This study evaluated an expert system that analyzes pacemaker diagnostic data to identify atrial tachyarrhythmias in patients without a prior history of them. The study found that the expert system detected previously unknown atrial tachyarrhythmias in 31% of 239 patients. Detection was highest in patients with high-degree AV block who had VDD pacemakers. The expert system's device programming suggestions were adopted in 30% of cases. Detection of arrhythmias by the system led to adjusted pharmacological management in most follow-up visits. The results show the expert system provides a valuable tool for detecting atrial tachyarrhythmias during pacemaker follow-ups.
This study evaluated an expert system that analyzes pacemaker diagnostic data to identify atrial tachyarrhythmias in patients without a prior history of them. The study found that the expert system detected previously unknown atrial tachyarrhythmias in 31% of 239 patients. Detection was highest in patients with high-degree AV block who had VDD pacemakers. The expert system's device programming suggestions were adopted in 30% of cases. Detection of arrhythmias by the system led to adjusted pharmacological management in most follow-up visits. The results show the expert system provides a valuable tool for detecting atrial tachyarrhythmias during pacemaker follow-ups.
atrial tachyarrhythmias in patients with no prior history of them A. Schuchert a, *, S. Lepage b , J.J. Ostrander c , R.J. Bos d , M. Gwechenberger e , A. Nicholls f , G. Ringwald g a Heart Center, University Hospital Hamburg-Eppendorf, Hamburg, Martinistrasse 52, D-20251 Hamburg, Germany b Centre Hospitalier Universitaire de Sherbrooke, Sherbrooke, QC, Canada c Grey Bruce Health Services Corporation, Owen Sound, ON, Canada d Franciscus Ziekenhuis, Roosendaal, The Netherlands e AKH, Wien, Austria f Leeds General Inrmary, Leeds, UK g Gemeinschafts Praxis, Bruchsal, Germany Submitted 27 September 2004, and accepted after revision 30 January 2005 KEYWORDS pacemaker diagnostics; expert system; atrial tachyarrhythmia Abstract Aims Modern pacemakers provide a large amount of diagnostic data. Given the limited time available during a pacemaker follow-up visit essential information may be overlooked. This registry was conducted to assess the utility of an expert system that analyses the diagnostic data collected by an implanted pacing device and noties and advises the physician about suspected technical issues and arrhythmias that need further attention. Methods Patients with various standard indications for pacing were included in this registry and received single or dual chamber pacemakers. Data were collected and analysed by the expert system during at least two subsequent follow-up visits. The evaluation of this system focused on data obtained from patients with a dual chamber pacing device without prior history of atrial arrhythmias. Results A total number of 239 patients without prior history of atrial tachyar- rhythmias were included in this analysis. Atrial tachyarrhythmias were detected in 73 (31%) of these patients. The highest incidence of newly detected arrhythmias occurred in the group of patients with high-degree AV block and VDD pacemakers. * Corresponding author. Tel.: C49 404 717 5304; fax: C49 404 717 7420. E-mail address: [email protected] (A. Schuchert). 1099-5129/$30 2005 The European Society of Cardiology. Published by Elsevier Ltd. All rights reserved. doi:10.1016/j.eupc.2005.01.005 Europace (2005) 7, 242e247
b y
g u e s t
o n
O c t o b e r
3 ,
2 0 1 3 h t t p : / / e u r o p a c e . o x f o r d j o u r n a l s . o r g / D o w n l o a d e d
f r o m
b y
g u e s t
o n
O c t o b e r
3 ,
2 0 1 3 h t t p : / / e u r o p a c e . o x f o r d j o u r n a l s . o r g / D o w n l o a d e d
f r o m
b y
g u e s t
o n
O c t o b e r
3 ,
2 0 1 3 h t t p : / / e u r o p a c e . o x f o r d j o u r n a l s . o r g / D o w n l o a d e d
f r o m
b y
g u e s t
o n
O c t o b e r
3 ,
2 0 1 3 h t t p : / / e u r o p a c e . o x f o r d j o u r n a l s . o r g / D o w n l o a d e d
f r o m
b y
g u e s t
o n
O c t o b e r
3 ,
2 0 1 3 h t t p : / / e u r o p a c e . o x f o r d j o u r n a l s . o r g / D o w n l o a d e d
f r o m
b y
g u e s t
o n
O c t o b e r
3 ,
2 0 1 3 h t t p : / / e u r o p a c e . o x f o r d j o u r n a l s . o r g / D o w n l o a d e d
f r o m
Furthermore, newly detected arrhythmias predominantly occurred in the period shortly after implantation. Device programming suggestions by the expert system were adopted in 30% of the cases. Following detection of atrial tachyarrhythmias by the expert system, pharmacological management was adjusted at 71% of the rst follow-up visits and at 27% of later follow-up visits. Conclusion Results of this registry show that this expert system provides a valuable tool for the detection of atrial tachyarrhythmias during pacemaker follow-up visits. 2005 The European Society of Cardiology. Published by Elsevier Ltd. All rights reserved. Introduction Modern pacemakers are capable of storing a large amount of diagnostic data related to various ar- rhythmias. One of the most striking advantages of having such diagnostic data is that it is collected on a 24-h day and 7-day week basis. This allows physicians to evaluate the appropriate function of the device, the status and progression of the patients arrhythmias and the effects of pacing and pharmacological therapies [1e4]. However, it also means that the data have to rst be retrieved and carefully analysed, which adds considerably to the length of each follow-up visit. Furthermore, correct interpretation of the retrieved data re- quires extensive experience with pacemaker man- agement, especially since specialized information in the form of textbooks or original reports is rare. One solution to this problem is to develop software that is capable of automatically analysing the data obtained during pacemaker interrogation. Such software, which is also referred to as an expert system, combines the expert knowledge that is embedded in the software with individual patient diagnostic data. It is designed to provide relevant messages if technical issues arise such as sensing failures or if signicant arrhythmic events occur. The progressively increasing functionality of modern pacemaker programmers allows for the introduction of such an expert system, known as Diagnostic Observations (DOs), into daily clinical routine. The DORE registry (Diagnostic Observation REg- istry) was designed to carry out a prospective evaluation of the performance of an expert system for the automatic analysis of diagnostic pacemaker data in a large patient population. Only patients with a newly implanted pace- maker that had the DOs feature were included in the registry. Since newly implanted devices in particular often require reprogramming, auto- matic systems that evaluate device performance and alert the physician to signicant arrhythmic events are expected to have the most utility during the immediate post implantation period. Methods Patients and devices Patients with a wide variety of pacemaker indica- tions, who received both single and dual chamber pacemakers, were included in the registry. No specic inclusion criteria applied to this registry. For the study we used the Clarity DDDR, Clarity SSIR and the Collection 3 series of pacemakers (Saphir VVD/R, Diamond DDDR, Ruby DDD and Topaz SSIR, Vitatron BV, Arnhem, The Netherlands). These are state-of-the-art pacemakers equipped with the DO function. These pacemakers continuously collect data on cardiac events and carry out measurements by, for example, periodically measuring the amplitude of atrial signals. This information is then stored in the pacemaker memory for later retrieval and auto- matic analysis by the DOs function. The diagnostic information used by the DOs function is identical to the information provided to the physician; the DOs function interprets these data and presents con- densed information and recommendations to the physician. Information related to possible atrial tachycardia is based on diagnostics, emanating from the mode switching algorithm, such as the amount of atrial sensed rhythm that is classied as pathological and the percentage of atrioventricular (AV) dissociation. Mode switching itself is an algo- rithm that has been used for many years in these devices and is based on a beat-to-beat analysis of the atrial rhythm [5]. Messages such as Low paced atrial rates (indicative of chronotropic incompetence), or Suspected atrial undersensing can be solved by reprogramming sensing parameters or activat- ing rate adaptive pacing. Messages indicat- ing signicant arrhythmic events may require additional intervention or adjustment of pacing parameters and/or pharmacological therapy. This is very important if the events are new or have not previously been seen. Particularly rele- vant is any suggestion of the presence of atrial Automatic analysis of pacemaker diagnostic data 243 tachyarrhythmias (AT) that the expert system detects on the basis of a relatively low percentage of AV synchrony (Less than 96% AV synchrony) in combination with mode switching (percentage of pathological atrial rates). Design One year of postoperative pacemaker follow-up data were used for the DORE registry. This could include up to four follow-up visits: one week after implantation, and after 3, 6 (as an option), and 12 months. Unscheduled follow-up visits were sepa- rately reported. At the start of each follow-up visit, patient symptoms were recorded and pacemaker data were retrieved. Upon retrieval, the expert system analysed the diagnostic information and recog- nized and responded to several events. The phys- ician was alerted to any notable events by an appropriate message (a Diagnostic Observation), which was displayed on the programmer screen. All DOs were documented at each follow-up. The physician was asked to conrm the presented message by reviewing the underlying diagnostic data, such as 24 h heart rate trend and various histograms and event counters. Subsequently, any changes in pacing and drug therapy and their relationship to the DOs were documented. Analysis The DORE registry included patients with a wide variety of pacemaker indications and included single and dual chamber pacemakers. The data analysis presented in this manuscript focuses on the detection of AT, the most frequent observation indual chamber devices. Our particular interest was in patients with no known history of AT and the proportion of these patients in whom the expert system detected the presence of this tachyarrhyth- mia. In addition we stratied the patients by in- dication for pacing. Results Patients Multiple centres from eight European countries and from Canada contributed to the registry. A total of 358 pacemaker patients were entered, comprising 192 (53%) males and 166 (47%) females. In the rst year after implantation, all patients underwent at least two follow-up examinations, with 46% of them having at least three. Hospital discharge data were available for 351 patients. The 1e3 months follow-up was reported for 276 pa- tients, the 5e7 months follow-up for 175 patients and the 10e14 months follow-up for 146 patients. Of all included patients, 280 patients with a dual chamber device had at least two scheduled follow- up visits from which diagnostic pacemaker data were available. Of these 280 patients, 145 were males (age 62.5 G12 years), and 135 were females (age 82.0 G8.4 years). At the time of pacemaker implantation, 36% of patients had bradycardiaetachycardia syndrome, 33% were in sinus rhythm, 30% had sinus node dysfunction and 1% had other atrial rhythms. With respect to the ventricular rhythm, 59% of patients had normal ventricular activation, 25% presented with idioventricular rhythms, 6% had frequent premature ventricular complexes and 10% had other ventricular rhythms. The indication for pacemaker implantation was high-degree AV block in 52% of patients, sinus node dysfunction in 47%, and other diagnoses in 1%. As mentioned earlier, we analysed only patients without a history of AT. Of the 280 dual chamber patients with diagnostic data from at least two follow-up visits, 239 patients had no prior history of atrial tachyarrhythmias. These patients were strat- ied according to the indication for pacing. We considered two main subgroups: those with a high- degree AV block (n Z168) and those without AV block, i.e. with intact AV conduction (n Z71). The rst group was subdivided into patients receiving a dual lead (DDD, n Z128) and a single lead (VDD; n Z40) dual chamber pacemaker. Frequency of diagnostic observations A total of 1433 follow-up visits were reported and in 323 (23%) one or more DOs were presented. During the follow-up period, 43% of all patients presented with at least one DOs. Fig. 1 shows that 43% of all DOs messages were related to a low percentage of AV synchrony (Less than 96% AV synchrony) and that 12% were related to High sensed atrial rates. Both mes- sages are indicative of the presence of AT. The device responds to high sensed atrial rates by means of mode switching; mode switching itself reduces the percentage of AV synchrony. Fig. 2 illustrates the distribution of DOs over the various follow-up visits. Although the absolute number of DOs decreases over time, the percent- age of patients with DOs does not substantially decline. This may be explained by the fact that during early follow-ups many DOs are related to 244 A. Schuchert et al. technical issues such as over- or undersensing. Once these issues are resolved, other issues that are possibly related to the occurrence of AT remain and will trigger the expert system. Non-scheduled follow-ups were usually due to signicant symp- toms. However, DOs obtained during these non- scheduled follow-ups did not frequently result in the detection of AT that was not previously known. Atrial tachyarrhythmia detection Patients with high-degree AV block and a DDD(R) pacemaker Table 1 shows the proportion of patients with a DDD(R) pacemaker and high-degree AV block and no prior history of AT in whom the expert system issued a message indicating that such an arrhythmia had indeed been detected. At the rst scheduled follow-up after approximately 3 months, AT was detected in 24 patients (19%) and within 1 year of follow-up AT was rst detected in 36 of these patients (almost 30%). Patients with high-degree AV block and a VDD pacemaker Table 2 shows the number of patients with high- degree AV block, a single lead VDD pacemaker and no prior history of AT in whom the expert system issued a message indicating that such a tachyar- rhythmia had been detected. At the rst scheduled follow-up visit after approximately 3 months, AT was detected in 14 patients (35%). In 50% of these patients, the presence of AT was detected for the rst time within 1 year. Patients without high-degree AV block Table 3 shows the number of patients with intrinsic AV conduction and no prior history of AT in whom the expert system issued a message indicating that such a tachyarrhythmia had been detected. At the rst scheduled follow-up visit after approximately 3 months, AT was detected in 10 patients (14%). In 22.5% of these patients, the presence of AT was detected for the rst time within 1 year. The cumulative incidence of newly detected AT stratied according to the indication for pacing is shown in Fig. 3. Combining all the data shows that in 73/239 patients (31%) with no prior history of AT, such an arrhythmia was detected over a 1-year follow-up period. AT is predominantly detected in the period shortly after implantation and dimin- ishes after the 3-month follow-up visit. These ndings are in line with earlier reports [6]. Response to diagnostic observations Responses to DOs were either pacemaker reprog- ramming, adaptation of the drug regimen or di- agnostic strategy. In response to DOs, the suggested programming recommendations were followed in 30% of cases. In 71% of the rst follow-ups and in 27% of the later follow-ups, the pharmacological manage- ment was adjusted after atrial tachyarrhythmias had been detected. < 96% AV synchrony Low paced atrial rates High sensed atrial rates Low paced ventricular rates Suspected atrial undersensing Others Figure 1 Incidence of diagnostic observations. Number of DO % of patients with DO D i s c h a r g e 1 - 3
m o n t h s 5 - 7
m o n t h s 1 0 - 1 4
m o n t h s U n s c h e d u l e d
f o l l o w - u p 0 20 40 60 80 100 120 140 0% 5% 10% 15% 20% 25% 30% 35% Figure 2 Distribution of DOs over the various follow-up visits. Table 1 Frequency of newly detected atrial tachyarrhythmias in patients with AV block and a DDD(R) pacemaker and no known previous history of atrial tachyarrhythmias (n Z128) Follow-up Patients with newly detected AT 1e3 months 24 (19%) 5e7 months 6 (5%) 10e14 months 6 (5%) Unscheduled follow-up 1 (0.8%) Automatic analysis of pacemaker diagnostic data 245 Discussion In 43% of all follow-up visits, the DOs function reported signicant diagnostic events. A major result of this study was the detection of AT in 31% of patients with no known history of them. AT may be asymptomatic and unrecognized in many patients as has been reported in previous studies. For instance, in the AIDA trial, 65% of the patients in whom atrial tachyarrhythmias were detected had no documented history of them. Detection of AT in asymptomatic patients is im- portant, given the risk of AT related stroke [7,8]. Most rst detections of AT occur shortly after implantation and diminish after the 3-month fol- low-up visit. These ndings are in line with earlier reports [9]. Since this registry was not designed to investi- gate the effects of atrial and ventricular pacing on the incidence of AT we did not statistically analyse our data with respect to this issue. However, the trends observed in this study with regard to the percentage of patients with AT in relation to the amount of atrial and ventricular pacing are consistent with earlier reports. Patients treated with dual chamber pacing have been reported to show a lower incidence of AT, compared with patients receiving ventricular pacing [10]. Consis- tently, the percentage of patients with AT de- tected by the pacemaker in combination with the expert system was considerably higher in the subgroup with a single lead VDD pacemaker, compared with patients with a dual lead dual chamber device. It is possible that the ventricular pacing which occurs at a lower rate in VDD mode contributed to this AT detection. In addition, in patients with a high-degree AV block, who are expected to receive a high amount of ventricular pacing, the percentage of newly discovered AT over a 1-year follow-up period was higher than in the group with intact AV conduction (30% versus 22.5%). This observation is in line with previous reports that conclude that frequent ventricular pacing correlates with a higher incidence of AT [11,12]. Atrial undersensing is a common issue with a VDD lead and may have been related to in- appropriate detection of atrial tachyarrhythmias. This would suggest that the amount of AT detected by the expert system in patients with a VDD device is an underestimation. The expert system noties the user with regard to suspected undersensing, based on the available diagnostic information. The expert system evaluated in this registry proved to be valuable in detecting new onset or asymptomatic episodes of AT. It needs to be emphasized that pacemaker DOs were particularly frequent during non-scheduled follow-up examina- tions of patients with signicant symptoms. How- ever, apart from one follow-up (see Table 1), the expert system did not detect AT in patients without a history of atrial tachyarrhythmias during these non-scheduled follow-up visits. This registry was conducted in order to assess the ability of the expert system to alert the user of issues potentially requiring adjustment in the therapy. The relation- ship between these adjustments in response to DOs and the effects on symptoms has not been in- vestigated during this study. Future studies should address to what extent the expert system may be helpful in effectively reducing symptoms. Table 2 Frequency of newly detected atrial tachyarrhythmias in patients with AV block and a VDD pacemaker and no known history of atrial tachyarrhythmias (n Z40) Follow-up Patients with newly detected AT 1e3 months 14 (35%) 5e7 months 4 (10%) 10e14 months 2 (5%) Unscheduled follow-up 0 (0%) Table 3 Frequency of newly detected atrial tachyarrhythmias in patients without AV block and no known history of atrial tachyarrhythmias (n Z71) Follow-up Patients with newly detected AT 3 months 10 (14%) 6 months 5 (7%) 12 months 1 (1.5%) Unscheduled follow-up 0 (0%) I n c i d e n c e
o f
n e w l y
d e t e c t e d
A T
[ % ] AVB / VDD AVB / DDD Non-AVB 0 10 20 30 40 50 1 - 3 mo 10 - 14 mo Follow up 5 - 7 mo Figure 3 Cumulative incidence of newly detected AT in patients with no prior history of AT. 246 A. Schuchert et al. Conclusion The expert system evaluated in this registry proved to be valuable in detecting new onset or asymptomatic episodes of AT in patients without prior history of AT. Acknowledgements The authors would like to acknowledge the collab- oration and commitment of all local investigating physicians and their staff. Austria: F. Rauscha, AKH Wien, Wien; A. Teubl, A.o. Krankenhaus Wr. Neustadt, Wr. Neustadt; K.H. Tscheliessnig, LKH Graz e Univ.-Klinikum, Graz; E. Vesely, A.o. Krankenhaus Waidhofen a.d. Thaya, Waidhofen a.d. Thaya; G. Zach, A.o. Krankenhaus Bruck a.d. Mur, Bruck a.d. Mur. Canada: R.I.G. Brown, Royal Columbian Hospi- tal, New Westminster, QC; R. Dong, Surrey Memo- rial Hospital, Surrey, BC; G. Evans, Hamilton H.S.C., Hamilton, ON; W.G. Hughes, The Kawartha Cardiology Clinic, Peterborough, ON; P. Polasek, Kelowna General Hospital, Kelowna, BC. France: C. Besson, Clinique Saint-Joseph, Alen- c on; H. Bonnet, Clinique dArgonnay, Pringy; J. Covillard, Clinique de Chenove, Dijon; Ch. Delaire, Centre Hospitalier, Cholet; E. Favre, Centre Hos- pitalier, Saint Vallier; J. Pellet, Clinique Belle- done, Grenoble; F. Senellart, Clinique des Ce`dres, Grenoble. Germany: U. Abdel-Rahman, Klin. d. J.-W.- Goethe Universitat, Frankfurt; A. Brattstrom, Kli- nikum Quedlinburg, Quedlinburg; H.-P. Rebeski, Kard. Gemeinschaftspraxis, Kiel; G. Ringwald, Gemeinschaftspraxis, Bruchsal; T. Unger, St. Sal- vator-Krankenhaus, Halberstadt. Sweden: E. Baselier, Malarsjukhuset, Eskilstuna. Switzerland: P. Dubach, Kantonsspital Chur, Chur. The Netherlands: L.H.M. Bouwens, Deventer Ziekenhuis, Deventer; D.J. van Doorn, Spaarne Ziekenhuis, Heemstede; H. Drost, Slingeland Zie- kenhuis, Doetinchem; G.M. Jochemsen, Ziekenhuis Tjongerschans, Heereveen; H.C. Klomps, Zorgnet- werk De Nieuweborgh, Weert; M.R. van der Linde, Ziekenhuis Nij Smellinghe, Drachten; F.J. de Nooyer, Ziekenhuis Bethesda, Hoogeveen; M.H.H. de Vaan, IJsselmeer Ziekenhuizen, lokatie Zuider- zee, Lelystad; C.J.P.J. Werter, St. Laurentius Ziekenhuis, Roermond. United Kingdom: J. Campbell-Cowan, J. Mc- Lenachan, J. Perrins, G. Reynolds, Leeds General Inrmary, Leeds; A. Cunnington, M. Payne, Manor Hospital, Walsall; J. Flint, P. Forsey, Wordsley Hospital, West Midlands; A.N. Sulke, Eastbourne District General Hospital, Eastbourne; W. McCrea, Great Western Hospital, Swindon; P. Rees, Great Ormond Street Hospital, London; N. Spyrou, Battle Hospital, Reading. References [1] Fitts SM, Hill MR, Mehra R, Gillis AM. High rate atrial tachyarrhythmia detections in implantable pulse genera- tors: low incidence of false-positive detections. The PA clinical trial investigators. Pacing Clin Electrophysiol 2000; 23:1080e6. [2] Sack S, Mouton E, Defaye P, et al. Automatic interpretation for diagnosis assistance. Improved detection and analysis of sensed and paced events in dual chamber pacemakers with extended memory function. A prospective multicen- ter trial in 626 patients. Herz 2001;26:30e9. [3] Plummer CJ, Henderson S, Gardener L, McComb JM. The use of permanent pacemakers in the detection of cardiac arrhythmias. Europace 2001;3:229e32. [4] Glotzer TV, Hellkamp AS, Zimmermann J, Lamas GA. Atrial high Rrte episodes detected by pacemaker diagnostics predict death and stroke. Circulation 2003;107:1614e9. [5] Lau CP, Leung SK, Tse HF, Barold SS. Automatic mode switching of implantable pacemakers: II. clinical perfor- mance of current algorithms and their programming. Pacing Clin Electrophysiol 2002;25:1094e113. [6] Defaye P, Dournaux F, Mouton E. Prevalence of supraven- tricular arrhythmias from the automated analysis of data stored in the DDD pacemakers of 617 patients: the AIDA study. Pacing Clin Electrophysiol 1998;21:250e5. [7] Ezekowitch MD, Kenneth E, Nazarian SM, et al. Silent cerebral infarction in patients with nonrheumatical atrial brillation. Circulation 1995;92:2178e82. [8] Savelieva I, Camm AJ. Clinical relevance of silent atrial brillation: prevalence, prognosis, quality of life, and management. J Int Card Electrophysiol 2000;4:369e82. [9] Garrigue S, Cazeau S, Ritter P, Lazarus A, Gras D, Mugica J. Incidence des arythmies atriales chez patients stimules au long cours en mode double chamber. Arch Mal Coeur Vaiss 1996;89:873e81. [10] Connolly SJ, Kerr C, Gent M, Yusuf S. Dual chamber versus ventricular pacing. Critical appraisal of current data. Circulation 1996;94:578e83. [11] Kristensen L, Nielsen JC, Mortensen PT, Pedersen OL, Pedersen AK, Andersen HR. Incidence of atrial brillation and thromboembolism in a randomized trial of atrial versus dual chamber pacing in 177 patients with sick sinus syndrome. Heart 2004;90:661e6. [12] Sweeney M, Hellkamp A, Ellenbogen A, et al. Adverse effect of ventricular pacing on heart failure and atrial brillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation 2003;107:2932e7. Automatic analysis of pacemaker diagnostic data 247