Pre-Installation Guide
For Millennia eV 5W/10W/15W Systems
Spectra-Physics, Inc.
3635 Peterson Way
Santa Clara, CA 95054
July 2014
Spectra-Physics Confidential Information
Table of Contents
Introduction .................................................................................................................................................. 3
1. When the System Arrives ....................................................................................................................... 4
1.1. Inspection ........................................................................................................................... 4
1.2. Review Instruction Manuals ............................................................................................... 4
1.3. Laser Safety Considerations ................................................................................................ 4
1.4. Diagnostics .......................................................................................................................... 4
2. Pre-installation considerations for Millennia eV Systems
2.1. Location and Environment .................................................................................................. 5
2.2. Physical Description ............................................................................................................ 6
2.3. General Specifications ........................................................................................................ 7
3. Pre-Installation Checklist ...................................................................................................................... 10
Spectra-Physics Confidential Information
Introduction
Congratulations on your purchase of a Spectra-Physics system. This guide describes preinstallation information for your laser system.
Some preliminary planning is essential to avoid unnecessary delays during installation and to
ensure easy operation and access to your system. You are requested to carefully consider your
operating environment prior to installation. Proper water, power and room temperature are
required for each system.
A checklist of pre-installation considerations is provided in this guide. You are responsible for
meeting these requirements prior to installation, with due consideration given to all applicable
building and safety codes.
We at Spectra-Physics intend to provide you with responsive support so that you can derive great
satisfaction and value in using our systems for your applications. We are available to you at
1-800-456-2552.
Spectra-Physics Confidential Information
When the system Arrives
1.1 Inspection
When the system arrives, any sign of damage to the shipping crates should be brought to the
attention of the delivering freight company. A claim must be filed with that commercial carrier
(usually within 30 days). Notify the originating Spectra-Physics office of any shipping damage.
Shipping damage is not covered by Spectra-Physics.
Your packing list will show all items that you have ordered. Open all the packages and check
each item for possible damage during shipping. Check the items against your packing list. Some
items may have been installed at the factory.
Each system comes with a manual, verify that you have received this item.
Please report any missing or damaged items to Spectra-Physics, or you may contact your SpectraPhysics Sales Engineer.
1.2 Review Instruction Manuals
Please read the manual to get vital information about your system. Familiarize yourself with the
system. You are encouraged to spend as much time as possible reviewing the system components
before your Spectra-Physics Service Engineer arrives for the installation and training.
1.3 Laser Safety Considerations
In addition to reviewing the sections in the manual regarding laser safety, be sure to have the
proper safety glasses available for ALL lab personnel present during the installation and testing of
your system. For more information, please call Spectra-Physics at 1-800-456-2552.
1.4 Diagnostics
During the course of installation, power measurements will be demonstrated on all of the
appropriate wavelengths. Your Service Engineer can identify which specifications will be
demonstrated and the equipment necessary to conduct such tests. A non-standard system will
require special consideration.
To have other published specifications demonstrated, consult with your Service Engineer to
determine whether additional diagnostic equipment will be required.
Spectra-Physics Confidential Information
Pre-Installation Considerations
2.1 Location and Environment
The location of the system and environment of your lab should have the following features:
A. A safe location that meets all applicable building codes.
B. Easy access with adequate clearance around the instrument.
C. Proper air conditioning could be critical for the performance of the laser. For certain
applications ambient room temperature changes may be an important factor for the laser
systems performance. Air ducts should not blow directly on laser or optical path.
D. In some applications vibration isolation may be required for your system. Structural integrity
of the flooring could play an important role.
E. Please note that the laser head requires a sturdy breadboard or optical table for mounting
purposes.
F. Consider room requirements for future maintenance and upgrades by your Spectra- Physics
Field Service Engineer.
Spectra-Physics Confidential Information
2.2 Physical Description
Dimensions shown are in inches over millimeters.
Figure 3-2
Millennia eV laser system outline drawing
Spectra-Physics Confidential Information
2.3
General Specifications
Table 3-1 lists the general characteristics of the Millennia eV laser. Due to our continuous
product improvement program, specifications can change without notice. For the most current
specifications and performance characteristics, visit the Newport website at
http://www.newport.com/Millennia-eV-High-Power-CW-DPSS-Lasers/1011478/1033/
info.aspx.
Complete environmental specifications for operating and non-operating conditions are listed in
Environmental Specifications.
Table 3-1
Millennia eV laser system general specifications
Feature
Specification
Operating Conditions
Warm-up time (from cold start)
< 30 min .
Temperature range
18 C to 35C
Altitude
0 to 3,000 m
Relative humidity
< 8 0%, non-condensing
Non-operating Conditions
Temperature range
-20C to 50C
Altitude
0 to 12,000 m
Relative humidity
< 90%, non-condensing
Physical Characteristics (Laser Head)
Dimensions (L x H x W)
37.47 x 10.38 x 15.24 cm (14.75 x 4.09 x 6 in.)
Weight
< 7 kg (< 16 lb.)
Electrical/Mechanical Specifications
Operating voltage
100240 VAC 10%, 50/60 Hz, single phase
Power consumption, end of life
< 300 W
Other Product Features
Constant power
Yes
RoHS
Yes
EternAlign optics
Yes
Data logging
Yes (service personnel access only)
Spectra-Physics Confidential Information
Table 3-1
Millennia eV laser system general specifications (continued)
Feature
Specification
Chiller Requirements
Water temperature range
20 0.1C
Water flow rate
0.5 gpm @ 15 psi
Thermal load
300 W
All performance specifications shown in Table 3-2 are guaranteed at specified output power only.
Due to our continuous quality improvement program, specifications are subject to change without
notice.
Table 3-2
Millennia eV laser output characteristics
Feature
Specification
Power
5 W, 10 W, or 15 W
Wavelength
532 nm
Spatial mode1
TEM00
Beam diameter at 1/e2 points2
2.3 mm 10%
Beam divergence, full angle2
< 0.5 mrad 10%
Polarization3
> 100:1 vertical
Power stability4
1%
Beam pointing stability5
< 2 rad/C
Noise6
< 0.04% rms
Boresighted
Yes
1. M2 < 1.1; beam ellipticity < 10%.
2. Measured at the exit port.
3. Vertical polarization standard; horizontal polarization available on request.
4. Measured over a 2-hour period after a 15 minute warm-up.
5. Measured as far-field x and y positions, after a 30 minute warm-up.
6. Measured over a 10 Hz to 0.1 GHz bandwidth at the specified output power.
Spectra-Physics Confidential Information
Chiller Specifications
It is critical that the chiller meet the requirements for cooling the laser head, as described in Table
3-1. Failure to do so will result in poor performance, and possibly damage the laser. Such damage
is not covered by the warranty.
Refer to the manual provided with the chiller for its dimensions, specifications, and utility
requirements.
Here are some basic requirements:
The chiller should use distilled (not de-ionized) water.
The chiller should have at least 300W thermal capacity at maximum ambient air temperature.
The chiller output flow must use a partial flow liquid filter of 10um or smaller filter size.
Minimum Software Operating Parameters
The Millennia eV system can be controlled using the Millennia eV graphical user interface
(GUI) control software that is included with the system, or it can be controlled by sending
commands one at a time using the Communication window in the GUI control program.
Installation instructions for the Millennia eV control software will be provided in the User
Manual. The GUI control software is provided on a USB memory stick along with the USB
driver.
To run the GUI control software, the control computer must meet these minimum requirements
Intel or AMD 32- or 64-bit, single or multi-core processor with > 1 GHz clock speed
1 GB RAM
100 MB available disk space for installation
Mouse or other Windows-compatible pointing device
Video display with 1366 x 768 (SVGA) or higher resolution
Available USB or RS-232 serial port
Microsoft Windows XP, Vista, or 7 operating system
A type A/B USB cable (preferred) or a standard 9-pin serial cable can be used for
communication between the host computer and the laser system.
Spectra-Physics Confidential Information
Before the arrival of your Spectra-Physics Service Engineer please review the following preinstallation requirements. When all the requirements have been met, initial the boxes and fax a
signed copy to the Spectra-Physics Service department at (408) 980-6921.
Pre-Installation Checklist
Physical Location:
A location with adequate clearance around system to conduct service and accessible by
Spectra-Physics personnel.
A temperature controlled room.
Utility services have been installed.
Local building and safety codes are in compliance and have been verified.
When Your System Arrives:
Check crates for damage.
(If damaged, file a claim with the carrier and notify Spectra-Physics.)
Uncrate and place the system on your work surface.
(Two or more people may be required to lift some equipment.)
Compare the packing list with your quotation. Call your Spectra-Physics office about any
discrepancies.
Check that all manuals were received.
Save all packing and shipping material until the installation has been completed.
Obtain the correct safety glasses and a power meter.
Customer Signature
Date
Phone Number
Fax Number
E-Mail
Sales Order Number
Spectra-Physics Confidential Information
10