CM44P

The document provides information on the Liquiline CM44P universal four-wire multichannel controller for process photometers and Memosens sensors. It details the device's benefits including flexibility, process safety, fast commissioning, and minimum inventory. Application examples are given for measuring systems using photometers and Memosens sensors in various industries. The document also covers the device's function, system design, ordering, configuration, and extension/modification options.

Uploaded by

Sadot Gutierrez
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
324 views

CM44P

The document provides information on the Liquiline CM44P universal four-wire multichannel controller for process photometers and Memosens sensors. It details the device's benefits including flexibility, process safety, fast commissioning, and minimum inventory. Application examples are given for measuring systems using photometers and Memosens sensors in various industries. The document also covers the device's function, system design, ordering, configuration, and extension/modification options.

Uploaded by

Sadot Gutierrez
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 52

TI01262C/07/EN/01.

16
71304152

Products

Solutions

Services

Technical Information
Liquiline CM44P
Universal four-wire multichannel controller for process
photometers and Memosens sensors

Field device or cabinet device


Application

Food and beverages


Life science
Power stations
Chemical industry
Other industrial applications

Your benefits
Highly flexible:
Able to connect up to 2 process photometers
Mathematics functions calculate new measured values
Digital fieldbuses (HART, PROFIBUS, Modbus,
Ethernet/IP) and integrated web server
Choice of cleaning function, controller and alarm relay
Optional digital or analog inputs/outputs

Maximum process safety thanks to standardized operating


concept across all devices in the Liquiline, sampler and
analyzer platform
Fast commissioning thanks to:
Memosens: lab-calibrated sensors & hot plug-and-play
Preconfigured Liquiline transmitter
Easy extension and adaptation
Minimum inventory:
Cross-platform, modular concept (e.g. identical modules
irrespective of parameters)
Integration into Fieldcare and W@M facilitates effective
asset management

Liquiline CM44P

Function and system design


Measuring system with
photometer

An optical measuring system comprises:


Liquiline CM44P transmitter
Sensor (photometer), e.g. OUSAF11/12/21/22/44/46, OUSTF10 or OUSBT66
Cable set CUK80
Assembly to suit the sensor, e.g. OUA260
The following are optional:
Post retainer
Protective cover
Memosens sensors ( 3)

Example of a measuring system with a photometer sensor

pipe

Flow assembly OUA260

Transmitter CM44P

Sensor: light source (lamp)

Cable set CUK80

Cable set CUK80

Sensor: detector

You can combine your measuring point with a variety of Memosens sensors and suitable
assemblies ( 3). For more information, visit www.endress.com/cm44p

Endress+Hauser

Liquiline CM44P

Measuring system with


optional Memosens sensors

The overview shows examples of measuring systems. Other sensors and assemblies can be ordered
for conditions specific to your application (www.endress.com/products).
Measuring point

pH value or ORP

A measuring system comprises:


Liquiline transmitter
Optional display (for cabinet device)
Sensors with Memosens technology
Assemblies to suit the sensors used
Post or rail mounting (optional, for field device)
Weather protection cover (optional, for field device)

pH measurement in the
pharmaceutical industry
Retractable assembly Cleanfit
CPA871
Sensor Orbisint CPS11D
Measuring cable CYK10
ORP in drinking water
Immersion assembly Dipfit
CYA112
Sensor Orbisint CPS12D
Measuring cable CYK10
Conductivity
Inductive conductivity
measurement in the food industry
Sensor Indumax CLS54D
Sensor fixed cable
Conductive conductivity
measurement in power plant
cooling water
Sensor Condumax CLS15D
Measuring cable CYK10
Oxygen

2
1
2
3
4
5

Cabinet installation (without sensor and signal


cables)
Optional display (back)
Liquiline
External power unit (CM444R and CM448R only)
Power supply cable (to be provided by the
customer)
Display cable

Oxygen in aeration basins


Immersion assembly Dipfit
CYA112
Holder CYH112
Sensor
COS61D (optical) with fixed
cable ( Fig.)
COS51D (amperometric),
cable CYK10

Nitrate and SAC

Turbidity and interface

Nitrate in wastewater
Sensor CAS51D-**A2 with fixed
cable
Immersion assembly Dipfit
CYA112
Holder CYH112

Turbidity in industrial water


Sensor Turbimax CUS51D with
fixed cable ( Fig.)
Assembly Flowfit CUA250
Spray head CUR3 (optional)

SAC in the wastewater treatment


outlet
Sensor CAS51D-**2C2 with
fixed cable
Immersion assembly Dipfit
CYA112
Holder CYH112

Interface in the primary clarifier


Sensor Turbimax CUS71D
Assembly CYA112
Holder CYH112

Chlorine

Ion-selective electrodes

Chlorine (and pH) in drinking


water
Sensor CCS142D
Sensor CPS11D
Measuring cable CYK10
Flow assembly CCA250

Ammonium and nitrate


measurement in the aeration basin
Sensor CAS40D with fixed cable
Holder CYH112

If mounting outdoors, always use the weather protection cover (see "Accessories") to protect the
transmitter against weather conditions.

Endress+Hauser

Liquiline CM44P

Application example

Measuring point in chromatography monitoring


Transmitter CM44P-AADINP1M22A1FG15BAEA+PK (cabinet device) with:
1 photometer input, 2 Memosens inputs, PROFIBUS, 2 analog outputs and 2 digital inputs
Optional display
Photometer OUSAF44 (item 4)
Flow assembly OUA260-AA1C05B1A3A with 2 mm path length and POPL, Triclamp 1/2" , quartz
window, item 3 (www.endress.com/oua260)
Flow assembly CYA680 with 2x Pg 13.5 process connection for Memosens sensors, item 1
pH and temperature with CPS71D, item 2 (www.endress.com/cps71d)
Conductivity, conductive four-pin sensor CLS81D, item 5 (www.endress.com/cls81d)
5

Chromatography Monitoring
11:12:13 29.06.2013

CH 1:1 pH

6.99 pH

CH 1:1 Temperature

19.1 C

CH 2:1 Cond c 4-pol

13.41 mS/cm

CH 3:1 Photometer

2.07 AU

MENU

Field Data Manager

8.00
mS/cm

29.05.2016
10:02:20

29.06.2016
11:09:15

29.05.2016
10:02:20

29.06.2016
11:09:15

12.00
pH

6.00
pH

A0028707

Measuring point in chromatography monitoring

Data retention
Storage of all measured values, incl. values of external sources, in the non-volatile memory
(data logbook)
Data called up on site via user-defined measuring menu and load curve display of the data logbook
Transmission of data by ethernet, CDI interface or SD card and storage in a tamper-proof database
(Field Data Manager)
Data export to csv file (for Microsoft Excel)

Endress+Hauser

Liquiline CM44P

Device architecture
Slot and port assignment
Slots
4R

2AO

2AI

Ports

2
3
4

Sensor 1
Sensor 2

2
Sensor 1
Sensor 2

Sensor 2

Sensor 1

2DS

2DS

2DS

BASE-E

Slot and port assignment of the hardware modules


Inputs are assigned to measuring channels in the
ascending order of the slots and ports.
Adjacent example:
"CH1: 1:1 pH glass" means:
Channel 1 (CH1) is slot 1 (basic module) : Port 1 (input
1), pH glass sensor
Outputs and relays are named according to their
function, e.g. "current output", and are displayed with the
slot and port numbers in ascending order

Order of the modules

Slot and port assignment on the


display

Depending on the version ordered, the device is supplied with a number of electronic modules, which
are assigned in a specific sequence in ascending order to slots 0 to 7. If you do not have a particular
module, the next moves up automatically:
The basic module (which is always present) always occupies slots 0 and 1
Fieldbus module 485 or Ethernet module ETH (only one of the two modules can be used)
Photometer module PEM
Memosens input module 2DS (DS = digital sensor)
Extension module for digital inputs and outputs DIO (DIO = digital input and output)
Current input module 2AI (AI = analog input)
Current output module 4AO or 2AO (AO = analog output)
Relay modules AOR, 4R or 2R (AOR = analog output + relay, R = relay)
Modules with 4 ports are connected before modules of the same type with 2 ports.

Basic rule for hardware


upgrades

Endress+Hauser

Please note the following if upgrading the device:


The sum of all current inputs and outputs may not exceed 8!
A maximum of two "DIO" modules may be used.

Liquiline CM44P

Determining the hardware


delivery status

You must be aware of the type of modules and the number of them supplied with the device you
have ordered to determine the delivery status of your Liquiline.
Basic module
One basic module in all versions. Always occupies slots 0 and 1.
Fieldbus module
Optional, and only one fieldbus module is possible.
Input modules
Must be clearly assigned to the number of optional inputs ordered.
Examples:
2 current inputs = module 2AI
2 inputs for photometer sensors = module PEM
4 Memosens inputs = 2 inputs with basic module + module 2DS with 2 further inputs
Current outputs and relays
Various module combinations can exist.
The following table will help you determine which modules your device has, depending on the type
and number of outputs.
Relays
Current outputs

1 x 2R

1 x 4R

1 x 2AO

1 x AOR

1 x 2AO + 1 x 4R

1 x 4AO

1 x 4AO + 1 x 2R

1 x 4AO + 1 x 4R

1 x 4AO + 1 x 2AO

1 x 4AO + 1 x 2AO + 1 x 2R

1 x 4AO + 1x 2AO + 1 x 4R

Sum up the number of modules and sort them according to the specified sequence 5.
This will give you the slot assignment for your device.
The unique terminal name is derived from:

Terminal diagram

Slot no. : Port no. : Terminal


Example, NO contact of a relay
Device with 2 inputs for digital sensors, 4 current outputs and 4 relays
Base module BASE-E (contains 2 sensor inputs, 2 current outputs)
Module PEM (1 photometer sensor)
Module 2AO (2 current outputs)
Module 4R (4 relays)

43 42 41

Relay 3

43 42 41

Relay 4

43 42 41

Relay 1

Slot 4

Port 1 Pin 41

Relay 2

43 42 41

Slot

A0028206

Creating a terminal diagram using the example of the NO contact (terminal 41) of a relay

Endress+Hauser

Liquiline CM44P

Device configuration using


the example of a CM44P-**

Ordered basic device (example)

CM44P-**DINP1M22A1FA*(cabinet device)
Functionality:
1x photometer (module PEM)
2x Memosens (module BASE-E)
PROFIBUS communication (module 485)
2 current outputs without HART (on BASE-E module)
2 current inputs (module AI)
3 slots are still free in this example. More or fewer slots can be free in
other versions.

Extension options without additional


modules

None

Modification options without


additional modules

Communication type changed by entering activation code. This


disables the communication type used previously!
Modbus RS485 (71140889)
Modbus TCP (71140890)
EtherNet/IP (71219868)
Retrofit to HART by removing module 485 and entering activation
code for HART (71128428)

Extension options by using extension


modules in free slots 5-7

Only the following is possible for the example above:


Module 2R (71125375) or 4R (71125376): 2 or 4 relays
Module DIO (71135638): 2 digital inputs and 2 digital outputs
If extended to four Memosens channels:
Module 2DS (71135631): 2 Memosens inputs
Use of the 2 current outputs in the basic module by entering
activation code (71140891)
Additional inputs or outputs and relays if fieldbus module 485 is
removed:
Module 2AO (71135632): 2 current outputs
Module AOR (71111053): 2 current outputs, 2 relays
Module 2R (71125375) or 4R (71125376): 2 or 4 relays
you replace module 485 with ETH, you can operate up to 6
Ifcurrent
outputs in addition to the ETH module's ethernet or
Modbus function. Only two current outputs are possible with
485.

Endress+Hauser

Basic rule for extensions

The sum of all current inputs and outputs may not exceed 8!

Restrictions if using CUS71D sensors


for interface measurement

If CUS71D sensors are used, the maximum number of Memosens


inputs is limited to two. Any combination of CUS71D or other sensors
is possible.

Product Configurator

www.endress.com/cm44p

Liquiline CM44P

Block circuit diagram of


CM44P-**

31

HART
mA

L L+

optional

N L-

32

PE

31

mA

32
optional

12

12

P+
S+
SP-

A (1)

A (2)

C (1)

C (2)

SH (1)

SH (2)

P+
S+
SP-

A (1)

A (2)

C (1)

C (2)

SH (1)

SH (2)

optional

3
optional

87
88
97
98

85

87
88
97
98

85

87
88
97
98

85

87
88
97
98

85

86

86

86

86

41
42

43

optional
31

41

mA

42

32
31

10

43
41
mA

42

32

10

43

optional

optional
31

41

mA

42

32
31

10

43
41
mA

42

32

10

43

optional

optional
31

mA

mA

31

23

24

32
mA

mA

23

24

32

optional

optional

45

47

46

48

45

47

46

48

91

+
11

92
91
92

optional
A0029203

7
1
2
3
4
5
6

Block circuit diagram of CM444 P


Current output 1:1, + HART (both optional)
Max. 7 x current output (optional)
Memosens input (2 x standard + 2 x optional)
PROFIBUS DP/Modbus/Ethernet (optional)
2 x current input (optional)
Power supply

7
8
9
10
11
12

Service interface
Power supply, fixed cable sensors
Alarm relay
2 or 4 x relays (optional)
2 digital inputs and outputs (optional)
Photometer: 2x lamp voltage and detector

Endress+Hauser

Liquiline CM44P

Communication and data processing


Types of communication:
Fieldbuses
HART
PROFIBUS DP (Profile 3.02)
Modbus TCP or RS485
EtherNet/IP
Only one type of Fieldbus communication can ever be active. The last activation code entered
decides which bus is used.
The device drivers available make it possible to perform a basic setup and display measured
values and diagnostics information via the fieldbus. A full device configuration via the fieldbus
is not possible.
Extension module 485 and current outputs
For PROFIBUS DP, Modbus and Ethernet communication protocols:
Max. of 2 current outputs can be used in parallel.
Extension module ETH and current outputs
Communication via Ethernet or EtherNet/IP
Max. of 6 current outputs can be used in parallel.
Max. of 4 current outputs can be used in parallel.
Bus termination on the device
Via slide switch at bus module 485
Displayed via LED "T" on bus module 485

Dependability
Reliability

Memosens
Memosens makes your measuring point safer and more reliable:
Non-contact, digital signal transmission enables optimum galvanic isolation
No contact corrosion
Completely watertight
Sensor can be calibrated in a lab, thus increasing the availability of the measuring point in the
process
Intrinsically safe electronics mean operation in hazardous areas is not a problem.
Predictive maintenance thanks to recording of sensor data, e.g.:
Total hours of operation
Hours of operation with very high or very low measured values
Hours of operation at high temperatures
Number of steam sterilizations
Sensor condition

MEMO

SENS

Sensor Check System (SCS)


The Sensor Check System (SCS) monitors the high impedance of the pH glass. An alarm is issued if a
minimum impedance value is undershot or a maximum impedance is exceeded.
Glass breakage is the main reason for a drop in high impedance values
The reasons for increasing impedance values include:
Dry sensor
Worn pH glass membrane
Process Check System (PCS)
The process check system (PCS) checks the measuring signal for stagnation. An alarm is triggered if
the measuring signal does not change over a specific period (several measured values).
The main causes of stagnating measured values are:
Contaminated sensor, or sensor outside of medium
Sensor defective
Process error (e.g. through control system)

Endress+Hauser

Liquiline CM44P

Sensor Condition Check (SCC)


This function monitors the condition of the electrode and the degree of electrode aging. The status is
indicated by the messages "SCC bad" or "SCC sufficient". The condition of the electrode is updated after
every calibration.
Self-monitoring functions
Current inputs are deactivated in the event of overcurrent and reactivated once the overcurrent
stops. Board voltages are monitored and the board temperature is also measured.
USP and EP
The limit functions for pharmaceutical water in accordance with USP and EP specifications are
implemented in the software for conductivity measurements:
"Water for Injection" (WFI) as per USP <645> and EP
"Highly Purified Water" (HPW) as per EP
"Purified Water" (PW) as per EP
The uncompensated conductivity value and the temperature are measured for the USP/EP limit
functions. The measured values are compared against the tables defined in the standards. An alarm
is triggered if the limit value is exceeded. Furthermore, it is also possible to configure an early
warning alarm that signals undesired operating states before they occur.
ChemocleanPlus
Freely programmable sequence control
e.g. for automatic sensor cleaning in retractable assemblies for reliable measurement results in
processes with a high risk of contamination
Individual, time-based activation of 4 outputs e.g. relays
Starting, stopping or pausing of activities via digital input or fieldbus signals e.g. from limit
position switches
Maintainability

Modular design
The modular transmitter design means it can be easily adapted to suit your needs:
Retrofit extension modules for new or extended range of functions, e.g. current outputs, relays and
digital communication
Upgrade to maximum 2 photometers and 4 Memosens inputs
Optional: M12 sensor connector for connecting any kind of Memosens sensor
Optional: CDI connector for external access to the service interface (avoids having to unscrew the
housing cover)

A0016027

10

CM444: example

A0012440

Extension module

Endress+Hauser

Liquiline CM44P

Data logger function


Adjustable scan time: 1 to 3600 s (1 h)
Data logbooks:
Max. 8 data logbooks
150,000 entries per logbook
Graphic display (load curves) or numerical list
Calibration logbook: max. 75 entries
Hardware version logbook:
Hardware configuration and modifications
Max. 125 entries
Version logbook:
E.g. software updates
Max. 50 entries
Operation logbook: max. 250 entries
Diagnostics logbook: max. 250 entries

A0015032

10

Data logbook: Graphic display

Logbooks remain unchanged even after a software update.


SD card
The exchangeable storage medium enables:
Quick and easy software updates and upgrades
Data storage of internal device memory (e.g. logbooks)
Transfer of complete configurations to a device with an identical setup (backup function)
Transfer of configurations without the TAG and bus address to devices with an identical setup
(copy function)
Saving of screenshots for documentation purposes
Endress+Hauser offers industry-approved SD cards as accessories. These memory cards provide
maximum data security and integrity.
Other SD cards can also be used. However, Endress+Hauser does not accept any responsibility for the
data security of such cards.
External signals for device control and for activating external devices
Hardware options, e.g. module "DIO" with 2 digital inputs and 2 digital outputs or fieldbus module
"485" enable the following:
via a digital input signal
measuring range switching for conductivity (upgrade code required, see accessories)
switching between different calibration datasets in the case of optical sensors
an external hold
a cleaning interval to be triggered
switching on and off a PID controller, e.g. via the proximity switch of the CCA250
the use of the input as an "analog input" for pulse-frequency modulation (PFM)
via a digital output signal
the static transmission (similar to a relay) of diagnostic states, point level switch states etc.
the dynamic transmission (comparable to a non-wearing "analog output") of PFM signals, e.g. to
control dosing pumps.

Endress+Hauser

11

Liquiline CM44P

FieldCare and Field Data Manager


FieldCare
Configuration and asset management software based on FDT/DTM technology
Complete device configuration when connected via FXA291 and service interface
Access to a number of configuration parameters and identification, measuring and diagnostic data
when connected via HART modem
Logbooks can be downloaded in CSV format or binary format for "Field Data Manager" software
Field Data Manager
Visualization software and database for measuring, calibration and configuration data
SQL database which is protected against manipulation
Functions to import, save and print out logbooks
Load curves to display measured values

A0016009

11

Field Data Manager: Load curves

Virtual process values (mathematical functions)


In addition to "real" process values, which are provided by connected physical sensors or analog
inputs, mathematical functions can be used to calculate a maximum of 6 "virtual" process values.
The "virtual" process values can be:
Output via a current output or a fieldbus
Used as a controlled variable
Assigned as a measured variable to a limit switch
Used as a measured variable to trigger cleaning
Displayed in user-defined measuring menus
The following mathematical functions are possible:
pH calculation based on two conductivity values acc. to VGB 405 RL, e.g. in boiler feed water
Difference between two measured values from different sources, e.g. for membrane monitoring
Differential conductivity , e.g. monitoring the efficiency of ion exchangers
Degassed conductivity, e.g. for process controls in power plants
Redundancy for monitoring two or three redundantly measuring sensors
rH calculation based on the measured values of a pH and an ORP sensor
Calculation of the remaining capacity of a cation exchanger

12

Endress+Hauser

Liquiline CM44P

Concentration tables
When the device is delivered from the factory, tables are saved in the device that allow inductive
conductivity measurements to be converted to concentrations of certain substances. 4 user-defined
tables are also possible.
The following factory concentration tables are available:
NaOH
NaOH
HCl
HNO3
H2SO4
H2SO4
H3PO4
NaCl
Safety

0 ... 15 %
25 ... 50%
0 to 20 %
0 to 30 %
0 to 28 % and 35 to 85 %
93 to 100 %
0 ... 40 %
0 ... 26 %

0 to 100 C (32 to 212 F)


2 to 80 C (36 to 176 F)
0 to 65 C (32 to 149 F)
2 to 80 C (36 to 176 F)
0 to 100 C (32 to 212 F)
0 to 100 C (32 to 212 F)
2 to 80 C (36 to 176 F)
2 to 80 C (36 to 176 F)

Real-time clock
The device has a real-time clock, which is buffered by a button cell battery if the power supply fails.
This ensures that the device continues to keep the correct date and time when it is restarted and that
the time stamp for the logbooks is correct.
Data security
All settings, logbooks etc. are stored in a non-volatile memory to ensure that the data are retained
even in the event of a disruption to the power supply.
Measuring range switching for conductivity
Can be used in CIP processes e.g. for safe monitoring of phase separations
Switching between 4 complete parameter sets:
Conductivity operating mode
Concentration tables
Temperature compensation
Output signal range
Limit value switch
Via digital inputs or fieldbus
Measured value compensation for oxygen and conductivity
Pressure or temperature compensation
Input signals from external sensors via current input or fieldbus
Signals from connected temperature sensors
Password protection
Password-protected login
For remote operation via web server
For local operation
Process safety
Two independent PID controllers
One- or two-sided control
Limit switches
4 cleaning programs which can be programmed independently of each other
IT security
We only provide a warranty if the device is installed and used as described in the Operating
Instructions. The device is equipped with security mechanisms to protect it against any inadvertent
changes to the device settings.
IT security measures in line with operators' security standards and designed to provide additional
protection for the device and device data transfer must be implemented by the operators themselves.

Endress+Hauser

13

Liquiline CM44P

Input
Measured variables

Photometer
Absorption (UV, color, NIR, cell growth)
Turbidity
Memosens sensors
Documentation of the connected sensor

Measuring ranges

Photometer
OUSAF12, OUSAF21, OUSAF22, OUSAF44, OUSAF46
OUSAF11
0 to 3 AU
0 to 6 OD (depending on the optical path length)
OUSTF10
0 to 200 FTU
0 to 200 ppm DE
OUSBT66
0 to 4 AU
0 to 8 OD (depending on the optical path length)
Memosens sensors
Documentation of the connected sensor

Types of input

Input signal

Depending on version:
Max. 2 x analog photometers
Max. 4 x binary sensor signal
2 x 0/4 to 20 mA (optional), passive, potentially isolated from one another and from the sensor
inputs
0 to 30 V

Cable specification

Cable type
OUK80 cable set for photometer sensors
Memosens data cable CYK10 or sensor fixed cable, each with cable end sleeves or M12 round-pin
connector (optional, for field housing)

Digital sensor inputs for sensors with Memosens protocol


Analog current inputs (optional)
Digital inputs (optional)
Analog photometer inputs

Cable length
Max. 100 m (330 ft)

Digital inputs, passive


Electrical specification

drawing power (passive)


Galvanically isolated

Span

High:
Low:

Nominal input current

Max. 8 mA

PFM function

Minimum pulse width: 500 s (1 kHz)

Test voltage

500 V

Cable specification

Max. 2.5 mm2 (14 AWG)

14

11 to 30 V DC
0 to 5 V DC

Endress+Hauser

Liquiline CM44P

Current input, passive


Span

> 0 to 20 mA

Signal characteristic

Linear

Internal resistance

Non-linear

Test voltage

500 V

Output
Output signal

Depending on version:
2 x 0/4 to 20 mA, active, galvanically isolated from one another and from the sensor circuits
4 x 0/4 to 20 mA, active, galvanically isolated from one another and from the sensor circuits
6 x 0/4 to 20 mA, active, galvanically isolated from one another and from the sensor circuits
8 x 0/4 to 20 mA, active, galvanically isolated from one another and from the sensor circuits
Optional HART communication (only via current output 1:1)
HART
Signal encoding

FSK 0.5 mA above current signal

Data transmission rate

1200 baud

Galvanic isolation

Yes

Load (communication resistor)

250

PROFIBUS DP/RS485
Signal encoding

EIA/TIA-485, PROFIBUS DP-compliant acc. to IEC 61158

Data transmission rate

9.6 kBd, 19.2 kBd, 45.45kBd, 93.75 kBd, 187.5 kBd, 500 kBd,
1.5 MBd, 6 MBd, 12 MBd

Galvanic isolation

Yes

Connectors

Spring terminal (max. 1.5 mm), bridged internally (T-function),


optional M12

Bus termination

Internal slide switch with LED display

Modbus RS485

Endress+Hauser

Signal encoding

EIA/TIA-485

Data transmission rate

2,400, 4,800, 9,600, 19,200, 38,400, 57,600 and 115,200


baud

Galvanic isolation

Yes

Connectors

Spring terminal (max. 1.5 mm), bridged internally (T-function),


optional M12

Bus termination

Internal slide switch with LED display

15

Liquiline CM44P

Ethernet and Modbus TCP


Signal encoding

IEEE 802.3 (Ethernet)

Data transmission rate

10/100 MBd

Galvanic isolation

Yes

Connection

RJ45

IP address

DHCP (default) or configuration via menu

EtherNet/IP
Signal encoding

IEEE 802.3 (Ethernet)

Data transmission rate

10/100 MBd

Galvanic isolation

Yes

Connection

RJ45

IP address

DHCP (default) or configuration via menu

Signal on alarm

Adjustable, as per NAMUR Recommendation NE 43


In measuring range 0 to 20 mA (HART is not available with this measuring range):
Error current from 0 to 23 mA
In measuring range 4 to 20 mA:
Error current from 2.4 to 23 mA
Factory setting for error current for both measuring ranges:
21.5 mA

Load

Max. 500

Linearization/transmission
behavior

Linear

Digital outputs, passive


Electrical specification

Passive
Open collector, max. 30 V, 15 mA

PFM function

Minimum pulse width: 500 s (1 kHz)

Auxiliary voltage

Electrical specification
Galvanically isolated
Unregulated, 24 V DC 20%
Max. 50 mA (per DIO module)

Test voltage

500 V

Cable specification

Max. 2.5 mm2 (14 AWG)

Current outputs, active


Span

0 to 23 mA
2.4 to 23 mA for HART communication

Signal characteristic

16

Linear

Endress+Hauser

Liquiline CM44P

Electrical specification

Output voltage
Max. 24 V
Test voltage
500 V

Cable specification

Cable type
Recommended: shielded cable
Cable specification
Max. 2.5 mm2 (14 AWG)

Relay outputs
Electrical specification

Relay types
1 single-pin changeover contact (alarm relay)
2 or 4 single-pin changeover contacts (optional with extension modules)
Relay switching capacity
Base module (Alarm relay)
Switching voltage

Load (max.)

Switching cycles (min.)

230 V AC, cos = 0.8 to 1

0.1 A

700,000

0.5 A

450,000

0.1 A

1,000,000

0.5 A

650,000

0.1 A

500,000

0.5 A

350,000

Switching voltage

Load (max.)

Switching cycles (min.)

230 V AC, cos = 0.8 to 1

0.1 A

700,000

0.5 A

450,000

2A

120,000

0.1 A

1,000,000

0.5 A

650,000

2A

170,000

0.1 A

500,000

0.5 A

350,000

2A

150,000

115 V AC, cos = 0.8 to 1

24 V DC, L/R = 0 to 1 ms

Extension modules

115 V AC, cos = 0.8 to 1

24 V DC, L/R = 0 to 1 ms

Cable specification

Endress+Hauser

Max. 2.5 mm2 (14 AWG)

17

Liquiline CM44P

Protocol-specific data
HART

PROFIBUS DP

Modbus RS485

Modbus TCP

18

Manufacturer ID

11h

Device type

155Dh

Device revision

001h

HART version

7.2

Device description files (DD/DTM)

www.endress.com/hart
Device Integration Manager DIM

Device variables

16 user-definable and 16 predefined device variables, dynamic


variables PV, SV, TV, QV

Supported features

PDM DD, AMS DD, DTM, Field Xpert DD

Manufacturer ID

11h

Device type

155Dh

Profile version

3.02

GSD files

www.endress.com/profibus
Device Integration Manager DIM

Output values

16 AI blocks, 8 DI blocks

Input variables

4 AO blocks, 8 DO blocks

Supported features

1 MSCY0 connection (cyclical communication, master class 1


to slave)
1 MSAC1 connection (acyclical communication, master class 1
to slave)
2 MSAC2 connections (acyclical communication, master class
2 to slave)
Device lock: The device can be locked using the hardware or
software.
Addressing using DIL switches or software
GSD, PDM DD, DTM

Protocol

RTU/ASCII

Function codes

03, 04, 06, 08, 16, 23

Broadcast support for function codes

06, 16, 23

Output data

16 measured values (value, unit, status), 8 digital values (value,


status)

Input data

4 setpoints (value, unit, status), 8 digital values (value, status),


diagnostic information

Supported features

Address can be configured using switch or software

TCP port

502

TCP connections

Protocol

TCP

Function codes

03, 04, 06, 08, 16, 23

Broadcast support for function codes

06, 16, 23

Output data

16 measured values (value, unit, status), 8 digital values (value,


status)

Input data

4 setpoints (value, unit, status), 8 digital values (value, status),


diagnostic information

Supported features

Address can be configured using DHCP or software

Endress+Hauser

Liquiline CM44P

EtherNet/IP

Protocol

EtherNet/IP

ODVA certification

Yes

Device profile

Generic device (product type: 0x2B)

Manufacturer ID

0x049Eh

Device type ID

0x109Ch

Polarity

Auto-MIDI-X

Connections

CIP

12

I/O

Explicit message 6
Multicast

3 consumers

Minimum RPI

100 ms (default)

Maximum RPI

10000 ms

System integration

EtherNet/IP

EDS

Rockwell

Add-on-Profile Level 3, Faceplate for Factory


Talk SE

Input (T O)

Device status and diagnostic message with


highest priority

IO data

Measured values:
16 AI (analog input) + Status + Unit
8 DI (discrete input) + Status
Output (O T)

Web server

Actuating values:
4 A0 (analog output) + status + unit
8 DO (discrete output) + Status

The web server enables full access to the device configuration, measured values, diagnostic
messages, logbooks and service data via standard WiFi/WLAN/LAN/GSM or 3G routers with a userdefined IP address.
TCP port

80

Supported features

Remote-controlled device configuration(1 session)


Save/restore device configuration (via SD card)
Logbook export (file formats: CSV, FDM)
Access to web server via DTM or Internet Explorer
Login
Web server can be switched off

Power supply
Supply voltage

CM44P
Depending on the version,:
100 to 230 V AC 15 %, 50/60 Hz
24 V DC +20 / -15 %
NOTICE
The device does not have a power switch
The customer must provide a protected circuit breaker in the vicinity of the device.
The circuit breaker must be a switch or power switch, and you must label it as the circuit breaker
for the device.
At the supply point, the power supply must be isolated from dangerous live cables by double or
reinforced insulation in the case of devices with a 24 V supply voltage.

Fieldbus connection

Endress+Hauser

Supply voltage: not applicable

19

Liquiline CM44P

Power consumption

CM44P
Depending on supply voltage
100 to 230 V AC:
Max. 73 VA (field device)
Max. 150 VA (cabinet device)
24 V DC:
Max. 68 W (field device)
Max. 59 W (cabinet device)

Fuse

Fuse not exchangeable

Overvoltage protection

Integrated overvoltage/lightning protection as per EN 61326


Protection category 1 and 3

Cable entries (field device


only)

Identification of the cable entry on housing base

Suitable gland

B, C, H, I, 1-8

M16x1.5 mm/NPT3/8"/G3/8

A, D, F, G

M20x1.5 mm/NPT1/2"/G1/2

M12x1.5 mm
Recommended assignment

1-8
A
B
C
D,F,G
H
I
E

C
D

A
1

Sensors 1-8
Power supply
RS485 In or M12 DP/RS485
Can be used freely
Current outputs and inputs, relays
Can be used freely
RS485 Out or M12 Ethernet
Do not use

A0018025

Cable specification

Length of display cable provided (cabinet device only):


3 m (10 ft)
Maximum permitted length of a display cable (cabinet device only):
5 m (16.5 ft)

20

Endress+Hauser

Liquiline CM44P

Connecting supply voltage

PK
GY

41
42

Power
(internal)

32 31

43

32 31

86

SD

Display

GY

2 87
1
+
BN BN
88 WH WH 88
97
97
A
A
GN GN
98
YE YE 98 B
B
+

87

31
+
32

31
+
32

Sensor

98 97 88 87

43 42 41

85

PK

Service

2+

86

0/4 ... 20 mA
HART

85

Sensor supply

86 85 86 85

1+

Alarm

Electrical connection

1
2

L+ N- PE

Power

L/+ N/ PE

L/+ N/

A0015873

A0015872

12

Overall wiring diagram BASE-E and


extension power unit (B)

Internal power supply cable


Extension power unit

86

GY
PK
GY

Power
(internal)

2 87
1
+
BN BN
88 WH WH 88
97
97
A
A
GN GN
98
98
YE YE
B
B
+

87

31
+
32

31
+
32

32 31

43

32 31

+ RD
- BK
L

Alarm

98 97 88 87

43 42 41

41
42

SD

Display

85

PK

Sensor

2+

86

Service

86 85 86 85

85

0/4 ... 20 mA
HART

1+

Sensor supply

A
B

13

Power supply connection on the BASE-E


(field device)

1
2

Power
L/+ N/ PE
L/+ N/

*
A0015873

A0025365

14
*

Power supply connection with BASE-E (cabinet


device)

15

Overall wiring diagram BASE-E and


external power unit (B)

Assignment depending on power unit, make sure


to connect correctly

The two device versions may only be operated with the power unit supplied and the power unit
cable. Also pay attention to the information in the operating manual supplied for the power
unit.

Endress+Hauser

21

Liquiline CM44P

Connecting optional modules

With extension modules you can purchase additional functions for your device.
NOTICE
Unacceptable hardware combinations (due to conflicts in power supply)
Incorrect measurements or total failure of the measuring point as a result of heat build-up or
overloading
If you are planning to extend your controller, make sure the resulting hardware combination is
permitted (Configurator on www.endress.com/).
Remember that the sum of all current inputs and outputs may not exceed 8!
Make sure not to use more than two "DIO" modules. More "DIO" modules are not permitted.
Please contact your Endress+Hauser sales center should you have any questions.
Overview of all the modules available
Module name
4R

2DS

2 x 0/4 to 20mA
analog outputs
2 relays
Order No.
71111053

2 relays
Order No.
71125375

4 relays
Order No.
71125376

2 digital sensor
2 digital inputs
inputs
2 digital outputs
2 power supply
with auxiliary
systems for digital
voltage
sensors
Order No.
Order No.
71135638
71135631

43

42
43

22

43

42
43
41

Relay 4

Relay

41

2+

86
85
86

PK
GY
PK
GY

47
+
48

47
+
48

1
2

42

41

85

42
43

87

88
97
A
98
B
+

87

88
97
A
98
B

45
+

BN
WH
GN

Sensor 1

43

43
41

41
42

42

1+

BN
WH
GN
YE

46

91
+
92

45
+
46

YE

Sensor 2

Relay 1

43

Relay 3

Relay

41

42

Relay 2

Relay 1

31
+
32

Relay 2

42

41

41

31
+
32

DIO

Sensor supply

2R

0/4 ... 20 mA

AOR

91
+
92

Endress+Hauser

Liquiline CM44P

Module name
4AO

2AI

485

2 x 0/4 to 20mA
analog outputs
Order No.
71135632

4 x 0/4 to 20mA
analog outputs
Order No.
71135633

2 x 0/4 to 20mA
analog inputs
Order No.
71135639

ETH

Service

82

Ethernet

RJ45

DIP switch

on

PWR
BF

Ethernet
DIP switch

23
+
24

SF

1
2
4
8
16
32
64
128/SW

COM

31
+
32

23
+
24

on

31
+
32

RJ45

1
2
4
8
16
32
64
128/SW
Service

DGND

31
+
32

81

VP

B'

A'

C'

DP/RS485

Ethernet (web
Web server and
server or Modbus
Ethernet/IP or
TCP)
Modbus TCP
5V power supply Order No.
for PROFIBUS DP
71272410
termination
RS485 (PROFIBUS
DP or Modbus
RS485)
Order No.
71135634

T COM SF BF PWR

31
+
32

31
+
32

0/4 ... 20 mA

0/4 ... 20 mA

31
+
32

0/4 ... 20 mA

0/4 ... 20 mA

24 23

24 23

2AO

PROFIBUS DP (module 485)


Contacts A - A, B - B and C - C are bridged in the connector. This ensures that PROFIBUS
communication is not interrupted if the connector is disconnected.

Endress+Hauser

23

Liquiline CM44P

Protective ground connection

4
A0025171

16

Cable mounting rail and associated function (field device)

17

Mounting rail for functional ground connections (cabinet device)

A0025366

Sensor connection

Cable mounting rail

Additional threaded bolts for ground connections

Threaded bolt (protective ground connection,


central grounding point)

Cable clamps (fixing and grounding the sensor


cables)

Photometer sensors
Sensor types

Sensor cable

Sensors

Analog photometer sensors without additional internal


power supply

CUK80

Fixed cable

OUSAF11

OUSAF12
OUSAF21
OUSAF22
OUSAF44
OUSTF10
OUSBT66

Sensors with Memosens protocol


Sensor types

Sensor cable

Sensors

Digital sensors without


additional internal power
supply

With plug-in
connection and
inductive signal
transmission

Fixed cable

Conductivity sensors with inductive measurement of


conductivity

Fixed cable

Turbidity sensors
Sensors for interface measurement
Sensors for measuring the spectral absorption
coefficient (SAC)
Nitrate sensors
Optical oxygen sensors
Ion-sensitive sensors

Digital sensors with


additional internal power
supply

pH sensors
ORP sensors
Combined sensors
Oxygen sensors (amperometric and optical)
Conductivity sensors with conductive measurement of
conductivity
Chlorine sensors

The following rule applies if connecting CUS71D sensors:


The maximum number of Memosens inputs is limited to two.
Any combination of CUS71D or other sensors is possible.
Types of connection
Direct connection of sensor cable to terminal connector of sensor module PEM and of Memosens
module 2DS or of base moduleE ( 18 ff.)(Memosens sensors only)
Optional for Memosens sensors: Sensor cable plug connected to the M12 sensor socket on the
underside of the device (field device)
With this type of connection, the device is already wired at the factory ( 22).

24

Endress+Hauser

Liquiline CM44P

PK

86 85 86 85

GY

32 31

GN
YE

32 31

32 31

32 31

98 97 88 87

43 42 41

BN
WH
GN
YE

98 97 88 87

43 42 41

86 85 86 85

Sensor cable connected directly

Sensor

Sensor

A0023038

18

Memosenssensors without additional supply


voltage

A0023039

19

Memosenssensors with additional supply


voltage

PK
GY

86

85

86

85

2DS

Sensor

98 97 88 87

Sensor 1

GN
YE

98 97 88 87

Sensor 2

BN
WH
GN
YE

Sensor

A0016197

20

Memosenssensors with and without additional


supply voltage on 2DS sensor module

A0028599

21

PEM module

Connecting photometer sensors to PEM module


Sensor

Cable color

PEM terminal

Assignment

OUSAF11
OUSAF12

YE (thick)

P+

Lamp voltage +

YE (thin)

S+

Recording lamp voltage +

BK (thick)

P-

Lamp voltage -

BK (thin)

S-

Recording lamp voltage -

A (1)

Sensor +

C(1)

Sensor -

SH (1)

Shield

RD
1)

BK / WH
GY

Endress+Hauser

2)

25

Liquiline CM44P

Sensor

Cable color

PEM terminal

Assignment

OUSAF21
OUSAF22
OUSTF10
OUSAF44

YE (thick)

P+

Lamp voltage +

YE (thin)

S+

Recording lamp voltage +

BK (thick)

P-

Lamp voltage -

BK (thin)

S-

Recording lamp voltage -

RD

A (1)

Measuring detector sensor +

BK

C(1)

Measuring detector sensor -

GY

SH (1)

Measuring detector screening

WH

A (2)

Sensor reference +

GN

C(2)

Sensor reference -

GY

SH (2)

Reference screening

BN

P+

Lamp voltage +

BN

S+

Recording lamp voltage +

BK

P-

Lamp voltage -

BK

S-

Recording lamp voltage -

RD

A (1)

Sensor +

OG

C(1)

Sensor -

TP

SH (1)

Shield

OUSBT66

1)
2)

OUSAF12
OUSAF11

PK
GY

85

86

85

Memosens connection via M12 connection

NC

86

4
5

Sensor 1

98 97 88 87

BN
WH
GN
YE

1
7

NC
6

Device versions with a pre-installed


M12 socket are ready-wired upon
delivery. Install an M12 socket, which
is available as an accessory, in a
suitable cable gland opening in the
base of the housing, and connect the
cables to the Memosens terminals of
the sensor or base module as per the
wiring diagram.
Connecting the sensor
Plug the connector of the sensor
cable (item 1) directly into the M12
socket.

Please note the following for these


device versions:
The internal device wiring is always
A0018021
the same regardless of what kind of
23 M12 assignment Top:
sensor you connect to the M12
socket Bottom: connector
socket (plug&play).
(top view in each case)
The signal and power supply cables
are assigned in the sensor plug-in
1
1
PK (24 V)
head in such a way that the PK and
2
GY (Ground 24 V)
A0018019
GY power supply cables are either
3
BN (3 V)
22 M12 plug-in connection (e.g. 4
used (e.g. optical sensors) or not
WH (Ground 3 V)
at sensor module)
(e.g. pH or ORP sensors).
5
GN (Memosens)
6
YE (Memosens)
1
Sensor cable with M12
7, Not connected
connector
NC
5

26

Endress+Hauser

Liquiline CM44P

Performance characteristics
Response time

Current outputs
t90 = max. 500 ms for an increase from 0 to 20 mA
Current inputs
t90 = max. 330 ms for an increase from 0 to 20 mA
Digital inputs and outputs
t90 = max. 330 ms for an increase from low to high

Reference temperature

25 C (77 F)

Measured error for sensor


inputs

Photometer
0 to 2.5 AU / to 50 OD
0.3 % of measuring range at 25 C (77 F)
Max. 1 % of measuring range
0 to 200 FTU / 0 to 200 ppm DE
Max. 2 % of measuring range
Memosens sensors
Documentation of the connected sensor

Measured error for current


inputs and outputs

Typical measured errors:


< 20 A (with current values < 4 mA)
< 50 A (with current values 4 to 20 mA)
at 25 C (77 F) each
Additional measured error depending on the temperature:
< 1.5 A/K

Frequency tolerance of
digital inputs and outputs

1%

Resolution of current inputs


and outputs

< 5 A

Repeatability

Documentation of the connected sensor

Endress+Hauser

27

Liquiline CM44P

Mounting

190 (7.48)

Mounting plate (field device)

170 (6.69)

Installation conditions

80 (3.15)
125 (4.92)

3 (0.12)

4 x 6.5 (0.26)

A0012426

24

Mounting plate in mm (inch)

Weather protection cover (field device)


300 (11.8)
170 (6.69)
80 (3.15)

320 (12.6)

170 (6.69)

80 (3.15)

270 (10.6)

8 x 7 (0.28)
A0012428

25

28

Weather protection cover in mm (inch)

Endress+Hauser

Liquiline CM44P

Installation

Post mounting
You require the post mounting kit (optional) to mount the unit on a pipe, post or railing (square
or circular, span range 20 to 61 mm (0.79 to 2.40")).
1

5
A0012665

26

Post mounting

Weather protection cover (optional)

Spring washers, nuts (post mounting kit)

Post mounting plate (post mounting kit)

Pipe or railing (circular/square)

Spring washers, nuts (post mounting kit)

Mounting plate

Pipe clamps (post mounting kit)

Threaded rods (post mounting kit)

Rail mounting
1

A0012668

Endress+Hauser

27

Rail mounting

Weather protection cover (optional)

Pipe or railing (circular/square)

Post mounting plate (post mounting kit)

Mounting plate

Spring washers, nuts (post mounting kit)

Threaded rods (post mounting kit)

Pipe clamps (post mounting kit)

Screws (post mounting kit)

Spring washers, nuts (post mounting kit)

29

Liquiline CM44P

Wall mounting

29
1
2
3
4

Wall mounting

Wall
4 drill holes 1)
Mounting plate
Screws 6 mm (not part of scope of supply)

A0012686

28

Mounting distance in mm (inch)

1)The

size of the drill holes depends on the wall plugs used. The wall plugs and screws must be provided by the
customer.

Mounting on DIN rail as per


IEC 60715

30

NOTICE
Incorrect mounting location in the cabinet, spacing regulations not observed
Possible malfunctions as a result of heat development and interference from neighboring devices
Do not position the device directly above sources of heat. Make sure to observe the temperature
specification.
The components are designed for convection-based cooling. Avoid the buildup of heat and
ensure openings are not covered over, e.g. by cables.
Make sure to keep the specified distance to other devices.
Physically separate the device from frequency converters and high-voltage devices.
Recommended installation direction: horizontal. The specified ambient conditions, and
particularly the ambient temperatures, only apply for horizontal installation.
Vertical orientation is also possible. However, this requires additional fixing clips at the place of
installation to hold the device in position on the DIN rail.
Recommended installation of power unit: to the left of the device.

Endress+Hauser

Liquiline CM44P

min. 20 (0.79)
min. 20 (0.79)

min. 50 (1.97)

min. 50 (1.97)

The following minimum spacing specifications must be observed:


Distance at the side to other devices including power units and distance to control cabinet wall:
at least 20 mm (0.79 inch)
Distance above and below the device and depth distance (to control cabinet door or other devices
installed there):
at least 50 mm (1.97 inch)

min. 50 (1.97)

A0027858

30

Minimum spacing in mm (inch)

31

Drilling pattern for wall mounting in mm (inch)

Wall mounting

A0027859

Endress+Hauser

31

Liquiline CM44P

The mounting plate also serves as the drilling template. The marks on the side help you mark
the position of the drill holes.

Mounting the external


display

140 (5.51)
105 (4.13)
4 x min. 8 (0.31)

105 (4.13)

140 (5.51)

min. 15 (0.59)

b
A0025371

32
a
b

Mounting plate of external display in mm (inch)

Retaining tab
Production-related recesses, no function for the user

Environment
Ambient temperature

CM44P
20 to 50 C (0 to 120 F)
External display (optional)
20 to 60 C (0 to 140 F)

Storage temperature

Field device
-40 to +80 C (-40 to 175 F)
Cabinet device
-25 to +85 C (-10 to +185 F)

Humidity

Field device
10 to 95%, not condensing
Cabinet device
DIN rail device
5 to 85%, not condensing
External display (in installed state)
5 to 95%, not condensing

Degree of protection

Field device
IP 66/67, impermeability and corrosion resistance in accordance with NEMA TYPE 4X

32

Endress+Hauser

Liquiline CM44P

Cabinet device
DIN rail device
IP20 shock protection
External display
IP66 front-panel, when installed correctly and with appropriate protective enclosure
Climate class (cabinet device
only)

As per IEC 60654-1: B2

Vibration resistance

Environmental tests
Vibration test based on DIN EN 60068-2, October 2008
Vibration test based on DIN EN 60654-3, August 1998
Post or pipe mounting
Frequency range
Amplitude
Test duration
Wall mounting
Frequency range
Amplitude
Test duration
1)

10 to 500 Hz (sinusoidal)
10 to 57.5 Hz:
0.15 mm
57.5 to 500 Hz:
2 g 1)
10 frequency cycles/ spatial axis, in 3 spatial axes (1 oct./min)
10 to 150 Hz (sinusoidal)
10 to 12.9 Hz:
0.75 mm
12.9 to 150 Hz:
0.5 g 1)
10 frequency cycles/ spatial axis, in 3 spatial axes (1 oct./min)

g ... gravitational acceleration (1 g 9.81 m/s2)

Electromagnetic
compatibility

Interference emission and interference immunity as per EN 61326-1:2013, Class A for Industry

Electrical safety

Field device
IEC 61010-1, Class I equipment
Low voltage: overvoltage category II
Environment < 3000 m (< 9840 ft) above MSL
Cabinet device
IEC 61010-1, Class I equipment
Low voltage: overvoltage category II
Environment < 2000 m (< 6562 ft) above MSL

Degree of contamination

Field device
The product is suitable for pollution degree 4.
Cabinet device
The product is suitable for pollution degree 2.
Optional display (for cabinet device)
The product is suitable for pollution degree 4.

Pressure compensation to
environment (field device
only)

Endress+Hauser

Filter made of GORE-TEX used as pressure compensation element


Ensures pressure compensation to environment and guarantees IP protection.

33

Liquiline CM44P

Mechanical construction

A
1

14 (0.55)
162 (6.38)

Field device

4 (0.16)

Dimensions

9 (0.35)
199 (7.38)

128 (5.04)

194 (7.64)

76 (2.99)

95 (3.74)

237 (9.33)
A0012396

33

34

Dimensions of field housing in mm (inch)

Endress+Hauser

Liquiline CM44P

cabinet device

75 (2.95)

100 (3.94)

114 (4.49)

205.5 (8.09)

TS 35

7.5 (0.30)

91 (3.58)

89 (3.50)

97 (3.82)

TS 35

220 (8.66)
A0025345

34

Endress+Hauser

Dimensions in mm (inch)

35

Liquiline CM44P

39.5 (1.56)

54 (2.13)

Optional display (for cabinet device)

144 (5.67)

144 (5.67)

A0025346

35

Dimensions in mm (inch)

External power units (for cabinet device)

121 (4.76)

90 (3.54)

Depending on the version ordered, a power unit for connection to 230 V or 24 V is supplied. There
are two delivery variants for each version (cannot be selected). The factory-preferred variant is
shown on the left in each case.

32 (1.26)

40 (1.58)

125 (4.92)

99 (3.99)

A0025738

36

Power unit 230 V

A0025739

37

Power unit 230 V (optional)


32 (1.26)

45 (1.77)

102 (4.02)

100 (3.94)

75 (2.95)

124 (4.88)

out

in

A0025784

38

36

Power unit 24 V

A0025786

39

Power unit 24 V (optional)

Endress+Hauser

Liquiline CM44P

Service display (accessories)


The service display comprises:
Portable display (same dimensions as under "Optional display")
Cover to protect the display and to hook it onto the (open) cabinet door
193 (7.60)
123 (4.84)

36 (1.42)
13.5 (0.53)

B-B

143 (5.63)

157 (6.18)

119 (4.69)

R1

40 (1.57)
B

47 (1.85)

144.5 (5.69)
A-A

A0025343

40

Weight

Dimensions of the service display cover in mm (inch)

Field device
Approx. 2.1 kg (4.63 lbs), depending on the version
Cabinet device
CM44P (fully configured)
Individual module
External display (excluding cables)
Service display cover
External power unit

Endress+Hauser

Approx. 0.95 kg (2.1 lbs)


Approx. 0.06 kg (0.13 lbs)
Approx. 0.56 kg (1.2 lbs)
0.46 kg (1 lbs)
0.27 to 0.42 kg (0.60 to 0.92 lbs), depending on the power
unit variant

37

Liquiline CM44P

Materials

Housing base and DIN rail housing

PC-FR

Display cover

PC-FR

Display foil and soft keys (field device)

PE

Housing seal
Display seal

EPDM

Soft keys (optional display)

EPDM

Module side panels

PC-FR

Module covers

PBT GF30 FR

Cable mounting rail (field device)


Terminal strip (cabinet device)
Clamps
Ground terminals

Stainless steel 1.4301 (AISI304)

Threaded fasteners

Stainless steel 1.4301 (AISI304)

Mounting plate (optional display)

Stainless steel 1.4301 (AISI304)

Securing screws (optional display)

Steel, galvanized

Cover for service display (accessories)

EPDM

Cable glands

Polyamide V0 as per UL94

Operability
display

Graphic display:
Resolution: 240 x 160 pixel
Back light with switch-off function
Red display background for alarms alerts users to errors
Transflective display technology for maximum contrast even in bright environments
User-definable measuring menus mean you can always keep track of the values that are important
for your application.

Operating concept

The simple and structured operating concept sets new standards:


Intuitive operation with the navigator and soft keys
Fast configuration of application-specific measurement options
Easy configuration and diagnosis thanks to plain-text display
All languages that can be ordered are available in every device

A0025228

41

38

42

Plain-text menu

Easy operation

Endress+Hauser

Liquiline CM44P

Local operation
1

2
A0011764

43
1
2
3

Remote operation

Overview of operation (using the example of the field device)

Display (with red display background in alarm condition)


Navigator (jog/shuttle and press/hold function)
Soft keys (function depends on menu)

Via HART (e.g. using HART modem and FieldCare)

A0028995

44
1
2
3
1)

Endress+Hauser

HART using modem

Device module Base L, H or E: current output 1 with HART


HART modem for connection to PC, e.g. Commubox FXA191 (RS232) or FXA195 1) (USB)
HART handheld terminal

Switch position "on" (substitutes the resistor)

39

Liquiline CM44P

Via PROFIBUS DP
Master class 1
(PLC, DCS)
Master class 2 (FieldCare, PDM)

PROFIBUS DP
(RS485)

FXA291

Ethernet

FieldCare /
Webbrowser

A0028991

45
T

PROFIBUS DP

Terminating resistor

Via Modbus RS485


Modbus Master / Gateway

Modbus RS485
(RTU, ASCII)

FXA291

Ethernet

FieldCare /
Webbrowser

A0028993

46
T

40

Modbus RS485

Terminating resistor

Endress+Hauser

Liquiline CM44P

Via Ethernet/Web server/Modbus TCP/EtherNet/IP

A0028994

47

Language packages

Modbus TCP and/or EtherNet/IP

The language selected in the product structure is the operating language preset at the factory. All
other languages can be selected using the menu.
English (US)
German
Chinese (Simplified, PR China)
Czech
Dutch
French
Italian
Japanese
Polish
Portuguese
Russian
Spanish
Swedish
Turkish
Hungarian
Croatian
Vietnamese
The availability of other languages can be checked via the product structure at www.endress.com/
cm44p.

Certificates and approvals


mark

The product meets the requirements of the harmonized European standards. As such, it complies
with the legal specifications of the EC directives. The manufacturer confirms successful testing of the
product by affixing to it the mark.

EAC

The product has been certified according to guidelines TP TC 004/2011 and TP TC 020/2011 which
apply in the European Economic Area (EEA). The EAC conformity mark is affixed to the product.

cCSAus

The product complies with "CLASS 2252 05 - Process Control Equipment" and "CLASS 2252 85 Process Control Equipment - Certified to US Standards" requirements.

Endress+Hauser

41

Liquiline CM44P

Ordering information
Product page

www.endress.com/cm44p

Product Configurator

On the product page there is a "Configuration" button to the right of the product image.
1.

Click this button.


The Configurator opens in a separate window.

2.

Select all the options to configure the device in line with your requirements.
In this way, you receive a valid and complete order code for the device.

3.

Export the order code as a PDF or Excel file. To do so, click the appropriate button on the right
above the selection window.
For many products you also have the option of downloading CAD or 2D drawings of the
selected product version. To do so, click the "CAD" tab and select the desired file type using dropdown lists.

Scope of delivery

The scope of delivery comprises:


1 multichannel controller in the version ordered
1 mounting plate
1 wiring label (attached at the factory to the inside of the display cover)
1 external display (if selected as an option) 1)
1 DIN rail power unit incl. Cable (cabinet device only)
1 printed copy of the Operating Instructions for DIN rail power unit (cabinet device only)
1 printed copy of the Brief Operating Instructions in the language ordered

Accessories
The following are the most important accessories available at the time this documentation was
issued. For accessories not listed here, please contact your service or sales office.
Measuring cable

CUK80 cable set


Pre-terminated and labeled cables for connecting analog photometer sensors
Product Configurator on the product page: www.endress.com/cuk80
CYK10 Memosens data cable
For digital sensors with Memosens technology
Product Configurator on the product page: www.endress.com/cyk10
Technical Information TI00118C
Memosens data cable CYK11
Extension cable for digital sensors with Memosens protocol
Product Configurator on the product page: www.endress.com/cyk11
Technical Information TI00118C

Sensors

Photometer sensors
OUSAF11
Optical sensor for VIS/NIR absorption
Stainless steel housing and sensor head made from dirt-repellent FEP
Product Configurator on the product page: www.endress.com/ousaf11
Technical Information TI00474C

1)

42

The external display can be selected as an option in the order structure or ordered as an accessory at a later stage.

Endress+Hauser

Liquiline CM44P

OUSAF12
Optical sensor for the measurement of absorbance
Variety of materials and process connections available
Product Configurator on the product page: www.endress.com/ousaf12
Technical Information TI00497C
OUSAF22
Optical sensor for measuring color concentrations
Variety of materials and process connections available
Product Configurator on the product page: www.endress.com/ousaf22
Technical Information TI00472C
OUSAF44
Optical sensor for measuring UV absorption
Variety of materials and process connections available
Hygienic design
Product Configurator on the product page: www.endress.com/ousaf44
Technical Information TI00416C
OUSTF10
Optical sensor for measuring turbidity and undissolved solids
Variety of materials and process connections available
Product Configurator on the product page: www.endress.com/oustf10
Technical Information TI00500C
OUSBT66
NIR absorption sensor for measuring cell growth and biomass
Sensor version suitable for pharmaceutical industry
Product Configurator on the product page: www.endress.com/ousbt66
Technical Information TI00469C
Glass electrodes
Orbisint CPS11D
pH electrode for process technology
Optional SIL version for connecting to SIL transmitter
With dirt-repellent PTFE diaphragm
Product Configurator on the product page: www.endress.com/cps11d
Technical Information TI00028C
Memosens CPS31D
pH electrode with gel-filled reference system with ceramic diaphragm
Product Configurator on the product page: www.endress.com/cps31d
Technical Information TI00030C
Ceraliquid CPS41D
pH electrode with ceramic junction and KCl liquid electrolyte
Product Configurator on the product page: www.endress.com/cps41d
Technical Information TI00079C
Ceragel CPS71D
pH electrode with double-chamber reference system and integrated bridge electrolyte
Product Configurator on the product page: www.endress.com/cps71d
Technical Information TI00245C
Memosens CPS171D
pH electrode for bio-fermenters with digital Memosens technology
Product Configurator on the product page: www.endress.com/cps171d
Technical Information TI01254C

Endress+Hauser

43

Liquiline CM44P

Orbipore CPS91D
pH electrode with open aperture for media with high dirt load
Product Configurator on the product page: www.endress.com/cps91d
Technical Information TI00375C
Orbipac CPF81D
Compact pH sensor for installation or immersion operation
In industrial water and wastewater
Product Configurator on the product page: www.endress.com/cpf81d
Technical Information TI00191C
Enamel pH electrodes
Ceramax CPS341D
pH electrode with pH-sensitive enamel
Meets highest demands of measuring accuracy, pressure, temperature, sterility and durability
Product Configurator on the product page: www.endress.com/cps341d
Technical Information TI00468C
ORP sensors
Orbisint CPS12D
ORP sensor for process technology
Product Configurator on the product page: www.endress.com/cps12d
Technical Information TI00367C
Ceraliquid CPS42D
ORP electrode with ceramic junction and KCl liquid electrolyte
Product Configurator on the product page: www.endress.com/cps42d
Technical Information TI00373C
Ceragel CPS72D
ORP electrode with double-chamber reference system and integrated bridge electrolyte
Product Configurator on the product page: www.endress.com/cps72d
Technical Information TI00374C
Orbipac CPF82D
Compact ORP sensor for installation or immersion operation in process water and wastewater
Product Configurator on the product page: www.endress.com/cpf82d
Technical Information TI00191C
Orbipore CPS92D
ORP electrode with open aperture for media with high dirt load
Product Configurator on the product page: www.endress.com/cps92d
Technical Information TI00435C
pH ISFET sensors
Tophit CPS441D
Sterilizable ISFET sensor for low-conductivity media
Liquid KCl electrolyte
Product Configurator on the product page: www.endress.com/cps441d
Technical Information TI00352C
Tophit CPS471D
Sterilizable and autoclavable ISFET sensor for food and pharmaceutics, process engineering
Water treatment and biotechnology
Product Configurator on the product page: www.endress.com/cps471d
Technical Information TI00283C

44

Endress+Hauser

Liquiline CM44P

Tophit CPS491D
ISFET sensor with open aperture for media with high dirt load
Product Configurator on the product page: www.endress.com/cps491d
Technical Information TI00377C
pH and ORP combined sensors
Memosens CPS16D
Combined pH/ORP sensor for process technology
With dirt-repellent PTFE diaphragm
With Memosens technology
Product Configurator on the product page: www.endress.com/cps16D
Technical Information TI00503C
Memosens CPS76D
Combined pH/ORP sensor for process technology
Hygienic and sterile applications
With Memosens technology
Product Configurator on the product page: www.endress.com/cps76d
Technical Information TI00506C
Memosens CPS96D
Combined pH/ORP sensor for chemical processes
With poison-resistant reference with ion trap
With Memosens technology
Product Configurator on the product page: www.endress.com/cps96d
Technical Information TI00507C
Conductivity sensors with inductive measurement of conductivity
Indumax CLS50D
High-durability inductive conductivity sensor
For standard and hazardous area applications
With Memosens technology
Product Configurator on the product page: www.endress.com/cls50d
Technical Information TI00182C
Indumax H CLS54D
Inductive conductivity sensor
With certified, hygienic design for foodstuffs, beverages, pharmaceuticals and biotechnology
Product Configurator on the product page: www.endress.com/cls54d
Technical Information TI00508C
Conductivity sensors with conductive measurement of conductivity
Condumax CLS15D
Conductive conductivity sensor
For pure water applications, ultrapure water applications and Ex applications
Product Configurator on the product page: www.endress.com/CLS15d
Technical Information TI00109C
Condumax CLS16D
Hygienic, conductive conductivity sensor
For pure water applications, ultrapure water applications and Ex applications
With EHEDG and 3A approval
Product Configurator on the product page: www.endress.com/CLS16d
Technical Information TI00227C

Endress+Hauser

45

Liquiline CM44P

Condumax CLS21D
Two-electrode sensor in plug-in head version version
Product Configurator on the product page: www.endress.com/CLS21d
Technical Information TI00085C
Memosens CLS82D
Four-electrode sensor
With Memosens technology
Product Configurator on the product page: www.endress.com/cls82d
Technical Information TI01188C
Oxygen sensors
Oxymax COS22D
Sterilizable sensor for dissolved oxygen
With Memosens technology or as an analog sensor
Product Configurator on the product page: www.endress.com/cos22d
Technical Information TI00446C
Oxymax COS51D
Amperometric sensor for dissolved oxygen
With Memosens technology
Product Configurator on the product page: www.endress.com/cos51d
Technical Information TI00413C
Oxymax COS61D
Optical oxygen sensor for drinking water and industrial water measurement
Measuring principle: quenching
With Memosens technology
Product Configurator on the product page: www.endress.com/cos61d
Technical Information TI00387C
Memosens COS81D
Sterilizable, optical sensor for dissolved oxygen
With Memosens technology
Product Configurator on the product page: www.endress.com/cos81d
Technical Information TI01201C
Chlorine sensors
CCS142D
Membrane-covered amperometric sensor for free chlorine
Measuring range 0.01 to 20 mg/l
With Memosens technology
Product Configurator on the product page: www.endress.com/ccs142d
Technical Information TI00419C
Ion-selective sensors
ISEmax CAS40D
Ion selective sensors
Product Configurator on the product page: www.endress.com/cas40d
Technical Information TI00491C

46

Endress+Hauser

Liquiline CM44P

Turbidity sensors
Turbimax CUS51D
For nephelometric measurements of turbidity and solids in wastewater
4-beam scattered light method
With Memosens technology
Product Configurator on the product page: www.endress.com/cus51d
Technical Information TI00461C
SAC and nitrate sensors
Viomax CAS51D
SAC and nitrate measurement in drinking water and wastewater
With Memosens technology
Product Configurator on the product page: www.endress.com/cas51d
Technical Information TI00459C
Interface measurement
Turbimax CUS71D
Immersion sensor for interface measurement
Ultrasonic interface sensor
Product Configurator on the product page: www.endress.com/cus71d
Technical Information TI00490C
Additional functionality

Hardware extension modules


Kit, extension module AOR
2 x relay, 2 x 0/4 to 20 mA analog output
Order No. 71111053
Kit, extension module 2R
2 x relay
Order No. 71125375
Kit, extension module 4R
4 x relay
Order No. 71125376
Kit, extension module 2AO
2 x 0/4 to 20 mA analog output
Order No. 71135632
Kit, extension module 4AO
4 x analog output 0/4 to 20 mA
Order No. 71135633
Kit, extension module 2DS
2 x digital sensor, Memosens
Order No. 71135631
Kit, extension module 2AI
2 x 0/4 to 20 mA analog input
Order No. 71135639
Kit, extension module DIO
2 x digital input
2 x digital output
Auxiliary voltage supply for digital output
Order No. 71135638
Kit, extension module 485
Ethernet configuration
Can be extended to PROFIBUS DP or Modbus RS485 or Modbus TCP or EtherNet/IP. This requires
an additional activation code which can be ordered separately.
Order No. 71135634

Endress+Hauser

47

Liquiline CM44P

Kit, extension module ETH


Ethernet configuration
Can be extended to Modbus TCP or EtherNet/IP. This requires an additional activation code which
can be ordered separately.
Order No. 71279810
Upgrade kit, extension module 485 with PROFIBUS DP
Extension module 485
PROFIBUS DP (+ Ethernet configuration)
Order No. 71140888
Upgrade kit, extension module 485 with Modbus RS485
Extension module 485
Modbus RS485 (+ Ethernet configuration)
Order No. 71140889
Upgrade kit, extension module 485 with Modbus TCP
Extension module 485
Modbus TCP (+ Ethernet configuration)
Order No. 71140890
Upgrade kit, extension module 485 with EtherNet/IP
Extension module 485
EtherNet/IP (+ Ethernet configuration)
Order No. 71219868
Upgrade kit, extension module ETH with Modbus TCP
Extension module ETH
Modbus TCP (+ Ethernet configuration)
Order No. 71279809
Upgrade kit, extension module ETH with EtherNet/IP
Extension module ETH
EtherNet/IP (+ Ethernet configuration)
Order No. 71279812
Firmware and activation codes
SD card with Liquiline firmware
Industrial Flash Drive, 1 GB
Order No. 71127100
You must quote the serial number of the device when ordering the activation code.
Activation code for digital HART communication
Order No. 71128428
Activation code for PROFIBUS DP
Order No. 71135635
Activation code for Modbus RS485
Order No. 71135636
Activation code for Modbus TCP for module 485
Order No. 71135637
Activation code for EtherNet/IP for module 485
Order No. 71219871
Activation code for Modbus TCP for module ETH
Order No. 71279813
Activation code for EtherNet/IP for module ETH
Order No. 71279830
Kit CM442: activation code for 2nd digital sensor input
Order No. 71114663
Kit CM444/CM448: upgrade code for 2 x 0/4 to 20 mA for BASE-E
Order No. 71140891
Activation code for feedforward control
Requires current input or fieldbus communication
Order No. 71211288

48

Endress+Hauser

Liquiline CM44P

Activation code for measuring range switch


Requires digital inputs or fieldbus communication
Order No. 71211289
Activation code for ChemocleanPlus
Requires relays or digital outputs or fieldbus communication and optional digital inputs
Order No. 71239104
Software

Memobase Plus CYZ71D


PC software to support laboratory calibration
Visualization and documentation of sensor management
Sensor calibrations saved in the database
Order as per product structure, www.endress.com/cyz71d
Technical Information TI00502C
Field Data Manager Software MS20
PC software for central data management
Visualization of series of measurements and logbook events
SQL database for secure data storage
Order No. 71129799

Other accessories

External display 2)
Graphic display
For installation in the control cabinet door or panel
Order No. 71185295
Service display
Portable, for commissioning
Order No. 71185296
SD card
Industrial Flash Drive, 1 GB
Order No. 71110815
M12 built-in socket and cable junction with Velcro strip
Kit CM42/CM442/CM444/CM448: external CDI socket
Socket with terminated connecting cables and counter nut
Order No. 51517507
Kit CM442/CM444/CM448/CSF48: M12 built-in socket for digital sensors
Pre-terminated
Order No. 71107456
Kit CM442/CM444/CM448/CSF48: M12 built-in socket for PROFIBUS DP/Modbus RS485
B-coded, pre-terminated
Order No. 71140892
Kit CM442/CM444/CM448/CSF48: M12 built-in socket for Ethernet
D-coded, pre-terminated
Order No. 71140893
Kit: external CDI socket, complete
Retrofit kit for CDI interface, with terminated connecting cables
Order No. 51517507
Cable junction with Velcro strip
4 pieces, for sensor cable
Order No. 71092051

2)

The external display can be selected as an option in the product structure or ordered subsequently as an accessory.

Endress+Hauser

49

Liquiline CM44P

Communication-specific accessories
Commubox FXA191
Intrinsically safe HART communication with FieldCare via the RS232C interface
Transforms HART signals on RS 232C interface
Technical Information TI00237F
Commubox FXA195
Intrinsically safe HART communication with FieldCare via the USB port
Technical Information TI00404F
Commubox FXA291
Connects the CDI interface of measuring devices with the USB port of the computer or laptop
Technical Information TI00405C
Wireless HART adapter SWA70
Wireless device connection
Easily integrated, offers data protection and transmission safety, can be operated in parallel with
other wireless networks, minimum cabling complexity
Technical Information TI00061S
Fieldgate FXA320
Gateway for the remote interrogation of 4-20 mA measuring devices via a Web browser
Technical Information TI00025S
FieldXpert SFX100
Compact, flexible and robust industrial handheld terminal for remote configuration and for obtaining
measured values via the HART current output
Operating Instructions BA00060S
System components
RIA14, RIA16
Field display unit for integration into 4-20 mA circuits
RIA14 in flameproof metal enclosure
Technical Information TI00143R and TI00144R
RIA15
Process display unit, Digital display unit for integration into 4-20 mA circuits
Panel mounting
With optional HART communication
Technical Information TI01043K

50

Endress+Hauser