Manual Book DO-5510
Manual Book DO-5510
OXYGEN METER
Model : DO-5510
ATTENTION :
Fill the Probe's Electrolyte at first.
2. FEATURES
1
* Auto power off saves battery life.
* Powered by 006P DC 9V battery.
* RS 232 PC serial interface.
* Build in temperature , measurement.
* Use the durable, long-lasting components, including a
heavy duty & compact ABS-plastic housing case.
3. SPECIFICATIONS
3
4. FRONT PANEL DESCRIPTION
Fig. 1
Calibartion Consideration :
5
5-2 Dissolved Oxygen ( DO) measurement
1)After the meter be calibrated ( above procedure 5-1 ),
the meter is ready for DO measurement.
2)Slide the " O2/DO selector " ( 4-11, Fig. 1 ) to the " DO "
position.
3)"% Salt" compensation of the probe :
a. Push the " % Salt button " ( 4-10, Fig. 1 )
first, and the display will show. :
0 %
S
b. Push the " Factor Adj. button " ( 4-8, Fig. 1 ) once
will add " 1 % " to the original salt % value until the
adjusting reading reach the desired salt values.
Then push the " % Salt button " ( 4-10, Fig. 1 )
again to finish the adjusting procedures. At this
moment the new % salt values will be executed.
c. If the measured liquid is the pure water or the
factor of salt % can be neglected, then adjust %
salt compensation value to 0%.
4)"Height" compensation of the probe :
Bear in mind that the DO measurement is considered
to be taken at sea level. However if the measuring
environment is not at sea level ( 0 meter ), then
should adjust the " Height " values for the probe
compensation when make the DO measurement.
6
a. Push the " MT ( Height) button " ( 5-13, Fig. 1 ) first,
then the display will show
0
H
b. Push the " Factor Adj. button " ( 4-8, Fig. 1 ) once
will add " 100 meters " to the original height values
until the adjusting reading reach the desired height
values ( meters ). Then push the " MT ( Height)
button " (4-13, Fig. 1) again to finish the adjusting
procedures. At this moment the new height values
(display unit is the " meters ") will be executed.
5)a. Immersed the probe to a depth at least 10 cm of
the measured liquid in order for the probe to be
influenced by the temperature & automatic
temperature compensation to take place.
b. As for the thermal equilibrium to occur between the
probe & the measurement sample must be allowed
to pass, which usually amounts to a few minutes if
the Temp. difference between the two is only
several Celsius degrees.
6)a. In order to measure the dissolved oxygen content in
any given liquid, it is sufficient to immerse the tip of
the probe in the solution, making sure that velocity
of the liquid coming into contact with the probe is at
least 0.2 - 0.3 m/s or to shake the probe.
b. During laboratory measurements, the use of a
magnetic agitator to ensure a certain velocity in the
fluid is recommended. In this way, errors due to the
diffusion of the oxygen present in the air in the
solution are reduced to a minimum.
7
7)Rinsed the probe accurately with normal tap water after
each series of measurement.
8
* With the " REC " symbol on the display :
(a) Push the " Memory CALL button " ( 4-7, Fig. 1 )
once, the " Max " symbol along with the maximum
value will appear on the display.
(b) Push the " Memory CALL button " again, the " Min "
symbol along with the minimum value will appear
on the display.
(c) Push the " Memory CALL button " again, the
" AVG " symbol along with the average value will
appear on the display.
(d) To exit the memory record function, push the
" RECORD " button once again. The display will
revert back to the current reading.
Calibration procedures
Power on, slide the " DO/O2 selector " to " O2 " position.
9
Dissolved Oxygen ( DO ) measurement
procedures
Determine the " % Salt " & " Height " values.
* For the general application, it is not necessary to make
the adjustment for the " % salt " & " Height " values.
Other function :
Power management
10
6. AUTO POWER OFF DISABLE
and LCD CONTRAST ADJUST
1)The instrument has built-in "Auto Power off "
in order to prolong battery life. The meter will switch
off automatically if none of the buttons are pressed
within 10 min.
To de-activate this feature, select the memory record
function during measurement by pressing the
" RECORD " button ( 4-6, Fig. 1 ).
2)The instrument also features the ability to adjust the
contrast of the display.
This is achieved by controlling the " LCD Contrast
Adjust " pot ( 4-5, Fig. 1 ).
11
7. RS232 PC SERIAL INTERFACE
The instrument features an RS232 output via 3.5 mm
Terminal ( 4-16, Fig. 1). The connector output is a 16
digit data stream which can be utilized to the user's
specific application.
Meter PC
(3.5 mm jack plug) (9W 'D" Connector)
12
Each digit indicate the following status :
D15 Start Word =02
D14 Reading Polarity for the Display.
0 = Both upper & lower display value are "+".
1 = Upper "-", Lower "+".
2 = Upper "+", Lower "-".
3 = Both upper & lower display value are "-".
D13 Anunuciator for Lower Display
0 =No Symbol 1 = 2 =
D12 & D11 Anunuciator for Upper Display
01 = 07 = mg/L
02 = 06 = % O2
D10 Decimal Point (DP) for lower display
0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP
D9 Decimal Point ( DP) for Upper display.
0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP
D8 to D5 Lower Display reading, D5=LSD, D8=MSD
D4 to D1 Upper Display reading, D1=LSD, D4=MSD
D0 End Word = 0D
RS232 setting
Baud rate 9600
Parity No parity
Data bit no. 8 Data bits
Stop bit 1 Stop bit
13
8. PROBE MAINTENANCE
User first time to use the meter :
Fig. 2
8-1 Probe handle
8-2 Temp. sensing metal
8-3 Probe head
15
9. BATTERY REPLACEMENT
1)When the left corner of LCD display show " LBT ", it is
necessary to replace the battery. However, in-spec
measurement may still be made for several hours
after low battery indicator appears.
2)Slide the " Battery Cover " ( 4-14, Fig. 1 ) away from
the instrument and remove the battery.
3)Replace with 9V battery (heavy duty type) and
reinstate the cover.
4)Make sure the battery cover is secured after changing
the battery.
16
0603-DO-5510