MR Sequences
MR Sequences
Click on one of the Red or Green buttons below to view more information
RARE: Overview
RARE: Rapid Acquisition with Relaxation Enchancement
Synonyms
Turbo spin echo (TSE)
Fast spin echo (FSE)
GO TO RARE: Physics
GO TO RARE: Applications
BACK TO Sequence Map
RARE: Physics
Timing diagram
A timing diagram for a 2D RARE acquisition is shown in Figure 1. Note the multiple 180
pulses during each TR and the way in which phase-encoding changes from TR to TR encode
each echo to a new line in k space. Every second lobe on the phase-encoding gradient is a
"rewinder": an equal and opposite gradient that undoes the phase encoding of the prior
lobe.
Effective TE
As in conventional spin-echo imaging, contrast in a RARE sequence is controlled by TE
and TR. It is clear what TR is in a RARE sequence: simply the time between 90 RF pulses.
But what about TE? Every echo in the echo train seems to have its own TE. In the RARE
sequence, the "effective TE" is the TE associated with the echo that is written closest to the
center of k space. Because the center lines of k space have the most effect on contrast, the
TE of this echo is the effective TE (TEeff) for the sequence (Fig 2).
Artifacts
Note how the echoes in a RARE echo train fall off in amplitude with T2 decay (Fig 3). This
effect is particularly pronounced where tissue T2 is short and the echo train long. This
results in non-uniform weighting of lines stored in k space and image blurring. In a T2weighted RARE sequence, the effective TE must be long. Later echoes fill the inner lines of k
space, and earlier echoes fill the outer lines. The echoes filling the outer lines have decayed
less. This overemphasizes the data stored in the outer lines in k space, which contain detail
on the edges of the image. As result, an artifact called edge enchancement may be seen.
GO TO RARE: Overview
GO TO RARE: Applications
BACK TO Sequence Map
RARE: Applications
Applications of the RARE sequence include T1- or T2-weighted high-resolution neurologic and
orthopedic imaging and T2-weighted breath-hold examinations of the abdomen (Figs 13).
Figure 1: T2-weighted image in the transverse plane at the level of the ventricles demonstrates a
necrotic tumor in the right frontal lobe with marked surrounding vasogenic edema.
Figure 3: Shared-echo RARE images of a carpal tunnel. Proton-density (bottom) and T2weighted (top) images are obtained in the same acquisition.
GO TO RARE: Overview
GO TO RARE: Physics
BACK TO Sequence Map
HASTE: Overview
HASTE: Half-Fourier-Acquired Single-Shot Turbo Spin Echo
Synonyms
HASTE
Hitachi
Toshiba
Background: Filling all of k space in a single shot with a RARE sequence is possible.
However, considerable T2 decay takes place over the relatively long period of acquisition,
and single-shot RAREis effectively restricted to imaging fluid (as in MR
cholangiopancreatography examinations). The HASTE sequence is an adaptation
of RARE that reduces total acquisition time by acquiring only half of k space. Half filling of k
space is sufficient to generate an image.
Features: The HASTE sequence allows acquisitions of 2D slices in less than a second.
Because of its long echo train (all the echoes take place in a single TR), HASTE images
will be T2-weighted.
Artifacts/Problems: Blurring artifact and limited resolution.
GO TO HASTE: Physics
GO TO HASTE: Applications
BACK TO Sequence Map
HASTE: Applications
The HASTE sequence provides strong T2 weighting and is used in MR
cholangiopancreatography and MR urography to demonstrate stones and
other filling defects as areas of low signal intensity within bright fluid-filled
structures (Figure). High speed makes the HASTE sequence a suitable choice
for uncooperative patients.
GO TO HASTE: Overview
GO TO HASTE: Physics
BACK TO Sequence Map
HASTE: Physics
Symmetry in k space
Figure 1 shows filled space after an MR image acquisition (the values in k space are plotted
as gray levels).This symmetric property of k space means it is necessary to fill only half of k
space; the rest of the data can be inferred. The HASTE sequence uses this property of k
space to its advantage.
Half-Fourier in phase
HASTE imaging uses a technique called half-Fourier in phase. It collects only half the lines of
k space (Fig 2) and infers the rest. (In practice, a little more that half of k space is filled to
correct for some errors in the data).
Single shot
HASTE is a single-shot technique, acquiring sufficient data for an
entire image in only one TR (Fig 3).
GO TO HASTE: Overview
GO TO HASTE: Applications
BACK TO Sequence Map
GO TO SPIN-ECHO: Physics
GO TO SPIN-ECHO: Applications
BACK TO Sequence Map
As a result of the dephasing, the signal picked up by the receiver coil decays in the form of a free
induction decay (FID). Dephasing is caused by true T2 dephasing and other dephasing effects caused by
magnet non-uniformities and susceptibility variations. The combination of all these effects causes signal
decay with a time constant T2* (ie, the FID has an exponential envelope with a time constant T2*).
The 180 rephasing flip is only fully effective when (as in Fig 5) the degree of dephasing before and after
the application of the 180 flip is the same. Dephasing caused by suceptibility effects and magnet nonuniformites are quite stable and so can be reversed by the 180 flip. Dephasing caused by true T2 decay
is random in time and is not reversed by the rephasing pulse. As a result, the echo peak lies on a T2
decay curve (Fig 6) .
Echo amplitude
The echo amplitude contributed by a particular tissue voxel determines how bright that voxel will appear in
the final image. In Figure 8, note how TE and TR influence echo amplitude. The echo peak lies on a T2
decay curve. T1 and TR determine how much longitudinal recovery occurs during TR.
Figure 9: Image contrast in spin-echo imaging. T1 weighting: Use the sliders to set the shortest possible
TE and TR. Note how the echo amplitude differs mainly between the two tissues with the different T1
values. T2 weighting: Use the sliders to set the longest possible TE and TR values. Note how the echo
amplitude differs mainly between the two tissues with the different T2 values.
GO TO SPIN-ECHO: Overview
GO TO SPIN-ECHO: Applications
BACK TO Sequence Map
GO TO SPIN-ECHO: Overview
GO TO SPIN-ECHO: Physics
BACK TO Sequence Map
GE Medical Systems
IR, Fast IR
IR, TIR (turbo inversion recovery)
Hitachi
Toshiba
Background: The term inversion recovery refers to a class of sequences that begin each
TR with an inverting 180 RF pulse. After a delay TI (the inversion time), a host sequence is
initiated. The host sequence may be one of several sequence types. In practice, IR is
frequently combined with a fast spin-echo (eg, RARE) host sequence, which gives rise to
names such as turboIR, turbo FSE, and fast IR.
Features: The longitudinal magnetization at the start of the host sequence is determined
by the degree of recovery of the inverted magnetization. Contrast in an IR image is strongly
influenced by TI and tissue T1 (ie, the image is T1-weighted).
Artifacts/Problems: Imaging times can be long, since the TR value must be long enough
to allow recovery between inverting pulses.
Oddities: Two reconstruction methods are possible with IR: standard "magnitude"
reconstruction and "real" reconstruction (also called true IR or real IR). Magnitude IR
sequences show a behavior called contrast reversal, in which the normal appearance of
tissues on a T1-weighted image (bright for short-T1 tissue and dark for long-T1 tissue) is
reversed as the TI is reduced.
GO TO IR: Physics
GO TO IR: Applications
BACK TO Sequence Map
In this real IR reconstruction, the background gray represents the "zero" gray level. Darker
areas on the image represent "negative" signal and brighter areas "positive" signal.
GO TO IR: Overview
GO TO IR: Physics
BACK TO Sequence Map
Figure 2: The delay between the inversion pulse and the first excitation pulse in the host sequence
is called the inversion time. The longitudinal magnetization available to the host sequence depends
on TI and tissue T1.
Contrast behavior
The TR of an IR sequence will normally be long. This is necessary for longitudinal magnetiztion to
recover between inversion pulses. If the echo time in the host sequence is short, the primary
determinant of contrast behavior is the relative longitudinal magnetisation at time TI. Images will
therefore tend to be T1-weighted with contrast behavior controlled by TI.
Magnitude versus real reconstruction
Two image reconstruction methods are possible in IR imaging: magnitude and real (also called
phase-sensitive reconstruction) (Fig 3).
Figure 3: Pixel intensites in a magnitude image are influenced only by the magnitude of the
longitudinal component of magnetization at time TI, not by its direction. Pixel intensites in a real
image are influenced both by the magnitude and direction of the longitudinal component at time TI.
Contrast reversal
Magnitude-reconstructed IR images show a type of behavior called contrast reversal (Fig 4). The
real image shown displays typical T1-weighted image behaviour for all values of TI: The shorter-T1
white matter looks brighter than the longer-T1 gray matter.
The magnitude image behaves differently. At a certain value of TI, the gray and white matter
become isointense. At longer values of TI, the image behaves as expected: The white matter
appears brighter than the gray matter. However, at shorter values of TI, contrast reversal is seen,
with white matter appearing darker than gray matter.
GO TO IR: Overview
GO TO IR: Applications
BACK TO Sequence Map
FLAIR: Overview
FLAIR: Fluid-attenuated Inversion Recovery
Synonyms
FLAIR
GO TO FLAIR: Physics
GO TO FLAIR: Applications
BACK TO Sequence Map
FLAIR: Applications
The FLAIR sequence makes periventricular changes easier to visualize by
suppressing bright cerebrospinal fluid (Figure). In addition, cortical changes,
which can also be hard to visualize with T2 sequences, are made more
prominent.
GO TO FLAIR: Overview
GO TO FLAIR: Physics
BACK TO Sequence Map
FLAIR: Physics
The basic elements of a FLAIR sequence are shown in Figure 1. A 180 pulse is applied, followed
by a TI delay. The host sequence could be a conventional spin-echo sequence. However, as TR
values are long (to let longitundinal magnetization recover), fast host sequences are preferred to
reduce imaging time. A RARE host sequence is typical.
Fluid Suppression
After the inverting pulse in FLAIR imaging, longitudinal magnetization begins to recover according
to T1. The magnetization of a given tissue passes through zero after a time equal to approximately
69% of the tissue T1 value. If the TI is selected to equal this time, there is no magnetization
available from that tissue at the start of the host sequence (Fig 2). Signal from the tissue is
therefore suppressed. Unlike in the STIR sequence, most non-nulled tissues are not inverted at
time TI, so contrast reversal is not apparent (see Fig 1, STIR Physics).
Figure 2: In FLAIR imaging, the TI is chosen to match the zero crossing point of fluid (often
cerebrospinal fluid). In a 1.5-T magnet, the T1 of cerebrospinal fluid is approximately 4200
msec, so a TI of approximately 2900 msec will suppress the fluid signal. (In practice, the
exact nulling time depends on the parameters of the host seqeuence; the given figures are
for a theoretically infinite TR.)
GO TO FLAIR: OVERVIEW
GO TO FLAIR: Applications
BACK TO Sequence Map
STIR: Overview
STIR: Short TI Inversion Recovery
Synonyms
Siemens Medical Solutions, Philips
Medical Systems, GE Medical
STIR
Systems, Hitachi, Toshiba
GO TO STIR: Physics
GO TO STIR: Applications
BACK TO Sequence Map
STIR: Applications
The STIR sequence is useful for suppressing marrow fat to reveal bone
abnormalities (Figure).
GO TO STIR: Overview
GO TO STIR: Physics
BACK TO Sequence Map
STIR: Physics
STIR sequence
The basic elements of a STIR sequence are shown in Figure 1. A 180 pulse is applied, followed by
a TI delay. The host sequence could be a conventional spin-echo sequence. However, since TR
values are long in spin-echo sequences (in order to let longitundinal magnetization recover), fast
host sequences are preferred to reduce imaging time. A RARE host sequence is typical.
Figure 2: In STIR imaging, TI is chosen to match the zero crossing point of fat. In a 1.5-T
imager, the T1 of fat is approximately 230 msec, so a TI of approximately 160 msec will
suppress fat signal. (In practice, the exact nulling time depends on parameters of the host
sequence; these figures are for a theoretically infinite TR). Note from the animation how the
Mz of other tissue (in this case white matter) is also reduced at time TI, and gives reduced
signal in the host sequence.
GO TO STIR: Overview
GO TO STIR: Applications
BACK TO Sequence Map
Hybrid EPI
Toshiba
Gradient echo
Hitachi
Field echo
Background: The term gradient echo refers to an entire class of sequences. These
sequences use a gradient to generate an RF echo, rather than a 180 refocusing pulse, as
is the case in spin-echo sequences. Gradient-echo sequences generally use an excitation
flip angle of less than 90.
Features: Gradient-echo sequences allow faster imaging than spin-echo sequences and
are frequently used in fast 3D imaging. The echo peak of a gradient echo lies on a T2*
curve, rather than on a T2 curve as in spin-echo imaging. Since the T2* value of tissue is
sensitive to variations in magnetic field strength, gradient-echo images are sensitive to
factors perturbing the static field. A benefit of this is good sensitivity to hemorraghic lesions,
since magnetic suceptibility is disturbed by hemoglobin degradation products.
Artifacts/Problems: Being sensitive to T2*, gradient-echo sequences are more prone to
susceptibility artifact. At a tissue/air boundary or in the presence of a metal implant, this
may manifest as a signal void on the image.
T2* decay
Immediately after the application of an RF flip in the gradient-echo sequence, a transverse
component appears. This transverse component dephases and decays according to T2*,
generating an FID (Fig 2). T2* is a combination of true tissue T2 and decay caused by magnetspecific non-uniformities and local variations in magnetic suceptibility (eg, at an air/tissue interface).
Echo formation
In a gradient-echo sequence, the frequency-encoding gradient generates the echo. The first lobe of
the gradient dephases the FID rapidly; the second lobe rephases the FID. Note now the echo peak
lies on the T2* decay curve (ie, the envelope of the original FID) (Fig 4). In spin-echo imaging, the
echo peak lies on the T2 decay curve. A basic gradient-echo sequence is sensitive to anything that
disturbs T2* (eg, poor magnetic uniformity, susceptibility effects due to metal prostheses, presence
of hemoglobin degradation products in hemorrhage).
Figure 1: T2* sensitivity as a source of artifact in gradient-echo images. This fatsuppressed T2-weighted gradient-echo image shows susceptibility artifact from a
left hip prosthesis.
The speed and variety of contrast behavior in gradient-echo imaging allow a range of
applications not possible with spin-echo imaging (Figs 3, 4).
Figure 3: MIP from a time of Flight (TOF) MR angiographic study of the circle of Willis. TOF
sequences use gradient echoes to generate high signal intensity from fresh blood flowing into a
slice or slab, while background stationary tissue becomes relatively saturated.
Siemens Medical
Solutions
Philips Medical Systems
GE Medical Systems
Hitachi
Toshiba
Figure 1: The animation shows the residual transverse magnetization remaining after
application of a single RF flip (flip angle is 35 in this case).
Spoiling
In RF spoiling, the phase of each RF pulse is varied from TR to TR. Use the animation in
Figure 2 to examine what is meant by the phase of the RF pulse. At each TR, the "new"
transverse component will be out of phase with any residual component from the last TR.
This prevents the RF flip from feeding the transverse component, and so a transverse
steady-state magnetization cannot build up. Gradient spoiling is an alternative method of
spoiling the transverse conponent. By applying a gradient, the transverse component can
be dephased. In practice, RF spoiling produces more uniform and effective spoiling.
Steady state
A normal gradient-echo sequence consists of repeated RF flips at intervals TR. In a spoiled
gradient-echo sequence, we need only consider the longitudinal component after each
applied flip, since the transverse component is deliberately destroyed. After a sufficient
number of RF pulses, the longitudinal magnetization settles into a steady state (Fig 3).
Figure 3: In the animation, the yellow graph shows the magnitude of the longitundinal
magnetization (Mz) in response to a series of RF pulses. Each sharp drop is due to an RF
pulse. The interpulse increases in magnitude are due to T1 recovery. Mz eventually settles
into a steady state, where recovery just offsets the effect of the applied flip at each TR. Try
varying the flip angle. Note how this affects the amplitude of the steady state.
Ernst angle
For any given ratio of TR/T1, there is an optimum flip angle that will give the strongest
transverse component at each applied RF pulse and therefore the strongest signal. This is
called the Ernst angle. The concept of the Ernst angle can be explored by using the
animation in Figure 4.
Figure 4: The yellow graph shows the longitudinal magnetization at the start of a spoiled
gradient-echo sequence. The orange spikes show the transverse magnetization
immediately after each flip. Try varying the flip angle. Note, for example, when the flip angle
is 90, all the longitudinal magnetization appears in the transverse plane after the first flip.
Very little longitundinal magnetization will have recovered for the next flip. The Ernst angle
in this case (TR/T1 = 0.1) is approximately 25. Set the flip angle to about 25 and note
how this flip angle gives the highest transverse signal once the system reaches steady
state.
Timing diagram
Figure 5 shows the basic elements of a 2D spoiled gradient-echo sequence.
Figure 5: Timing diagram of a FLASH sequence. Note that the phase-encoding gradient is
"rewound" at the end of each TR by an equal and opposite gradient. Without a rewinder,
the effectiveness of RF spoiling would vary across the image.
Figure 1: Source image (left) and MIP (right) in contrast-enhanced MR angiography of the
renal arteries. Short imaging times combined with centric k-space filling allow acquisition of
pure arterial images.
Figure 2: Transverse FLASH image of the brain above the level of the ventricles
demonstrates an old area of hemorrhage in the right frontal lobe.
Transverse vectors of fat and water fall in and out of phase periodically because of "chemical
shift." Chemical shift is the small difference in precession frequency between fat and water
(~210 Hz at 1.5 T). This may be applied usefully in in-phase and out-of-phase images
generated by running two FLASH sequences at slightly different TE values (2.4 and 4.8 msec
at 1.5 T). For voxels containing fat/water mixtures, the signal is reduced in the out-of-phase
image relative to the in-phase image. In spin-echo imaging, rephasing by the 180 pulse does
not allow this phenomenon to be demonstrated. The technique is useful for detecting fatcontaining structures such as the adrenal adenoma shown in Figure 3.
Figure 3: In-phase (left) and out-of-phase (right) FLASH images show adreanl adenoma.
GO TO SPIN-ECHO: Overview
GO TO Gradient Echo: Physics
BACK TO Sequence Map
Background: Gradient-echo sequences that acquire an image in less than a second are
considered ultrafast sequences. Achieving adequate signal and contrast in a simple very
short TE, very short TR gradient-echo sequence is a problem; for good T1 contrast, a
larger flip angle is needed, but the optimal flip angle for a good signal-to-noise ratio (SNR)
(ie, the Ernst angle) at short TRs is small. Ultrafast sequences overcome this problem by
incorporating a magnetization preparation phase prior to the gradient-echo host. Typically,
a 180 RF pulse is applied in the preparatory phase. The 180 flip introduces increased T1
weighting. All lines in k space are acquired after this single preparatory pulse.
Features: Used for high-speed T1-weighted imaging
Figure 1: Timing diagram for an ultrafast gradient-echo sequence (eg, turbo FLASH). The
basic structure of the sequence is a preparatory phase (shaded purple) followed by
repeated gradient-echo acquisitions (shaded green). In, for example, an image with 256
pixels in the phase-encoding direction, these acquisitions will be repeated 256 times to
complete k-space filling. The entire length of image acquisition will be on the order of a
second or less. In the animation, only the first three lines of k space are shown being filled.
TI in turbo FLASH
TI is the time to the midpoint of the acquisition phase of the sequence, where the middle
lines of k-space are being filled (Fig 2).
Figure 2: Note how the inversion pulse introduces T1 weighting. TI corresponds to the time
of filling of the middle lines in k space. (The plot of Mz is somewhat simplified: The effect of
the repeated RF flips on recovery during the acquisition period has been omitted).
3D RSSG (radiofrequency-spoiled
steady state acquisition with
rewound gradient echo)
3D FFE (fast field echo)
Background: This is the 3D variant of the ultrafast gradient-echo sequence (eg, turbo
FLASH [fast low-angle shot], TFE, FFE).
As in two dimensions, a preparatory pulse introduces T1 weighting. Usually, the imaging
time for a 3D gradient-echo sequence would be too long for a single preparatory pulse at
the start of the sequence to maintain its effect. Instead, several preparatory pulses are
applied over the course of the acquisition, with new lines in k space acquired after each
pulse.
Features: Used for high-speed 3D T1-weighted imaging.
GO TO SPIN-ECHO: Physics
GO TO SPIN-ECHO: Applications
BACK TO Sequence Map
Timing diagram
In the 3D ultrafast gradient-echo sequence, there is a preparatory phase at each repetition
time (TR), before the collection of several lines in 3D k space (Fig 1). Typically, the k-space
lines for an entire partition of k space may be collected in a single TR period (a partition is a
slice though 3D k space corresponding to all the slice-encoding steps for a given phaseencoding step).
Figure 1. MP-RAGE timing diagram. In the preparatory phase (shaded purple) an inversion
pulse is applied. Repeated gradient-echo acquisitions follow (shaded green). In each TR, the
slice (depth) encoding is varied to acquire all lines in the slice-encoding direction. The phaseencoding gradient is varied from TR to TR. For illustration, there are only four slices in the case
shown.
Figure 2. The upper graph shows the effect of the inversion pulse on longitudinal
magnetization (Mz) over the acquisition period. The inversion time (TI) is the time from the
inversion pulse to the midpoint of the acquisition, where the mean or middle partition is filled.
Note how the inversion pulse introduces T1 weighting. After data acquisition, there is a
recovery period for the longitudinal magnetization prior to application of the next preparatory
pulse.
EPI: Overview
EPI: Echo Planar Imaging
Synonyms
EPI
GO TO EPI: Physics
GO TO EPI: Applications
BACK TO Sequence Map
EPI: Applications
EPI sequences and sequence variants form the basis of practical diffusion-weighted
imaging. (Figs 1, 2), perfusion imaging, and functional MR imaging.
GO TO EPI: Overview
GO TO EPI: Physics
BACK TO Sequence Map
EPI: Physics
Timing diagram
A timing diagram for an EPI sequence is shown in Figure 1. Note how only one RF flip is applied
(90), and all k-space lines can be acquired in a single shot. In the sequence shown, the alternating
frequency-encoding gradient sweeps the location in k-space from side to side. A weak constant
phase-encoding gradient simultaneously sweeps the k-space location from bottom to top. The weak
pixel-to-pixel frequency difference in the phase-encoding direction emphasizes chemical shift
artifact. To avoid this, EPI sequences are normally used with fat supression.
Image distortion
The shallow phase encoding means there is a very narrow bandwidth across the tissue in the
phase-encoding direction. In other words, there is only a small pixel-to-pixel precession frequency
difference implementing phase encoding. In EPI, any effect that disturbs the local magnetic field by
even a small amount will have a proportionally large disruptive effect on these small frequency
differences. For example, at air/tissue interfaces, transistions in magnetic susceptibility will cause
geometric distortion of the image. Imperfect magnet shimming will also cause geometric distortion.
Ghosting
EPI is subject to a ghost arifact in which a ghost of the image appears shifted in the phaseencoding direction by a distance FOV (field of view)/2. This is called an N/2 or Nyquist ghost.
T2* decay
EPI generates gradient echoes under the FID curve created by the RF flip (Fig 2). As the curve
decays according to T2*, gradient EPI sequences willl be T2* weighted. Sensitivity to T2*, of
course, introduces sensitivity to artifacts caused by changes in magnetic susceptibility (eg,
air/tissue interfaces) and imperfect magnet shim.
In a single-shot EPI acquisition, all echoes for the image are acquired under a single T2* decay
curve. Echo spacings must be extremely short so that signal levels at the later echoes are not
completely degraded. Because all k-space lines are acquired at different stages of T2* decay, EPI
images are subject to a blurring effect in the phase direction.
Figure 2: Echoes in an EPI sequence form under the T2* decay curve.
Figure 3: The two frequency-encoding gradient lobes shown have the same area A under the curve
and so can perform the same function in the EPI sequence timing diagram. However, the upper
gradient is applied for a shorter time than the lower gradient and so can achieve shorter echo
spacing. In practice, this leads to a requirement for high-performance gradient hardware in EPI.
Blipped EPI
There are many different schemes for driving the phase- and frequency-encoding gradients during
EPI. The timing diagram for "blipped" EPI is shown in Figure 4. Here, a brief application of the
phase-encoding gradient between echoes nudges the trajectory in k space to a new row.
Figure 4: Blipped EPI. The trajectory corresponds directly to the rectilinear spacing of k-space
locations. Unlike the "zigzag'' trajectory, corrective measures such as reformatting of k space are
not necessary.
Spiral EPI
The timing diagram for spiral EPI is shown in Figure 5. The phase- and frequency-encoding
gradient waveforms are out of phase sinusoids.
Figure 5: Spiral EPI timing diagram. Sinusoidally driven gradients place less demands on the
gradient hardware. The spiral acquisition requires special k-space reformatting and recontruction
methods.
GO TO EPI: Overview
GO TO EPI: Applications
BACK TO Sequence Map
SSFP-FID: Overview
SSFP-FID: Steady State Free Precession FID
Synonyms
FISP (fast imaging with steady-state
precession)
GRASS (gradient-recalled acquisition in the
steady state)
GE Medical Systems
FFE (fast field echo)
SARGE (steady-state acquisition with
rewound gradient echo)
GO TO SSFP-FID: Physics
GO TO SSFP-FID: Applications
BACK TO Sequence Map
SSFP-FID: Physics
Timing diagram
The timing diagram in Figure 1 is for an SSFP-FID sequence. At first glance the diagram
does not appear to differ significantly from a spoiled gradient-echo (eg, FLASH)
sequence. However, there is a key difference: an absence of the spoiling mechansims (RF
or gradient) used in FLASH imaging.
Figure 2: The animation shows transverse components of spins immediately after an alpha
flip and the effect of the gradient on the spins (as used, say, in FISP or FLASH imaging).
Note how at the end of a TR period, a dephased set of spins is left. (Dephasing by T2, T2*,
etc, is ignored here: Only the effects of the gradients are considered.)
After many RF flips, the result is a complex interchange of the "histories" of dephasing,
rephasing, decay, and recovery between the transverse and longitundinal directions: The
remaining transverse magnetization feeds into the longitundinal magnetization at a
subsequent RF flip (Fig 3). "New" transverse magnetization from the longitudinal direction
mixes into the transverse plane at each flip. Any remaining transverse component can be
partially rephased by a subsequent RF flip. The conditions for the transverse component to
contribute to the steady state (ie, unspoiled gradient echo) are
1. A short TR
2. Constant dephasing during each TR
3. Same net application of gradients during each TR (to meet condition 2)
4. Constant flip angle
Signal pathways
A simplified view of the mixing of signal in an unspoiled SSFP sequence can be considered
by referring to Figure 4, which shows the response of tissue to a train of RF pulses during
an SSFP sequence. The action of the gradients has been ignored.
The signal received can be divided into two components. Not unexpectedly, there is an FID
component after each RF flip; but there is also a spin-echo component. This is because
each alpha flip has the potential to partially refocus magnetization from a prior FID. This is
analogous to the more complete refocusing of an FID seen with a 180 flip in spin-echo
imaging.
GO TO SSFP-FID: Overview
GO TO SSFP-FID: Applications
BACK TO Sequence Map
SSFP-FID: Applications
The SSFP-FID sequence is applied in 3D imaging during orthopedic
examinations (Figure).
GO TO SSFP-FID: Overview
GO TO SSFP-FID: Physics
BACK TO Sequence Map
SSFP-Echo: Overview
SSFP-Echo: Steady-State Free-Precession-Echo
Synonyms
PSIF (mirrored FISP)
CE-FFE-T2 (contrast-enhanced T2weighted FFE)
CE-GRASS (contrast-enhanced
gradient-recalled-echo in the steady
state)
GO TO SSFP-Echo: Physics
GO TO SSFP-Echo: Applications
BACK TO Sequence Map
SSFP-Echo: Physics
Timing diagram
In an SSFP-echo type sequence (eg, PSIF), the gradients act to rephase the echo part the
signal rather than the FID part as in SSFP-FID (eg, FISP). The timing diagram for an
SSFP-echo sequence is shown in the Figure.
SSFP-echo timing diagram. The echo produced is a spin echo and shows T2 rather than
T2* weighting. The echo is in effect a rephasing of transverse components generated by
the RF flip in the previous TR. This gives, effectively, a long TE (approximately twice the
TR) and heavy T2 weighting.
GO TO SSFP-Echo: Overview
GO TO SSFP-Echo: Applications
BACK TO Sequence Map
SSFP-Echo: Applications
The SSFP-echo sequence provides bright fluid, suitable for imaging
cerebrospinal fluid (Figure). Balanced SSFP sequences are now providing
similar bright fluid, but at an improved signal-to-noise ratio and with less
artifact sensitivity than SSFP-echo imaging.
GO TO SSFP-Echo: Overview
GO TO SSFP-Echo: Physics
BACK TO Sequence Map
DESS: Overview
DESS: Dual-Echo Steady State
Synonyms
Spin echo (SE)
Background: The DESS sequence generates both an SSFP-FID (eg, FISP) signal and an
SSFP-echo (eg, PSIF [mirrored FISP]) signal in a single TR period. The two generated
echoes are then combined to form a single image.
Artifacts/Problems: In the DESS sequence, the SSFP-echo component provides
additional T2 weighting to long-T2 tissues, showing bright fluid. The SSFP-FID component
shows other tissues with weighting similar to that obtained with a FISP sequence. The
sequence is used for high-resolution bright-fluid 3D examinations, particularly in
orthopedics.
GO TO DESS: Physics
GO TO DESS: Applications
BACK TO Sequence Map
DESS: Applications
The DESS sequence is used in orthopedic imaging, yielding high-resolution
images with bright fluid and good contrast between synovial fluid and cartilage
(Figure).
GO TO DESS: Overview
GO TO DESS: Physics
BACK TO Sequence Map
DESS: Physics
Timing diagram
In the DESS sequence, the gradients act to separate out the FISP element and the
PSIF echoes (see Fig 4, SSFP-FID Physics). The images can be combined to
form a single image. A DESS timing diagram is shown in the Figure.
GO TO DESS: Overview
GO TO Dess: Applications
BACK TO Sequence Map
Toshiba
Background: Balanced SSFP sequences are sequences in which the net phase change
caused by all three gradients (frequency, phase, and slice) is zero.
Features: Delivers contrast related to the ratio T2/T1. Tissues with a T2 value approaching
their T1 value appear brightest. Gives good blood/muscle contrast. Very fast sequence with
excellent signal-to-noise ratio.
Artifacts/Problems: The T2/T1 weighting has limited application. Prone to a banding
artifact caused by imperfect shimming or suceptibility effects; problem is minimized with
use of a short TR.
Oddities: While a balanced SSFP sequences settles into its steady state ("transient"
phase), its contrast will be a mix of proton-density and T2/T1 weighting. So a 2D image
may show different contrast than a 3D image (since a higher proportion of data acquisition
in the 2D case takes place during the transient phase). Balanced SSFP does not show high
sensitivity to T2* effects, as might a gradient-echo sequence.
GO TO SSFP: Physics
GO TO SSFP: Applications
BACK TO Sequence Map
The effects of balanced gradients can be studied by using the animation in Figure 2. Note
how net dephasing is zero at the end of TR. In a balanced SSFP sequence, all three
gradients are balanced in this way.
Figure 3: The animation shows the resultant vector in a steady-state balanced sequence
from TR to TR. The flip angle is alpha, with an alternating sign. The vector "settles down" to
a position alpha/2 around the longitudinal axis.
Banding artifact
The balanced steady state shown in Figure 4 is for conditions of perfect magnet uniformity.
In reality, small local differences in precession frequency cause small deviations from the
ideal case of zero net phase at the end of each TR. When the phase angle approaches
180, there is a rapid loss of the steady state and a decrease in signal. Areas of relatively
poor magnet uniformity or changes in susceptibility introduce sufficient dephasing to cause
a dark "banding artifact" on the image. The requirement to keep dephasing in any voxel to
less than 180 to avoid banding is stringent, and balanced SSFP requires good magnet
uniformity. If TR is kept short, less time is given for phase differences to develop and the
artifact will be reduced.
Figure 4: In practice, the ideal steady state (Fig 3) is disturbed by small local variations in
precession frequency. Where the phase change introuced by these varaiations is severe, a
rapid decrease in signal occurs and results in the appearance of a banding artifact.
GO TO SSFP: Overview
GO TO SSFP: Applications
BACK TO Sequence Map
Figure 1: Balanced SSFP image in a coronal plane through the liver. High signal intensity
is noted in the gall bladder and portal vein and within bowel loops. The sequence can be
used to demonstrate venous anatomy without contrast agent.
Figure 2: Balanced SSFP image in a coronal plane through the abdomen demonstrates
fetus
Good blood/myocardium contrast and high speed make balanced SSFP sequences
suitable for cardiac imaging applications (Fig 3).
GO TO SSFP: Overview
GO TO SSFP: Physics
BACK TO Sequence Map
CISS: Overview
CISS: Constructive Interference in the Steady State
Synonyms
CISS
FIESTA-C (fast imaging with steady-state
acquisition and phase cycling)
Background: Balanced SSFP sequences are subject to a "banding artifact." In the CISS
sequence, two balanced SSFP images in each acquisition are obtained. One image is
acquired with an "alternating phase" RF pulse as in a standard balanced SSFP sequence,
and the other is acquired with a nonalternating phase RF pulse. The banding artifact in
each image will be displaced relative to the banding artifact in the other image. By
combining the two images, the banding artifact is eliminated.
GO TO CISS: Physics
GO TO CISS: Applications
BACK TO Sequence Map
CISS: Physics
CISS and alternating phase
The CISS sequence is simply a combination of two balanced SSFP acquisitions.
The images are combined to remove banding artifact. The acquisitions differ in that
one is "phase alternating," and the other is not. This refers to the sign of the applied
RF pulse in each sequence (Fig 1): In one acquisition, the angle alternates at each
TR (eg, 70, -70, 70 etc), and in the other it does not. Phase alternation (or
cycling) is the "normal" case for a stand-alone balanced SSFP acquisition.
Figure 1: RF flip angle. In a phase-alternating acquisition, the sign of the flip angle
alternates.
GO TO CISS: Overview
GO TO CISS: Applications
BACK TO Sequence Map
CISS: Applications
CISS provides heavily T2-weighted images at high resolution. It is most
commonly used to demonstrate the seventh and eighth cranial nerves and the
cererbropontine angle (Figure).
GO TO CISS: Overview
GO TO CISS: Physics
BACK TO Sequence Map