0% found this document useful (0 votes)
200 views24 pages

TR88 Ac1e3x2c2000

The document discusses modular thermometers that can be installed in existing thermowells. It provides details on thermometer design, operation, and performance including measurement ranges and accuracy for RTD and thermocouple versions. Diagrams show examples of thermometer installations and measurement systems.

Uploaded by

Xavier Monroy
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
200 views24 pages

TR88 Ac1e3x2c2000

The document discusses modular thermometers that can be installed in existing thermowells. It provides details on thermometer design, operation, and performance including measurement ranges and accuracy for RTD and thermocouple versions. Diagrams show examples of thermometer installations and measurement systems.

Uploaded by

Xavier Monroy
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 24

TI01098T/09/EN/03.

15
71287921

Products

Solutions

Services

Technical Information
Omnigrad S TR88, TC88
Modular Thermometer

TR88 with resistance insert (RTD)


TC88 with thermocouple insert (TC)
Extension neck and threaded connection for
installation in an existing thermowell
Application
Universal range of application
Suitable for installation in already existing thermowells
Measuring range:
Resistance insert (RTD): 200 to 600 C (328 to 1 112 F)
Thermocouple (TC): 40 to 1 100 C (40 to 2 012 F)
Degree of protection up to IP68
Head transmitter
All Endress+Hauser transmitters are available with enhanced accuracy and reliability
compared to directly wired sensors. Easy customizing by choosing one of the
following outputs and communication protocols:

Analog output 4 to 20 mA
HART
PROFIBUS PA
FOUNDATION Fieldbus

Your benefits
High degree of flexibility thanks to modular design with standard terminal heads
as per DIN EN 50446 and customer-specific immersion lengths
Variable total length in suitable thermowells thanks to compression fitting on
extension neck
Types of protection for use in hazardous locations:
Intrinsic Safety (Ex ia)
Non-sparking (Ex nA)

Omnigrad S TR88, TC88

Function and system design


Measuring principle

Resistance thermometer (RTD)


These resistance thermometers use a Pt100 temperature sensor according to IEC 60751. The
temperature sensor is a temperature-sensitive platinum resistor with a resistance of 100 at
0 C (32 F) and a temperature coefficient = 0.003851 C-1.
There are generally two different kinds of platinum resistance thermometers:
Wire wound (WW): Here, a double coil of fine, high-purity platinum wire is located in a ceramic
support. This is then sealed top and bottom with a ceramic protective layer. Such resistance
thermometers not only facilitate very reproducible measurements but also offer good long-term
stability of the resistance/temperature characteristic within temperature ranges up to
600 C (1 112 F). This type of sensor is relatively large in size and it is comparatively sensitive to
vibrations.
Thin film platinum resistance thermometers (TF): A very thin, ultrapure platinum layer,
approx. 1 m thick, is vaporized in a vacuum on a ceramic substrate and then structured
photolithographically. The platinum conductor paths formed in this way create the measuring
resistance. Additional covering and passivation layers are applied and reliably protect the thin
platinum layer from contamination and oxidation, even at high temperatures.
The primary advantages of thin film temperature sensors over wire wound versions are their smaller
sizes and better vibration resistance. A relatively low principle-based deviation of the resistance/
temperature characteristic from the standard characteristic of IEC 60751 can frequently be observed
among TF sensors at high temperatures. As a result, the tight limit values of tolerance category A as
per IEC 60751 can only be observed with TF sensors at temperatures up to approx. 300 C (572 F).
For this reason, thin-film sensors are generally only used for temperature measurements in ranges
below 400 C (752 F).
Thermocouples (TC)
Thermocouples are comparatively simple, robust temperature sensors which use the Seebeck effect
for temperature measurement: if two electrical conductors made of different materials are connected
at a point, a weak electrical voltage can be measured between the two open conductor ends if the
conductors are subjected to a thermal gradient. This voltage is called thermoelectric voltage or
electromotive force (emf.). Its magnitude depends on the type of conducting materials and the
temperature difference between the "measuring point" (the junction of the two conductors) and the
"cold junction" (the open conductor ends). Accordingly, thermocouples primarily only measure
differences in temperature. The absolute temperature at the measuring point can be determined
from these if the associated temperature at the cold junction is known or is measured separately and
compensated for. The material combinations and associated thermoelectric voltage/temperature
characteristics of the most common types of thermocouple are standardized in the IEC 60584 and
ASTM E230/ANSI MC96.1 standards.

Endress+Hauser

Omnigrad S TR88, TC88

Measuring system

24V DC / 30 mA

4...20 mA

A
20-250V DC/AC

A0012641

1
A
B

Application example
Thermometer with fitted head transmitter mounted in an existing onsite thermowell
RIA16 field display unit - The display unit records the analog measuring signal from the head transmitter and
shows this on the display. The LC display shows the current measured value in digital form and as a bar graph
indicating a limit value violation. The display unit is looped into the 4 to 20 mA circuit and gets the required
energy from there. More information on this can be found in the Technical Information (see "Documentation").
Active barrier RN221N - The RN221N (24 V DC, 30 mA) active barrier has a galvanically isolated output for
supplying voltage to loop-powered transmitters. The universal power supply works with an input supply
voltage of 20 to 250 V DC/AC, 50/60 Hz, which means that it can be used in all international power grids.
More information on this can be found in the Technical Information (see "Documentation").

Equipment architecture

2
8

ML

ML

IL

A0012672

Thermometer design

1
2
3
4
5

Insert with head transmitter mounted (example with 3 mm (0.12 in))


Insert with terminal block mounted (example with 6 mm (0.24 in))
Complete thermometer with terminal head
Thermowell connection: Threaded connection on extension neck
Thermowell connection: Adjustable compression fitting on the extension neck. Possible maximum length of
the extension neck E as basic condition for the nominal calculation of the insertion length IL.
6
Thermowell connection: Adjustable compression fitting on the extension neck. The length E can be adjusted
during the installation.
7
Existing onsite thermowell located in the process
8
Version without extension neck, if thermowell and extension neck are present on site in the process (E = 0
mm)
E
Length of extension neck
IL Insertion length
ML Insertion length for existing onsite components

Endress+Hauser

Omnigrad S TR88, TC88

Thermometers from the Omnigrad S TR88 and TC88 series have a modular design. The terminal
head is used as a connection module for the mechanical and electrical connection of the insert. The
position of the actual thermometer sensor in the insert ensures that it is mechanically protected. If
installed in a thermowell the insert can be exchanged and calibrated without interrupting the
process. The insert has flying leads, a ceramic connection socket or mounted temperature
transmitter. The thermometers are designed for installation in an existing onsite thermowell.
Different threaded connections are available on the bottom of the extension neck for installation in
the thermowell. Insofar as the thermowell is suited to this purpose, the thermometer may also be
mounted using a suitable compression fitting on the extension neck. This means that thermometers
with a fixed insertion length (ML) can be used variably, even in thermowells of varying length, and
can be installed in such a way that an optimum thermal contact is guaranteed between the tip of the
insert and the bottom of the thermowell.
Measurement range

RTD: 200 to 600 C (328 to 1 112 F)


TC: 40 to 1 100 C (40 to 2 012 F)

Performance characteristics
Operating conditions

Ambient temperature range


Terminal head

Temperature in C (F)

Without mounted head transmitter

Depends on the terminal head used and the cable gland or fieldbus
connector, see 'Terminal heads' section

With mounted head transmitter

40 to 85 C (40 to 185 F)

With mounted head transmitter and 20 to 70 C (4 to 158 F)


display

Process pressure
The maximum process pressure depends on the thermowell into which the thermometer is screwed.
For an overview of the Endress+Hauser thermowells which may be used, see 22.
Permitted flow rate as a function of immersion length
The maximum permitted flow rate to which the thermometer can be subjected, reduces the greater
the immersion depth of the thermowell in the flowing medium. In addition, it is dependent on the
diameter of the tip of the thermowell, the medium type, process temperature and process pressure.
For an overview of the Endress+Hauser thermowells which may be used, see 22 .
Shock and vibration resistance
The Endress+Hauser inserts meet the requirements of IEC 60751 which specify shock and vibration
resistance of 3g in the range from 10 to 500 Hz. The vibration resistance at the measuring point
depends on the sensor type and design, see the following table:

Version

Vibration resistance for the sensor tip

Pt100 (WW or TF)

30 m/s (3g)

iTHERM StrongSens Pt100 (TF)

> 600 m/s (60g)

Endress+Hauser

Omnigrad S TR88, TC88

RTD resistance thermometer as per IEC 60751

Accuracy
Class

Max. tolerances (C)

Characteristics
1)

Cl. AA, former 1/3 (0.1 + 0.0017 |t| )


Cl. B

3.0

Cl. A

(0.15 + 0.002 |t|)

2.5

Cl. B

(0.3 + 0.005 |t|)

Max. deviation (C)

2.0

Temperature ranges for compliance with the


tolerance classes
Wire wound
sensor (WW):

Thin-film version
(TF):
Standard
iTHERM
StrongSens

B
1.5

Cl. A

Cl. AA

100 to
+450 C

50 to +250 C

Cl. A

Cl. AA

30 to +300 C

0 to +150 C

30 to +300 C

0 to +200 C

1.0

A
AA

0.5

-200

-100

100

200

300

400

500

600C

AA
- 0.5

A
- 1.0

- 1.5

- 2.0

- 2.5

- 3.0 Max. deviation (C)


A0008588-EN

1)

|t| = absolute value C

In order to obtain the maximum tolerances in F, the results in C must be multiplied by a factor
of 1.8.
Permissible deviation limits of thermoelectric voltages from the standard characteristic for
thermocouples as per IEC 60584 or ASTM E230/ANSI MC96.1:
Standard

Type

Standard tolerance

Special tolerance

Class Deviation

Class Deviation

J (Fe-CuNi)

2.5 C (40 to 333 C)


0.0075 |t| 1) (333 to 750 C)

1.5 C (40 to 375 C)


0.004 |t| 1) (375 to 750 C)

K (NiCr-NiAl)

2.5 C (40 to 333 C)


0.0075 |t| 1) (333 to 1 200 C)

1.5 C (40 to 375 C)


0.004 |t| 1) (375 to 1 000 C)

IEC 60584

1)

|t| = absolute value C

Standard
ASTM E230/ANSI
MC96.1

1)

Endress+Hauser

Type

Standard tolerance

Special tolerance

Deviation, the larger respective value applies


1.1 K or 0.004 |t| 1) (0 to
760 C)

J (Fe-CuNi)

2.2 K or 0.0075 |t| 1) (0 to 760 C)

K (NiCrNiAl)

2.2 K or 0.02 |t| 1) (200 to 0 C)


1.1 K or 0.004 |t| 1) (0 to
2.2 K or 0.0075 |t| 1) (0 to 1 260 C) 1 260 C)

|t| = absolute value C

Omnigrad S TR88, TC88

Response time

Tested in accordance with IEC 60751 in flowing water (0.4 m/s at 30 C):
Insert:
Sensor type

Diameter ID

Response time

Thin film (TF)

iTHERM StrongSens

6 mm (0.24 in)

t50

< 3.5 s

t90

< 10 s

t50

2.5 s

t90

5.5 s

t50

5s

t90

13 s

t50

2s

t90

6s

t50

4s

t90

12 s

t50

0.8 s

t90

2s

t50

2s

t90

5s

t50

1s

t90

2.5 s

t50

2.5 s

t90

7s

3 mm (0.12 in)
TF Sensor

6 mm (0.24 in)

3 mm (0.12 in)
WW Sensor

6 mm (0.24 in)

3 mm (0.12 in)
Thermocouple (TPC100)
grounded

6 mm (0.24 in)

3 mm (0.12 in)
Thermocouple (TPC100)
ungrounded

6 mm (0.24 in)

Response time for the sensor assembly without transmitter.

Insulation resistance

RTD:
Insulation resistance according to IEC 60751 > 100 M at 25 C between terminals and sheath
material measured with a minimum test voltage of 100 V DC
TC:
Insulation resistance according to IEC 1515 between terminals and sheath material with a test
voltage of 500 V DC:
> 1 G at 20 C
> 5 M at 500 C

Dielectric strength

Tested at a room temperature for 5 s:


6 mm (0.24 in): 1 000 V DC between terminals and insert sheath
3 mm (0.12 in): 250 V DC between terminals and insert sheath

Self heating

RTD elements are passive resistances that are measured using an external current. This
measurement current causes a self-heating effect in the RTD element itself which in turn creates an
additional measurement error. In addition to the measurement current, the size of the measurement
error is also affected by the temperature conductivity and flow velocity of the process. This selfheating error is negligible when an Endress+Hauser iTEMP temperature transmitter (very small
measurement current) is connected.

Calibration

Endress+Hauser provides comparison temperature calibration from


80 to +1 400 C (110 to +2 552 F) based on the International Temperature Scale (ITS90).

Endress+Hauser

Omnigrad S TR88, TC88

Calibrations are traceable to national and international standards. The calibration certificate is
referenced to the serial number of the thermometer. Only the insert is calibrated.
Insert:
6 mm (0.24 in) and 3 mm (0.12 in)

Minimum insertion length of insert in mm (in)

Temperature range

without head transmitter

80 to 40 C (110 to 40 F)

200 (7.87)

40 to 0 C (40 to 32 F)

160 (6.3)

0 to 250 C (32 to 480 F)

Material

with head transmitter

120 (4.72)

150 (5.91)

250 to 550 C (480 to 1 020 F)

300 (11.81)

550 to 1 400 C (1 020 to 2 552 F)

450 (17.72)

Extension neck, insert and process connection.


The temperatures for continuous operation specified in the following table are only intended as
reference values for use of the various materials in air and without any significant compressive load.
The maximum operation temperatures are reduced considerably in some cases where abnormal
conditions such as high mechanical load occur or in aggressive media.
Material
name

Short form

Recommended
Properties
max. temp. for
continuous use in
air

AISI 316L/
1.4404
1.4435

X2CrNiMo17-12-2
X2CrNiMo18-14-3

650 C (1 202 F) 1) Austenitic, stainless steel


High corrosion resistance in general
Particularly high corrosion resistance in chlorinebased and acidic, non-oxidizing atmospheres
through the addition of molybdenum (e.g.
phosphoric and sulfuric acids, acetic and tartaric
acids with a low concentration)
Increased resistance to intergranular corrosion and
pitting
Compared to 1.4404, 1.4435 has even higher
corrosion resistance and a lower delta ferrite
content

AISI 316Ti/ X6CrNiMoTi17-12-2 700 C (1 292 F)


1.4571

Alloy 600/
2.4816

1)

NiCr15Fe

Properties comparable to AISI316L


Addition of titanium means increased resistance to
intergranular corrosion even after welding
Broad range of uses in the chemical, petrochemical
and oil industries as well as in coal chemistry
Can only be polished to a limited extent, titanium
streaks can form

1 100 C (2 012 F) A nickel/chromium alloy with very good resistance


to aggressive, oxidizing and reducing atmospheres,
even at high temperatures
Resistant to corrosion caused by chlorine gas and
chlorinated media as well as many oxidizing
mineral and organic acids, sea water etc.
Corrosion from ultrapure water
Not to be used in a sulfur-containing atmosphere

Can be used to a limited extent up to 800 C (1472 F) for low compressive loads and in non-corrosive
media. Please contact your Endress+Hauser sales team for further information.

Components
Family of temperature
transmitters

Endress+Hauser

Thermometers fitted with iTEMP transmitters are an installation-ready complete solution to


improve temperature measurement by significantly increasing accuracy and reliability, when
compared to direct wired sensors, as well as reducing both wiring and maintenance costs.

Omnigrad S TR88, TC88

PC programmable head transmitters


They offer a high degree of flexibility, thereby supporting universal application with low inventory
storage. The iTEMP transmitters can be configured quickly and easily at a PC. Endress+Hauser offers
free configuration software which can be downloaded from the Endress+Hauser Website. More
information can be found in the Technical Information.
HART programmable head transmitters
The transmitter is a 2-wire device with one or two measuring inputs and one analog output. The
device not only transfers converted signals from resistance thermometers and thermocouples, it also
transfers resistance and voltage signals using HART communication. It can be installed as an
intrinsically safe apparatus in Zone 1 hazardous areas and is used for instrumentation in the
terminal head (flat face) as per DIN EN 50446. Swift and easy operation, visualization and
maintenance by PC using operating software, Simatic PDM or AMS. For more information, see the
Technical Information.
PROFIBUS PA head transmitters
Universally programmable head transmitter with PROFIBUS PA communication. Conversion of
various input signals into digital output signals. High accuracy over the complete ambient
temperature range. Swift and easy operation, visualization and maintenance using a PC directly from
the control panel, e. g. using operating software, Simatic PDM or AMS. For more information, see
the Technical Information.
FOUNDATION Fieldbus head transmitters
Universally programmable head transmitter with FOUNDATION Fieldbus communication.
Conversion of various input signals into digital output signals. High accuracy over the complete
ambient temperature range. Swift and easy operation, visualization and maintenance using a PC
directly from the control panel, e.g. using operating software such as ControlCare from Endress
+Hauser or NI Configurator from National Instruments. For more information, see the Technical
Information.
Advantages of the iTEMP transmitters:
Dual or single sensor input (optionally for certain transmitters)
Pluggable display (optionally for certain transmitters)
Unsurpassed reliability, accuracy and long-term stability in critical processes
Mathematical functions
Monitoring of the thermometer drift, sensor backup functionality, sensor diagnostic functions
Sensor-transmitter matching for dual sensor input transmitters, based on Callendar/Van Dusen
coefficients

Endress+Hauser

Omnigrad S TR88, TC88

Terminal heads

All terminal heads have an internal shape and size in accordance with DIN EN 50446 flat face and a
thermometer connection of M24x1.5, G1/2" or 1/2" NPT thread. All dimensions in mm (in). The
cable glands in the diagrams correspond to M20x1.5 connections. Specifications without head
transmitter installed. For ambient temperatures with head transmitter installed, see "Operating
conditions" section.

TA30A

Specification

15.5 (0.6)

68.5 (2.7)

107.5 (4.23)

28
(1.1)

78 (3.1)
A0009820

TA30A with display window

Specification
Available with one or two cable entries
Protection class: IP66/68 (NEMA Type 4x encl.)
Temperature: 50 to +150 C (58 to +302 F) without cable
gland
Material: aluminum, polyester powder coated
Seals: silicone
Threaded cable entry: G ", " NPT and M20x1.5
Protection armature connection: M24x1.5
Head color: blue, RAL 5012
Cap color: gray, RAL 7035
Weight: 420 g (14.81 oz)
with TID10 display
Ground terminal, internal and external
3-A marked

15.5 (0.6)

107.5 (4.23)

91.6 (3.61)

Available with one or two cable entries


Protection class: IP66/68 (NEMA Type 4x encl.)
Temperature: 50 to +150 C (58 to +302 F) without cable
gland
Material: aluminum, polyester powder coated
Seals: silicone
Threaded cable entry: G ", " NPT and M20x1.5;
Protection armature connection: M24x1.5
Head color: blue, RAL 5012
Cap color: gray, RAL 7035
Weight: 330 g (11.64 oz)
Ground terminal, internal and external
With 3-A symbol

28
(1.1)78 (3.1)
A0009821

TA30D

Specification

15.5 (0.6)

110 (4.3)

107.5 (4.23)

28
(1.1) 78 (3.1)
A0009822

Endress+Hauser

Available with one or two cable entries


Protection class: IP66/68 (NEMA Type 4x encl.)
Temperature: 50 to +150 C (58 to +302 F) without cable
gland
Material: aluminum, polyester powder coated
Seals: silicone
Threaded cable entry: G ", " NPT and M20x1.5
Protection armature connection: M24x1.5
Two head transmitters can be mounted. In the standard
version, one transmitter is mounted in the terminal head
cover and an additional terminal block is installed directly
on the insert.
Head color: blue, RAL 5012
Cap color: gray, RAL 7035
Weight: 390 g (13.75 oz)
Ground terminal, internal and external
With 3-A symbol

Omnigrad S TR88, TC88

TA30P

Spezifikation

114 (4.5)

41.5 (1.63)

!83

(3.3
)

11
(0.43)
68 (2.7)

Protection class: IP65


Max. temperature: 40 to +120 C (40 to +248 F)
Material: polyamide (PA), antistatic
Seals: silicone
Threaded cable entry: M20x1.5
Protection armature connection: M24x1.5
Two head transmitters can be mounted. In the standard
version, one transmitter is mounted in the terminal head
cover and an additional terminal block is installed directly on
the insert.
Head and cap color: black
Weight: 135 g (4.8 oz)
Types of protection for use in hazardous locations: Intrinsic
Safety (G Ex ia)
Ground terminal: only internal via auxiliary clamp

A0012930

TA20B

Specification

116 (6.54)

73 (2.87)

25 (1)

Protection class: IP65


Max. temperature: 40 to +80 C (40 to +176 F) without
cable gland
Material: polyamide (PA)
Cable entry: M20x1.5
Head and cap color: black
Weight: 80 g (2.82 oz)
3-A marked

84 (3.3)
A0008663

TA21E

Specification
Protection class: IP65 (NEMA Type 4x encl.)
Temperature: 40 to 130 C (40 to 266 F) silicone, up to
100 C (212 F) rubber seal without cable gland (observe
max. permitted temperature of the cable gland!)
Material: aluminum alloy with polyester or epoxy coating,
rubber or silicone seal under the cover
Cable entry: M20x1.5 or plug M12x1 PA
Protection armature connection: M24x1.5, G 1/2" or NPT
1/2"
Head color: blue, RAL 5012
Cap color: gray, RAL 7035
Weight: 300 g (10.58 oz)
3-A marked

57 (2.2)

95 (3.9)

122 (4.8)

28 (1.1)

84 (3.3)
A0008669

10

Endress+Hauser

Omnigrad S TR88, TC88

TA20J

Specification
Protection class: IP66/IP67 (NEMA Type 4x encl.)
Temperature: 40 to 70 C (40 to 158 F) without cable
gland
Material: 316L (1.4404) stainless steel, rubber seal under the
cover (hygienic design)
4 digits 7-segments LC display (loop powered with optional4
to 20 mA transmitter)
Cable entry: 1/2" NPT, M20x1.5 o plug M12x1 PA
Protection armature connection: M24x1.5 or 1/2" NPT
Head and cap color: stainless steel, polished
Weight: 650 g (22.93 oz) with display
Humidity: 25 to 95 %, no condensation
3-A marked

41 (1.61)

119 (4.7)*
89 (3.5)

114 (4.49)

The programming is executed through 3 keys at the bottom of


the display.

28 (1.1)
64 (2.52)
A0008866

* dimensions with optional display

TA20R

Specification

81 (3.2)
54 (2.13)

79 (3.11)

53 (2.1)

28 (1.1)

Protection class: IP66/67


Max. temperature: 40 to +100 C (40 to +212 F) without
cable gland
Material: SS 316L (1.4404) stainless steel
Cable entry: 1/2" NPT, M20x1.5 or plug M12x1 PA
Head and cap color: stainless steel
Weight: 550 g (19.4 oz)
LABS-free
3-A marked

!55 (2.17)

A0008667

Maximum ambient temperatures for cable glands and fieldbus connectors

Endress+Hauser

Type

Temperature range

Cable gland " NPT, M20x1.5 (non Ex)

40 to +100 C (40 to +212 F)

Cable gland M20x1.5 (for dust ignition-proof area)

20 to +95 C (4 to +203 F)

Fieldbus connector (M12x1 PA, 7/8" FF)

40 to +105 C (40 to +221 F)

11

Omnigrad S TR88, TC88

All dimensions in mm (in).


B

!42 (1.65)
!33 (1.3)

!44 (1.73)
33 (1.3)

10 (0.4)

10 (0.4)

M4x1

!6
(0.24)

!33 (1.3)
!42 (1.65)

ML

IL

ML

!11 (0.43) /
!12 (0.47)

D
180 (7.1) "
2x Pt100

90 (3.54) "
1x Pt100

Design

!ID

!ID

!ID

!ID
A0012662

3
A
B
C
D
E
IL
ML
ID

Dimensions of the Omnigrad S TR88 and TC88


Insert with terminal block mounted
Insert with head transmitter mounted
Insert with flying leads
Model without extension neck, intended for mounting in an existing onsite extension neck
Extension neck length
Total length of insert
Insertion length
Insert diameter

The insertion length (ML) must be selected on the basis of the total length and the type of
thermowell used.

12

Endress+Hauser

Omnigrad S TR88, TC88

Depending on the application different inserts are available for the assembly:

Insert
Sensor
Sensor design;
connection method
Vibration resistance of
the insert tip
Measuring range;
accuracy class

Standard thin-film

iTHERM StrongSens

1x Pt100, 3- or 4-wire, mineral 1x Pt100, 3- or 4-wire, mineral


insulated
insulated
Up to 3g

Wire wound
1x Pt100, 3- or 4-wire,
mineral insulated

Enhanced vibration resistance


> 60g

50 to +500 C
50 to +400 C
(58 to +752 F), Class A or AA (58 to +932 F), Class A or AA

2x Pt100, 3-wire, mineral


insulated

Up to 3g
200 to +600 C (328 to +1 112 F), Class A or AA

Diameter

3 mm ( in), 6 mm ( in)

6 mm ( in)

3 mm ( in), 6 mm ( in)

Insert type

TPR100

iTHERM TS111

TPR100

TC
Selection in order code

Sensor design; material

1x K; Alloy 600

2x K; Alloy 600

1x J; 316L

2x J; 316L

Measuring range acc. to:


DIN EN 60584

40 to 1 200 C

40 to 750 C

ANSI MC 96.1

0 to 1 250 C

0 to 750 C

TC Standard, accuracy

IEC 60584-2; class 1


ASTM E230-03; special

Insert type
Diameter

Weight

Endress+Hauser

TPC100
3 mm (0.12 in) or 6 mm (0.24 in), depending on the selected tip shape

0.5 to 2.5 kg (1 to 5.5 lbs) for standard options.

13

Omnigrad S TR88, TC88

Process connection

The thermometer is designed for installation in an existing onsite thermowell or in a thermowell


which can be ordered separately. The installation is done using the threaded connection on the
bottom of the extension neck or using a compression fitting.
Threaded connection

Cylindrical

Conical

Version

Thread length TL

Width
across flats
AF

12 mm (0.47 in)

17

M14x1.5
M18x1.5

SW/AF

M20x1.5

15 mm (0.6 in)

24

G 1/2"

15 mm (0.6 in)

27

NPT

NPT 1/2"

8 mm (0.32 in)

22

R 3/4"

8.5 mm (0.33 in)

27

TL

TL

24

ML,
L

R 1/2"

22

A0019445

Threaded compression
fitting (TA50)

G1/2" 47 mm
(1.85 in)

Clamping
ring
material

Max. process
temperature

Max. process
pressure

15 mm
(0.6 in)

SS316 1)

500 C (932 F)

40 bar at 20 C
(580 psi at 68
F)

F
A0016971

1)

SS316 compression ferrule: Can only be used once, the compression fitting cannot be repositioned on the
protection tube after loosening. Fully adjustable immersion length on initial installation.

When using a compression fitting the thermometer is pushed through a gland and fixed using a
metal ferrule (cannot be loosened).
Spare parts

The RTD insert is available as spare part TPR100 22


Der iTHERM StrongSens is available as spare part TS111 22
The TC insert is available as spare part TPC100 22
The inserts are made from mineral insulated cable (MgO) with a sheath in AISI316L/1.4404 (RTD)
or Alloy 600 (TC).
If spare parts are required, refer to the following equation:
Insertion length IL = E + L + 10 mm (0.4 in)
Extension neck welded with threaded connection to terminal head. DIN flat face, different
connections to separate thermowell, order number TN15-...
Compression fitting with thread, 12 mm (0.47 in), process connection thread G1/2", clamping
ring made of 316L stainless steel, order number TA50-KC

14

Endress+Hauser

Omnigrad S TR88, TC88

Wiring
Wiring diagrams for RTD

Type of sensor connection


Head mounted transmitter TMT18x (single input)
Power supply
head transmitter and
analog output 4 to 20 mA,
or bus connection
2

mA

1
3-wire

4-wire

RTD

RTD

6 (red)
5 (red)

6 (red)
5 (red)

4 (white)
3 (white)

3 (white)

A0016433-EN

Head mounted transmitter TMT8x (dual input)


Sensor input 2

Sensor input 1

RTD: 3-wire

RTD: 4- and 3-wire

Bus connection
and supply voltage

red
(black)
red
(black)

7
red

6
6

red

5
5

white
(yellow)

white

white

1
+
2
-

-2

Display connection
A0008848-EN

Terminal block mounted


1 x Pt100

2 x Pt100

1 x Pt100
red

red

red

red

white
white

4 wires

white

white

black

red

black

red

yellow

3 wires

3 wires
A0008591-EN

Wiring diagrams for TC

Endress+Hauser

Thermocouple wire colors


As per IEC 60584

As per ASTM E230

Type J: black (+), white (-)


Type K: green (+), white (-)

Type J: white (+), red (-)


Type K: yellow (+), red (-)

15

Omnigrad S TR88, TC88

Head mounted transmitter TMT18x (single


input)

Head mounted transmitter TMT8x (dual input)

Sensor
input 1
TC
7

Sensor
input 2
TC

Power supply
head transmitter and
analog output 4...20 mA
or bus connection

Bus connection
and supply voltage

1
+

5
5

1 +
2 -

1
+
-2

Display connection

A0012698-EN

A0012699-EN

Terminal block mounted

2 x TC

1 x TC
-

+
A0012700

16

Endress+Hauser

Omnigrad S TR88, TC88

Installation conditions
Orientation

No restrictions.

Installation constructions

ML
A

A0012639

Thermometer installation

The thermometer is designed for installation in an existing thermowell or in a thermowell which can
be ordered separately. Different threaded connections to suit the thermowell are available on the
thermometer's extension neck 14. The necessary insertion length (ML) of the insert depends
on the total length of the thermowell (A) and the type of thermowell used. It can be freely selected
within the range of 100 to 5 000 mm (3.94 to 197 in). Longer insertion lengths are available on
request. This also applies when ordering the insert as a spare part. More detailed information on
determining the insertion length (ML) required in each case can be found in the following table
(applies to Endress+Hauser thermowells with standard base thicknesses (D)).
Type of thermowell

ML in mm (in)

Type of thermowell

ML in mm (in)

TA535

ML = A

TA565

ML = A - 3 (0.12)

TA540

ML = A - 2 (0.08)

TA566

ML = A - 3 (0.12)

TA550

ML = A - 3 (0.12)

TA570

ML = A - 3 (0.12)

TA555

ML = A - 2 (0.08)

TA571

ML = A - 3 (0.12)

TA557

ML = A - 2 (0.08)

TA572

ML = A - 3 (0.12)

TW15

ML = A

TA575

ML = A - 3 (0.12)

TA560

ML = A - 3 (0.12)

TA562

ML = A - 3 (0.12)

TA576

ML = A - 2 (0.08)

In the case of thermowells with a noncompliant standard base thickness (D), the following formula
must be used: ML = A - D + 3 (0.12) in mm (in).

Endress+Hauser

17

Omnigrad S TR88, TC88

Extension neck length

The extension neck is the part between the process connection and the terminal head. As illustrated
in the following diagram, the extension neck length influences the temperature in the terminal
head. This temperature must remain within the limit values defined in the "Operating conditions"
section.

Terminal head heating

180
T in K
160
140
120
100
80
60
40
20
50
2

100
4

150
6

200
8

250
10

Process temperature
1000 C (1832 F)
800 C (1472 F)
600 C (1112 F)
400 C (752 F)
250 C (482 F)
300 mm
12 in

Extension neck length


A0012701-EN

Heating of the terminal head as a function of the process temperature. Temperature in terminal head =
ambient temperature 20 C (68 F) + T

Certificates and approvals


CE Mark

The device meets the legal requirements of the EC directives if applicable. Endress+Hauser confirms
that the device has been successfully tested by applying the CE mark.

Hazardous area approvals

For further details on the available Ex versions (ATEX, CSA, FM etc.), please contact your nearest
Endress +Hauser sales organization. All relevant data for hazardous areas can be found in separate
Ex documentation.

Other standards and


guidelines

EN 60079: ATEX certification for hazardous areas


IEC 60529: Degree of protection of housing (IP code)
IEC 61010-1: Protection Measures for Electrical Equipment for Measurement, Control, Regulation
and Laboratory Procedures
IEC 60751: Industrial platinum resistance thermometers
IEC 60584 and ASTM E230/ANSI MC96.1: Thermocouples
DIN EN 50446: Terminal heads
IEC 61326-1: Electromagnetic compatibility (EMC requirements)

PED approval

The thermometer complies with paragraph 3.3 of the Pressure Equipment Directive 97/23/EC and is
not marked separately.

Test report and calibration

The "Factory calibration" is carried out according to an internal procedure in a laboratory of Endress
+Hauser accredited by the European Accreditation Organization (EA) according to ISO/IEC 17025. A
calibration which is performed according to EA guidelines (SIT/Accredia or DKD/DAkks calibration)
may be requested separately. The calibration is performed on the replaceable insert of the
thermometer. In the case of thermometers without a replaceable insert, the entire thermometer from the process connection to the tip of the thermometer - is calibrated.

18

Endress+Hauser

Omnigrad S TR88, TC88

Ordering information
Detailed ordering information is available from the following sources:
In the Product Configurator on the Endress+Hauser web site: www.endress.com Choose your
country Products Select measuring technology, software or components Select product
(picklists: measurement method, product family etc.) Device support (right-hand column):
Configure the selected product The Product Configurator for the selected product is opened.
From your Endress+Hauser Sales Center: www.addresses.endress.com
Product Configurator - the tool for individual product configuration
Up-to-the-minute configuration data
Depending on the device: Direct input of measuring point-specific information such as
measuring range or operating language
Automatic verification of exclusion criteria
Automatic creation of the order code and its breakdown in PDF or Excel output format
Ability to order directly in the Endress+Hauser Online Shop

Endress+Hauser

19

Omnigrad S TR88, TC88

Accessories
Various accessories, which can be ordered with the device or subsequently from Endress+Hauser, are
available for the device. Detailed information on the order code in question is available from your
local Endress+Hauser sales center or on the product page of the Endress+Hauser website:
www.endress.com.
Communication-specific
accessories

Configuration kit TXU10

Configuration kit for PC-programmable transmitter with setup software and


interface cable for PC with USB port
Order code: TXU10-xx

Commubox FXA195
HART

For intrinsically safe HART communication with FieldCare via the USB interface.

Commubox FXA291

Connects Endress+Hauser field devices with a CDI interface (= Endress+Hauser


Common Data Interface) and the USB port of a computer or laptop.

For details, see "Technical Information" TI00404F


For details, see "Technical Information" TI00405C

HART Loop Converter


HMX50

Is used to evaluate and convert dynamic HART process variables to analog current
signals or limit values.
details, see "Technical Information" TI00429F and Operating Instructions
For
BA00371F

Wireless HART adapter


SWA70

Is used for the wireless connection of field devices.


The WirelessHART adapter can be easily integrated into field devices and existing
infrastructures, offers data protection and transmission safety and can be operated
in parallel with other wireless networks with minimum cabling complexity.

For details, see Operating Instructions BA061S


Fieldgate FXA320

Gateway for the remote monitoring of connected 4-20 mA measuring devices via a
Web browser.
details, see "Technical Information" TI00025S and Operating Instructions
For
BA00053S

Fieldgate FXA520

Gateway for the remote diagnostics and remote configuration of connected HART
measuring devices via a Web browser.
details, see "Technical Information" TI00025S and Operating Instructions
For
BA00051S

Field Xpert SFX100

Compact, flexible and robust industry handheld terminal for remote configuration
and for obtaining measured values via the HART current output (4-20 mA).

For details, see Operating Instructions BA00060S


Service-specific accessories

Accessories

Description

Applicator

Software for selecting and sizing Endress+Hauser measuring devices:


Calculation of all the necessary data for identifying the optimum measuring
device: e.g. pressure loss, accuracy or process connections.
Graphic illustration of the calculation results
Administration, documentation and access to all project-related data and
parameters over the entire life cycle of a project.
Applicator is available:
Via the Internet: https://wapps.endress.com/applicator
On CD-ROM for local PC installation.

20

Endress+Hauser

Omnigrad S TR88, TC88

Konfigurator+temperature

Software for selecting and configuring the product depending on the measuring
task, supported by graphics. Includes a comprehensive knowledge database and
calculation tools:
For temperature competence
Quick and easy design and sizing of temperature measuring points
Ideal measuring point design and sizing to suit the processes and needs of a wide
range of industries
The Konfigurator is available:
On request from your Endress+Hauser sales office on a CD-ROM for local PC
installation.

W@M

Life cycle management for your plant


W@M supports you with a wide range of software applications over the entire
process: from planning and procurement, to the installation, commissioning and
operation of the measuring devices. All the relevant device information, such as
the device status, spare parts and device-specific documentation, is available for
every device over the entire life cycle.
The application already contains the data of your Endress+Hauser device. Endress
+Hauser also takes care of maintaining and updating the data records.
W@M is available:
Via the Internet: www.endress.com/lifecyclemanagement
On CD-ROM for local PC installation.

FieldCare

FDT-based plant asset management tool from Endress+Hauser.


It can configure all smart field units in your system and helps you manage them. By
using the status information, it is also a simple but effective way of checking their
status and condition.

For details, see Operating Instructions BA00027S and BA00059S


System components

Accessories

Description

Field display unit RIA16

The display unit records the analog measuring signal from the head transmitter
and shows this on the display. The LC display shows the current measured value in
digital form and as a bar graph indicating a limit value violation. The display unit is
looped into the 4 to 20 mA circuit and gets the required energy from there.

For details, see the "Technical Information" document TI00144R/09/en


RN221N

Active barrier with power supply for safe separation of 4-20 mA standard signal
circuits. Offers bidirectional HART transmission.
details, see "Technical Information" TI00073R and Operating Instructions
For
BA00202R

RNS221

Supply unit for powering two 2-wire measuring devices solely in the non-Ex area.
Bidirectional communication is possible via the HART communication jacks.
details, see "Technical Information" TI00081R and Brief Operating
For
Instructions KA00110R

Endress+Hauser

21

Omnigrad S TR88, TC88

Documentation
Technical Information:
iTEMP Temperature head transmitter:
TMT180, PC-programmable, single-channel, Pt100 (TI088R/09/en)
PCP TMT181, PC-programmable, single-channel, RTD, TC, , mV (TI00070R/09/en)
HART TMT182, single-channel, RTD, TC, , mV (TI078R/09/en)
HART TMT82, two-channel, RTD, TC, , mV (TI01010T/09/en)
PROFIBUS PA TMT84, two-channel, RTD, TC, , mV (TI138R/09/en)
FOUNDATION FieldbusTM TMT85, two-channel, RTD, TC, , mV (TI134R/09/en)
Inserts:
Resistance thermometer insert Omniset TPR100 (TI268t/02/en)
Thermocouple insert Omniset TPC100 (TI278t/02/en)
iTHERM TS111 Insert for installation in thermometers (TI01014T/09/en)
Application example:
RN221N Active barrier, for supplying loop-powered transmitters (TI073R/09/en)
RIA16 Field display unit, loop-powered (TI00144R/09/en)
Technical information thermowells:
Type of thermowell
TA535

TI250T/02/en

TA565

TI160T/02/en

TA540

TI00166T/09/en

TA566

TI177T/02/en

TA550

TI153T/02/en

TA570

TI161T/02/en

TA555

TI154T/02/en

TA571

TI178T/02/en

TA557

TI156T/02/en

TA572

TI179T/02/en

TW15

TI00265T/02/en

TA575

TI162T/02/en

TA560

TI159T/02/en

TA576

TI163T/02/en

TA562

TI00230T/09/en

Supplementary ATEX documentation:


Omnigrad TRxx, TCxx, TSTxxx, TxCxxx; Omniset TPR100, TET10x, TPC100, TEC10x, iTHERM
TS111 ATEX II 3GD Ex nA (XA00044R/09/a3)
RTD/TC Thermometer Omnigrad TRxx, TCxx, TxCxxx, ATEX II 1GD or II 1/2GD Ex ia IIC T6...T1
(XA00072R/09/a3)
Inserts Omniset TPR100, TPC100, ATEX II 1G (XA087R/09/a3)
iTHERM TS111, TM211 Omnigrad TST310, TSC310 Omniset TPR100, TPC100 IECEx Ex ia IIC
T6T1 (XA00100R/09/a3)

22

Endress+Hauser

www.addresses.endress.com

You might also like