu Menu Selection
Enable positive
ramp
Not Available Not Available
FM Enable FM mode
Enter Storage
menu
Attenuation
Selection
ms, mVpp
AM Enable AM mode Enter Recall menu Low Pass Select MHz, Vrms
Sweep
Enable Sweep
mode
Enter Counter
Mode
Freq. Meas./
Totalize Enable
kHz, mVrms
Burst
Enable Burst
mode
DC Offset Select Gate Select Hz, dBm
Other Keys:
Key Name Main Function Other Function
Output Main OUTPUT signal On/Off
Generate single trigger in sweep and
burst mode
Shift Select secondary function Enter units in s, Vpp, N
3.1.2 Display Annunciators:
Display areas:
Waveform Indication
Main Alphanumeric Display
Frequency Measurement/Totalize Settings
Instrument states
4080 series Function & Arbitrary/Function Generators Instruction Manual
13
Waveform states
Sine waveform is enabled (,main waveform)
Square waveform is enabled (main waveform)
Triangle waveform is enabled
Ramp waveform is enabled
Arb Arbitrary waveform mode is enabled
Frequency Measurement/Totalize states
Filter: Low Pass Filter is enabled
ATT: Input Attenuator is enabled
GATE: Gate received trigger
Function Generator states
Adrs: The instrument is in remote state.
Trig: Function Generator is waiting for a single trigger or external trigger.
FM: FM modulation is enabled
AM: AM modulation is enabled
Sweep: Sweep mode is enabled
Ext: Generator is configured for external signal input
Freq: Frequency measurement function is enabled
Count: Totalize function is enabled
FSK: FSK (frequency shift keying) modulation is enabled
FSK: PSK modulation is enabled
Burst: Burst mode enabled.
Offset: DC offset of output signal is not 0.
Shift: Shift key has been pressed, Shift mode is active
Rmt: The instrument is in remote state.
3.13 Description of front panel keys
shiftkey
Access secondary functions or to enter units seconds, Vpp and N
Numerical key pad:
Keys are used for direct entry of numerical values. Keys0-are assigned secondary
functions to enter the systems menu, reset the unit to its default values or revert to Standard
Waveform (SW) mode.
arrow keys
The primary function is to move the flashing digit left or right or to select the desired arbitrary
waveform from the Arb List (secondary function). When in Counter Mode, these 2 keys are
used to start/stop or reset/resume the counter.
Freq/Periodkey:
Toggle between frequency and period display (primary function) or to enable sine waveform.
Ampl/Pulse Widthkey:
Display and adjust the amplitude of waveforms or, when in pulse mode, toggle the display
4080 series Function & Arbitrary/Function Generators Instruction Manual
14
MODE > START F > STOP F > TIME >TRIG
between amplitude and pulse width. The secondary function activates the square waveform
FSK/PSKkey:
Toggle between FSK and PSK modulation (primary function). Activate triangle waveform
Menukey:
Enter modulation parameters for FSK, PSK, FM, AM and burst modulation and for sweep
mode. When in standard waveform mode (no modulation, main waveforms enabled), this key
can also be used to toggle the units for the amplitude display value between Vpp, Vrms and
dBm. (Press the Amplitude key first, then the Menu key to toggle between the units). The
secondary function enables the ramp waveform.
FMkey:
Activate FM modulation (primary function). Enter Storage mode (secondary function). Enter
units ms or mVpp after entering the desired value by numerical key pad. In Counter Mode, this
key turns the input attenuator on or off.
AMkey:
Activate AM modulation (primary function). The secondary function is used to recall and
recreate signals stored in status memory. Enter units MHz and Vrms after entering the
desired value by numerical key pad. Enable the low pass filter when in Counter mode.
Sweep key:
Activate sweep mode. Select frequency measuring and totalize mode (secondary function) .
Enter units kHz or mVrms after entering the desired numerical value directly via the keypad.
Use the Shift key to toggle between frequency measurements and totalize mode.
Burstkey:
Activate burst mode (primary function). Enter DC offset mode (secondary function). Enter
units Hz or dBm. When in frequency measurement mode, press this key to enter the gate time.
Outputkey
Press this key to toggle the main OUTPUT signal between the ON and OFF state. By default
the output is turned on, indicated by the green LED and the currently active wave form is
available at the OUTPUT terminal. In Burst or Sweep mode, this key is also used to generate a
single trigger.
3.14 Description of menu parameters
Use this key to configure modulation parameters, sweep mode parameters and system parameters.
Modulation and Sweep mode: After enabling modulation or sweep mode, press menu to
configure the related parameters. Each time you press the menu key, the parameter will flash for 1
second, followed by the currently active value of that parameter. Use the knob or numerical keys to
enter a new value. Once the parameter is set, press menu to advance to the next parameter.
Continue pressing the menu key to cycle through all parameters. Press ShiftSWto return to
the main waveform mode and to set the carrier waveform parameters.
Sweep Mode:
4080 series Function & Arbitrary/Function Generators Instruction Manual
15
FM DEVIA> FM FREQ > FM WAVE > FM SOURCE
AM LEVEL > AM FREQ > AM WAVE > AM SOURCE
TRIG > COUNT > SPACE T > PHASE
START F > STOP F> SPACE T > TRIG
P1 > P2 > SPACE T > TRIG
MODE: Select LINEAR or LOGarithmic sweep
START F :Sweep start frequency
STOP F: Sweep stop frequency
TIME: Sweep time
TRIG: Select trigger source, INTernal or EXTernal
FM modulation:
FM DEVIA: Peak frequency deviation
FM FREQ: Modulating signal frequency
FM WAVE: Modulating signal waveform (sine, square, triangle rising or falling ramp)
FM SOURCE: Toggle between internal and external modulating signal.
AM modulation:
AM LEVEL: Modulation depth
AM FREQ: Frequency of modulating signal
AM WAVE: Modulating signal waveform (sine, square, triangle rising or falling ramp)
AM SOURCE Select internal or external modulating signal
Burst modulation:
TRIG Select trigger source, internal or external
COUNT: Number of burst cycles
SPACE T:Burst time spacing
PHASE: starting phase of the burst
FSK modulation:
START F Primary frequency (same as carrier wave)
STOP F the second frequency (hop frequency)
SPACE T FSK rate
TRIG Select trigger source, external or internal
PSK modulation
P1 phase value #1
P2: phase value #2
4080 series Function & Arbitrary/Function Generators Instruction Manual
16
POWER ON > ADDRESS > OUT Z > INTERFACE>..
BAUD > PARITY > BEEP
SPACE T: PSK rate
TRIG: trigger mode, internal or external
System Function Mode:
POWER ON Power on state
ADDRESS: set GPIB address (option)
OUT Z: Configure amplitude display value for 50 or high impedance load termination
INTERFACE: Select RS232 or GP-IB (IEEE-488) interface (option)
BAUD: Baud rate for the RS232 interface
PARITY: Parity and Data Bits configuration for RS232
BEEP: Toggle the key beep on/off
4080 series Function & Arbitrary/Function Generators Instruction Manual
17
3.2 REAR PANEL
Ext. Trig/FSK/Burst:
Input Terminal for external trigger signals for FSK/PSK Burst modulation and sweep mode
MOD In
Apply modulating signal for AM and FM to this Input terminal
Meas Freq/TOT In
Input terminal for Universal Counter which operates in frequency measurement or Totalize
mode
MOD OUT
The internally generated modulating signal when in AM mode will be available at this output
RS232C
RS232 interface for remote control of instrument (all models) or for download of custom
arbitrary waveforms (models 4084AWG and 4086AWG only)
AC Socket and fuse compartment
Connect the supplied power line cord to this receptacle. Make sure to install the appropriate
fuse according to the selected AC voltage.
AC Input Selector switch
Set this switch to the corresponding AC voltage in your area
4080 series Function & Arbitrary/Function Generators Instruction Manual
18
4. OPERATING INSTRUCTIONS
4.1 Installation
This section contains installation information, power requirements and initial inspection and signal
connections for the 4080 series generators.
Mechanical Inspection
This instrument was carefully inspected before shipment. Upon receipt, inspect the instrument for
damage that might have occurred in transit. If any sign of damage is found, notify your B+K
Precision distributor.
Instrument Mounting
The 4080 Series is intended for bench use. The instrument includes a handle adjustable for
optimum panel viewing angle. The instrument does not require special cooling when operated
within conventional temperature limits.
Power Requirements
The 4080 Series can be operated from any source of 99V to 242V AC and frequency of 48Hz to
63Hz. The maximum power consumption is 35 VA. Replace fuses with the same type, according to
the rating indicated on the rear panel of the instrument.
The instrument power fuse is located in a fuse compartment below the AC input receptacle. To
access the fuse, first disconnect the power cord and then remove the fuse cartridge.
Power-on procedure
Turn on the instrument by pressing the power key on the front panel of the unit. The display should
be flashing BK PRECISION for 2 seconds followed by the model number. e.g. .4084AWG for 1
second. By default, the instrument will enter the standard waveform (SW) mode with the
frequency set to 10.00000000 kHz and the waveform annunciator displaying the ~ symbol. If
the Power ON configuration in the systems menu was modified, the parameters of the last operation
before power-down will be displayed.
4.2. Main operating modes
SW standard waveform mode
In this mode, the generator outputs any of the 27 build-in waveforms, including the 2 main
waveforms sine and square. This also includes the 8 user programmable arbitrary waveforms
(models 4084AWG and 4086AWG only). For most waveforms, you can set frequency, amplitude
and DC offset. In this mode, modulation and sweep is not active and all annunciators in status area 4
are turned off. When modulation or sweep is enabled, pressshiftSWto return to this mode.
Modulation, Pulse and Sweep Mode
In this mode, the generator operates like a conventional function generator and the Arb annunciator
is turned off. The user can select one of the modulation modes AM, FM, FSK, PSK, Burst or sweep
mode. The 2 main waveforms sine and square can be used as carrier signal.
There is a close, reciprocal relationship between the modulation/sweep mode and the standard
4080 series Function & Arbitrary/Function Generators Instruction Manual
19
waveform mode used to set the sine or square wave parameters. When one of the modulation modes
is active, the parameters of the carrier wave are carried over from the Standard Waveform Settings
(for sine or square). PressshiftSWto toggle between any of the modulation/sweep modes (to
set the modulation parameters) and the standard waveform mode (to define the carrier settings for
frequency, amplitude and DC offset). On the other hand, the carrier frequency can also be set in the
modulation/sweep menus and if changed, will overwrite the standard waveform frequency setting.
Example: When transitioning from standard waveform mode to FM modulation mode, (by pressing
the FM button) the parameters set previously in SW mode carry over to the FM mode and are now
defining the FM carrier. If the FM carrier frequency in the FM modulation menu is changed, it will
also overwrite the frequency setting of the standard waveform settings.
To adjust the duty cycle, enable pulse mode by pressing shift . Note that the duty cycle
can only be adjusted in pulse mode.
4.3 Waveform Selection:
Enabling common waveforms
Pressshift followed by the waveform key to select one of the 5 common waveforms sine, square,
triangle, ramp and pulse. The corresponding waveform annunciator will be displayed in the
waveform display area. Note that the instrument operates in Function Generator mode when
selecting sine and square and in ARB mode for the remaining functions. (Arb annunciator is lit).
Note: The 5 common waveforms can also be output by selecting the corresponding number from the
ArbList in this section.
Example: Press ShiftFSK/PSKto enable the triangle waveform
Enabling stored waveforms from the Arb List
Make sure you are in SW mode. (All area 4 status annunciators are off). If modulation or sweep
is active, pressshift SWto return to the standard waveform mode.
Press shiftto enter the ArbList: Waveform,6: NOISE is enabled and appears on the display.
Use the knob or the numeric keypad to select one of the 27 stored waveforms from the ArbList
according to the table in this section.
Example: To select negative ramp DOWN_RAMP, do the following:
Press shift (enter ArbList)
Press 5N(enable DOWN_RAMP)
or
select waveform with the knob
Models 4084AWG and 4086AWG only: Locations 28-35 are reserved for the storage of user
defined arbitrary waveforms. The display name for these waveforms is ARB1-ARB8. These
memory locations can be accessed in one of 2 ways:
a) Enter the ArbList, then use the knob or the numeric keypad to enable waveforms ARB1 ARB8
b) Press Shiftfollowed by any number between 1-8.
Example: Press Shift2to enable waveform stored in memory location ARB2
4080 series Function & Arbitrary/Function Generators Instruction Manual
20
Table of stored waveforms (ArbList) and their memory locations
No. Waveform Display Name No. Waveform Display Name
1 Sine wave SINE 15 Half-wave rectified COMMUT_HA
2 Square wave SQUARE 16 Sine transverse cut SINE_TRA
3 Triangle wave TRIANG 17 Sine vertical cut SINE_VER
4 Ramp UP_RAMP 18 Sine phase modulation SINE_PM
5 Falling ramp DOWN_RAMP 19 Logarithmic function LOG
6 Noise NOISE 20 Exponential function EXP
7 Pulse wave PULSE 21 Half-round function ROUND_HAL
8 Positive pulse P_PULSE 22 SINX/X function SINX/X
9 Negative pulse N_PULSE 23 Square root function SQU_ROOT
10 Positive DC P_DC 24 Tangent function TANGENT
11 Negative DC N_DC 25 Cardiac wave CARDIO
12 Stair wave STAIR 26 Earthquake wave QUAKE
13 Coded pulse C_PULSE 27 Combination wave COMBIN
14
Full wave
rectified
COMMUT_FU 28~35*
User programmable
Arbitrary waveforms
ARB1~ARB8
Note:
*No 28~35: memory location for user programmable waveforms, models 4084AWG and 4086AWG only. Refer to
chapter 7 for more details
4.4 Data entry
Using the arrow keys and the knob
Use the knob and arrow keys to modify the displayed number.
Use the keys to move the flashing digits left or right then adjust the value with the knob.
Using this method of entry, the output signal will be updated immediately. Move the arrow keys left
for coarse adjustment and right for fine adjustment.
To disable the knob, use the key to move the cursor all the way to the left or right until
the digits stop flashing. Now data entry via the knob is disabled.
Direct entry using the numerical key pad
Use the numerical keypad to enter a number with the appropriate unit.
Enter numbers from left to right. Use the key to enter a decimal point. Enter -for negative
numbers. Repeatedly pressing this key will toggle between positive (no sign visible) and negative
numbers. Numerical entries do not update the output signal until a unit key has been pressed. Once
you entered the correct numerical value, press the appropriate unit key to assign a unit and to make
the entry effective. The instrument will now output a signal according to the displayed data.
For entries not associated with a unit, press the shift key (N =no unit) to make the entry
effective.
Note: Numerical values entered via the keypad are not effective until a valid unit key or shift (N)
has been pressed.
4080 series Function & Arbitrary/Function Generators Instruction Manual
21
Entry of invalid key presses or invalid values
If the entered value exceeds the rated range, a beep sound will be heard. If the entered value is below
the lower limit, the instrument will automatically change the entry to the lowest possible value. If
the entered value exceeds the upper limit, the instrument will automatically revert to the maximum
value.
Example: When trying to entering 90MHz in a model 4086 (80 MHz max), a beep will be heard and
the value will be forced to 80 MHz.
Invalid keys: A beep sound will inform the user that the key entry is invalid. The instrument will
simply ignore the key pressed.
Example: Key-is pressed when trying to enter a frequency value. The instrument will ignore the
entry and respond with a beep sound.
4.5 Output Configuration
4.5.1 Set Frequency and Period
Thefrequency/periodkey is used to toggle between Frequency and Period display of the standard
waveform or the carrier waveform.
Frequency
Press the frequency key to display the current frequency value. The value can be modified using
the numerical keypad or the knob.
Example: To set a frequency value of 5.8 kHz, enter the following key sequence:
frequency58kHz
or
frequency5800Hz
or
use the knob and keys
The display will be 5.80000000 kHz.
Period Setting:
The signal frequency can also be displayed or entered as a period value. If the current display is
frequency, press thefrequency/periodkey to display the current period value. Values can be
entered with the numerical key pad or using the knob.
Example: To set a period value of 10ms, enter the following sequence of keys:
Period10ms
Or
use the knob and keys
4.5.2 Set Amplitude
Press the Amplitudekey to display the current amplitude value. Modify the value using the knob
or the numerical keypad.
Example: Set amplitude to 4.6V peak-to-peak:
PressAmplitude46Vpp
or
modify values using the knob and arrow keys
4080 series Function & Arbitrary/Function Generators Instruction Manual
22
In case of standard waveforms sine, square, triangle, rising ramp and pulse, numbers can
be entered and displayed as Peak-to-peak value (Vpp or mVpp) , root mean square value (Vrms and
mVrms) or dBm value.
All other waveforms can only be edited or displayed using Vpp or mVpp units only.
4.5.3 Set DC Offset Voltage
Pressshiftoffsetto display the current DC offset value. If the current DC offset value is not
equal to zero, the annunciator Offset will turn on. The DC offset value can be entered directly or
using the knob.
Example: Set an offset value -1.6V peak-to-peak
Pressshiftoffset-16Vpp
or
shiftoffset16-Vpp
or
use the knob for value entry
Zero Point Adjustment:
For zero point adjustment of the output signal, using the knob is more convenient than direct entry
via the numerical keypad. The transition of the DC offset Voltage from plus to minus sign will be
automatic when passing through the zero point. The input range of the signal amplitude and DC
offset should satisfy the following equation: |Voffset| +Vpp/2 Vmax, with the parameters
defined as followed:
Vpp is the peak-to-peak value of the amplitude
|Voffset| is the absolute value of the DC offset
Vmax is 10V at high impedance and 5V at 50 load.
The following table shows the corresponding relationship between the Vp-p value of the amplitude
and the absolute DC offset value at high impedance:
Vp-p value of AC signal Absolute value of DC offset
1.001 V ~20.00 V 0 ~(10.000-Vpp/2) V
316.1 mV ~1.000 V 0 ~2.000 V
100.1 mV ~316.0 mV 0 ~632.9 mV
31.01 mV ~100.0 mV 0 ~200.9 mV
2.000 mV ~31.00 mV 0 ~62.99 mV
4.5.4 Adjustment of duty cycle
If the current waveform is pulse, and the current display value is amplitude, use the Ampl/Duty
button to toggle between Amplitude and Duty cycle display. When the pulse width is displayed,
enter a value via the numeric keypad or the knob. The valid range is 0.1% ~99.9% for frequencies
below 10kHz with a maximum resolution of 0.1%. For frequencies between 10kHz~100kHz the
range is 1% ~99% and the maximum resolution is 1%.
Example: Enter a duty cycle value of 60.5%
Press Pulse605N
or
use the knob and arrow keys
4080 series Function & Arbitrary/Function Generators Instruction Manual
23
4.5.5 TTL Signal
A TTL signal output is provided on the front-panel TTL terminal. All signals, including modulated
signals and arbitrary waveforms (except DC and noise) have an associated TTL signal.
This signal is a TTL high when the waveforms output (on the main terminal) is positive, relative
to zero volts. The signal is a TTL low when the output is negative. The TTL signal is generated by
passing the main output signal through a comparator configured as a Schmitt Trigger. The TTL
signal will transition to high or low once the main output signal exceeds a certain threshold value,
which causes a time delay between the main output signal and the TTL signal.
4.5.6 Signal Store and Recall
Up to 10 different instrument states can be stored in non-volatile memory. This enables you to recall
the entire instrument configuration with a single command from the remote interface or with just a
few key presses from the front panel.
The state storage feature remembers the function (including arbitrary waveforms), frequency,
amplitude, DC offset, duty cycle, as well as any modulation parameters. To recall a stored state, you
must use the same memory location used previously to store the state.
The last state of the instrument before power-off is automatically stored in location #0, therefore a
total of 11 groups locations ranging from 0 ~10 can be recalled.
Example:
Store the current output signal in group location #1:
ShiftStore1N
The following prompt will be displayed for a few seconds: STORE 1
Previously stored signal configurations will be overwritten.
To recall group location #1 and make it the active output signal:
PressShiftrecall1N
The following prompt will be displayed for a few seconds: RECALL: 1
The stored signals can be continuously recalled and reproduced by scrolling through each location
using the knob.
4080 series Function & Arbitrary/Function Generators Instruction Manual
24
MODE > START F > STOP F > TIME >TRIG
4.6 Set Modulation and Sweep Parameters
4.6.1. Sweep mode
In the frequency sweep mode, the function generator steps from the start frequency to the stop
frequency at a sweep rate which you specify. You can sweep up or down in frequency, and with
either linear or logarithmic spacing. You can also configure the function generator to output a single
sweep (one pass from start frequency to stop frequency) by applying an external trigger. The
function generator can produce a frequency sweep for sine or square waveforms.
Press Sweep to enable the frequency sweep mode. The display shows a preset frequency (start
frequency) and the Sweep symbol appears in the state display area. The carrier frequency will be
displayed and the frequency, amplitude, waveform and DC offset of the carrier signal can be set as
described in previous sections of this chapter. The main functions sine or square wave can be
selected as a carrier signal.
Press the menukey repeatedly to cycle through the sweep menu parameters listed below.
MODE Select LINEAR or LOGarithmic sweep mode
START F sweep start frequency
STOP F: sweep stop frequency
TIME: sweep time
TRIG: Select trigger source INTernal or EXTernal
Each time you press the menu key, the parameter will flash for 1 second, followed by the value
of that parameter. Sweep mode parameters can be entered via numerical keypad or the knob.
Once the parameter is set, press menu to advance to the next parameter.
Sweep mode
MODE parameter: Select LINear (No.1) or LOGarthimic (No.2). In linear sweep mode, the output
frequency changes in a linear fashion during a sweep, whereas in LOG mode the frequency changes
exponentially. The spacing is calculated automatically based on start frequencies, stop frequencies
and sweep time.
Start frequency START F:
The frequency where the sweep starts is the start frequency.
After displaying START F for 1 second, the current start frequency is displayed automatically and
can be modified via knob or direct data entry. By default, the start frequency will carry over from
the setting for the main wave form
.
Stop frequency STOP F
The sweep stops at the frequency STOP F
When the start frequency is lower than the stop frequency, the frequency sweep increases gradually
from start frequency (low frequency) to stop frequency (high frequency); When the start frequency
is higher than stop frequency, the frequency sweep decreases gradually from the start frequency
(high frequency) to stop frequency (low frequency).
4080 series Function & Arbitrary/Function Generators Instruction Manual
25
The frequency range from start frequency to stop frequency is 1Hz ~Fmax in linear sweep mode
and 1mHz ~Fmax in log sweep mode. (Fmax see specification section for main waveform)
Sweep time TIME
The time needed to complete one sweep from start to stop frequency is called sweep time.
The sweep time range is 1ms ~800s in linear sweep mode or 100ms ~800s in LOG sweep mode.
Trigger mode TRIG:
A sweep can be triggered internally or externally. The corresponding display values are 1: INT
and 2: EXT. The default value is INTernal trigger. In internal trigger mode, the instrument will
continuously sweeps from the start frequency to the stop frequency according to the Sweep
parameters. An external trigger signal can be generated in one of two ways.
a) Press the outputkey to trigger a single sweep. The signal frequency will in/decrease from start
frequency to stop frequency, then sweep will stop.
b) Apply an input trigger signal to the EXT Trig terminal on the rear panel. A rising edge will
trigger a single sweep. In external trigger mode, symbols Trig and Ext are displayed in the
status field.
Start and stop of sweep
The sweep will start automatically once sweep mode is selected. If you dont want to output the
sweep signal, simply press the outputkey to disable the signal output (LED above Output button
is OFF). Once all the parameters are set, you can make the sweep signal available at the output by
pressing the Output button again. In external trigger mode, the Outputkey functions only as a
single pulse trigger key, the Output On/Off functionality is now disabled and the output LED will
always be lit.
Sweep configuration example
Sweep parameters are as followed: Frequency range 100Hz~200kHz, sweep time 10s, linear sweep,
internal trigger mode.
Set the carrier frequency parameters
Presssweep
Pressmenu
Select MODE, wait, then enter1N(linear sweep)
Pressmenuto select START F, press100Hz
Pressmenuto select STOP F, press200kHz
Pressmenuto select TIME, press10s
Pressmenu to select TRIG, press1N
Hint: The frequency displayed immediately after pressing [Sweep] is the start frequency. Any
changes to that value will automatically update parameter START F
4080 series Function & Arbitrary/Function Generators Instruction Manual
26
FM DEVIA > FM FREQ > FM WAVE > FM SOURCE
4.6.2 FM modulation
A modulated waveform consists of a carrier waveform and a modulating waveform. In FM, the
frequency of the carrier is varied by the amplitude of the modulating waveform. The function
generator will accept an internal or external FM modulating signal.
Press the FMkey to enter into FM mode. The display will indicate the carrier frequency and the
FM annunciator will be lit. The frequency, amplitude, waveform and DC offset of the carrier
signal can be set as described in the previous section. Only the main functions sine or square wave
can be selected as a carrier signal.
Press the menu key repeatedly to cycle through the FM modulation parameters listed below.
FM DEVIA Peak frequency deviation
FM FREQ: Frequency of modulating signal
FM WAVE: Waveform of modulating signal
FM SOURCE Toggle between internal or external modulating signal
Each time you press the menu key, the parameter will flash for 1 second, followed by the value of
that parameter. FM modulation parameters can be entered directly or via the knob. Once the
parameter is set, press menu to advance to the next parameter.
Peak frequency deviation FM DEVIA
Explanation: The variation in frequency of the modulating waveform from the carrier frequency
(center frequency).
Range of deviation: For internal FM modulation, the maximum value should not exceed 50% of the
carrier frequency. In external FM mode, the maximum deviation is 10% of the carrier frequency.
Additionally, the frequency deviation plus carrier frequency should not exceed the maximum
operating frequency of the instrument. (F
c
+F
d
<=F
max
+100 kHz) For Fmax refer to the
specifications for main waveforms.
This parameter applies only when FM SOURCE is set to INTernal.
Modulating signal frequency FM FREQ:
The frequency range of the modulating signal is 100Hz ~10kHz.
This parameter applies only when FM SOURCE is set to INTernal.
Modulating signal waveform FM WAVE
Waveform of the modulating signal: Waveforms sine, square, triangle, rising and falling ramp can
be used as the modulating signal. Waveforms are selected by entering the corresponding number,
numbers 1 5.
This parameter applies only when FM SOURCE is set to INTernal.
Modulating signal source FM SOURCE:
The modulating signal could be an internal signal and external input signal. The corresponding
prompts are 1: INT and 2: EXT, the default is INTernal signal. The external modulating signal
is applied to terminal Mod In on the rear panel (max signal amplitude is 3Vp-p). When
4080 series Function & Arbitrary/Function Generators Instruction Manual
27
AM LEVEL> AM FREQ > AM WAVE>AM SOURCE
modulating signal source external is selected, the symbol Ext is displayed and parameters FM
DEVIA, FM FREQ and FM WAVE do not apply (disabled)
Start and stop of FM modulation
An FM signal is generated once FM function mode is selected. The instrument will automatically
output a signal according to the preset parameters. If you do not want to output the FM signal,
simply press the outputkey to disable the signal output (LED above Output button is OFF). Once
all the parameters are set, you can turn make the FM signal available at the output by pressing the
Output button again.
FM example:
Example configuration: Carrier signal is square, frequency is 1MHz, amplitude is 2V, modulating
signal is generated internally. Carrier waveform is sine (No. 1), Frequency is 5kHz. Peak frequency
deviation is 200kHz.
PressFM
Pressfrequencythen1MHz (set carrier frequency)
Pressamplitude, then2V (set carrier amplitude)
PressShiftand square (set carrier waveform)
Pressmenu, select FM DEVIA, enter200kHz (set FM deviation)
Pressmenu, select FM FREQ, press5kHz (set FM frequency)
Pressmenu, select FM WAVE, press1N (set FM waveform as sine)
Pressmenu, select FM SOURCE, press1N (set FM source as internal)
4.6.3 AM modulation
AM stands for amplitude modulation.
PressAMto enable AM modulation. The carrier frequency and the AM annunciator is displayed.
Frequency, amplitude, waveform and DC offset of the carrier signal can be set as described in the
previous section of this chapter. The parameters carry over from the parameter settings of the main
waveforms sine and square. In AM only sine and square waves can be selected for carrier.
Press the menu key repeatedly to cycle through the AM modulation parameters listed below.
AM LEVEL: Modulation depth
AM FREQ: Frequency of modulating signal
AM WAVE: Waveform of modulating signal
AM SOURCE: internal or external modulating signal
In AM mode, to ensure normal signal output at 100% modulation depth, the instrument reduces the
peak-to-peak value of the carrier by 50%. Only sine and square waves can be selected as AM
carrier.
4080 series Function & Arbitrary/Function Generators Instruction Manual
28
Modulation depth AM LEVEL:
The range of modulation depth is 1% ~120%.
Modulating frequency AM FREQ
Frequency of the modulating signal. Range is 100Hz ~20kHz. Parameter does not apply when
AM SOURCE is set to EXTernal
Modulating signal waveform AM WAVE:
Waveforms sine, square, triangle, rising and falling ramp can be used as the modulating signal.
Waveforms are selected by entering the corresponding number, numbers 1 5. Parameter does not
apply when AM SOURCE is set to EXTernal
Modulating signal source AM SOURCE
Select internal signal and external input signal. The number and prompt symbols are 1: INT, 2: EXT.
The default of the instrument is internal signal. The external modulating signal is input through the
rear panel Modulation Input terminal (with a maximum signal amplitude of 3Vp-p).
Enabling and disabling the AM signal:
An AM signal is generated once AM modulation is selected. The instrument will automatically
output a signal according to the preset parameters. If you do not want to output the AM signal,
simply press the outputkey to disable the signal output (LED above Output button is OFF). Once
all the parameters are set, you can make the AM signal available at the output by pressing the Output
button again.
AM example:
Configuration: carrier signal is square wave, frequency 1MHz, amplitude 2V, internal signal,
modulating waveform sine wave (No.1), modulating signal frequency 5kHz, modulation depth
50%:
PressAM (activate AM modulation)
Pressfrequency, then1MHz (set carrier frequency)
Pressamplitude, then2V (set carrier amplitude)
Pressshift thensquare (set carrier waveform)
Pressmenu, select AM LEVEL, press50N (set modulation depth)
Pressmenu, select AM FREQ, press5kHz (set modulating signal frequency)
Pressmenu, select AM WAVE, press1N (set AM wave as sine)
Pressmenu select AM SOURCE, press1N (set AM source to internal)
4.6.4 Burst modulation
You can configure the function generator to output a waveform with a specified number of cycles,
called a burst. You can output the burst at a rate determined by the internal rate generator or an
external signal applied to the rear-panel connector. The function generator can produce a burst using
sine or square waveforms.
PressBurstto enter into burst mode. The carrier frequency and the Burst annunciator are
displayed. Frequency, amplitude, waveform and DC offset of the carrier signal can be set as
4080 series Function & Arbitrary/Function Generators Instruction Manual
29
TRIG > COUNT > SPACE T > PHASE
described in the previous section of this chapter. The parameters carry over from the parameter
settings of the main waveforms sine and square. Press the menu key repeatedly to cycle through
the burst menu parameters listed below.
TRIG: Trigger source
COUNT: Number of cycles
SPACE T: Burst time spacing
PHASE: The starting phase of the burst
Select trigger source TRIG
Three trigger sources are available. Internal trigger and 2 types of external triggers sources, external
gated and single. The corresponding menu parameters are 1: INT, 2: EXT and 3: SINGLE. The
default is internal triggering.
a) Internal trigger
When the internal trigger source is selected, the frequency at which the burst is generated depends
on parameter SPACE_T, burst count and the burst carrier frequency.
b) Single trigger mode
Generate a single trigger event by pushing the outputkey once or by applying a single TTL pulse
at the Ext.Trig input terminal in the rear panel. A single burst with predefined parameters will be
generated.
c) External gated burst mode
The TTL signal applied to Ext. Trig input terminal in the rear panel will enable or disable the
output of the generator. When the TTL level of the gate signal is high, the generated will generate
burst pulses according to the parameters of the set carrier frequency. Burst parameters COUNT and
SPACE_T are ignored in this mode. Anunciators Trig and Ext turn on when this mode is active.
Burst count COUNT:
Definition: The number of cycles to be output per burst.
Range: 1 ~10000 cycles in 1 cycle increments. Additionally, the minimum number of cycles must
also satisfy F
C
/50khz +1Parameter does not apply when TRIG is set to EXTernal.
Burst spacing time SPACE T:
Definition: Time interval between the 2 consecutive groups of bursts (time interval during which
there is no burst signal present) Parameter does not apply when TRIG is set to EXTernal or
SINGle.
Range: 0.1ms ~800s.
Starting phase of the burst PHASE
Burst phase: 0.0 ~360.0 in 0.1 increments
4080 series Function & Arbitrary/Function Generators Instruction Manual
30
START F > STOP F> SPACE T > TRIG
Enabling/disabling the burst signal
The generator outputs a burst as soon as the burst key is pressed. The instrument will automatically
output a signal according to the preset parameters. If you do not want to output the burst signal,
simply press the outputkey to disable the signal output (LED above Output button is OFF). Once
all the parameters are set, you can make the burst signal available at the output by pressing the
Output button again. Note that the output on/off function does not apply when single trigger mode
is selected.
Burst example:
Configuration: Burst is sine wave with frequency 20kHz, amplitude 2V, 10 cycles per group,
spacing time between each group 10 ms, start phase 90.0
Pressburst (enter into burst mode)
Pressfrequency, press20kHz (set wave form frequency)
Pressamplitude, press2V (set waveform amplitude)
Pressshiftandsine (set waveform)
Pressmenu, select TRIG, press1N (set trigger mode as internal)
Pressmenu, select COUNT, press10N (set number of bursts/cycles)
Pressmenu, select SPACE T, press10ms (set space time)
Pressmenu, select PHASE, press90N (set the start phase of burst)
4.6.5 FSK modulation
You can configure the function generator to shift its output frequency between two preset values
using FSK modulation. The rate at which the output shifts between the two frequencies (carrier
frequency and hop frequency) is determined by the internal rate generator or the signal level on
the rear-panel ext Trig/FSK/Burst terminal.
PressFSK/PSK to enable FSK modulation. A preset frequency and the FSK annunciator is
displayed. The preset frequency is identical to the START F parameter. Any changes
automatically update START-F and vice versa. Frequency, amplitude, waveform and DC offset of
the carrier signal can be set as described in the previous section of this chapter. In FSK/PSK only
sine and square waves can be selected as carrier wave.
Press the menu key repeatedly to cycle through the FSK menu parameters listed below.
START F: First frequency or carrier frequency
STOP F: Second frequency (hop frequency)
SPACE T: FSK rate
TRIG: Trigger source
Trigger source TRIG
Select internal signal and external input signal, 1: INT or 2: EXT. The default of the instrument is
4080 series Function & Arbitrary/Function Generators Instruction Manual
31
P1 > P2 > SPACE T > TRIG
internal.
In internal mode, the rate at which the signal shifts between the carrier and hop frequency is
determined by the FSK rate.
In external trigger mode, the trigger signal applied to the trigger input terminal on the rear panel
determines the FSK rate. Logical high of the trigger signal is associated with frequency 2, while
the Low level of the trigger signal is associated with frequency 1.
Frequency 1 START F
The first frequency or carrier frequency.
Frequency 2 STOP F
Second frequency or hop frequency
Frequency input range of frequency 1 and frequency 2 is 1Hz ~Fmax.
Spacing time SPACE T:
This parameter sets the FSK rate with a range of 1ms ~800s.
FSK example:
Configuration: Sine signal of 2V output amplitude, carrier frequency 20kHz and hopping frequency
600 kHz, FSK rate 10ms
PressFSK/PSK (enter into FSK function mode)
PressAmpl/Duty, press2V (set waveform amplitude)
Pressshiftandsine (set waveform)
Pressmenu, select TRIG, enter1N (set trigger mode as internal)
Pressmenu, select START F, enter20kHz (set carrier frequency F1)
Pressmenu, select STOP F, enter600kHz (set hop frequency F2)
Pressmenu, select spacing time SPACE T, press10ms (set spacing time)
4.6.6 PSK modulation
You can configure the function generator to shift its output phase between two preset values using
PSK modulation. The rate at which the output shifts between the two phases is determined by the
internal rate generator or the signal level applied to the trigger input terminal on the rear-panel.
Press FSK/PSK twice to enter phase shift keying (PSK) mode. The carrier frequency will be
displayed along with the symbol and FSK. Frequency, amplitude, waveform and DC offset
of the carrier signal can be set as described in the previous section of this chapter. The parameters
carry over from the parameter settings of the main waveforms sine and square. If FSK mode was
already active, press this key only once, (this key toggles the modulation between FSK and PSK
mode). In FSK mode only sine and square waves can be selected as a carrier signal.
Press the menu key repeatedly to cycle through the burst menu parameters listed below.
P1 Phase value #1
4080 series Function & Arbitrary/Function Generators Instruction Manual
32
POWER ON > ADDRESS > OUT Z > INTERFACE>.
BAUD > PARITY > BEEP
P2: phase value #2
SPACE T PSK rate
TRIG Trigger source for PSK
Trigger source TRIG
Select internal signal and external input signal 1: INT or 2: EXT. The default of the instrument is
internal. In internal mode, the phase of the output signal shifts according to the PSK rate parameter.
In external trigger mode, the trigger signal is applied to the Ext. Trig Input terminal on the rear
panel. In external trigger mode, the trigger signal applied to the trigger input terminal on the rear
panel determines the PSK rate. Logical high of the trigger signal is associated with phase value #2,
while the Low level of the trigger signal is associated with phase value #1.
Phase 1 P1
The first start phase value of the PSK output signal.
Input range of phase 1 and phase 2 is 0.0~360.0.
Phase 2 P2
The second start phase value of the PSK output signal.
Input range of phase 1 and phase 2 is 0.0~360.0.
Spacing time SPACE T
PSK rate: Range is 0.1ms ~800s.
PSK example
Configuration: Sine signals with 600 kHz output frequency, 2V amplitude, phase alternating
between 90.0 and 180.0, PSK rate is 10ms
PressFSK/PSK (2 times if necessary to enter PSK mode)
PressFreq/Period, enter600kHz (set waveform frequency)
PressAmpl/Duty, enter2V (set waveform amplitude)
Pressshiftsine (set waveform)
Pressmenu, select TRIG, enter1N, (set trigger mode as internal)
Pressmenu, select P1, enter90N (set phase 1)
Pressmenukey, select P2, enter180N (set phase 2)
Pressmenu, select SPACE T, enter10ms (set PSK rate)
4.7 Set System Parameters
PressShiftandsystemto enter the Systems menu. SYSTEM will flash on the display.
Repeatedly pressmenuto cycle through the system parameters indicated below.
4080 series Function & Arbitrary/Function Generators Instruction Manual
33
POWER ON Power on state
ADDRESS: set GPIB address (option)
OUT Z: Configure amplitude display for 50 or high impedance load termination
INTERFACE: Select RS232 or GP-IB (IEEE-488) interface (option)
BAUD: Baud rate for the RS232 interface
PARITY: Parity and Data Bits configuration for RS232
BEEP: Toggle the key beep on/off
After the selected system parameter flashes for 1 second, the parameter value will be displayed and
the value can be edited using the knob or via direct entry using the numerical keypad.
Power on state POWER ON:
This parameter has 2 possible values, 1: DEFAULT or 2: LAST STATE. In determines the
initial setting of the instrument after power up. The factory setting is 1: DEFAULT. The default
state is main waveform sine wave, 10kHz, 2Vpp which is identical to the state the instrument is in
after pressing Reset. When selecting 2: LAST STATE, the instrument will remember the last
state the unit was in before power was turned off. The power on state is stored in non-volatile
memory, location 0 (see Store/Recall section)
GPIB address ADDRESS
The default of GPIB interface address is 1. It can be set within 0 ~30.
Output impedance OUT Z
The generator has a fixed output impedance of 50 ohms on the OUTPUT terminal. You can specify
whether you are terminating the output into a 50 ohm load or an open circuit. Incorrect impedance
matching between the function generator and your load will result in amplitude or offset which does
not match the specified signal level.
Select 1: HIGH Z or 2: 50 OHM according to your load configuration.
INTERFACE selection
Select RS232 (2: RS232), GPIB (1: GP-IB) or USB (3: USB). The default is RS232.
BAUD rate setting
Set the baud rate for the RS232 interface. Possible baud rates are 9600, 4800, 2400, 1200, 600 and
300. The default rate is 9600 (1: 9600)
PARITY setting
Set parameters parity bit and number of data bits. Possible configurations are:
None/ 8 data bits (1: NONE 8 BITS)
Odd/ 7 data bits (2: ODD 7 BITS
Even/ 7 data bits (3: EVEN 7 BITS).
The default setting is None/ 8 data bits.
4080 series Function & Arbitrary/Function Generators Instruction Manual
34
BEEP setting
Turns the key beep on or off. Select OFF(1: CLOSE) or ON (2: OPEN). The default is
ON.
Example for system function setting
Configuration: Set power on state [POWER ON] to default and output impedance to 50
PressShiftsystem (enter systems menu)
Pressmenu, POWER ON, press1N (set power on state to default)
Pressmenu, select OUT Z, press2N (configure for load impedance of 50)
4.8 Universal Counter
Overview of Counter functions
This instrument contains a counter with frequency measurement and totalize functionality which is
completely independent from the Arbitrary/function generator section. The range of the frequency
measurement function is 1Hz ~100MHz. Repeatedly pressing ShiftCountwill toggle between
Frequency measurement and Totalize mode.
PressShiftandF.C./TOTonce to activate the frequency measurement mode. Symbols Ext
and Freq will be displayed in the state field and WAITING will flash on the main display. The
unit is now ready to measure and display the frequency of a signal applied to the input terminal
Meas Freq/TOT IN in the rear panel. When totalize mode is active, symbols Ext and Count
will be displayed on the status field along with the result for the count on the main display (0 is
displayed until a valid signal is detected). The unit is now ready to totalize events represented by the
signal applied to input terminal Meas Freq/TOT IN in the rear panel
In event counting mode, press the key to start and stop the event counter. When stopped, the
current count will be displayed. Presskey to reset the event counter to zero and resume
counting.
Gate time
Only applies to frequency measurement mode. PressShiftgateto enter a gate time setting.
Either the numerical keypad or the knob can be used to enter a gate time value. If the gate is open,
the symbol GATE is displayed in the counter state field on the right. The range of gate times is
10ms ~10s
Low pass filter
By pressing ShiftLPF, the input signal will be passed through a low pass filter before being
passed onto the counter circuit. The word Filter appears on the right display field to indicate that
the Low Pass Filter is active.
Attenuation
PressShiftATT to activate the attenuator which attenuates the input signal by a factor 10:1.
ATT is displayed on the right to indicate that the input attenuator is active.
4080 series Function & Arbitrary/Function Generators Instruction Manual
35
5. REMOTE INTERFACE REFERENCE
5.1 Introduction
An Introduction to the SCPI Language
SCPI (Standard Commands for Programmable Instruments) is an ASCII-based instrument
command language designed for test and measurement instruments. SCPI commands are based on
a hierarchical structure, also known as a tree system. In this system, associated commands are
grouped together under a common node or root, thus forming subsystems. A portion of the
SOURCE subsystem is shown below to illustrate the tree system.
SOURce:
FREQuency:
STARt {<frequency>|MINimum|MAXimum}
FREQuency:
STARt? [MINimum|MAXimum]
FREQuency:
STOP {<frequency>|MINimum|MAXimum}
FREQuency:
STOP? [MINimum|MAXimum]
SOURce is the root keyword of the command, FREQuency is the second-level keywords, and
STARt and STOP are third-level keywords. A colon (: ) separates a command keyword from a
lower-level keyword.
Preparations for Remote Operation
The Arbitrary/Function Generator series 4080 can be remotely controlled from a PC via the
standard RS232 interface. Make sure to configure the RS232 interface to match the settings on your
PC by setting the parameters in the Systems menu accordingly. The instrument enters the remote
state after any remote command is received at which time all keys except of the Localkey (Shift)
are locked When in Remote mode, the instrument can be returned to Local mode by pressing the
Local key or by sending a LOCAL instruction via the RS232 interface.
5.2 Overview and syntax of SCPI instructions
This section summarizes the SCPI commands available to program the function generator over the
remote interface. Refer to the later sections in this chapter for more details on each command.
Throughout this manual, the following conventions are used for SCPI command syntax.
Square brackets ([ ]) indicate optional keywords or parameters.
Braces ({ }) enclose parameters within a command string.
Triangle brackets (<>) indicate that you must substitute a value for the enclosed parameter.
A vertical bar (|) separates multiple parameter choices.
Command Summary
APPLY instruction
4080 series Function & Arbitrary/Function Generators Instruction Manual
36
APPLy:SINusoid [<frequency>, <amplitude>, <offset>]
APPLy:SQUare [<frequency>, <amplitude>, <offset>]
APPLy:TRIangle [<frequency>, <amplitude>, <offset>]
APPLy:UP_RAMP [<frequency>, <amplitude>, <offset>]
APPLy:DOWN_RAMP [<frequency>, <amplitude>, <offset>]
APPLy:NOISe [<frequency>*, <amplitude>, <offset>]
APPLy:P_DC [<frequency>,* <amplitude>, <offset>]
APPLy?
*: if waveform is Noise or DC, the frequency data contained in the command is not invalid.
[SOURce:]
FUNCtion SHAPe {SINusoid|SQUare|TRIangle|UP_RAMP|DOWN_RAMP|
NOISe|PULSe|P_PULSE|N_PULSE|P_DC|N_DC|STAIR|
C_PULSE|COMMUT_FU|COMMUT_HA|SINE_TRA|
SINE_VER|SINE_PM|LOG|EXP|ROUND_HAL|
SINX/X|SQU_ROOT|TANGent|CARDIO|QUAKE|COMBIN}
FUNCtion:SHAPe?
[SOURce:]
FREQuency <frequency>
FREQuency?
[SOURce:]
PULSe:DCYCle <percent>
PULSe:DCYCle?
[SOURce:]
VOLTage <amplitude>
VOLTage?
VOLTage: OFFSet <offset>
VOLTage: OFFSet?
[SOURce:]
OUTPut:LOAD {50|INFinity}
OUTPut:LOAD?
INPUT:FILTER <on|off>
INPUT:FILTER?
INPUT:ATTenuator <on|off>
INPUT:ATTenuator?
*SAV {0|1|2|3|4|5|6|7|8|9|10}. State 0 is the working state of the instrument before power off
*RCL {0|1|2|3|4|5|6|7|8|9|10}. States 110 are user defined working states
RF:STATe <ON/OFF>
RF:STATe?
[SOURce:]
MODULATION instructions
OUTPUT instructions
4080 series Function & Arbitrary/Function Generators Instruction Manual
37
AM:DEPTh <depth in percent>
AM:DEPTh?
AM:INTernal:FUNCtion {SINusoid|SQUare|TRIangle|UP_RAMP|DOWN_RAMP|}
AM:INTernal:FUNCtion?
AM:INTernal:FREQuency <frequency>
AM:INTernal:FREQuency?
AM:SOURce {INTernal|EXTernal}
AM:SOURce?
AM:STATe {ON|OFF}
AM:STATe?
[SOURce:]
FM:DEViation <peak deviation in Hz>
FM:DEViation?
FM:INTernal: FUNCtion {SINusoid|SQUare|TRIangle|UP_RAMP|DOWN_RAMP|}
FM:INTernal: FUNCtion?
FM:INTernal: FREQuency <frequency>
FM:INTernal: FREQuency?
FM:SOURce {INTernal|EXTernal}
FM:SOURce?
FM:STATe {ON|OFF}
FM:STATe?
[SOURce:]
BM:NCYCles <#cycles>
BM:NCYCles?
BM:PHASe <degrees>
BM:PHASe?
BM:INTernal: Space <time in second>
BM:INTernal: Space?
BM:SOURce {INTernal|EXTernal|SINGle}
BM:SOURce?
BM:STATe {ON|OFF}
BM:STATe?
[SOURce:]
FSKey:FREQuency <frequency>
FSKey:FREQuency?
FSKey:INTernal: Space <time in second>
FSKey:INTernal: Space?
FSKey:SOURce {INTernal|EXTernal}
FSKey:SOURce?
FSKey:STATe {ON|OFF}
FSKey:STATe?
[SOURce:]
PSKey:PHASe1 <degrees>
PSKey:PHASe1?
PSK instructions
FSK instructions
4080 series Function & Arbitrary/Function Generators Instruction Manual
38
PSKey:PHASe2 <degrees>
PSKey:PHASe2?
PSKey:INTernal: Space <time in second>
PSKey:INTernal: Space?
PSKey:SOURce {INTernal|EXTernal}
PSKey:SOURce?
PSKey:STATe {ON|OFF}
PSKey:STATe?
[SOURce:]
FREQuency:STARt <frequency>
FREQuency:STARt?
FREQuency:STOP <frequency>
FREQuency:STOP?
[SOURce:]
SWEep:SPACing {LINear|LOGarithmic}
SWEep:SPACing?
SWEep:TIME <time in second>
SWEep:TIME?
SWEep:SOURce {INTernal|EXTernal}
SWEep:SOURce?
SWEep:STATe {ON|OFF}
SWEep:STATe?
[SOURce:]
FUNCtion:TOTal INITial
FUNCtion:TOTal STARt
FUNCtion:TOTal STOP
FUNCtion:TOTal CLEAr
FUNCtion:TOTal?
[SOURce:]
FUNCtion:FREQuency MEASure
FUNCtion:FREQuency?
FUNCtion:FREQuency GATE <time>
FUNCtion:FREQuency GATE?
TRIGger:SOURce {IMMediate|EXTernal|BUS}
TRIGger:SOURce?
FREQUENCY MEASUREMENT instructions
TRIGGER instructions
SWEEP instructions
COUNT instructions
4080 series Function & Arbitrary/Function Generators Instruction Manual
39
*IDN
*RST
*SAV {0|1|2|3|4|5|6|7|8|9|10}
*RCL {0|1|2|3|4|5|6|7|8|9|10}
SYSTem:LOCAL
SYSTem:REMOTE
5.3 Detailed description of SCPI Instructions
APPLy instructions
APPLy instructions are used to directly set the output of the function generator through the remote
interface including waveform, frequency and offset. For example, to output a sine wave of 5kHz,
3Vp-p with 2.5V DC offset, the following commands need to be sent:
APPL:SIN 5.0E+3,3, -2.5
Instructions of lower level can also be used:
FUNC:SHAPE SIN, to output a sine wave
FREQ 5.0 E+3, to set frequency as 5kHz
VOLT 3.0, to set voltage as 3 VPP
VOLT:OFFSET -2.5, to set output offset as -2.5 V
APPLy?
Queries the waveform, frequency, amplitude and offset of current function output. The returned data
format is:
SIN 5.000000000000E+03, 3.000000E+00, -2.500000E+00
OUTPUT instructions
FUNCtion:SHAPe {SINusoid|SQUare|TRIangle|UP_RAMP|DOWN_RAMP|NOISe}
Select output waveform. 27 waveforms are available. In fixed frequency mode, each of the 27 kinds
can be selected. In other modes, only sine and square waves can be selected.
FUNCtion:SHAPe?
Check the waveform of current function output. The returned data are SIN, SQU, TRI, UP_RAMP,
etc.
FREQuency <frequency>
Set the frequency of function output
FREQuency?
Check the frequency of current main function output. The returned data unit is Hz.
PULSe:DCYCle <percent>
Set the duty cycle of pulse output.
Duty cycle: 0.1%99.9%, in step of 0.1% (f 10 KHz)
1%99% in step of 1% (10 kHz <f 100 kHz), default value 20%.
PULSe:DCYCle?
RS232 specific instructions
SYSTEM instructions
4080 series Function & Arbitrary/Function Generators Instruction Manual
40
Check the duty cycle of current pulse output. The returned value is percentage.
VOLTage <amplitude>
Set the amplitude of the currently active waveform. The amplitude setting should be consistent with
the technical specifications. The output amplitude ranges of P_PULSE, N_PULSE, P_DC and
N_DC should be 2mVpp ampl 10Vpp in high impedance output setting and 1mVpp ampl
5Vpp in 50 setting. The above value is given based on a DC offset of 0V. If a DC offset is used,
refer to the setting of DC offset described in Chapter 4 in this manual
VOLTage?
Check the output amplitude of current main function output.
VOLTage OFFSet <offset>
Set the DC offset. For the relationship between the DC offset and output amplitude, refer to section
4.5.3.
VOLTage OFFSet?
Check the DC offset of main function output.
OUTPut:LOAD {50|INFinity}
Configure the amplitude display value for 50 or high impedance (INFinity) load termination. The
output amplitude should be consistent with the technical specifications.
OUTPut:LOAD?
Read the current load impedance setting of the function generator.
INPUT: FILTER {ON|OFF}
Enable/disable the low pass filter for the counter
INPUT:FILTER?
Read the current on/off state of the low pass filter.
INPUT:ATTenuator {ON|OFF}
Enable/disable Set the input attenuator of the counter
INPUT:ATTenuator?
Read the state of the input attenuator (on/off)
*SAV {0|1|2|3|4|5|6|7|8|9|10}
Save as many as 11 input states of the instrument. Among the 11 states, state 0 is used to
automatically store the power off setting of the instrument. States 1~10 are user defined.
*RCL {0|1|2|3|4|5|6|7|8|9|10}
Recall one of the states 0 -10.
RF:STATe <ON/OFF>
Enables/Disables main output
RF:STATe?
Reads the current state of the main output
AM instructions
1To set the carrier of output using APPLy instructions or FUNCtion, FREQuency, VOLTage,
VOLTAGE: OFFSet instructions.
2To set modulating waveform of AM using AM:INTernal: FUNCtion
{SINusoid|SQUare|TRIangle} instructions
3To set modulating frequency of AM using AM: INTernal: FREQuency <frequency>instructions.
4080 series Function & Arbitrary/Function Generators Instruction Manual
41
4To set the modulation depth of AM using AM: DEPTh {depth in percent} instructions
5To set modulating source of AM using AM: SOURce {INTernal|EXTernal} instructions
6To start AM using AM: STATe ON instructions.
AM: DEPTh {depth in percent}
Set modulation depth of AM. MIN=1%, MAX=120%.
AM: DEPTh?
Read the current modulation depth of AM.
AM:INTernal:FUNCtion {SINusoid|SQUare|TRIangle}
Set the modulating waveform for AM,using modulating waveforms SIN, SQU, TRI UP, RAMP and
DOWN_RAMP.
AM:INTernal:FUNCtion?
Check current modulating waveform of AM. Return parameters are SIN, SQU, TRI,
UP_RAMP, DOWN_RAMP
AM:INTernal: FREQuency <frequency>
Set frequency of modulating signal of AM. When internal modulating source is selected, the
modulating frequency should be10kHz.
AM:INTernal:FREQuency?
Read the frequency of the current modulating signal for AM.
AM:SOURce {INTernal|EXTernal}
Set modulating source of modulating signal of AM as internal (INTernal) or external (EXTernal).
AM:SOURce?
Read the current modulating source of AM.
AM:STATe {ON|OFF}
Turn AM mode on/off
AM:STATe?
Check the on/off state of AM mode. The returned data are 0 (OFF) or 1 (ON).In off state,
the instrument reverts back to fixed frequency mode.
FM instructions
1To set the carrier of output using APPLy or FUNCtion, FREQuency, VOLTage, VOLTAGE:
OFFSet instructions.
2To set modulating waveform of FM using FM: INTernal: FUNCtion
{SINusoid|SQUare|TRIangle} instructions
3To set modulating frequency of FM using instruction FM:INTernal: FREQuency <frequency>
4To set the modulating deviation of FM using instruction FM: DEViation <peak deviation in Hz>
5To set modulating source of FM using the FM: SOURce {INTernal|EXTernal} instruction
6To start FM using FM: STATe ON instruction.
FM:DEViation <peak deviation in Hz>
Set the modulating deviation of FM. The maximum value should be half of carrier frequency in
internal modulating source and 10% of carrier frequency in external modulating frequency. (Should
also satisfy the following equation: F
c
+F
d
<= F
max
+100khz
FM:DEViation?
Read the current modulating deviation of FM.
4080 series Function & Arbitrary/Function Generators Instruction Manual
42
FM:INTernal:FUNCtion {SINusoid|SQUare|TRIangle}
Set the modulating waveform of FM using SIN, SQU, TRI UP, RAMP or DOWN_RAMP waveforms.
FM:INTernal:FUNCtion?
Read the current modulating waveform of FM. The returned data are SIN, SQU, TRI,
UP_RAMP, DOWN_RAMP.
FM:INTernal:FREQuency <frequency>
Set the frequency of modulating signal of FM. The modulating frequency should be 10KHz when
internal modulating source is selected.
FM:INTernal:FREQuency?
Read the frequency of current modulating signal of FM.
FM:SOURce {INTernal|EXTernal}
Set modulating source of FM signal as internal (INTernal) or external (EXTernal).
FM:SOURce?
Read the current modulating source of FM.
FM:STATe {ON|OFF}
Turn on FM mode (ON), turn off FM mode (OFF).
FM:STATe?
Check the on/off state of FM mode. The returned data are 0 (OFF) or 1 (ON). It returns to
fixed frequency mode in off state.
Burst Mode instructions
1. Set the carrier using APPLy instructions or FUNCtion, FREQuency, VOLTage, VOLTAGE:
OFFSet instruction.
2. Set number of burst using BM: NCYCles <#cycles>instruction.
3. Set start phase of burst using BM: PHASe <degrees>instruction.
4. Set time interval between two bursts using BM:INTernal: SPACe <time>instruction.
5. Set trigger source of burst using BM: SOURce {INTernal|EXTernal|SINGle} instruction.
6. To turn on burst use the BM: STATe ON instruction.
BM:NCYCles <#cycles>
Set waveform numbers of burst. MAX10000.
BM:NCYCles?
Check number of bursts.
BM:PHASe <degrees>
Set start phase of burst in 0360 in step of 0.1.
BM:PHASe?
Check the start phase of burst.
BM:INTernal: SPACe <time>
Set time interval between two bursts.
BM:INTernal: SPACe?
Check time interval between two bursts.
BM:SOURce {INTernal|EXTernal|SINGle}
Set trigger source of burst as internal (INTernal), external (EXTernal) or single (SINGle).
4080 series Function & Arbitrary/Function Generators Instruction Manual
43
BM:SOURce?
Check current trigger source of burst.
BM:STATe {ON|OFF}
Turn on burst mode (ON) or turn off burst mode (OFF).
BM:STATe?
Check on/off state of burst mode. The returned data are 0 (OFF) or 1 (ON). The instrument
reverts back to fixed frequency mode in off state.
FSK Mode instructions
1To set the output waveform, amplitude, DC offset and frequency 1 using APPLy instructions or
FUNCtion, FREQuency, VOLTage, VOLTAGE:OFFSet instructions.
2set frequency 2 of FSK using FSKey: FREQuency <frequency>instruction.
3set time interval of frequency switching using FSKey:INTernal: SPACe <time>instruction.
4set trigger source of FSK using FSKey: SOURce {INTernal|EXTernal} instructions.
5turn on FSK using FSKey: STATe ON instruction.
FSKey:FREQuency <frequency>
Set frequency 2 of FSK.
FSKey:FREQuency?
Check frequency 2 of FSK.
FSKey:INTernal:SPACe <time>
Set time interval of frequency switching
FSKey:INTernal:SPACe?
Check time interval of frequency switching
FSKey:SOURce {INTernal|EXTernal}
Set trigger source of FSK as internal (INTernal) or external (EXTernal).
FSKey:SOURce?
Check current trigger source of FSK.
FSKey:STATe {ON|OFF}
Turn on FSK mode (ON) or turn off FSK mode (OFF).
FSKey:STATe?
Check the on/off state of FSK mode. The returned data are 0 (OFF) or 1 (ON). It returns to
fixed frequency mode in off state.
PSK Mode instructions
1. To set the output waveform, amplitude, DC offset and frequency using APPLy instructions or
FUNCtion, FREQuency, VOLTage, VOLTAGE:OFFSet instructions.
2. To set phase 1of PSK using PSKey: PHASe1 <degrees>instruction.
3. To set phase 2 of PSK using PSKey: PHASe2 <degrees>instruction.
4. To set time interval of phase switching using PSKey:INTernal: SPACe <time>instruction.
5. To set trigger source of PSK using PSKey: SOURce {INTernal|EXTernal} instructions.
6. To turn on PSK using PSKey: STATe ON instruction.
FSKey:PHASe1 <degrees>
Set phase 1 of PSK in 0--360.
PSKey:PHASe1?
Check phase 1of PSK
4080 series Function & Arbitrary/Function Generators Instruction Manual
44
PSKey:PHASe2 <degrees>
Set phase 2 of PSK in 0--360.
PSKey:PHASe2?
Check phase 2 of PSK.
PSKey:INTernal:SPACe <time>
Set time interval of phase switching
PSKey:INTernal:SPACe?
Check time interval of phase switching
PSKey: SOURce {INTernal|EXTernal}
Set trigger source of PSK as internal (INTernal) or external (EXTernal).
PSKey:SOURce?
Check current trigger source of PSK.
PSKey:STATe {ON|OFF}
Turn on PSK mode (ON) or turn of f PSK mode (OFF).
PSKey:STATe?
Check the on/off state of PSK mode. The returned data are 0 (OFF) or 1 (ON). It returns to
fixed frequency mode in off state.
Sweep Mode instructions
1. To set the output waveform, amplitude, DC offset and frequency using APPLy instructions or
FUNCtion, FREQuency, VOLTage, VOLTAGE:OFFSet instructions.
2. To set start frequency of sweep using FREQuency: STARt <frequency>instruction.
3. To set stop frequency of sweep using P FREQuency: STOP<frequency>instruction.
4. To set sweep mode using SWEep: SPACing {LINear|LOGarithmic} instruction.
5. To set sweep time using SWEep: TIME <time>instructions.
6. To set trigger source of sweep using SWEep: SOURce {INTernal|EXTernal}instructions.
7. To turn on sweep using SWEep: STATe ON instruction.
FREQuency:STARt <frequency>
Set start frequency of sweep.
FREQuency:STARt?
Check startting frequency of sweep.
FREQuency:STOP <frequency>
Set stop frequency of sweep.
FREQuency:STOP?
Check stop frequency of sweep.
SWEep:SPACing {LINear|LOGarithmic}
Set sweep mode as linear (LINear) or logarithmic (LOGarithmic)
SWEep:SPACing?
Check sweep mode.
SWEep:TIME <time>
Set sweep time.
SWEep:TIME?
4080 series Function & Arbitrary/Function Generators Instruction Manual
45
Check the sweep time.
SWEep:SOURce {INTernal|EXTernal}
Set trigger source of sweep as internal (INTernal) or external (EXTernal).
SWEep:SOURce?
Check current trigger source.
SWEep:STATe {ON|OFF}
Turn on sweep mode (ON) or turn off sweep mode (OFF).
SWEep:STATe?
Check the on/off state of sweep mode. The returned data are 0 (OFF) or 1 (ON). It returns to
fixed frequency mode in off state.
Counting instructions
FUNCtion:TOTal INITial
Initialize the counter.
FUNCtion:TOTal STARt
Set the counter to start counting.
FUNCtion:TOTal STOP
Set the counter to stop counting.
FUNCtion:TOTal CLEAr
Clear the counted value of the counter.
FUNCtion:TOTal?
Record the current value of the counter.
Frequency Measurement instructions
FUNCtion:FREQuency MEASure
Enable frequency measurement function for the instrument to start a new frequency measurement.
FUNCtion:FREQuency?
Record last frequency value after frequency measurement is finished.
FUNCtion:FREQuency GATE <time>
Set gate time in the range of 10ms to 10s for frequency measurement.
FUNCtion:FREQuency GATE?
Check current gate time for frequency meadurement.
Trigger instructions
TRIGger:SOURce {IMMediate|EXTernal|BUS}
Set trigger source or modulating source for available functions as internal (IMMediate), external
(EXTernal) or single (BUS).
TRIGger:SOURce?
Check trigger source or modulating source of current function as internal (IMM), external, (EXT)
or single (BUS).
System related instructions:
*IDN
This query will return the instruments manufacturer, model number and firmware version.
Example: B+K Precision 4084AWG V1.00
4080 series Function & Arbitrary/Function Generators Instruction Manual
46
*RST
Reset instrument to the default state
*SAV {0|1|2|3|4|5|6|7|8|9|10}
Save as many as 11 input states of the instrument. Among the 11 states, state 0 is used to
automatically store the power off setting of the instrument. States 1~10 are user defined.
*RCL {0|1|2|3|4|5|6|7|8|9|10}
Recall on of the states 0 -10.
Special instructions for RS232
SYSTem:LOCAL
This instruction causes the instrument to enter into local state. After receipt of this command, all
front panel keys are enabled.
SYSTem:REMOTE
This instruction causes the instrument enter remote state from local state. After receipt of this
command, all front panel keys are disabled except of the local key (shift key) which can be used to
manually revert to local state.
NOTE: All command strings send to the instrument must terminate with a <new line>
character. (0x0A or \n).
Wiring diagram of RS232 cable required for remote control (null modem configuration)
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
DB9 connector
DB9 connector
4080 series Function & Arbitrary/Function Generators Instruction Manual
47
6. USER PROGRAMMABLE ARBITRARY WAVEFORM
MODULE
Product Overview
In addition to the 27 build-in waveforms, models 4084AWG and 4086AWG also provide the
capability to generate and store up to 8 user-defined waveforms in non-volatile memory.
This Windows-based software tool offers a graphical user interface that you can use to create, edit,
and transfer custom waveforms to the internal memory of the 4084AWG or 4086AWG
Function/Arbitrary Generator.
Waveforms can be created in several ways: Users can draw new waveforms with a mouse, edit one
of the predefined waveforms, read in waveform parameters from a text file or download waveforms
from a TEKTRONIX