100% found this document useful (1 vote)
911 views54 pages

Ziehm Dicom PDF

Ziehm NetPort implements the necessary DICOM services to download work lists from an information system, save acquired XA, RF, CR, OT, DX, RTIMAGE and CT images to a network storage device or CD-R, Floppy, MOD, memory stick, print to networked hardcopy device, query patient -, study-, seriesand image data and display moved images.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
911 views54 pages

Ziehm Dicom PDF

Ziehm NetPort implements the necessary DICOM services to download work lists from an information system, save acquired XA, RF, CR, OT, DX, RTIMAGE and CT images to a network storage device or CD-R, Floppy, MOD, memory stick, print to networked hardcopy device, query patient -, study-, seriesand image data and display moved images.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 54

DICOM Conformance Statement

DICOM Conformance Statement

Form: EW_DICOM_CONF.doc

Ziehm NetPort
Company Name:
Product Name:
Devices and Software Version:

Ziehm Imaging GmbH


Ziehm NetPort
Ziehm 8000, Ziehm 7000 Plus SW 2.05
Ziehm Vista (/Plus) SW 22.004 or 1.05/2.04
Ziehm Vision (FD, R, RFD) SW 5.25
Ziehm Vision (FD) SW 5.25
Ziehm Vario/3D SW 2.06
Ziehm Vision Vario/3D (FD) SW1.20
Ziehm Vision Vario/3D (FD) SW1.20
Ziehm Solo SW 5.25

Version:
Completion Date:

2.015 (previous versions SW17, SW18, etc.)


October 29, 2009

Document No.:
Release Date:

28415
02/11/2009

Ziehm Imaging GmbH


Donaustr. 31, D-90451 Nrnberg
Fon +49-911-2172-0

www.ziehm.com
File: conforma215.doc
Page 1 of 54

DICOM Conformance Statement


A.1 CONFORMANCE STATEMENT OVERVIEW
The product Ziehm NetPort implements the necessary DICOM services to download work
lists from an information system, save acquired XA, RF, CR, OT, DX, RTIMAGE and CT
images to a network storage device or CD-R, Floppy, MOD, memory stick, print to networked
hardcopy device, query patient -, study-, series- and image data and display moved images,
inform the information system about the work actually done.
Table A. 1-1 provides an overview of the network services supported by Ziehm Netport:
Table A. 1-1
NETWORK SERVICES
SOP Classes

User of
Provider of
Service (SCU) Service (SCP)

Transfer

Computed Radiography Image Storage


X-Ray Angiographic Image Storage
X-Ray Radiofluoroscopic Image Storage (only Vision)
Secondary Capture Image Storage
RT Image Storage ( Vision, Vario 3D)
Digital X-Ray Image Storage for Presentation (only VARIO 3D)
CT Image Storage (only VARIO 3D)
Workflow
Modality Worklist
Storage Commitment Push Model
Modality Performed Procedure Step
Query/Retrieve
Patient Root Q/R Information Model FIND and MOVE
Study Root Q/R Information Model FIND and MOVE
Print Management
Basic Grayscale Print Management
Presentation LUT

Yes
Yes
Yes
Yes
Yes
Yes
Yes

No
No
No
No
No
No
No

Yes
Yes
Yes

No
No
No

Yes
Yes

No
No

Yes
Yes

No
No

Table A. 1-2 provides an overview of the Media Storage Application Profiles supported by
Ziehm Netport:
Table A. 1-2
MEDIA SERVICES
Media Storage Application Profile
Write Files (FSC or FSU)

Form: EW_DICOM_CONF.doc

General Purpose CD-R


Magneto Optical Disk
Floppy (not Vision)
General Purpose USB
General Purpose DVD ( only Vision)

Yes
Yes
Yes
Yes
Yes

Read Files (FSR)

Yes
Yes
Yes
Yes
Yes

No (Vision)

File: conforma215.doc
Page 2 of 54

DICOM Conformance Statement

Form: EW_DICOM_CONF.doc

A.2 TABLE OF CONTENTS


A.1
CONFORMANCE STATEMENT OVERVIEW .......................................................................... 2
A.2
TABLE OF CONTENTS ............................................................................................................ 3
A.3
GENERAL ................................................................................................................................. 6
A.3.1
REVISION HISTORY......................................................................................................... 6
A.3.2
AUDIENCE ........................................................................................................................ 7
A.3.3
REMARKS ......................................................................................................................... 7
A.3.4
DEFINITIONS, TERMS AND ABBREVIATIONS ............................................................... 8
A.3.5
REFERENCES .................................................................................................................. 8
A.4
NETWORKING ......................................................................................................................... 9
A.4.1
IMPLEMENTATION MODEL ............................................................................................. 9
A.4.1.1
Application Data Flow................................................................................................. 9
A.4.1.2
Functional Definition of Ziehm NetPort AE............................................................... 10
A.4.1.2.1 Ziehm NetPort as a Storage Application Entity.................................................... 10
A.4.1.2.2 Ziehm NetPort as a Query/Retrieve Application Entity ........................................ 10
A.4.1.2.3 Ziehm NetPort as a Worklist Application Entity.................................................... 10
A.4.1.2.4 Ziehm NetPort as a Print Application Entity ......................................................... 10
A.4.1.3
Sequencing of Real-World Activities ........................................................................ 11
A.4.2
AE SPECIFICATIONS ..................................................................................................... 12
A.4.2.1
Storage Application Entity Specification................................................................... 12
A.4.2.1.1 SOP Classes ........................................................................................................ 12
A.4.2.1.2 Association Policies ............................................................................................. 12
A.4.2.1.2.1 General .......................................................................................................... 12
A.4.2.1.2.2 Number of Associations................................................................................. 12
A.4.2.1.2.3 Asynchronous Nature .................................................................................... 12
A.4.2.1.2.4 Implementation Identifying Information ......................................................... 12
A.4.2.1.3 Association Initiation by Real-World Activity ........................................................ 12
A.4.2.1.3.1 Activity Net Store, DICOM Store, NetCommit, DICOM Commit................. 12
A.4.2.1.3.1.1 Sequencing of Activities.......................................................................... 13
A.4.2.1.3.1.2 Proposed Presentation Contexts ............................................................ 14
A.4.2.1.3.1.3 SOP Specific Conformance DX and CT Image Storage on Zvario3D.... 14
A.4.2.1.3.1.4 Behavior of SCU in response to C_STORE_RSP Status....................... 14
A.4.2.1.3.1.5 SOP Specific Conformance for Storage Commitment SOP Class ......... 14
A.4.2.1.3.1.5.1 Storage Commitment Operations (N-ACTION)................................ 14
A.4.2.1.3.1.5.2 Behavior of SCU in response to N_EVENT_REPORT_RQ Status . 15
A.4.2.2
Query/Retrieve Application Entity Specification ....................................................... 15
A.4.2.2.1 SOP Classes ........................................................................................................ 15
A.4.2.2.2 Association Policies (see A.4.2.1.2)..................................................................... 15
A.4.2.2.3 Association Initiation Policy.................................................................................. 15
A.4.2.2.3.1 Activity Query/Retrieve ............................................................................... 15
A.4.2.2.3.1.1 Description and Sequencing of Activities................................................ 15
A.4.2.2.3.1.2 Proposed Presentation Contexts ............................................................ 17
A.4.2.2.3.1.3 SOP Specific Conformance for Query/Retrieve...................................... 17
A.4.2.2.4 Accepted SOP Storage Classes for Image Retrieve ........................................... 19
A.4.2.3
Workflow Application Entity Specification................................................................. 20
A.4.2.3.1 SOP Classes ........................................................................................................ 20
A.4.2.3.2 Association Policies (see A.4.2.1.2)..................................................................... 20
A.4.2.3.3 Association Initiation Policy.................................................................................. 20
A.4.2.3.3.1 Activity Worklist .......................................................................................... 20
A.4.2.3.3.1.1 Description and Sequencing of Activities................................................ 21
A.4.2.3.3.1.2 Proposed Presentation Contexts ............................................................ 22
A.4.2.3.3.1.3 SOP Specific Conformance Statement................................................... 22
A.4.2.3.3.2 Activity MPPS ............................................................................................. 24
A.4.2.3.3.2.1 Description and Sequencing of Activities................................................ 24
A.4.2.3.3.2.2 Proposed Presentation Contexts ............................................................ 25
A.4.2.3.3.2.3 SOP Specific Conformance for MPPS.................................................... 25
A.4.2.3.4 Association Acceptance Policy ............................................................................ 27
File: conforma215.doc
Page 3 of 54

Form: EW_DICOM_CONF.doc

DICOM Conformance Statement


Print Management Application Entity Specification .................................................. 27
A.4.2.4
A.4.2.4.1 SOP Classes ........................................................................................................ 27
A.4.2.4.2 Association Policies (see A.4.2.1.2)..................................................................... 27
A.4.2.4.3 Association Initiation Policy.................................................................................. 27
A.4.2.4.3.1 Activity DICOM Print................................................................................... 27
A.4.2.4.3.2 ........................................................................................................................... 28
A.4.2.4.3.2.1 Description and Sequencing of Activities................................................ 28
A.4.2.4.3.2.2 Proposed Presentation Contexts ............................................................ 29
A.4.2.4.3.2.3 SOP Specific Conformance all Print Management SOP Classes .......... 29
A.4.2.4.3.2.3.1 Printer SOP Class Operation (N-Get) .............................................. 29
A.4.2.4.3.2.3.2 Film Session SOP Class Operations ............................................... 30
A.4.2.4.3.2.3.3 Basic Film Box SOP Class Operations ............................................ 30
A.4.2.4.3.2.3.4 Basic Grayscale Image Box SOP Class Operation (N-SET) ........... 31
A.4.2.4.3.2.3.5 Presentation LUT SOP Class Operations ........................................ 31
A.4.2.4.4 Association Acceptance Policy ............................................................................ 32
A.4.3
NETWORK INTERFACES............................................................................................... 32
A.4.3.1
Physical Network Interface ....................................................................................... 32
A.4.3.2
Additional Protocols.................................................................................................. 32
A.4.4
CONFIGURATION........................................................................................................... 33
A.4.4.1
AE Title/Presentation Address Mapping................................................................... 33
A.4.4.1.1 Local AE Title(s) and Port(s) ................................................................................ 33
A.4.4.1.2 Remote AE Titles and Ports................................................................................. 33
A.4.4.1.2.1 Storage and Storage Commitment SCP AE Titles and Ports ....................... 34
A.4.4.1.2.2 Query/Retrieve SCP AE Titles and Ports ...................................................... 35
A.4.4.1.2.3 Worklist and MPPS SCP AE Titles and Ports ............................................... 35
A.4.4.1.2.4 Print SCP AE Titles and Ports ....................................................................... 36
A.4.4.2
Configuration Files.................................................................................................... 37
A.5
MEDIA INTERCHANGE.......................................................................................................... 37
A.5.1
IMPLEMENTATION MODEL ........................................................................................... 37
A.5.1.1
Application Data Flow............................................................................................... 37
A.5.1.2
Functional Definition of AEs ..................................................................................... 38
A.5.1.2.1 Functional Definition of Offline-Media Application Entity ..................................... 38
A.5.1.3
Sequencing of Real World Activities ...................................................................... 38
A.5.1.4
File Meta Information Options .................................................................................. 38
A.5.2
AE SPECIFICATION........................................................................................................ 38
A.5.2.1
Offline-Media Application Entity Specification .......................................................... 38
A.5.2.1.1 File Meta Information for the Application Entity ................................................... 38
A.5.2.1.2 Real World Activities ............................................................................................ 39
A.5.2.1.2.1 Activity Export to Medium and Import from Medium .................................. 39
A.5.2.1.2.1.1 Media Storage Application Profiles......................................................... 39
A.5.2.1.2.1.1.1 Options ............................................................................................. 39
A.5.3
AUGMENTED AND PRIVATE APPLICATION PROFILES.............................................. 39
A.5.4
MEDIA CONFIGURATION .............................................................................................. 39
A.6
SUPPORT OF CHARACTER SETS ....................................................................................... 40
A.7
SECURITY .............................................................................................................................. 40
A.8
ANNEXES ............................................................................................................................... 40
A.8.1
IOD CONTENTS .............................................................................................................. 40
A.8.1.1
Created SOP Instances............................................................................................ 40
A.8.1.1.1 X-Ray Angiographic Image IOD........................................................................... 40
A.8.1.1.2 Computed Radiography Image IOD..................................................................... 41
A.8.1.1.3 Secondary Capture Image IOD............................................................................ 41
A.8.1.1.4 Digital X-Ray Image IOD...................................................................................... 41
A.8.1.1.5 COMPUTED TOMOGRAPHY Image IOD ........................................................... 42
A.8.1.1.6 RT Image IOD ...................................................................................................... 42
A.8.1.1.7 Common Modules ................................................................................................ 43
A.8.1.1.8.................................................................................................................................. 51
A.8.1.1.8.1 Attributes of DICOMDIR ................................................................................ 51
A.8.1.1.8.2 Example README.TXT (Vision) ................................................................... 52
File: conforma215.doc
Page 4 of 54

DICOM Conformance Statement

Form: EW_DICOM_CONF.doc

A.8.1.1.9 Storage Commitment Request N-ACTION Information ....................................... 52


A.8.2
DATA DICTIONARY OF PRIVATE ATTRIBUTES........................................................... 53
A.8.3
ZIEHM NAVIPORT 2D ..................................................................................................... 53
A.8.3.1
Configuration ............................................................................................................ 53
A.8.3.2
Navigation Flag......................................................................................................... 53
A.8.3.3
Behavior.................................................................................................................... 53
A.8.3.4
Form of UDP messages ........................................................................................... 53
A.8.4
IHE INTEGRATION STATEMENT................................................................................... 54

File: conforma215.doc
Page 5 of 54

DICOM Conformance Statement


A.3 GENERAL

Form: EW_DICOM_CONF.doc

A.3.1 REVISION HISTORY


Document
Version
SW11-SW14

Date of Issue

Author

Description

before 2000

Stahl

SW 15ff

Jan-Aug 2000

Rotermund

SW 16

Okt 2000

Rotermund

SW 17ff
SW 18
SW 19
SW 20ff
SW 21ff
SW 22

2001
2001
Jul 2001
Aug 2002-Feb 2004
Jun 2004
Jan 2005

Rotermund
Rotermund
Rotermund
Rotermund
Rotermund
Rotermund

2.0

May 30, 2005

Rotermund

2.001

July 14, 2005

Rotermund

2.002

August 18, 2005

Rotermund

2.003

December 05, 2005

Rotermund

2.004

March 01, 2006

Rotermund

2.005

Januar 30, 2007

Rotermund

2.006

September 10, 2007 Rotermund


Heuss

Document
Version
2.007

Date of Issue

Author

Storage released
Query Patient-Level released
Worklist released
ISO IR 100 completed
new Instance UID
Verification released
Basic Print released
XA-Multiframe released Media Storage released
CD-Rom
Query/Retrieve all levels and model MOVE
Storage Commitment released MPPS released
common version with Vision
CD on Vario3D
new chapter Security
SW 4.12 Vision released
Conformance Statement completely reviewed
with DICOM 3.2 2004
FSR for Vision SW 4.14
Export single images Vario3D
Vision Flat SW 4.15:
Collimator RECTANGLE
Vista SW 22.002:
Worklist matching keys
Tag 0018,1500 (cine)
Tag 0018,115E in Gydm
Ziehm Media Viewer on Vision/Vario3D
IHE Integration Profiles
Scheduled Workflow
Actor: Acquisition Modality (Vision only)
Portable Data for Imaging
Actor: Portable Media Creator (Vision only)
Description Ziehm NaviPort
IHE Integration Statement
SW 5.03 Vision
SW 22.003 DOS = SW 1.01 Linux
SW 1.03 V2-0.5K
SW 22.004 only bug elimination
SW 5.05 Vision Family
- UIDs new mechanism
SW 1.04/2.01 Vista Family
Revised by technical writer
Description

April 30, 2008

Rotermund

SW 5.16 Vision Family


IHE Profile Patient Information Reconciliation
MPPS Append updated
Storage Commitment updated
RF Image Storage
SW 1.05 Vista Step1
Workaround for E93 time synchronization

File: conforma215.doc
Page 6 of 54

DICOM Conformance Statement


Document
Version
2.008
2.009
2.010

Date of Issue

Author

August 27, 2008

Heckmann
Hauner
September 30, 2008 Hauner
November 20, 2008 Rotermund
Hauner

2.011

November 21, 2008

Rotermund

2.012

February 11, 2009

Hauner
Rotermund

2.013
2.014

May 20, 2009


October 6, 2009

Rotermund
Rotermund

2.015

October 29, 2009

Rotermund

Description
Added Vision RFD
Added Ziehm Solo
DICOM by merge added
Deleted Ziehm Solo
Changed Ziehm Vision Vario/3D (FD) SW1.09
Changed Ziehm Vision Vario/3D (FD) SW1.09
Added Ziehm Solo
Vision SW 5.20
RT Image
DICOM Multiframe restricted
DICOM Print and Worklist Configuration
Changed SW 5.20 -> SW 5.22
SW 2.04 Ziehm Vista
A: A/08/189
Vision without merge-lib
SW 5.25 Ziehm Vision
More than 1 navigation server
WLAN
Ziehm 8000 (7000 Plus) without merge-lib
A: A2.05/09/145
Ziehm Vista is out of production

A.3.2 AUDIENCE
This document is intended for hospital staff, health system integrators, software designers or
implementers. It is assumed that the reader has a working understanding of DICOM.

A.3.3 REMARKS
DICOM, by itself, does not guarantee interoperability. However, the Conformance Statement
facilitates a first-level validation for interoperability between different applications supporting
the same DICOM functionality.
In the last 10 years the Ziehm Netport has gained experience in successful interconnectivity
with other vendors medical equipment.
For the benefit of readability, the product name Vision will be used for both systems Ziehm
Vision and Ziehm Vision throughout this document.

Form: EW_DICOM_CONF.doc

This product uses the OFFIS DICOM Toolkit DCMTK 1993-2009 OFFIS e.V., Oldenburg
Germany.

File: conforma215.doc
Page 7 of 54

DICOM Conformance Statement


A.3.4 DEFINITIONS, TERMS AND ABBREVIATIONS
Definitions, terms and abbreviations used in this document are defined within the different
parts of the DICOM standard.
Abbreviations and terms are as follows
AE
DICOM Application Entity
AET
Application Entity Title
CD-R
Compact Disk Recordable
DICOM
Digital Imaging and Communications in Medicine
FSC
File Set Creator
FSU
File Set Updater
FSR
File Set Reader
IHE
Integrating the Healthcare in Enterprises
IOD
(DICOM) Information Object Definition
ISO
International Standard Organization
MPPS
Modality Performed Procedure Step
R
Required Key Attribute

Optional Key Attribute


PACS
Picture Archiving and Communication System
PDU
(DICOM) Protocol Data Unit
SCU
(DICOM) Service Class User (= DICOM Client)
SCP
(DICOM) Service Class Provider (= DICOM server)
SOP
(DICOM) Service Object Pair
RIS
Radiology Information System
UID
unique identifier
VR
Value Representation

A.3.5 REFERENCES

Form: EW_DICOM_CONF.doc

DICOM Standard, NEMA PS 3.1-3.18, 2004 http://medical.nema.org/dicom


MergeCom-3 Library Linux Version 3.30 IB25, https://www.merge.com
DCMTK Version 3.5.4, http://dicom.offis.de

File: conforma215.doc
Page 8 of 54

DICOM Conformance Statement


A.4 NETWORKING
A.4.1 IMPLEMENTATION MODEL
A.4.1.1 Application Data Flow

Remote AE
receives
Images

Send Images

Update Worklist
Application
Entity
Ziehm
NetPort

Remote AE
provides
Worklist
It

Remote AE sends
Patient Data and
Images

Query Retrieve
Patient Data
and Images

Acquire
Images

Remote AE
receives MPPS
Create/Completed

Film Images

Remote AE
prints Film
Sheets

DICOM Standard Interface

Form: EW_DICOM_CONF.doc

Figure A.4.1-1
APPLICATION DATA FLOW DIAGRAM
The Ziehm NetPort Application Entity sends images to a remote AE, receives Worklist information
from and sends MPPS information to a remote AE, queries Patient Data and Images from a remote
AE and displays images, prints images on a remote AE.

File: conforma215.doc
Page 9 of 54

DICOM Conformance Statement


A.4.1.2 Functional Definition of Ziehm NetPort AE
A.4.1.2.1 Ziehm NetPort as a Storage Application Entity
The Ziehm NetPort AE is a SCU that is responsible for sending digital images from the Ziehm C-arm
to a Storage SCP. Upon successful negotiation of a Presentation Context the image transfer is
started. If the association cannot be opened, the Ziehm NetPort displays an error message. The user
can send single images, multiframe images, also single images from a multiframe (Vision creates a
new multiframe), a set of marked images, a whole patient folder or a set of patient folders (only
Vision). If the Storage SCP is configured as an archive device Ziehm NetPort AE will request Storage
Commitment and if a commitment is successfully obtained it will record this information in the local
database. If MPPS is activated the Ziehm NetPort AE sends MPPS Completed to the MPPS SCP.
Maximum size of Multiframe is ca. 600 Mbyte.

A.4.1.2.2 Ziehm NetPort as a Query/Retrieve Application Entity


The Ziehm NetPort AE is a SCU that is responsible for querying patient data, study data, series data,
image data and retrieving and displaying images from Query SCP. The Ziehm NetPort AE is able to
retrieve single images or a complete series from 2 up to 4 (Vision) other SCPs. The Association
Negotiation works as specified in A.4.1.2.1.

A.4.1.2.3 Ziehm NetPort as a Worklist Application Entity


The Ziehm NetPort AE is a SCU that is responsible for downloading a Worklist from a Worklist SCP.
The results will be displayed in a list, where the user can select an item which is automatically saved in
the local database. If MPPS is activated the Ziehm NetPort AE sends a MPPS Create to the MPPS
SCP. An existing patient can be overwritten by worklist data (Vision).

A.4.1.2.4 Ziehm NetPort as a Print Application Entity

Form: EW_DICOM_CONF.doc

The Ziehm NetPort AE is a SCU that is responsible printing digital images from the Ziehm C-arm to a
Print SCP. An Association is established with the printer and the printers status is determined. If the
printer is operating normally, the Ziehm NetPort AE sends the print job, otherwise an error message
will be displayed.

File: conforma215.doc
Page 10 of 54

DICOM Conformance Statement


A.4.1.3 Sequencing of Real-World Activities
Storage

Print

Worklist

RIS

Printer

PACS

1. Query Worklist
2. Receive Worklist

3. Select Worklist
Item (MSPS)
Create MPPS
(automatically)

4. Retrieve of Images, if desired

5. Start Procedure Step, acquire


images, select images, store
images, end o f procedure
requests Storage Commitment
(if activated) - automatically
6. Send MPPS Completed
(automatically)

7. Print Images, if desired

8. Storage Commitment, if
supported

Form: EW_DICOM_CONF.doc

Figure A.4.1-2
SEQUENCING CONSTRAINTS
Under normal scheduled workflow conditions the sequencing constraints illustrated in Figure A.4.1-2
apply:

Query Worklist
2. Receive Worklist
3. Select Worklist Item, Send MPPS Create ( automatically, if activated)

If desired Retrieve former images of selected patient from PACS


5. Start Procedure Step, Acquire images, Store images , Send N_ACTION for Commitment
6. Send MPPS Completed (automatically, if activated)
7. Print images, if desired

Ask Storage Commitment, if supported by server


File: conforma215.doc
Page 11 of 54

DICOM Conformance Statement


A.4.2 AE SPECIFICATIONS
A.4.2.1 Storage Application Entity Specification
A.4.2.1.1 SOP Classes
The Ziehm NetPort AE provides Standard Conformance to the following SOP Classes:
Table A.4.2-1
SOP CLASSES FOR AE STORAGE

SOP Class Name


Computed Radiography Image Storage
X-Ray Angiographic Image Storage
X-Ray Radiofluoroscopic Image Storage
(only Vision)
Secondary Capture Image Storage
RT Image Storage
Digital X-Ray Image Storage for
Presentation (only VARIO 3D)
CT Image Storage (only VARIO 3D)
Storage Commitment Push Model (not
VARIO 3D workstation)
Verification

SOP Class UID


1.2.840.10008.5.1.4.1.1.1
1.2.840.10008.5.1.4.1.1.12.1
1.2.840.10008.5.1.4.1.1.12.2

SCU
Yes
Yes
Yes

SCP
No
No
No

1.2.840.10008.5.1.4.1.1.1.7
1.2.840.10008.5.1.4.1.1.481.1
1.2.840.10008.5.1.4.1.1.1.1

Yes
Yes
Yes

No
No
No

1.2.840.10008.5.1.4.1.1.2
1.2.840.10008.1.20.1

Yes
Yes

No
No

1.2.840.10008.1.1

Yes

No

A.4.2.1.2 Association Policies


A.4.2.1.2.1 General
The DICOM standard application context name for DICOM 3.0 is always proposed:
Table A.4.2-2
DICOM APPLICATION CONTEXT FOR AE STORAGE
Application Context Name
1.2.840.10008.3.1.1.1

A.4.2.1.2.2 Number of Associations


Ziehm NetPort initiates one Association at a time for each destination to which a transfer request is
being processed. Ziehm NetPort initiates and accepts only one association establishment at a time.

A.4.2.1.2.3 Asynchronous Nature


Multiple outstanding transactions are not supported.

A.4.2.1.2.4 Implementation Identifying Information


Table A.4.2-3
DICOM IMPLEMENTATION CLASS AND VERSION FOR AE STORAGE
Implementation Class UID (Ziehm ROOT_UID)
2.16.840.1.113669.632.6.1
Implementation Version Name
ZiehmNetP_22.00

A.4.2.1.3 Association Initiation by Real-World Activity

Form: EW_DICOM_CONF.doc

A.4.2.1.3.1 Activity Net Store, DICOM Store, NetCommit, DICOM Commit


The user of the Ziehm 8000, Ziehm 7000 PLUS, Ziehm Vista, Ziehm Vista PLUS, Ziehm Vario3D
selects one or more images on the appliance and presses the key: Net Store. On Ziehm Vision the
user can also select a whole patient folder or a set of folders and presses the DICOM Store button.

File: conforma215.doc
Page 12 of 54

DICOM Conformance Statement


A.4.2.1.3.1.1 Sequencing of Activities

Storage
AE

Image
Manager
1. Open Association
2. C-STORE first image
3. Close Association (*)
4. Open Association
5. C-STORE next image
6. Close Association

All other images .......

7. Open Association
8. N_ACTION Storage Commitment Request

9. N_EVENT_REPORT Storage Commitment Response


10. Close Association

Form: EW_DICOM_CONF.doc

Figure A.4.2-4
SEQUENCING OF ACTIVITY
Net Store, DICOM Store, NetCommit, DICOM Commit

Ad 9. Ziehm Vision AE waits asynchronously for N_EVENT_REPORT


Ad (*) as configured 7-10. are executed after each DICOM Store

File: conforma215.doc
Page 13 of 54

DICOM Conformance Statement


A.4.2.1.3.1.2 Proposed Presentation Contexts
Each time an association is initiated, the Ziehm NetPort proposes one single Presentation Context to
be used on that association.
Table A.4.2-5
Proposed Presentation Contexts for Real-World Activity NetStore, DICOM Store, NetCommit

Abstract Syntax
Name
UID
CR Image Storage 1.2.840.10008.
5.1.4.1.1.1
XA Image Storage 1.2.840.10008.
5.1.4.1.1.12.1
RF Image Storage 1.2.840.10008.
5.1.4.1.1.12.2
Secondary Capture 1.2.840.10008.
Image Storage
5.1.4.1.1.7
RT Image Storage 1.2.840.10008.
5.1.4.1.1.481.1
DX Image Storage 1.2.840.10008.
for Presentation
5.1.4.1.1.1.1
CT Image Storage 1.2.840.10008.
5.1.4.1.1.2
Storage Commit1.2.840.10008.
ment Push Model 1.20.1

Presentation Context Table


Transfer Syntax
Name
UID
Little endian explicit
and implicit VR
Little endian explicit
and implicit VR
Little endian explicit
and implicit VR
Little endian explicit
and implicit VR
Little endian explicit
and implicit VR
Little endian explicit
VR
Little endian explicit
VR
Little endian explicit
and implicit VR

1.2.840.10008.1.2.1
1.2.840.10008.1.2
1.2.840.10008.1.2.1
1.2.840.10008.1.2
1.2.840.10008.1.2.1
1.2.840.10008.1.2
1.2.840.10008.1.2.1
1.2.840.10008.1.2
1.2.840.10008.1.2.1
1.2.840.10008.1.2
1.2.840.10008.1.2.1

SCU

Ext.
Negotiati
on
None

SCU

None

SCU

None

SCU

None

SCU

None

SCU

None

1.2.840.10008.1.2.1

SCU

None

1.2.840.10008.1.2.1 SCU
1.2.840.10008.1.2

None

Role

Ziehm Vision only supports Transfer Syntax Implicit VR. The Ziehm NetPort client does not attempt
any extended negotiation.

A.4.2.1.3.1.3 SOP Specific Conformance DX and CT Image Storage on Zvario3D


The user acquires a set of isocentric or non isocentric images (30 or more), these images are saved in
one series as DX images. On user demand a 3D image is generated and saved as 256 CT image
series. Both series can be stored to the PACS.

A.4.2.1.3.1.4 Behavior of SCU in response to C_STORE_RSP Status


Table A.4.2-6
STORAGE C-STORE RESPONSE STATUS HANDLING BEHAVIOUR

Error Code

Service Status

Behavior

0000
B000
B007

Success
Warning: Element Coercion
Warning: Data Set and SOP
Class not consistent
Warning: Elements
discarded
Failure

Image successfully transmitted and flagged up with D


Image successfully transmitted and flagged up with D
Image successfully transmitted and flagged up with D

B006

Form: EW_DICOM_CONF.doc

all other
Codes

Image successfully transmitted and flagged up with D


Aborts association with error message: Failure!!
Response from Server: <Status Code>

A.4.2.1.3.1.5 SOP Specific Conformance for Storage Commitment SOP Class


A.4.2.1.3.1.5.1 Storage Commitment Operations (N-ACTION)
The Ziehm NetPort AE will request storage commitment for instances of the XA Image Storage SOP
Class, CR Image Storage SOP Class and Secondary Capture Image Storage SOP Class if the
Remote AE is configured as an archive device and a presentation context for the Storage
Commitment Push Model has been accepted. On Vision a special Commitment SCP is asked.
File: conforma215.doc
Page 14 of 54

DICOM Conformance Statement


The Ziehm NetPort AE requests the safekeeping of a set of SOP Instances created with the Storage
Service Class.

Table A.4.2-7
STORAGE COMMITMENT N-ACTION RESPONSE STATUS HANDLING BEHAVIOUR
Error Code
Service Status Behavior
0000
Success
Continues waiting for N_EVENT_REPORT, committed images are
flagged up with C.
all other
Failure
Aborts association with error message: Failure!! Response from
Codes
Server: <St. Code>

A.4.2.1.3.1.5.2 Behavior of SCU in response to N_EVENT_REPORT_RQ Status


The Remote AE informs the Ziehm NetPort AE through the N_EVENT_REPORT Response whether
or not it has accepted Storage Commitment for the requested SOP Instances. The Ziehm NetPort AE
only waits for N_EVENT_REPORT over the existing association (not Vision). Ziehm Vision waits
asynchronously for N_EVENT_REPORT.
Table A.4.2-8
STORAGE COMMITMENT N-EVENT_REPORT RESPONSE STATUS HANDLING BEHAVIOUR
Error Code
Service Status Behavior
0000
Success
Examines Event Type ID:
1 = SUCCESS: image or image set is committed
3 = FAILURE, aborts association with Failure Reason tag
(0008,1197)
2 = UNKNOWN
in case 2,3 Ziehm NetPort shows a failure list
all other
Failure
Aborts association with error message: Failure!! Response from
Codes
server:<Failure Code>

A.4.2.2 Query/Retrieve Application Entity Specification


A.4.2.2.1 SOP Classes
The Ziehm NetPort AE provides Standard Conformance to the following SOP Classes:

SOP Class Name

Table A.4.2-9
SOP CLASSES FOR AE QUERY/RETRIEVE
SOP Class UID

Patient Root Q/R Information Model FIND


Patient Root Q/R Information Model MOVE
Study Root Q/R Information Model FIND
Study Root Q/R Information Model MOVE
Verification

1.2.840.10008.5.1.4.1.2.1.1
1.2.840.10008.5.1.4.1.2.1.2
1.2.840.10008.5.1.4.1.2.2.1
1.2.840.10008.5.1.4.1.2.2.2
1.2.840.10008.1.1

SCU

SCP

Yes
Yes
Yes
Yes
Yes

No
No
No
No
No

Form: EW_DICOM_CONF.doc

A.4.2.2.2 Association Policies (see A.4.2.1.2)


A.4.2.2.3 Association Initiation Policy
A.4.2.2.3.1 Activity Query/Retrieve
A.4.2.2.3.1.1 Description and Sequencing of Activities
The request of Patient Data, Study Data, Series Data, Image Data and the displaying of a selected
image is initiated by user interaction. The user select a patient ID and presses the Query button to get
patient data, after saving the patient, the user can step through the DICOM information model to get
File: conforma215.doc
Page 15 of 54

DICOM Conformance Statement


data of each level ( i.e. STUDY, SERIES, IMAGE). On level SERIES or IMAGE the user can move
images to be displayed by pressing the button Retrieve.

Image Manager

Query/Retrieve AE

1. Open Association
2. C-FIND Request Patient Data with selected Patient ID
3. C-FIND Response Patient Data
4. Save Patient, C-FIND Request Study Data
5. C-FIND Response Study Data
6. Select Study, C-FIND Request Series Data and
7. C-FIND Response Series Data
8. Select Series, C-FIND Request Image Data
(automatically on Ziehm 8000, Ziehm 7000 PLUS,
Ziehm Vista, Ziehm Vista PLUS)
9. C-FIND Response Image Data
10. C-MOVE Request SOP Instances
11. C-MOVE Response
12. Open Association
13. C-STORE Request
14. C-Store Response

Form: EW_DICOM_CONF.doc

15. Close Association

Figure A.4.2-5
SEQUENCING OF ACTIVITY Query/Retrieve

File: conforma215.doc
Page 16 of 54

DICOM Conformance Statement


A.4.2.2.3.1.2 Proposed Presentation Contexts
Each time an association is initiated, the Ziehm NetPort SCU proposes one single Presentation
Context to be used on that association.
Table A.4.2-10
Proposed Presentation Contexts for Application Entity Ziehm NetPort
and Real-World Activity Find and Move Request
Presentation Context Table
Abstract Syntax
Transfer Syntax
Name
UID
Name
UID
Patient Root Q/R
1.2.840.10008.5. DICOM Implicit
1.2.840.10008.1.2
Information Model 1.4.1.2.1.1
VR Little Endian
FIND
Patient Root Q/R
1.2.840.10008.5. DICOM Implicit
1.2.840.10008.1.2
Information Model 1.4.1.2.1.2
VR Little Endian
MOVE
Study Root Q/R
1.2.840.10008.5. DICOM Implicit
1.2.840.10008.1.2
Information Model 1.4.1.2.2.1
VR Little Endian
FIND
Study Root Q/R
1.2.840.10008.5. DICOM Implicit
1.2.840.10008.1.2
Information Model 1.4.1.2.2.2
VR Little Endian
MOVE

Role
SCU

Extended
Negotiation
None

SCU

None

SCU

None

SCU

None

A.4.2.2.3.1.3 SOP Specific Conformance for Query/Retrieve


The Tables A.4.2-11 A.4.2-14 below contain the C-FIND message lists of attributes which is. Sent to
the SCP. Each new level contains data from the level before. At last Table A.4.2-16 contains the
attributes of the C-MOVE Request.
Table A.4.2-11
QUERY/RETRIEVE C-FIND REQUEST Attributes Level PATIENT Root

Name

Tag

VR

Value

Description

Form: EW_DICOM_CONF.doc

Query/Retrieve Level
0008,0052 CS
PATIENT
Constant
Accession Number
0008,0050 SH
*
Constant 1)
Patients Name
0010,0010 PN
*
Constant 1)
Patient ID
0010,0020 LO
1
User supplied
Patients Birth Date
0010,0030 DA
<null>
Constant
Patients Sex
0010,0040 CS
<null>
Constant
Wildcard matching is displaced by single value matching with empty string (not Version
22.004)
Table A.4.2-12
QUERY/RETRIEVE C-FIND REQUEST Attributes Level STUDY
Tag
Name
VR Value
Description
Query/Retrieve Level
0008,0052 CS STUDY
Constant
Study Date
0008,0020 DA <null>
Constant
Study Time
0008,0030 TM <null>
Constant
Accession Number
0008,0050 SH <null>
Constant
Referring Physicians Name 0008,0090 PN <null>
Constant
Study Description
0008,1030 LO <null>
Constant
Patients Name
0010,0010 PN test^person* From PATIENT level, only STUDY ROOT
Patient ID
0010,0020 LO 1
From PATIENT level
Study Instance UID
0020,000D UI <null>
Constant
Study ID
0020,0010 SH <null>
Constant

File: conforma215.doc
Page 17 of 54

DICOM Conformance Statement


Table A.4.2-13
QUERY/RETRIEVE C-FIND REQUEST Attributes Level SERIES

Name

Tag

VR Value

Description

Query/Retrieve Level
Series Date
Series Time
Modality
Series Description
Patient ID
Study Instance UID
Series Instance UID
Series Number

0008,0052
0008,0021
0008,0031
0008,0060
0008,103E
0010,0020
0020,000D
0020,000E
0020,0011

CS
DA
TM
CS
LO
LO
UI
UI
IS

constant
constant
constant
constant
constant
Patient Root only
From Study Level
constant
constant

SERIES
<null>
<null>
<null>
<null>
1
2.16.840.1.113669.623.6.1.....
<null>
<null>

Table A.4.2-14
QUERY/RETRIEVE C-FIND REQUEST Attributes Level IMAGE

Name

Tag

VR Value

Description

Query/Retrieve Level
SOP Instance UID
Retrieve AE Title
Patient ID
Study Instance UID
Series Instance UID
Image Number
Image Comments
Storage Media File Set ID
Storage Media File Set UID

0008,0052
0008,0018
0008,0054
0010,0020
0020,000D
0020,000E
0020,0013
0020,4000
0088,0130
0088,0140

CS
UI
AE
LO
UI
UI
IS
LT
SH
UI

Constant
Constant
Constant
Only PATIENT ROOT
From level STUDY
From level SERIES
Constant
Constant
Constant
Constant

IMAGE
<null>
<null>
1
2.16.840.1.113669.632.6.1....
2.16.840.1.113669.632.6.1.....1
<null>
<null>
<null>
<null>

Table A.4.2-15
QUERY/RETRIEVE C-FIND RESPONSE STATUS HANDLING BEHAVIOUR
Error Code Service Behavior
Status
0000
Success The SCP has completed the matches. C-Find messages are available for
display. If no Pending-Status is sent, Ziehm NetPort displays an error
message: No data on level ...
FF00,FF01 Pending The SCP continues sending messages
All other
Failure
The Ziehm NetPort aborts association with error message: Failure!!
codes
Response from server:<Failure Code>

Form: EW_DICOM_CONF.doc

Table A.4.2-16
QUERY/RETRIEVE C-MOVE REQUEST Attributes Level IMAGE

Name

Tag

VR

Value

Description

Move Destination
Query/Retrieve Level
SOP Instance UID
Patient ID
Study Instance UID
Series Instance UID

0000,0600
0008,0052
0008,0018
0010,0020
0020,000D
0020,000E

AE
CS
UI
LO
UI
UI

Ziehm NetPort AE
IMAGE
2.16.840.1.113669.6.1.....17
1
2.16.840.1.113669.632.6.1....
2.16.840.1.113669.632.6.1.....1

Ziehm NetPort AE
Constant
From level IMAGE
Only PATIENT ROOT
From level STUDY
From level SERIES

File: conforma215.doc
Page 18 of 54

DICOM Conformance Statement


Table A.4.2-17
QUERY/RETRIEVE C-MOVE RESPONSE STATUS HANDLING BEHAVIOUR
Error Code Service Behavior
Status
0000
Success The Ziehm NetPort has successfully got the image from SCP. The Ziehm
NetPort downsizes larger images to 576x576 matrix (or 1024x1024 matrix for
Vision) or fills up to 576x576 (or 1024x1024) matrix, displays the image and
saves into local data base. Retrieved images are flagged up with R (only
Vision).
FF00,FF01 Pending The SCP continues sending messages
All other
Failure The Ziehm NetPort aborts association with error message: Failure!! Response
codes
from server: <Failure Code>
A801= Destination unknown,
Ziehm NetPort IP-address or hostname is not known by SCP
C001= Unable to process,
Ziehm NetPort AE-Title or port-address is not known by SCP
SW5.24/SW2.05:
Failed: MoveDestinationUnknown = AE-Title Ziehm NetPort is not configured
on server
Failed : UnableToProcess = Port and/or IP-address Ziehm NetPort is not
configured on server

A.4.2.2.4 Accepted SOP Storage Classes for Image Retrieve


The following table A.4.2-18 shows the Storage SOP Classes accepted by Ziehm NetPort,
the Table A.4.2-19 the accepted attributes of the retrieved image. Other images will not be
displayed.

Form: EW_DICOM_CONF.doc

Table A.4.2-18
Valid SCU Store SOP Classes for Ziehm NetPort

Name of SOP Class

UID

MR Image Storage
Computed Radiography Image Storage
CT Image Storage
Secondary Capture Image Storage
Ultrasound Image Storage
Ultrasound Multi-Frame Image Storage
Nuclear Medicine Image Storage
Standard X-Ray Angiographic Image Storage
Standard X-Ray Angiographic Bi-Plane Image Storage
Standard X-Ray Radiofluoroscopic Image Storage
Standard RT Image Storage
Standard Positron Emission Tomography Image Storage
Digital X-Ray Image Storage- For Presentation
Digital X-Ray Image Storage- For Processing
Digital Intra-oral X-Ray Image Storage- For Presentation
Digital Intra-oral X-Ray Image Storage- For Processing
Digital Mammography Image Storage- For Presentation
Digital Mammography Image Storage- For Processing
VL Endoscopic Image Storage
VL Microscopic Image Storage
VL Photographic Image Storage
VL Slide-Coordinates Microscopic Image Storage

1.2.840.10008.5.1.4.1.1.4
1.2.840.10008.5.1.4.1.1.1
1.2.840.10008.5.1.4.1.1.2
1.2.840.10008.5.1.4.1.1.7
1.2.840.10008.5.1.4.1.1.6.1
1.2.840.10008.5.1.4.1.1.3.1
1.2.840.10008.5.1.4.1.1.20
1.2.840.10008.5.1.4.1.1.12.1
1.2.840.10008.5.1.4.1.1.12.3
1.2.840.10008.5.1.4.1.1.12.2
1.2.840.10008.5.1.4.1.1.481.1
1.2.840.10008.5.1.4.1.1.128
1.2.840.10008.5.1.4.1.1.1.1
1.2.840.10008.5.1.4.1.1.1.1.1
1.2.840.10008.5.1.4.1.1.1.3
1.2.840.10008.5.1.4.1.1.1.3.1
1.2.840.10008.5.1.4.1.1.1.2
1.2.840.10008.5.1.4.1.1.1.2.1
1.2.840.10008.5.1.4.1.1.77.1.1
1.2.840.10008.5.1.4.1.1.77.1.2
1.2.840.10008.5.1.4.1.1.77.1.4
1.2.840.10008.5.1.4.1.1.77.1.3

File: conforma215.doc
Page 19 of 54

DICOM Conformance Statement


Table A.4.2-19
Valid Image Attributes to display image

Tag

Name

VR

Description

0008,0018
0008,0023
0008,0033
0008,0080
0008,1050
0008,1090
0018,0060
0018,1151
0018,1000
0018,1020
0020,0013
0020,4000
0028,0002
0028,0004
0028,0008
0028,0010
0028,0011
0028,0100
0028,0101
0028,0102
0028,0103
0028,0006
0028,0106
0028,0107
0028,1050
0028,1051
0028,1052
0028,1053
0028,1055

SOP Instance UID


Image Date
Image Time
Institution Name
Performing Physicians Name
Manufacturers Model Name
KVP
X-Ray Tube Current
Device Serial Number
Software Versions
Image Number
Image Comments
Samples per pixel
Photometric Interpretation
Number of Frames
Rows
Columns
Bits Allocated
Bits Stored
High Bit
Pixel Representation
Planar Configuration
Smallest Image Pixel Value
Largest Image Pixel Value
Window Center
Window Width
Rescale Intercept
Rescale Slope
Window Center&Width
Explanation
Pixel Data

UI
DA
TM
LO
PN
LO
DS
IS
LO
LO
IS
LT
US
CS
IS
US
US
US
US
US
US
US
US
US
DS
DS
DS
DS
LO

displayed in case of error


Displayed
Displayed
Displayed
Displayed
||
Displayed
Displayed
||
||
C-arm creates own number
max 64 character displayed
1,3
MONOCROME1,MONOCHROME2,RGB
max 3800
max 3800
8 or 16
all smaller Bits Allocated
Bits Stored 1
0 (unsigned integer), 1(2s complement)
0, if Samples per pixel =3.. R1G1B1, ....
||
||
only first value
only first value
||
||
||

OB/OW

only IMPLICIT LITTLE ENDIAN

7FE0,0010

A.4.2.3 Workflow Application Entity Specification


A.4.2.3.1 SOP Classes

The Ziehm NetPort AE provides Standard Conformance to the following SOP


Classes:
Table A.4.2-19
SOP CLASSES FOR AE WORKLIST and MPPS

Form: EW_DICOM_CONF.doc

SOP Class Name


Modality Worklist Information Model FIND
Modality Performed Procedure Step

SOP Class UID


1.2.840.10008.5.1.4.31
1.2.840.10008.3.1.2.3.3

SCU
Yes
Yes

SCP
No
No

A.4.2.3.2 Association Policies (see A.4.2.1.2)


A.4.2.3.3 Association Initiation Policy
A.4.2.3.3.1 Activity Worklist

File: conforma215.doc
Page 20 of 54

DICOM Conformance Statement


A.4.2.3.3.1.1 Description and Sequencing of Activities
The request for a Worklist is initiated by user interaction, i.e. pressing the button Worklist.
On the device Ziehm 8000, Ziehm 7000 PLUS, Ziehm Vista, Ziehm Vista PLUS or Ziehm Vario3D :
If the user wants to get the whole list of worklist items, he has to press the key Worklist. The Ziehm
NetPort AE request all items for a Scheduled Procedure Step Date (actual date not configurable),
Modality (XA, CR,..), Scheduled Station AE Title (Ziehm NetPort AE or all) and Location (string or
empty string), furthermore an interval of hours Scheduled Procedure Start Time and Scheduled
Procedure End Time is possible. Modality, Scheduled Station AE Title, Location and timespan are
configurable by a Service Engineer. The user can restrict the worklist items by a given patient ID and
patient family name (wildcard matching).
On the device Vision:
If the user wants to get the whole list of worklist items he has to press the worklist button and in the
next dialog the button delete and the worklist button once more. The Ziehm NetPort AE request all
items for a Scheduled Procedure Step Date, Modality (XA, CR, DX,.. or all), Scheduled Station AE
Title (Ziehm NetPort AE or all) and Location ( string or empty string). Modality, Scheduled Station AE
Title, Location are configurable by a Service Engineer. The user can restrict the worklist items with a
patient name wildcard matching is possible -, a patient ID, the accession number and the requested
procedure ID. Furthermore he can request a worklist from today, yesterday, today and yesterday, and
a fixed interval +/- 8 hours from actual time.
Upon initiation of the request, the Ziehm NetPort AE will build an Identifier for the C-FIND request, will
initiate an Association to send the request and will wait for Worklist responses. After retrieval of all
responses, Ziehm NetPort AE will display a list of all worklist items and the user can select one of
them. With pressing Return or Apply the patient is saved in the local database, i.e. a new patient
folder is created. Pressing Overwrite the current patient will be overwritten by the worklist data.
The list can be sorted by patient name, first name, ID, date and time of scheduled procedure step.
(Vision).

Worklist AE

Department
Scheduler
1. Open Association
2. C-FIND Request (Worklist Query)
3. C-FIND Response (Worklist Item) Status = Pending
4. C-FIND Response (Worklist Item) Status = Pending
5. C-FIND Response Status = Success
6. Close Association

Form: EW_DICOM_CONF.doc

7. Select Worklist Item

Figure A.4.2-5
SEQUENCING OF ACTIVITY Worklist

File: conforma215.doc
Page 21 of 54

DICOM Conformance Statement


A.4.2.3.3.1.2 Proposed Presentation Contexts
The Ziehm NetPort SCU proposes one single Presentation Context to be used on that association.
Table A.4.2-20
Proposed Presentation Contexts for Application Entity Ziehm NetPort and Real-World Activity
Find Request
Presentation Context Table
Transfer Syntax
Name
UID
Name
UID
Worklist Information 1.2.840.10008.5.1.4.31 DICOM Implicit
1.2.840.10008.1.2
Model FIND
VR Little Endian
Abstract Syntax

Role
SCU

Ext.
Neg.
None

A.4.2.3.3.1.3 SOP Specific Conformance Statement


The behavior of Ziehm NetPort when encountering status codes in a Modality Worklist C-FIND
response is summarized in the Table below.
Table A.4.2-21
MODALITY WORKLIST C-FIND RESPONSE STATUS HANDLING BEHAVIOUR
Error Code Service Status
Behavior
0000
Success
provides valid Worklist Data, if exist
else display message: No data
FF00,FF01 Pending
Ziehm NetPort is waiting for next item
A700
Refused-Out of Resources
provides valid Worklist Data, if exist and
DICOM-Error A700 then aborts association
A900
Error: Data Set and SOP Class not
DICOM-Error A900
consistent
C001
Error: unable to process , data from SCP as A700 with DICOM Error C001
not DICOM conform
FE00
user cancels
as A700

Form: EW_DICOM_CONF.doc

The table below provides a description of the Ziehm NetPort Worklist Request Identifier and specifies
the attributes that are copied into the images. Unexpected attributed returned in a C-FIND response
are ignored.

File: conforma215.doc
Page 22 of 54

DICOM Conformance Statement

Table A.4.2-22
WORKLIST REQUEST IDENTIFIER
Tag
VR M

Form: EW_DICOM_CONF.doc

Module Name
Attribute Name
SOP Common
Specific Character Set
Scheduled Procedure Step
Scheduled Procedure Step Sequence
Scheduled Station AET
Scheduled Procedure Step Start Date
Scheduled Procedure Step Start Time
Modality
Scheduled Performing Physicians Name
Scheduled Procedure Step Description
Scheduled Station Name
Scheduled Procedure Step Location
Scheduled Protocol Code Sequence
(Scheduled Action Item Code Sequence)
>> Code Value
>> Code Scheme Designator
>> Code Meaning
Scheduled Procedure Step ID
Requested Procedure
Requested Procedure ID
Requested Procedure Description
Requested Procedure Code Sequence
>> Code Value
>> Code Scheme Designator
>> Code Meaning
Study Instance UID
Referenced Study Sequence
>> Referenced SOP Class UID
>> Referenced SOP Instance UID
Requested Procedure Priority
Patient Transport Arrangements
Imaging Service Request
Accession Number
Referring Physicians Name
Patient Identification
Patients Name
Patient ID
Patient Demographic
Patients Birth Date
Patients Sex

IOD Note
ISO IR
100

(0008,0005) CS
(0040,0100)
(0040,0001)
(0040,0002)
(0040,0003)
(0008,0060)
(0040,0006)
(0040,0007)
(0040,0010)
(0040,0011)
(0040,0008)

SQ
AE
DA
TM
CS
PN
LO
SH
SH
SQ

(0008,0100)
(0008,0102)
(0008,0103)
(0040,0009)

SH
SH
LO
SH

X
X
X
X

X
X

X
X
X
X

(0040,1001)
(0032,1060)
(0032,1064)
(0008,0100)
(0008,0102)
(0008,0103
(0020,000D)
(0008,1110)
(0008,1150)
(0008,1155)
(0040,1003)
(0040,1004)

SH S
LO 3)
SQ
SH
SH
SH
UI
SQ
UI
UI
SH
LO

X
X

X
X

X
X

X
X
X
X

X
X
X
X

(0008,0005) SH S
(0008,0090) PN 3)

X
X

X
X
X 3) X

(0010,0010) PN *
(0010,0020) LO S

X
X

X
X

X
X

(0010,0030) DA
(0010,0040) CS

X
X

X
X

X
X

S
RM
RM
S

X
X
X
X
X
X
X
X
X
X

X 2)
X
X
X
X
X
X
X

1)
6)
5)
4)

2 Items
(Vision)

3)
X
X
X
X

X
X
X
X
3)

File: conforma215.doc
Page 23 of 54

DICOM Conformance Statement


The above table should be read as follows:
Module Name
The name of the associated module for supported worklist attributes
Attribute Name
Attributes supported by Ziehm NetPort Worklist Request Identifier
Tag
DICOM Tag for this attribute
VR
DICOM Value Representation for this Attribute
M
Matching Keys, i.e. S= Single Value Matching, RM= Range Matching,
*= Wildcard Matching
R
Return Keys with zero length for Universal Matching
D
Displayed Keys
WL
Worklist
IOD
Tag will be saved in local data base for MPPS and DICOM Storage
1)
Single Value Matching Ziehm NetPort AE or Universal Matching
2)
Not Vision
3)
Only Vision
4)
Single Value Matching for XA, CR or Universal Matching (additional
on Ziehm Vision: DX, ES, RF, RG, PX, CT)
5)
If TIME range expires midnight, a second request to SCP is started
6)
Today + Yesterday is possible

A.4.2.3.3.2 Activity MPPS


A.4.2.3.3.2.1 Description and Sequencing of Activities
After selecting a worklist item, the patient is saved in the local database and a new patient folder is
created. If MPPS is activated the Ziehm NetPort establishes automatically an Association to the
configured MPPS SCP, sends an N_CREATE to set the Procedure Step into the State IN
PROGRESS.
Ziehm NetPort supports the creation of unscheduled cases on all devices.
Ziehm NetPort supports 0-1, 1-1 relationship between Scheduled and Performed Procedure Steps, on
device Ziehm Vision the 1-N relationship is supported (Append case).
With a successful Storage of one or more images to the PACS Ziehm NetPort establishes a second
Association to the MPPS SCP, sends an N_SET to set the Procedure Step into the state
COMPLETED. On Ziehm Vision it is also possible to set state DISCONTINUED.
Department
Scheduler

Ziehm NetPort AE

1. Open Association
2. N-CREATE (MPPS) IN PROGRESS
3. Close Association

4. DICOM Storage

5. Open Association

Form: EW_DICOM_CONF.doc

6. N-SET (MPPS) - COMPLETED


7. Close Association

Figure A.4.2-6
SEQUENCING OF ACTIVITY MPPS AUTOMATICALLY!
File: conforma215.doc
Page 24 of 54

DICOM Conformance Statement

A.4.2.3.3.2.2 Proposed Presentation Contexts


Table A.4.2-23
Proposed Presentation Contexts for Application Entity Ziehm NetPort and Real-World Activity
MPPS
Presentation Context Table
Abstract Syntax
Transfer Syntax
Role Extended
Name
UID
Name
UID
Negotiation
Modality Performed 1.2.840.10008.3.1.2. DICOM Implicit 1.2.840.10008.1.2 SCU None
Procedure Step
3.3
VR Little Endian

A.4.2.3.3.2.3 SOP Specific Conformance for MPPS


The behavior of Ziehm NetPort when encountering status code in an MPPS N-CREATE or N-SET
response is summarized in Table A.4.2-24.

Table A.4.2-24
MPPS N-CREATE/ N-SET RESPONSE STATUS HANDLING BEHAVIOUR
Error Code
Service Status
Behavior
0000
Success
Returns tag 0000,1000 in data-File
all other Codes
Failure
Aborts association with error message: Failure!!
Response from server:<Error Code>

Form: EW_DICOM_CONF.doc

Table A.4.2-25 provides a description of the MPPS M-CREATE and N-SET request identifiers sent by
Ziehm NetPort.

Table A.4.2-25
MPPS N-CREATE/ N-SET REQUEST IDENTIFIER
Attribute Name
Tag
VR N-CREATE
N-SET
Specific Character Set
(0008,0005) CS ISO IR 100
Modalit
(0008,0060) CS XA, CR
Procedure Code Sequence
(0008,1032) SQ From WL
>>Code Value
(0008,0100) SH
>>Coding Scheme Designator
(0008,0102) SH
>>Code Meaning
(0008,0104) LO
Referenced Patient Sequence
(0008,1120) SQ Zero Length
Patients Name
(0010,0010) PN From WL or input:
Family^Given
Patient ID
(0010,0020) LO from WL or user
input
Patients Birth Date
(0010,0030) DA from WL or user
input
Patients Sex
(0010,0040) CS from WL or user
input
Image Area Dose Product
(0018,115E) DS Zero Length
1) in dGy*cm*cm
Distance Source to Detector(SID) (0018,1110) DS Zero Length
970/970/1110 mm 3)
Study ID
(0020,0010) SH =Tag (0040,1001)
Performed Station AE Title
(0040,0241) AE Ziehm NetPort AE
Title
Performed Station Name
(0040,0242) SH Ziehm NetPort AE
Title
Performed Location
(0040,0243) SH from configuration
Performed Proc. Step Start Date (0040,0244) DA CreatDate of pat.
Folder
File: conforma215.doc
Page 25 of 54

Form: EW_DICOM_CONF.doc

DICOM Conformance Statement


Attribute Name
Performed Proc. Step Start Time

Tag
VR
(0040,0245) TM

Performed Proc. Step End Date


Performed Proc. Step End Time
Performed Procedure Step
Status
Performed Procedure Step ID

(0040,0250) DA
(0040,0251) TM
(0040,0252) CS

Performed Procedure Step Desc.


Performed Proc. Type Descr.
Performed Protocol Code Seq.
>>Code Value
>>Coding Scheme Designator
>>Code Meaning
Scheduled Step Attribute Seq.
> Referenced Study Sequence
>>Referenced SOP Class UID
>>Referenced SOP Instance UID
> Study Instance UID

(0040,0254)
(0040,0255)
(0040,0260)
(0008,0100)
(0008,0102)
(0008,0104)
(0040,0270)
(0008,1110)
(0008,1150)
(0008,1155)
(0020,000D)

LO
LO
SQ
SH
SH
LO
SQ
SQ
UI
UI
UI

> Accession Number


> Requested Procedure Descr.
> Scheduled Proc. Step Descr.
> Scheduled Protocol Code Seq.
>>Code Value
>>Coding Scheme designator
>>Code Meaning
> Scheduled Procedure Step ID
> Requested Procedure ID
Performed Series Sequence
> Retrieve AE Title

(0008,0050)
(0032,1060)
(0040,0007)
(0040,0008)
(0008,0100)
(0008,0102)
(0008,0104)
(0040,0009)
(0040,1001)
(0040,0340)
(0008,0054)

SH
LO
LO
SQ
SH
SH
LO
SH
SH
SQ
AE

> Series Description


> Performing Physicians Name
> Operators Name
> Referenced Image Sequence
>> Referenced SOP Class UID

(0008,103E)
(0080,1050)
(0080,1070)
(0008,1140)
(0008,1150)

LO
PN
PN
SQ
UI

>> Referenced SOP Instance


UID
> Protocol Name

(0008,1155) UI

> Series Instance UID


> Series Description
> Referenced Standalone SOP
Instance Sequence
Total Time of Fluoroscopy
Total Number of Exposures
Distance Source to Entrance
Exposure Dose Sequence
> KVP
> Radiation Mode

(0020,000E) UI
(0008,103E) LO
(0040,0220) SQ
(0040,0300)
(0040,0301)
(0040,0306)
(0040,030E)
(0018,0060)
(0018,115A)

> Filter Type

(0018,1160) SH

(0040,0253) SH

N-CREATE
CreatTime of pat.
Folder
Zero Length
Zero Length
IN PROGRESS

Actual end date


Actual end time
COMPLETED
DISCONTINUED (Vision)

Input or created
by device
Zero Length
Zero Length
from (0040,0008)

2)
From WL

From WL or
created by Ziehm
NetPort
From WL or input
From WL
From WL
From WL

From WL
From WL
Zero Length
AE Title of Image
Manager
(0032,1060) from Worklist
User input
Configuration
One or max 10 items
SOP Class from Storage
Configuration
SOP Instance of image

(0018,1030) LO

US
US
DS
SQ
DS
CS

N-SET

Zero Length
(not supported)
Zero length
Zero length
Zero length
Zero length
-

Code Meaning from


(0040,0260)
Created by Ziehm NetPort
User input
Zero Length
(not supported)
Total Time in sec
Number Exposures (4
SID 200 mm
From Image Acquisition
CONTINUOUES/PULSED
/PULSED (3
FLAT

File: conforma215.doc
Page 26 of 54

DICOM Conformance Statement


Attribute Name
> Filter Material
> X-Ray Tube Current in uA

Tag
VR
(0018,7050) CS
(0018,8151) DS

N-CREATE
-

N-SET
ALUMINIUM/COPPER
From Image Acquisition

*1)WARNING for Ziehm 8000, Ziehm 7000 PLUS, Ziehm Vista, Ziehm Vista PLUS, Ziehm Vario 3D:
to get consistent value, save image after last x-ray
*2) Ziehm Vision supports IHE requirements (Manual Modality Setting)
*3) Ziehm Vision / Ziehm Vision FD / Ziehm Vista
*4) only Ziehm Vista

A.4.2.3.4 Association Acceptance Policy


Since the Ziehm NetPort AE is a Service Class User only, it will never accept associations.

A.4.2.4 Print Management Application Entity Specification


A.4.2.4.1 SOP Classes
Ziehm NetPort provides Standard Conformance to the following SOP Classes:
Table A.4.2-26
SOP CLASSES FOR AE PRINT MANAGEMENT
SOP Class Name
SOP Class UID
Basic Grayscale Print Management Meta
1.2.840.10008.5.1.1.9
Presentation LUT (only Vision)
1.2.840.10008.5.1.1.23

SCU
Yes
Yes

SCP
No
No

A.4.2.4.2 Association Policies (see A.4.2.1.2)


A.4.2.4.3 Association Initiation Policy

Form: EW_DICOM_CONF.doc

A.4.2.4.3.1 Activity DICOM Print

File: conforma215.doc
Page 27 of 54

DICOM Conformance Statement


A.4.2.4.3.2
A.4.2.4.3.2.1 Description and Sequencing of Activities
A user composes images onto film sheets by marking them on the appliance. On Ziehm 8000, Ziehm
7000 PLUS, Ziehm Vista, Ziehm Vista PLUS or Ziehm Vario 3D the user can select the number of
copies, the Film Destination, the Size ID, Number of Images per Page and can insert a Film Session
Label. Otherwise these data are configurable by a service engineer.

Ziehm NetPort AE

Printer
1.Open Association
2. N-GET Request PRINTER
3.N-GET Response PRINTER Status
4. Close Association

If PRINTER Status = NORMAL

5. Open Association
6. N-GET (Printer)
7. N-CREATE ( Film Session)
8. N-CREATE (Presentation LUT)
9. N-CREATE ( Film Box)
10. N-SET ( Image Box)
11. N-ACTION (Film Box)

12. Print Film sheets

13. N-DELETE (Film Session)


14. Close Association

Form: EW_DICOM_CONF.doc

Figure A.4.2-7
SEQUENCING OF ACTIVITY DICOM PRINT

The Ziehm NetPort AE opens one association at a time, first it sends a Printer N_GET Request to get
the Printer Status. If the Status is NORMAL, it sends a Basic Film Session N_CREATE message to
the SCP, followed by a Basic Film Box, N_CREATE message. Then it sends a Basic Grayscale Image
Box, N_SET message. At the end an N_ACTION message is sent to instruct the Print SCP to print at
the Basic Film Session (1 page) or at the Basic Film Box level (more then 1 page).
If Printer supports Presentation LUT, Ziehm NetPort sends Presentation LUT N_CREATE.

File: conforma215.doc
Page 28 of 54

DICOM Conformance Statement


A.4.2.4.3.2.2 Proposed Presentation Contexts
The Ziehm NetPort provides Standard Conformance to the DICOM Basic Grayscale Print
Management Meta SOP class as a SCU. The proposed context is only the bold printed line in the
table.
Table A.4.2- 27
PROPOSED PRESENTATION CONTEXTS FOR ACTIVITY DICOM PRINT
Presentation Context Table
Role Ext.
Abstract Syntax
Transfer Syntax
Neg.
NAME
UID
NAME
UID
1.2.840.10008.5.1.1.9 DICOM Implicit
1.2.840.10008.1.2 SCU None
Basic Grayscale
VR Little Endian
Print Management

Meta
Basic Film
Session
Basic Film Box
Basic Grayscale
Image Box
Printer
Presentation LUT
(only Vision)

1.2.840.10008.5.1.1.1

DICOM Implicit
VR Little Endian
1.2.840.10008.5.1.1.2 DICOM Implicit
VR Little Endian
1.2.840.10008.5.1.1.4 DICOM Implicit
VR Little Endian
1.2.840.10008.5.1.1.16 DICOM Implicit
VR Little Endian
1.2.840.10008.5.1.1.23 DICOM Implicit
VR Little Endian

1.2.840.10008.1.2 SCU None


1.2.840.10008.1.2 SCU None
1.2.840.10008.1.2 SCU None
1.2.840.10008.1.2 SCU None
1.2.840.10008.1.2 SCU None

A.4.2.4.3.2.3 SOP Specific Conformance all Print Management SOP Classes


A.4.2.4.3.2.3.1 Printer SOP Class Operation (N-Get)
Ziehm NetPort uses Printer SOP Class N-GET operation to obtain information about the current printer
status. The attributes obtained via N-GET are listed in the Table below:
Table A.4.2-28
PRINTER SOP CLASS N-GET REQUEST ATTRIBUTES
Attribute Name
Tag
VR Value
Printer Status
Printer Status Info

(2110,0010)
(2110,0020)

CS
CS

Provided by Printer
Provided by Printer

Presence
of Value
ALWAYS
ALWAYS

Source
Printer
Printer

Form: EW_DICOM_CONF.doc

The Printer Status information is evaluated as follows:


1. If Printer Status (2120,0010) is NORMAL, the print-job continues to be printed.
2. If not, the Printer Status Info is reported to the user and the print-job is stopped, see Table below:

Table A.4.2-29
PRINTER SOP CLASS N-GET RESPONSE STATUS HANDLING BEHAVIOUR
Error Code
Service Status
Behavior
0000
Success
continues creating Film Session
All other Codes and Printer
Failure
aborts association with error message: Failure!!
Status is not NORMAL
Response from Printer: <Status Code>
Vision aborts only in case of FAILURE

File: conforma215.doc
Page 29 of 54

DICOM Conformance Statement


A.4.2.4.3.2.3.2 Film Session SOP Class Operations
The Ziehm NetPort supports the following DIMSE operations for the Film Session SOP Class:

N-CREATE
N-ACTION
N-DELETE

The attributes supplied in an N_CREATE Request are listed in the Table below:

Table A.4.2-30
FILM SESSION SOP CLASS N-CREATE REQUEST ATTRIBUTES
Attribute Name
Tag
VR
Value
Number of Copies
(2000,0010)
IS
1..10
Print Priority
(2000,0020)
CS
user defined : HIGH, MED, LOW or default of SCP
Medium Type
(2000,0030)
CS
user defined: PAPER, BLUE FILM, CLEAR FILM or
default of SCP
Film Destination
(2000,0040)
CS
user defined: PROCESSOR, MAGAZINE, BIN_i
(i=1..10) or default of SCP
Film Session Label
(2000,0050)
LO
user defined
Table A.4.2-31
FILM SESSION SOP CLASS N-CREATE, N-ACTION, N-DELETE RESPONSE
STATUS HANDLING BEHAVIOUR
Error Code
Service Status
Behavior
0000
Success
continues creating Film Session
All other Codes
Failure
aborts association with error message: Failure!!
Response from Film Session ..: <Error Code>

A.4.2.4.3.2.3.3 Basic Film Box SOP Class Operations


The Ziehm NetPort supports the following DIMSE operations for the Basic Film Box SOP Class:

N-CREATE
N-ACTION

The attributes supplied in an N-CREATE Request are listed in the Table below:

Form: EW_DICOM_CONF.doc

Table A.4.2-32
FILM BOX SOP CLASS N-CREATE REQUEST ATTRIBUTES
Attribute Name
Image Display Format
Referenced Film Session Sequence
Ref SOP Class UID
Ref SOP Instance UID

Tag
(2010,0010)
(2010,0500)
(0008,1150)
(0008,1155)

VR
CS
SQ
UI
UI

Film Orientation
Film Size ID

(2010,0040) CS
(2010,0050) CS

Magnification Type (only Vision)


Min Density
Max Density
Referenced Presentation LUT Sequence
> SOP Class UID

(2010,0060)
(2010,0120)
(2010,0130)
(2050,0500)
(0008,1150)

CS
US
US
SQ
UI

Value
STANDARD\C,R C=1..4 R=1..6
1.2.840.10008.5.1.1.1
returns from SCP when FILM Session
is created
PORTRAIT/LANDSCAPE
user defined: 8INX10IN,
8_5INX11IN, 10INX12IN, 10INX14IN,
11INX14IN, 11INX17IN, 14INX14IN,
14INX17IN,24CMX24CM,
24CMX30CM, A4, A3 or default of
SCP
REPLICATE/BILINEAR/CUBIC/NONE
0..65535
0..65535
If Presentation LUT is created
1.2.840.10008.5.1.1.23

File: conforma215.doc
Page 30 of 54

DICOM Conformance Statement


Attribute Name
> SOP Instance UID

Tag
VR
(0008,1155) UI

Illumination

(2010,015E) US

Reflected Ambient Light

(2010,0160) US

Value
From SCP when Presentation LUT is
created
150 cd/m for Medium Type=PAPER
2000 cd/m others
10 cd/m for Media Type Film

Table A.4.2-33
FILM BOX SOP CLASS N-CREATE, N-ACTION RESPONSE STATUS HANDLING
BEHAVIOUR
Error Code
Service Status
Behavior
0000
Success
Continues creating Film Session
All other Codes
Failure
Aborts association with error message: Failure!!
Response from Film Box ..: <Error Code>

A.4.2.4.3.2.3.4 Basic Grayscale Image Box SOP Class Operation (N-SET)


The attributes supplied in an N-SET Request are listed in the Table below:
Table A.4.2-34
BASIC GRAYSCALE IMAGE BOX SOP CLASS N-SET REQUEST ATTRIBUTES
Attribute Name
Tag
VR
Value
Image Position
(2020,0010)
US
1..max24
Basic Grayscale Image Sequence (2020,0110)
SQ
> Samples Per Pixel
(0028,0002)
US
|1|
> Photometric Interpretation
(0028,0004)
US
|MONOCHROME2|
> Rows
(0028,0010)
US
|576| or |1024|
> Columns
(0028,0011)
US
|576| or |1024|
> Bits Allocated
(0028,0100)
US
|8|
> Bits Stored
(0028,0101)
US
|8|
> High Bit
(0028,0102)
US
|7|
> Pixel Representation
(0028,0103)
US
|0|
> Pixel Data
(7FE0,0010)
OW
Table A.4.2-35
BASIC GRAYSCALE IMAGE BOX SOP CLASS N-SET RESPONSE STATUS HANDLING
BEHAVIOUR
Error Code
Service Status
Behavior
0000
Success
continues creating Film Session
All other Codes
Failure
Aborts association with error message: Failure!!
Response from Image Box ..: <Error Code>

A.4.2.4.3.2.3.5 Presentation LUT SOP Class Operations

Form: EW_DICOM_CONF.doc

The Ziehm NetPort supports the following DIMSE operations for the Presentation LUT Session SOP
Class:

- N-CREATE
If the server supports Presentation LUT SOP Class, the Affected SOP Instance UID is
expected in the response message.
The attributes supplied in an N_CREATE Request are listed in the Table below:
Table A.4.2-36
PRESENTATION LUT SOP CLASS N-CREATE REQUEST ATTRIBUTES
Attribute Name
Tag
VR
Value
Presentation LUT Shape (2050,0020)
CS
IDENTITY
File: conforma215.doc
Page 31 of 54

DICOM Conformance Statement


Table A.4.2-37
PESENTATION LUT N-CREATE RESPONSE
STATUS HANDLING BEHAVIOUR
Error Code
Service Status
Behavior
0000
Success
continues
All other Codes
Failure
aborts association <Error Code>

A.4.2.4.4 Association Acceptance Policy


The Ziehm NetPort AE does not accept Associations.

A.4.3 NETWORK INTERFACES


A.4.3.1 Physical Network Interface
Ziehm NetPort supports a single network interface. One of the following physical network interfaces
will be available depending on installed hardware options:
Table A.4.3-1
SUPPORTED PHYSICAL NETWORK INTERFACES
Ethernet 10/100base T
Fiber Optic connection
Wireless LAN WPA2 with AES

A.4.3.2 Additional Protocols

Form: EW_DICOM_CONF.doc

Ziehm NetPort does not support DHCP, LDAP.


Only on Ziehm Vario3D DNS is supported for address resolution of server hostname.
Table A.4.3-2
NETWORK PARAMETERS
Parameter
Value
Ziehm NetPort (or Vario) IP Address
Input by service engineer
Vision Vario: IP Address Vision +1 (automatically)
Please avoid leading zeros on the Vision unit!
Ziehm NetPort Hostname
Vision: configurable
Other devices: set to AE-Title Ziehm NetPort!
DNS Server
Ziehm Vario 3D Workstation
Others: no
Gateway
Input by service engineer
Device Vision/V2-0.5K: if no Gateway take IP address of
Ziehm NetPort
Please avoid leading zeros on the Vision unit!
Subnet mask
Input by service engineer
Please avoid leading zeros!
IP Address SCP
Ziehm Vario3D Workstation: no
Others: Input by service engineer
Please avoid leading zeros on the Vision unit!
Hostname SCP
Not configurable
Domain name
Not configurable
Broadcast address
Not configurable

File: conforma215.doc
Page 32 of 54

DICOM Conformance Statement


A.4.4 CONFIGURATION
A.4.4.1 AE Title/Presentation Address Mapping
A.4.4.1.1 Local AE Title(s) and Port(s)
The service engineer can configure the Ziehm NetPort AE Title(s) and the TCP/IP Port(s) in the
service dialog. The dialog is protected with a password.

Table A.4.4-1
LOCAL AE-Title and Port Ziehm NetPort
Device
AE-Title
Port
Ziehm Vision
Configurable, same for
Configurable, please use value
every Application
> 1024
(including Vision Vario)
Ziehm Vario 3D Workstation
Configurable
Configurable
All other
Configurable, for each
104, not configurable
Application
1025 V2-0,5K
Please use always the
same, because it is set to
Hostname!

A.4.4.1.2 Remote AE Titles and Ports

Form: EW_DICOM_CONF.doc

The service engineer can configure the SCP AE Title(s) and the TCP/IP Port(s) in the service dialog.
The dialog is protected with a password. On the device Ziehm Vision the service engineer has the
possibility to CLONE a server: Press button Clone, Save, leave the dialog DICOM , return to the
dialog DICOM, you will find a new section net_xxx_server_CLONE. You cannot delete cloned
servers. On all other devices the number of servers is hard coded.
With the button Verify or Test (Ziehm Vision) first the network connection is tested with ping.
If the ping-command fails, the error message server not reachable or ping failed is displayed for 2
seconds, then Ziehm NetPort starts the Association Negotiation. Table A.4.4-2 lists the list of possible
error-messages:
Table A.4.4-2
ERROR MESSAGES DICOM VERIFICATION
Error Message
Reason(s)
No connection to server
1. Server is not switched on
Ping failed (Ziehm Vision)
2. Or network parameter not correct Table 4.3-2
SW5.03: Echo failed for security reason
Please check!
SW1.03: ... (Ping failed)
Failed to connect to remote host
1. IP address of SCP or Port of SCP is wrong
2. Or Subnet Mask is wrong
Until SW15: No connection to server
With SW5.24: Failed to establish association
Please check!
Unable to resolve host name to IP address
1. Please write hostname of SCP to IP address of
SCP
Until SW15: No connection to server
With SW5.22: - no longer Association negotiation failed
1. AE Title or Port of SCP not correct
2. AE Title Ziehm NetPort is not correct
Until SW15: System Error 23
With SW5.24: Association rejected:
Please check!
Wrong server (client) aetitle
Connection aborted before association
1. Application SCP is not started
negotiation was completed
2. Port SCP not correct
Please check!
Until SW15: No response from server
File: conforma215.doc
Page 33 of 54

DICOM Conformance Statement


Error Message
With SW5.24: Failed to establish association
Required Configuration Info Missing
Message service unacceptable for this
association
Until SW15: No response from server
With SW5.24: - no longer Network unexpectly shutdown

Reason(s)
1. wrong mergecom.app -> please order Software
Update

1. Association Negotiation is ok! SCP sends no


ECHO-Response
MERGE.INI cannot be loaded
1. DOS-Version: Wrong autoexec.bat
2. LINUX-Version: Software Bug
Until SW15: System Error 23
With SW5.24: - no longer Please order Software Update
PC/TCP resident module is not loaded; aborting Only in DOS-Versions:
program
1. Driver of network card is not loaded or wrong
autoexec.bat or pctcp.ini
Until SW15: No connection to server
- no longer in linux versions Please call service hotline

Errors are logged to the file DICOM.log/merge.log, retrieval from the device to floppy/usb disc is
possible in the Service dialog.

A.4.4.1.2.1 Storage and Storage Commitment SCP AE Titles and Ports

Form: EW_DICOM_CONF.doc

Table A.4.4-2
REMOTE AE-Title and Port STORAGE/Commitment SCP
Device
AE-Title
Port
Number of Name of servers 1)
servers
Ziehm Vision
Configurable
Configurable
5
Configurable in field Storage Server
= 1 active Storage Server
= 2 active Navigation Server
= 3 Commitment Server
= 0 inactive Storage Server
= 4 inactive Navigation Server
versions 4.xx: net_storage_server = 1
Ziehm Vario3D Configurable
Configurable
1
Workstation
All other
Configurable
Configurable
2
Storage Server
Print Server with Function Storage2
Table A.4.4-3
REMOTE AE-Title and Port NAVIGATION
Device
AE-Title
Port
Name of Storage SCP
Ziehm Vision
Configurable
Configurable
<Storage Server> = 2
see A.8.3 NaviPort 2D:
XnavistudyX = 2
XnaviimageX = 2
X free text
Ziehm Vario3D Workstation Configurable
Configurable
All other
Configurable
Configurable
Print Server with Function
Navigation
In case of navigation the Ziehm NetPort sends 2 UDP-messages to configured navigation server.

File: conforma215.doc
Page 34 of 54

DICOM Conformance Statement

Device

Table A.4.4-4
Attributes Storage SCP
Attributes
Tag
Name

Ziehm Vario3D Workstation Modality


All other
Ziehm Vision
Image Type

Ziehm Vario3D Workstation


All other
Ziehm Vision
Ziehm Vision

Bits Allocated
Image Size

Value

(0008,0060) DX,CT ( hard coded)


XA or CR or RF or SC=RG=OT or RT
(0008,0008) ORIGINAL = Image attributes off
DERIVED = Image attributes on
With Annotations
Or with Text overlay
ORIGINAL (hard coded)
ORIGINAL = all image attributes Off
DERIVED = one of the image attributes
(0028,0100) 16 or 8
(0028,0010) 1024 or 512
(0028,0011)

A.4.4.1.2.2 Query/Retrieve SCP AE Titles and Ports


The service engineer can select between STUDY_ROOT and PATIENT_ROOT.

Table A.4.4-5
REMOTE AE-Title and Port QUERY/RETRIEVE SCP
Device
AE-Title
Port
Number of
Name of servers 2)
servers
Ziehm Vision
Configurable
Configurable
1 for Query
Configurable in field Query Server
= 1 Query Server
= 0 Retrieve Servers
4 for Retrieve versions 4.xx: net_query_server = 1
Ziehm Vario3D Not supported Not supported Workstation
All other
Configurable
Configurable
1
Query Server
2
Retrieve Server 1
Retrieve Server 2

Form: EW_DICOM_CONF.doc

A.4.4.1.2.3 Worklist and MPPS SCP AE Titles and Ports


Table A.4.4-6
REMOTE AE-Title and Port WORKLIST AND MPPS SCP
Device
AE-Title
Port
Number of Name of servers 3)
servers
Ziehm Vision
Configurable
Configurable
1 for
Configurable in field Worklist Server
Worklist
= 1 Worklist Server
= 2 MPPS Server
1 for MPPS versions 4.xx: net_worklist_server = 1
net_mpps_server = 1
Ziehm Vario3D Not supported Not supported Workstation
All other
Configurable
Configurable
1
Worklist Server
1
MPPS Server

File: conforma215.doc
Page 35 of 54

DICOM Conformance Statement

Device
all
all

all
Not Vision

Table A.4.4-7
MATCHING KEYS FOR WORKLIST
Attributes Name
Tag
Value
Modality
(0008,0060)
XA or CR or all or DX, ES, RF, RG, PX, CT
(Vision)
Station AE Title
(0040,0001)
Yes = Ziehm NetPort AE Title
No = all
Empty (Vision only)
Any AE Title (Vision only)
Location
(0040,0011)
<string> or empty
Time
(0040,0003)
Range of hours
Date
(0040,0002)
Input: 24 24 = yesterday and today
00 -- 24 = today
- 24 -- 00 = yesterday
+09 15 = today from 9 until 15 oclock
- 10 10 = yesterday from 14 oclock until
today 10 oclock

A.4.4.1.2.4 Print SCP AE Titles and Ports

Form: EW_DICOM_CONF.doc

Table A.4.4-8
REMOTE AE-Title and Port PRINT SCP
Device
AE-Title
Port
Number of
Name of servers 4)
configurable
servers
Ziehm Vision
Configurable
Configurable
5
Default for first server:
net_print_server
Activate in field Print Server = 1
versions 4.xx: net_print_server = 1
Ziehm Vario3D Not supported Not supported Workstation
All other
Configurable
Configurable
2
Print Server 1 with Function Print
Print Server 2
Table A.4.4-9
CONFIGURABLE ATTRIBUTES FOR PRINT SCP
Attribute Name
Tag
VR
Value
Number of Copies
(2000,0010)
IS
1..10
Print Priority
(2000,0020)
CS
user defined : HIGH, MED, LOW or default of SCP
Medium Type
(2000,0030)
CS
user defined: PAPER, BLUE FILM, CLEAR FILM
or default of SCP
Film Destination
(2000,0040)
CS
user defined: PROCESSOR, MAGAZINE, BIN_i
(i=1..10) or default of SCP
Film Session Label
(2000,0050)
LO
user defined
Film Size ID
(2010,0050)
CS
user defined: 8INX10IN,
8_5INX11IN, 10INX12IN, 10INX14IN,
11INX14IN, 11INX17IN, 14INX14IN,
14INX17IN,24CMX24CM, 24CMX30CM, A4, A3
or default of SCP
Min Density
(2010,0120)
US
0..65535
Max Density
(2010,0130)
US
0..65535
Image Display Format
(2010,0010)
CS
STANDARD\C,R C=1..4 R=1..6
(Number per Page)
1 -> STANDARD\1,1
...
24 -> STANDARD\4,6
File: conforma215.doc
Page 36 of 54

DICOM Conformance Statement


Ziehm 8000, Ziehm 7000 PLUS, Ziehm Vista, Ziehm Vista PLUS allows to change Tag (2000,0030),
(2000,0040), (2000,0050) and (2010,0010), (2010,0050) also on user level.
Note 1)- 4) Name of Servers: DONT take reserved names: Client, Servers, Storage, Query,
Worklist, Print, echo, dicom_media

Please D O N T take a name twice!

A.4.4.2 Configuration Files


The Ziehm NetPort application uses the MergeCOM-3 subroutine library from Merge Healthcare (see
A 3.5) Linux Version 3.30 IB 25, Ziehm Vario 3D Workstation uses the DCMTK subroutine library
Version 3.5.4.
The Ziehm NetPort application references 6 configuration files. The file merge.ini, is found through the
MERGE_INI environment variable. They are as follows:
hosts

Specifies IP addresses of Ziehm NetPort and query/retrieve


and commitment servers.
ae_titel.dat(new)
Specifies the AE titel and the network parameter for the Ziehm
NetPort application and the remote SCP applications.
This file is configured with a dialog on the C-arm.
Merge.ini
Specifies the names of 3 configuration files and also
contains message logging parameters.
Mergecom.pro
Specifies run-time parameters for the Ziehm NetPort
application
mergecom.srv
Service and sequence definitions.
Mergecom.app
Specifies DICOM Services the will be negotiated with the
SCPs
With SW 5.24 and SW 2.05 the MergeCOM-3 library is replaced by OFFIS DICOM Toolkit ( Vision
only), no configuration files are used.

A.5 MEDIA INTERCHANGE


A.5.1 IMPLEMENTATION MODEL
A.5.1.1 Application Data Flow

Export to
Medium

Offline-Media
AE

Removable
Storage
Medium

Figure A.5.1-1
APPLICATION DATA FLOW DIAGRAMM FOR MEDIA STORAGE
Form: EW_DICOM_CONF.doc

The Offline- Media Application Entity exports images (single or multiframe, a set of marked
images, a whole patient folder or a set of patient folders only Vision) to a floppy (not Vision), a
MOD, a CD-R, DVD (only Vision) and USB Storage medium. It is associated with the local realworld activity Export <Medium Type>. The export is performed upon user request for the selected
images. Maximum size of Multiframe is 2 GByte.

File: conforma215.doc
Page 37 of 54

DICOM Conformance Statement

A.5.1.2 Functional Definition of AEs


A.5.1.2.1 Functional Definition of Offline-Media Application Entity
First the user has to set the format of the desired medium to DICOM or DICOM 512x512 8bit
(Vision). On the MOSAIC-screen he can select a set of images and send the export job to the desired
medium.

A.5.1.3 Sequencing of Real World Activities


At least one image must exist before Offline Media Application Entity can be invoked. The operator
must insert a new or not empty medium before invocation of the Offline-Media Application Entity. The
Offline-Media Application first finds out the free space on the medium and compares it with the size of
data to be written. Then Offline-Media Application starts to write on the medium, exception: CD-R and
DVD the data are collected in a mirror-directory and burned after user confirmation.

A.5.1.4 File Meta Information Options


The implementation information written to the File Meta header in each file is see Table 4.2-3.

A.5.2 AE SPECIFICATION
A.5.2.1 Offline-Media Application Entity Specification
The Offline-Media Application Entity provides standard conformance to the DICOM Interchange
Option of the Media Storage Service Class. The Application Profiles and roles are listed below:

Form: EW_DICOM_CONF.doc

Table A.5.2-1
APPLICATION PROFILES, ACTIVITIES AND ROLES FOR OFFLINE-MEDIA
Application Profiles
Real World Activity
FSC/FSU
FSR on
Supported
on device
device
STD-GEN-CD
Export/Import CD-R
all
all

SC Option
Interchange

STD-XA1K-CD

Export/Import CD-R

Vision

Vision

Interchange

STD-GEN-DVD (Vision only)

Export/Import to DVD

Vision

Vision

Interchange

STD-GEN-USB

Export/Import to USB-Stick

Vision
V2-0.5K

Vision
V2-0.5K

Interchange

FLOP (not Vision)

Export/Import Floppy-Disk
1.44MB

Vista

Vista

Interchange

MOD650

Export/Import MOD 650MB

all

all

Interchange

A.5.2.1.1 File Meta Information for the Application Entity


The Source Application Entity included in the File Meta Header is configurable ( see section 5.4).

File: conforma215.doc
Page 38 of 54

DICOM Conformance Statement

A.5.2.1.2 Real World Activities


A.5.2.1.2.1 Activity Export to Medium and Import from Medium
The Offline Media Application Entity acts as an FSC and FSR using the interchange option when
requested to export SOP Instances from and to the local database.
Acting as an FSC the selected images and the corresponding DICOMDIR will be written together to
the desired medium, multisession is supported. As an optional feature a DICOM viewer is burned on
the CD, DVD or written on USB-Stick. The Ziehm Vision also writes a file README.TXT.
Acting as an FSR the contents of DICOMDIR will be displayed, the user steps through the whole
DICOM information model STUDY, SERIES and IMAGE. At last it shows the image on the monitor
(see DICOM Query/Retrieve MOVE). It only supports uncompressed Transfer Syntaxes (Little
Endian). On the device Vision the contents of the whole Medium is displayed without user interaction!

A.5.2.1.2.1.1 Media Storage Application Profiles


The Offline-Media Application Entity supports the Application Profiles listed in Table A.5.2-1.

A.5.2.1.2.1.1.1 Options
The Offline-Media Application Entity supports the SOP Classes and Transfer Syntaxes listed in the
Table below:
Table A.5.2-2
IODS, SOP CLASSES AND TRANSFER SYNTAXES FOR OFFLINEMEDIA
Information Object Definition SOP Class UID
Transfer
Transfer Syntax
Syntax
UID
Media Storage Directory
1.2.840.10008.1.3.10
DICOM Explicit 1.2.840.10008.1.2.1
Storage
VR Little Endian
Computed Radiography Image 1.2.840.10008.5.1.4.1.1.1
DICOM Explicit 1.2.840.10008.1.2.1
Storage
VR Little Endian
X-Ray Angiographic Image
1.2.840.10008.5.1.4.1.1.12.1 DICOM Explicit 1.2.840.10008.1.2.1
Storage
VR Little Endian
Secondary Capture Image
1.2.840.10008.5.1.4.1.1.7
DICOM Explicit 1.2.840.10008.1.2.1
Storage
VR Little Endian
RT Image Storage
1.2.840.10008.5.1.4.1.1.481. DICOM Explicit 1.2.840.10008.1.2.1
1
VR Little Endian
Digital X-Ray Image Storage
1.2.840.10008.5.1.4.1.1.1.1 DICOM Explicit 1.2.840.10008.1.2.1
for Presentation (Vario 3D)
VR Little Endian
CT Image Storage (Vario 3D)
1.2.840.10008.5.1.4.1.1.2
DICOM Explicit 1.2.840.10008.1.2.1
VR Little Endian
FSR: all other SOP Classes
See table: A.4.2-18

A.5.3 AUGMENTED AND PRIVATE APPLICATION PROFILES

Form: EW_DICOM_CONF.doc

No private application profiles are supported.

A.5.4 MEDIA CONFIGURATION


All local applications use AE Title configured in the service dialog.

File: conforma215.doc
Page 39 of 54

DICOM Conformance Statement


A.6 SUPPORT OF CHARACTER SETS
The Ziehm NetPort DICOM applications support the
ISO-IR 100

(ISO 8859-1:1987 Latin Alphabet No. 1 supplementary set)

A.7 SECURITY
The Ziehm NetPort does not support any specific security measures.
It is assumed that the Ziehm NetPort is used within a secured environment. It is assumed that a
secured environment includes at a minimum:
a. Firewall or router protections to ensure that only approved external hosts have network access to
Ziehm NetPort.
b. Firewall or router protections to ensure that Ziehm NetPort only has network access to approved
external hosts and services.
c. Any communication with external hosts and services outside the locally secured environment use
appropriate secure network channels.
d. That configuration parameters are changed by technical specialists with due diligence, in
particular with registered port numbers, twice distributed ip-addresses.

A.8 ANNEXES
A.8.1 IOD CONTENTS
A.8.1.1 Created SOP Instances
The following tables use a number of abbreviations. The abbreviations used in the Presence
of.column are:
VNAP
Value Not Always Present
ANAP
Attribute Not Always Present
ALWAYS
Always Present
EMPTY
The abbreviations used in the Source ..column are:
MWL
Attribute from Modality Worklist
USER
Attribute from User input
AUTO
Attribute generated automatically
CONFIG
Attribute from configurable parameter

Form: EW_DICOM_CONF.doc

A.8.1.1.1 X-Ray Angiographic Image IOD

IE
Patient
Study
Series
Equipment
Image

Table A.8.1-1
IOD OF CREATED XA/XRF SOP INSTANCES
Module
Reference
Patient
Table 8.1-6
General Study
Table 8.1-7
General Series
Table 8.1-8
General Equipment
Table 8.1-10
General Image
Table 8.1-12
Image Pixel
Table 8.1-13
Cine
Table 8.1-14

Presence of Module
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
Only if Multiframe

File: conforma215.doc
Page 40 of 54

DICOM Conformance Statement


IE

Module
Multi-Frame
Mask
X-Ray Image
X-Ray Acquisition
X-Ray Collimator
XA/XRF Positioner
VOI LUT
SOP Common

Reference
Table 8.1-15
Table 8.1-16
Table 8.1-17
Table 8.1-18
Table 8.1-19
Table 8.1-35
Table 8.1-29
Table 8.1-30

Presence of Module
Only if Multiframe
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS

A.8.1.1.2 Computed Radiography Image IOD

IE
Patient
Study
Series
Equipment
Image

Table A.8.1-2
IOD OF CREATED CR SOP INSTANCES
Module
Reference
Patient
Table 8.1-6
General Study
Table 8.1-7
General Series
Table 8.1-8
CR Series
Table 8.1-20
General Equipment
Table 8.1-10
General Image
Table 8.1-12
Image Pixel
Table 8.1-13
CR Image
Table 8.1-21
VOI LUT
Table 8.1-29
SOP Common
Table 8.1-30

Presence of Module
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS

A.8.1.1.3 Secondary Capture Image IOD

IE
Patient
Study
Series
Equipment
Image

Table A.8.1-3
IOD OF CREATED OT SOP INSTANCES
Module
Reference
Patient
Table 8.1-6
General Study
Table 8.1-7
General Series
Table 8.1-8
General Equipment
Table 8.1-10
SC Equipment
Table 8.1-11
General Image
Table 8.1-12
Image Pixel
Table 8.1-13
SC Image
Table 8.1-22
VOI LUT
Table 8.1-29
SOP Common
Table 8.1-30

Presence of Module
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS

Form: EW_DICOM_CONF.doc

A.8.1.1.4 Digital X-Ray Image IOD

IE
Patient
Study
Series
Equipment
Image

Table A.8.1-4
IOD OF CREATED DX SOP INSTANCES
Module
Reference
Patient
Table 8.1-6
General Study
Table 8.1-7
General Series
Table 8.1-8
DX Series
Table 8.1-9
General Equipment
Table 8.1-10
General Image
Table 8.1-12
Image Pixel
Table 8.1-13
DX Anatomy Imaged
Table 8.1-24
DX Image
Table 8.1-23

Presence of Module
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS

File: conforma215.doc
Page 41 of 54

DICOM Conformance Statement


IE

Module
DX Detector
VOI LUT
Acquisition Context
SOP Common
Private Application

Reference
Table 8.1-25
Table 8.1-29
Table 8.1.31
Table 8.1-30
Table 8.1-32

Presence of Module
ALWAYS
ALWAYS
ALWAYS

A.8.1.1.5 COMPUTED TOMOGRAPHY Image IOD


Table A.8.1-5
IOD OF CREATED CT SOP INSTANCES
IE
Module
Reference
Patient
Patient
Table 8.1-6
Study
General Study
Table 8.1-7
Series
General Series
Table 8.1-8
Frame of Reference
Frame of Reference
Table 8.1-26
Equipment
General Equipment
Table 8.1-10
Image
General Image
Table 8.1-12
Image Plane
Table 8.1-27
Image Pixel
Table 8.1-13
CT Image
Table 8.1-28
VOI LUT
Table 8.1-29
SOP Common
Table 8.1-30

Presence of Module
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS

A.8.1.1.6 RT Image IOD

Form: EW_DICOM_CONF.doc

IE
Patient
Study
Series
Equipment
Image

Table A.8.1-37
IOD OF CREATED RT Image SOP INSTANCES
Module
Reference
Patient
Table 8.1-6
General Study
Table 8.1-7
General Series
Table 8.1-8
General Equipment
Table 8.1-10
General Image
Table 8.1-12
Image Pixel
Table 8.1-13
Cine
Table 8.1-14
Multi-Frame
Table 8.1-15
Mask
Table 8.1-16
RT Image
Table 8.1-38
VOI LUT
Table 8.1-29
SOP Common
Table 8.1-30

Presence of Module
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
Only if Multiframe
Only if Multiframe
ALWAYS
ALWAYS
ALWAYS
ALWAYS

File: conforma215.doc
Page 42 of 54

DICOM Conformance Statement


A.8.1.1.7 Common Modules
Table A.8.1-6
PATIENT MODULE OF CREATED SOP INSTANCES
Attributes Name
Tag
VR
Value
Patients Name

(0010,0010)

PN

Patient ID
Patient's Birth Date

(0010,0020)
(0010,0030)

LO
DA

Patients Sex

(0010,0040)

CS

See *)
Max 64 char
Max 64 char
Default value if not
present
M,F,O

Presence
of Value
VNAP

Source

ALWAYS
ALWAYS

MWL,User
MWL,Query,User

ALWAYS

MWL,Query,User

MWL,Query,User

*) Ziehm Vision exports Family^Given^Middle if sent via DICOM Worklist


Ziehm Vista exports Family^Given^Middle^Prefix^Suffix if sent via DICOM Worklist
Table A.8.1-7
GENERAL STUDY MODULE OF CREATED SOP INSTANCES
Attributes Name
Tag
VR Value
Presence
of Value
Study Instance UID
(0020,000D) UI See 1)
ALWAYS
Study Date
(0008,0020) DA <yyyymmdd>
ALWAYS
Study Time

(0008,0030) TM <hhmmss>

ALWAYS

Referring Physicians
Name
Study ID
Accession Number

(0008,0090) PN From Worklist

VNAP

(0020,0010) SH Req.Proc. ID
(0008,0050) SH

VNAP
VNAP

(0008,1030) LO Subject from dialog or VNAP


Scheduled Procedure
Step Description
Ref. Study Sequence
(0008,1110) SQ From Worklist
VNAP
> Ref. SOP Class UID
(0008,1150) UI
> Ref. SOP Instance UID (0008,1155) UI

Form: EW_DICOM_CONF.doc

Study Description

Table A.8.1-8
GENERAL SERIES MODULE OF CREATED SOP INSTANCES
Attributes Name
Tag
VR Value
Presence
of Value
Modality
(0008,0060) CS XA,CR,OT,DX,CT, ALWAYS
RTIMAGE
Series Instance UID
(0020,000E) UI See 1)
ALWAYS
Series Number
(0020,0011) IS 1
ALWAYS
Series Date
(0008,0021) DA <yyyymmdd>
ALWAYS
Performing Physicians
(0008,1050) PN
VNAP
Name
Protocol Name
(0018,1030) LO With MPPS
VNAP
Series Description
(0008,103E) LO
VNAP
Operators Name
(0008,1070) PN Only family name VNAP
Ref. Study Comp. Seq.
(0008,1111) SQ From MPPS
VNAP
> Ref. SOP Class UID
(0008,1150) UI 1.2.840.10008.3.1. VNAP
2.3.3
> Ref. SOP Instance UID (0008,1155) UI <MPPS UID>
VNAP
Request Attribute Seq.
(0040,0275) SQ From Worklist
VNAP

Source
MWL,AUTO
Creation Date of
patient folder
Creation Time of
patient folder
MWL
MWL
MWL, Query or
User input
MWL, User input

MWL

Source
AUTO
AUTO 1) 2)
AUTO 2)
Same as Study Date
User input
MWL
Same as 0008,1030
From Configuration
If MPPS is created

AUTO

File: conforma215.doc
Page 43 of 54

DICOM Conformance Statement


Attributes Name

Tag

VR Value

>Req. Proc. ID
>Sched Proc. Step ID
>Sched Proc. Step Desc.
>Sched Prot.Code Seq
>> code sequence

(0040,1001)
(0040,0009)
(0040,0007)
(0040,0008)

SH
SH
LO
SQ

Attributes Name

Tag

VR Value

From Worklist
From Worklist
From Worklist
From Worklist

Presence
of Value
VNAP
VNAP
VNAP
VNAP

Presence
of Value
ALWAYS

Source
MWL
MWL
MWL
MWL
See A.4.2-25

Source

Perf. Proc Step Start


(0040,0244) DA <yyyymmdd>
AUTO, see Study
Date
Date
Perf. Proc Step Start
(0040,0245) TM <hhmmss>
ALWAYS
AUTO, see Study
Time
Time
Perf. Proc. Step ID
(0040,0253) SH
ALWAYS
AUTO or input 2)
1) Overwriting of patient family name, first name or ID causes creation of new
Study Instance UID, Series Instance UID and SOP Instance UID
2) input of Tag (0040,0253) on Vision only, causes creation of a new Series (UID and
Number)
RTIMAGE Series does not export Protocol Name, Requested Attribute Sequence and Performed
Procedure Data
Table A.8.1-9
DX SERIES MODULE OF CREATED SOP INSTANCES
Attributes Name
Tag
VR
Value
Modality
Presentation Intent Type

(0008,0060)
(0008,0068)

CS
CS

DX
FOR
PRESENTATION

Presence
of Value
ALWAYS
ALWAYS

Form: EW_DICOM_CONF.doc

Table A.8.1-10
GENERAL EQUIPMENT MODULE OF CREATED SOP INSTANCES
Attributes Name
Tag
VR
Value
Presence
of Value
Manufacturer
(0008,0070)
LO
Ziehm imaging
ALWAYS
Institution Name
(0008,0080)
LO
From Hospital
VNAP
Station Name
(0008,1010)
SH
<AE Title
ALWAYS
ZiehmNetPort>
Manufacturers Model
(0008,1090)
LO
Ziehm NetPort
ALWAYS
Name
Device Serial Number
(0018,1000)
LO
<Serial Number of ALWAYS
device>
Software Version
(0018,1020)
LO
<SW-Version>
ALWAYS
Table A.8.1-11
SC EQUIPMENT MODULE OF CREATED SOP INSTANCES
Attributes Name
Tag
VR
Value
Presence
of Value
Conversion Type
(0008,0064)
LO
DI
ALWAYS

Source
AUTO
AUTO

Source
AUTO
User input
CONFIG
AUTO
AUTO
AUTO

Source
AUTO

File: conforma215.doc
Page 44 of 54

DICOM Conformance Statement


Table A.8.1-12
GENERAL IMAGE MODULE OF CREATED SOP INSTANCES
Attributes Name
Tag
VR Value
Presence
of Value
Instance Number
(0020,0013) IS
Generated by device
ALWAYS
Patient Orientation
(0020,0020) CS R/F default
ALWAYS
Content Date

(0008,0023) DA

Content Time

(0008,0033) TM

Image Comments

(0020,4000) LT

Generated by device
ALWAYS
(Creation date of image)
Generated by device
ALWAYS
(creation time of image)
From Notice
VNAP

Table A.8.1-13
IMAGE PIXEL MODULE OF CREATED SOP INSTANCES
Attributes Name
Tag
VR
Value
Samples per Pixel
Photometric
Interpretation
Rows
Columns
Bits Allocated
Bits Stored
High Bit
Pixel Representation
Pixel Data

Smallest Image Pixel


Value
Largest Images Pixel
Value

Form: EW_DICOM_CONF.doc

AUTO
User can
change
AUTO
AUTO
User input

Presence
of Value
ALWAYS
ALWAYS

Source

(0028,0002) US
(0028,0004) CS

1
MONOCHROME2

(0028,0010)
(0028,0011)
(0028,0100)
(0028,0101)
(0028,0102)
(0028,0103)
(7FE0,0010)

ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS

AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO

(0028,0106) US

576 or 1024
576 or1024
8 or 16
8 or 10
7 or 9
0
OB for Bits allocated 8
OW for Bits allocated
16
0

ALWAYS

AUTO

(0028,0107) US

255 or 1023

ALWAYS

AUTO

US
US
US
US
US
US
OW/OB

Table A.8.1-14
CINE MODULE OF CREATED XA SOP INSTANCES
Attributes Name
Tag
VR Value
Frame Time
Cine Rate

Source

(0018,1063) DS
(0018,0040) IS

If multiframe
If multiframe

Presence
of Value
ANAP
ANAP

AUTO
AUTO

Source
AUTO
User selects

Table A.8.1-15
MULTI-FRAME MODULE OF CREATED XA SOP INSTANCES
Attributes Name
Tag
VR Value
Presence
of Value
Number of Frames
(0028,0008) IS
If multiframe
ANAP

Source

Frame Increment Pointer (0028,0009) AT

AUTO

Pointer to 0018,1063

ANAP

AUTO

File: conforma215.doc
Page 45 of 54

DICOM Conformance Statement


Table A.8.1-16
MASK MODULE OF CREATED XA SOP INSTANCES
Attributes Name
Tag
VR Value
Mask Subtraction
Sequence
>Mask Operation
Recommended Viewing
Mode

(0028,6100) SQ

If multiframe

Presence
of Value
ALWAYS

(0028,6101) CS
(0028,1090) CS

NONE
NAT

ALWAYS
ALWAYS

Table A.8.1-17
X-RAY IMAGE MODULE OF CREATED XA SOP INSTANCES
Attributes Name
Tag
VR Value

Form: EW_DICOM_CONF.doc

AUTO
AUTO
AUTO

Source

(0028,1040) CS

ORIGINAL\PRIMARY\SINGLE
(acquired images)
or DERIVED\PRIMARY\
SINGLE PLANE
(post-processed images)
LIN

Presence
of Value
ANAP
ALWAYS

ALWAYS

AUTO

(0028,0002) US
(0028,0004) CS

1
MONOCHROME2

ALWAYS
ALWAYS

AUTO
AUTO

(0028,0010)
(0028,0011)
(0028,0100)
(0028,0101)
(0028,0102)
(0028,0103)
(0018,0022)
(0008,2111)
(0018,1400)

576 or 1024/512
576 or 1024/512
8 or 16
8 or 10
7 or 9
0
ROTA
Data from post processing
Data from post processing

ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ANAP
ANAP

AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO

Frame Increment Pointer (0028,0009) AT


Image Type
(0008,0008) CS

Pixel Intensity
Relationship
Samples per Pixel
Photometric
Interpretation
Rows
Columns
Bits Allocated
Bits Stored
High Bit
Pixel Representation
Scan Options
Derivation Description
Acquisition Device
Processing Description

Source

US
US
US
US
US
US
CS
ST
LO

Table A.8.1-18
X-RAY ACQUISITION MODULE OF CREATED XA SOP INSTANCES
Attributes Name
Tag
VR Value
Presence
of Value
KVP
(0018,0060) DS From Acquisition
ALWAYS
parameters
Field of View Shape
(0018,1147) CS RECTANGLE
ALWAYS
ROUND
Radiation Setting
(0018,1155) CS SC
ALWAYS
X-Ray Tube Current
(0018,1151) IS
In milliampere (mA)
ALWAYS
Type of filters
(0018,1161) LO 4 mm aluminium
ALWAYS
6.5 mm alu - 0.1 mm cop
6 mm..
5 mm..
Grid
(0018,1166) CS IN
ALWAYS
Radiation Mode
(0018,115A) CS PULSED (Vision)
ALWAYS
CONTINUOUS (Vista)
Image Area Dose
(0018,115E) DS Vision from patient folder ALWAYS

AUTO
AUTO

Source
AUTO
VISION FLAT
OTHERS
AUTO
AUTO
VISTA
VISION
VISION FLAT
VISION R
AUTO
AUTO
AUTO

File: conforma215.doc
Page 46 of 54

DICOM Conformance Statement


Attributes Name

Tag

VR

Product
Focal Spots

(0018,1190) DS

Value
Vista from image only
with MPPS Completed
0.6

Presence Source
of Value

ALWAYS

AUTO

Table A.8.1-19
X-RAY COLLIMATOR MODULE OF CREATED XA SOP INSTANCES
Attributes Name
Tag
VR Value
Presence of Source
Value
Collimator Shape
(0018,1700) CS RECTANGULAR ALWAYS
VISION FLAT
CIRCULAR
OTHERS
POLYGONAL
VISION (SW5.25)
Center of Circular
(0018,1710) IS
287\287
ALWAYS
Not VISION FLAT
Collimator
or 511\511
Radius of Circular
(0018,1712) IS
288
ALWAYS
Not VISION FLAT
Collimator
or 512
Collimator Left Vertical
(0018,1702) IS
0
ALWAYS
VISION FLAT
Edge
Collimator Right Vertical (0018,1704) IS
1023
ALWAYS
VISION FLAT
Edge
Collimator Upper
(0018,1706) IS
0
ALWAYS
VISION FLAT
Horizontal Edge
Collimator Lower
(0018,1708) IS
1023
ALWAYS
VISION FLAT
Horizontal Edge
Vertices of the Polygonal (0018,1720) IS
<row,column>\. ALWAYS
VISION (SW5.25)
Collimator

Table A.8.1-20
CR SERIES MODULE OF CREATED CR SOP INSTANCES
Attributes Name
Tag
VR Value
Presence of
Value
View Position
(0018,5101) CS Zero Length
ALWAYS
Focal Spots
(0018,1190) DS 0.6
ALWAYS

Form: EW_DICOM_CONF.doc

Table A.8.1-21
CR IMAGE MODULE OF CREATED CR SOP INSTANCES
Attributes Name
Tag
VR Value
KVP

(0018,0060) DS

X-Ray Tube Current


Distance Source to
Detector
Acquisition Device
Processing Description

(0018,1151) IS
(0018,1110) DS
(0018,1400) LO

Source
AUTO
AUTO

Presence Source
of Value
ALWAYS AUTO

From Acquisition
parameters
In milliampere (mA)
ALWAYS
970 Vision Vista
ALWAYS
1110 Vision FD
Data from post processing ANAP

AUTO
AUTO
AUTO

File: conforma215.doc
Page 47 of 54

DICOM Conformance Statement

Attributes Name

Table A.8.1-22
SC IMAGE MODULE ATTRIBUTES
Tag
VR Value

Attributes Name

Table A.8.1-23
DX IMAGE MODULE ATTRIBUTES
Tag
VR Value

Image Type
Pixel Intensity
Relationship
Pixel Intensity
Relationship Sign
Samples per Pixel
Photometric
Interpretation
Rows
Columns
Bits Allocated
Bits Stored
High Bit
Pixel Representation
Rescale Intercept
Rescale Slope
Rescale Type
Presentation LUT Shape
Lossy Image
Compression
Patient Orientation
Burned in Annotation
Window Center
Window width

Presence Source
of Value
-

(0008,0008) CS
(0028,1040) CS

ORIGINAL\PRIMARY
LIN

Presence Source
of Value
ALWAYS AUTO
ALWAYS AUTO

(0028,1041) SS

-1

ALWAYS

AUTO

(0028,0002) US
(0028,0004) CS

1
MONOCHROME2

ALWAYS
ALWAYS

AUTO
AUTO

(0028,0010)
(0028,0011)
(0028,0100)
(0028,0101)
(0028,0102)
(0028,0103)
(0028,1052)
(0028,1053)
(0028,1054)
(2050,0020)
(0028,2110)

US
US
US
US
US
US
DS
DS
LO
CS
CS

568
568
8
8
7
0
0
1
US
IDENTITY
0

ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS

AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO

(0020,0020)
(0028,0301)
(0028,1050)
(0028,1051)

CS
CS
DS
DS

L\H
20302
128
256

ALWAYS
ALWAYS
ALWAYS
ALWAYS

AUTO
AUTO
User input
User input

DX Images are Projection images for Vario 3D and support private tags for construction of 3D Images.

Attributes Name

Form: EW_DICOM_CONF.doc

Image Laterality

Attributes Name
Detector Type

Table A.8.1-24
DX ANATOMY IMAGED MODULE ATTRIBUTES
Tag
VR Value
(0020,0062) CS

Table A.8.1-25
DX DETECTOR MODULE ATTRIBUTES
Tag
VR Value
(0018,7004) CS

SCINTILLATOR

Presence Source
of Value
ALWAYS AUTO

Presence Source
of Value
ALWAYS AUTO

File: conforma215.doc
Page 48 of 54

DICOM Conformance Statement


Table A.8.1-26
FRAME OF REFERENCE MODULE ATTRIBUTES (CT SOP INSTANCES)
Attributes Name
Tag
VR Value
Presence Source
of Value
Frame of Reference UID (0020,0052) UI
Not supported!
-

Table A.8.1-27
IMAGE PLANE MODULE ATTRIBUTES (CT SOP INSTANCES)
Attributes Name
Tag
VR Value
Presence
of Value
Pixel Spacing
(0028,0030) DS 2.000\2.000
ALWAYS
Image Position (Patient)
(0020,0037) DS 0.0\0.0\0.0
ALWAYS
Image Orientation (Patient)
(0020,0037) DS 1\0\1\0\1\0
ALWAYS
Slice Thickness
(0018,0050) DS 2
ALWAYS

Attributes Name

Table A.8.1-28
CT IMAGE MODULE ATTRIBUTES
Tag
VR Value

Image Type
Pixel Intensity
Relationship
Samples per Pixel
Photometric
Interpretation
Rows
Columns
Bits Allocated
Bits Stored
High Bit
Rescale Intercept
Rescale Slope

Attributes Name

Form: EW_DICOM_CONF.doc

Attributes Name

(0008,0008) CS
(0028,1040) CS

DERIVED
LIN

(0028,0002) US
(0028,0004) CS

1
MONOCHROME2

ALWAYS
ALWAYS

AUTO
AUTO

(0028,0010)
(0028,0011)
(0028,0100)
(0028,0101)
(0028,0102)
(0028,1052)
(0028,1053)

256
256
16
16
15
0
1

ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS
ALWAYS

User input
User input
AUTO
AUTO
AUTO
AUTO
AUTO

Presence
of Value
ALWAYS
ALWAYS
ALWAYS

Source

US
US
US
US
US
DS
DS

(0028,1050) DS
(0028,1051) DS
(0028,1055) LO

0..255 or 0..1023
1..256 or 1..1024
[C-W/2; C+W/2] -> [0;255]
{x| y=(255/W)*(x-C+W/2)}

Table A.8.1-30
SOP COMMON MODULE ATTRIBUTES
Tag
VR Value

Specific Character Set


SOP Class UID
SOP Instance UID

?
?
?
?

Presence Source
of Value
ALWAYS AUTO
ALWAYS AUTO

Table A.8.1-29
VOI LUT TABLE MODULE ATTRIBUTES
Tag
VR Value

Window Center
Window Width
Window Center & Width
Explanation

Source

(0008,0005)
(0008,0016)
(0008,0018)

CS
UI
UI

ISO_IR 100
See Table A.4.2-1
Generated by device

Presence
of Value
ALWAYS
ALWAYS
ALWAYS

User input
User input
AUTO

Source
AUTO
AUTO
AUTO

File: conforma215.doc
Page 49 of 54

DICOM Conformance Statement

Table A.8.1-31
AQUISITION CONTEXT MODULE ATTRIBUTES (DX SOP INSTANCES)
Tag
VR Value
Presence Source
of Value
-

Attributes Name

Table A.8.1-35
XA/XRF POSITIONER MODULE ATTRIBUTES
Tag
VR Value

Form: EW_DICOM_CONF.doc

Attributes Name

Distance Source To Detector

(0018,1110)

CS

Positioner Motion
Positioner Primary Angle
Positioner Secondary Angle

(0018,1500)
(0018,1510)
(0018,1511)

CS
DS
DS

970 Vision Vista


1110 Vision FD
STATIC
Zero Length
Zero Length

Presence of
Value
ALWAYS

Source

If Cine-Loop
ALWAYS for XA
ALWAYS for XA

AUTO
AUTO
AUTO

Table A.8.1-37
RT IMAGE MODULE OF CREATED RT SOP INSTANCES (not Vista)
Attributes Name
Tag
VR Value
Presence
of Value
Image Type
(0008,0008)
CS ORIGINAL\PRIMARY\D ALWAYS
RR (acquired)
DERIVED\PRIMARY\
DRR (post-processed)
Samples per Pixel
(0028,0002)
US 1
ALWAYS
Photometric Interpretation
(0028,0004)
CS MONOCHROME2
ALWAYS
Rows
(0028,0010)
US 1024/512
ALWAYS
Columns
(0028,0011)
US 1024/512
ALWAYS
Bits Allocated
(0028,0100)
US 8 or 16
ALWAYS
Bits Stored
(0028,0101)
US 8 or 10/16
ALWAYS
High Bit
(0028,0102)
US 7 or 9/15
ALWAYS
Pixel Representation
(0028,0103)
US 0
ALWAYS
RT Image Label
(3002,0002)
SH Vario 3D
ALWAYS
Conversion Type
(0008,0064)
CS DI
ALWAYS
RT Image Plane
(3002,000C)
CS NORMAL
ALWAYS
X-Ray Image Receptor Angle (3002,000E)
DS Rotation angle of image ALWAYS
Image Plane Pixel Spacing
(3002,0011)
DS <Depends from unit>
ALWAYS
RT Image Position
(3002,0012)
DS From Vario Unit
ALWAYS
Radiation Machine Name
(3002,0020)
SH VARIO
ALWAYS
Primary Dosimeter Unit
(300A,00B3)
CS MINUTE
ALWAYS
Radiation Machine SAD
(3002,0022)
DS From Vario
VNAP
RT Image SID
(3002,0026)
DS See (0018,1110) in XA
ALWAYS
Image
Exposure Sequence
(3002,0030)
SQ
ALWAYS
>KVP
(0018,0060)
DS See XA Image
ALWAYS
>X-Ray Tube Current
(0018,1151)
IS
See XA Image
ALWAYS
>Number Of Blocks
(300A,00F0)
IS
0
ALWAYS
Gantry Angle
(300A,011E)
DS From Vario
VNAP

AUTO

Source
AUTO

AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO
AUTO

File: conforma215.doc
Page 50 of 54

DICOM Conformance Statement

A.8.1.1.8
A.8.1.1.8.1 Attributes of DICOMDIR
Table A.8.1-32
Tag
VR
Value/Examples
0002,0001
OB
1

Description
Constant

0002,0002

UI

1.2.840.10008.1.3.10

Constant

0002,0003

UI

0002,0010
0002,0012

UI
UI

2.16.840.1.113669.632.6.1.22 constant, x=2 floppy,


00.20030514.4101500.x0
x=3 MOD/Stick,
x=4 CD
1.2.840.10008.1.2.1
Constant
see A.4.2.1.2.4
Constant

0002,0013

SH

see A.4.2.1.2.4

Constant

0002,0016

AE

Ziehm NetPort AE

user defined

0004,1130

CS

File Set Consistency Flag 0004,1212


Directory Record Type
0004,1430

US
CS

Referenced FileID
Referenced SOP Class
UID In File
Referenced SOP
Instance UID In File
Patient's Name
Patient ID
Patient's Birth Date
Patient's Sex
Study Date
Study Time
Study Description
Study Instance UID

0004,1500
0004,1510

CS
UI

NETPORT_F<serial number> automatically


generated
0x0000
Constant
PATIENT, STUDY, SERIES, Constant
IMAGE
XA00001 or XA\XA00001
Filename on medium
1.2.840.10008.5.1.4.1.1.7
SOP Class UID

0004,1520

UI

0010,0010
0010,0020
0010,0030
0010,0040
0008,0020
0008,0030
0008,1030
0020,000D

PN
LO
DA
CS
DA
TM
LO
UI

Study ID

0020,0010

SH

Requested Proc. Step ID

Modality
Series Instance UID

0008,0060
0020,000E

CS
UI

Series Number

0020,0011

IS

Series Date
Series Description
Image Type
Image Date
Image Time
Image Number

0008,0021
0008,103E
0008,0008
0008,0023
0008,0033
0020,0013

DA
LO
CS
DA
TM
CS

XA,CR,OT,RF,RT,RG
2.16.840.1.113669.632.6.1.22
00.20030514.1100200.10
100000
Number of patient
folder
20030415
Date of Patient Folder
Right hand thumb
user supplied
see Table A.8.1-17
20030514
Image Date
101502
Image Time
11
Constant

Attributes Name
File Meta Information
Version
Media Storage SOP
Class UID
Media Storage SOP
Instance UID

Form: EW_DICOM_CONF.doc

Transfer Syntax UID


Implementation Class
UID
Implementation Version
Name
Source Applicaion Entity
Title
File Set ID

2.16.840.1.113669.632.6.1.22
00.20030514.1101502.11
Doe^John^^
51
19400101
M
20030514
100200
right hand thumb
2.16.840.1.113669.632.6.1.22
00.20020514.11002000

SOP Instance UID


user supplied
user supplied
user supplied
user supplied
Date of patient folder
Time of patient folder
user supplied
generated
automatically or
supplied by MWL
generated
automatically or
supplied by MWL
Constant

File: conforma215.doc
Page 51 of 54

DICOM Conformance Statement

A.8.1.1.8.2 Example README.TXT (Vision)


Institution:
This disk was created by:
<Institution Name>
Media Creator Application:
This disk was created using:
dcmtk 3.5.4
(for the DICOMDIR and DICOM files)
MKISOFS ISO 9660/HFS FILESYSTEM BUILDER & CDRECORD CD-R/DVD CREATOR (C)
1993 E.YOUNGDALE (C) 1997 J.PEARSON/J.SCHILLING
(for burning CDs)
Vision Software 5.0.5 10.02.2006
and a manual process.
Contact Information:
If you have any questions about how this disk or the data on
it were generated, please contact:
Ziehm Imaging GmbH
Donaustr. 31
D-90451 Nuernberg
ph: +49-911-2172-500
email: [email protected]
Web: www.ziehm.com

Media Organization:
Dicom images are located in folders named XA, CR, OT, RT, RG, RF.

A.8.1.1.9 Storage Commitment Request N-ACTION Information

Form: EW_DICOM_CONF.doc

Table A.8.1-33
Attributes Name
Requested SOP Class UID
Action Type ID
Transaction UID

Tag
0000,1001
0000,1008
0008,1195

Referenced SOP Sequence


Referenced SOP Class UID
Referenced SOP Instance UID

VR
UI
US
UI

Value
1.2.840.1008.1.20.1.1
1
|.16.840.1.113669.632
.6.1.1621.20030515.3
1258340

Description
constant
constant
constructed from :
Ziehm ROOT.UID.
<serialNo>.<date>.<3time>

0008,1199
SQ
>0008,1150 UI

1.2.840.5.1.4.1.1.12.1

Selected Storage SOP


Class

>0008,1155 UI

=(0008,0018)

File: conforma215.doc
Page 52 of 54

DICOM Conformance Statement

A.8.2 DATA DICTIONARY OF PRIVATE ATTRIBUTES


The Private Attributes added to created SOP Instances are listed in the Table below:
Table A.8.1-34
PRIVATE APPLICATION MODULE ATTRIBUTES (DX SOP INSTANCES)
Attributes Name
Tag
VR Value
Presence Source
of Value
PT
(0029,0010)
18761
ALWAYS AUTO
PT: X-ray Image Position (0029,1000)
104.704114/
ALWAYS AUTO
PT: X-Ray Spot Position (0029,1001)
0.729651/...
ALWAYS AUTO
PT: Pixel Value for
(0029,1002)
48
ALWAYS AUTO
background mask
PT: Image shift
(0029,1003)
3169328
ALWAYS AUTO
PT
(0031,0010)
4935753
ALWAYS AUTO
PT: X-ray Image Position (0031,1000)
1...
ALWAYS AUTO
PT: X-Ray Spot Position (0031,1001)
ALWAYS AUTO
PT: Pixel Value for
(0031,1002)
ALWAYS AUTO
background mask

A.8.3 ZIEHM NAVIPORT 2D


The Ziehm NaviPort 2D has been created in the Year 2003.

A.8.3.1 Configuration
The configuration is described in A.4.4.1.2.1 Table A.4.4-3.

A.8.3.2 Navigation Flag


Ziehm Vision:
Others:

switch on Navigation in dialog Config Operation Settings


set Function to Navigation in dialog DICOM Settings Print Server

A.8.3.3 Behavior
If Navigation is on the C-arm sends two Udp messages to the configured navigation server at port
56666, the first after x-ray-off, the second at the end of stack filter.
Pressing STORE at the mobile stand, the image will be sent by DICOM Store to the navigation server.
The corresponding image archives the date and time of the udp-messages into the DICOM tags
Content Date (0008,0023) and Content Time (0008,0033) (Vision: net_naviImage = 2)
and
Study Date (0008,0020) and Study Time (0008,0030)
(Vision: net_naviStudy = 2)
Only XA-single images without image attributes are stored.

Form: EW_DICOM_CONF.doc

A.8.3.4 Form of UDP messages


<AETitle Ziehm NetPort>;<SW-Version>;<No of stack filter>;<No of message>;<date>;<time>;
i.e: first message:
ZiehmNP-F2526;21.000;4;1;20060930;120411;
second message: ZiehmNP-F2526;21.000;4;2;20060930;120411;

File: conforma215.doc
Page 53 of 54

DICOM Conformance Statement


A.8.4 IHE INTEGRATION STATEMENT

IHE Integration Statement

Date:
30.04.2008

Vendor

Product Name

Version

Ziehm Imaging GmbH

Ziehm Vision

2.0

This product implements all transactions required in the IHE Technical Framework to
support the IHE Integration Profiles, Actors and Options listed below:
domain: Radiology
Integration Profiles
Implemented

Actors Implemented

Scheduled Workflow (SWF)

Acquisition Modality

Patient Information
Reconciliation (PIR)

Acquisition Modality

Options Implemented

Internet address for vendor's IHE information :


www.ziehm.com
Links to Standards Conformance Statements for the implementation
Links to general information on IHE

Form: EW_DICOM_CONF.doc

In North America :
http://www.ihe.net

In Europe :
http://www.ihe-europe.org

In Japan :
http://www.jira-net.or.jp/ihe-j

File: conforma215.doc
Page 54 of 54

You might also like