0% found this document useful (0 votes)
157 views1 page

Termostat

This document provides specifications for the MT4N series panel meter. It can measure DC/AC voltage, current, and frequency. It has a 4-digit LCD display and measurement accuracy of 0.1% for DC signals. Outputs include relay, open collector, RS-485, and 4-20mA. It operates from 12-24VDC/AC or 100-240VAC power. The panel cut-out size is 45x37mm. It has functions for preset limits, initialization, error display, and parameter settings accessed via the front keys.

Uploaded by

DIANTORO
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
157 views1 page

Termostat

This document provides specifications for the MT4N series panel meter. It can measure DC/AC voltage, current, and frequency. It has a 4-digit LCD display and measurement accuracy of 0.1% for DC signals. Outputs include relay, open collector, RS-485, and 4-20mA. It operates from 12-24VDC/AC or 100-240VAC power. The panel cut-out size is 45x37mm. It has functions for preset limits, initialization, error display, and parameter settings accessed via the front keys.

Uploaded by

DIANTORO
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

Specifications

PANEL METER

MT4N SERIES
M

Parameter

Model

Power supply

12-24 VDC/AC

Power consumption
Display method

DC: 3W, AC: 5VA


5VA
7 Segment LCD Display, Character height: 9mm
23 5 DC Type: F.S.0.1% rdg2digit / AC Type: F.S.0.3% rdg3digit
DC/AC Type: Within F.S.0.3% rdg3digit only for Current 5A terminal
-10 to 50 DC/AC Type: F.S.0.5% rdg3digit
DC Voltage/Current, AC Voltage/Current, AC Frequency
110% F.S. for input spec.
Practical oversampling using successive approximation ADC
50ms(DC), 16.6ms(AC)(1/12,000)
-1999 to 9999(4 Digit)
Relay output Contact capacity: 125VAC 0.3A, 30VDC 1A/Contact composition: N.O(1a)
NPN/PNP Open Collector output 12-24VDC 2V 50mA Max.(Load resistance)
RS485 communication output
Baud rate: 1200/2400/4800/9600, Transmission method: 2 wires half duplex, Tuning
method: Sub-synchronization, Protocol: Modbus type
DC4-20mA output Resolution: 12,000 division(Load resistance max. 600)
Selectable RMS or AVG

Display accuracy
Input
Max. allowable input
A/D conversion method
Sampling cycle
Max. display range

: Upgrade feature

Parameter

MT4N
100-240 VAC

PA1
(Parameter 1)

Preset output

Sub output
(Transmission output)

Thank you very much for selecting Autonics products.


For your safety, please read the following before using.

AC measuring function
Frequency measuring
function
Hold function1
Insulation resistance

Caution for your safety


Please keep these instructions and review them before using this unit.

Please observe the cautions that follow;


Warning Serious injury may result if instructions are not followed.
Caution Product may be damaged, or injury may result if instructions are not followed.
The following is an explanation of the symbols used in the operation manual.
Caution: Injury or danger may occur under special conditions.

Warning
1.In case of using this unit with machinery(Nuclear power control, medical equipment, vehicle, train, airplane,
combustion apparatus, entertainment or safety device etc), it is required to install fail-safe device.
It may cause a fire, human injury or property loss.
2.It must be mounted on panel.
It may give an electric shock.
3.Do not connect, inspect or repair terminals when it is power on.
It may give an electric shock.
4.Do not disassemble or modify this unit. If needs, please contact us.
It may cause a fire and give an electric shock.
5.Please check the number of terminal when connecting power or measured input.
It may cause a fire.

Caution
1.This unit shall not be used outdoors.
It might shorten the life cycle of the product or give an electric shock.
Use this product indoors only. Do not use the product outdoors or at locations subject to the
temperatures or humidity outside.(Example: rain, dirty, frost, sunlight, condensation, etc.)
2.When connecting wire, use AWG 20(0.50) be used and tighten screw bolt on terminal block
with 0.74 to 0.90N.m strength.
It may cause a malfunction or fire due to contact failure.
3.Please observe the rated specification.
It might shorten the life cycle of the product and cause a fire.
4.Do not use beyond of the rated switching capacity of relay contact.
It may cause insulation failure, contact melt, contact failure, relay broken and fire etc.
5.In cleaning the unit, do not use water or an oil-based detergent.
It may cause a fire and give an electric shock.
6.Do not use this unit in place where flammable or explosive gas, humidity, direct ray of the
light, radiant heat, vibration or impact, etc. exists.
It may cause a fire or explosion.
7.Do not inflow dust or wire dregs into the unit.
It may cause a fire or mechanical malfunction.
8.Please wire properly after checking the polarity of measuring terminals.
It may cause a fire or explosion.

Ordering information
DV

E N

N
0
1
2
3
4

Control output

Power supply

Without output function


Relay output(2 Contacts)
NPN Open collector output(OUT1,GO,OUT2)
PNP Open collector output(OUT1,GO,OUT2)
Relay(OUT1)+PV transmission(DC4-20mA)output
Relay(OUT2)+RS485 communication output

E
4

12-24VDC/AC
100-240VAC

DV
DA
AV
AA

Input

Size
Digit
Item

DC Volt
DC Ampere
AC Volt
AC Ampere

9999(4 Digit)
Multi Meter

Panel cut-out
(Unit: mm)

Min. 55

OUT1: Preset output of OUT1


GO: Preset Go output of OUT1/OUT2
OUT2: Preset output of OUT2
MODE : MODE key
: Shift key
mV, V Unit
Min. 37
: Down key
mA, A Unit
: Up key
Hz Unit

MT4N-DV-

5V/1V

45

OUT1

SOURCE
12-24VDC/AC
100-240VAC

OUT2

HOLD/ZERO

OUT1

OUT1

GO

OUT2

8
GO

HOLD/ZERO

HOLD/ZERO

DC4-20mA

10 11 12
OUT1

HOLD/ZERO

10 11 12
OUT1

This function is to display setting(-1999 to 9999) of particular High/Low-limit value in order to display
High/Low-limit value of measured input. If measured inputs are 'a' and 'b' and particular values are 'A'
and 'B', it will display a=A, b=B as below graphs.
B
A

A
a

b
Input value

Display
B
a
b
Input value

B
Display
A

Input value

a
A
Display
B
a
b
Input value

Error display function


Display

HHHH
LLLL

0.00 to 50.00(Fixed)
0.00 to 10.00(Fixed)
0.000 to 5.000(Fixed)
0.000 to 1.000(Fixed)
0.0 to 250.0(Fixed)
0.00 to 50.00(Fixed)

0-500mA

[500mA]

0.1

0.0 to 500.0(Fixed)

0-200mA
0-50mA
4-20mA
0-5mA
0-2mA
0-250V
0-125V

[200mA]
[50mA]
[4-20mA]
[5mA]
[2mA]
[250V]
[125V]

0.1
1.1
1.1
11.1
11.1
1.109
1.109

0.0 to 200.0(Fixed)
0.00 to 50.00(Fixed)
4.00 to 20.00(Fixed)
0.000 to 5.000(Fixed)
0.000 to 2.000(Fixed)
0.0 to 250.0(Fixed)
0.0 to 125.0(Fixed)

0-50V

[50V]

222

0.00 to 50.00(Fixed)

0-25V
0-5V
0-2.5V
0-5A
0-2.5A
0-500mA
0-250mA
0-100mA
0-50mA

[25V]
[5V]
[2.5V]
[5A]
[2.5A]
[500mA]
[250mA]
[100mA]
[50mA]

222
22
22
0.01
0.01
0.1
0.1
0.5
0.5

0.00 to 25.00(Fixed)
0.000 to 5.000(Fixed)
0.000 to 2.500(Fixed)
0.000 to 5.000(Fixed)
0.000 to 2.500(Fixed)
0.0 to 500.0(Fixed)
0.0 to 250.0(Fixed)
0.0 to 100.0(Fixed)
0.00 to 50.00(Fixed)

Description
Flashes when measured input is exceeded
the max.allowable input(110%)
Flashes when measured input is exceeded
the min.allowable input(-10%)
Flashes when display input is exceeded H-SC setting value

D-HH
D-LL Flashes when display input is exceeded L-SC setting value
Flashes when input frequency is exceeded the max. display

F-HH value of measuring range


OVER Flashes when it exceeds zero range(99)

Zero adjusting error is returning to measuring mode after


[ OVER] flashes twice.
Refer to " Measuring AC frequency function" for frequency
measuring range.

Display
B
Input value
Input value

Initialization function
It initializes parameter setting state. Press
+ + keys simultaneously for over 5 sec. in
RUN status, it returns to factory default value.
Ex)

PA1

-1999 to 9999(Variable)
-199.9 to 999.9(Variable)
-19.99 to 99.99(Variable)
-1.999 to 9.999(Variable)
(Display position will be
different depending on
decimal point position.)
Please wire proper terminal
to its max. input voltage
within 30 to 100% of input
terminal.
When it is higher than
input voltage, it may cause
breakdown of terminal and
over display range and
the accuracy is decreased
when it is connected to
the terminal under 30%.

Press MODE key for 2 sec. in RUN mode, [ PA1](Parameter 1) is displayed.


Press MODE key for 4 sec. in RUN mode, [ PA2] is displayed after [ PA1].
When press MODE key is pressed continually, it stops displaying at [ PA2].
It is advanced to current display parameter releasing MODE key at [ PA1] or [ PA2].
Press MODE key for 3 sec., it is returned to RUN at any position.
If any key is not touched for 60 sec. in each parameter, it will return to RUN mode.
After return to RUN mode, press MODE key within 2 sec., it will return to
previous parameter.(Refer to the below descriptions for set parameter.)

PA2

Change the parameter


setting value

Setting
value

Ex) Mode

IN-R

Press one

Setting
value

Mode

Press one

Input
Opercorrection Front key
ation
value

MODE reset.

+ + Press
for over 5 sec.

IN1 T
NO

0.2 0.4 0.6 0.8 1V Input value

measured input and range".)

MODE

Set display type for measured input as


STND SCAL FREQ and FREQ.
It is only available for AC type.

DISP
MODE

Select measuring type for AC input type.

<When DISP is STND>


value of standard
specification.
Display value is fixed.

MODE

DOT

MODE

H-SC
MODE

L-SC
MODE

D.UNT
MODE

Mode

OUT1 output

This time function limits the operation of an output until the measured input(overvoltage or inrush
current) is stable at moment of power on. All outputs are off during startup compensation time
setting after power is applied.
Setting range: 00.0 to 99.9 (Unit: sec.) Factory default: 00.0

0.000

5.000

0.000

3.750

0.000

3.000

0.000

HI

Hysterisis

LO

OUT1.L
OUT1 output

HL

OUT1.H
OUT1.L
OUT1 output

HL-G

OUT1.H
OUT1.L
OUT1 output

Set display unit.


Set range:mV/V/mA/A/Hz/off
Adjust High-limit display value
gradient for max. input.

INB.H

Adjust Low-limit display value


deviation for min. input.

INB.E

Parameter 2
PA2
Select Preset output mode of OUT1 as
OFF HI LO HL HL-G
MODE But, it is only displayed in OUT1 output
included model.

OU1.T

Select Preset output mode of OUT2 as


OFF HI LO HL HL-G
MODE But, it is only displayed in OUT2 output
included model.

OU2.T

Select Preset hysteresis of OUT1 within 10%


of F.S. But, it is not displayed when OU1.T
MODE mode is
OFF.

HYS.1

HYS.2
MODE

STA.T
MODE

Select Preset hysteresis of OUT2 within


10% of F.S. But, it is not displayed when
OU2.Tmode is OFF.
Set startup compensation time.
Set range is 0.0 to 99.9 sec.
Set monitoring delay time within 00 to 30 sec.

MODE

DIS.T
MODE

X
Input value

Execute variable set display period by 0.1


sec. within 0.1 to 5.0 sec.
Select color with 5 modes.
RED GRN YEL R-G G-R

RED, GRN, YEL: Display with 1 color.

R-G, G-R : Color is changed when error is occurred.


Ex) R-G : Red is standard and green when error is
occurred.
Color is changed only when error in" Error
display function" is occurred excluding 'over'.

Period ON: Display valueOUT1.H


Period OFF: Display valueOUT1.H-Hys
Period ON: Display valueOUT1.L
Period OFF: Display valueOUT1.L+HYS
Period ON: Display valueOUT1.L
or Display valueOUT1.H
Period OFF: Display valueOUT.L+Hys or
Display valueOUT.H+Hys
Period ON: OUT1.LDisplay valueOUT1.H+Hys
Period OFF: Display valueOUT.L+Hys or
Display valueOUT.H+Hys

Set output mode separately for each OUT1/OUT2.


OUT1/OUT2 are operated individually depending on output operation mode.
Setting value mode of parameter group 0 is displayed depending on output operation mode.
GO outputs when the period both OUT1/OUT2 are off.(NPN/PNP Open collector output type)

Adjust High-limit display value


gradient for max. input.

MODE Set range: 0.100 to 9.999

MODE

Set index for frequency display.


Set range:10-210-1
10 010 1

Enable zero adjustment by front key operation


to select YES. Press both + keys simultaneously in 3 sec. The deviation value is saved in
MODE INB.L automatically.

ZERO

EVIN
MODE

Select between hold input by terminal 11, 12


and zero set by external signal.

HOLD: Hold display value

ZERO: Zero adjustment by external signal

Set High-limit value, 20mADC output position of PV output.


When input range and prescale mode are
changed, the set value of FS-H and FS-L is
MODE
change automatically as max/min. value of input
range.
Set Low-limit value, 4mADC output position
FS-L
of PV output.
MODE

FS-H

ADRS
MODE

BPS
MODE

PRTY
MODE

Remark

Operation

<When DISP is FREQ>


Set frequency measuring range.
(Refer to " Measuring AC
frequency function" .)

DOT

MODE

Set display value for min.


input of measured input.

MODE Set range: -99 to 99

MODE

No output

OUT1.H

INB.L

COLR

No output

Display cycle delay function [PA 2: DIS.T mode]

Startup compensation timer function [ PA 2: STA.T mode]

0.000

7.500

Output operation

OUT1 output

In some applications the measured input may fluctuate which in turn causes the display to fluctuate. By
adjusting the display cycle delay function time in the DIS.T mode in parameter 2, the operator can adjust
the display time within a range of 0.1 sec to 5 sec. For example, if the operator sets the display cycle time
to 4.0 sec., the display value displayed will be the average input value over 4 sec. and also will show any
changes if any every 4 sec.

Unavailable

Set display value for max.


input of measured input.

MODE Set range: 0.100 to 5.000

Time

1.000 2.000
3.000 In this case, any setting
methods display same
4.000
display value.
5.000

OFF

INB.H

Time

INB.H

Measured input range

MT4N-DV 50V 10V 5V 1V 250mV 50mV 50V


500mA 200mA 50mA 4-20mA 5mA
MT4N-DA
2mA 500mA
MT4N-AV 250V 125V 50V 25V 5V 2.5V 250V
SA 25A 500mA 250mA 100mA 50mA
MT4N-AA
5A

<When DISP is SCAL>


Set decimal point position.
0 0.0 0.00 0.000

STND It shows max. display

PEK.T

L-SC

If PEK.T monitoring delay time in PA2 is '00', H.PEK and


L.PEK are not displayed.

IN-T Select RM5 AU6

MODE

(Figure1)

Preset output mode [ PA 2: OU1.T / OU2.T mode ]

YES
MODE

Display value for


measured input

H-SC

MODE reset.

<Measured input specification for each model>

Select measured input specification.

It displays Low peak value in RUN mode


and
keys are pressed, it is

L.PEk

Model

IN-R (Refer to "Specification of

Short-circuit
external Hold
terminal no.11,
12 over min.50m.
It is enable to
use in option
mode.

Y
Y

keys.

It displays High peak value in RUN mode


and
keys are pressed, it is

H.PEk

PA1

Refer to description [ Error correction function],


[ Error indication function] and [ Parameter 2]
for function and error.

Ex) To display "3.000" in DC 200mV input for measured input specification as 0 to 1V,
Select 0-1VDC for measured input in Parameter 1
Standard specification in input: 0-1VDC and 1.000 therefore it has to be 15.000( H-SC)
for 1VDC(Input) in order to display 3.000 for 200mVDC(input).
But it is unable due to setting range is 9.999.
In this case, please check below chart. Please set as INB.H H-SC = 15.000
Setting
method

Change the value by

Set Low-limit preset value of OU2.T


(It is not displayed when OU2.T mode
MODE value of PA2 is OFF.)
Change the value by
keys.

OU2.L

Parameter 1

Input external
signal

Press both
,
keys for 3
sec. at the
measuring
mode.

It corrects a gradient of prescale value and display value.


(Figure 1)Display value Y can be adjusted as , times against X input value by
correction function [ INB.H ] and used as correction function of max. display value
[ H-SC ]. Adjustment range is 0.100 to 5.000 and multiply current gradient by the
value.

15.000
12.000
9.000
6.000
3.000
1.000

MODE value of PA2 is OFF.)

125V

Zero adjustment function


It adjusts the display value of the optional configured
input value as zero by force, zero point error can be
adjusted with 3 ways as below.
When zero point adjustment with front key and
Hold terminal is finished normally, zero point
of measuring terminal is displayed and the
adjusted value is saved in INB.L automatically.

Set High-limit preset value of OU2.T.


(It is not displayed when OU2.T mode

OU2.H

Setting
value MODE

250V

PA1: Direct
input
correction
Desvalue
cription
method
at INB.
Lmode

keys.
Change the value by
Set Low-limit preset value of OU1.T.
(It is not displayed when OU1.T mode
MODE value of PA2 is OFF.)
Change the value by
keys.

OU1.L

4.Change the set value by


or
key when setting
value is flashed.
Ex) Change AC type measured input from 250V to 125V.

IN-R

Set High-limit preset value of OU1.T.


(It is not displayed when OU1.T mode

MODE value of PA2 is OFF.)

Saved setting value

0.100 to 5000Hz: Display accuracy error within


0.3%F.S.2Digit
5000 to 9999Hz: Display accuracy error within
1%F.S.3Digit
INB.H: 0.100 to 9.999 (Gradient adjustment of high value)
INB.E:10-2, 10-1, 10 0, 10 1 (Index adjustment of INB.H)

Display value

MODE

OU1.H

250V flashes every 0.5 sec.

Decimal point
0.000
0.00
0.0
0
position
Measurement 0.100 to 0.10 to 0.1 to
1 to
range
9.999Hz 99.99Hz 999.9Hz 9999Hz

FS-H

RUN

1.Advance to the parameter to be changed when


press MODE key continuously in RUN mode and
release MODE key at the parameter.(Refer to "
Parameter setting")
2.When press MODE key in each parameter, the
initial mode of the parameter is displayed.(Refer
to the description of each parameter.)
,
keys in display mode.
3.When press one of ,

DISP

It measures input signal frequency when it is an AC input


using fixed decimal point[PA1: DOT mode]. Measuring
range can be changed by setting and measuring range
of decimal point position is as below chart. It is available
to adjust upper gradient in [PA 1: INB.H mode] and [PA
1: INB.E mode]. In order to measure frequency normally,
input signal, over 10% F.S. of measuring range, should be
supplied. Please wire the proper measuring terminal.
Measuring range

Min. setting range


10% F.S.

Parameter 0

5. When press MODE key to complete the change and


it is advanced to the next mode after flashes 2 times.
6. When press MODE key for 3 sec. after change, it
returns to RUN mode.

Measuring AC frequency
function [ PA 1: DISP mode ]

It corrects display value error of measured input.


INB.L:99 (Adjust deviation of low value)
INB.H: 5.000 to 0.100(Correct gradient(%) of high value)
Display value=(Measured value INB.H)+ INB.L
Ex) Low value correction
When there is an application where there is a residual
voltage of 1.2V, but a 0V display is desired, then it
is possible by adjusting the INB.L parameter setting
to 12(offset correcting value or equal to 1.2V without
decimal) that the desired display value of 0 can be
achieved.
Ex) High value correction
When there is an application where the high actual value
of display is 501 and exceeds the 500V display range,
then it is possible by adjusting the INB.H parameter
setting to 0.998(calculated by desired value of 500/actual
value of 501), that the desired value can be achieved.
The offset correction range of INB.L is within -99 to
99 for D-0, D-1 digit regardless of decimal point.

RUN

The above specifications are subject to change without notice.

MODE key

2 sec.

Gradient correction function [ PA 1: INB.H mode ]

HOLD/ZERO

Prescale function[PA1: H-SC/ L-SC mode]


Display
B

434.35
434.35
43.35
43.35
2.15
2.15

[ PA 1: INB.H/ INB.Lmode ]

A(+)

Display

[50V]
[10V]
[5V]
[1V]
[250mV]
[50mV]

Error correction function

Relay+RS485 communication output [MT4N--4]

B(-)

0-50V
0-10V
0-5V
0-1V
0-250mV
0-50mV

FS-L

Relay+Current(DC4-20mA) output [MT4N--3]

10 11 12

OU2.H
OU2.L

MODE key

Praescale [SCAL]

[ PA 2: FS-H/ FS-Lmode ]

10 11 12

OUT2

ZERO

Set monitoring delay time for peak value(sec) Set range: 00 to 30 sec.
Set sampling time(sec)
0.1 to 5.0 sec.(Variable by 0.1 sec.)
Select color
RED / GRN / YEL / R-G / G-R
No: Enable/disable zero adjustment key
Zero key
Enable zero adjustment key
Yes: Enable zero adjustment key
Hold: Use external terminal as Hold terminal
Zero: Use external terminal as zero adjustment terminal
Event input
Set external terminal(11, 12) function
It is enable to use in option mode.
Full scale high
Set High-limit value output position of PV output Min. set range: Min. 10% F.S.
Full scale low
Set Low-limit value output position of PV output Max. set range: Max. FS-H 10%
Address
Set communication address
Set range: 01 to 99
Bit per second
Set baud rate(bps)
Selectable:1200/2400/4800/9600
Parity bit
Set parity bit
Selectable: None/ Even / Odd
Stop bit
Set stop bit
Set range: 1/2
Response wating time Set response wating time
Set range: 5 to 99
Lock
Enable lock status
Selectable: OFF/LOC1/LOC2/LOC3
Set the range within display range of
OUT1 high preset
Set value of OUT1 High-limit output
STND/SCAL.
OUT1 low preset
Set value of OUT1 Low-limit output
For MT4N-DV/DA Type, set range of
OUT2 high preset
Set value of OUT2 High-limit output
OU1.H/OU2.H and OU1.L/OU2.L is within -5
OUT2 low preset
Set value of OUT2 Low-limit output
to 110%.
High peak
Max. value by data monitoring
key is pressed, it will be returned
to initial status.
Low peak
Min. value by data monitoring

NPN Open Collector output [MT4N--1]

Set range: 0.0 to 99.9 sec.

2 sec.

Display range [Variable]

Current output(DC4-20mA)
Scale adjustment function

4mA

SOURCE
12-24VDC/AC
100-240VAC

5A/2.5A

PNP Open Collector output [MT4N--2]

500mA/250mA

10 11 12

Set startup compensation time

RUN

It monitors Max./Min. value of display value based on current display value and then display the data in H.PEK
mode and L.PEK mode of parameter 0. Set delay time(0 to 30 sec.) in PEK.T mode of parameter 2 in order to avoid
caused by initial overcurrent or overvoltage, when monitoring the peak value. Delay time is 0 to 30 sec. and it starts
to monitor the peak value after set time.
keys are pressed at H.PEK and L.PEK mode of parameter 0, it will be initialized.
When
Monitoring function is not indicated when setting the PEK.T of parameter 2 as "0".

SOURCE
12-24VDC/AC
100-240VAC

MT4N-AA-

<Option>
Relay output [MT4N--0]

AC
Ampere

Set range: 10-2/10-1/100/101


OFF/ HI/ LO / HL/ HL-G
OFF/ HI/ LO / HL/ HL-G
Within 1 to F.S. 10%(Variable depending
on set of input range and prescale.)

Parameter setting

Standard [STND]

Output
20mA

100mA
/50mA

250V/125V

500mA/200mA

5V/
2.5mV
50V/25V

50mA/4-20mA

SOURCE
12-24VDC/AC
100-240VAC

MT4N-AV-

AC
Volt

Set display index of frequency mode


Select output mode of OUT1
Select output mode of OUT2
Select hysteresis of OUT1
Select hysteresis of OUT2

H.PEK

Display rang [Fixed]

5mA
/2mA

50V/10V

DC
Ampere

Input
impedance

It sets current output for preset display value at the output


current 4-20mA DC.
It sets display value for 4mA in FS.L and 20mA in FS.H
and set range between FS-H and FS-L should be 10%
F.S.(When it set as under 10% F.S., it changed as over
10% F.S. automatically.)Preset display value is fixed to
output as 4mA at under FS-Land 20mA at over FS-H.

+0.6
-0

MT4N-DA-

250mV
/50mV

DC
Volt

Correct Low-limit gradient of display value

L.PEK

Measured input
and range

These are displayed in SCAL only and set


max/min. display value(-1999 to 9999).

INB.L Input bias low


INB.E Input bias exponent
OU1.T Out1 type
OU2.T Out2 type
HYS.1 Out1 hysteresis
HYS.2 Out2 hysteresis
Startup ompensation
STA.T time
PEK.T Peak time
DIS.T Display time
COLR Color

OU1.L

Specification of measured input and range


Type

Note
Available AC type only.
Selectable: STND/SCAL/FREQ
Display max. display value of STND
Available AC type only.

It is displayed in SCAL/FREQ only and set the position.


Set range: mV / V / mA / A / Hz / off
STND/SCALCorrection range: 0.100 to 5.000
FREQ Correction range: 0.100 to 9.999
Correction range: -99 to 99

FS-H
FS-L
ADRS
BPS
PRTY
STP
RSY.T
LOC
OU1.H

Approx. 64g

1: " " Mark indicated that equipment protected throughout by double insulation or reinforced insulation.
2: The indicator has no Hold function.
Environment resistance is rated at no freezing or condensation.

22.2+0.3
-0

Terminal connection

Unit weight

Function
Selectable RMS/AVG in AC type
Input range selection
Display type selection
Standard scale range
Frequency display
Scale range
Set max. value of display range
Set min. value of display range
Set decimal point position
Set display unit
Correct High-limit gradient of display value

EVIN

PA0
(Parameter 0)

Display
Input type
Input range
Display
Standard
Frequency
Scale
High scale
Low scale
Dot
Display unit lamp

INB.H Input bias high

PA2
(Parameter 2)

[PA 0: H.PEK/ L.PEKmode, PA 2: PEK.Tmode ]

DIN W48H24

Front panel identification

Approval

Monitoring peak display value function

MT

Insulation type2

Display value

Includes(Outer hold function)


Min. 20(at 500VDC megger)
1000VAC for 1 minute
2000VAC for 1 minute
Dielectric strength
(Between external terminal and case)
(Between external terminal and case)
Noise strength
2kV the square wave noise(pulse width: 1) by the noise simulator
Mechanical 0.75mm amplitude at frequency of 10 to 55Hz in each of X, Y, Z direction for 2 hours
Vibration
Malfunction 0.5mm amplitude at frequency of 10 to 55Hz in each of X, Y, Z direction for 10 minutes
Mechanical 100m/s2(Approx. 10G) in X, Y, Z directions for 3 times
Shock
Malfunction 300m/s2(Approx. 30G) in X, Y, Z directions for 3 times
Ambient
-10 to 50, Storage: -20 to 60
Environ- temperature
ment
Storage
35 to 85%RH, Storage: 35 to 85%RH
humidity

Display value

MT 4

Measurement range: 0.100 to 9999Hz(Differ according to decimal point position)

IN-T
IN-R
DISP
STND
FREQ
SCAL
H-SC
L-SC
DOT
D-UNT

STP
MODE

RSM.T
MODE

LOC
MODE

Set address of RS485 communication output


within 01 to 99.
Set baud rate of RS485 communication
output within 9600480024001200.
Set parity bit.
Set range: None / Even / Odd
Set stop bit.
Set range: 1/2
Set response waiting time.
Set range: 5 to 99
Enable Lock function with 4 mode,
OFF LOC1 LOC2 LOC3 OFF .

OFF Disable Lock function LOC2 Lock parameter 1,2


LOC1 Lock parameter 1 LOC3 Lock parameter 0,1,2

Caution for using


1. Allowable installation environment
If shall be used indoor Altitude Max. 2,000m Pollution Degree 3 Installation Category II
2. Use compression terminal(M3, Max. 6.0mm) to connect AC power.
Max.
3. Separate from high-tension line, power line to avoid inductive noise.
0.6mm
4. Install power switch or circuit breaker to on / off the power at once.
5. The switch or circuit breaker should be installed nearby users for safety.
6. Avoid to use the unit nearby machinery with high frequency noise, such as high frequency welder / sewing machine and high capacity SCR
controller etc.
Using line filter
Using varistor
7. "HHHH" or "LLLL" is displayed, off the power and check
lines.
Install it closely from DPM
8. Noise inflowing from power line can cause serious problem
HI
110/
for D.P.M.(Digital Panel Meter) of AC power. It is hard
D.P.M. INPUT
110/
to install protection circuit in the small unit even there is
D.P.M. 220VAC
LOW
220VAC
condenser to avoid noise between lines at primary of power
transformer. Use noise absorber circuit such as line filter,
Earth ground
varistor at external lines when abnormal voltage is occurred
by power relay, magnet switch, high frequency equipment are Using double shield wire
operated in same lines.
HI
+
9. Input line : Use shield wire when measured input line is
Vin
D.P.M.
extended or in a place with noise and open the non-used
LOW
terminals.
It may cause malfunction if above instructions are not followed.

Using Single shield wire


HI
D.P.M.
LOW

+
Vin
-

Main products

Proximity sensors
Counters
Area sensors
Timers
Photoelectric sensors
Display units
Fiber optic sensors
Panel meters
Door/Door side sensors
Pressure sensors
Rotary encoders
Power controllers
Sensor controllers
Graphic/Logic panels
Tachometer/Pulse(Rate) meters
Temperature controllers
Temperature/Humidity transducers
Switching power supplies
Stepping motors/drivers/motion controllers
Field network devices
Laser marking system(CO, Nd:YAG)
Laser welding/soldering system

http://www.autonics.com

Satisfiable Partner For Factory Automation


HEAD QUARTERS :
41-5. Yongdang-dong, Yangsan-si, Gyeongnam, 626-847,
Korea
OVERSEAS SALES :
Bldg. 402 3rd FL., Bucheon Techno Park, 193, Yakdae-dong,
Wonmi-gu, Bucheon-si, Gyeonggi-do, 420-734, Korea
TEL : 82-32-610-2730 / FAX : 82-32-329-0728
E-mail : [email protected]

The proposal of a product improvement


and development :[email protected]

EP-KE-77-0012G

You might also like