HYDROSTATIC LEVEL TRANSMITTER
General
Expert 3400 Hydrostatic level transmitter with ceramic capacitive
measuring system and built-in breather tube for direct air pressure
compensation, designed for use in waterworks, sewage treatment plants
and industrial applications.
The transmitter is accurate and has long term stability. It is very robust
and designed for applications in rough environments with aggressive
chemicals and fluids.
A major design feature is the programming facility using a standard
USB interface and a laptop computer, the servicing transmitter can be
programmed on-the-fly to the required measuring range.
The design without sharp edges prevents particles, textiles and paper
from sticking to the housing or the diaphragm.
Features
Ceramic diaphragm, PPS body,
2-wire 4 - 20 mA loop powered, 10-30VDC
Surge resistance
Accuracy better than 0.1 % and stability 0.1 % per year
Field re-programmable for custom ranges and zero offset calibration
Steel reinforced PUR cable with high tensile strength (1000kg)
Measuring ranges from 0 - 0,3 m to 0 - 300 m
Applications
Waste water
Process water
Chemical tanks with a range of acids, bases and ferric chloride.
Approved for mounting in explosion hazardous areas.
DATASHEET
EN 2.77 EXPERT 3400 DATASHEET 1412
2.77
Expert 3400
Expert 3400 Level Transmitter
Specifications
Expert 3400 Level Transmitter
Nominal measuring range
0-0,3 m
Measuring principle
Ceramic capacitive, relative pressure
0-1 m.
Min. programmable range
0-0,3 m
0-0,3 m
0-1 m.
0-3 m
0-3 m
0-10 m
0-30 m
0-90 m
Max. programmable range
0-1 m
0-1 m
0-3 m
0-10 m
0-10 m
0-30 m
0-100 m
0-300 m
Max. overpressure
4 bar
4 bar
6 bar
10 bar
10 bar
18 bar
40 bar
40 bar
Temperature range
-10 C to 70 C
Temp. deviation, zero point
Better than 0.02 % / C
Temperature deviation, FS
Better than 0.01 % / C
Linearity / Stability
Better than 0.2 % FS / 0.2 % FS
Measurement accuracy
Better than 0.1 % FS @ +10 to 30 C
Better than 0.2 % FS @ full temperature range
Long time stability
Better than 0.1 % FS per year
Materials
Housing: PPS (Ryton)
Diaphragm: Ceramic 99.9% AL2O3
Cell Packing: Viton
Supply voltage
10 - 30 V DC (12 - 30 V DC for cable lengths above 110m)
Output signal
2-wire 4 - 20 mA (passive transmitter)
Cable
2 0,5 mm2 (pressure) + 5 x 0,15 mm2 (data), shielded, PUR insulation, 1000kg pull strength
Cable length
12 m
Enclosure
IP68, withstands a static pressure equal to max. overpressure.
Safety Approvals
UL Class I, II and II, Division 1 Hazardous (Classified) Locations, UL 913
12 m
0-3 m
12 m
0-5 m
0-10 m
0-30 m
0-100 m
0-300 m
Ceramic cap., Sealed gate
12 m
12 m
35 m
110 m
320 m
ATEX: II 2G EEx ia IIC T4...T6
Test Approvals
CE directive: 89/336/EEC, 2004/108/EC, 1999/EC
EN 61326-1, 61326-2-3,
EN 55011, 61000-4-2, 61000-4-3, 61000-4-4, 61000-4-5,61000-4-6, 61000-4-8
EN 61000-6-4 2007-02-19, EN 61000-6-2 2005-09-08,
Can be delivered with other cable lengths on request.
Electrical Connections
Designations
1
Red
Positive (+) wire
Green
CS (Chip Select)
Brown
Negative (-) wire
Brown
GND
Black
Shield (NOT signal ground!)
Yellow
Data In
Tube
Air pressure compensation tube and Moisture filter
White
Clk
Hidden
Programming wires
Grey
Data Out
36 mm
135 mm
236 mm
32 mm
1 ISO
135 mm
Dimensions
Programming (5)
60 mm
Dia. 50 mm
Dia. 50 mm
MJK Expert 3400 Level Transmitter
Cable bracket for MJK Expert 3400 Level
Transmitter
(1 pcs. is delivered with each transmitter)
Expert 3400 Level Transmitter
Isolator /
Zener barriere
When using model Expert level transmitters
1400/3400 in explosion hazardous environments
Category 2 (Zone 1 and 2), an isolator or a zener
barrier must be used.
Isolator
The isolator is an integrated unit including active
components with low power consumption. A
power supply provides power to the pressure
transmitter through a galvanic separator such
that neither current nor voltage exceeds the
specified values. A ground connection is not
required. Item No. 202993
Hazardous area
Zener Barrier
The Zener barrier is a passive unit requirering a
safe grounding connection based on the ATEX
regultions.
The Zener barrier contains of a number of Zener
diodes for voltage protection. A safety fuse will
blow if the current exceeds the limits. A Zener
barrier requires a safe ground connection.
Item No. 202993
Accessories
202922 Transmitter connection box IP 67
enclosure for a safe connection of 4-20 mA
cables
200126 Field indicator 531,display insert,
4-20mA loop powered, IP65 enclosure.
For instant local readout.
200117 Rainroof/Sun Shield for 531, weather
protection for display and connection box
840100 For support of the Expert 1400 og 3400
series, use with 203990 PC Interface For Win XP-7
Use with 840100-2.03 USB driver for Win 7.
840100-2.03 USB driver for Expert Link (Win XPWin 7) from download center
Safe area
POWER SUPPLY
Wire 1 Red
Wire 2 Brown
Zener barriere
or EX isolator
+
-
Shield
Wire 3
Black
Control Drawing XXX XXX
Apparatus
Expert 3400 Level Transmitter
203951
Expert 3400, range 0-0,3 m
203952
Expert 3400, range 0-1 m
203953
Expert 3400, range 0-3 m
203954
Expert 3400, range 0-5 m
203955
Expert 3400, range 0-10 m
203956
Expert 3400, range 0-30 m
203957
Expert 3400, range 0-100 (SG) m
203958
Expert 3400, range 0-300 (SG) m
203961
Expert 3400, range 0-0,3 m with 1 ISO thread top
203962
Expert 3400, range 0-1 m with 1 ISO thread top
203963
Expert 3400, range 0-3 m with 1 ISO thread top
Order Numbers Accessories
200117
Sun shield / rain roof for connection box / field mounted temperature transmitter
200126
Display insert type 531 for connection box / field mounted temperature transmitter, 4-20mA loop powered
202920
Cable length, not standard
202922
Connectionbox for Expert level transmitter cable
202924
1 Thread top for Expert 7060/3400
202925
Expert 3400, with special measuring range
203990
USB interface for Expert 1400/3400 level transmitters with calibration software
202993
Loop supplied insulator for DIN rail mounting.
202995
Zener barrier for hydrostatic transmitter 9002/13-280-110-00
560915
Cable mounting bracket for Expert 7060/3400
691004
Cable for Expert 7060/3400, heavy duty, high tensile PUR insulated cable
One of these is mandatory when using model 3400 in explosion hazardous environments.
ER
Tel +45 45 56 06 56
Fax +45 45 56 06 46
www.mjk.com
Connect, MConnect, Chatter, MagFlux, Oxix, pHix compact, Shuttle and SuSix
are registrered trademarks of MJK Automation A/S.
2014 Xylem, Inc. Datasheet Expert 3400 11 2014
9 0 0 1- 2 0
08
MJK Automation
Byageren 7
DK-2850 Nrum
Denmark
Order Numbers
Accessories
Expert 3400, Capasitive Ceramic, Relative Pressure
IS
Order Numbers
TIFIE