DETERMINATION OF COLOR L*a*b* BY SPECTROCOLORIMETER
               METHOD: T4400.315.01.WP
                          1
                                    TITANIUM DIOXIDE
                                        (Ti-Pure)
                         Determination of Color by Spectrocolorimeter
                                          (TP-512.1)
 I. Principle
   Color can be described and quantified in a number of different ways.
   The system used in this method is based on the CIE 1976 L*a*b* color
   scale. These color scales give measurements of color in units of
   approximate visual uniformity throughout the color solid. Where L*
   measures lightness approximately as the eye would evaluate it, and
   varies from 100 for perfect white to zero for black; and where a*
   measures redness when positive, gray when zero and greenness when
   negative; and b* measures yellowness when positive, gray when zero and
   blueness when negative.
II. Applicability
   A. General
      This procedure can be used to measure the "color" of titanium
      dioxide (TiO2) pigment.
   B. Specific
      This procedure applies to the determination of "color" of in process
      and finished titanium dioxide samples for process control and
      finished product specification. It also can be used in
      characterizing competitive samples. L* of plastic grades normally
      runs from 97 to 99.5; all other grades run from 97.5 to 99.7. TiO2
      samples will read close to the top of the scale and on occasions,
      readings of 100+ are possible.
III. Limitations
      None
IV. Sensitivity, Precision and Accuracy
      A. Sensitivity
      No data is available.
                                               2
B. Precision
  1. Single Operator
                          _
    The average analysis (X), standard deviation (s) and 95%
    confidence limits (95% CL) established for the single operator
    precision of the method were as follows:
                  _
                  X            s              95% CL
       L*                      99.32          0.035   ± 0.083
       a*        -0.81         0.012                  ± 0.028
       b*                      3.06           0.023   ± 0.054
    The above data were calculated from eight replicate analyses of
    one sample performed by one technician over a period of two
    days.
    2. Multiple Operator
                           _
    The average analysis (X), standard deviation (s) and 95%
    confidence limits (95% CL) established for the multiple
    operator precision of the method were as follows:
                  _
                  X             s              95% CL
       L*                      99.35          0.045   ± 0.092
       a*        -0.79         0.026                  ± 0.053
       b*                      2.96           0.071   ± 0.145
    The above data were calculated from thirty-two replicate
    analyses of one sample performed by eight technicians over a
    period of several days.
C. Accuracy
    No study has been made; however, the method has proved to be
    satisfactory based on experience.
                                          3
V. Special Apparatus
  1. Spectrocolorimeter using 0/45 degree geometry (typically 0 degree
     incident, 45 degree viewing) to exclude specular reflection, with
     light source filtered to approximate CIE illumination D65. Prefer a
     circumferential ring of optical sensors or illumination to eliminate
     effect of sample orientation in measurement. Sample port size
     variable from 10 to 50 mm.
     Instrument must be capable of generating CIE XYZ and CIE L*a*b*
     color measurements calculated relative to illuminant D65 and the CIE
     1964 10 degree standard observer. Preferred operator interface is
     PC control/cathode ray tube with keyboard input.
     Preferred optical processors should be capable of full spectrum
     scanning from 400 to 700 nm, with tristimulus integration based up to
     31 point spectral reflectance data with an effective bandpass of
     10 nm.
     Preferred Supplier/Model:
     Labscan spectrocolorimeter Model LS-5100, available from Hunter
     Associates Laboratory, Inc., 11495 Sunset Hills Road, Reston, VA
     22090.
     Alternate instruments must be qualified. Qualification standard is
     ability to pass ILC.
  2. Aluminum or brass rings, ~1-1/16 inch (27 mm) ID or larger, ~1-1/4
     inch (32 mm) OD, 3/8" (9-10 mm) high, with square and parallel ends
     (important). Supplier: Fabricated locally from aluminum pipe (See
     Appendix 1).
  3. Sample ring jig, fabricate locally; sketch attached in Appendix 1.
  4. Hydraulic lab press, minimum 3 ton capacity, platen size 4 to 6
     inches, stroke range 2 to 6 inches. Two column design, with
     pressure gauge.
     Press modified with steel or brass ram, 3 to 4 inches long, 1"
     diameter (fits cleanly inside sample rings) screwed into top
     stationary portion of press. (See attached sketch) Steel ram to
     be fabricated from 1" OD rod.
                                              4
      Supplier/Model:
            Carver Catalog No. 3889 15-ton auto “C” press (preferred) OR Fisher Scientific 13-
            872/Carver 12-ton mini “C” press or standard “C” press. From Carver, Inc., 1569
            Morris Street, P.O. Box 544, Wabash, Indiana 46992-0544, www.carverpress.com.
            Modified as purchased locally.
    5. Filter paper, Whatman #50, 5.5 cm or equivalent.
    6. Hand ram, constructed of brass stock or Delrin, about 3 to 4" long,
       1" dia., fits cleanly inside sample rings. (See Appendix 1)
 VI. Reagents
     None
VII. Special Safety Considerations
    A. Product Hazards
       No product hazards have been identified. However, titanium dioxide
       is regulated as an air contaminant and care should be exercised to
       minimize product dusting. Wipe up all spills with a damp paper
       towel. See MSDS #2816CR for more information on Titanium Dioxide.
    B. Procedure Hazards
       None
VIII. Procedure
    A. Operating Conditions
      1. When pressing exhibits of titanium dioxide using the automatic
         hydraulic press, the Dwell Time should be set at 2 seconds and the
         Pressure at 2000 psig. When pressing using the manual hydraulic
         press (Section VIII-D), apply approximately 2000 + 5 psig for
         2 + 1 seconds.
      2. Read the manufacturer's operating manual for both the spectro-
         colorimeter and the hydraulic press carefully. Also read the
         procedure T4400.042.01.WP (Calibration and Use of the
         Spectrocolorimeter).
                                                5
B. Calibration
   Calibrate the Labscan at the beginning of every shift and whenever the software is
   restarted. A complete calibration consists of calibrating the Labscan to the black and
   white tiles, hitching to the alternate ‘hitch’ standard, and checking the calibration by
   running the reference standard. All three of these steps should be taken each time the
   instrument is calibrated.
   Press new reference and alternate hitch standards at the beginning of each shift prior to
   calibrating the instrument. Refer to Standard Sample Management, T4400.010.09.WP
   for current sample and values. The standard is prepared the same as the samples in
   Sample Analysis D. 1-6.
   Use D-65 illuminant and 10 degree observer according to the procedure described in
   T4400.042.01.WP, Calibration and Use of the Labscan (5000 & 5100)
   Spectrocolorimeter or equivalent procedure for another instrument.
C. Sampling
   Samples are normally supplied by manufacturing in a paper bag but
   any clean dry container is satisfactory. No mixing is required.
D. Sample Analysis
   1. Clean all parts of the jig and sample ring before using with a clean dry paper towel.
   2. Place a new piece of the filter paper on the milled flat surface of the base of the jig.
      Do not reuse filter paper sides in pressing pellets. Assure that the base of the jig is
      smooth and free of indentations.
   3. Place a sample ring in the top section of the jig; place it over the bottom section
      And tighten so that the sample ring is lying flat. The top section can be secured
      So that the ring lies flat by tightening the thumb screws of opposite corners
      simultaneously.
   4. Fill sample ring with sample or standard and compress using the hand ram. Add
      sample and compress again so that the sample ring is almost full.
   5. Transfer the jig to hydraulic press and compact by applying 2000 + 5 psig pressure
      for 2 + 1 seconds. L*, a*, and b* are sensitive to pressure differences. If using a
      manual press, take precautions to ensure that the same amount of pressure is applied
      by all analysts and by the same analyst each time a color pellet is pressed.
                                             6
        6. Remove the jig from the press, remove the sample ring and examine the exhibit.
           There should be no pigment on the ring or visible flaws such as voids or cracks on the
           bottom face of the exhibit. Any imperfect samples should be removed from the ring
           and preparation repeated.
        7. Assure that Labscan port is free of contamination inside and outside and that pellet is
           level on base before placing sample pellet on reflectance port. If necessary, wipe the
           port with a clean dry paper towel or cloth.
        8. Read L*, a*, and b* values of the samples according to T4400.042.01.WP
           "Calibration and Use of Labscan (5000 & 5100) Spectrocolorimeter" or equivalent.
           The values are displayed digitally and should be read and recorded to the nearest 0.01
           unit.
      E. Calculations
                L*, a*, and b* values are reported directly as measured as described above.
                Color differences can be calculated from analysis of samples and a standard
                following the procedure outlined above. When a comparison against a color
                standard is required, ∆L*, ∆a*, and ∆b*, and ∆E (∆E is the total color difference)
                are calculated by the following four equations respectively:
                ∆L* = L*sample − L*s tan dard
                ∆a * = a *sample − a *s tan dard
                ∆b * = b sample
                         *
                                − bs* tan dard
                ∆E = (∆L* ) 2 + (∆a * ) 2 + (∆b * ) 2
IX.     Quality Control
        A. General
            1. Measure color (L*a*b*) on the current color reference standard each
               shift or as required by individual sites.
            2. Plot the values obtained from this analysis using statistical process control
               Techniques
                If the results of the reference standard are off aim, refer to the troubleshooting
                guide in Section IX, B.
                                                        7
B. Troubleshooting Guide
    Most color analysis problems are related to improper instrument calibration,
    Alternate Standard/Reference Standard pellet condition, or incorrect pellet press
    pressure.
      •   Always press a new alternate standard and reference standard at the beginning
          of each shift and insure that all pellets are pressed to 2000 pounds!
    Problem:
    The Reference Standard is out-of-CUSUM tolerance
    Action:
      • Remove and clean the port and lens with lens cleaning solution and kimwipes.
          Replace the port. Close and restart UNIVERSAL software. Refer to site SJP
          for operation of the Labscan. Standardize the Labscan with the Black and
          White tiles. Press a new Alternate Standard pellet and hitch the instrument.
          Read the Alternate Standard (Hitch Standard) as an unknown sample. The
          Alternate Standard should read very close to its Hitch value. Press a new
          Reference Standard pellet and analyze. If the Reference Standard values are
          still out-of-tolerance, proceed to the step below.
      •   Verify that the instrument is properly configured. The color scale should be
          CIELAB, illuminant should be D65, and observer should be 10°. If the
          Reference Standard values are still out-of-tolerance, proceed to the step
          below.
      •   Verify that the Alternate Standard and Reference Standard is the correct
          material by pressing new Standards from different containers. If the Reference
          Standard values are still out-of-tolerance, proceed to the step below.
      •   Set up and operate a backup Labscan.
    Problem:
    Low color on in-process/packed TiO2 samples or if a delta of consecutive sample
    results is +/-0.3 L*, +/- 0.3 a*, or +/- 0.8 b*.
    Action:
      • Verify instrument calibration by analyzing the Reference Standard. If the
          Reference Standard is out-of-tolerance, follow the steps in the above problem
          “The Reference Standard is out-of-tolerance”. Record the Reference Standard
          analysis result in the logbook and enter into CQMS.
      •   Insure that the sample is prepared correctly (if slurry) and proper pellet
          pressure is applied (2000 pounds). See the procedure T4400.315.06WP for
          TiO2 slurry preparation and T4400.315.01WP for color analysis.
                                        8
              •   Recheck the TiO2 samples that are low in color.
          Remember that proper instrument operation can be verified by pressing and
          analyzing the color reference standard. If the reference standard analysis result is
          typical, the instrument is functioning correctly.
X. Comments
   A. General
      None
   B. Specific
      None
XI. References
   A. General
     1. Instruction Manual for the Hunterlab Spectrocolorimeter.
     2. ASTM D-1535 Specifying Color by the Munsell system.
     3. ASTM D-1729 Visual Evaluation of Color Differences of Opaque
        Materials.
     4. ASTM D-2244 Instrumental Evaluation of Color Differences of
        Opaque Materials.
     5. ASTM E-308 Recommended Practice for Spectrophotometry and
        Description of Color in CIE 1931 System.
     6. MSDS No. 2816CR.
    B. Specific
     1. TP-512.1 (3/7/1975) "Instrumental Color Determination Using
        Du-Color Color Difference Meter" the first dry TiO2 method.
                                               9
       2. T4400.042.01.WP "Calibration and Use of Labscan (5000 & 5100)
          Spectrocolorimeter."
       3. T4400.010.09.WP "Standard Sample Management."
       4. It was decided by the TiO2 Test Committee, at the meeting held
          April 28 and 29, 1993, at Delisle, that a pigment standard would
          be used in the determination of Color L*a*b* rather than a
          standard tile.
          See Standard Sample Management, T4400.010.09.WP, for current
          sample identification and values.
XII. Appendix
       1. Jig for Pressing TiO2 Samples for Optical Measurement
       2. Revision Sheet
XIII. Approval/Certification
       Origin: Johnsonville Plant, March 10, 1975
       Prepared by: JV R&D
       Approved by: TiO2 Test Committee
       Supersedes: T4400.315.01.WP, dated October 22, 2001